GE Healthcare Signa HDx & HDi 1.

5T System Installation
REV 7 DIRECTION 5166001

1–5 BASIC SYSTEM (Continued)

MAGNET MONITOR BLOWER BOX


(MSM) (MG6)

SPT PHANTOM SET


STORAGE CART
PENETRATION
PANEL
(PP1)

MAGNET ENCLOSURE
(MG2)

LIGHTWEIGHT
WATER CHILLER FOR PATIENT TRANSPORT
SHIELD COOLER (PT1)
GRADIENT COIL COOLING COMPRESSOR CABINET
(WC1) (MS5)

MNS Option

Î
RFS HFD  1.5T MNS AMP
CABINET CABINET  CABINET
(MR2) (MR3) (MR10)

SIGNA EXCITE 1.5T SYSTEM


ILLUSTRATION 1–1

1–5
1 FILTERS FOR EXAM Signa HDx & HDi 1.5T (14.x) SYSTEM INTERCONNECT
FACILITY POWER, ROOM 1
1 EMERGENCY POWER, MS4 LIGHTS – HiSpeed 8 Channels or
A12 AND AUXILIARY (MRU)
OW1 A6 2037 MNTR 1 USB 1055 A18 10 PORT 6 1056 (SERVICE LAPTOP) 5 WARNING DEVICE 1 606
– EchoSpeed 8 or 16 Channels
(HOST MONITOR) (SERVICE KEY) 1 OM1 1
FACILITY WIDE SCREEN
1065 OW1 A15 RS232
PORT 8 11 TO SITE (OXYGEN 1 MG3 (FIXED SITE)
(HOST COMPUTER) 12 FACILITY EXHAUST FAN (REAR PED)
EMERGENCY 1066 11 DB–9 OW1 A13 ETHERNET POWER MONITOR) 1
POWER OW1 A7 2038 USB
(MODEM) PORT 7 MS1
(KEYBRD) PCI (C) 1059 PORT 2 942 1254 RF SHIELD ROOM 040 (MAGNET) NOTES:
USB COMMON GROUND
1 SCSI 2 1067 OW1 A5 (DASM) T OW1 A17 1 1 Customer furnished
OW1 A8 5 5 OM3
(MOUSE) 3 1379 (LAN MS10 1
USB HUB) (REMOTE MRU) (REMOTE O2 SENSOR) Magnet and superconducting shim power supplies are not included with
2
1371 PCI (D) USB system, but may remain in place if purchased as an option.
OW1 A20 TO 5 457 J17 458 J3P
(SCIM) 1064 SCSI 1 LINE OUT 603 J3R MS9 3
PDU MS7 1266 J16 1265 (REMOTE MRU BOX) Bar Code Reader Option connection.
Power LINE IN
OW1 A2 (S/C SHIM P.S.)
(GOC 604 4 Typical connection for normally closed emergency off switch
OW1 A16
COMPUTER 603 MS1 A3 A1 J3
T (SCSI 1071 J1 J10 (TERMINAL BOX) P403
EXPANSION CABINET) MS6 RED (2) 6 604 5 Optional: May not be present on all systems.
DEVICE) J2 J7 (BROWN) (MAGNET P.S.) 601 BLK (2)
2 602
1 1070 J11 J12 601 MS1 6 Magnet ramping cables pass through ramp port on PP1 plate.
OW1 A1 1 (MAGNET SERV TURRET)
(OW TABLE) FAN J3 J6 F1 602 7 RUNS (600 SERIES) Furnished with GE Magnet.
(GREEN) OW1 A11 1075 J4 J5 623 F2 624
(POWER USB–DB9
MS5A1 F3
SLEEVE
8 Required for 16 Channel Systems only.
OW1 A10 STRIP) 1074 J2 (COLD F4 MS1 A2
1 (SPKR) ADAPTER HEAD
1 WATER (SHIELD COOLER 9 Body Hybrid To Body Coil Connections depend on magnet field polarity.
2053 J6 OW1 A21 J12
OW1 A19 MS5A5 CMPRSSR) 621 COLD HEAD)
(CABINET 1320 (WATER
J15 (GOCAA) J14 PDU) FLOW)
A6 JR 622 COLDHEAD 10 Customer supplied telephone line for Insite Modem.
IN HUB
J7 1060 827 A1 J1 FEED THRU 832 Ethernet cable is customer supplied. RUN 1066 and modem may not be
J17 J3
OUT C2
833 (J60 & J61) MS1 A5 A2 PREAMP 11
J18 J10 J16 FJ4 PP1 (PRESSURE J2 J1 required if connection to InSite is provided thru site ethermet connection.
J1
(PENETRATION PANEL) TRANSDUCER)
FJ3 12 Network must be available for Magnet Monitor 24 hours a day/7days a week.
TEMP/HUMID
(YELLOW) MG3 TB1
1 1085 STEREO MUSIC 5 SENSOR
1321 1322 826 J9 MSM1 J8 825 J48 829 830 (REED SWITCH)
1081 (MAGNET
1 f RACK IN BORE
942 12 J5 MONITOR) J7 831 824 J10
FJ2
828
848 J5 J1 862 J1 MG2 A5
FACILITY 823 J10 FJ1 711 (RIGHT OPER.
POWER PP1 A17 A1 J91 711 (J11,12,13) J4 897 J3
MG2 A4 CONTROL)
1045 712
3 f F/O REPEATER BD
712 (J9,10) J4 BULB
(OPERATOR
MG2 A5
DISPLAY)
J1 PP1 A17
MG2 A11
(PAC P. GATING J5 1261 J4 J2 863 J2 (LEFT OPER.
PATIENT (REPEATER PNL) 2 F/O I/F) CONTROL)
1196 (PAC 3) RESP
MG2 A33
ALERT (SRI–3)
CONTROL ALERT AIR LINE ECG J2 422 J4 MG3 A1 (MAGNET I/F #1)
BOX
716 J2 POWER
PPG
AIR HOSE J10 1260 MG2 A36 (BORE TEMP SENSOR)
1/2 1/2 1/2 FROM POWER
J3 (X) 762 762 PP1 A7 762 J1
BLOWER
J4 (Y) 763 763
3/4 (GRADIENT 3/4
763 3/4 BOX
J3 421 J1 MG2 A31
FILTER)
715 J20 (DOCK LIMIT SWITCH)
5/6 5/6 J2
J5 (Z) 764 764 764 5/6

