SYRINGE PUMP OPERATING MANUAL

0 SP-8800 / UME-02-REV10
Thank you for purchasing our SP-8800 Syringe Pump.
In order to use this pump correctly and safely, read this manual carefully before operating SP-8800 Syringe Pump.
If you have any questions as you are reading through this manual, call the local authorized dealer in your country.
Retain this manual together with the unit for future reference.
This SP-8800 Syringe Pump is intended for the infusion of chemicals such as vasopressor drugs, depressor
drugs, anti-coagulants, anesthetic drugs, anti-cancer drugs, oxytocic, nutrition, drug for chemotherapy medication
and blood infusion at ICU, CCU, NICU, or operation room.
This device is designed for high flow-rate accuracy and ease of handle in the infusion of solutions with the
equipped audio-visual alarms and state-of-the-art sensors.

Features

⚫ Self testing – Every time the unit is turned on, self testing is proceeded.
⚫ Automatic syringe detection – Detect syringe size automatically. (10, 20, 30, 50ml )
⚫ Shortcut key – Syringe brand set, Occlusion set, Buzzer level set
⚫ K.V.O.(Keep Vein Open) – When the delivery volume has been reached, flow rate turns into K.V.O.
rate (adjustable from 0.1ml/h to 10ml/h) automatically.
⚫ Keypad lock function – Keypad is locked with or without password depends on its using
condition.
⚫ Infusion setup – Flow rate/ Delivery volume/ Infusion time
⚫ Retain memory – Last setting value is retained.
⚫ Infusion remaining time display
⚫ Alarm repeat function
⚫ Open system – Calibration up to 10 syringe brands in a single unit.
⚫ Buzzer level – 3steps
⚫ Occlusion level – 9steps (3.8psi~19.3psi)
⚫ Purge rate – Adjustable (0.1~1200ml/h)
⚫ Bolus rate – ON/OFF, Adjustable (0.1~1200ml/h, 0.1~1000ml)
⚫ Anti-bolus – Reduce significantly bolus after occlusion release. (ON/OFF)
⚫ History call back – Call back last 10 infusion data.
⚫ Dosage mode(Body weight mode) – Automatically calculate and set a proper flow rate when dose
rate, body weight, drug mass and solution volume are entered.
⚫ Unit setup for Dosage mode – ml/h, mg/kg/h, mg/kg/min, ㎍/kg/h, ㎍/kg/min
⚫ Nurse call – Each unit can be connected with nurse call system (DC 24V, 0.5A) (Option)
⚫ Profile function – Program infusion condition (flow rate, delivery volume, delivery time)
differently for every hour, up to 24hours. (Option)
⚫ Central system – Connect with main PC for monitoring (Option)

0 SP-8800 / UME-02-REV10
INTRODUCTION / FEATURES ············································· 1

1. DESCRIPTION OF SYRINGE PUMP ································ 4


1-1.Front View ·································································································· 5~6

1-2.Rear View ·································································································· 7

1-3.Side View ··································································································· 7

1-4.Syringe View ······························································································ 8

1-5.Components ······························································································ 8

2. PRIOR TO PUMP USE ······················································ 9


2-1.Explanation of Symbols ············································································· 10
2-2.Warnings ···································································································· 10

2-3.Precautions ································································································ 11

2-4.Cleaning & Sanitation ················································································ 11

2-5.Storage ······································································································ 12

2-6.Maintenance & Repair ··············································································· 12~13

3. OPERATION ······································································ 14
3-1.Initial setup ································································································· 15

3-2.Syringe installation ····················································································· 16

3-3.Setup flow rate (ml/h) ················································································· 17

3-4.Setup delivery volume (ml) ········································································ 18

3-5. Confirm total infused volume (Σ㎖/CLEAR)··············································· 18

3-6.Setup delivery time (hours) ········································································ 19

3-7.Insert the needle into the patient ······························································· 19

3-8.Start infusion ······························································································ 19

3-9.Infusion completion ···················································································· 20

3-10.Stop infusion ···························································································· 21

3-11.[POWER] button ······················································································ 21

3-12.Battery remain ························································································· 21

3-13.Error alarm ······························································································· 22

3-14.DC Input operation ···················································································· 22

3-15.Battery operation ····················································································· 22

3-16.Nurse call function (Option) ···································································· 22

3-17.Keypad lock ····························································································· 23

1 SP-8800 / UME-02-REV10
CONTENT
INTRODUCTION / FEATURES
NTS
4. SYSTEM SETUP ······························································· 24
4-1.Buzzer level setup (Short-cut key) ···························································· 25

4-2.Occlusion level setup ················································································ 26

4-3.Syringe selection ······················································································· 26

4-4.Infusion set ································································································ 27

4-5.Display ······································································································· 30

4-6.Config ········································································································ 31

5. SPECIAL FUNCTIONS ····················································· 33


5-1.Dosage ····································································································· 34

5-2.History ········································································································ 35
5-3.Profile (Option spec) ·················································································· 36

5-4.Drug label (Option spec) ············································································ 37

5-5.Unit setup ··································································································· 37

5-6.Pro. Mode ·································································································· 37

6. TROUBLE SHOOTING – ALARM ···································· 38~41

SPECIFICATIONS ································································· 42

SYMBOLS ············································································· 43
WARRANTY ·········································································· 44~45

2 SP-8800 / UME-04-REV.12
1. DESCRIPTION
OF PUMP

1-1.Front View
1-2.Rear View
1-3.Side View
1-4.Syringe View
1-5.Components

3 SP-8800 / UME-04-REV.12
1-1. FRONT VIEW

4 SP-8800 / UME-02-REV10
1. DESCRIPTION OF PUMP

제품 설명
① LCD : Display operating state.
② FLOW RATE: Set up and adjust Flow rate(㎖/h)
③ DEL. VOL : Set up and adjust Delivery volume(㎖)
④ TOTAL VOL: Display total infusion volume(㎖)
⑤ NUMERICAL BUTTONS
⑥ SEL : Save setup value / Set up infusion time.
⑦ ESC/2sCLR : Move to previous state/ Delete setup value by press the button for 2seconds
⑧ MENU/2s : Move to “MENU” mode/ Lock keypad by press the button for 2seconds
⑨ SILENCE :Hold alarm sound for 2minutes when alarm goes off. After 2minutes, alarm foes off again.
⑩ START
⑪ STOP
⑫ POWER : Press for 2seconds to turn on or off the unit.
⑬ SLIDER : Push the plunger of the syringe to infuse.
⑭ HANDLE LEVER
⑮ SLIDER CLUTCH
16 SLIT
17 CLAMP
18 PURGE - STANDBY mode : Remove air bubble in IV tube line.
- During infusion : BOLUS function, Bolus can be set in system menu mode.
19 DIRECTION BUTTONS
20 Line Guide : Fix extension tube.
21 Clamp State
22 Slider Clutch State : When Slider Clutch does not fix the Push Button of syringe, blinked in red.
23 Handle Lever State : When Clutch is opened(Handle Lever is up), blinked in red.
24 Display FLOW RATE
25 Display DEL VOLUME
26 Display TOTAL VOLUME
27 Display INFUSION TIME
28 Display Syringe BRAND
29 Syringe Size: 10㎖, 20㎖, 30㎖, 50㎖
30 Display state of battery
31 Display state of BUZZER level
32 Display state of lock : Lock / Unlock
33 Power supply : AC / BATTERY
34 BATTERY REMAIN : Display remaining battery operating time.
※ Battery remaining time can be very different depends on battery condition.
35 CHARGING : ( Fully charged : Green / Charging : Orange)

