Intermit Suction
Intermit Suction
L
A BOC Health Care
Company
Critical Care
Worldwide
User Responsibility
This Product will perform in conformity with the without the prior written approval of Ohmeda’s
description thereof contained in this manual Safety Department. The user of this Product
and accompanying labels and/or inserts, when shall have the sole responsibility for any mal-
assembled, operated, maintained and repaired function which results from improper use, faulty
in accordance with the instructions provided. maintenance, improper repair, damage, or
This Product must be checked periodically. A alterations by anyone other than Ohmeda.
malfunctioning Product should not be used.
Important: Federal law in the U.S.A. and
Parts that are broken, missing, plainly worn,
Canada restricts this device to sale by or on the
destroyed or contaminated, should be replaced
immediately. Should such repair or replacement order of a licensed medical practitioner.
become necessary, Ohmeda recommends that a Important: This document is not to be repro-
telephonic or written request for service advice duced in any manner, nor are the contents
be made to the nearest Ohmeda Service Office. herein to be disclosed to anyone without the
This Product or any of its parts should not be express authorization of Ohmeda, Columbia
repaired other than in accordance with written MD, USA.
instructions provided by Ohmeda, or altered
Technical Competence
The procedures described in this service man- Genuine replacement parts manufactured or
ual should be performed by trained and author- sold by Ohmeda must be used for all repairs.
ized personnel only. Maintenance should only
be undertaken by competent individuals who Read completely through each step in every
procedure before starting the procedure; any
have a general knowledge of and experience
exceptions may result in a failure to properly
with devices of this nature. No repairs should
ever be undertaken or attempted by anyone not and safely complete the attempted procedure.
having such qualifications.
i
Notes
ii
1/Precautions
1.1 Definitions If the screws on the timing valves are turned all
the way clockwise, the ISU and Posi-Pulse ISU
Note: A Note provides additional information to will not cycle.
clarify a point in the text.
If the Vacuum Regulator is repaired or disas-
Important: An Important statement is similar to sembled in any manner, the Service Checkout
a note but of greater emphasis. Procedure must be performed before using the
equipment on the patient.
wCAUTION: A CAUTION statement is used
when the possibility of damage to the equip- The fitting port of the regulator must be oc-
ment exists. cluded when setting the prescribed suction
level. This prevents the patient from receiving
wWARNING: A WARNING statement is used
higher than required suction levels.
when the possibility of injury to the patient or
the operator exists. The Posi-Pulse ISU must be mounted vertically
for proper operation. The regulator may not
1.2 Warnings operate correctly in any other position.
After patient use, regulators may be contami- The Pre-Use Checkout Procedure must be
nated. Handle in accordance with your performed before using this equipment on each
hospital’s infection control policy. patient. If the regulator fails any part of the Pre-
Use Checkout Procedure, it must be removed
Clean and sterilize all multiple use suction from service and repaired by qualified service
equipment before shipment to ensure transpor- personnel.
tation personnel and/or service personnel are
not exposed to any hazardous contamination. When using the Posi-Pulse ISU with a Posi-
Pulse Device, avoid using suction over 200 mm
Clean and sterilize all multiple use suction Hg/26 kPa or Posi-Pulse operation may not
equipment if contaminated before disassembly, allow effective drainage.
to ensure service personnel are not exposed to
hazardous contamination.
Connect the Vacuum Regulator to the vacuum
source only. Connection to pressure sources
even momentarily could injure the patient or
operator and damage the equipment.
Always connect the regulator to the vacuum
source and check its operation before attaching
the patient connection.
Do not use this device in the presence of flam-
mable anesthetics. Static charges may not
dissipate and possible explosion hazard exists
in the presence of these agents.
Following sterilization with ethylene oxide,
parts should be quarantined in a well ventilated
area to allow dissipation of residual ethylene
oxide gas absorbed by the material. Follow
sterilizer manufacturer’s recommendations for
specific aeration periods required.
For proper Posi-Pulse operation, the Posi-Pulse
ISU’s “On” and “Off” timing cycles must be a
minimum of 10 seconds.
If the flow control valve is rotated fully clock-
wise, the ISU and Posi-Pulse ISU will not cycle
in the intermittent mode. Drainage will cease.
1-1 1-1
1/Precautions
1.3 Cautions
A supply vacuum of less than 15 in Hg/51 kPa
(380 mm Hg) may impair Posi-Pulse ISU and
Posi-Pulse Device operation.
Cleaning the gauge may result in damage.
Do not lubricate any internal components of the
regulator module.
Do not over-tighten the timing valve stem.
Needle portion of the valve may be damaged.
Do not steam autoclave or liquid sterilize the
ISU or Posi-Pulse ISU. Severe impairment to the
operation of the regulator will result. The only
acceptable method of sterilization is with gas
(ethylene oxide).
Do not use any 1Loctite® products to seal the
fitting and adapter port threads (or products
which contain Methacrylate Ester as an active
ingredient).
Only competent individuals trained in the repair
of this equipment should attempt to service it.
Sterilization with ethylene oxide mixtures may
cause crazing (minute superficial cracking) of
some plastic parts. Crazing will be more pro-
nounced when mixtures containing 2Freon® are
used.
The cap screws can strip the regulator module
housing threads if they are screwed in too tight.
The gauge assembly must be handled with
utmost care to retain its precision. If the lens is
removed, do not rest the gauge on its face.
To help prevent aspirate from entering the
regulator, an Overflow Safety Trap should be
attached prior to its use. Aspirate in the regula-
tor will impair the operation. The use of the
Overflow Safety Trap will help prevent this and
extend the life of the suction equipment.
To prevent stripping the plastic threads, place
the screw in the hole and turn counterclockwise
until it drops into the original threads. Tighten
screw.
Use of lubricants other than recommended, may
degrade plastic or rubber components.
1
Loctite is the registered trademark of the Loctite Corporation.
2
Freon is the registered trademark of the DuPont Co.
1-2 1-2
2/Scope
This service manual contains service, mainte-
nance and parts information on four models of
the ISU and four models of the Posi-Pulse ISU.
North American (2)
English (2)
French (2)
Spanish (2)
2-1 2-1
2/Scope
2.3 French Vacuum Regulators
Note: Part numbers given are for Vacuum
Regulators without fittings or adapters.