WC1 COOLANT FEED THRU 536 MG2 A12 A1 387


COOLANT LINES 1215 MG2 A29 (DOCK ASM)
MR3 (WATER CHILLER)
(BLUE/RED) COOLANT FEED THRU 537 (GRADIENT COIL)
P1
(HFD 4 F/O FLEXIBLE
J2 1179 TEMP/HUMID SENSOR
CABINET) EXTENSION CABLES
J7 1180 970 MG3 A17
(ORN/YEL) (CABLE I/F)
J8 1181 RF AMP J3 1249 1238 J4 1328 J2
CRYO
TS1 J4 1259 235 J44 J3 1153 J1 (MNS XMT)
COOLER J2 (MNS RCVE)
COMPRESSOR 18,19,GRN/YEL,N 1081 J202 1187 J202 1202 J4 1206 MG2A20 MG2 A16
MR2
GRN/YEL 1254 (RFS (MC (LOW PROFILE
MAIN 10,11,12, CABINET) 257 1331 1370 J18 I/F CARRIAGE
DISCONNECT
1204 J36 J35 J20 BOX) ASSEMBLY)
1 GRN/YEL, N 1234 (ORANGE) J23
J9
1 A 1179 J17 J24 823 J1 MG3 A36 J21 1332 J20
MG3 A8
THRU
J19 A (MULTICOIL)
f IN 1180
B J56 (16 CHANNEL J16 (QUICK
3 A11 J13 1045 5 F/O (UTNS SWITCH 1) J1–8 CONN DISCONNECT)
C 1255 J26 ANTENNA) J18 1333 J21 J17 A1
(HEAD COIL)
GRN/YEL 1371 J63 J15 1214 J20
(1–8) J200 1185 J200 1200 J1
J22 8 1334 J5 J4 1338 J3
(9–16) J201 1186 8 J201 8 1201 J2
J6
J17 1335 J1 8 1339 J23 8
PP1 A15 (QUICK
J10 1329 J78 (MC SW FLTR) 1217 J10 J24 8 1336 J7 J2 1340 8 J22
B DISCONNECT)
(MULTICOIL)
1 J19 1337 J8 J3 1341 J21 CONN
PP1 A16 8
1 J3 1188 J77 (MC BIAS FLTR) 1216 J5
MG3 A19 MG2 A14 A1 (MIC)
(IMAGE
1330 1203
SPLICE