5 SP-8800 / UME-04-REV.12
1. DESCRIPTION OF PUMP

제품 설명
1-2. REAR VIEW
① ② ③ ④ ⑤ ⑥

①Communication Connector (RS232 Connector)


②Nurse Call Connector (Nurse Call Connector): Connect alarm terminal cord installed in the ward.
③Pole Clamp
④Handle
⑤DC Input Port
⑥AC Power Supply Connector

1-3. SIDE VIEW

① Coupler: Joins two syringe pumps when one syringe pump is laid on another syringe pump

6 SP-8800 / UME-04-REV.12
1. DESCRIPTION OF PUMP

제품 설명

1-4. SYRINGE

Barrel

Finger-grips

Plunger

1-5. COMPONENT

AC power cord

7 SP-8800 / UME-04-REV.12
2. PRIOR TO
PUMP USE

2-1.Explanation of Symbols
2-2.Warnings
2-3.Precautions
2-4.Cleaning & Sanitation
2-5.Storage
2-6.Maintenance & Repair

8 SP-8800 / UME-02-REV10
2. PRIOR TO PUMP USE

2-1. EXPLANATION OF SYMBOLS 안전 및 올바른 사용 정보


Warning is used to indicate the presence of a hazard which can cause severe
personal injury, death or substantial property damage if the warning is ignored.

Caution is used to indicate the presence of a hazard which will cause minor personal
injury or property damage if the warning is ignored.

NOTE Note is used to notify the user of installation, operation or maintenance information
which is important but not hazard-related.

2-2. WARNINGS
• If used in the vicinity of the surgical operation equipment which generates a high frequency current such as
mobile (cellular) phone, radio, or defibrillator, the pump may malfunction since electrical interference.
Please carefully check for any sources of electrical interference in the vicinity before use.
• Before using the pump, make sure that the manufacturer of the syringe to be used is the same as the one
specified for this pump. If it is not matched, the flow rate accuracy and alarm system cannot be guaranteed.
In order to change the syringe manufacturer, please contact your local authorized dealer.
• When using the pump concurrently with the surgical operation equipment, please note the following:
- Do not use the pump together with any surgical operation equipment that generates high noise level.
- Be sure that the pump is kept a sufficient distance from the surgical operation equipment.
- The pump and such device should not be powered from the same outlet.
- Check and confirm the normal operation of the pump periodically.
※ In case of malfunction, immediately turn off the power and remove the infusion set from the patient’s
skin. After this action, please contact your local authorized dealer.
• Avoid using the pump in presence of flammable gases and flammable anesthetic mixture with air, oxygen or
nitrous oxide.
• The use of any mobile (cellular) phone near the pump is not allowed since the high frequency noise during the
conversation could cause malfunction of the pump.
• The use of the pump in MRI rooms such as high-pressure rooms or places where high electromagnetic radiation
is generated is not allowed.
• In case of using new brand of syringe other than suggested by AMPall, new brand of syringe must be calibrated
by AMPall or local authorization. If not, new syringe may cause problem on infusion accuracy and alarm state.
※ Please contact your local authorized dealer if you wish to use syringe that are not supported.
• This pump does not detect ‘over-infusion’ or ‘under-infusion’. Please check infusion state with patient regularly.
• Make sure about correct syringe installation, drug may flow backward if syringe is installed incorrect way.
• Eliminating a needle from the patient to prevent a release-bolus is required when the tube line got an occlusion.
• The pump does not detect damage to the infusion tube such as a leak in the tube or a rupture in the filter due to
pressure exertion. Therefore, regularly check for any damage to the infusion line during infusion.
9 SP-8800 / UME-04-REV.12
2. PRIOR TO PUMP USE
• When the flow is obstructed due to kinking of the tube or clogging of the needle or filter, it can increase pressure
안전 및 올바른 사용 정보
in the infusion tube and cause the tube to be inflated with the solution. Complete removal of the obstruction will
allow the solution to be delivered to the patient. If the flow is obstructed, take appropriate actions.
• The syringe is disposable. Therefore, the used syringe should not be reused.
• Never modify the pump by yourself.
• The pump is connected to the AC power outlet to be operated. If there is no available AC power outlet, pump
can be operated with only its internal battery. (Alternatively, the pump may be operated by DC power).
• The spill of the solution on the AC power inlet may cause a short circuit.
• Make sure that there is any damage with the pump and components. In case of the unit and components that
were shocked, do not use them even if visible damages are not observed. Please contact your local authorized
dealer.
• In case of malfunction, do not try to take the unit apart or attempt to repair by yourself. Please contact your local
authorized dealer immediately. If the user does not comply with these warnings, AMPall does not take liable and
the warranty does not apply.

2-3. PRECAUTIONS
• The pump does not detect about correct infusion state with blood vessel on patient. Please check the puncture
site and monitor the patient’s condition carefully.
• Fix the pump securely to a pole stand and check its stability. If the pump is dropped or given a hard shock, do
not use it. Please contact your local authorized dealer.
• The pump must be used in accordance with this instruction manual by trained medical personnel.
• When alarm sounds, please take corrective actions.
• Do not apply a strong force or damage to the clamp or slider.

If the user does not comply with these warnings, the company is not responsible for its
result and the warranty is not applied.

2-4. CLEANING & SANITATION


• Before cleaning the pump, make sure to power off the pump and disconnect the AC power cord.
• Do not immerse the pump in water.
• Do not immerse the pump in any liquid nor allow any liquid to leak into the pump.
• Do not use alcohol, thinner, benzene, ammonia, acetone or any other organic solvents.
• Do not sterilize the pump.

10 SP-8800 / UME-04-REV.12
2. PRIOR TO PUMP USE
< Casing >
안전 및 올바른 사용 정보
• When the pump casing is stained, use a gauze cloth or similar, moistened with cold or tepid water and
let it dry, especially the AC power inlet before use. Clean AC power inlet, DC INPUIT PORT, Slider hook, Clamp,
Slit and Clutch regularly.
• In case of disassemble Clutch or Clamp for cleaning, contact local authorization.
• Do not use a drier to dry casing parts.