2-2 2-2
3/Description and Specifications
3.1 Description The Posi-Pulse ISU also contains a pause
module and Posi-Pulse attachment port. It can
wWARNING: Do not use this device in the be used with or without a disposable Posi-Pulse
presence of flammable anesthetics. Static Drainage Assist Device®.
charges may not dissipate and possible explo-
In use, the vacuum source is connected through
sion hazard exists in the presence of these
agents. the regulator module which functions as an
automatic valve. Rotating the suction control
The Intermittent Suction Unit (ISU) and Posi- knob adjusts the position of the regulator
Pulse Intermittent Suction Unit (Posi-Pulse ISU) module and selects a predetermined level of
are dual purpose Vacuum Regulators which suction.
provide either intermittent or continuous
During use, as the flow requirement increases,
suction. They can be used for N.G. (intermit-
tent) or pharyngeal/tracheal (continuous) the valve automatically opens to maintain
suction at the preset level. Conversely, when
suctioning throughout the hospital.
the flow requirement decreases, the valve
Each unit contains a regulator module to automatically closes to maintain suction at the
regulate and adjust suction, a vacuum gauge preset level. The same mechanism compen-
which indicates suction supplied, a patented sates for changes in supply vacuum to auto-
Unilogic Module to silently and reliably switch matically maintain the pre-set suction level.
suction on and off, and timing valves which
Note: Posi-Pulse Drainage Assist Device® is a
select and adjust the length of the on and off
registered trademark of The BOC Group, Inc.
cycles.
Pause
Module
Cover *
Vacuum
Gauge
Mode
Selector Adapter
Knob Port
Fitting Port
Posi-Pulse
Suction Attachment Port *
Control Knob
3-1 3-1
3/Description and Specifications
1. Suction Control Knob - Allows easy adjust- When used as an Intermittent Suction
ment of suction to the patient. Unit, both the “On” and “Off” timing
cycles are independently adjustable within
2. Mode Selector Knob - Allows quick and easy a range of 3 to 30 seconds. These are pre-
mode changes.
set at the factory at a supply vacuum of
Note: In the text of this manual, the interna- 450 mm Hg/60 kPa to provide approxi-
tional graphic symbol for the intermittent mately 15 seconds “On” and 8 seconds
mode normally written as: “Off” for the ISU and 15 seconds “On” and
15 seconds “Off” for the Posi-Pulse ISU. If
| the supply vacuum is different, the timing
O cycles will vary. Adjustment procedures
| are outlined in the Operations Section of
O this manual under the heading “Setting
| the Timing Cycles”.
will be written as |O|O|. b. O (OFF) - No suction is supplied to the
a. IOIOI (INTER) or IOIOI P-P (INTER) - patient.
Suction is intermittent (cycles on and off) c. I (CONT) - Suction is continuous and can
and the suction level can be adjusted with be adjusted with the suction control knob.
the suction control knob. This is the
selection used for Posi-Pulse N.G. Drainage
System operation. 3. Vacuum Gauge - The suction level to the
patient is displayed during use.
3.2 Specifications
Gauge: Accuracy: ±5% of full scale deflection
Posi-Pulse Signal Line Pause: 6 seconds ± 4 seconds from the start of the OFF time cycle
3-2 3-2
4/Operation
4.1 Equipment Set-up If the regulator is equipped with an adapter for
wall outlets, insert the adapter into the vacuum
wWARNING: Connect the Vacuum Regulator wall outlet. If the regulator is mounted else-
to the vacuum source only. Connection to where, connect a vacuum supply hose between
pressure sources even momentarily could the regulator’s adapter port and the wall outlet.
injure the patient or operator and damage the Connect the Collection Bottle’s vacuum port to
equipment. the regulator’s fitting or Overflow Safety Trap
with the appropriate vacuum tubing.
wWARNING: Always connect the regulator to
the vacuum source and check its operation Note: For proper installation of Adapters/Probes
before attaching the patinet connection. and Fittings see Appendix.
wWARNING: The Posi-Pulse ISU must be For standard ISU operation, mount the Collec-
mounted vertically for proper operation. The tion Bottle above the patient.
regulator may not operate correctly in any
For optimal Posi-Pulse system operation, mount
other position.
the Collection Bottle at or below patient level.
wCAUTION: To help prevent aspirate from
entering the regulator, an Overflow Safety An Ohmeda High Flow Disposable Filter should
be used between the Collection Bottle and the
Trap should be attached prior to its use.
Vacuum Regulator to prevent contamination of
Aspirate in the regulator will impair the
operation. The use of the Overflow Safety the regulator.
Trap will help prevent this and extend the life Suction Filters
of the suction equipment.
Carton of 20 ................................... 6730-0350-800
Carton of 200 ................................. 6730-0351-800
Vacuum
Port Connective
Tubing
Vacuum
Regulator
Patient
Port
Disposable
Suction Filter
Overflow
Safety Trap
Collection
Bottle
Equipment Set-up
4-1 4-1
4/Operation
O (OFF)
I (CONT)
4-2 4-2
4/Operation
4.4 Pre-Use Checkout Procedure
wWARNING: The Pre-Use Checkout Procedure
must be performed before using this equip-
ment on each patient. If the regulator fails any
part of the Pre-Use Checkout Procedure, it
must be removed from service and repaired
by qualified service personnel.
wWARNING: Always connect the regulator to
the vacuum source and check its operation O
before attaching the patinet connection. (OFF)
Before the vacuum regulator is used, it must be
tested for correct operation as outlined below.
All tests must be performed with the regulator
in its normal (vertical) operating position and
with a minimum supply vacuum of 15 in Hg/51
kPa (380 mm Hg).
1. Turn the mode selector knob to O (OFF). Clamp
Rotate the suction control knob one full turn
clockwise (increase). Clamp the connective
tubing to occlude the fitting port. The gauge
needle should not move.
Step 1
Clamp
Step 2
4-3 4-3
4/Operation
Clamp
Release
Clamp
Step 3
Clamp
Step 4
4-4 4-4
4/Operation
5. For Posi-Pulse ISU ONLY with a Posi-Pulse d. During the “On” cycle, ensure that the
Device pistons lift inside the Posi-Pulse Device’s
clear chamber.
a. Connect the Posi-Pulse signal line (blue
tubing) to the attachment port. e. When the “Off” cycle begins, ensure that
there is a delay of 2 to 10 seconds before
b. Turn the mode selector knob to |O|O| P-P
the pistons drop.
(INTER).
c. Clamp the connective tubing.
ON OFF
Cycle Cycle
I
O PP
I (INTER)
O
I
Posi-Pulse Device
Clamp
Step 5
O
(OFF)
Step 6
4-5 4-5
4/Operation
4.5 Patient Set-up
1. Make sure the Pre-Use Checkout Procedure
has been performed.
2. Turn the mode selector knob to | (CONT) and
clamp the connective tubing.
3. Set the prescribed suction level.
wWARNING: The fitting port of the regula-
tor must be occluded when setting the I
(CONT)
prescribed suction level. This prevents the
patient from receiving higher than required
suction levels.
Clamp
Steps 1 to 3
Patient
4. Turn the mode selector knob to O (OFF) Connective
5. Attach the patient tubing to the Collection Tubing
Bottle’s patient port.
