Maintenance Instructions
© S ie me ns A G 1 99 6
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
Register 7
Print No.: U3.LH300.101.01.01.02 16 December 1996
Replaces:
0-2 Revision
Page
1 _______Introduction ___________________________________________________ 1 - 1
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Preventive Parts Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Functional System Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Peripheral Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Concluding the Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Appendix: Mandatory System Modifications . . . . . . . . . . . . . . . . . . . . . .2-5
1 Introduction
1.2.1 Equipment
• Standard Field Service Tools
• Electrical Safety Analyzer
• Ultrasound Phantom (see ARTD)
• Formatted 3.5’’ HD diskette for saving system preset data
1.2.2 Materials
• 2 lamps for control panel lighting (P/N 48 30 977 LE300)
The same type of lamps is used in different SONOLINE systems.
• Lithium battery (P/N 99 06 496 LE007)
A battery in a system should be replaced every 3 years.
The same type of battery is used in different SONOLINE systems.
• printer cleaning sheets2 (depends on printer installed in a system)
1. Document “SONOLINE Prima Imaging System - Planned Maintenance Certificate”, U3.LH300.105.01.01.02 (or
newer).
2. See document “OEM Products & Accessories - Spare Parts List”, US-OEM.081.01.02 (or newer).
1. This document is a revised version of the former “TI 236” document, which is no longer valid.
2 Planned Maintenance
2.1 Preparations
WARNING When handling parts that may have had patient contact, take
appropriate precautions against exposure to bloodborne patho-
gens (see TI 246, “Prophylactic Measures Against Infectious Dis-
eases in Servicing”).
CAUTION Do not clean the transducer head with isopropyl alcohol or other
organic solvents.
Do not allow the transducer cable and connector to become really
wet.
6. While cleaning, check the transducer heads for physical damage such as cracks,
splits, air bubbles (sector transducers), or any other malfunction that may affect
their performance. Inspect the cable and the connection at the strain relief on both
ends of the cable.
7. Inspect all transducers and user accessible ports for damage, loose connectors,
bent pins, etc.
PM Clean the Air Flow Filter
8. Rinse the filter, using flowing cold water, to remove any lint. 1
Shake excess water from the filter, and allow it to dry before reinstalling.
1. You may also refer to the Operator’s Manual for more information on cleaning the system, the transducers, and the
filter.
CAUTION When working on the system with the cover removed, take appro-
priate precautions to prevent damage to static-sensitive parts due
to electrostatic discharge (ESD).
CAUTION To make sure that configuration data is not lost, do not switch off
the system while the battery pack cable is disconnected.
3. If the battery on the A21 Battery Pack PCB has been in the system for more than 3
years (see page 1 - 1):
- Apply power to the system,
- disconnect the cable from the A21 Battery Pack PCB to the A44 System
Controller board,
- replace the battery on the A21 Battery Pack PCB,
- write the replacement date onto the battery,
- and reconnect the battery pack cable.
(The cable may be connected in either polarity.)
4. Turn the system power off.
5. Replace the two lamps of the control panel lighting.
If necessary, insert a spare lamp.
2.4 Maintenance
PM Boards and cables
1. Check that all boards are properly seated and that all cables are properly
connected.
CAUTION Do NOT remove the battery pack cable from the A44 System Con-
troller board, or configuration data may be lost.
1. See document “SONOLINE Versa Imaging System - Repair Instructions”, U4.LH400.091.02.01.02 (or newer).
2. The message text depends on the selected language. With system software 2.2.0 or higher, automatic calibration
is performed only once a day (when the system is powered on for the first time).
3. See document “SONOLINE Prima Imaging System - Troubleshooting Guide”, U3.LH300.072.01.01.02 (for system
software 1.9.0 and lower).
4. See document “SONOLINE Prima Imaging System - Troubleshooting Guide”, U3.LH300.072.01.02.02 (or newer;
for system software 2.2.0 and higher).
1. See document “Video Imager Camera CCM 601 - Preventive Maintenance Instructions”, U0.LHOEM.101.01.01.02
(or newer).
2. See document “SONOLINE Versa Imaging System - Installation Instructions for Video Imager Camera CCM 601”,
U4.LH400.031.06.01.02 (or newer).
Mandatory
System Modification H/W S/W Completion Documentation Verification of Modification
Date
Verify that product status on PS
label is PS3B or PS3D or PS4.0 (or
Upgrade of A33 RBF Boards ■ 31 January 1997 UT 96-09 higher)
Verify that system SW version is
1.8.0 (or higher)
Maintenance Certificate
Customer:
Department: Room:
Address:
© S ie me ns A G 1 99 6
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
Register 7
Print Number: U3.LH300.105.01.01.02 16 December 1996
Replaces:
KS-Nr.: . . . . . . . . Maintenance Certificate
Siemens office
Address:
Region: Country:
Contact person: Telephone:
FSE in charge: Telephone:
Explanation of abbreviations
Abbreviation Explanation
SI Safety Inspection
PM Preventive Maintenance
Q Quality Check
Measurement devices
Comments
Open issues
System Status
The equipment has minor problems that do not restrict the use of it. The equipment is
okay to use. Siemens recommends scheduling service to repair the minor problems.
The equipment has major problems that restrict the use of it. The equipment is not okay
to use. Siemens strongly recommends that the equipment not be used until the problem
has been resolved.
Date
Signature
OK not n.a.
OK
Page
1 Introduction 1-1
1.1 Hints on Performing the Maintenance 1-1