2-5. STROAGE
• Avoid the following environment for storage and transport for SP-8800 Syringe Pump.
- Where the unit is exposed to dirt or heavy dust.
- Where the unit is exposed to salty atmosphere.
- Where the unit is exposed to severe vibration or corrosive GAS.
- Where the unit is exposed to rough handling.
- Where the unit is exposed to direct sunlight or UV light.
- Where the unit is exposed to water.
- Where the unit is exposed to extreme temperature and humidity.
• The optimum storage conditions for the unit

- Temperature is between -10℃ ~ 45℃.


- Humidity is between 10% ~ 95%

2-6. MAINTENANCE & REPAIR


• If any irregularity and failure are detected, stop operation of the pump immediately and contact your local
authorized dealer to repair or replace by supplying the details of the situation. Never try to disassemble or
repair by yourself because it could cause further serious failure.
• Make sure that there is not any damage with the pump and components. In case that the unit and components
were shocked, do not use them even if visible damages are not observed. Please contact your local authorized
dealer.
• Contact your local authorized dealer for periodical inspection of the pump for safety and longer product life.
• Operate the pump with the internal battery once a month to check its performance because the internal battery
is subject to aging. If the operation time is getting short after it is normally recharged, contact your local
authorized dealer to replace with a new battery. Please be sure that your local authorized dealer checks it
annually.
• Please recharge the internal battery fully for more than 6 hours by connecting the pump to an AC power outlet
before the pump is used for the first time or after a long interval.
• If the battery is low, the pump may stop running if there is no way to connect the pump to an AC power outlet.

11 SP-8800 / UME-04-REV.12
2. PRIOR TO PUMP USE

안전 및 올바른 사용 정보
the product and its electronic accessories (e.g., Power cord) should not be disposed
of with other household waste at the end of their working life. To prevent possible
harm to the environment or human health from uncontrolled waste disposal, please
separate these items from other types of waste and recycle them responsibly to
promote the sustainable reuse of material resources.

12 SP-8800 / UME-04-REV.12
3. OPERATION

3-1. Initial setup


3-2. Syringe installation
3-3. Setup flow rate (ml/h)
3-4. Setup delivery volume (ml)
3-5. Confirm total infused volume (Σ㎖/CLEAR)
3-6. Setup delivery time (hours)
3-7. Insert the needle into the patient
3-8. Start infusion
3-9. Infusion completion
3-10. Stop infusion
3-11. [POWER] button
3-12. Battery remains
3-13. Error alarm
3-14. DC Input operation
3-15. Battery operation
3-16. Nurse call function (Option)
3-17. Keypad lock

13 SP-8800 / UME-04-REV.12
3. OPERATION

3-1. INITIAL SETUP


3-1-1 ATTACHING THE PUMP TO THE IV POLE STAND
• Fix the pump securely on a pole stand, using the pole clamp on the back of the pump.

3-1-2 CONNECT AC POWER


• Insert AC power cord to the AC power inlet on the back side of the pump and connect the power cord to AC
power outlet. Charging lamp will be on in red or green depends on its battery charging state.
( Fully charged : Green / Charging : Orange)
• After turn on the unit,

- LCD displays on the left side of LCD when it connected with AC power.
- LCD displays connected with external DC power, and connected with battery for power
supply.

NOTE • Alternatively, the pump may be operated by DC power.

• When AC power is connected, the internal battery is automatically being recharged.

3-1-3 TURN ON THE UNIT


• Press [POWER] button for 2 seconds to turn on the unit.
- LCD, FND and LED state lamp turn on with confirming sound and SELF TESTING is proceeded for 3seconds.
- LCD divides left side and right side as below picture. Left side displays setup conditions, and right side
displays operating state.

* Power supply - : AC power


: DC Input
: Battery
- Syringe brand is displayed on the left side of LCD. Please confirm syringe brand, if it is the one you are going
to use with the unit.

Syringe brand State of syringe installation


Syringe size
Battery state FLOW RATE
Buzzer level
DEL VOL
Lock / Unlock TOTAL VOL
Power supply
Delivery time
Remaining battery
operating time

[SETUP CONDITION] [OPERATING CONDITION]

NOTE • Please contact a local authorized dealer or agency when you have problem with the
unit.
14 SP-8800 / UME-04-REV.12
3. OPERATION

3-1-4 TURN OFF THE UNIT


• Press [POWER] button for 2 seconds on “STAND BY” mode to turn off the unit.
- The unit is turned off after 2seconds with confirming sound and logo in LCD.
- The unit is not turned off in “MENU” mode.

3-2. SYRINGE INSTALLATION


1. Push up the handle lever and pull at the slider to the right.
2. Pull at the clamp and turn it to the left.(Open the clamp.)
3. Install syringe filled with solution and connected with infusion tube. Fix infusion tube into the slit.
4. Pull at the clamp and turn up to close the clamp and check the syringe is fitted correctly.

NOTE • If the syringe is not properly placed in the slit, the flow rate accuracy and alarm
system are not guaranteed.

5. After infusion tube is placed in Line Guide, push the slider slowly until it hits push-button of the syringe.
When the slider hits the push-button, slider hook automatically gets down to fix the syringe.
When the syringe is set in the pump, [SYRIGNE SIZE] LED lights on according to the size of syringe.
6. Press and hold [PURGE] button until a drop of solution at the end of the infusion set appears to remove
air bubble in the infusion tube.

Syringe Size Initial Purge Rate


10ml 300ml/h
20ml 400ml/h
30ml 500ml/h
50ml 1200ml/h

NOTE • The priming volume is added to the volume delivered.


• Refer to page 25 to set up purge rate.

15 SP-8800 / UME-04-REV.12
3. OPERATION

3-3. SETUP FLOW RATE(ml/h)


1. Press [FLOW RATE] button to get into FLOW RATE setup mode as below picture.

State of menu

Previous setup value

Flow rate (ml/h)

Flow rate setup range

[SETUP FLOW RATE]

2. Input desired value for FLOW RATE(ml/h) with numerical buttons and press [SEL] button to save it.
Display turns into previous display automatically.

3. Press [ESC/2s CLR] button for 2seconds to delete complete value, press button to delete previous

digit only.
4. When setup value exceeds the range, “RANGE OVER” will come up and maximum value is saved
automatically. Make sure setup value before start.

Syringe size Flow Rate


10ml 0.1ml/h ~ 300ml/h (0.1ml/h increasement)
20ml 0.1ml/h ~ 400ml/h (0.1ml/h increasement)
30ml 0.1ml/h ~ 500ml/h (0.1ml/h increasement)
0.1ml/h ~ 1200ml/h
50ml (0ml/h~999.9ml/h with 0.1ml/h increasement)

(1000ml/h~1200ml/h with 1ml/hncreasement)

NOTE • Please do not change the flow rate during the infusion.
To change the flow rate, the infusion should be stopped by pressing 'Stop' button first.
• During flow rate setup, if press [STOP] button for more than 2seconds, the value will be
reset as “0000”.
• The pump does not start when the flow rate is setup at 0 ml/h. The message “NO SET”
will come up on the LCD.

