Pasco $pectrometer Èxp
Pasco $pectrometer Èxp
SPECTROMETER
Introduction
In principle, a spectrometer is the simplest of scientific very sensitive detection and precise measurement, a real
instruments. Bend a beam of light with a prism or diffraction spectrometer is a bit more complicated. As shown in Figure
grating. If the beam is composed of more than one color of 1, a spectrometer consists of three basic components; a
light, a spectrum is formed, since the various colors are collimator, a diffracting element, and a telescope.
refracted or diffracted to different angles. Carefully measure
The light to be analyzed enters the collimator through a
the angle to which each color of light is bent. The result is a
narrow slit positioned at the focal point of the collimator
spectral "fingerprint," which carries a wealth of information
lens. The light leaving the collimator is therefore a thin,
about the substance from which the light emanates.
parallel beam, which ensures that all the light from the slit
In most cases, substances must be hot if they are to emit strikes the diffracting element at the same angle of inci-
light. But a spectrometer can also be used to investigate cold dence. This is necessary if a sharp image is to be formed.
substances. Pass white light, which contains all the colors of
The diffracting element bends the beam of light. If the beam
the visible spectrum, through a cool gas. The result is an
is composed of many different colors, each color is dif-
absorption spectrum. All the colors of the visible spectrum
fracted to a different angle.
are seen, except for certain colors that are absorbed by the
gas. The telescope can be rotated to collect the diffracted light at
very precisely measured angles. With the telescope focused
The importance of the spectrometer as a scientific instrument
at infinity and positioned at an angle to collect the light of a
is based on a simple but crucial fact. Light is emitted or
particular color, a precise image of the collimator slit can be
absorbed when an electron changes its orbit within an
seen. For example, when the telescope is at one angle of
individual atom. Because of this, the spectrometer is a
rotation, the viewer might see a red image of the slit, at
powerful tool for investigating the structure of atoms. It's
another angle a green image, and so on. By rotating the
also a powerful tool for determining which atoms are present
telescope, the slit images corresponding to each constituent
in a substance. Chemists use it to determine the constituents
color can be viewed and the angle of diffraction for each
of molecules, and astronomers use it to determine the
image can be measured. If the characteristics of the diffract-
constituents of stars that are millions of light years away.
ing element are known, these measured angles can be used to
In its simplest form, a spectrometer is nothing more than a determine the wavelengths that are present in the light.
prism and a protractor. However, because of the need for
EYE PIECE
TELESCOPE
RED LIGHT
COLLIMATOR
SLIT COLLIMATOR
ANGLE OF
LIGHT DIFFRACTION
SOURCE
GREEN LIGHT
PARALLEL BEAM
DIFFRACTION GRATING
(OR PRISM)
COLLIMATOR
VERNIER
TABLE ROTATION,
VERNIER FINE ADJUST /
MAGNIFYING GLASS LOCK-SCREW
FOR READING VERNIER
2. Focus the telescope at infinity. This is best accom- 1. The telescope and collimator pivot about a fulcrum on
plished by focusing on a distant object (e.g.; out the their respective mounting pillars (see Figure 4). Use the
window). aluminum rod provided with the accessory equipment to
adjust the leveling screws. Loosen one as the other is
3. Check that the collimator slit is partially open (use the tightened until the unit is level and secure.
slit width adjust screw).
4. Align the telescope directly opposite the collimator as
shown in Figure 3.
FULCRUM
TELESCOPE COLLIMATOR
LEVELING
SCREWS
MOUNTING
PILLAR
190 170
180
20 15 10 5 0
12' 30" 172° 20'
172° 20' + 12' 30" = 172° 22' 30"
Caution: The Diffraction Grating is a delicate Important: Stray light can obscure the images. Use
component. Be careful not to scratch the surface the spectrometer in a semi-darkened room or drape a
and always replace it in the protective foam sheet of opaque material over the spectrometer.
wrapping when it is not being used.
Perform steps 6-9 with reference to Figure 8.
Maintenance
Periodically clean the telescope aperture, the collimator Important: Always handle the spectrometer and
aperture, and the prism with a nonabrasive lens paper its accessories with care to avoid scratching the
(available at any camera store). No other regular mainte- optical surfaces and throwing off the alignment.
nance is required. Also, when not in use, the spectrometer should be
stored in its hardwood case.