J5 J203 PP1 J11


EO2 REJECT
4,5,6,GRN/YEL,N 968 J1 AC IN MC SW FLTR
FILTER)
(EMERGENCY 1206 RCV FLTR OUTPUT MG3 A3 A5
GRN/YEL 037 (A12) (BODY
OFF SWITCH) DRIVER MG2 A12 A2 FLEXIBLE INPUT PREAMPLIFIER)
MODULE (RF BODY COIL) EXTENSION 893
1 1
MR3 PD1 J2 971 J23 J21 1191 J92 PP1 A11 J72 778 J3 FRONT
CABLE J2
(POWER J22 1192 J93 (FILTER) J73 779 J4 FRONT 257 1250 J1
4 DISTRIBUTION 701 J15 J23 1193 J94 J75 781 J5 BACK J1 744 J4 ISO. BOX J4
UNIT) 9
RF DOOR J24 1194 J95 J76 782 J6 BACK J2 743 J3 ISO. BOX J3 MG3 A3
(RED) SWITCH J25 1195 J96 J74 780 (BODY
1 1
780 J6 HYBRID SPLITTER)
2
1196 J35 PP1 J89
J125 1184 J32 PHPS J14 715
T J155(R)
J47 387
J20 385 JP1 MG2 A30
1181 J154(P) J15 300 (PATIENT ALIGNMENT
J11 318 LIGHTS)
16,17, GRN/YEL 970 MG3 A7 MG2 A15 FIBER–OPTIC
F5/F6 (LONGITUDINAL
7, N, GRN/YEL 1182 1182 MOTOR DRIVE)
(+,–) (BORE LIGHT) BUNDLE
784 MG2 A14
30,GRN/YEL,N 1183 1183 F7 786 MG6 AIR HOSE
(PATIENT COMFORT
F8 (BLOWER MODULE)
BOX AIR HOSE J7
GND 045 TO GRADIENT COIL MG3 A2
J8 MG2 A14 A2 (SPEAKER)
1085 (PATIENT
J12 282 J6
COMMUNICATION
296 296 J18 297 BOX) J9 1378 MG2 A35
EO1 (EMERGENCY OFF SWITCH) 1 (FRONT BELL MICROPHONE)

REV 7 6-3 DIRECTION 5166001


GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2 CABLE IDENTIFICATION


Cables delivered to site for installation have specifications dependent on its usage. The following tables contain
information on cable runs for this system as it was originally installed.
The table below provides definitions for the following cable tables.

COLUMN HEADING DESCRIPTION SPECIAL NOTES


RUN The run number as seen on the delivered cable The vendor cable may not have run numbers.
“FROM” & “TO” Labels the destination of the cable
DESCRIPTION Could include information on length, number of Total length of cable delivered to site, type of
conductors, type of cable, and end plug type cable specifies power from data.
CABLE TYPE Type designation for the cable
VOLTAGE RATING The highest voltage that may be continously applied Specific to the core cable manufacturer given.
to a wire
ACTUAL VOLTAGE Nominal for continuous; maximum for variable If voltage varies during use, the maximum is
voltage given.
TEMP RATING Maximum temperature insulating material my be Specific to the core cable manufacturer given.
used without loss of basic properties
FLAME RATING Measure of material’s ability to support combustion Specific to the core cable manufacturer given.
REMARKS Describe the cable’s use in the MR system

10–2–1 PDU POWER AND GROUND CABLES


The cables listed in Table 10–1 are connected at one end to the PDU module in the ACGD/PDU Cabinet.

TABLE 10–1
PDU POWER AND GROUND CABLES
VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP

RUN “FROM” “TO” DESCRIPTION REMARKS

037 PD1 GND MR2 A12 GND 36 ft Ground Wire (#10) SO 600 0 105 VW–1 System Cabinet ground
040 MSI RF COM GND 69 ft Ground Wire (#1/0) SO 600 0 105 VW–1 Magnet Ground Cable
045 PP1 GND MG6 GND 41 ft Ground Wire (#10) SO 600 0 105 VW–1 Blower Cabinet Separate
ground
968 MR1 A12 J1 MR3 PD1 35 ft Power (5–#8) SO 600 208Y 75 FT–4 System Cabinet Power

970 WC1 MR3 PD1 82 ft Power (3–#10) STO 600 208 90 FT–4 Water Chiller Power

1081 OW1 A19 J1 MR3 PD1 65.5 ft Power (4–#10) SO 600 120 75 FT–4 Operator Workspace Power

1182 PP1 PHPS PWR MR3 PD1 61 ft. Power (3–#10) AWM 600 120 90 FT–4 PHPS Module Power

1183 PP1 F7, F8 MR3 PD1 41 ft Power (3–#10) SO 600 208 105 VW–1 Blower Cabinet Power
1234 MR2 RF AMP MR3 PD1 35.5 ft Power (4–#8) SO 600 208 90 FT–4 RF Amplifier Power

1254 PD1 GND RF COM GND 69 ft Ground Wire (#1/0) SO 600 0 105 VW–1 RF Shield Ground Cable
N/A PD1 MAG/SHIM MAG/SHIM 50 ft Power (5–#8) With Cable for Compact PDU to GE
OUTLET POWER SUPPLY 30A Plug and Receptacle Magnet Power Supply or SC
Shim Supply Power

10-5 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–2 MAGNET ROOM CABLES NOT CUT TO LENGTH


The cables listed in Table 10–2 are routed and connected in the Magnet Room and are not cut to length.

