G Recorder User Manual
G Recorder User Manual
INTRODUCTION........................................................................................................................................... 5
INTENDED USE ............................................................................................................................................ 6
INDICATIONS FOR USE.................................................................................................................................. 6
RELATED PRODUCTS................................................................................................................................... 7
RELATED DOCUMENTS ................................................................................................................................ 8
CONVENTIONS ............................................................................................................................................ 9
RELEASE NOTES ...................................................................................................................................... 10
NEW FEATURES......................................................................................................................................... 10
CHANGES ................................................................................................................................................. 10
INSTALLATION AND CONFIGURATION ................................................................................................. 11
HARDWARE AND SOFTWARE REQUIREMENTS ............................................................................................. 11
INSTALLATION FROM A CD ......................................................................................................................... 13
UNINSTALL................................................................................................................................................ 14
FILES ON YOUR COMPUTER ....................................................................................................................... 15
BEFORE RUNNING G.RECORDER .......................................................................................................... 16
RUNNING G.RECORDER IN USER MODE .............................................................................................. 17
RECORDER MAIN W INDOW ........................................................................................................................ 18
Status Bar ........................................................................................................................................... 20
Toolbar ............................................................................................................................................... 21
DATA VIEWER ........................................................................................................................................... 24
FEATURE VIEWER ..................................................................................................................................... 25
The Feature Viewer review mode during data acquisition ................................................................. 26
EVOKED POTENTIALS VIEWER ................................................................................................................... 28
LOAD CONFIGURATION .............................................................................................................................. 29
RECORDING INFORMATION......................................................................................................................... 33
DISPLAY SETTINGS.................................................................................................................................... 36
CHANNEL SETTINGS .................................................................................................................................. 37
VIDEO CAMERA ......................................................................................................................................... 39
START AND STOP RECORDING ................................................................................................................... 40
MARKERS ................................................................................................................................................. 41
DIGITAL INPUTS......................................................................................................................................... 43
Individual Triggers .............................................................................................................................. 43
Combined Triggers ............................................................................................................................. 43
LOAD ELECTRODE MONTAGE ..................................................................................................................... 45
LOAD, FIND, CONVERT AND REPLAY RECORDED FILES ............................................................................... 46
Load, Find and Replay HDF5 files ..................................................................................................... 46
CONVERT G.MOBILAB+ BIN FILES TO HDF5 FORMAT................................................................................. 54
PRINT RECORDED FILES ............................................................................................................................ 55
CLOSING THE REVIEW MODE IN G.RECORDER ............................................................................................. 57
CLOSING G.RECORDER ............................................................................................................................. 58
G.USBAMP SPECIFIC FEATURES ................................................................................................................ 59
Impedance Measurement with g.USBamp and g.Recorder ............................................................... 59
Calibration of g.USBamp with g.Recorder ......................................................................................... 60
G.HIAMP SPECIFIC FEATURES.................................................................................................................... 62
Impedance Measurement with g.HIamp and g.Recorder................................................................... 62
Calibration of g.HIamp with g.Recorder ............................................................................................. 63
G.NAUTILUS SPECIFIC FEATURES............................................................................................................... 65
Impedance Measurement with g.Nautilus and g.Recorder ................................................................ 65
Calibration of g.Nautilus with g.Recorder ........................................................................................... 66
RUNNING G.RECORDER IN ADMINISTRATOR MODE .......................................................................... 68
Our team is prepared to find the best solution for your needs.
Reduce development time for sophisticated real-time applications from month to hours.
Integrate g.tec's open platform seamlessly into your processing system.
g.tec's rapid prototyping environment encourages your creativity.
Scientist
Are you planning an experimental study in the field of brain or life sciences? We can offer consultation in
experimental planning, hardware and software selection and can even do the measurements for you. If
you have already collected EEG/ECG/EMG/EOG, g.tec can analyze the data starting from artifact control,
do feature extraction and prepare the results ready for publication.
g.Recorder is a novel, user-friendly software package for the recording of biosignal data such as EEG,
ECoG, ECG, EOG, EMG and sensor data. Trigger information and video sequences can also be
captured. An easy-to-use Replay mode allows for the inspection of data after recording.
g.Recorder has been designed to work with the g.tec biosignal amplifiers, g.HIamp and g.USBamp, and
with the portable devices g.MOBIlab+ and g.Nautilus.
g.Recorder offers raw data inspection and also the extraction and presentation of different parameters.
