HEXON QUALITY ASSURANCE

LABORATORY
Hexon Quality Assurance Laboratory

Measurement Uncertainty Form


Document No: HEXON/QAL/7.6/F-16

Test Title Validation of Method of Imidacloprid determinations


Test Method CIPAC MODIFIED IMIDACLOPRID METHOD
Method Reference 8.1. CIPAC Vol K page # 70
Test Date 27-03-2019,24-04-2019
Equipment HPLC Method

S # Observations repeatability Reproducibility SD (R)


1 Reading 1 25.03 25.04 -
2 Reading 2 25.06 25.06 -
3 Reading 3 25.04 25.02
4 Reading 4 25.07 24.99
5 Reading 5 25.01 25.04 -
4 Average 25.042 25.030 0.008
5 SD (r) 0.024 0.026 -

Uncertainty of Repeatability i.e. SD (r) 0.026 units


Uncertainty of Reproducibility i.e. SD (R) 0.008 units
Average from MAX SD (r) 25.030 units
Average from SD (R) 25.036 units

UNCERTAINTY MEASUREMENT (Budget)

S# Sources Uncertainty Type A/B Divisor Unc Cont

1 Repeatability 0.024 A 1 0.024


2 Reproducibility 0.026 A 1 0.026
3 Calibration Certificate 0.000 B 2 0.000
4 CRM 0.005 B 1.73 0.003
5 1/2 Resolution 0.0001 B 2 0.00005
6 Dilution (weight/Volume) 0.004 A 1.73 0.00231213873
7
Combined Uncertainty (Uc) 0.036 @ 68 % CL
Confidence Level (K) 2 = 95% CL
Expanded Uncertainty (Ue) 0.071 @ 95% CL
0.100

Result Summary
Expanded Uncertainity 0.100
Upper Result Limit (+) 25.136 units
Average 25.036 units
Lower Result Limit (-) 24.936 units

no need
Imidacloprid 25.036 ± 0.1 this is reported results

Prepared by Lab Assistant Approved by Quality Assurance Manager


Formulas Used in Calculation of Measurement of Uncertainty

S_x = √((∑_( i=1)^( n)▒ 〖


Standard Deviation (x_i-x ̅ ) 〗 ^2 )/(n-1))
_____________________________________

x ̅ = The mean of x_i


n = The number of data points

x_i = Each of the values of the data

CU = √(∑▒ 〖〖 U 〗 _r^2+ 〗
Combined Uncertainty
U_R^2+U_e^2+U_m^2 )
U_r = Uncertainty of Repeatability
____________________________________________

U_R = Uncertainty of
U_e =
Reproducibility

U_m =
Uncertainty of Equipment
Uncertainty of Method