TABLE 10–2
MAGNET ROOM RUNS (NOT CUT TO LENGTH)

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

282 PP1 J12 MG3 A2 J6 45 ft. 25 pin Sub–D CI3 300 <30 80 FT–4 Patient Intercom

297 PP1 J18 EO1 90 ft. 9 pin Sub–D CI3 300 <30 80 FT–4 Emergency off switch

300 PP1 J15 MG2 A15 60 ft. 25 pin Sub–D CI3 300 <30 80 FT–4 Bore Light power

318 PP1 J11 MG3 A7 45 ft. 9 pin Sub–D CI3 300 38.5 80 FT–4 Long Drive Motor power
VDC
385 PP1 J20 MG2 A30 67 ft. 37 pin Sub–D CI3 300 <30 80 FT–4 Patient Alignment Light power

387 PP1 J47 MG2 A29 J1 60 ft. 25 pin Sub–D CI3 300 38.5 80 FT–4 Table Elevation motor power
VDC
421 MG2 A33 J1 MG3 A31 J1 15 ft. 15 pin Sub–D CI3 300 <30 80 FT–4 Dock Limit

422 MG2 A33 J2 MG3 A1 J4 25 ft. 37 pin Sub–D CI3 300 <30 80 FT–4 Long Drive Encoder/Limit
Switch
458 PP1 J17 OM3 9 pin Sub–D CI3 300 <30 80 FT–4 Optional Remote O2 Sensor

715 PP1 J14 MG2 A33 J1 65 ft. 37 pin Sub–D CI3 300 <30 80 FT–4 SRI Power

716 PP1 J89 MG2 A11 A1 J2 65 ft. 15 pin Sub–D CI3 300 <30 80 FT–4 PAC Power

743 MG2 A12 A1 J2 MG3 A3 A1 J3 10 ft Coax w/M&F “N” STO 2500 950 85 VW–1 Body Coil Drive
VDC
744 MG2 A12 A1 J1 MG3 A3 A1 J4 10 ft Coax w/M&F “N” STO 2500 950 85 VW–1 Body Coil Drive
VDC
778 PP1 A11 J72 MG2 A12 A2 J3 70 ft. MHV Coax CI2 1900 1000 80 VW–1 Dynamic Disable (body coil
VRMS VDC frnt)
779 PP1 A11 J73 MG2 A12 A2 J4 70 ft. MHV Coax CI2 1900 1000 80 VW–1 Dynamic Disable (body coil
VRMS VDC frnt)
780 PP1 A11 J74 MG3 A3 A3 J6 60 ft. MHV Coax CI2 1900 1000 80 VW–1 Inductive Drive
VRMS VDC
781 PP1 A11 J75 MG2 A12 A2 J5 60 ft. MHV Coax CI2 1900 1000 80 VW–1 Dynamic Disable (body coil
VRMS VDC rear)
782 PP1 A11 J76 MG2 A12 A2 J6 60 ft. MHV Coax CI2 1900 1000 80 VW–1 Dynamic Disable (body coil
VRMS VDC rear)
784 PP1 F5, F6 MG6 J1 40 ft. 2 cond power SO 600 120 80 FT–4 Patient cooling blower
cable power
786 PP1 F7, F8 MG6 J2 40 ft. 2 cond power SO 600 120 80 FT–4 Gradient cooling blower
cable power
848 MG2 A11 J4 MG2 A4 J5 16 ft. 25 pin Sub–D/rt. CI3 300 <30 80 FT–4 Operator Display to PAC II
angl
862 MG2 A4 J1 MG2 A5 J1 2 ft. 26 pin Rt. angle CI2 300 <30 80 VW–1 Operator Display to Rt. Op
Cntrl
863 MG2 A4 J2 MG2 A5 J2 2 ft. 26 pin Rt. angle CI2 300 <30 80 VW–1 Operator Display to Lt. Op
Cntrl
869 MG3 A2 J8 MG2 A14 A2 SPK 10.5 F ft. 9 pin Sub–D CI3 300 <30 80 FT–4 Speaker Cable

893 MG3 A3 J2 MG3 A3 A5 INPUT 6.6 Inch BNC 1900 <30 80 VW–1 Spliiter to Preamp Cable
VRMS
897 MG2 A4 J3 MG2 A33 J4 20 ft. 37 pin Sub–D/26 CI3 300 <30 80 FT–4 Operator Display to SRI
pin

10-6 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–2 MAGNET ROOM CABLES NOT CUT TO LENGTH (Continued)

TABLE 10–2 (CONT’D)