These include, but are not limited to, heart rate variability, heart rate, compressed spectral arrays, and
cerebral function.
g.Recorder data can be exported seamlessly to g.tec’s biosignal analysis software, g.BSanalyze, for in-
depth offline evaluation and data processing.
The device must not be used for patient monitoring. The device must not be used for the determination of
brain death. Additional examinations are needed for diagnosis, and no diagnosis may be done only based
on using this device.
No part of this manual may be copied or reproduced without the express permission of the authors.
g.tec provides several biosignal analysis elements that are especially relevant to the kinds of tasks you
perform with g.Recorder and g.tec’s biosignal amplifiers.
For more detailed information on any of our elements, updates or new extensions please visit our
homepage www.gtec.at or just send us an email at office@gtec.at.
g.Recorder Parallel Port Qickstart – A short introduction on how to use stimulus delivery software
E-Prime® (Psychology Software Tools, Inc., www.pstnet.com) and Presentation® (Neurobehavioral
Systems, Inc., www.neurobs.com) over the parallel port with g.HIamp or g.USBamp devices and
g.Recorder.
g.Recorder Evoked Potentials Quickstart – A short introduction on calculation and display of evoked
potentials in g.Recorder with g.USBamp devices using g.HIsys for stimulus presentation.
Furthermore, g.tec can provide tutorials on how to load g.Recorder data into EMSE® Suite, BESA®, and
EEGLAB/ERPLAB Toolbox.
New features
• Biosignal amplifier g.Nautilus is now supported.
Changes
• Windows 10, 64-bit is now supported; Windows 7 support retired.
• Updated g.HIamp impedance measurement window and corresponding description according to
new driver version.
• The g.HIamp configuration dialog shows an information message now if channel-specific settings
should be applied but no channels are highlighted.
• Added performance hint to feature definition section in manual (regarding CFM/CSA).
• The system requirements have changed.
Hardware Properties
CPU 2 GHz or faster processor
Harddisk 30 GB or higher
RAM 4 GB or higher
USB 2.0 port One free USB port for the hardlock dongle
One free USB port for each g.HIamp device
One free USB port for each g.USBamp device
USB 2.0 port (optional)
One free USB port for each g.Nautilus device
One free USB port for attaching a color printer
Bluetooth® interface2 (optional) Bluetooth® interface for g.MOBIlab+ devices
Screen resolution 1920 x 1080 pixels
Video camera Webcam
g.Recorder acquires data from g.tec’s biosignal amplifiers. Depending on the used devices, the
corresponding drivers for g.HIamp, g.USBamp, g.Nautilus (g.NEEDaccess) or g.MOBIlab+ must be
installed. Depending on the Windows® operating system, administrator rights might be necessary for the
installation.
Software Version
Windows 10 Pro
Microsoft® Windows® English
64-bit
g.HIamp driver 2.16.00
g.USBamp driver 3.16.00
g.MOBIlab+ driver 1.16.00
g.MONcreator 5.16.00
g.NEEDaccess 1.16.01
Sentinel HASP® Runtime Environment3 7.50
Adobe® Acrobat® Reader® 4 DC 2015
1 Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other
countries.
2 The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. Other trademarks and trade names
are those of their respective owners.
3HASP®, Sentinel HASP®, Hardlock®, are registered and/or unregistered trademarks of SafeNet, Inc., or Aladdin Knowledge
Systems, Ltd., in the United States and other countries.
4Adobe, Acrobat, and Reader are either registered trademarks or trademarks of Adobe Systems Incorporated in the United States
and/or other countries.
