Clarity User Guide
Clarity User Guide
Code/Rev.: M021/90H
Date: 2024-05-14
DataApex Ltd.
Phone: +420 251 013 400 Petrzilkova 2583/13
clarity@dataapex.com 158 00 Prague 5
www.dataapex.com Czech Republic
Clarity®, DataApex® and ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are
trademarks of Microsoft Corporation.
DataApex reserves the right to make changes to manuals without prior notice. Updated manuals can be
downloaded from www.dataapex.com.
Author: DR
User Guide Table of Contents
Contents
1 Installation 1
1.1 Installing Clarity Software 1
1.2 Updating Clarity Software 6
1.3 Installing Colibrick 13
1.4 Installing a Rockey USB dongle 14
1.5 Connecting a chromatograph analogue output to Clarity 16
1.6 Installing Multicom 17
2 Configuring the Chromatography Station 20
2.1 Obtaining information about Clarity configuration 20
2.2 Setting the number and type of instruments 21
2.3 Adding a new device 22
2.4 Configuring a device 24
2.5 Setting the options for sending the method to the instrument 26
2.6 Assigning digital input and output to start acquisition 27
2.6.1 Data Inputs & Outputs 28
2.6.1.1 External Start Digital Input 28
2.6.1.2 Ready Digital Output 29
2.6.1.3 Settings Examples 29
2.7 Setting a custom image and a name for an instrument 30
2.8 Tablet mode 30
3 Configuring the User Accounts 33
3.1 Configuring the User Accounts 33
3.2 Creating a new user account 34
3.3 Deleting a user account 35
3.4 Sharing user settings among users 36
3.5 Restricting access 37
3.6 Setting a password for the first time 38
3.7 Changing a user password 39
3.8 Logging in without a password 40
4 Proposed workflow for routine analysis 41
5 Method Setup 42
5.1 Setting up a method 42
5.2 Setting up a slow flow rate increase and decrease on your LC pump 47
5.3 How to make a mark in a chromatogram during acquisition 50
6 Data Acquisition 53
6.1 Running a single analysis 53
-i-
User Guide Table of Contents
- ii -
User Guide Table of Contents
- iii -
User Guide Table of Contents
- iv -
User Guide Table of Contents
-v-
User Guide Table of Contents
To facilitate the orientation in the User Guide manual and Clarity chromatography station, different
fonts are used throughout the manual. Meanings of these fonts are:
Open File (italics) describes the commands and names of fields in Clarity, parameters that
can be entered into them or a window or dialog name.
WORK1 (capitals) indicates the name of the file and/or directory.
ACTIVE (capital italics) marks the state of the station or its part.
Chromatogram (blue underlined) marks clickable links referring to related chapters.
The bold text is sometimes also used for important parts of the text and the name of the Clarity
station. Moreover, some sections are written in format other than normal text. These sections are
formatted as follows:
- vi -
User Guide 1 Installation
1 Installation
Topics covering installation of Clarity software, Colibrick, Multicom, etc. Also the
connection between Clarity and chromatograph is explained.
3. Confirm the License Agreement. It is possible to continue only in case you agree
with the statement.
-1-
User Guide 1 Installation
4. Choose the destination folder. The user must have Read/Write/Modify access to
the installation directory.
5. Set location for your data files. C:\CLARITY\DataFiles is set by default. Notice
that Data location folder name cannot contain following characters / : * ? " < > |
and also cannot start or end with a space and cannot end with a dot.
-2-
User Guide 1 Installation
6. Enter the User code corresponding to your hardware key or select Enter User
Code later to start 30-day Trial.
Note: The user code can be found on the back side of the card provided with
the Installation USB. Alternatively, you can contact DataApex support to
obtain one.
7. Select the type of installation. Make sure that control module for your device is
selected to be installed.
Note: In most cases "Typical" should be selected. "Custom"/"Full" installation is
necessary e.g., for Agilent and other devices controlled via ICF, DANI
devices and few others.
Note: List only contains names of original devices. If you can't find your device,
even though it is stated as controlled, it is most likely an OEM version of
different device.
-3-
User Guide 1 Installation
8. Select the Start Menu folder for the shortcut or create a new one. After clicking
Install installation process will start.
9. Wait for installation to finish. In the end you might be prompted to confirm
installation of several hardware drivers.
10. When installation is finished Notes and Tools window will be opened.
-4-
User Guide 1 Installation
11. Finally the last window will offer you several actions which may be executed after
the installation is completed.
More Info:
l The Launch Manager allows you to start Clarity using different profiles that
correspond to different configurations and combinations of instruments,
projects, methods, etc. For example, different users or groups of users can
have the chromatography station configured on the same computer.
l The Installation Qualification - IQ (Make IQ report option) is a procedure that
confirms that the software has been installed successfully and that the files
are in the correct version.
-5-
User Guide 1 Installation
4. Confirm the License Agreement. It is possible to continue only in case you agree
with the statement.
-6-
User Guide 1 Installation
Note: Selecting Install to different location will result in presence of two different
Clarity version at once (current version will be left as it is and new version
will be installed elsewhere).
-7-
User Guide 1 Installation
-8-
User Guide 1 Installation
9. Wait for uninstall to finish and click Finish in the following window.
-9-
User Guide 1 Installation
11. Set destination folder for your data files. If data location is changed all data from
current version will be moved accordingly.
12. Select type of installation. Installation type which was used for current version
will be preselected. Make sure that control module for your device is selected to
be installed
Note: In most cases "Typical" should be selected. "Custom"/"Full" installation is
necessary e.g., for Agilent and other devices controlled via ICF, DANI
devices and few others.
Note: List only contains names of original devices. If you can't find your device,
even though it is stated as controlled, it is most likely an OEM version of
different device.
- 10 -
User Guide 1 Installation
13. Select the Start Menu folder for the shortcut or create a new one. After clicking
Install installation process will start.
- 11 -
User Guide 1 Installation
16. Finally the last window will offer you several actions which may be executed after
the installation is completed.
More Info:
l The Launch Manager allows you to start Clarity using different profiles that
correspond to different configurations and combinations of instruments,
projects, methods, etc. For example, different users or groups of users can
have the chromatography station configured on the same computer.
l The Installation Qualification - IQ (Make IQ report option) is a procedure that
confirms that the software has been installed successfully and that the files
are in the correct version.
Caution: When upgrading from Clarity 6.2 or older to Clarity 7.0 or newer be aware
that there is change in the installation structure - the content of the
original installation folder is separated to three new subfolders BIN, CFG
and DataFiles. The respective files are moved automatically during
update to the new locations if you selected Update existing installation. In
the rare cases this fails, some files may need to be moved manually.
- 12 -
User Guide 1 Installation
3. Connect the CANNON SUB D 27-pin connector on the (INT7) cable to Colibrick
back panel.
- 13 -
User Guide 1 Installation
4. Connect the cables (Starting, Digital output and Analog Signal) to the
chromatograph as explained in Connecting a chromatograph analogue output to
Clarity.
5. Start Clarity and then add the Colibrick channels to specific Clarity instruments
as explained in Adding a new device.
6. Check the LED's on the front panel to find out about the status of Colibrick and
whether it has been installed properly.
More Info:
l Ready (orange) LED status Indicates correct installation.
l Data (blue) LED status Indicates connection to the chromatography data
station.
l Digital Input (green) LEDs status
l LED ON - the input status is High (logical "1") or not connected.
l LED OFF - the input status is Low (logical "0") or connected to the
ground (GND).
l Digital Output (red) LEDs status
l LED ON - the output status is High (logical "1"), the relay contact is
opened.
l LED OFF - the output status is Low (logical "0"), the relay contact is
closed.
- 14 -
User Guide 1 Installation
5. Verify that the driver has been installed correctly. Meaning that the Device
Manager ⓐ has the item "Universal Serial Bus Controllers" - "Rockey4 USB"
ⓑ.
- 15 -
User Guide 1 Installation
If this does not work, try the following procedure after the installation of Clarity:
After connecting the dongle, Windows will detect a new Plug and Play device and
the Found New Hardware Wizard will appear.
1. Select "Search for a suitable driver for my device."
2. Select "Specify a location" and then select the C:\CLARITY \BIN\HW_
DRIVERS\ROCKEY\folder. The rest of the installation will be carried out
automatically.
Note: On Windows 10 and Windows 11, the driver for older HW keys (Rockey4USB.sys)
can only be installed if Memory Integrity/Core isolation function of Windows
Security is turned off. This function is turned on by default. If an incompatible
driver is already installed on the computer, it cannot be turned on. But if it is
enabled, the driver installation fails with Error code 39.
- 16 -
User Guide 1 Installation
1. Install your external A/D converter like Colibrick (refer to Installing Colibrick) or
an A/D converter card (refer to the specific HW manual).
2. Switch off your chromatograph.
3. Connect the bare wires to the chromatograph depending on your equipment,
configuration and the following guidelines.
More Info:
l The Signal cables "DET 1" to "DET 4" carry the main signal from the
chromatograph to the computer. The connection can be asymmetrical or
symmetrical.
l The Starting cables "IN1" to "IN4" come in pairs, one part connected to the
27-pin connector and ending on a female RCA connector and the other with a
male RCA connector and free leads for connection to a starting contact or a
button for a manual start.
l The Digital Output cables "OUT 1R" to "OUT 4R" end on free leads and they
can be used for synchronizing autosamplers.
Caution: The shielding must be connected. It works not only as the shielding, but also as
the analogue ground against which measurement takes place. In the case of asymmetrical
output of a detector (only two leads/terminals/pins/screws) the shielding must be
connected to the white lead! No lead of the signal cable may remain unconnected.
4. Connect the CANNON SUB D 27-pin connector to the A/D convertor.
5. Switch on your computer and your chromatograph.
- 17 -
User Guide 1 Installation
- 18 -
User Guide 1 Installation
6. Check the LEDs on the top panel to find out about the status of Multicom and
whether it has been installed properly.
More Info:
Green LED status:
l OFF – not connected to USB, the driver is not installed or Multicom is in
suspend mode.
l ON (Constant) – idle state, no communication.
l BLINKING:
l Two short consecutive blinks – only sending data from USB to
the COM port.
l Turned off twice consecutively – only receiving data from the
COM port into USB.
l Constant blinking – both sides are receiving and sending data.
- 19 -
User Guide 2 Configuring the Chromatography Station
- 20 -
User Guide 2 Configuring the Chromatography Station
- 21 -
User Guide 2 Configuring the Chromatography Station
2. If the device you want to add is not in the list, click on Add ⓐ and the following
window will open. Here you can filter the list by typing some text in the filter field
- 22 -
User Guide 2 Configuring the Chromatography Station
4. Add new devices to the appropriate instrument: drag and drop the device from
the left pane ⓓ to the Instrument pane on the right ⓔ or select the device and
click on ⓕ.