Appendix: Using the Gaussian Eyepiece
The optional Gaussian eyepiece (Model SP-9285) simplifies 9. Rotate the spectrometer table 180 ° and, using the table
the task of aligning and focusing the spectrometer and rotation fine adjust knob, align the vertical cross-hair
aligning the diffraction grating. One Gaussian eyepiece can with the reflected image.
be used to align and focus any number of spectrometers, so
10. Adjust the table leveling screws to remove half the
only one is generally needed per lab.
separation between the horizontal cross-hair and the
reflected image. Adjust the telescope leveling screws to
remove the remaining error, so the cross-hairs and their
To Align and Focus the Spectrometer reflected images are superimposed.
Using the Gaussian Eyepiece:
11. Repeat steps 9 and 10 until the cross-hairs and their
1. Remove the telescope eyepiece and replace it with the reflected images are superimposed from both sides of
Gaussian eyepiece. the diffraction grating.
2. While looking through the telescope, slide the eyepiece 12. Unplug the Gaussian eyepiece. Adjust the slit of the
in and out until the cross-hairs come into sharp focus. collimator so it is open and vertical.
Loosen the graticule lock ring, and rotate the graticule 13. Illuminate the slit with an external light source. Rotate
until one of the cross-hairs is vertical. Retighten the lock the telescope directly opposite the collimator and focus
ring and then refocus if necessary. the collimator only (do not disturb the telescope focus)
3. Plug in the power supply of the Gaussian eyepiece. until the illuminated slit is in sharp focus. If the collima-
The light from the eyepiece is reflected along the optical tor slit is not vertical, loosen the lock ring, align the slit
axis of the telescope by a half-silvered mirror. Looking vertically, and then retighten the lock ring. Then align
through the eyepiece, you'll see the cross-hairs lighted the fixed edge of the slit with the vertical cross-hair.
up as they scatter some of the light back into the 14. Adjust the collimator leveling screws until the slit is
eyepiece. vertically centered in the field of view of the telescope.
4. Mount the grating holder to the spectrometer table and (As with the telescope, you may need to adjust the
insert the diffraction grating. collimator so that its optical axis is square to the axis of
rotation.) The telescope, collimator, and spectrometer
5. Looking through the telescope, rotate the table until a table are now properly aligned.
patch of light is reflected back through the telescope
from the glass surfaces of the grating. The spectrometer 15. If you are going to use the grating, plug the Gaussian
table and the telescope must be at least roughly level to eyepiece back in and rotate the spectrometer table until
achieve this reflection. If they are not, see Realigning the vertical cross-hair is again aligned with its reflected
the Spectrometer, earlier in the manual. image. This insures that the grating is perpendicular to
the optical axis of the spectrometer.
6. Adjust the focus of the telescope until the cross-hairs
and their reflected images are in sharp focus. The glass 16. If you wish, you may replace the Gaussian eyepiece
slides of the grating are not efficient reflectors, so you with the original eyepiece. The focus of the telescope
must look carefully to see them. will be maintained if you slide in the original eyepiece
until the cross-hairs are in sharp focus.
IMPORTANT: The grating is sandwiched
between two glass slides so, depending on how
parallel the slides are, you may see as many as four Alignment Error
reflected images of the cross-hairs. In the following
The multiple reflections from the glass slides of the grating
steps, you will be instructed to superimpose the
introduce some error into the alignment procedure. Nor-
graticule with its reflected image. If there is more
mally, centering the cross-hairs between the reflected images
than one image, just center the cross-hairs as
will reduce the error below the 30-second resolution that is
accurately as possible between the images.
obtainable when reading the vernier scales.
7. Use the table rotation fine adjust knob to align the To verify the alignment, use a light source with discrete
vertical cross-hair with its reflected image. spectral lines such as a sodium or mercury vapor lamp. If
the alignment is correct, corresponding spectral lines on
8. Adjust the spectrometer table leveling screws until both opposite sides of the optical axis will have equal angles of
cross-hairs are superimposed on the reflected image. diffraction. If necessary, adjust the rotation of the spectrom-
eter table until the measurements are the same.
Using the Diffraction
Grating in a Student
Spectrometer
f a sodium lamp is used, the
doublet may be viewed when
θ 10.5° and θ2 21°.