O
(OFF)
Steps 4 to 5
4-6 4-6
4/Operation
6. Turn the mode selector knob to | (CONT) or
|O|O| (INTER)
I
O I
I (INTER) or (CONT)
O (CONT)
I
Step 6
4-7 4-7
4/Operation
4.7 Setting the Intermittent
Mode Flow Rate
wWARNING: If the flow control valve is
rotated fully clockwise, the ISU and Posi-Pulse
ISU will not cycle in the intermittent mode.
Drainage will cease.
A change in the intermittent mode flow rate is
rarely needed. If there is a desire to change the
factory set flow rate, perform the following
procedure.
1. Connect the supply vacuum to the adapter
port.
2. Pull the Ohmeda nameplate wrench from
the front of the regulator to access the flow
control valve.
3. Occlude the regulator fitting port.
4. Turn the mode selector knob to | (CONT).
5. Set 120 mm Hg/16 kPa on the regulator
gauge.
6. Turn the mode selector knob to O (OFF).
7. Connect the regulator fitting port to the
10 LPM flowmeter.
8. Turn the mode selector knob to |O|O|
(INTER) or |O|O| P-P (INTER).
9. Locate the flow control valve in the regula-
tor.
10. The flow rate must be adjusted during the
“On” cycle of the intermittent mode. Using
the proper tool, either a Phillips head
screwdriver or a ball hex key wrench, adjust
the flow control valve to produce a flow rate
of 8 LPM. To increase the flow rate turn the
valve counterclockwise. To decrease the
flow rate turn the valve clockwise.
4-8 4-8
5/Cleaning and Sterilization
5.1 Cleaning 5.2 Sterilization
wWARNING: After patient use, regulators wWARNING: Following sterilization with
may be contaminated. Handle in accordance ethylene oxide, parts should be quarantined in
with your hospital’s infection control policy. a well ventilated area to allow dissipation of
residual ethylene oxide gas absorbed by the
5.1.1 Routine Exterior Cleaning material. Follow sterilizer manufacturer’s
Routine cleaning of the regulator is recom- recommendations for specific aeration periods
mended as a standard procedure after each use. required.
Wipe all exterior surfaces with a solution of wCAUTION: Do not steam autoclave or liquid
water and mild detergent. sterilize the ISU or Posi-Pulse ISU. Severe
impairment to the operation of the regulator
5.1.2 Internal Component Cleaning will result. The only acceptable method of
wCAUTION: Cleaning the gauge may result in sterilization is with gas (ethylene oxide).
damage. wCAUTION: Sterilization with ethylene oxide
1. Refer to the Service - Disassembly and mixtures may cause crazing (minute superfi-
Assembly instructions. cial cracking) of some plastic parts. Crazing
will be more pronounced when mixtures
2. All internal components with the exception containing Freon® are used.
of the gauge, may be cleaned with a solution
of warm water and mild detergent. Note: ISU or Posi-Pulse ISU should only be
sterilized if it is contaminated or maintenance is
3. Dry all components with a lint free cloth to be performed.
before assembly.
1. The regulator should be sterilized with the
mode selector knob in the |O|O| (INTER) or
|O|O| P-P (INTER) position.
2. The only acceptable method of sterilization is
with ethylene oxide. Ethylene oxide mixtures
can be used at temperatures of 125 to 135°F
(52 - 57°C ). If this temperature cannot be
obtained, room temperature sterilization with
100% ethylene oxide can also be used.
Sterilization is not recommended as a stan-
dard procedure after each use.
3. After each sterilization check the condition of
the internal filters. Replace any shrunken
filters before returning the regulator to
service.
5-1 5-1
5/Cleaning and Sterilization
Notes
5-2 5-2
6/Troubleshooting
Place the regulator in the vertical position and
connect to a supply vacuum of 15 in Hg/51 kPa
(380 mm Hg).
6-1 6-1
6/Troubleshooting
Problem Possible Causes Remedy
F. Insufficient flow 1. Partial blockage in 1. Confirm wall supply open
through regulator wall supply air flow meets minimum
hospital requirements
Unblock if necessary
2. Partial blockage in 2. Perform the Flow Test in
regulator the Service Checkout
Procedure section of this
manual
Clear blockage if necessary
H. With the fitting port 1. Blocked filter and/ 1. Clear orifice and replace
occluded, unable to or orifice filter
decrease the suction
level and gauge needle 2. Damaged gauge 2. Replace gauge
does not return to zero
when switched to 3. Blocked unilogic 3. Replace filter
O (OFF) module filter
4. Regulator module 4. See “Regulator module
malfunction malfunction” under Problem
column in this section of the
manual
6-2 6-2
6/Troubleshooting
Problem Possible Causes Remedy
K. Regulator cycles in the 1. Blue signal line 1. Reconnect blue signal line
|O|O| P-P (INTER) mode, disconnected
but the Posi-Pulse Device
pistons do not lift 2. Posi-pulse connector 2. Tighten connector
not tightly secured to
attachment port
M. Posi-Pulse pistons lift 1. “Off” timing cycle is too 1. Adjust “Off” timing to
but do not drop short 15 seconds
2. Blue signal line tubing 2. Straighten or replace tubing
is kinked
6-3 6-3
6/Troubleshooting
Notes
6-4 6-4
7/Disassembly and Assembly
7.1 Service Tools and Equipment 7.2 ISU and Posi-Pulse ISU
wCAUTION: Use of lubricants other than 7.2.1 Disassembly
recommended, may degrade plastic or rubber
wWARNING: If the Vacuum Regulator is
components.
repaired or disassembled in any manner, the
The following items should be on hand during Service Checkout Procedure must be per-
any service procedure: formed before using the equipment on the
patient.
• Supply Vacuum: 500 mm Hg/67 kPa minimum
& 50 LPM Open Air Flow minimum wWARNING: Clean and sterilize all multiple
use suction equipment if contaminated before
• Supply Vacuum Regulator with Gauge,
disassembly, to ensure service personnel are
760 mm Hg/101.3 kPa Full Scale
not exposed to hazardous contamination.
• Low Vacuum Calibration Gauge, 225 mm Hg/ wCAUTION: The gauge assembly must be
30 kPa Full Scale (*)
handled with utmost care to retain its preci-
(Ohmeda P/N 6700-0353-800)
sion. If the lens is removed, do not rest the
• 50 LPM Flowmeter gauge on its face.