- 23 -
User Guide 2 Configuring the Chromatography Station
- 24 -
User Guide 2 Configuring the Chromatography Station
- 25 -
User Guide 2 Configuring the Chromatography Station
- 26 -
User Guide 2 Configuring the Chromatography Station
If you select the Send Method to Instrument option, the method selected in Single
analysis window will be sent every time you log into the instrument or this method is
modified, or when the sequence is finished.
- 27 -
User Guide 2 Configuring the Chromatography Station
3. Clarity gets the digital input marked as START from device over:
a. Communication line (LAN, serial (RS-232), USB/GPIB)
b. Wire as TTL signal (this connection requires an external A/D converter or
other device offering digital input to Clarity)
The last case (3) is most typical way and this chapter describes how to assign
digital input and output in Clarity for the most common wiring of devices (typically an
autosampler). The setting is performed in System Configuration dialog through Ext.
Start Dig. Input and Ready Dig. Output functions.
- 28 -
User Guide 2 Configuring the Chromatography Station
For some chromatographs, the settings of the module and its behavior are set in the
module's configuration based on the actual wiring, and it is upon the person
installing Clarity and setting up the configuration to set the whole system correctly.
For correct option always consider the source of the analysis start first (sampler,
sampling valve, Start button on a instrument) and ask a question "What will happen
if Clarity does not signal anything? Will the instrument run anyway?"
l If Yes - option should be set to This Device Starts the Run in Clarity
l If No - option should be set to Clarity Starts This Device
Ready Dig. Output defines the device and its specific pin through which Clarity
informs other parts of the system that injection can be performed. After starting a
sequence, controlled modules READY states are verified, Clarity triggers the
Ready Dig. Output and changes its state, thus broadcasting to other modules of the
system that sequence can be run. READY signal is received by the sampler that
performs the injection and sends START signal. Then Clarity detects START signal
and triggers again Ready Dig. Output and change its state, thus preventing the
sampler to perform another injection.
Caution: Using the Event Table in Method Setup other actions can be configured changing
the parameters of Digital Output. Note, that if the same Digital Output is modified, it
may cause conflicts in synchronization. Avoid using these inputs and outputs in the
Event Table.
- 29 -
User Guide 2 Configuring the Chromatography Station
5. Check the Custom Image check-box and then on to select your image. Click
OK to save your changes.
6. Repeat for every Instrument you wish to change the name or the pictures.
- 30 -
User Guide 2 Configuring the Chromatography Station
application. It is designed for devices with small displays. The optimum resolution is
1920 x 1080.
In Tablet mode, Instrument window is narrower than in standard layout and is
positioned on the left side of the monitor, other windows open on top of each other
and fill the remaining space on the monitor. Tablet mode allows the use of higher
scale in Windows (up to 200 %) which improves work with the software and
readability of parameters.
2. After logging in to the Instrument, new windows open in tablet layout. Instrument
window is positioned to the left side of the monitor, other windows open on top of
each other and fill the remaining space on the monitor.
More Info:
Instrument window cannot be fully maximized. Its width can be expanded up to a
maximum of 50% of the monitor width and is stored.
Other windows always fill the remaining space of the display.
Method Setup and Single Analysis windows are opened maximized.
If the scale in Windows is set to more than 100%, some of the icons and analysis
status line can break into a new line.
- 31 -
User Guide 2 Configuring the Chromatography Station
- 32 -
User Guide 3 Configuring the User Accounts
- 33 -
User Guide 3 Configuring the User Accounts
- 34 -
User Guide 3 Configuring the User Accounts
- 35 -
User Guide 3 Configuring the User Accounts
- 36 -
User Guide 3 Configuring the User Accounts
1. Open the User Accounts dialog: click on or choose System - User Accounts.
2. Configure the file access rights for other users ⓐ .
3. Check the instruments the user will have access to ⓑ .
4. Check the Clarity procedures the user will have access to ⓒ .
5. Click OK to accept the changes.
- 37 -
User Guide 3 Configuring the User Accounts
2. Type in and confirm the new password and then click OK.
- 38 -
User Guide 3 Configuring the User Accounts
3. Alternatively or if you are creating a new user, you can follow the same
procedure as explained in section Changing a user password and create a new
one.
- 39 -
User Guide 3 Configuring the User Accounts
- 40 -
User Guide 4 Proposed workflow for routine analysis
- 41 -
User Guide 5 Method Setup
5 Method Setup
Following chapters include useful tips regarding the Method Setup and working with
method.
- 42 -
User Guide 5 Method Setup
Send Method
l Only a method that is saved can be sent.
l Pressing the Send Method button will result in two actions:
l Method will be sent to all connected hardware and thus
displayed in the information table of the Instrument window -
just as before.
l Method will be set as Method for Single Analysis - you can
start the single analysis form the instrument window using the
icon.
l When measuring sequence, each row can have different method. It is set
while creating or editing the sequence tab in the Method Name column.
Method Development
Method currently used for acquisition or present on already locked row of the
sequence can't be edited, and can be viewed in read-only mode only.
When developing a method, you can configure the following settings among others:
1. Set the measurement conditions.
Here you can disable or enable Autostop and set the run time for the
analysis ⓐ as well as configure the external signal start and stop settings
ⓑ (default setting Start Only should be used in majority of applications).
Autostop can be set within interval 0.2 - 9999 min, such value is not
influenced and does not influence any device specific time program settings
on different tabs (like LC Gradient or GC). When time defined in Autostop
passes, acquisition is finished, chromatogram is created, remaining time
from time programs will be spent in so called Control state.
You can also describe some parameters of the method and add a note ⓒ .
Such information is purely informative and do not influence analysis in any
way. They are saved into the resulting chromatogram and are showed in the
Chromatogram window.
- 43 -
User Guide 5 Method Setup
- 44 -
User Guide 5 Method Setup
- 45 -
User Guide 5 Method Setup
Note: Integration parameters are often modified manually after the first acquired
chromatogram directly in the Chromatogram window. Then you can manually
rewrite them into any method in the Method Setup dialog or you can use the
Method - Save as Template menu command in the Chromatogram window to
easily copy them into your method. For more details see the chapter Saving the
chromatogram method as a template method on pg. 74..
4. Set the calculation options.
l
You can create New... calibration file or Set... the one created previously ⓐ
and configure the different settings related to it.
l When calibration file is set, resulting chromatograms will be calibrated
according to it and calibration standards will be used to automatically re-
calibrate the attached calibration file.
l
Clone... ⓑ button will create a copy of a current calibration file. In the
following Save As dialog you can select the name and location of the new
calibration file. The copy will then be linked to the method.
l If it is desired to have no calibration in the method you can use the None
button ⓒ .
l In this tab you can also choose how you want the results to appear in the
results ⓓ and set calculation type ⓔ .
l
Name of calibration clone ⓕ must be set here to use calibration cloning in
sequences, refer to chapter "Calibrating using clone on first recalibration" on
page 96.
- 46 -
User Guide 5 Method Setup
Note: For more information on any of the settings above, press F1 in such Clarity
window to go the corresponding section of the Help or use the Clarity Reference
Guide.
- 47 -
User Guide 5 Method Setup
Startup Method
1. From the Instrument window open the Method Setup window using the Method -
Analysis Method
7. Repeat steps 2 to 6 but this time:
l Save the Method under the name Analysis.
l Set the Idle State to Initial ⓓ .
l Set the Initial Flow, Flow on the second row ⓔ and Standby Flow
value ⓕ to the value of Standby Flow from Startup method.
l Set the Time to the total duration of your analysis ⓖ .
- 48 -
User Guide 5 Method Setup
Shutdown Method
8. Repeat steps 2 to 6 but this time:
l Save the Method under the name Shutdown.
l Set the Idle State to Initial - Standby ⓗ .
l On the second row, set the Flow to 0 and the Time to obtain the
appropriate flow rate decrease for your column ⓘ .
l Set the Standby Flow to 0 ⓙ .
- 49 -
User Guide 5 Method Setup
Sequence
9. Set up a Sequence as explained in Running a Sequence and:
l On the first row set the Sample Type column as Bypass and the
Method Name to Startup.
l On the last row set the Sample Type column as Bypass and the
Method Name to Shutdown.
l Set as many rows as you need in between according to the
conditions of your analysis or sequence and set the Method Name to
Analysis.
- 50 -
User Guide 5 Method Setup
labeled with name and RT) in the Data Acquisition and Chromatogram windows.
The digital input can be triggered by pressing the start button when using HW or by
pressing the corresponding Digital Output in the Device Monitor when using the
Virtual Digital Input Output Loop module.
The following settings are necessary:
l HW with a digital input or the Virtual Digital Input Output Loop module configured
to an Instrument.
l Method Setup
l Measurement tab - External Start/Stop enabled and set to Start Only
l Event Table tab - set according to the image below, where:
l Name = the label of the event
l Source = the module providing the digital input
l Input = digital input recording the event
l To display the marks, enable the Show Events option in the Graph Properties
dialog accessible from the Data Acquisition and Chromatogram windows (the
dialog can be invoked by right-clicking the graph area in the respective window
and clicking Properties...).
- 51 -
User Guide 5 Method Setup
- 52 -
User Guide 6 Data Acquisition
6 Data Acquisition
Chapters describing how to perform a measurement using Single Analysis or
Sequence and how to evaluate chromatograms during run.
6. Start the analysis by clicking the Run button ⓓ . Analysis can also be triggered
by External Start from the chromatograph. The Instrument will get to the
- 53 -
User Guide 6 Data Acquisition
RUNNING state and the Single Analysis dialog will close. The state is visible in
the status line in the Instrument window.
7. It is possible to monitor the analysis from Data Acquisition window, see Pre-
evaluating a chromatogram during acquisition for more.
- 54 -
User Guide 6 Data Acquisition
resulting name. For more info regarding variables see "Chromatogram File
Name" in the Clarity Reference Guide.
7. The SV (Starting Vial) and EV (End Vial) rows ⓕ will be pre-filled with numbers
corresponding to the sample position in the autosampler tray, if you are using an
autosampler. It is possible to change these as needed, injection from multiple
vials can be made on one sequence row.
8. Fill in volume of the injection in the Inj. Vol. column ⓖ .
9. Select the method to be used in the Method Name column.
10. Tick any of the Open, Open Calib., Print etc., columns ⓗ in case you wish to
open measured chromatogram in chromatogram or calibration window or print
the results after each measurement of a sample.
Caution: To correctly include chromatograph graphs in the reports it is necessary
to check both Open and Print checkboxes. It is also necessary to fill in
Report Style when Print or Print to PDF is used.
11. Repeat the steps 4-10 for rows you need to add to the Sequence Table.
Note: It is possible to display additional columns by right-clicking the sequence
table, selecting Setup Columns and moving items from Hide Columns to
Show Columns list by Show button.
12. Check the validity of the sequence by selecting Sequence - Check Sequence or
clicking on icon. In valid sequences, all rows will show the symbol in
the Status column. Invalid sequences will issue a warning message with the
cause of the problem.