MAGNET ROOM RUNS (NOT CUT TO LENGTH)

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

1153 LPCA J1 PED I/O J3 3400mm RG142

1200 PP1 J200 MG3 16CHSW J1 8–Coax 8W8 Combo D CATV 1500 <30 85 FT–4
R
1201 PP1 J201 MG3 16CHSW J2 8–Coax 8W8 Combo D CATV 1500 <30 85 FT–4 16 Channel only
R
1202 PP1 J202 MG3 A17 J3 2–Coax 8W8 Combo D CATV 1500 <30 85 FT–4
R
1203 PP1 MC SW MG3 16CHSW 49 ft. 37 pin Sub–D CI3 300 <30 80 Ft–4
FLTR J203 J11
1206 MG3 IPED I/F J4 MG3 A36 J35 1550mm/600mm 2 x 50
MG3 A3 A5 J11 Ohm Coax Rd316–DS
1214 LPCA A16 J20 16CHSW A36 J15 3950mm 9 pin Sub–
D/Molex
1215 MG3 16CHSW MG2 SRI J20 37 PIN SUB–D CI3 300 <30 80 FT–4
J20
1216 PP1 A16 J77 MG3 16CHSW J5 25/15 pin Sub–D CI3 300 <30 80 FT–4

1217 PP1 A15 J78 MG3 16CHSW 37 pin Sub–D CI3 300 <30 80 FT–4
J10
1250 MG3 A3 J1 CONNECT TO Times Microwave LMR CATV 6000 8 FTR
RUN 257 600FR R VDC

1260 MG2 A33 J10 MG2 16CHSW 35 ft 3 cond w/ 9–Sub D CI3 300 <30 105 FT–4 Bore Temperature Sensor

1261 MG2 A4 J4 MG2 A33 J5 20 ft. Data Cable CI2 300 <30 80 VW–1 SRI Splitter to Oper Display
(80–#28)
1331 LPCA J2 16 CHSW #1 J23 3400mm

1332 MC–9 TO MC–16 16 CHSW #1 J21 3100mm 8 x 50 Ohm


Coax
1333 MC J21 16 CHSW #1 J18 3500mm 37 pin Sub–D

1334 IMAGE REJECT 16 CHSW #1 J22 1000mm 8 x 50 Ohm 16 Channel only


#1 J5 Coax Rg316–DS
1335 IMAGE REJECT 16 CHSW #1 J17 1000mm 37 pin Sub–D
#1 J6
1336 IMAGE REJECT 16 CHSW #1 J24 1000mm 8 x 50 Ohm 16 Channel only
#1 J7 Coax Rg316–DS
1337 IMAGE REJECT 16 CHSW #1 J19 1200mm 25 pin Sub–D
#1 J8
1338 IRF #1 J4 LPCA J3 3500mm

1339 IRF #1 J1 LPCA J23 3400mm 8 x 50 Ohm 16 Channel only


Coax
1340 IRF #1 J2 LPCA J22 3500mm 37 pin Sub–D 16 Channel only

1341 IRF #1 J3 LPCA J21 3400mm 8 x 50 Ohm 16 Channel only


Coax
1370 MC j18 PED I/O J2 3250mm 9 pin Sub–D

1378 MG3 A35 MG3 A2 J9 30 ft. 9–pin Sub–D CI3 300 <30 80 FT–4 Front Cover Microphone

10-7 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–3 MAGNET MONITORING CABLES


The Magnet Monitoring cables listed in Table 10–3 are delivered with the Magnet.

TABLE 10–3
MAGNET MONITORING CABLES DELIVERED WITH MAGNET

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

823 MSM1 A1 J10 MR2 A11 J24 Data Cable (15–#22) CI3 300 <30 80 FT–4 Monitor to System Cabinet I/F

824 PP1 J10 FJ1 Data Cable (9–#22) CI3 300 <30 80 FT–4 Monitor Sensor I/F to
Penetration Panel
825 PP1 J48 MSM1 A1 J8 Data Cable (25–#22) CI3 300 <30 80 FT–4 Magnet Monitor connector to
Penetration Panel
826 MSM1 A1 J9 FJ3 Data Cable (15–#22) CI3 300 <30 80 FT–4 Monitor to Compressor I/F

827 FJ3 MS5 A5 A1 J1 / Data Cable (8–#22) & CI3 300 <30 80 FT–4 Compressor I/F
FJ4 Data Cable (4–#22)
828 PP1 J10 MA1 A3 A1 P403 Data Cable (9–#22) CI3 300 <30 80 FT–4 Penetration Panel to Sensor