1. Insert the g.tec product CD, open the g.Recorder directory on the CD, open the correct
directory for the architecture of your PC (Win64) and double-click setup.exe. If User Account
Control is turned on, additional dialogs may ask for permission. Confirm the dialogs to allow
installation of g.Recorder through User Account Control. The installation starts and displays the
welcome message.
2. Click Next to read the License Agreement. If you agree with the terms, click I Agree and Next.
3. Choose the installation folder (default is C:\Program Files\gtec\) where the install routine
should copy all necessary files and press Next.
4. Follow the instructions on the screen to install g.Recorder. When the following window informs
you about completion of installation, click Close to complete.
If the computer is connected to the internet, let Windows Update download and install the correct
driver for the hardlock. For manual installation, run the HASPUserSetup.exe from the
prerequisites/HASP HL folder on your g.tec CD.
Uninstall
To remove g.Recorder in Windows 10, use the Programs and Features tool in the Control Panel.
Note that you have to uninstall the hardlock driver software separately if you uninstall the g.Recorder
software.
C:\Program Files\gtec\gRecorder
C:\Program Files\gtec\gRecorder\Config
C:\Program Files\gtec\gRecorder\Help
In the g.Recorder Menu and Toolbar you can select and set different recording parameters, filenames
and start/stop the actual measurement, etc.
The Data Viewer window shows the raw biosignal data. Using the menu and toolbar window, the number
of channels and the seconds on screen can be adapted, etc. The channel information (number and
name) and the channel scaling is displayed on the left side of the window.
Selected parameters, for example, heart rate and cerebral function are displayed in the Feature Viewer
window. The parameters are displayed from left to right. On the left side of the window the y-axis scaling
as well as the specific feature name are displayed.
The Evoked Potentials viewer displays calculated evoked potentials for each acquired raw data channel,
if configured. Zooming, scrolling and scaling for the evoked potential plots can be adapted directly in this
viewer.
You can resize all windows according to your preferences and needs. Individual windows can be hidden
to enlarge for example the area of the Feature Viewer.
Note: The actual layout of the screen and the specific settings are stored to a default configuration file
when g.Recorder is closed. This default configuration is used the next time you start g.Recorder. This
allows easy continuation of your studies. You can of course load another configuration.
Data Viewer
Feature Viewer
Evoked
Potentials
Video Viewer
Status Bar
The actual status of g.Recorder is displayed in the Status Bar at the bottom of the main window. The
Status Bar shows information that relates to the current recording. The first field shows the current
g.Recorder Mode. When data are stored to the hard disk the mode is set to Recording. If data is loaded
for off-line visualization it is set to Replay and otherwise to Standard.
The Toolbar is always shown below the Menu and includes functions for file opening, printing, starting
and stopping the data acquisition and video capturing, setting markers and navigating to a certain instant
in the data.
Open…
Load an already recorded file for offline replay
Print Features…
Print selected features
➢ Note: All data are still stored to the hard disk if this
button is pressed in Recording Mode
Record
Start data acquisition to harddisk. Depending on the
selected options raw data from the amplifiers, the video
and features are saved.
Stop
Stop data acquisition
Zoom Out and Zoom In into the raw data in the Data
Viewer.
The percentage can also be inserted directly by typing in
the desired zooming level
Signal Cropping
Enable or disable signal cropping. If signal cropping is
enabled, the signal will be cropped at its displayed vertical
channel range. If signal cropping is disabled, the data of
the displayed channels may overlap the other channels'
ranges. Examples are given below. Signal cropping is
disabled by default.
Visualization-only Filters
Customizable digital filters can be applied to the analog
input channels in the data viewer for visualization purposes
only. The filters are not applied to the recorded or analyzed
data.
To apply filters on recorded and analyzed data, use the
configuration dialog of the selected amplifier instead.
+ 185.0 µV
0 µV
- 189.0 µV
The number on the left side of the green box represents the channel number used on the amplification
device. This information is useful if more than one biosignal amplifier is used. Here, channel 4 of the first
amplifier 1 is displayed. The number on the right side of the green box is the logical channel number in
the actual recording setup. In this case the logical channel number is also 4.