13. Save the sequence by selecting File - Save or clicking on icon.
14. Run the prepared sequence by selecting Sequence - Run or clicking on icon
ⓘ.
Note: The sequence state will change to WAITING FOR INJECTION or
INJECTING state, depending on READY state of other controlled
modules. The state is visible at the bottom of the Sequence window.
- 55 -
User Guide 6 Data Acquisition
without lamp on. Proper combination of the possible actions needs to be set
accordingly.
1. To setup shutdown open the Sequence Options in your opened sequence by
selecting Sequence - Options or clicking on icon.
2. Now, you need to decide whether you only need to send shutdown method to
controlled devices to prepare them for shutdown (step 3), or if you need also to
run such method, e.g. for lowering the gradient (steps 3 and 4). As mentioned in
the beginning of chapter this setting is highly depended on individual control
module capabilities.
3. Send shutdown method: checkbox: Check this checkbox to be able to select and
send shutdown method to controlled devices. You can also edit such method, if
needed. This option will only send the selected method to devices.
4. Run shutdown method checkbox: Use this option to run selected method.
Resulting chromatogram will be saved to the current project as
[SEQUENCENAME - SHUTDOWN - METHODNAME - %R.PRM], where %R
stands for current date and time.
Note: Use Bypass to omit injection from the vial, or Unknown for injection from
specified vial. Valid Vial position and Injection volume may be required by
some autosamplers, even if the injection will not be performed (i.e. when
method is not run or Bypass injection is selected).
5. Perform shutdown checkbox: Use this option to shutdown all controlled devices.
Shutdown command is sent after sending and completed run of the shutdown
method. Note that the reaction of some devices to the Shutdown may be
amended in their respective Setup window in System Configuration.
- 56 -
User Guide 6 Data Acquisition
Note: The parameters of the After sequence is finished section are also executed when
the user stops the sequence by Stop command or the sequence is stopped by
an error. If it is necessary to stop the running sequence without sending (and
running) the shutdown method, use the Abort command.
Note: When the sequence is already finished and the shutdown is also already
performed, if new lines are added to the finished sequence and such sequence is
then resumed, the whole After sequence is finished will be applied again.
- 57 -
User Guide 7 Device Monitor
7 Device Monitor
The Device Monitor window serves firstly to show the current status of each control
module and secondly for direct control of the device. The layout and possible
control options greatly vary based on the used control modules.
The window can be personalized. To change the order of individual monitors right
click on the module’s name ⓐ and choose desired action in the context menu ⓑ .
The individual monitor panel can be collapsed or expanded using the arrow icon ⓒ
in the header of the panel.
- 58 -
User Guide 7 Device Monitor
- 59 -
User Guide 7 Device Monitor
Further commands are available for selected pumps (typically those controlled
by Clarity in real time). They are enabled only during runs with a gradient
program.
l Hold halts the gradient at its current state. The button changes to Resume which
can be used to resume the gradient from the point where it was halted.
l Modify Gradient invokes the LC Control Manual Flow dialog (similar to the
Method Setup - LC Gradient) which can be used to adjust the gradient for the
current analysis. This does not influence the method used for analysis in any
way.
- 60 -
User Guide 8 Chromatogram Editing
8 Chromatogram Editing
This chapter focuses mainly on adjusting integration. For proposed general
workflow see chapter How to integrate chromatogram
Clarity Tips&Tricks videos covering Integration topics can be found in Clarity
Integration playlist.
- 61 -
User Guide 8 Chromatogram Editing
3. Set Global Peak Width parameter. Choose the narrowest peak which should
be integrated in your chromatogram and select his start and end. This setting
is used for calculation of Global Bunching.
4. Set Global Bunching parameter. This is a filter which is calculated based on
Global Peak Width and sample rate.
a. Alternatively, for chromatogram with high noise, it is possible to use
other filters - FFT Filter, Savitzky- Golay Filter or Moving Average
Filter.
5. Set Global Threshold parameter. This is influenced by the Global Bunching
or other filter set previously. Select an interval containing only noise, not
peaks. A more detailed description of the global parameters is covered by
topic Modifying Global parameters.
- 62 -
User Guide 8 Chromatogram Editing
6. (Optional) Set Detect Negative parameter. This has two potential uses, firstly
when you want to integrate also negative peaks. Secondly it may help when
signal has inconsistencies such as dips. This will prevent the integration
algorithm from placing peak starts/ends into them.
- 63 -
User Guide 8 Chromatogram Editing
7. Set Global Baseline Slope parameter. Change this parameter after you have
all the peaks integrated, it servers to optimize their start/end positions.
Note: Global Slope is available only in Wave Integration Algorithm.
- 64 -
User Guide 8 Chromatogram Editing
8. If you still aren't satisfied with the integration there are another possibilities
for manual changes.
a. Try to use Interval parameters such as Baseline Together, Baseline
Valley, Forward Horizontal etc..
b. Or try to use Peak parameters such as Peak Start, Peak End, Peak
Both, Add Positive etc..
Note: Add Positive operation should be used as a last resort because is
prone to every Retention time shifts. Peak Start and Peak End should
be preferred parameters as they are defined relative to peak apex.
9. When you are happy with the integration. Save method as Template and use
it for further measurements of these sample types.
10. If you have already measured some chromatograms and want to use this
integration in them you can do it using Batch - Reprocess by Method.
- 65 -
User Guide 8 Chromatogram Editing
2. To set the interval on which to apply the integration use Chromatogram -
Integration - Integration Interval or icon ⓐ . It is possible to set multiple
intervals within a single chromatogram.
3. You can check, modify the interval or delete this operations in the Integration
Table ⓑ .
Note: If you want to exclude specific interval from the integration refer to Remove peak
from integration.
- 66 -
User Guide 8 Chromatogram Editing
Global Threshold:
l
To adjust minimum height use Chromatogram - Global Threshold or use
icon ⓐ and select an interval containing only noise, not peaks.
Global Bunching:
l Works as a smoothing filter that automatically averages number of data
points in order to preserve at least 30 data points per narrowest peak. It is
dependent on Global Peak Width value. To use it select Chromatogram -
Global Bunching or use the icon ⓐ after you done adjusting the Global
Peak Width parameter.
Global Slope:
l To adjust minimum slope, where the peak should start/end use
Chromatogram - Global Slope or use icon ⓐ and select the point where
the peak should start/end.
- 67 -
User Guide 8 Chromatogram Editing
If you are still not happy with the integration suggested by selection from the graph
you can modify the parameter values in the Integration Tableⓑ .
Note: Global bunching is a filter that is always applied to the whole chromatogram. To
filter data locally use other filers from Chromatogram - Integration menu.
- 68 -
User Guide 8 Chromatogram Editing
- 69 -
User Guide 8 Chromatogram Editing
Fig. 1: Difference between Peak Hide (left) and Baseline Lock (right).
Note the two vertical guidelines marking the beginning and end of the peak. More
than one peak can be removed from integration at once.
- 70 -
User Guide 8 Chromatogram Editing
3. You can check, modify the position or delete this operation in the Integration
Table ⓒ .
Note the vertical guideline indicating the currently selected peak separator position.
- 71 -
User Guide 8 Chromatogram Editing
1. Select the rider peaks you want to separate on the leading side of the mother
peak: use Chromatogram - Baseline - Front Tangent or icon ⓐ .
2. Select the rider peaks you want to separate on the trailing side of the mother
peak: use Chromatogram - Baseline - Tail Tangent or icon ⓐ .
3. You can check, modify the interval or delete this operations in the Integration
Table (Baseline - Front tangent and Baseline - Tail tangent rows) ⓑ .
Note: You can also use the Tangent Slope Ratio and Tangent Area Ratio from
Chromatogram - Separation menu to set a threshold for the separation of rider
peaks based on those two parameters.
- 72 -
User Guide 8 Chromatogram Editing
- 73 -
User Guide 8 Chromatogram Editing
Note the two vertical guidelines marking the beginning and end of the peak. After
finishing the operation new peak is added ⓒ .
- 74 -
User Guide 8 Chromatogram Editing
accordingly, you can create template method from the method in chromatogram,
which is useful after optimizing integration parameters.
1. Open the chromatogram with optimized method.
2. Select the Method - Save as Template… to save the method (including the
changes made by you) from the current chromatogram as a new method, or for
example overwrite (update) your acquisition method.
Note: It is not possible to overwrite any method that is currently in use, for example a
method opened on any Instrument.
2. Either select an existing group from the list or create a new one by inserting a
letter to ID (groups are defined by a single letter) field and click the Add button.
Note: Group Name is based on the calibration file, similarly to the Compound
Name.
3. Click in the chromatogram for the first time to select the start point and second
time to select the end point of the group interval. Peaks with apexes found in the
interval will be added to the new group.
- 75 -
User Guide 8 Chromatogram Editing
- 76 -
User Guide 8 Chromatogram Editing
- 77 -
User Guide 8 Chromatogram Editing
9. Click and drag the text if you wish to move it to a different location.
10. Double-click the text to open Text Label dialog to adjust the label settings or to
delete it.
9. Click and drag the line if you wish to move it to a different location. Alternatively
drag one of the ends to change its position.
10. Double-click the line to open Line Label dialog to adjust the label settings or to
delete it.
- 78 -
User Guide 8 Chromatogram Editing
- 79 -
User Guide 9 Calibration
9 Calibration
Following chapters contains multiple topics which will guide you through the basic
principles of calibrating in Clarity and also introduces you to advanced solution, for
example using the Bracketing.
Clarity Tips&Tricks videos covering Calibration topics can be found in Clarity
Calibration playlist.
- 80 -
User Guide 9 Calibration
4. Fill in the units in the Compound Units section ⓒ to suit your analysis
conditions.
5. Set the Mode to Calibrateⓓ .
6. Set the Calibration option to Automatic to add the peaks without modification or
Manual to modify them one by one ⓔ .
7. Open integrated chromatogram of a standard: choose File - Open Standard… or
click on ⓕ in the Calibration window.
8. Add peaks belonging to the compounds of interest from the chromatogram of the
standard to the calibration file.
Note:
Select Calibration - Add All or click on to add all integrated peaks or
the Add Peak / Add Group icons to add specific peaks ⓗ.
Regardless of the set Current Level ⓖ the peaks will be added to the
first free level.
- 81 -
User Guide 9 Calibration
- 82 -
User Guide 9 Calibration
10. If you selected Manual calibration: Fill in the Amount, the Compound Name and
set any other parameters related to the peak on the Calibration - Add Peak
window. This window will open once for each one of the peaks processed.
- 83 -
User Guide 9 Calibration
- 84 -
User Guide 9 Calibration
Note: In case more peaks than expected emerge in the calibration, the surplus
peaks can be deleted by selecting them in the Calibration Summary
Table and deleting them using Calibration - Delete Compound or clicking
on ⓕ.
7. Set the amounts of the particular compounds into the Calibration Summary
Table, into the Amount column ⓖ of the respective calibration level.