16 SP-8800 / UME-04-REV.12
3. OPERATION

3-4. SETUP DELIVERY VOLUME(ml)


1. Press [DEL.VOL] button to get into DELIVERY VOLUME setup mode as below picture.

State of menu
Previous setup value

Delivery volume(ml)

Delivery volume setup range

[SETUP DELIVERY VOLUME]

2. Delivery volume range is from 0.1 to 1000.0ml with 0.1ml increasement.

3. Press [ESC/2s CLR] button for 2seconds to delete complete value, press button to delete previous

digit only.

NOTE • If delivery volume is set with 0ml, it is considered as maximum value, 1000ml and
operates delivery volume with 1000ml.
(When remaining time is exceeded 99hours, infusion remaining time is not
displayed.)
• During infusion state, delivery setup is not available. Setup is available only “STAND
BY” mode.

3-5. CONFIRM TOTAL INFUSED VOLUME (Σ㎖/CLEAR)


• Total volume(Σ㎖/CLEAR) shows the delivered volume so far.
• Press TOTAL VOL button.
• Delete total infused volume by press [ESC / 2sCLR] button for 2seconds.

State of menu
Previous value

Total infused volume(ml)

[TOTAL INFUSED VOLUME]

17 SP-8800 / UME-04-REV.12
3. OPERATION

3-6. SETUP DELIVERY TIME(hour)


• Press [SEL] button, DELIVERY TIME on the bottom of LCD is blinked. Set up delivery time with numerical
buttons and save it with [SEL] button.

NOTE • Third value is calculated and displayed automatically when two setup values are
entered among FLOW RATE, DELIVERY VOLUME, DELIVERY TIME.
• Please make sure about total infused volume when you setup infusion by time,
previous total infused volume is included if it is not deleted.
• DELIVERY TIME range: 1min~99hour 59min

3-7. INSERT THE NEEDLE INTO THE PATIENT


• Insert the needle into the patient.

• Alarm does not go off when needle comes off the vein, resulting in an out-of-the-vein
infusion. Therefore, regularly check the puncture site and monitor the patient’s
condition carefully.

3-8. START INFUSION

NOTE • Before operating the pump, make sure to check the flow rate, delivery volume and
the condition of infusion tube again.

1. Press [START] button to start infusion.

2. During infusion, is blinked in order. “INFUSING” and DELIVERY TIME blinked by


turns on the bottom line.

[INFUSING DISPLAY]
18 SP-8800 / UME-04-REV.12
3. OPERATION

3-9. INFUSION COMPLETION


1. When delivery volume is reached, flow rate turns into K.V.O. rate to avoid occlusion of infusion by coagulation.
2. The unit turns into K.V.O. rate automatically, and LCD shows “CAUTION FINISH K.V.O” as below picture with
alarm sound.

[K.V.O. RATE]
3. Press [STOP] button to stop K.V.O. rate. LCD display turns into initial display.

• K.O.V. rate is adjustable, please make sure K.V.O. rate before start infusion.
Otherwise over infusion may be caused.
• Over infusion may be caused by user’s setup condition, manufacturer does not take
any responsibility on the problem.

4. When drug in the syringe is left about 3~4cc, “NEAR EMPTY” alarm goes off with alarm sound
depends on setup flow rate. OCC alarm indicator is blinked in red.

[NEAR EMPTY]
5. When the syringe is empty, “ WARNING SYRINGE OCCLUSION” alarm goes off with alarm sound and
OCC alarm indicator is on red.

[SYRINGE OCCLUSION]
6. Press [STOP] button to stop. LCD display turns into initial display.
19 SP-8800 / UME-04-REV.12
3. OPERATION

3-10. STOP INFUSION


• Press [STOP] button to stop infusion during the operation.
• Before restart infusion, FLOW RATE, DELIVERY VOLUME and infusion tube should be confirmed. After
confirming the condition, press [START] button to start infusion.

NOTE • Allam goes off if do not restart in 2minutes after stop infusion temporally.

3-11. [POWER] BUTTON


• Power is turned on and off by press [POWER] button.
• During infusion, [POWER] button does not function.
• Press [STOP] button first to stop infusion, then press [POWER] button to turn off the unit.

NOTE • Reset Total volume as “0” when restart infusion with new infusion condition.

3-12. BATTERY REMAIN


• Press button in initial mode, display turns into number ③ picture and display the remaining battery
operating time.
① ②

Remaining battery
operating time

• Remaining battery operating time is different as its infusion setup condition and
battery condition.
• Check about battery condition and operating condition with you refer to “BATTERY
REMAIN”.
• Charging LED on the front should be green when you refer to “BATTERY REMAIN”.

20 SP-8800 / UME-04-REV.12
3. OPERATION

작동 순서

3-13. ERROR ALARM


• When alarm goes off with error, press [SILENCE] button to hold alarm sound for 2seconds then settle the
problem. Otherwise, alarm sound goes off again.

3-14. DC INPUT OPERATION


• This unit can be operated with DC input. DC 24V 1A, DV 12V 1A.
• Connect the DC INPUT code (Cod No.: 05LN51﹡) to the DC INPUT PORT and press [POWER] bottom.
• “ “ comes up in right side of LCD.

NOTE • DC input cable length is up to 3m.

3-15. BATTERY OPERATION


• This unit operates with AC power supply and built-in rechargeable battery. When AC power supply is
disconnected, the unit turns into battery operating mode automatically.
• Battery is charged when AC power supply is connected when ever it is on or off.
• Battery operating time is about 4 hours.
• New rechargeable battery should be charged more than 6hours.

• Battery state displays on LCD as its remain state.


• With battery state, if there is no AC power supply connected, the unit will be off with alarm sound and
stop operation.

• Battery state should be checked every 6month to inspect battery life.


• Battery capacity may be different depends on it using condition such as fully
charging and fully discharge cycle. For safety of longer battery operating time,
battery should be fully charged one a month.
• If the unit has been kept without using it for more than 1months, or it is first time to
use the unit, try to charge the built-in battery for more than 6hours.

3-16. NURSE CALL FUNCTION (OPTION)


• Connect nurse call connector on the back side of the unit and NURSE CALL SYSTEM built in a hospital with
nurse call cord.

21 SP-8800 / UME-04-REV.12
3. OPERATION

작동 순서
NOTE • When alarm goes off, error message shows up in LCD, alarm signals are transferred
to NURSE CALL SYSTEM every 5 second.
• Nurse call code number: 06LA548﹡
• Nurse call cable length is up to 3m

3-17. KEYPAD LOCK


• Lock keypad by pressing MENU/2s button.
• Keypad lock can be done with password so user can control keypad lock.
• Under Keypad LOCK, all key do not function.
NOTE • Refer to page 28 for setup KEYPAD LOCK.