829 PP1 J48 MS1 FJ2 Data Cable (25–#22) CI3 300 <30 80 FT–4 Penetration Panel to Magnet
I/F
830 MS1 FJ2 MS1 A2 J2/ 3cond/3cond/25cond CI3 300 <30 80 FT–4 Magnet I/F
MS1 A5 A2 J1/
MG3 TB1
831 FJ1 MSM1 A1 J7 Data Cable (8–#24) CI3 300 <30 80 FT–4 Monitor to Sensor I/F

832 MS1 A2 J1 MS1 SLEEVE/ Data Cable (8–#22) CM 300 <30 80 FT–4 Buffer Amp to Cooling Sleeves
MS1 COLDHEAD
833 FJ4 MS5 A1 A6 JR Data Cable (6–#24) CM 300 <30 80 FT–4 Compressor to Compressor
I/F
940 MSM3 RS232 MSM1 J1 6 ft. Data Cable (9–#22) CI3 300 <30 80 FT–4 Monitor to Modem
MR2 A11 J24 MR2 IPG J9/ MR2 System Cabinet I/F
PDU J5

10-8 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–4 GE MAGNET CABLES


The Magnet cables listed in Table 10–5 arrive on site with the magnet and are installed at the time of Magnet delivery.

TABLE 10–4
GE MAGNET CABLES

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

601 MS8 MS1 Power Cable (#4/0) SO 600 208 80 VW–1 Ramp cables, Power Assembly
Positive 4/0
602 MS8 MS1 Power Cable (#4/0) SO 600 208 80 VW–1 Ramp cables, Power Assembly
Negative 4/0
603 MS7 MS1 A3 A1 Power Cable (#4/0) SO 600 208 80 VW–1 Ramp and Shim Cable Assem-
bly
604 MS7 or MS8 MS1 A3 A1 Data Cable (12–#22) CI3 150 60 80 FT–4 Switch Heaters
VDC
605 MS1 MR2 A8 Data Cable (4–#22) CI3 300 <30 105 FT–4 Instrumentation Cable, 80 feet

606 MS4 MS1 A3 A1 Data Cable (4–#22) CI3 300 <30 105 FT–4 Emergency Run Down Unit

608 MS1 MR2 A8 Data Cable (4–#22) CI3 300 <30 105 FT–4 Instrumentation Cable, 15 feet

609 MS7 OR MS8 PDU Power Cable (5–#16) SO 300 208Y 105 FT–4 Feeder Cables

621 MS5 A1 MS1 A2 7/8” Gas Flex Line STO N/A N/A 85 VW–1 Supply Flexible Gas Line (PP1
J60 or J61)
622 MS5 A1 MS1 A2 7/8” Gas Flex Line STO N/A N/A 85 VW–1 Return Flexible Gas Line (PP1
J60 or J61)
623 MS5 A1 PP1 F1, F2, F3, Power Cable (4–#18) SO 600 200 105 VW–1 Compressor to Pen Panel
F4 Cable (equipment room)
624 MS1 A2 PP1 F1, F2, F3, Power Cable (4–#18) SO 600 200 105 VW–1 Cold Head Control Cable
F4

10–2–5 FIBER OPTIC CABLES


The Fiber Optic cables listed in Table 10–5 are routed and connected in the Magnet Room and the Equipment Room.

TABLE 10–5
FIBER OPTIC CABLES
VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
CABLE
TYPE

FLAME
TEMP

RUN “FROM” “TO” DESCRIPTION REMARKS

711/ PP1–J91 MG2 A33 and 100 ft. 3/4 in. fiber N/A N/A 0 85 N/A One duplex fiber optic cables
712 MG2 A11 A1 optic conduit and one simplex fiber optic
cable and One duplex plastic
fiber optic cable for PAC–II
1045 MR2 A11 J13 PP1 A17 A1 3/4 in. fiber optic con- N/A N/A 0 85 N/A Two duplex fiber optic and one
duit simplex fiber optic cables

1179 MR2 A11 J17 MR3 A11 J2 20 ft. 3/4 in. fiber optic N/A N/A 0 85 N/A Four glass fiber optic cables
conduit

10-9 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–6 EQUIPMENT ROOM AND MAGNET ROOM CABLES AND HOSES CUT TO LENGTH
The Gradient cables and the RF Head and Body cables listed in Table 10–6 will be routed, cut, and terminated at
designated cable ends according to instructions found in the CABLE INSTALLATION section of this manual.