8. Save the calibration file: choose File - Save or click ⓗ
9. Click any tab below ⓘ and you will be able to see the calibration curve ⓙ with
all the levels added for one specific compound ⓚ .
- 85 -
User Guide 9 Calibration
2. Check whether the Calibration File (Peak Table) field ⓑ is set to (None). If that
is the case, then the chromatogram does not have a calibration file linked to it.
3. Also check the Compound Names ⓒ in the Result Table section. This column
must be empty.
4. To link the calibration file to the Chromatogram, click the Set… button ⓓ in the
right section of the Results tab. You will get a list of all calibrations available in
the present project.
5. Select the correct calibration file from the list and click OK. The content of the
Chromatogram window will change.
- 86 -
User Guide 9 Calibration
6. Check that the Calibration File (Peak Table) field contains the name of the
calibration file. ⓔ
7. The Compound Name column ⓕ in the Result Table, as well as the identified
peaks in the graph ⓖ , will now have the names of the identified peaks from the
calibration file.
8. Check the Calculation field ⓗ to see the type of calculation performed on the
chromatogram.
9. Save the chromatogram: select File - Save or click .
- 87 -
User Guide 9 Calibration
3. Click the Set… button ⓐ to select a calibration file for the method, or create a
new calibration file by clicking the New… button ⓑ .
4. Change the default calibration type in the Calculations field ⓒ .
5. Click OK ⓓ to save the changes to the opened method.
6. You can modify the calibration file after acquisition. For more info go to Apply the
calibration to a Chromatogram.
- 88 -
User Guide 9 Calibration
3. Check that the Compound Units ⓒ are set correctly, Mode to Calibrate ⓓ and
Calibration option is set to Automatic ⓔ .
4. Open an integrated chromatogram of a standard (containing peaks of
compounds of interest with a known concentration): select File - Open
Standard… or click ⓕ on the Calibration window.
5. Add all peaks in the chromatogram of the calibration standard to the calibration
file. Choose Calibration - Add All or click on ⓖ.
- 89 -
User Guide 9 Calibration
8. Type in the Manual Response Factor for each one of the compounds ⓙ .
- 90 -
User Guide 9 Calibration
Caution: Calibration curve cannot be constructed when amount is set to zero for all levels.
In case calibration curve could not be constructed for one of the compounds, no
results are calculated for all of the compounds (as the total amount could be
wrong). If you do not want to calculate amounts for all identified compounds, you
could use free calibration with zero response factor.
- 91 -
User Guide 9 Calibration
5. Set the Calibration option to Automatic to add the peaks without modification or
to Manual to modify them one by one ⓓ .
- 92 -
User Guide 9 Calibration
- 93 -
User Guide 9 Calibration
3. Connect the calibration file to the method by using the Set… ⓐ button.
4. Save the method file - click OK ⓑ or select File - Save Method or click .
- 94 -
User Guide 9 Calibration
The file is now prepared so that the calibration standards measured according to
their sequence rows will automatically recalibrate the calibration file.
4. Select the Save As option. In the Save As dialog check the Clear Responses
checkbox and save the calibration under a new name.
- 95 -
User Guide 9 Calibration
- 96 -
User Guide 9 Calibration
2. Click on Set... button and select a calibration file ⓐ to be used during cloning at
first recalibration.
Note: This calibration will remain unchanged as newly created clone of the
calibration will be used with new responses.
3. Create a custom name for the calibration files in Calibration Cloning in Sequence
ⓑ as explained in Creating customized file names automatically.
Note: The name of the final calibration file will match just the content of this field -
if you wish to include the name of the template calibration, include the
name in this field again (e.g. "test - %s %L %R").
4. Follow the steps in Creating and running a sequence to create your
sequence based on the example below.
More Info:
l
Set the row/s for the standard/s at the beginning. ⓒ
l Add a row for a blank, if you wish to.
l Set the row/s for the unknown samples.
- 97 -
User Guide 9 Calibration
Note: The measured sequence can be reprocessed using Batch. For more info see the
topic Reprocessing whole sequence while using calibration cloning.
- 98 -
User Guide 9 Calibration
Once you added all chromatograms to your calibration and filled Amount values for
all compounds and all levels, you can start to adjust it. Note that the Amount values
are the values you know, because they originate from the concentration levels of
your calibration solutions. Further adjustments of the calibration are available on
the corresponding Compound Tabs at the bottom of the window. In our case, the
Comp No. 3 ⓐ is edited.
- 99 -
User Guide 9 Calibration
Modifying Origin
Another option how to improve your calibration is setting of Origin ⓒ and its
incorporation or exclusion in calibration calculations. In Clarity there are three
available setting options, as described on the following images. There is also
demonstrated an effect of the Origin setting on calculated Equation and Correlation
factor ⓓ .
- 100 -
User Guide 9 Calibration
- 101 -
User Guide 9 Calibration
- 102 -
User Guide 9 Calibration
As you can see from the Correlation factor ⓓ values, changing the Origin setting
from the Curve Passes through Origin to Compute with Origin improved the curve
fit. However, the option Ignore Origin has proved to be the best match for the used
detector.
When changing Curve Fit Type ⓔ pay also attention to the values of the calculated
Equation and Correlation Factor. Increasing value of the Correlation Factor
indicates better selected Curve Fit Type for the measured data.
- 103 -
User Guide 9 Calibration
- 104 -
User Guide 9 Calibration
As you can see, changing the Curve Fit Type from Linear to Cubic increased the
Correlation Factor from 0.954 to 0.992.
- 105 -
User Guide 9 Calibration
As you can see, excluding the incorrectly measured calibration point resulted in
more accurate calibration.
- 106 -
User Guide 9 Calibration
the responses from the last series of calibration standards (immediately preceding
the current unknown samples).
1. Open the method that will be used in the sequence: click File - Open Method…
from the Instrument window.
2. Navigate to Calculation tab.
3. Set the template calibration as Calibration File: click Set... ⓐ and select the
calibration. Note that this calibration will remain unchanged, newly created clone
of the calibration will be used with new responses.
4. Set the name of the cloned calibration: set the name in Calibration Cloning In
Sequence ⓑ . You can use predefined parameters (refer to Creating
customized file names automatically).
- 107 -
User Guide 9 Calibration
- 108 -
User Guide 9 Calibration
10. Open Sequence window and create a new sequence: click Analysis -
Sequence and then icon.
11. Set the sequence according to the following steps: (for more details about
creating the sequence refer to Creating and running a sequence)
l
Set the row/s for the standard/s. ⓘ
l Add a row for a blank, if you wish to.
l Set the row/s for the unknown samples.
l Repeat the row/s for the standard/s.
Note: The sequence must start and end with a row with Standard Sample Type.
l Repeat the previous four steps for every "bracket" of unknown samples you
wish to add.
- 109 -
User Guide 9 Calibration
13. Run the sequence (for more details about running the sequence refer to
Creating and running a sequence) .
14. The results shown while the sequence is running are recalculated at the end
of each bracket, when the standard after the unknown sample is acquired.
- 110 -
User Guide 9 Calibration
The calibration used is an average of the two calibrations, before and after
the unknown.
Note: The measured sequence can be reprocessed using Batch. For more info see the
topic Reprocessing whole sequence while using calibration cloning.
Prerequisites:
l Integrated chromatograms of standards.
- 111 -
User Guide 9 Calibration
- 112 -
User Guide 9 Calibration
- 113 -
User Guide 9 Calibration
12. Fill the Calibration Summary Table with peaks from the currently selected
(second) signal in chromatogram using Add All icon. Dialog will be invoked
after clicking Add All icon questioning if an already used chromatogram
should be reused for this calibration. It is necessary to confirm the reuse by
clicking Yes button.
Note: In case of constructing calibration with more than two signals, steps 12 and
14 have to repeated as many times as necessary in order to fill in peaks
from all signals to Calibration Summary Table.
Note:
- 114 -
User Guide 9 Calibration
Note: Which signal of chromatogram is active and is being worked with is given
by the Calibration Summary Table title's color and displayed number of
currently active signal.
13. Rename the automatically pre-filled names in the Calibration Summary Table
column.
14. Enter values into Amount column. The first concentration level is finished now. It
is possible to proceed building up to next levels of calibration.
Note: All rows of Amount column have to be filled in.
- 115 -
User Guide 9 Calibration
- 116 -
User Guide 9 Calibration
- 117 -
User Guide 9 Calibration
19. When calibration is finished do not forget to save it using select File - Save or
click on .
- 118 -
User Guide 9 Calibration
20. For calculating result in sample open the Chromatogram window select
Window - Chromatogram on the Instrument window or click on and open
chromatogram of sample and link the calibration to chromatogram.
21. Review results on Results Table of each individual signal or review results for all
signals on All Signals Results Table.
- 119 -
User Guide 9 Calibration
- 120 -
User Guide 9 Calibration
- 121 -
User Guide 9 Calibration
- 122 -
User Guide 9 Calibration
6. Open the Calibration Options dialog. Choose Calibration - Options… or click on
. Select the STDADD in the Display Mode drop-down list and click OK.
Note: In case you want to use a blank sample too, such sample shall be always put in
the sequence before the unknown sample.
- 123 -
User Guide 9 Calibration
9. Click to open the Sequence Options dialog ⓔ and select Standard Addition
Measurement and click OK.
10. Run the sequence and wait until the sequence is finished.
11. Open the Calibration window: choose Window - Calibration in the Instrument
window or click on .
12. Go to File - Open and open the cloned calibration file for the desired sample.
13. Click on the Compound tab ⓕ to see the calibration curve. Fill in the amounts of
the standard samples.
- 124 -
User Guide 9 Calibration
14. Open chromatogram of the desired unknown sample. The Result Table now
contains amount of the unknown sample, calculated using Standard Addition.
Note: The measured sequence can be reprocessed using Batch. For more info see the
topic Reprocessing whole sequence while using calibration cloning.
- 125 -
User Guide 9 Calibration
- 126 -
User Guide 9 Calibration
3. Edit the Left and Right window values to define the range within which the peak
should appear ⓒ . This window may include other peaks if the selected
reference peak meets the Peak Selection criteria.
4. Select the Peak Selection criteria: biggest, nearest, first or last peak ⓓ .
More Info:
Ordinary peaks are identified by the nearest option by default. For reference peaks
the biggest will be set as default. Note that the biggest refers to the selected
Response Base, i.e. if the Area is the Response Base, the detected peak may not be
the highest one. In specific cases selecting First or Last may be advantageous.
To add the Peak Selection column:
l Right click anywhere on the table
l Select Set Up Columns to open the relevant window.
l Select Peak Type on the right and click on Show and then Ok.
5. Repeat steps 3 to 5 to add more reference peaks.
Note: In case of multisignal calibration, where compound used as reference does not
match the selection criteria on some signal(s), create a variant of the compound
only for that signal(s). The name must be different - e.g. Oxalic - UV, Oxalic - RI),
copy the rest of the original row, check the used checkbox only for one variant on
each signal. Now change the Peak Selection criteria value, or if suitable, change
the Peak type to Ordinary, and select other Reference peak on that signal.