(Ohmeda P/N 6700-0355-800)
• 10 LPM Flowmeter
(Ohmeda P/N 6700-0354-800)
• Phillips Head Screwdriver, No. 2
• Flat Head Screwdriver, 1/4 inch
• Open End Adjustable Wrench
• Ball Hex Key Wrench, 3/16 inch
• Open End Wrench, 11/16 inch
• Pipe Cleaner
• Wooden Toothpick (O-ring Remover)
• Tweezers (Filter Remover)
3
• Dow® 111 Grease
(Ohmeda P/N 6700-0074-200)
• Tubing Clamp
• 1/2 Gallon Reservoir
• Bubble Leak Tester
• Stop Watch
7-1 7-1
7/Disassembly and Assembly
27
21 25
17
24
23 2018 26
22 17 19 16
27
18 15
10
9
2 3 8
7 14
6
5 13
4 12
1
11
28
29
Description
1. Gauge Assembly (Includes Gauge, O-rings and 15. Flow Control Valve
Lens) 16. O-ring, 2-011 Buna-N
2. Gauge Lens 17. Timing Valve (2)
3. O-ring, 2-012 Buna-N (2) 18. O-ring, 2-006 Buna-N (4)
4. Screw 19. O-ring, 2-014 Buna-N (2)
5. Retaining Washer 20. Filter (2)
6. Spiral Spring Washer 21. Unilogic Module
7. Detent Plate 22. Filter, Green
8. Ball (4) 23. Filter, Beige
9. Switch Main Valve Body 24. Cap (2)
10. O-ring, 2-010 Buna-N, (4) 25. Tube (6)
11. Regulator Module 26. Backplate Assembly
12. O-ring, 2-016 Buna-N 27. Cover Screw (3)
13. O-ring, 2-004 Buna-N 28. Ohmeda Nameplate Wrench
14. Gauge Filter 29. Cover & Knob Assembly
7-2 7-2
7/Disassembly and Assembly
Posi-Pulse ISU
3
ISU 10
3 8
7
4
2 5
10
6 4
4
2
1
Description
1. Mounting Plate 1
2. Gasket
3. Serial Number Label
4. Screw (4 or 5)
5. Pause Cover
6. Posi-Pulse Attachment Port
7. Silastic Tube
8. Attachment Port/Pause Module Tube
9. Pause Module
10. Backplate Assembly
7-3 7-3
7/Disassembly and Assembly
5. To remove the regulator module, grasp and 7.2.2 Assembly
pull it from the backplate.
wCAUTION: To prevent stripping the plastic
Note: To disassemble the regulator module, threads, place the screw in the hole and turn
refer to the Regulator Module Disassembly counterclockwise until it drops into the origi-
section. nal threads. Tighten screw.
6. Remove the unilogic module by carefully 1. Lubricate all o-rings and the gasket with a
pulling the unit away from the backplate. small amount of Dow® 111 Grease before
This releases the tubing connections. assembly.
Replace beige and green filters and silastic
tubes if required. 2. Place the lubricated mounting plate gasket
onto the mounting plate properly aligning
7. Remove the timing valves by pulling the the holes and install the mounting plate onto
valve stem out of the backplate. Replace the the backplate. Ensure the protruding posts
filters if required. on the mounting plate align into the
backplate properly.
Note: To disassemble the timing valves,
refer to the Timing Valve Disassembly 3. Install the four mounting plate screws and
section. tighten.
8. If the mode selector switch needs service, 4. For the Posi-Pulse ISU ONLY.
carefully unscrew the center retaining
screw while holding the main valve body. a. Install the attachment port into the mount-
ing plate and tighten by hand. Rotate an
Remove the washers, detent plate, ball
additional 1/8th revolution (45 degrees)
bearings, and valve body. Replace o-rings if
required. with the 11/16 inch open end wrench. Do
not over tighten.
9. Remove the gauge bleed hole filter from the
b. Install the pause module by aligning the
backplate. Replace if required.
large silastic tube with the mounting plate
10. Remove the flow control valve by turning port and small silastic tube with the
counterclockwise until it is out of the attachment port barb. Press until the
backplate. Take extra care when removing tubes seat. The pause module will be
this valve to ensure that the threads are supported by the tubes and the backplate.
engaged properly. Replace o-ring if re-
c. Place the pause cover over the pause
quired.
module and slide over the mounting plate.
11. For the ISU ONLY - Remove the ISU mount- Install the screw.
ing plate by removing the four screws and
5. Install the flow control valve into its proper
pulling the plate and gasket from the
backplate. Replace gasket if required. location. Use the proper tool to rotate the
valve clockwise approximately 3 full turns,
12. For the Posi-Pulse ISU ONLY. taking extra care to ensure that the threads
are engaged properly.
a. Remove the pause cover by removing the
screw. 6. Place the gauge bleed hole filter into its
proper location. Push until it bottoms on the
b. Remove the pause module by grasping
backplate.
and pulling it away from the mounting
plate and attachment port. This releases 7. Mode Selector Switch Assembly
the tubing connections. Replace tubing if
a. Place the four selector switch o-rings into
required.
their proper location on the backplate.
c. Remove the attachment port from the
b. Install the main valve body on the the
mounting plate by rotating it counter-
backplate post with the protruding ear at
clockwise with the 11/16 inch open end
wrench. the 3 o’clock position. Ensure the valve
body seals on the four o-rings.
d. Remove the Posi-Pulse ISU mounting
c. Place the four ball bearings on top of the
plate by removing the four screws and
pulling the plate and gasket from the valve body. Place the detent plate on top
of the four ball bearings. Place the spiral
backplate. Replace gasket if required.
7-4 7-4
7/Disassembly and Assembly
spring washer on top of the detent plate. 14. Place and align the cover and knob assem-
Place the flat washer on top of the spiral bly onto the backplate with the mode
spring washer. selector knob in the O (OFF) position. Rotate
the suction control knob clockwise (in-
d. Install the screw through the washer and
crease) until the cover is flush with the
into the backplate post. Tighten the
backplate.
screw.
15. Install the three cover screws.
8. Grasp the main valve body and rotate the
switch in all three positions. Ensure that the 16. Install all fittings.
protruding ear is near the 3 o’clock position
after testing the switch. Note: For proper installation of Adapters/
Probes and Fittings see Appendix.
9. Install the timing valves into their proper
locations and push fully in so that the o- 7.3 Regulator Module
rings seal in the backplate.
wCAUTION: Do not lubricate any internal
10. Install the unilogic module by aligning the components of the regulator module.
silastic tubes with the ports on the
backplate and push fully on. If necessary, wCAUTION: The cap screws can strip the
wet the tubes with water for easier installa- regulator module housing threads if they are
tion. screwed in too tight.
11. Install the regulator module into its proper 7.3.1 Disassembly
location. Rotate the module so that the 1. Remove the cap screws (2) and remove the
protruding ears are at the 5 o’clock and 11 cap.
o’clock positions.
2. Remove the o-rings (2).
12. Install the snap-fit lens by placing one of the
retaining tabs over the edge of the gauge 3. Observe the position of the diaphragm
face, and then pressing lightly until the convolution to aid in assembly.
other tab snaps onto the gauge face. Rotate
4. Withdraw the diaphragm/stem assembly
the lens to confirm proper installation. If the from the housing.
lens is not the snap-fit type, simply place
the lens onto the gauge ensuring it rotates 5. Remove the spring.
easily and is resting properly.