TABLE 10–6
EQUIPMENT ROOM AND MAGNET ROOM CABLES & HOSES CUT TO LENGTH

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

536 WC1 MG2 A12 A1 Flexible Tubing N/A N/A N/A N/A N/A Gradient Coil Cooling Line

537 WC1 MG2 A12 A1 Flexible Tubing N/A N/A N/A N/A N/A Gradient Coil Cooling Line

235 CONNECT TO PP1 J44 Times Microwave CATV 5000 85 FT–4 1.5T Body Coil Run created
RUN 1259 LMR–900FR R VDC from kit

257 PP1 J44 CONNECT TO Times Microwave CATV 5000 85 FT–4 1.5T Body Coil Transmit
RUN 1250 LMR–900FR R VDC

762 MR3 CABI/F J3 PP1 A7 1, 2 2 Conductor #2 STO 1000 600 90 FT–4 X Gradient output

762 MG3 A32 1/2 PP1 A7 1, 2 2 Conductor #2 STO 1000 600 90 FT–4 X Gradient output

763 MR3 CABI/F J4 PP1 A7 3, 4 2 Conductor #2 STO 1000 600 90 FT–4 Y Gradient output

763 MG3 A32 3/4 PP1 A7 3, 4 2 Conductor #2 STO 1000 600 90 FT–4 Y Gradient output

764 MR3 CABI/F J5 PP1 A7 5, 6 2 Conductor #2 STO 1000 600 90 FT–4 Z Gradient output

764 MG3 A32 5/6 PP1 A7 5, 6 2 Conductor #2 STO 1000 600 90 FT–4 Z Gradient output

1238 PP1 J4 CONNECT TO Times Microwave CATV 6000 85 FT–4 RF Head Transmit
RUN 1249 LMR–600FR R VDC

1328 PP1 J4 MG3 PEDI/F J2 Times Microwave CATV 4000 85 FT–4 RF Head Transmit
LMR–600FR R VDC

10-10 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–7 EQUIPMENT AND OPERATOR ROOM CABLES

The cables listed in Table 10–7 are routed and connected between Penetration Pnael, cabinets in equipment room and
Operator Workspace.

TABLE 10–7
EQUIPMENT & OPERATOR ROOM RUNS

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

296 FACILITY PP1 J18 45 ft 9 pin Sub–D CI3 300 <30 80 FT–4 Emergency off switch
DISCNCT
457 OM1 PP1 J17 9 pin Sub–D CI3 300 <30 80 FT–4 Optional O2 Monitor cable

701 MR2 A11 J15 RF Door switch 70 ft 9 pin Sub–D CI3 300 <30 80 FT–4 RF Door switch

971 MR2 A11 J23 MR3 PD1 J2 40 ft 15 pin Sub–D CI3 300 <30 80 FT–4 PDU Status (ACGD)

1056 OW1 A17 PRT6 OW1–A12 3.3 ft. Ethernet Cat 5 CI3 30 <30 80 FT–4 Laptop Access cable
ETHRNET
1059 OW1 A17 PRT2 OW1–A15 5.9 ft. Ethernet Cat 5 CI3 30 <30 80 FT–4 Lan Hub cable
ETHRNET
1060 OW1 A19 J3 OW1 A21 J17 3.9 ft. 9 pin Sub–D CI3 300 <30 80 FT–4 WIM Power cord

1064 OW1 A15 SCSI 1 OW1 A16 SCSI 9.9 ft. SCSI cable 50 CI3 300 <30 80 VW–1 SCSI Tower Data cable
TOWER
1065 OW1 A11 J2 OW1 A6 Pwr 9.9 ft. Power (3–#18) SVT 300 120 80 VW–1 Host LCD Display Power
cable
1066 OW1 A15 OW1 A13 3.9 ft. 25/9 pn Sub–D CM 300 <30 80 FT–4 Modem I/O cable

1067 OW1 A15 SCSI 2 OW1 A5 DASM 5.9 ft. SCSI cable 50 CI3 30 <30 80 VW–1 DASM I/O cable

1070 OW1 A11 J11 OW1 A16 Pwr 9.9 ft. Power (3–#18) SVT 300 120 80 VW–1 SCSI Tower Power cable

1071 OW1 A11 J1 OW1 A15 Pwr 9.9 ft. Power (3–#18) SVT 300 120 80 VW–1 Host Computer Power cable

1074 OW1 A19 J2 OW1 A11 3.3 ft. Power (3–#14) SVT 600 120 105 VW–1 Power Strip cord

1075 OW1 A19 J4 OW1 A11 3.3 ft. Power (3–#14) SVT 600 120 105 VW–1 Power Strip cord

1085 OW1 A21 J7 PP1 J12 65 ft 25 pin Sub–D CI3 300 <30 80 FT–4 Intercom

1180 MR3 J7 MR2 A11 J56 26 ft. 9 pin Sub–D CI3 300 <30 80 FT–4 Can

1181 MR3 J8 PP1 PHPS J154 65 ft 9 pin Sub–D CI3 300 <30 80 FT–4 Can

1184 MR2 A11 J125 PP1 PHPS J32 65 ft 25 pin Sub–D CI3 300 <30 80 FT–4

1185 MR2 A11 J200 PP1 J200 8–Coax 8W8 Combo D CATV 1500 <30 85 FT–4
R
1186 MR2 A11 J201 PP1 J201 8–Coax 8W8 Combo D CATV 1500 <30 85 FT–4 16 Channel only
E
1187 MR2 A11 J202 PP1 J202 2–Coax 8W8 Combo D CATV 1500 <30 85 FT–4
R
1188 MR2 DRV J3 PP1 A1 J77 65 ft 15 pin Sub–D CI3 300 <30 80 FT–4 16 Channel SW Cntrl Power