- 127 -
User Guide 9 Calibration
- 128 -
User Guide 9 Calibration
- 129 -
User Guide 9 Calibration
- 130 -
User Guide 9 Calibration
5. The resulting amount in Chromatogram Results Table will be calculated
according to the calibration curve of compound selected to Calculate By, and
multiplied by the Correction Factor.
Note: A compound marked as ISTD should neither be used as a compound that
is used to calculate another compound nor as a compound that is
calculated using another compound.
Note: You can set up to 10 compounds as ISTDs in the Is ISTD column. When more
ISTDs are present, you can specify which one to use for quantification of which
compound in the Use ISTD column.
- 131 -
User Guide 9 Calibration
5. Now, the first calibration level is set. Continue as usual to set the remaining
calibration levels using the measured calibration standards (as shown in Adding
a new calibration level). Enter the ISTD amounts according to the mode you plan
to work in.
The created ISTD calibration may now be linked to an already measured
chromatogram for direct results assessment (see Applying calibration to a
chromatogram), and it may be also linked to a method. This causes that the
chromatograms measured using this method will be automatically evaluated using
the pre-selected ISTD calibration.
- 132 -
User Guide 9 Calibration
6. To link the ISTD calibration to a method, in Method Setup, open the Calculation
tab ⓐ and set the created ISTD calibration file ⓑ . The ISTD parameter ⓒ
should be selected in the Calculations option. Press the OK button to save
the Method and close the dialog.
7. To use the ISTD method in single analysis, set the ISTD1 Amount ⓐ in the
Single Analysis window to the desired value. In the case of more than one ISTD
compound present, set the amounts in the dialog which opens by clicking the
triple dot menu ⓑ .
8. Select the previously created method ⓒ . You can Run Acquisition directly
from the Single Analysis window.
- 133 -
User Guide 9 Calibration
The sequence analysis setup is analogous to the single analysis. The ISTD2–
ISTD10 amount columns are not shown by default and can be added via the Edit -
Setup columns... dialog.
Note: In the case the amount of internal standard is the same for all standards and
samples, it is possible to enter zero instead of the amount (which may not be
known). In such case, the correction will be based on the ratio of the ISTD peak
response in the standards and unknowns. Note that the ISTD amounts must be
entered or set to zero both in Calibration and in the Sample Header. A mismatch
will be detected and reported as error.
- 134 -
User Guide 9 Batch Processing
9 Batch Processing
This chapter focuses on processing more chromatograms using Batch dialog.
There are 3 main ways of workflow Reprocessing whole sequence, Reprocessing
by Method and Post Run options, all will be explained in more detail in this chapter.
Clarity Tips&Tricks videos covering Batch topics can be found in Clarity Batch
playlist.
4. Open the Batch dialog by using Analysis - Batch in the Instrument window.
5. Switch File Type to Sequence Files ⓓ .
6. Select the Sequence you wish to reprocess.
7. Check the Complete Processing ⓔ checkbox.
8. To preserve manually updated integration (this mainly considers small
individual adjustments where optimized parameters weren't satisfactory)
select the Preserve ⓕ in the Integrationpart of Options section.
9. To perform recalibration based on updated integration select Update ⓖ in
- 135 -
User Guide 9 Batch Processing
the Calibration part of Options section.
10. Proceed ⓗ with the batch processing.
The recalibration will be performed and Post Run options defined in the sequence
will be performed.
Caution: The operations during Batch reprocessing are done row after row, injection after
injection. In some situations it is necessary to perform the Complete Processing in
two steps - first just the recalibrations, second the post-run actions (either using
the Complete Processing again with the Integration a Calibration settings in the
Options section to Preserve, or through running the post- process on selected
chromatograms only through procedure described in Performing Post Run
Options from Batch dialog section). For example, if the sequence is using
calibration bracketing the unknowns are measured before second standards set
and if the unknowns were reported during the recalibration step, the responses
from the second standards set would be missing in the report.
- 136 -
User Guide 9 Batch Processing
Please refer to the Calibrating using clone on first recalibration, Compensating for
response drift using bracketing or Improving quantification with the standard
addition method topics to learn more about Calibration Cloning in Clarity.
The difference from the standard Reprocessing procedure described in
Reprocessing whole sequence topic is shown below.
l In the Batch dialog you have additional option for calibration behavior, the
meanings of options are:
l If you select Update ⓐ in Calibration part of Options section the
calibration(s) that are currently linked to measured chromatogram(s) will
be recalibrated (new cloning is NOT performed).
l If you select Clone New ⓑ in Calibration part of Options section new
calibration(s) will be created as if the sequence was run again (new clone
(s) are created and linked to chromatograms).
Note: Reprocess by Method can be combined with Post Run Options if you are sure that
changes in chromatograms do not have to be reviewed before exporting/printing.
- 137 -
User Guide 9 Batch Processing
Now your chromatograms are reprocessed, with linked calibration file and
amended integration parameters from the used method.
- 138 -
User Guide 9 Batch Processing
Note: It is also possible to process data before printing them by using Complete
Processing or Reprocess by Method options. When Complete Processing is
selected Post Run options are governed by settings in reprocessed sequence. For
more info see chapter regarding Batch dialog for more info.
- 139 -
User Guide 10 Results and Calculations
3. Fill in the Title ⓑ of the new column. In our case Ratio Area/ISTD.
4. Check/Uncheck Calculate Total ⓒ (depends on whether sum of calculated
values has a meaning).
5. Fill in Expression ⓓ line , which presents the user's defined calculation. In our
case:
- 140 -
User Guide 10 Results and Calculations
l In Columns: ⓔ list double-click Area.
l In Opers: ⓕ list double-click /.
lIn Columns:list select Area, then click on Special Values ⓖ , select
Compound ⓗ , choose ISTD ⓘ and click OK.
6. Fill in appropriate Units ⓙ based on the formula.
7. Click OK ⓚ in the Add User Column dialog.
Note: It is possible to fill Expression by typing if you know the correct syntax.
Note: You can edit existing User Columns by right-clicking into them and selecting User
Columns - Edit Selected....
- 141 -
User Guide 10 Results and Calculations
- 142 -
User Guide 10 Results and Calculations
- 143 -
User Guide 10 Results and Calculations
5. Defined Custom Name of the variable is used directly in the column header ⓘ .
Fill in the values for each row (different values can be used).
- 144 -
User Guide 10 Results and Calculations
Note: To use variables in custom calculations add User Column as described in chapter
"User Columns" and while formulating the Expression select the desired variable
from Variables list ⓚ .
- 145 -
User Guide 10 Results and Calculations
- 146 -
User Guide 10 Results and Calculations
- 147 -
User Guide 10 Results and Calculations
simplify changes of the value - this variable can be then used instead of entering
the numerical value in the Expression edit box.
Note: In the Expression edit box, variables are treated as numbers, therefore operators
and numbers can be used to modify the formula. The resulting formula can be e. g.
2* [Height]/ [Noise] or 2* [Height]/ [UserVariableNoise] (where the
[UserVariableNoise] represents user variable to which the Noise value
determined from chromatogram blank is stored) or 2*[Height]/0.1080 (where the
0.1080 represents the actual Noise value determined from chromatogram blank).
Note: Setting the Analysis User Variable or the Method User Variable is described in the
section User Variables on pg. 142.
- 148 -
User Guide 10 Results and Calculations
Note: To insert [malic$Reten. Time] select Reten. Time in Columns: list, click Special
Values ④ , in the drop-down list select Compound ⑤ , and in the following dialog,
select the compound ⑥ you want to use for calculation of RRT (the list is based
on the used calibration file).
Note: You can do a quick check that the calculation is correct: look at the malic
compound, the RRT shoud be 1.
- 149 -
User Guide 10 Results and Calculations
- 150 -
User Guide 10 Results and Calculations
- 151 -
User Guide 10 Results and Calculations
Note: Value is only valid for peaks, where peak end is not on the baseline.
- 152 -
User Guide 10 Results and Calculations
- 153 -
User Guide 10 Results and Calculations
You can either use items from the Columns: list with Special Values or type it
manually.
6. Apply the iferr0 function on the items that may be missing from the result table.
The resulting function will look like the following: (iferr0 ([α-
Tocopherol$Amount])+0.5*iferr0 ([β- Tocopherol$Amount])+0.25*iferr0 ([γ-
Tocopherol$Amount])+0.01*iferr0([δ-Tocopherol$Amount]))/10 ③ .
7. Close the dialog by clicking OK.
Note that the All Peaks in Calibration ⑤ option is selected. Otherwise, the missing
compound would not be present in the Result Table, and the calculation would not
be possible.
- 154 -
User Guide 10 Results and Calculations
- 155 -
User Guide 10 Results and Calculations
2. Using the Open with stored calibration option. Such option is accessible from
multiple dialogs, depending on whether Single Analysis or Sequence is
measured (automated approach) or this option can be selected upon opening
the chromatogram.
l In the Single Analysis dialog - Post Run Settings tab select Open with
stored calibration option.
- 156 -
User Guide 10 Results and Calculations
from history and show the results according to the stored calibration.
Any changes in the linked calibration will not affect those results. To
open the chromatogram with the stored calibration this way it is
necessary to re- open the chromatogram using the Open
Chromatogram dialog - re-opening the Chromatogram window is not
sufficient.
Note: This setting is only applied and saved to the currently opened
chromatogram. It will not be transferred to the next (different)
opened chromatogram.
- 157 -
User Guide 10 Results and Calculations
- 158 -
User Guide 10 Results and Calculations
Note: It is possible to automatize this process. Calibration can be linked to the used
method so it is automatically linked. Amount and Dilution can be pre- filled in
Single Analysis or Sequence windows. Beware that columns in Sequence have
slightly different names - Sample Amount and Sample Dilut.
- 159 -
User Guide 10 Results and Calculations
- 160 -
User Guide 10 Results and Calculations
3. Right click on the Summary Table, if you want to change the visualization of the
table or add a custom column, etc.
4. To see all signals, select the Show All Signals checkbox in the Summary Table
Options dialog accessible from the pop- up menu of the Summary Table. By
default, only signals containing calibrated peaks are visible in the Summary
Table.
- 161 -
User Guide 10 Results and Calculations
result in flat line when the peak is pure compound, impurities with different signal
ratio would be represented as response change on the Virtual Detector signal.
1. Add the Virtual Detector as well as your dual wavelength detector to the
instrument, if not already present, based on chapter Adding a new device.
2. Open the Method Setup dialog and navigate to Acquisition tab.
3. Select VD: Channel 1 in the Select Detector drop-down list ⓐ .
4. On the Argument settings tab, tick the Argument X checkbox, select an Ext.
Source ⓑ , then click on the Settings button ⓒ , select the appropriate detector
Channel ⓓ and click OK .
5. Repeat the same steps for Argument Y and select the second channel from the
detector.