22 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

4-1. Buzzer level setup (Short-cut key)


4-2. Occlusion level setup (Short-cut key)
4-3. Syringe selection
4-4. Infusion setup
4-5. Display
4-6. Config.

23 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

작동 순서

4. SYSTEM SETUP
• Press MENU button to get into Menu mode and select “SYSTEM SETUP” with [SEL] button.
• SYSTEM SETUP MENU
1. BUZZER
2. OCCLUSION
3. BRAND SET
4. INFUSION SET
Menu move buttons Total menu number
5. DISPLAY
6. CONFIG
• It you do not press any button for 10seconds, LCD turns back to previous display with alarm sound.
• Press [ESC / 2sCLR] button to go back previous display. And press [SEL] button to save.

4-1. BUZZ. LEVEL SETUP (SHORT-CUT KEY)


• Get into “BUZZ. LEVEL” mode, select menu in order to MENU➔SYSTEM SETUP➔BUZZER.

•. Adjust buzzer level with UP , DOWN buttons. There are three levels, “HIGH” , “MIDDLE” and “LOW”.
Select and save with [SEL] button. After save buzzer level display turns into previous mode.

Initial BUZZ. LEVEL HIGH

※ Short-cut key for buzzer level: Press + [1] button in initial display.

24 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

4-2. OCC. LEVEL SETUP (SHORT-CUT KEY)SPECIAL FUNCTION


• Get into ”OCC. LEVEL” mode, select menu in order to MENU➔SYSTEM SETUP➔OCCLUSION

• OCCLUSION is adjusted with UP , DOWN buttons, it is from LEVEL 1 to LEVEL 9. Press [SEL] button
to select and save. After select and save it, it turns into previous menu.

NOTE • “H”: 800±200mmHg (106.7±26.7 kPa) or 1.09±0.27 kgf/㎠


• “M”: 500±100mmHg (66.7±13.3 kPa) or 0.68±0.14 kgf/㎠
• “L”: 300±100mmHg (40.0±13.3 kPa) or 0.41±0.14 kgf/㎠
• Initial set is LEVEL 5

※ Short-cut key for occlusion level: Press + [2] button in initial display.

4-3. SYRINGE SELECTION


• Get into “BRAND SET” menu in order to MENU➔SYSTEM SETUP➔BRAND SET.

• Syringe should be calibrated by supplier, select the syringe brand will use with UP , DOWN button.
Select and save the syringe brand with [SEL] button. After select and save it, it turns into previous menu.

25 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

SPECIAL FUNCTION
• Here is the list of basic calibration syringe brands.
(Calibrated syringe brand may be different depends on local authorized dealer.)

NO Brand Manufacturer
1 B-D Beckton
2 MEDI Medihut
3 KOVA Korean Vaccine
4 SHIN Shinchang
5 HWAJ Hwajin
6 SUNG Sungshin

NOTE • Initial setup is “B-D”.


• If you want to use other syringe brand beside the list, please contact our local
authorization dealer to get confirm about calibration.
바랍니다.
• Un calibrated syringe may occur error or accuracy problem. Please make sure the
syringe brand you are using is calibrated and approved by local authorized dealer or
manufacturer.
• Manufacturer or local authorized dealer do not cover the result from using syringe
brand which is not calibrated or approved by manufacturer or local authorized dealer.

※ Short-cut key for syringe brand: Press + [3] button in initial display.

4-4. INFUSION SET


There are “PURGE SETUP” , “BOLUS SETUP” , “KVO SETUP” in “INFUSION SET” menu.

26 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

SPECIAL FUNCTION
4-4-1. PURGE SET UP
• PURGE is setup from 0.1~1200ml/h.

Previous setup

PURGE (㎖/h)

Range

• If PURGE rate is not setup, beep sound will come out and infusion will not start.

NOTE • Purge: To remove air in tube(line).


• Purge volume is included in total infused volume.
• Purge is working only “STAND BY” mode.

• When Purge rate is less then flow rate, Purge rate works as setup flow rate.

4-4-2. BOLUS SET


There are “BOLUS FLOWRATE” , “BOLUS D.VOLUME” for “BOLUS SET” menu.
※ BOLUS: Infuse sure amount of solution or drug by sure infusion speed.

- BOLUS FLOWRATE SET : Setup flow rate of BOLUS with numeral buttons. Save desired value with [SEL]

button. Press [ESC / 2sCLR] buttons for 2seconds to delete the total value and press to delete previous
digit in order.
※ Press [ESC / 2sCLR] button to move previous menu.

Previous setup

BOLUS F/R (㎖/h)


BOLUS VOL(㎖)
Range

27 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

SPECIAL FUNCTION
- BOLUS VOLUME : Setup total amount of BOLUS. Save desired value with [SEL] button. Press [ESC /

2sCLR] buttons for 2seconds to delete the total value and press to delete previous digit in order.

NOTE • BOLUS ON/OFF: Get into the menu in order to MENU➔SYSTEM


SETUP➔6.CONFLG➔BOLUS ON/OFF. You can select BOLUS ON/OFF. (Refer to
page 29 for setup). (Initial setup is BOLUS OFF mode.)
• BOLUS and PURGE button is same one. It functions differently as its mode as blow.
※ BOLUS function: Only during the infusion
※ PURGE function : Only STANDBY mode
• Please make about BOLUS ON/OFF. When you do not use BOLUS function, please
setup BOLUS OFF so it does not function. And make sure when you use BOLUS
function about its setup condition (flow rate, delivery volume) before use it.

4-4-3. K.V.O. SETUP


Setup K.V.O. (Keep Vein Open) rate from 0.1~10.0㎖/h.
※ K.V.O. (Keep Vein Open): When the delivery volume has been reached, flow rate turns into K.V.O. rate
(adjustable from 0.1ml/h to 10.0ml/h) automatically to avoid clog by blood coagulation.

• Make sure K.V.O rate setup before use it to avoid excess infusion during K.V.O.
function.
• Manufacturer does not cover the result from K.V.O. setup failure by user. So please
make sure about K.V.O. setup condition before start infusion.

28 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP

4-5. DISPLAY SPECIAL FUNCTION


There are “TIME SETUP” , “B.LIGHT TIME” , “B.LIGHT BRIGHT” for “DISPLAY” menu.

4-5-1. TIME SETUP


• Get into “TIME SET” menu in order to MENU➔SYSTEM SETUP➔DISPLAY➔TIME SETUP.

• Move with , , , buttons and input with numerical buttons for time setup. Press [SEL] button to
select and save time set.

• There is no extra delete button because you can move with , , , buttons at any point.

NOTE • Once time setup, the unit keeps time whenever it turns on or off.