The gray box shows the scaling of the channel. The minimal and maximal input amplitude range
displayed on the screen is given by the 2 numbers. The horizontal dotted gridline represents 0 µV. The
minimal and maximal values can be adapted via the menu Settings/Channels.
The light blue box shows the channel name and channel type. The text on the left side gives the chosen
channel name. Here the selected channel name was FZ. The text on the right side shows the selected
channel type (e.g. EEG). The channel name and channel type information can be set via the menu
Settings/Channels.
The zooming level of the Data Viewer as well as the color for the paper, pen and grid can also be defined
via a context menu by clicking with the right mouse button of a specific channel in the Data Viewer.
The actual time is given on top of individual feature panels as the time on screen can vary depending on
the specific user settings. The format of the time is hh:mm:ss. The y-axis scaling and units depend on the
specific selected feature. The minimal and maximal values can be adapted via the feature specific context
menu. The y-scaling for the CFM feature is fixed to 0 and 100 µV. The scaling for the HR feature is set to
0 and 200 beats per minute [bpm].
Running time
Name of selected
feature: CFM [µV]
Name of selected
feature: HR [bpm]
The scaling values of the Heart Rate feature can be set via a context menu by clicking on the right mouse
button in the feature panel.
The Feature Viewer review mode allows for the inspection of previous epochs of feature signals. If the
time on screen is elapsed, then a menu entry appears in the upper right corner of the Feature Viewer
window.
The Previous Output review panel allows the user to scroll trough past feature signals. The upper panel
area displays the already recorded features and the lower panel area yields more specific information
about the time interval Window, the File name of the actual recording and Subject specific information.
It is possible to configure the scale settings, to scale the data to the individual minimum/maximum of
each acquired channel or to the global minimum/maximum of all acquired channels.
It is possible to change channel names, signal type, units and sensitivity range in the channel settings
dialog. Channel names can be loaded from montage files.
By default, all plots are arranged in a square matrix. It is possible to change rows columns and
arrangement in the arrangement settings. Topographic arrangement of the plots can be loaded from
montage files.
The color settings allow the user to change the appearance of all elements of the evoked potential plots.
c:\Program Files\gtec\gRecorder\Config
g.Recorder will be initialized with all necessary hardware and software settings.
Previously saved configurations can be loaded by clicking on the Load Setup… button in the File menu.
The picture above displays the front side of the g.USBamp. The front side is divided into four groups
(A,B,C,D from left to right). Each group consists of four red connectors (1-4, 5-8, 9-12, 13-16) for
connecting biosignal electrodes (EEG, ECG, EMG, etc.), one blue connector for the reference electrode
(R) and one yellow connector for the ground electrode (G).
For EEG measurements, all electrodes for EEG are connected to Group A (first group on the left side of
the g.USBamp). One EEG electrode is connected to channel 1 of Group A (red connector), the reference
electrode to reference R of group A (blue connector) and the ground electrode is connected to ground G
of group A (yellow connector).
A Path for the data file can be entered or the desired path can be chosen by clicking on the Browse
button.
The Filename defines the name of the data file. The default filename is RecordSession_ with date and
time added. When the measurement is started the data and the time are added automatically to the
filename (e.g. RecordSession_2016.11.09_11.36.02.hdf5). The data is stored in the HDF5 file
format. For more information about this format see www.hdfgroup.org or refer to the MATLAB help.
The Record Raw Data checkbox toggles whether the raw data from the amplifier is stored to the HDF5
file or not. For offline processing after measurement it is recommended to record raw data.
Recording Time
IA time limitation for the measurement can be set with the Enable Time Limitation checkbox in the
Recording Time section. After this time, the measurement is stopped automatically.
Information about the subject like Last Name(s), First Names(s) and Date of Birth can be entered in the
Subject section. Additional information (e.g. medications) can be added in the Comment section.
Session Data
Note: Mandatory fields (the keyword mandatory is displayed next to this field) must be filled out correctly.
Otherwise the recording cannot be started.