6. If want to use the same signal ratio for entire analysis you can fill in the Resulting
Formula field ⓔ with the proper formula. If you need multiple formulas for
different times skip this step and continue with the next one.
7. To set multiple formulas for different time intervals navigate to the Advanced
settings, check the Use Advanced Settings ⓕ checkbox and fill the time table as
needed.
- 162 -
User Guide 10 Results and Calculations
8. On the Detector settings tab, set the Time to 0 ⓖ to avoid distortion and fill
Sampling Rate ⓗ equal to the actual detector sample rate.
Method prepared this way can be used to check compound identity as mentioned in
the beginning of this chapter.
- 163 -
User Guide 11 Data Reports
11 Data Reports
Following chapters describe how to create a report, adjust it and print.
- 164 -
User Guide 11 Data Reports
To Print a Report:
1. Select File - Print in the Instrument, Calibration, Chromatogram or Sequence
windows. In Method Setup click Report Setup icon and in the following dialog
click Print....
- 165 -
User Guide 11 Data Reports
2. Print dialog will open for you to setup your printing options and confirm the
printout.
To Print to PDF:
1. Select File - Print to PDF in the Instrument, Calibration, Chromatogram or
Sequence windows. In Method Setup click Report Setup icon and in the
following dialog click Print to PDF.
2. Print to PDF dialog will open for you to enter file name confirm the saving of the
file.
- 166 -
User Guide 11 Data Reports
- 167 -
User Guide 11 Data Reports
Sequence
1. Procedure is similar to the single analysis. In Sequence window some of the
required columns are hidden by default, to display them right-click the sequence
table and use Setup Columns in the following dialog move desired columns to
the Show Columns list.
2. Check checkboxes in columns Open, Print and/or Print to PDF on rows where
report should be generated.
3. Select Report Style for each row where report is generated. It is possible to
modify specif report style by selecting its cell and clicking .
4. It is also possible automatically export data, to do so check Export Data option.
What should be exported can be set by using Setting - Export Data in the
Instrument window.
- 168 -
User Guide 11 Data Reports
- 169 -
User Guide 11 Data Reports
- 170 -
User Guide 11 Data Reports
- 171 -
User Guide 11 Data Reports
- 172 -
User Guide 11 Data Reports
8. Set the Chromatogram section. Only the format of printed graphs is set here,
result tables are printed based on Results tab, see next step.
l Click the Chromatogram tab ⓕ .
l Check the options Print and Fixed Height and set the height to 110 mm.
l Select the options Signals - All and Print range - As On Screen.
- 173 -
User Guide 11 Data Reports
- 174 -
User Guide 11 Data Reports
12. You can preview your report by clicking Preview.... For more info see Printing or
previewing a report.
13. Click the OK to save the report style.
- 175 -
User Guide 11 Data Reports
- 176 -
User Guide 12 File Management
12 File Management
How to set project directories, create new projects, preset file names of measured
chromatograms based on variables, store files into subfolders etc.
Caution: Be aware that when updating Clarity the COMMON files outside of the installation
directory must be updated manually.
- 177 -
User Guide 12 File Management
1. To create new project navigate to the Project Setup dialog by selecting
Instrument - Project... in the Instrument window.
Note: You can also open a project through the Login Dialog opened from the Main
Clarity window by selecting the New Project option and clicking OK.
2. In the Project Setup dialog click the New button ⓐ to create a new project.
Note: Entered project name must not contain invalid characters, i. e. \ /:*?"< >|.
- 178 -
User Guide 12 File Management
- 179 -
User Guide 12 File Management
- 180 -
User Guide 12 File Management
- 181 -
User Guide 12 File Management
4. Set the File Name of the first row to %s\%q_%Q_%R. (To see detailed
description on how to create the name using variables see chapter "Creating
customized file names automatically" on page 179.)
5. Right click on the File Name column and select Fill down to apply the name to all
rows.
l To create subfolder for each month use the %m or %B variables and for
each day use the %a or %A variables.
l Same procedure can be also used in Single Analysis window.
- 182 -
User Guide 13 Import and Export Data
4. Set the parameters as needed in the subsequent dialogs depending on the file
format.
5. For all file formats you can select imported file name and which method to apply
ⓒ . This will apply integration, calibration etc.
l Import AIA File - when importing AIA file (*.CDF suffix).
l You have option to inspect data from the source file and override some of
them ⓓ e.g, Detector Unit.
l Import Text File - when importing Text, Multidetector or EZChrom ASCII file
(*.TXT, *.CHR, *.ASC and *.CSV suffixes).
l You can set information regarding given sample ⓔ like Dilution or
Injection Volume etc. You can also set information regarding source file
- 183 -
User Guide 13 Import and Export Data
format ⓕ .
Note: File format related settings is typically detected automatically
and there is no need to change it.
l Save As - when importing *.RAW file, as there is no need to set other
parameters.
- 184 -
User Guide 13 Import and Export Data
- 185 -
User Guide 13 Import and Export Data
l This setting will be used for automated data export from Sequence or Single
Analysis.
l To manually export data you can invoke the same dialog from Chromatogram
window by using File - Export - Export Data... command or icon. In this case
dialog also includes Export button which can be used the export data from
currently opened chromatogram.
Export settings
1. Open Clarity Instrument window and select - Setting - Export Data....
2. In the Export Data dialog set the required Export Content ⓐ .
Note: While exporting Result Table it is recommended to check In Fixed Format.
Otherwise the content will change according to changes on screen.
Beware that In Fixed Format is only applied to the In Result Table.
3. Select desired file formatting in Text Format section.
4. Select Export to - Text File ⓑ and choose preferred suffix.
5. Select Full Format option ⓒ . This will precede each result table row with a file
name, date and time, thus allowing easy sorting after import.
- 186 -
User Guide 13 Import and Export Data
6. Set File Name ⓓ and check Append option, in this way all results will be
exported to a single file which can then be simply imported directly to your LIMS.
Alternatively, leave the field empty in which case each export will be performed
to a separate file. Exported files will have automatically generated file name from
the chromatogram name.
- 187 -
User Guide 13 Import and Export Data
Single Analysis
l In the Post Run Settings tab, of the Single Analysis dialog, select
Export Data ⓕ .
Note: Export of data will be performed automatically after a single
analysis is finished.
- 188 -
User Guide 13 Import and Export Data
Sequence
l In the Sequence window, select checkbox in the Export Data ⓖ
column for row(s) to be exported.
Note: Export of data will be performed automatically after the row
has been measured.
More Info:
l By default, Export Data column is hidden. Right mouse click in
the sequence table and choose Setup Columns... to show it.
l In the Setup Columns dialog, choose Export Data from the left
list and click on the Show button - it will be added to the show list.
Once you click on the OK button, you will return to the Sequence
window and new Export Data column will be present.
- 189 -
User Guide 13 Import and Export Data
Batch
l Export multiple chromatograms at once using the Batch dialog,
accessible from the Instrument window by Analysis - Batch...
command.
l Select chromatograms to be exported and check the option to Export
Dataⓗ .
l Clarity is able to start external program with a parameter ⓘ . Specify
a program to run using the and specify a command line parameter
- typically the file name of the exported text file %e. Once ready to
export the data, click the Proceed button.
- 190 -
User Guide 13 Import and Export Data
- 191 -
User Guide 13 Import and Export Data
- 192 -
User Guide 13 Import and Export Data
8. Set the mapping so that the columns used in Clarity (Sequence Column) will
correspond to the columns from the imported file (Imported Column) ⓖ .
Columns that are highlighted in bold letters are required. When the imported
table contains column headers (the First Row is Header checkbox from the
Import Sequence Step 1 dialog has been checked), Clarity will attempt to
automatically map the columns according to their names. Manual mapping will
override the default name mapping. Clarity will store these names for future
imports .
9. Select whether you want to save imported sequence as a new file or just append
it to currently opened one ⓗ .
10. Three checkboxes allows you to do following ⓘ :
l Original file will be erased after successful import.
l Settings of this dialog will be saved for future imports.
l Sequence Option will be shown immediately after finishing import.
11. Press the Finish button to accomplish import and close the dialog ⓙ .
- 193 -
User Guide 13 Import and Export Data
- 194 -
User Guide 14 Mathematical Operations
14 Mathematical Operations
This functionality was implemented into Clarity to give users a tool to manipulate
already measured data without modifying original raw data. Mathematical
operations present a very convenient approach to perform various actions such as
extraction of a selected signal(s) from multiple signal chromatogram or subtraction
of various chromatograms from each other. You can display three examples of
applications of Mathematical Operation in following chapters. Many other
applications may be developed by Clarity users on their own.
- 195 -
User Guide 14 Mathematical Operations
other chromatogram. It is possible to set Subtraction Chromatogram directly in
Chromatogram window on Measurement Conditions, however, if you want to have
both original and subtracted chromatograms as individual files you can use
Mathematical Operations.
1. Open your chromatograms that you want to work with in Overlay.
2. In Chromatogram window select Chromatogram - Overlay - Mathematics to
open Mathematical Operations dialog.
3. Select operation A - B ⓐ . From the drop-down menu in the Operand A section,
select the chromatogram you want to subtract from ⓑ .
4. From the drop-down menu in the Operand B section, select the chromatogram
you want to subtract ⓒ .
5. Check span Save As Chromatogram ⓓ , fill a name ⓔ and click OK to save the
new subtracted chromatogram.
- 196 -
User Guide 14 Mathematical Operations
- 197 -
User Guide 14 Mathematical Operations
- 198 -
User Guide 15 Archive and Restore
- 199 -
User Guide 15 Archive and Restore
- 200 -
User Guide 15 Archive and Restore
- 201 -
User Guide 15 Archive and Restore
- 202 -
User Guide 15 Archive and Restore
- 203 -
User Guide 16 Managing the Chromatography Station
- 204 -
User Guide 16 Managing the Chromatography Station
Options described below are optional and are not obligatory for correct
functionality. They are revealed by clicking the Advanced button.
l Unregister All Instruments - disables the monitoring in the Clarity2Go
application. Instruments that have been registered will no longer be available
for monitoring. If you will later change your mind, you will have to generate
new Instrument ID.
l Web Server Address - do not change this field. It defines the address of
Clarity2Go web server. Address other than default will result in the
monitoring to be not functional! Press the Default button to set functional web
address of the server.
l Proxy Server Address - consult with your local administrator if a proxy server
is applied in your local network and then provide the proxy server address.
l Protect by Password - provided password will be valid for all Instruments.
The same password needs to be provided in the Clarity2Go application to
unlock the monitoring.
l Click the OK button to save the configuration and continue with steps
described in How to set up Clarity2Go.
- 205 -
User Guide 16 Managing the Chromatography Station
2. In the Settings, tap on Demo - Switch to demo mode which turns the Clarity
Demo mode OFF.
3. Return back. In case you are configuring Clarity2Go for the first time, you will
see that there is no instrument.