4-5-2. LCD BACK LIGHT TIME

• Get into “B.LIGHT TIME” menu in order to MENU➔SYSTEM SETUP➔DISPLAY➔B.LIGHT TIME.

• There are 4modes for LCD black light mode.


-ON : Always LCD black light on.
-30SEC : Turns off after 30seconds.
- 60SEC : Turns off after 60seconds.
-90SEC : Turns off after 90seconds.

Move with , buttons and press [SEL] button to select and save the mode. After select and save it, it
turns into previous menu.

NOTE Initial set is ON.

29 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP
4-5-3. LCD BLACK LIGHT BRIGHT SPECIAL FUNCTION
• Get into “BLIGHT BRIGHT” menu in order to MENU➔SYSTEM SETUP➔DISPLAY➔B.LIGHT BRIGHT.

• There are brightness levels from LEVEL 1 to LEVEL 9. As level number is up, LCD goes brighter. Adjust it with

, buttons and press [SEL] button to select and save it.

NOTE Initial set is LEVEL 9.

4-6. CONFIG
There are “KEY LOCK P/W” , “BOLUS ON/OFF” and “ANTI-BOLUS” for CONFIG menu.

Problems from incorrect setup by user, are charged on user. It does not cover by
manufacturer.

4-6-1. KEY LOCK P/W


• Get into KEYLOCK P/W menu in order to MENU➔SYSTEM SETUP➔CONFIG➔KEYLOCK P/W.

• Move with , buttons, select and save with [SEL] button.

30 SP-8800 / UME-04-REV.12
4. SYSTEM SETUP
4-6-2. BOLUS ON/OFF SPECIAL FUNCTION
• Get into BOLUS INFUSE menu in order to MENU➔SYSTEM SETUP➔CONFIG➔BOLUS ON/OFF.

• Move with , buttons, select and save with [SEL] button.


NOTE Initial setup is OFF. (KEY LOCK P/W, BOLUS INFUSE)

4-6-3. ANTI BOLUS ON/OFF


• Get into ANTI BOLUS INFUSE menu in order to MENU➔SYSTEM SETUP➔CONFIG➔ANTI – BOLUS.
※ ANTI BOLUS: Reduce significantly bolus after occlusion release. O

• Move with , buttons, select and save with [SEL] button.

NOTE Initial setup is OFF. [KEY LOCK P/W] , [BOLUS INFUSE] , [ANTI BOLUS]

4-6-4. KEY SOUND


Get into “KEY SOUND” menu in order to MENU➔SYSTEM SETUP➔CONFIG➔KEY SOUND.

• Move with , buttons, select and save with [SEL] button.

NOTE • if “KEY SOUND” is OFF, there is no sound when the buttons are pressed
• Preset is ON for KEY SOUND

31 SP-8800 / UME-04-REV.12
5. SPECIAL
FUNCTIONS

5-1.Dosage
5-2.History
5-3.Profile (Option spec)
5-4.Drug label
5-5.Unit setup
5-6.Pro. mode

32 SP-8800 / UME-02-REV10
5. SPECIAL FUNCTION
4. SYSTEM SETUP
5. SEPCIAL FUNCTION SPECIAL FUNCTION
• Press MENU button to get into Menu mode and select SPEICAL FUNCTION with [SEL] button.
• SPEICAL FUNCTION MENU
1. DOSAGE
2. HISTORY
3. PROFILE (OPTION SPEC)
4. DRUG.LABEL (OPTION SPEC)
5. UNIT SETUP
6. PRO.MODE

NOTE • Contact local authorized dealer for OPTION SEPC.

5-1.DOSAGE

• DOSAGE SETUP
• Press buttons in order to MENU➔SPECIAL FUNCTION➔DOSAGE and there come up “DOSE RATE” on the
LCD.
• DOSEGE: Automatically calculate and set the proper flow rate when the dose rate, body weight, drug mass and
solution volume are entered.
• After input each parameter, press [SEL] button. Each mode is confirmed and turns into next parameter
automatically. Please refer to picture No. ① to ④.
• After setup complete parameter, it turns into picture No. ⑤ automatically and display complete information for
each parameter value.

• Each parameter, you can delete or change the value with [ESC / 2sCLR] or , , , buttons.
• Press [ESC / 2sCLR] button to back previous mode.

① ② ③ ④

• As soon as finish input each parameter value, complete data is displayed as picture No. ⑤. Select YES to
confirm the data for DOSAGE mode, it turns into STANDBY mode with confirm sound. Flow rate and Delivery
volume are displayed on FND automatically. Press [START] button to start. If you select NO on the picture No.
33 SP-8800 / UME-04-REV.12
5. SPECIAL FUNCTION
⑤, it goes back DOSE RATE mode automatically to setup each parameter again. 4. SYSTEM SETUP
⑤ SPECIAL FUNCTION

* Range for each parameter

Dose rate 0.01 ~ 99.99㎍/㎏/min. (0.01㎍/㎏/min. increment)


Body weight 0.1 ~ 300.0㎏ (0.1㎏ increment)
Solution volume 0.1 ~ 999.9㎖ (0.1㎖ increment)
Drug volume 0.1 ~ 999.9㎎ (0.1㎎ increment)

* Change unit for DOSE RATE : Unit can be changed page 36 “UNIT SET”

1. ㎍/㎏/min
2. ㎍/㎏/hour
DOSE RATE
3. ㎎/㎏/hour
4. ㎎/㎏/min

NOTE • Each parameter value should be more than “0”.


• DOSAGE Mode
Flow rate (ml/h) = [{Dose rate(㎍/kg/min.) * Body weight(kg) * Solution volume(ml)}
/ {Drug volume(mg) * 1000}]* 60

5-2. HISTORY
5-2-1. Check HISTORY
• Get into HISTORY VIEW menu in order to MENU➔SPECIAL FUNCTION➔HISTORY.

• HISTORY : Display last 10 infusion data, infusion starting time, infusion completing time, FLOW RATE ,
34 SP-8800 / UME-04-REV.12
5. SPECIAL FUNCTION
INFUSED VOLUME. 4. SYSTEM SETUP

• Move with , , ,
SPECIAL FUNCTION
buttons and press [SEL] button to see detail information for each data.
• HISTORY DATA: Last 10 infusion data only.

5-3. PROFILE (OPTION SPEC)

• Get into PROFILE SET menu in order to MENU➔SPECIAL FUNCTION➔PROFILE.

5-3-1. PROFILE LOAD

• PROFILE LOAD : Call back the previous setup data. When you select this mode, there is YES or NO selection
mode to start same profile.
• Select “YES” to start same profile with previous setup. The unit turns into STAND BY mode. Press [START]
button to start PROFILE, the unit start PROFILE mode with displaying “PROFILE” on LCD.
• Select “NO” to setup new PROFILE.