6. Grasp the stem with your fingers and un-
13. Install the gauge assembly into its proper screw the stem screw.
location so that the o-rings seal fully. Ensure
7. Remove the retainer, diaphragm and piston
the gauge is properly aligned.
from the stem.
11
10
9 11
8
7
6
5
Description
1. O-ring, Stem
2. O-ring 4
3. Housing 3
4. Stem
5. Spring
6. Piston 2
7. Diaphragm
8. Retainer 1
9. Stem Screw
10. Cap
11. Cap Screw Parts and Descriptions
7-5 7-5
7/Disassembly and Assembly
7.3.2 Assembly 6. Place the spring in the housing.
1. Return the diaphragm to its original position 7. Insert the retainer/diaphragm/piston/stem
(convoluted), with the molded number facing assembly through the spring and into the
the retainer. cavity in the housing. Fit the diaphragm
bead into the groove in the housing.
2. Assemble the diaphragm onto the retainer by
lining up the center holes as shown in the 8. Install the cap and the two cap screws.
diagram..
9. Apply a small amount of Dow® 111 Grease
3. Insert the piston into the diaphragm and to the o-rings only.
ensure that the retainer mates with the
recess in the piston. 10. Install o-rings onto the stem and housing.
Retainer
Stem
Stem Screw
Piston
Diaphragm
Diaphragm Assembly
7-6 7-6
8/Service Checkout Procedure
wWARNING: If the Vacuum Regulator is 8.2 Leak Test - Supply Side
repaired or disassembled in any manner, the
Service Checkout Procedure must be per- 1. Connect the supply vacuum tubing to port
formed before using the equipment on the “A” of the Bubble Leak Tester.
patient.
2. Connect port “B” of the Bubble Leak Tester
Important: This entire procedure must be to the regulator adapter port.
performed in numerical order.
3. Set the supply vacuum to 300 mm Hg ± 30
mm Hg/40 kPa ± 4 kPa on the supply vacuum
8.1 Set-up gauge.
1. Verify that a minimum of 500 mm Hg/67 kPa 4. Turn the mode selector knob to O (OFF).
vacuum exists at the supply vacuum gauge. Allow the fitting port to be open to air.
2. The supply open flow must be 50 LPM 5. Wait 5 seconds.
minimum.
6. No more than 5 bubbles should appear in the
Bubble Leak Tester for the next 10 seconds.
Important: Prior to venting Port “A” of the
Bubble Leak Tester to atmosphere (ie: turn-
ing the Supply Regulator off) ensure the tub-
ing from Port “B” has been disconnected
from the Adapter Port of the Vacuum Regula-
tor.
Vacuum Regulator
Port B Port A
Sealed
Fitting
Port
1.0 In
Water
8-1 8-1
8/Service Checkout Procedure
8.3 Flow Test 8.3.2 Intermittent Mode Flow Test
1. Using the proper tool, rotate the flow control
1. Remove the Bubble Leak Tester from the
supply vacuum connection and connect the valve two turns counterclockwise.
supply vacuum directly to the regulator 2. Clamp the fitting port tubing.
adapter port.
3. Set 120 mm Hg/16 kPa on the regulator
2. Set the supply vacuum to 450 mm Hg ±25 gauge.
mm Hg/60 kPa ±3.3 kPa on the supply vac-
uum gauge. 4. Connect the regulator fitting port to the 10
LPM flowmeter.
8.3.1 Continuous Mode Flow Test
5. Turn the mode selector knob to |O|O| (INTER)
1. Connect the regulator fitting port to the 50 or |O|O| P-P (INTER).
LPM flowmeter.
6. During the “On” cycle, adjust the intermit-
2. Turn the mode selector knob to | (CONT). tent mode flow rate to 8 LPM ± 1 LPM.
3. Rotate the suction control knob fully clock- Note: If the regulator does not cycle “On”
wise (increase). after 20 seconds in the intermittent mode,
4. Verify that the flowrate exceeds 35 LPM. use a flat head screwdriver and rotate the
“Off” timing valve stem counterclockwise
until it cycles “On”.
Vacuum “On”
Regulator Timing
Valve
“Off”
Timing
Valve Flow Supply Vacuum Gauge
Posi-Pulse Control 0-760 mm Hg/0-101.3 kPa
Signal Line Valve
Supply Regulator
Set 450 mm Hg/60 kPa
Mode
Posi-Pulse Device Selector Suction
Knob Control
Knob Supply Vacuum
500 mm Hg/67 kPa
Minimum and
Open Flow of 50 LPM
Minimum
Closed
Port
Occluded
Low Vacuum
Calibration Gauge 1/2 Gallon
Flowmeter Flowmeter 225 mm Hg/30 kPa Reservoir
50 LPM 10 LPM
8-2 8-2
8/Service Checkout Procedure
8.4 Gauge Test
Note: All Ohmeda Gauges are supplied with an 4. Adjust the timing cycles:
accuracy of ±5% of full scale deflection through-
a. ISU -
out their range. The ISU and Posi-Pulse ISU
vacuum gauges have an accuracy of ±10 mm On cycle 15 seconds ± 3 seconds
Hg/1.3 kPa. Off cycle 8 seconds ± 3 seconds
Note: All gauge needles should return to the b. Posi-Pulse ISU -
stop pin when no suction is being supplied.
Gauges which do not comply may be out of On cycle 15 seconds ± 3 seconds
Off cycle 15 seconds ± 3 seconds
calibration.
Note: Adjustment procedures are outlined in
When checking gauge accuracy, be sure that
the calibration gauge has an accuracy of 1% of the Operations Section of this manual under
the heading “Setting the Timing Cycles”.
full scale deflection or better.
1. Connect the regulator fitting port to the low 8.7 Vacuum Build-up/Bleed-down
vacuum calibration gauge with tubing.
Test - Intermittent Mode
2. Turn the mode selector knob to | (CONT).
1. Connect the regulator fitting port to the 1/2
3. Assure that the regulator gauge is in agree- gallon reservoir.
ment with the low vacuum calibration gauge
within the ±10 mm Hg/1.3 kPa tolerance. 2. During the “On” cycle, check that the suction
Recommended test points are: increases to the preset 120 mm Hg/16 kPa on
the regulator gauge in 5 seconds or less.
40, 80, 140, and 200 mm Hg
3. During the “Off” cycle, check that the suc-
or
tion decreases to zero on the regulator gauge
5, 10, 18, and 26 kPa (zero stop pin) in 5 seconds or less.
4. Rotate the suction control knob fully clock- 4. If the regulator fails either of the Build-up or
wise (increase) and verify the gauge’s read- Bleed-down tests, replace the regulator
ing is in the FULL VAC range. module.
5. Rotate the suction control knob fully counter-
clockwise (decrease) and verify the gauge’s 8.8 Posi-Pulse Test
reading decreases to zero. For Posi-Pulse ISU ONLY with a Posi-Pulse
Device.