4. Tap on the blue "plus" button at the bottom right corner to add a new instrument.
5. Enter Instrument ID that has been generated by Clarity and enter password
(only if you have set it up in Clarity). Tap on the OK button to start monitoring
this instrument.
6. The newly configured instrument will be added to the list of instruments that are
being monitored.
Note: You can invoke the application menu by tapping on the 3 horizontal lines -
the menu contains Settings, built-in Help and About options.
- 206 -
User Guide 16 Managing the Chromatography Station
Manual Lock
l Instrument can be manually locked either from Instrument window by clicking
Instrument - Lock Instrument_Name.
or
l From the Main Station window by clicking Instruments - Lock Instrument_Name.
Auto Lock
It is possible to set the automatic lock function so that all opened instruments will be
locked after a period of inactivity.
1. Open the User Accounts window by clicking on or choose System - User
Accounts.
2. Check the Auto Lock function ⓐ .
3. Set the period of inactivity in minutes after which all opened Instruments will be
locked ⓑ .
- 207 -
User Guide 16 Managing the Chromatography Station
Caution: When an Instrument is Auto Locked, all eventual unsaved changes in modal
dialogs (Method Setup, Single Analysis, Report Setup, etc) are discarded and
those dialogs are closed.
- 208 -
User Guide 16 Managing the Chromatography Station
1. Unlock the Instrument using credentials for a user with administrator rights
over the locked Instrument according to chapter Taking control over locked
instrument.
2. Restart Clarity. Instrument can again be opened by the previous user.
Note: It is recommended to set one common Desktop File for all users with this option
enabled. Otherwise desktop of the first logged user will be used for all users that
take control over the Instrument, until the Instrument is closed again.
Note: It is also recommended to set the same Edit... rights for users that plan to take
over Instruments as different rights may limit the options of second user.
Note: It is possible to set the Auto Lock function so that all opened instruments will be
locked after a period of inactivity.
- 209 -
User Guide 16 Managing the Chromatography Station
Caution: All unsaved changes in files made by the previous user will be discarded.
Note: It is possible to take control of locked Instrument using the Command line
parameters. Always include the user (user=…) and password (p=…) parameters
for each command, because the instrument is kept locked.
- 210 -
User Guide 16 Managing the Chromatography Station
2. Click on the Session tab ⓐ if you would like to see the log from the time Clarity
was started or on the Daily Audit Trial tab for the present day events.
Note: The Daily Station Audit Trail is stored in one separate file every day the
station is running. These files can be opened by using the command File -
Open Audit Trail (Append).
3. Click on the Instruments or System icons ⓑ to filter out events and operations
based on instrument where they occurred.
4. Inspect the Description column to find out about operations and events that have
taken place ⓒ .
- 211 -
User Guide 16 Managing the Chromatography Station
5. Click on the Properties icon ⓓ to set up which events and operations should
be recorded in the Session or Daily Audit Trail.
- 212 -
User Guide 16 Managing the Chromatography Station
- 213 -
User Guide 17 Clarity in Network
17 Clarity in Network
Clarity might be used in network and following chapters describe different
approaches of such usage.
- 214 -
User Guide 17 Clarity in Network
More Info:
l
Note that it is possible to have more than one Clarity ⓑ in the network.
l
Shared Network Disk ⓒ can be on the same PC as well and not as a separate unit as
seen in the picture above.
Clarity in network is a solution that consists of at least one Clarity, at least one
Clarity Offline and a reliable computer network as the most basic setup.
Simplified scheme of the possible configuration is displayed in the diagram above.
LC and GC instruments ⓐ are controlled via Clarity ⓑ . Chromatograms,
calibrations and methods are all saved (using directory configuration) on a shared
network disk ⓒ . Clarity Offline ⓓ could then be used for evaluation of acquired
chromatograms and preparation of methods which are saved (using directory
configuration) on the shared network disk. Clarity ⓑ is able to send those prepared
methods to corresponding instruments.
Note that this shared network disk is accessible to all computers within this
computer network therefore a much wider configuration can be implemented than
the one described above.
Following step-by-step guide will help you configure the Clarity in network solution.
- 215 -
User Guide 17 Clarity in Network
8. When you try to login for the first time with the new directory configuration, you
will be asked to create a new project.
- 216 -
User Guide 17 Clarity in Network
If you have followed the steps correctly, your Clarity in network is configured. If you
are not sure, you can test it by measuring some chromatogram and evaluate it on
Clarity Offline. Once you see measured chromatogram in the Data directory you
know it has been configured correctly.
- 217 -
User Guide 17 Clarity in Network
Principle behind migrating Clarity project into a different location is straightforward.
It is necessary to move the whole project directory (e.g. WORK1) as well as the
project file itself (e.g. WORK1.PRJ).
More Info:
l Note that it is possible to move the project directory and the project file also using File
Explorer when Clarity is turned off.
1. Switch Directories back to the previous (default) location, so you can see
previously used projects.
2. Log to the project you don't want to migrate, e.g. one of the demo projects.
3. In the Instrument window, go to menu File - Archive....
4. Backup dialog opens on the Create Archive tab ⓐ .
5. Because we want to migrate whole project, change File Type using the drop
down list ⓑ to Projects.
6. Section File List now contains all projects in Clarity. Click the project to be
migrated (e.g. WORK1) ⓒ .
7. As a Target select the destination of the shared network disk using the at ⓓ .
8. Check the options Without Compressing and Move to Archive ⓔ .
9. Check also Including Common, if you use customized report style, template
sequence and other files stored in Common folder ⓕ (you will be prompted to
confirm overwriting the default files created during Procedure A).
10. When everything is set as described above, you can click the Archive ⓖ button
which migrates project WORK1 to the D:\SHARED\.
11. Now change the directories back to the network shared drive.
12. Login to the Instrument which has set directory in the network. Notice that in your
Clarity Login Dialog the Select Project drop down list offers your migrated project
(WORK1) and project created after setting the new directory only.
- 218 -
User Guide 17 Clarity in Network
13. Your Clarity project has been successfully migrated and you can start working.
This step by step guide will help you to connect remotely to a PC with Clarity ⓒ
installed from your home or office ⓐ . It will allow you to connect remotely to that
PC and take control over the whole computer and thus control Clarity and
connected instruments. The connection is realized over the internet or a local
computer network. Connection over the local computer network is more secure
since the transferred data never enter the internet.
Requirements:
l ⓐ PC needs to have Remote Desktop Connection installed
l ⓑ Internet connection or a reliable computer network
l ⓒ PC with Clarity needs to have Remote Desktop Connection installed
More Info:
l Remote Desktop Connection is a standard application installed in Windows operating
systems.
- 219 -
User Guide 17 Clarity in Network
l Under Remote Desktop, select one of the three options.
l Click Select Users.
l In the Remote Desktop Users dialog box, click Add.
l In the Select Users or Groups dialog box, do the following:
l To specify the search location, click Locations and then
select the location you want to search.
l In Enter the object names to select, type the name of the
user that you want to add and then click Check Names.
This will check whether the user exists. If not, it will trigger
a not found dialog. Check the name once again. Note that
this user must have a profile on this computer. If the user
name is correct, click OK.
l The name will be displayed in the list of users in the
Remote Desktop Users dialog box. Click OK and then
click OK again.
l In case Remote Desktop options are grayed out your computer is
probably in a domain and due to domain policies you may not be able
to change the settings. Contact your network administrator to resolve
the situation.
2. Look up the name of the remote computer ⓒ . You will to provide this
information in step 5.
l Go to System - About.
l There is the Device Name, which is the info you need.
l Alternatively your network administrator might also be able to give you the
name of the computer.
3. Set a password for your user account on PC with Clarity ⓒ . Your user account
must be password protected before you can use Remote Desktop to connect to
- 220 -
User Guide 17 Clarity in Network
another computer. Password protection is now default setting in Windows, so
you may have that already set up.
l Go to the User Accounts - Sign In Options.
4. To allow Remote Desktop connections through a Windows Firewall on the
remote PC ⓒ follow the steps below:
If you're having trouble connecting, Remote Desktop connections might be
getting blocked by the firewall. Here's how to change that setting on a PC. If
you're using another firewall, make sure the port for Remote Desktop
(usually 3389) is open.
l Go to the Control Panel.
l Click System and Security.
l Click Allow an app through Windows Firewall under Windows Defender
Firewall section.
l Click Change settings and then check the box next to Remote Desktop.
l Click OK to save the changes.
5. Start Remote Desktop from the computer you want to work from ⓐ .
l Open Remote Desktop Connection.
l In the Computer box, type the name of the computer that you want to connect
to, and then click Connect. (You can also type the IP address instead of the
computer name.)
l Note that the remote PC cannot be in sleep mode or hibernating.
More Info:
l This text has been taken from the How- to: "Connect to another computer using
Remote Desktop Connection" created by Microsoft Windows.
Once you successfully connect to the remote PC you can work as if you were sitting
in the lab and working with Clarity. The remote desktop will be presented in the
normal window. To terminate the session, close the window.
This solution then enables you to:
l Control instruments that are directly connected to Clarity.
l Monitor data acquisition.
l Evaluate chromatograms in Clarity.
l Work on other projects and leave the remote session open and check once
in a while if everything is running smoothly.
- 221 -
User Guide 17 Clarity in Network
Possible situations that may arise using the Remote Desktop Connection:
l If you remotely connect to a PC where you are currently logged in, you will be
automatically put through and you can start working.
l However, if you try to connect to a PC when there is logged in someone else,
e.g. another analyst, he will be asked if he allows the remote connection to
put through. If he declines the remote connection you will not be able to
connect.
l PC that you are connecting to must be turned on, it is not possible to connect
to a PC that is off or in sleep mode.
- 222 -
User Guide 18 Utilities
18 Utilities
Clarity installation contains various utilities for validating the installation or
predefining various Clarity profiles which helps you in using more configurations of
laboratory instruments.
Caution: Some driver packages have standalone IQ that must be performed separately by
clicking here in given section. Clarity IQ is NOT valid unless IQ of all components
passed and reports are stored together.
- 223 -
User Guide 18 Utilities
Note: The most common reason for a "Failed" result is the installation of an updated
version over an existing installation of Clarity. This itself does not produce any
errors but since some of the files are preserved from the original installation, the
checksums will not match.
- 224 -
User Guide 18 Utilities
3. Select a profile ⓐ to modify it or click on the New button to create a new profile
ⓑ.
4. If you created a new profile, fill its name. For easier orientation it is
recommended to fill in the Description which eases the distinguishing of different
profiles.
5. Select the configuration file that will be loaded after the start of Clarity ⓒ .
Note: The list of configuration files is retrieved from the installation directory
(configuration files are located in C:\CLARITY\CFG by default). Clicking
on button opens menu for managing the configuration files, see
Creating a duplicate configuration using the Launch Manager for one
example of usage.
6. If you select the <Last Used> option, Clarity will start with the last configuration it
was opened with or if Clarity is running, the present configuration will be
preserved.
7. For each Instrument, choose if it should be opened at the Clarity start by Open
Instrument checkbox. Select the User, Desktop, Project, Method and Sequence
files that will be loaded when opening instruments.
Note: In the desktop file, all layout setting is saved (the width of the table
columns, the custom toolbars), but also custom calculations created in
User Columns.
Note: If <From Project> is selected, the Method and Sequence that were last
opened in selected Project will be used.
8. You can use Refresh Files button ⓔ to reload all the files displayed in the Edit
Profiles dialog.
9. Once the profile is configured, Close the Edit Profiles dialog and Launch the
profile either by double clicking on the profile or by selecting a profile and clicking
on the Launch.
- 225 -
User Guide 18 Utilities
Note: You can use Create Shortcut ⓕ to place shortcut on the desktop which
can be directly used to launch given profile.
- 226 -
User Guide 18 Utilities
2. Open Launch Manager either from the Windows Start menu or by using
C:\CLARITY\BIN\LAUNCHMANAGER.EXE.
3. In the Select Clarity Profile dialog click Edit Profiles....
4. In the Edit Profiles dialog, click the New button ⓐ to create a new profile and
name your profile (e.g. RI + DAD). Click OK to save the profile. Newly created
profile is displayed in the small table on left ⓑ .
5. From the drop-down list ⓒ , select CLARITY.CFG, click the button, from the
selection choose Duplicate and in the following dialog name your configuration
(e.g. ri+dad.cfg).
Note: This step will create a duplicate configuration (.cfg file) based on the
current Clarity setup (as described in step 1.). It prevents Clarity from
overwriting a default configuration.
- 227 -
User Guide 18 Utilities
- 228 -
User Guide 18 Utilities
9. Clarity has launched with the selected profile, open System Configuration. Since
the configuration has been duplicated from the initial configuration, both
detectors are present. Remove the RI detector from instrument and save the
configuration by click OK in the System Configuration window.
Note: This step may differ based on used control modules. In our case simply
drag the RI module from right side to the left one. In other modules e.g.,
when using ICF you have to invoked the control module setup and
change it there.
- 229 -
User Guide 18 Utilities
10. Launch each profile using the Launch Manager to make sure that correct
configuration is loaded. Make necessary changes to the configuration if needed.
- 230 -
User Guide 19 Extensions
19 Extensions
Chapters describing topics related to specific Extensions.
- 231 -
User Guide 19 Extensions
Caution: Once selected, the Calibration Type can't be changed later. More on GPC
Calibration Options can be found in the GPC Extension manual, accessible on
www.dataapex.com.
4. Click OK to save the calibration. To fill calibration name open the File - Save As
dialog window.
5. Open integrated chromatogram of a standard: select File - Open Standard… or
click on ⓑ in the Calibration window.
6. Add peaks in the chromatogram of the calibration standard to the calibration file:
select Calibration - Add All or click on ⓒ (if you have multiple peaks in your
chromatogram) or select Calibration - Add Narrow Peak or click on ⓓ to add
desired peak from your standard.
Note: If you have multiple chromatograms of standards, you can repeat these steps to
add the desired peaks: open the standard and click Add Narrow Peak, then open
another standard and click Add Narrow Peak, repeat for all standards. The
number in field ⓔ is connected to the used standard and is not connected to the
concentration level. Setting a peak on already used number will overwrite the
values with a newly added one.
Note: When using any type of Broad calibration, Add Broad Peak will be enabled instead
of Add Narrow Peak.
7. Fill in the appropriate molecular weight values for the respected peaks into the M
column ⓕ .
8. Save the calibration file: from File - Save or click on ⓖ.
- 232 -
User Guide 19 Extensions
- 233 -
User Guide 19 Extensions
2. To link the calibration file to the Chromatogram, click the Set… button ⓑ , so that
the Open GPC Calibration dialog window will emerge. You can select calibration
file from your current project or you can navigate to other folders.
3. Select the desired calibration file from the list and click OK.
4. Check that the Calibration File (Peak Table) field contains the name of the
calibration file ⓒ .
Note: When no calibration file is linked to a chromatogram, the field Calibration
File (Peak Table) contains inscription (None).
- 234 -
User Guide 19 Extensions
3. Click the Set… button ⓑ to set a GPC calibration file for the template method, or
create a new one by clicking the New… button ⓒ .
4. Click OK ⓓ to apply and save the changes made to the template method.
- 235 -
User Guide 19 Extensions
- 236 -
User Guide 19 Extensions
- 237 -
User Guide 19 Extensions
- 238 -
User Guide 19 Extensions
3. To change a view of a pane, right mouse click and select one of the views. For a
more specific example, see the chapter "How to search in PDA library" on pg.
244..
4. Numbers ① to ④ correspond to different views in the picture above.
- 239 -
User Guide 19 Extensions
- 240 -
User Guide 19 Extensions
Caution: Increasing the zoom may cause that the markers to be out of the current view. To
move them to the current zoom, right mouse click in the pane and use Move
Markers Here from context menu.
- 241 -
User Guide 19 Extensions
- 242 -
User Guide 19 Extensions
Manage libraries:
l
Create a new library: click the New Library icon.
l
Open existing library: click the Open Library button and in the dialog
choose your desired library.
Add compounds:
1. Add a spectrum into the library: click the Add Spectrum icon which opens
Spectrum Property. In the dialog you can specify compound name and
additional comment.
- 243 -
User Guide 19 Extensions
2. Add all identified peaks: click the command from the menu Spectrum - Add All
Identified Peaks.
Note: Identified peaks are those that correspond to peaks identified and named in the
calibration file. Peaks are added into the currently opened library.
3. Save library: to keep any changes made in the library, click the Save Library
icon.
4. Close library: click the Close Library icon. If you have unsaved changes you
will be prompted to either save the changes or discard them.
View library:
1. Spectral Library View: to view contents of your library, right mouse click in any
pane of PDA Chromatogram windo and choose Spectral Library View. It
contains a table with spectrum name, retention time and other parameters.
- 244 -
User Guide 19 Extensions
- 245 -
User Guide 19 Extensions
- 246 -
User Guide 19 Extensions
1. Used views: Chromatogram View, Spectral Library, Spectral View and Isoplot
View.
Note: How to change views is shown in "How to work with PDA chromatogram"
on page 237
2. Move the marker to your desired position ① .
3. In the Spectral Library View check the Spectrum Name for each ② spectrum
that you want to show in the Spectral View.
4. Spectral View displays current spectrum along with other spectra ③ checked in
the PDA library.
5. You can move the marker ④ in the Isoplot View to move alongside both
wavelength and retention time axes.
Note that those steps can be also performed on results from the Search in Library -
thus having spectra from the search in overlay.
Note: To compare spectra it is recommended to normalize their view. Right click into
Spectral View and select Properties. In PDA Properties dialog navigate to
Chrom&Spectral View tab and select Normalization: At Current Range ⑤ .
- 247 -
User Guide 19 Extensions
19.3 MS Operation
Pick the desired topic in the following chapters.
- 248 -
User Guide 19 Extensions
To be able to add selected compounds (or spectra) to own library, such compounds
must be selected in the Librarian tab. To do so, you can either perform the Single
Compound Search or use Add Spectrum to Library icon or command from the
MS Chromatogram window.
- 249 -
User Guide 19 Extensions
Single Compound Search
After clicking the Single Spectrum Search icon or command, the view will lock in the
graph and will let you select the spectrum. After setting the desired parameters and
clicking the Search button the MS Search program will open. Switch to the
Librarian tab and using the Add to Library button add the spectrum into the
library.
Note: For more details on Single Search Compound please see the chapter "MS
Libraries" in MS Extension manual.
Quantification Signal
Quantification Signal serves for quantification of compounds mentioned in MS
Method. Integration table is unique for each quantification signal. When multiple
compounds are quantified on the same quantification signal they will share the
same table.
l Switch to MS Integration tab in Chromatogram window.
l In the drop-down list select desired compound (defined by m/z ion and the
name). Chromatogram will be focused on the compound's peak.
l Apply integration operations as desired.
Note: Even if you select Result Table Signal, e.g. TIC also as your Quantification Signal
for any of your compounds, the integration table from Result Table Signal will not
be used for such Quantification Signal. Each quantification signal has its own
integration table.
- 250 -
User Guide 19 Extensions
- 251 -
User Guide 19 Extensions
l Select retention time or whole chromatogram to be searched in and hit the
Search button.
l You will get results - exact retention times of peaks with the highest
probability.
l You can filter the results by Filter by Min Match Factor slider/editbox, based
on Match probability.
l Select the desired compound and add such compound to your MS Method.
Background subtraction
Defines exact points and/or intervals of retention time to be omitted from the
search. Such values can be inserted manually or selected directly in the graph.
- 252 -
User Guide 19 Extensions
Here we provide an example on how to create SST method which can be used to
check if the control samples are within expected limits and how to set up sequence
to be checked by different SST methods.
Caution: In order to get desired reactions to passed or failed limits for this scenario it is
necessary to have chromatogram Overlay Mode switched off.
1. Create a SST method in the Chromatogram window using the:
l SST - SST Result to display the SST tab
l SST - New to create a new SST method
l SST - Update from Calib to load the list of peaks from a calibration file
(There has to be calibration linked to the chromatogram on the Results
tab)
l Than the screen will look similar to this:
2. As the second step, fill in the necessary parameters for the limits. (In this case
Tetrachlormethane peak is the one used to check actual against expected value
in Control sample., Expected Amount is 1.2, lower limit is 1.1, upper limit 1.3.)
l Select the checkbox of the Tetrachlormethane in the table on the left.
l Double-click the Amount column in the table on the right to activate it.
l Set the 1.1 value in the Lower Limit cell of the column, 1.3 into the Upper
Limit cell.
l Right-click the table and select the SubParameters item. On it uncheck
the %RSD Limit item (as you do not need it) and select the Each
Individual Value.
- 253 -
User Guide 19 Extensions
More Info:
l When the SST Result is based on Each Individual Value than it
compares each evaluated value with the Upper Limit or Lower Limit.
l When the SST Result is based on Mean of All Values than it
compares average value of all opened chromatograms with linked
calibration with the Limits. So pay attention to all opened
chromatograms.
l You may hide the inactive rows and columns by using the SST - Show All
Columns and SST - Show All Rows items in the menu.
l The result of the check is displayed by the green tick mark or red cross.
You can validate several different parameters at the same time.
l You should see similar result now:
- 254 -
User Guide 19 Extensions
lIt is possible to trigger any external program wanted and set different
events on failed or passed check etc.
4. Save the created method by using the SST - Save item (I used HighControl.sst
file name).
5. You can create other SST methods in the same manner. For example if you use
multiple calibration levels it is possible to create methods which control each
level against its own limits.
l This will force Clarity to open the correct SST method for the row and
perform the check.
Caution: Parameters like %Self% are case sensitive. All 3 checkboxes
Open, Run Program and Include in SST must be checked for
SST to work correctly.
- 255 -