NOTE • During PROFILE function, you can not setup Flow rate, Delivery volume. To stop
PROFILE function, make sure PROFILE LOAD setup as “ NO”.

5-3-2. PROFILE SETUP

• PROFILE: Program infusion condition (flow rate, delivery volume, delivery time) differently for every hour, up to
24hours. (Option).
• Select “PROFILE SET UP” and press [SEL] button to get into below picture mode. Press [SEL] button to setup
time for each profile, input each time with numerical buttons. Press [SEL] button to save each time and infusion
setup (Flow rate, Delivery volume).
• Setup profile data up to 24hours in order to same process.
• Press [START] button to save.
35 SP-8800 / UME-04-REV.12
5. SPECIAL FUNCTION
4. SYSTEM SETUP

SPECIAL FUNCTION

NOTE • Time setup increment is hour only (not by minute).


• Maximum profile time is 24hour and 24 kinds of infusion setup available.
• Press up is “OFF”.
• Problem occurred by user setup is charged on user. Manufacturer does not cover
the problem occurred by incorrect setup or use by user.

5-4. DRUG LABEL (OPTION SPEC)


This unit does not include “DRUG LABEL” function.

5-5. UNIT SETUP


• Get into UNIT SET menu in order to MENU➔SPECIAL FUNCTION➔UNIT SETUP.
• This is changing the unit on DOSAGE RATE mode.

• Move with , , , buttons to select the unit and press [SEL] button to save it.

NOTE
• Initial setup is “㎍/㎏/min”.

5-6.PRO.MODE
Contact local agency or local authorized dealer for this mode.

36 SP-8800 / UME-04-REV.12
6. TROUBLE SHOOTING

37 SP-8800 / UME-02-REV10
6.TROUBLE SHOOTING - ALARM
Take the following actions if any trouble occurs. When the troubles could not be solved with the following
actions, please contact your local authorized dealer immediately.

NOTE • Whenever alarm sounds, the pump stops infusion. That is, alarm sounds only in error
situation during infusion.

Symptom Cause Action

Pump cannot be • AC or DC power cord is not ▶Check the AC or DC power cord connection.

switched on. inserted properly • Never connect both AC and DC power to the pump

at the same time

• Internal battery has deteriorated ▶Stop the operation of the pump and replace with a new battery through

your local authorized dealer.

• The voltage of the internal battery ▶Recharge the battery fully for more than 24 hours by connecting the

is low pump to an AC power outlet.

• Because of an industrial waste, used

Ni-MH battery must be returned to

AMPall or distributor.

[OCCLUSION] LED • The infusion line is kinked or ▶1. Turn the alarm off by pressing [SILENCE] button.

flashes and alarm twisted ▶2. Check the infusion line and take a corrective action like untwisting or

sounds. replacing with a new one to solve the problem of occlusion

▶3. Press [STOP] button to show “STAND-BY” on the LCD.

▶4. Restart infusion by pressing [START] button.

• When infusion is restarted, the ▶The system is in trouble.

alarm sounds again, and the ▶Turn the alarm off by pressing [SILENCE] button.

pump stops ▶Please contact your local authorized dealer.

[FINISH] LCD display • The syringe is completely empty ▶1. Turn the alarm off by pressing [SILENCE] button.

and alarm sounds ▶2. To continue the infusion, replace the syringe with a new one.

▶3. Press [STOP] button to show “STAND-BY” on the LCD.

▶4. Restart infusion by pressing [START] button.

[NEAR EMPTY] LED • The syringe is nearly empty ▶1. Press [STOP] button to stop infusion.

flashes and alarm (Finishing injection in 5 minutes ▶2. To continue the infusion, replace with the new syringe filled with

sounds. before, the flash light and solution.

warning sound occurring) ▶3. Restart infusion by pressing [START] button.

38 SP-8800 / UME-04-REV.12
6.TROUBLE SHOOTING - ALARM

Symptom Cause Action

[SLIDER HOOK] LED • Slider Hook does not fix Push- ▶1. Turn the alarm off by pressing [SILENCE] button.

flashes and alarm Button of syringe ▶2. Make Slider Hook fix Push-Button of syringe.

sounds ▶3. Press [STOP] button to show “STAND-BY” on the LCD.

▶4. If re-infusion is needed, restart infusion by pressing [START]

button.

[CLAMP] LED flashes • Clamp does not fix Barrel of syringe ▶1. Turn the alarm off by pressing [SILENCE] button.

and alarm sounds ▶2. Pull the clamp, turn it to the left up to 90° and put the finger

grips of syringe into the slit.

▶3. Press [STOP] button to show “STAND-BY” on the LCD.

▶4. If re-infusion is needed, restart infusion by pressing [START]

button.

[CLUTCH] LED flashes • Clutch is not properly engaged ▶1. Turn the alarm off by pressing [SILENCE] button.

and alarm sounds ▶2. Make the clutch engaged properly.

▶3. Press [STOP] button to show “STAND-BY” on the LCD.

▶4. If re-infusion is needed, restart infusion by pressing [START]

button.

“BATT-LOW” on the • The voltage of internal battery is low ▶1. Press [STOP] button to stop infusion.

LCD and alarm sounds ▶2. Recharge the battery fully for 24 hours by connecting the

pump to an AC power outlet.

▶3. Restart infusion by pressing [START] button.

“KEEP-VEIN” on the • In the operation, the total volume ▶ Stop K.V.O. rate by pressing [STOP] button

LCD and alarm sounds delivered reaches the delivery limit

• The total volume delivered exceeds ▶ Stop K.V.O. rate by pressing [STOP] button

the delivery limit ▶ When infusion is newly needed, set the delivery volume again or

clear the total volume delivered.

When [PURGE] button • The syringe is not set properly ▶ Set the syringe correctly.

is pressed, the alarm • Even when the syringe is set again, ▶ The system is in trouble.

sounds. purging does not occur Please contact your local authorized dealer

“ERROR” is on the LCD • The system is in trouble ▶ Please contact your local authorized dealer

39 SP-8800 / UME-04-REV.12
6.TROUBLE SHOOTING - ALARM

Symptom Cause Action

Even when [START] • The value of “flow rate, delivery ▶ Confirm the values that were set.

button is pressed, the limit, dose rate, body weight, drug

alarm sounds, and the volume, or solution volume” is not

pump does not start. set correctly

• Clutch is not properly engaged ▶ Engage the clutch correctly.

• The syringe is not set properly ▶ Set the syringe correctly.

• The low voltage of the internal ▶ Recharge the battery fully for more than 24 hours by connecting

battery the pump to an AC power outlet.

[SYRINGE SIZE] LED • The syringe, which is used, may be ▶ Make sure that the syringe, which is used, is the same as the

and [CLAMP] LED flash different from the one specified in one specified in the pump in terms of manufacturer.

the pump in terms of manufacturer To use the syringe of different manufacturer, please contact your

• When infusion is restarted, the local authorized dealer.

alarm sounds again, and the pump ▶ The system is in trouble.

stops ▶1. Turn the alarm off by pressing [SIL ENCE] button.

▶2. Turn off the power.

▶3. Please contact your local authorized dealer for repair

The accuracy of flow • The syringe, which is used, may be ▶ Make sure that the syringe, which is used, is the same as the

rate is questionable. different from the one specified in one specified in the pump in terms of manufacturer.

the pump in terms of manufacturer To use the syringe of different manufacturer, please contact your

local authorized dealer.

• Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.

to be in trouble

The operation hour by • Internal battery has deteriorated ▶ Stop the operation of the pump and replace with a new battery

battery is too short after through your local authorized dealer.

recharging the battery • Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.

to be in trouble

Battery is not recharged • AC power cord is not inserted ▶ Check the AC power cord connection.

even when it is properly

connected to AC power • AC power cord has deteriorated ▶ Replace with a new AC power cord.

outlet. • Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.

to be in trouble

Battery is not recharged • DC power cord is not inserted ▶ Check the DC power cord connection.

even when it is properly

connected to DC power • DC power cord has deteriorated ▶ Replace with a new DC power cord.

outlet. • Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.

to be in trouble

40 SP-8800 / UME-04-REV.12
SPECIFICATION

DELIVERY PERFORMANCE

Syringe 10ml: 0~300ml/h

Syringe 20ml: 0~400ml/h


FLOW RATE RANGE
Syringe 30ml: 0~500ml/h

Syringe 50ml: 0~1200ml/h

VOLUME LIMIT 0.1~1000ml ( 0.1ml steps)

PURGE RATE 0.1~1200ml/h (Adjustable)

K.V.O. RATE 0.1~10ml/h ( Adjustable)

Mechanical accuracy: ±2%


ACCURACY WITH APPROVED SYRINGES
Accuracy including syringe: ±2%

ALARM
OCCLUSION, NEAR EMPTY, LOW BATTERY, K.V.O. RATE, CLUTCH OPEN, SLIDER HOOK OPEN,

CLAMP OPEN, REPEAT ALARM, STAND BY

FEATURES
CLASSIFICATIONS ClassⅠ, CF, IPX3

OCCLUSION DETECTION PRESSURE 1.3kgf/㎠

NURSE CALL DC12V, 1A, The Cable length is up to 3m

KEYPAD LOCK,SHORT-CUT KEY, HISTORY CALL-BACK, RETAIN MEMORY, REMAINNING TIME,

OPEN SYSTEM, PURGE, BOLUS, K.V.O., BUZZER VOLUME / OCCLUSION PRESSURE CONTROL,

DOSAGE MODE, NURSE CALL(OPTION), DRUG LABEL(OPTION), PROFILE(OPRION),

CENTRALL SYSTEM(OPTION)

OTHER PARAMETERS
POWER REQUIREMENTS / CONSUMPTION AC100 / 240V, 50 / 60Hz or DC9.7V, DC12V 1A / 45VA

The DC Input cable length is up to 3m.

DIMENSIONS (W × D × H) 220×130×108 (mm)

WEIGHT Approximately 2kg

BATTERY / OPERATION / CHARGING Ni-MH9.7V / 4 hours / 6 hours

BATTERY LIFE 1.5~2 years

OPERATION CONDITIONS 5~40℃, 30~90% RH (no condensation)

STORAGE CONDITIONS -20~45℃, 10~95% RH (no condensation)

WARRANTY PERIOD 1 year

Expected Service Life 5 years

※Specifications and design are subject to change for improvement without prior notice

41 SP-8800 / UME-02-REV10
SYMBOL

Meaning Symbol
Description Symbol

CE Marking of conformity

Model Number.

Type CF applied part

Classification Class I

Medical Device

The product should be recycled


separately from household waste.
When this product reaches its end of
life, follow the local laws and
regulations of disposal. The improper
disposal of waste electronic
equipment from the consumer may be
subject to fines.

Rated power input, A.C.

Rated power input, D.C.

Battery, general

42 SP-8800 / UME-04-REV.12
Manufacture information

Manufacturing Date and year

Serial Number

Manufacturer ’ s EU representative
information

Refer to instructions for use

Keep dry

Fragile

This way up

Use no hooks

To indicate that transport package


shall not be exposed to sunlight.

43 SP-8800 / UME-04-REV.12
Indicates the temperature limits to
which the medical device can be
safely exposed.

Indicates the range of humidity to


which the medical device can be
safely exposed.

Indicates the range of atmospheric


pressure to which the medical device
can be safely exposed.

44 SP-8800 / UME-04-REV.12
AMPall warrants that the product shall be free from defects in workmanship
and materials for the warranty period. The warranty is void if failure of the
product has resulted from accident, alteration, abuse or misapplication to
include unauthorized opening of the product.

Under the warranty, AMPall shall repair or replace at its option and expense
any products found defective by the customer during the warranty period
and returned to an authorized AMPall representative.
Return of the repaired or replaced product to the customer’s original location
shall be at the expense of AMPall unless AMPall determines that the
products are not defective within the terms of the warranty.
In such case the customer shall pay AMPall all costs of handling,
transportation and labor.

This warranty is in lieu of all other warranties whether expressed or implied.

• Manufacturer : AMPall Co.,Ltd.


3F, Annex Hankook Junja Hyeopdong B/D (Gasan-Dong)
114 Gasan digital 2-ro, Geumcheon-gu, Seoul, 08506, Korea
TEL: 82-2-858-2839~41
FAX: 82-2-858-2442
E-mail: ampall@ampall.com

• EC Representative : AR Experts B.V.


Amerlandseweg 7
3621 ZC Breukelen, The Netherlands
TEL: +31 (0)88 995 1333
E-mail: info@ar-experts.eu

45 SP-8800 / UME-04-REV.12
Warranty Card

Thank you for your purchasing for AMPall product. The warranty period
Of SP-8800 is valid from the purchasing date. Please notice the purchasing
date on the warranty card and keep this Warranty Card with the device
all the time to get full customer service.
Product Name Infusion Pump

Model Name □ SP-8800

Manufacturing Date

Warranty Period 1 years

Purchasing Date / / (year/ month/ date)

Name: Tel.:
Customer
Add.:

Name: Tel.:
Retailer
Add.:

Repairing Record

Date Contents of Repair Confirmation

☆ Please show this Warranty Card when you request repairing service.

3F. Annex Hankook Junja Hyeopdong B/D,


114, Gasan digital 2-ro, Geumcheon-gu, Seoul, Korea.
TEL : +82 - 2 - 858 - 2839~41 / FAX : +82 - 2 - 858 – 2442
E-mail : ampall@ampall.com / Website: www.ampall.com

46 SP-8800 / UME-04-REV.12