1191 MR2 DRV J21 PP1 A11 J92 65 ft. RG 58C/U Coas CI2 1900 1000 80 VW–1 Dynamic Disable
VRMS VDC
1192 MR2 DRV J22 PP1 A11 J93 65 ft. RG 58C/U Coas CI2 1900 1000 80 VW–1 Dynamic Disable
VRMS VDC
1193 MR2 DRV J23 PP1 A11 J94 65 ft. RG 58C/U Coas CI2 1900 1000 80 VW–1 Dynamic Disable
VRMS VDC
1194 MR2 DRV J24 PP1 A11 J95 65 ft. RG 58C/U Coas CI2 1900 1000 80 VW–1 Dynamic Disable
VRMS VDC
1195 MR2 DRV J25 PP1 A11 J96 65 ft. RG 58C/U Coas CI2 1900 1000 80 VW–1 Dynamic Disable
VRMS VDC

10-11 APPENDIX
GE Healthcare Signa HDx & HDi 1.5T System Installation
REV 7 DIRECTION 5166001

10–2–7 EQUIPMENT AND OPERATOR ROOM CABLES (Continued)

TABLE 10–7 (CONT’D)


EQUIPMENT & OPERATOR ROOM RUNS

VOLTAGE

VOLTAGE
RATING

ACTUAL

RATING

RATING
FLAME
CABLE
TYPE

TEMP
RUN “FROM” “TO” DESCRIPTION REMARKS

1196 PP1 A17 J1 PP1 PHPS J35 2 ft 15 pin Sub–D CI3 300 <30 80 FT–4 Repeater Board Power

1249 MR8 J3 CONNECT TO 5 ft. Times Microwave CATV 6000 85 FT–4 RF Extension cable
RUN 1238 LMR R VDC
1255 OW1 A21 J18 MR2 A11 J26 60 ft 9 pin Sub–D CI3 300 <30 80 FT–4 E–stop

1259 MR2 RF AMP J4 CONNECT TO 5 ft. Times Microwave CATV 2500 85 FT–4 RF Extension cable
RUN 235 LMR 400FR R VDC

1320 OW1 A15 USB HUB IN 2000mm RJ11 Temp/Humidity Sensor

1321 OW1 A2 USB HUB OUT 2000mm RJ11 Temp/Humidity Sensor

1322 MR2 A11 Cabi- HUB C2 100 ft. RJ11 Temp/Humidity Sensor
net Sensor
1329 MR2 DRV J10 PP1 A17 J78 65 ft 37 pin Sub–D CI3 300 <30 80 FT–4 MC Drive

1330 MR2 DRV J5 PP1 J203 65 ft 37 pin Sub–D CI3 300 <30 80 FT–4 MC Drive

1371 OW1 A15 ETH D MR2 A11 J63 75 ft. Ethernet Cat 5 CI3 300 <30 80 FT–4 EXCITE Ethernet C‘cable

1379 Bar Code Reader OW1 A15 KYBRD 10 ft. Mini Din 6 Male Bar Code Reader Extension
to Femalre extender Cable (option)
2037 OW1 A15 Video 1 OW1 A6 MAIN 9.9 ft. 29 pin DVI, 15 pin CI3 300 <30 80 VW–1 Host LCD Display Video cable
LCD VGA
2038 OW1 A7 KYBD OW1 A15 USB 1000mm USB or Keyboard extension cable
2000mm USB
2039 MOUSE HOST PC 1000mm USB Mouse extension cable

2053 OW1 A20 OW1 A21 J6 13.2 ft. 50 pin SCSI, 9 CI3 300 <30 80 VW–1 SCIM cable
pin Sub–D

10–2–8 CABLE SPARES (FRU’S)


The generic cables listed below are designated as replacement cables for the associated cable Runs.

CABLE PART NUMBER CABLE RUN


46–244710G30 Run 1227
Run 1236
46–244712G30 Run 1188
46–244713G30 Run 1181
46–244714G30 Run 1180
Run 1181
Run 1233
46–258877G30 Run 1189
Run 1190