8.5 Regulation Test
1. Connect the Posi-Pulse signal line (blue
1. Disconnect the low vacuum calibration gauge tubing) to the Posi-Pulse attachment port.
from the regulator fitting port.
2. Connect the regulator fitting port to the 1/2
2. Occlude the fitting port. gallon reservoir.
3. Set 90 mm Hg/12 kPa on the regulator gauge. 3. Turn the mode selector knob to |O|O| P-P
4. Open and close the fitting port several times. (INTER).
5. Check that the regulator gauge reads be- 4. Set 120 mm Hg/16 kPa on the regulator
tween 80 mm Hg/10.7 kPa and 100 mm Hg/ gauge during the “On” cycle.
13.3 kPa when the regulator fitting port is 5. Ensure that the pistons inside the Posi-Pulse
occluded. Device’s chamber lift during the “On” cycle.
8-3 8-3
8/Service Checkout Procedure
8.9 Bleed Test 8.10 Leak Test - Patient Side
1. Turn the mode selector knob to I (CONT). 1. Turn the mode selector knob to O (OFF).
2. Occlude the regulator fitting port and set the 2. Connect the regulator fitting port tubing to
vacuum level to 120 mm Hg/16 kPa. port “A” of the Bubble Leak Tester.
3. Turn the mode selector knob to O (OFF) and 3. Rotate the suction control knob fully counter-
observe the gauge needle. It must return to clockwise (decrease).
zero within 10 seconds.
4. Turn the mode selector knob to | (CONT).
5. Wait 5 seconds.
6. No bubbles should appear in the Bubble Leak
Tester for the next 10 seconds.
Port B
Port A
Sealed
1/4” I.D.
Clear Tubing
Mode Suction With Knife
Selector Control Edge
Knob Knob 1.0 in
Water
8-4 8-4
9/Maintenance
9.1 General Maintenance of Routine maintenance and inspection are impor-
tant to the performance of suction equipment.
Suction Equipment The following is a recommended list for care of
suction equipment after each patient use.
wWARNING: The Pre-Use Checkout Procedure
must be performed before using this equip- 1. Perform a thorough cleaning by washing all
ment on each patient. If the regulator fails any bottles, tubing, metal connectors, etc, and
part of the Pre-Use Checkout Procedure, it removing all residue.
must be removed from service and repaired
by qualified service personnel. 2. Wipe all exterior surfaces with a solution of
water and mild detergent.
wWARNING: Clean and sterilize all suction
equipment if contaminated before disassem- 3. Perform a careful visual inspection.
bly, to ensure service personnel are not 4. Check that the High Flow Disposable Suction
exposed to hazardous contamination. Filter is clean and in good condition.
ISUs and Posi-Pulse ISUs should be kept in use 5. Check that all tubing is in good condition and
or used on a rotating basis. Internal parts of connected securely to the correct ports.
unused equipment may tend to deteriorate.
6. Check the floats in the Overflow Safety Trap
Maintenance of the vacuum piping system is as and Collection Bottle for correct operation.
important as maintenance of the suction equip-
ment. The use of Collection Bottles with reliable 7. Perform the Pre-Use Checkout Procedure.
shut-off valves, Overflow Safety Trap assem-
blies and disposable suction filters will protect 9.2 Recommended Maintenance
the regulator and piping system. The flow rate Schedule
at the wall outlet should be checked on a yearly
basis and suitable cleaning of the outlets In addition to the Pre-use Checkout Procedure,
should be performed. The flow rate measure- the following periodic maintenance should be
ment should meet local standards. performed.
9-1 9-1
9/Maintenance
9.3 Repair Policy 9.5 Return Instructions
wWARNING: Clean and sterilize all multiple 1. Clean and sterilize the Vacuum Regulator.
use suction equipment before shipment to
2. Package the Vacuum Regulator securely for
ensure transportation personnel and /or
protection, preferably in the original con-
service personnel are not exposed to any
hazardous contamination. tainer.
3. Include a letter describing in detail any
wCAUTION: Do not steam autoclave or liquid
difficulties experienced with the Vacuum
sterilize the ISU or Posi-Pulse ISU. Severe
impairment to the operation of the regulator Regulator. Include the person, title, and
telephone number to contact for functional
will result. The only acceptable method of
questions.
sterilization is with gas (ethylene oxide).
wCAUTION: Only competent individuals 4. If the Vacuum Regulator is less than one year
old, include warranty information that came
trained in the repair of this equipment should
with the device and a copy of the invoice.
attempt to service it.
Do not use malfunctioning equipment. Make all 5. Include a purchase order to cover repair of a
regulator not under warranty.
necessary repairs. Have the equipment repaired
by qualified service personnel or by Ohmeda. 6. Ship the Vacuum Regulator prepaid. Write
Parts listed in this service manual may be your return address and billing address
repaired or replaced by a competent, trained information on the package or letter that
person who has experience in repairing devices comes with the package.
of this nature. After repair, perform the Service
Checkout Procedure to ensure that it is func- For U.S.A Warranty Repairs (less than one
tioning properly, and complies with the year from the purchase date) mail the pack-
manufacturer’s published specifications. age to:
Ohmeda
9.4 Technical Assistance 3411 Commercial Avenue
Northbrook, IL 60062 USA
If technical assistance is required, contact
Ohmeda technical support or field operations For U.S.A Nonwarranty Repairs (more than
listed on the back cover. one year from the purchase date) mail the
package to:
Ohmeda
Service and Distribution Center
7750 The Bluffs NW
Austell, GA 30001 USA
In other locations contact your nearest
Ohmeda office or authorized Ohmeda dis-
tributor.
9-2 9-2
10/Ordering Information
10.1 Illustrated Parts
10.4.4 Spanish
Instructions: Select the vacuum regulator and fittings
to form an 11-digit part number.