The Color Settings can be set in the Data Viewer dialog. The Pen color sets the color of the data
recorded, the Paper color sets the background color and Grid color sets the color of the grid. It is
possible to change the settings of each channel individually in the Individual color settings group, or for
multiple channels at once in the Common color settings group.
In the Grid settings area the grid can be enabled and the number of vertical grid lines can be selected.
The normal time scale of the Data Viewer is the elapsed measurement time. If the Show absolute time
checkbox is checked, then the system time is shown.
Individual settings
The first column, Device-Ch Nr., shows the device number and also the channel number (e.g. 1-1 for the
first amplifier and channel 1, 2-1 for the second amplifier and channel 1). The second column, Logical
Ch. Nr., shows the logical channel number. If two amplifiers with each 16 channels are used then the
Logical Ch Nr. goes from 1 to 32.
In the Name column a name for the channel can be entered and in the Type column EEG, ECG, EMG,
EOG, ECoG, Resp., TRIG can be selected. With the channel type selection the Sensitivity of the
channel is changed automatically.
The Sensitivity and an Offset of each channel can be entered manually. The Unit can be µV, mV or V.
This Sensitivity affects only the visualization of the raw data.
The same settings can be applied to multiple channels at once in the Common settings section.
To only apply a Type to a range of channels, select the Start channel and the End channel and the
Type in the upper half and click Apply next to it.
To apply a common sensitivity, an offset and a unit to a specific Type of channels, select the lower
sensitivity (Sens L.), the upper sensitivity (Sens H.), the Offset and the Unit and choose the Type of the
channels you want to apply those settings on, and click Apply.
To see the video images in the main window of g.Recorder, select Video Viewer from the View menu.
The Data Viewer, the Video Viewer and the Feature Viewer can be automatically arranged by selecting
Arrange in the Window menu.
To display data in the Data Viewer without saving it, click on the View Data button in the Toolbar or
select Start Data Viewing from the File menu.
Note: This button is also used to start streaming to SD Card with g.MOBIlab+.
To record data to the hard disk on your PC, click the Record button or select Record from the File menu.
To stop data acquisition in Standard mode or pause the data display in the Recording mode, press the
Pause Data Viewing button or select Pause Data Viewing from the File menu.
Note: If the button is pressed in Recording mode, only the visualization of the data on the display is
stopped, but the g.Recorder continues to store data to the HDF5 file.
Click on the Stop button or select Stop from the File menu to stop data acquisition.
The Pause Data Viewing button is the only button activated if a g.MOBIlab+ is streaming to SD Card.
If g.Recorder is used to reconnect to a streaming g.MOBIlab+, only the Start Data Viewing button is
activated.
Note: For g.MOBIlab+, data can either be stored the local hard disk of your PC or streamed to the built-in
SD card.
Here, two markers are defined that can be set either by clicking on the marker button in the Toolbar of
g.Recorder or by pressing the keys on the keyboard. Marker e – Event is a Toggle marker, marker m –
Movement is a time marker.
The window below shows the visualization of the markers in the Data Viewer:
The set markers for the different events are also indicated by small bars near the top of the individual
features channels in the Feature Viewer:
Some types of devices allow the user to choose if each of those trigger channels should be treated
individually or if several trigger channels should be combined.
Individual Triggers
By default, trigger channels selected for acquisition are treated individually. Depending on their
configuration, each occurrence of such a trigger channel is represented as a separate event with a value
of HIGH (H) for rising edges or LOW (L) for falling edges.
Combined Triggers
Combined triggers are supported by g.USBamp and g.HIamp devices only. In this mode, all trigger
channels of a device are represented by the same event for that device and the value is a binary coded
decimal made up of the current high or low values of the individual trigger channels. The triggers are
represented as vertical lines including a description consisting of device's serial number, combined trigger
tag, event code and name.
This enables reception and interpretation of binary coded trigger values sent via the parallel port.
g.Recorder supports external stimulation software like Presentation® (Neurobehavioral Systems, Inc.,
www.neurobs.com) and E-Prime® (Psychology Software Tools, