Specification
Specimen Whole blood, Serum, Plazma, Dialysate
Test available (measured) 1810 1820 1830 1815 1825 1835
Item Measurement range
1810aqc 1820aqc 1830aqc 1815aqc 1825aqc 1835aqc
pH 6.000 - 8.000
pCO2 10.0 - 200.0 Torr
pO2 10.0 - 760.0 Torr
cNa+
cK+
80.0 - 200.0 mmol/L
1.00 - 20.00 mmol/L Blood Gas System
cCl- 50.0 - 200.0 mmol/L
cCa2+ 0.5 - 5.0 mmol/L
GASTAT-1800 Series
cGlucose 20.0 - 1200.0 mg/dL
cLactate 0.4 - 30.0 mmol/L
cHb 2.0 - 23.0 g/dL
ctHb 2.0 - 23.0 g/dL
sO2 0.0 - 100.0 %
FO2Hb 0.0 - 100.0 %
FCOHb 0.0 - 100.0 %
FMetHb 0.0 - 100.0 %
FHHb 0.0 - 100.0 %
Test available (calculated) cHCO3 act
-
cHCO3 std
Made in Japan product guarantees platinum quality
-
cBE(ecf)
cBE(B)
cBB
ctCO2(P)
ctCO2(B)
sO2(est)
BO2
p50
p50(T)
p50(st)
pO2/FiO2
pO2(A,T)
pO2(A)
pO2(A-a)
pO2(A-a,T)
pO2(a/A)
pO2(a/A,T)
RI
RI(T)
ctO2(a)
ctO2(B)
ctO2(v)
ctO2(a-v)
ctO2([a-v]/a)
Hct
AG
AG(K)
cCa2+(7.4)
Sample volume Model Syringe Capillary
GASTAT-1810/1810aqc 120 µL 60 µL
GASTAT-1820/1820aqc 120 µL 80 µL
GASTAT-1830/1830aqc 160 µL 100 µL
GASTAT-1815/1815aqc 155 µL 60 µL
GASTAT-1825/1825aqc 155 µL 80 µL
GASTAT-1835/1835aqc 195 µL 100 µL
30 µL (Gas, pH/Electolyte parameter)
Micro mode (All model)
NOTE: Metabolic and OXY parameter cannot be measured by Micro mode
Analysis time GASTAT-1810/1820(aqc) About 50 seconds
(from press measurement button to result appears GASTAT-1830(aqc) About 55 seconds
on the display) GASTAT-1815/1825/1835(aqc) About 75 seconds
Cycle time GASTAT-1810/1820(aqc) About 105 seconds
(measurement sample + flush cleaning) GASTAT-1830(aqc) About 170 seconds
GASTAT-1815/1825 About 155 seconds
GASTAT-1835(aqc) About 190 seconds
Sample aspiration time GASTAT-1810/1820(aqc) Within 4 seconds
GASTAT-1830/1815/1825/1835(aqc) Within 5 seconds
Display 12 inch full color display/touch panel
Entered parameter
Calibration
Patient ID, Patient temperature, Patient Hb, Barometric pressure, FiO2, Height, Weight, Sex, Age
Automatic, programmable 1 and 2 point calibration
12 models cover the various requirements of critical care sites
Reagent, waste volume detection Automatic intergrade detection method
Data storage 10,000 samples
Printer Line thermal
Interface TCP/IP
1810 1820 1830 1815 1825 1835 1810 1820 1830 1815 1825 1835
aqc aqc aqc aqc aqc aqc
Barcode reader Built-in (12 kinds of barcode are available)
Operating temperature 10 - 30 °C
Humidity 5 - 80 % pH, pCO2
Power 85 - 264V 50/60Hz 150W, Built-in Battery pO2, ctHb
Dimensions 330(W) - 240(D) - 570(H) mm
Weight 18.5 kg cNa+, cK+
Specifications subject to change without notice cCl–,cCa2+
cGlucose
Manufacturer Distributor cLactate
CO-Oximetry
5-5-1 Nakamachidai, Tsuzuki-ku, Yokohama, Kanagawa, 224-0041 Japan
Phone +81-45-948-1961 F a x +81-45-948-1962 AutoQC
E-mail overseas@technomedica.co.jp
W e b http://technomedica.co.jp
VER1 Published December 2014
Blood Gas System
GASTAT-1800 Series
Features
Short analysis time: 50 seconds *
*Model GASTAT-1820
Small sample volume:
80 µL for capillary and 120 µL for syringe *
*Model GASTAT-1820
Acid-base graph available
Ultrafast Sample Aspiration Touch-Sensitive Panel
Throw-in-type printer realizes easy paper roll 4 seconds’ sample aspiration time realizes short The 12 inch color touch panel conveniently displays
Turn-Around-Time. * easy-to follow guidance for every operation.
replacement *Model GASTAT-1820
Save mode reduces reagent consumption for
users having less samples in the midnight
Automatic Quality Control conveniently
guarantees accuracy of test results
Built-in PC functionality for intercommunicate
without using external PC
Built-in battery enables 30 minutes stable
operation or 5 samples analysis under power
failure Built-in Barcode Reader Injection Detect Function
Built-in barcode reader reduces errors and save time The system starts automatic cleaning when sample is
for users. injected.
Easy 2 steps operation
STEP 1 STEP 2 Zero-maintenance Electrodes Minimal Reagents Replacement Required
Mount either syringe or capillary on the Select sample type and then press Test results are reported by both display The stable high-precision performance electrodes Only two kinds of reagents required (CAL Cartridge
sample inlet. Analysis icon. and paper print out. enable absolutely maintenance free. and FLUSH SOLUTION)