Example: 6700-1237-901 = Three Mode High
Continuous Regulator w/Tubing Nipple
27
21 25
17
24
23 2018 26
22 17 19 16
27
18 15
10
9
2 3 8
7 14
6
5 13
4 12
1
11
28
29
Description Part Number
1. Gauge Assembly (Includes Gauge, 18. O-ring, 2-006 Buna-N (4)* ............... 0210-0593-300
O-rings and Lens) 19. O-ring, 2-014 Buna-N (2)* ............... 6700-0135-500
North American Posi-Pulse ISU .... 6700-0089-200 20. Filter (2)* .......................................... 0221-5883-300
International ISU/Posi-Pulse ISU . 6700-0090-200 21. Unilogic Module .............................. 6700-0093-200
North American ISU ....................... 6700-0050-200 22. Filter, Green* .................................. 0221-5879-300
2. Gauge Lens ...................................... 6700-0087-500 23. Filter, Beige* ................................... 0221-5880-300
3. O-ring, 2-012 Buna-N (2)* ............... 6700-0130-500 24. Cap (2) ............................................... 0203-0165-300
4. Screw ................................................ 6700-0150-400 25. Tube (6) ............................................. 0211-0675-300
5. Retaining Washer ........................... 0402-0748-300 26. Backplate Assembly ....................... 6700-0043-500
6. Spiral Spring Washer ..................... 6700-0168-400 27. Cover Screw (3) ............................... 6700-0152-400
7. Detent Plate ..................................... 0214-3213-300 28. Ohmeda Nameplate Wrench
8. Ball (4) ............................................... 0409-1684-300 Posi-Pulse ISU ................................. 6700-0033-400
9. Switch Main Valve Body ................ 0221-5891-300 North American/English ISU ......... 6700-0043-400
10. O-ring, 2-010 Buna-N, (4)* .............. 6700-0131-500 French ISU ....................................... 6700-0044-400
11. Regulator Module ............................ 6700-1225-800 Spanish ISU ...................................... 6700-0045-400
12. O-ring, 2-016 Buna-N* .................... 6700-0136-500 29. Cover & Knob Assembly
13. O-ring, 2-004 Buna-N* .................... 0210-0527-300 Posi-Pulse ISU ................................. 6700-0016-700
14. Gauge Filter* ................................... 0210-6540-300 ISU ..................................................... 6700-0009-700
15. Flow Control Valve ......................... 0207-0132-300
16. O-ring, 2-011 Buna-N* .................... 6700-0133-500 (*) These parts are available as service kits. See
17. Timing Valve (2) ............................. 0207-8295-300 section 10.2.
10-1 10-1
10/Ordering Information
Posi-Pulse ISU 3
ISU 10
3
8
7
4
2 5
10
6 4
4
2
1
10-2 10-2
10/Ordering Information
10.2 Service Kits
1. Regulator Module Replacement 7b. ISU Backplate Replacement Kit,
Kit ............................................. 6700-0030-700 International ............................ 6700-0003-700
Includes the following parts: Includes the following parts:
O-ring, Stem Backplate Assembly
O-ring Gasket
Diaphragm ISU Mounting Plate
Stem Screw Screw (4)
Cap Screw (2)
8a. Posi-Pulse ISU Backplate Replacement
These parts are available only in kit form. If Kit, North American ................ 6700-0032-700
other parts are needed for replacement, the
entire Module must be ordered. (6700-1225- Includes the following parts:
Backplate Assembly
800)
Gasket
2. Complete Filter and O-ring Posi-Pulse ISU Mounting Plate
Replacement Kit ...................... 0221-5887-300 Screw (4)
Includes all O-rings (16) and Filters (5) for 8b. Posi-Pulse ISU Backplate Replacement Kit,
servicing the regulator International ............................ 6700-0033-700
3. Complete O-ring Replacement Includes the following parts:
Kit ............................................. 0210-0474-870 Backplate Assembly
Gasket
Includes all O-rings (16) for servicing the Posi-Pulse ISU Mounting Plate
regulator
Screw (4)
4. Complete Filter Replacement
9. Posi-Pulse Attachment Port
Kit ............................................. 0221-5884-870 Kit ............................................. 6700-0088-700
Includes all Filters (5) for servicing the
Includes the following parts:
regulator
Posi-Pulse Attachment Port (4)
5. Mode Selector Knob Replacement Attachment Port/Pause Module Tube (4)
Kit ............................................. 0221-5897-300
Includes the following parts:
Mode Selector Knob
Extension
Hardware
6. Suction Control Knob Replacement
Kit ............................................. 0221-5898-300
Includes the following parts:
Suction Control Knob
Threaded Stud
Hardware
7a. ISU Backplate Replacement Kit,
North American ....................... 6700-0002-700
Includes the following parts:
Backplate Assembly
Gasket
ISU Mounting Plate
Screw (4)
10-3 10-3
10/Ordering Information
10.3 Fittings and Adapters
6
13 14 15 16 17 18
7
10
11
12
10-4 10-4
10/Ordering Information
10.4 North American
10.4.1 Regulator Options
Instructions: Select the adapter, vacuum regulator, and fitting to form an 11-digit part number.
Example:
6701 6701-1251-901 = 6701 1224 1251 901 901
Diamond Intermittent Tubing
Diamond Standard
Adapter Tubing Nipple
Suction Unit (ISU)
Adapter Three Mode Nipple
6704 904
Schrader Locking Gland
905
6705 Locking Gland and
NCG Overflow Safety Trap
Rectangle
1252
Posi-Pulse ISU 906
Locking Gland and
6706 Overflow Safety Trap
NCG Round and 1/2 Gallon Bottle,
with Wall Slide
(Straight)
6707 907
Puritan- DISS Male
Bennett
908
6708 DISS Male & Overflow
OES Safety Trap, DISS
Wing Nut & Gland
6709
MedStar 909
DISS Male, Overflow
Safety Trap and 1/2
6711 Gallon Bottle, with
90° Tubing Wall Slide (Straight)
Nipple
10-5 10-5
10/Ordering Information
10.5 English Regulator Options
Instructions: Select the vacuum regulator and fittings to form an 11-digit part number.
Example: 6700-1253-901 = Intermittent Suction Unit (ISU) w/Tubing Nipple
6700-12XX-900 0221-2685-802
w/o Inlet Fittings Diamond
6700-12XX-901 0204-6717-800
w/Tubing Nipple Hand-I-Twist
6700-12XX-902 0221-0205-800
w/Tubing Nipple, NCG (Rectangle)
12" Tubing,
1/2 gal. Bottle,
6700-1253-9XX Cap & Float 0221-0204-802
Intermittent Assembly and Wall NCG (Round)
Suction Unit (ISU) Slide (Straight)
0221-0106-300
6700-12XX-904 Puritan
w/Locking Gland
6700-0051-700
6700-12XX-905 Drager
w/Locking Gland and
Overflow Safety Trap
6700-0057-700
BOC IV
6700-12XX-906
w/Locking Gland,
Overflow Safety Trap, 6700-0091-400
6700-1256-9XX 12" Tubing, 1/2 gal. Air Liquide
Posi-Pulse ISU Bottle, Cap & Float
Assembly and Wall
Slide (Straight) 0221-0163-300
MedStar
6700-12XX-911
w/Locking Gland, 0204-9020-535
Overflow Safety Trap, Tubing Nipple
12"Tubing, 1/2 gal.
Bottle, Cap & Float
Assembly and Wall 0221-0152-300
"V" Slide OES
(International)
0221-0690-731
6700-12XX-912 Schrader
w/Tubing Nipple,
12" Tubing, 1/2 gal.
Bottle, Cap & Float 6700-0094-700
Assembly and Wall DISS Nut &
"V" Slide Gland
(International)
6700-0095-700
90° Tubing
Nipple
10-6 10-6
10-7
10/Ordering Information
10.6 French Regulator Options
Instructions: Select the vacuum regulator and fittings to form an 11-digit part number.
Example: 6700-1254-901 = Intermittent Suction Unit (ISU) w/Tubing Nipple
6700-12XX-900 0221-2685-802
w/o Inlet Fittings Diamond
6700-12XX-901 0204-6717-800
w/Tubing Nipple Hand-I-Twist
6700-12XX-902 0221-0205-800
w/Tubing Nipple, NCG (Rectangle)
12" Tubing,
1/2 gal. Bottle,
Cap & Float 0221-0204-802
6700-1254-9XX Assembly and Wall NCG (Round)
Intermittent Slide (Straight)
Suction Unit (ISU)
0221-0106-300
6700-12XX-904 Puritan
w/Locking Gland
6700-0051-700
6700-12XX-905 Drager
w/Locking Gland and
Overflow Safety Trap
6700-0057-700
BOC IV
6700-12XX-906
w/Locking Gland,
Overflow Safety Trap, 6700-0091-400
12" Tubing, 1/2 gal. Air Liquide
6700-1257-9XX
Bottle, Cap & Float
Posi-Pulse ISU
Assembly and Wall
Slide (Straight) 0221-0163-300
MedStar
6700-12XX-911
w/Locking Gland, 0204-9020-535
Overflow Safety Trap, Tubing Nipple
12"Tubing, 1/2 gal.
Bottle, Cap & Float
Assembly and Wall 0221-0152-300
"V" Slide OES
(International)
0221-0690-731
6700-12XX-912 Schrader
w/Tubing Nipple,
12" Tubing, 1/2 gal.
Bottle, Cap & Float 6700-0094-700
Assembly and Wall DISS Nut &
"V" Slide Gland
(International)
6700-0095-700
90° Tubing
Nipple
10-8
10-7 10-7
10/Ordering Information
10.7 Spanish Regulator Options
Instructions: Select the vacuum regulator and fittings to form an 11-digit part number.
Example: 6700-1255-901 = Intermittent Suction Unit (ISU) w/Tubing Nipple
6700-12XX-900 0221-2685-802
w/o Inlet Fittings Diamond
6700-12XX-901 0204-6717-800
w/Tubing Nipple Hand-I-Twist
6700-12XX-902 0221-0205-800
w/Tubing Nipple, NCG (Rectangle)
12" Tubing,
1/2 gal. Bottle,
6700-1255-9XX Cap & Float 0221-0204-802
Intermittent Assembly and Wall NCG (Round)
Suction Unit (ISU) Slide (Straight)
0221-0106-300
6700-12XX-904 Puritan
w/Locking Gland
6700-0051-700
6700-12XX-905 Drager
w/Locking Gland and
Overflow Safety Trap
6700-0057-700
BOC IV
6700-12XX-906
w/Locking Gland,
Overflow Safety Trap, 6700-0091-400
6700-1258-9XX 12" Tubing, 1/2 gal. Air Liquide
Posi-Pulse ISU Bottle, Cap & Float
Assembly and Wall
Slide (Straight) 0221-0163-300
MedStar
6700-12XX-911
w/Locking Gland, 0204-9020-535
Overflow Safety Trap, Tubing Nipple
12"Tubing, 1/2 gal.
Bottle, Cap & Float
Assembly and Wall 0221-0152-300
"V" Slide OES
(International)
0221-0690-731
6700-12XX-912 Schrader
w/Tubing Nipple,
12" Tubing, 1/2 gal.
Bottle, Cap & Float 6700-0094-700
Assembly and Wall DISS Nut &
"V" Slide Gland
(International)
6700-0095-700
90° Tubing
Nipple
10-8 10-8
10-9
10/Ordering Information
10.4.2 Other Options (North American)
Instructions: Select the desired configuration and write down the first four and the last three digits
of the part number. Select the vacuum regulator desired (below) and insert the middle four digits in
the "XXXX" section of the selected 11-digit part number.
Example: 6713-1251-907 = Intermittent Suction Unit with Rail Bracket and DISS Male fitting
1251
Intermittent
Suction Unit (ISU)
Regulator and
Regulator with 30° Dovetail Bracket,
Fitting, Overflow Overflow Safety
Safety Trap Trap
6711-XXXX-930* 6711-XXXX-935*
* 1252 Posi-Pulse ISU is not available in this configuration. It must be mounted vertically.
10-9
10-6 10-9
10/Ordering Information
10.8 English, French and Spanish Other Options
Instructions: Select the desired configuration and write down the first four and the last three digits
of the part number. Select the vacuum regulator desired (below) and insert the middle four digits in
the "XXXX" section of the selected 11-digit part number.
Example: 6711-1253-913 = Intermittent Suction Unit (English) with MPTS (Slide) Bracket and
Overflow Safety Trap
1253
Intermittent
Suction Unit (ISU) Regulator & MPTS
(English) (Slide) Bracket,
Overflow Safety
Trap
6711-XXXX-913*
1254
Intermittent
Suction Unit (ISU)
(French)
1255
Intermittent
Suction Unit (ISU)
(Spanish)
* 1256, 1257, 1258 Posi-Pulse ISU is not available in this configuration. It must be mounted vertically.
10-10 10-10
Appendix
Installation procedure for
Adapters/Probes and Fittings.
All adapters/probes and fittings should be wCAUTION: Do not use any Loctite® products
sealed and installed properly to prevent leaks to seal the threads (or products which contain
and to support the equipment when mounted. Methacrylate Ester as an active ingredient).
Both vacuum regulator ports are 1/8-27 NPTF
tapered pipe threads. It is important to note The torque range for installing adapters/probes
that adapters/probes and fittings seal on the and fittings is 4.0 ft-lb (5.4 N-m) minimum to
thread and may have threads exposed after 10.0 ft-lb (13.6 N-m) maximum.
they have been tightened properly. Adapters/probes and fittings which are not
Prior to installing the adapter/probe or fitting, keyed for specific orientation, should be
seal the thread with 4Teflon® (PTFE) tape or one torqued to approximately 6.0 ft-lb (8.1 N-m).
of the following lubricants: Adapters/probes and fittings that are keyed to
®
Dow 111 (Ohmeda P/N 6700-0074-200) specific orientation, must be torqued initially to
4.0 ft-lbs. Additional torque is applied only until
5
Ball Vac-Kote (37951M) (Ohmeda P/N 0220- orientation is correct.
0091-300)
4
Teflon is a registered trademark of the DuPont Company.
5
Ball Vac-Kote is a registered trademark of the Ball Aerospace Systems Division.
A-1 A-1
Ohmeda L
Ohmeda
9065 Guilford Road
Columbia MD 21046 1801
USA
A Division of The BOC Group Inc
In the USA, please call Customer Service at 800 345 2700 for additional information or to place an order.
L Cvr Chg
Date
A BOC Health Care
Company 6700 0082 000
05 90 0 02 23 13
Critical Care Printed in USA
Worldwide © The BOC Group Inc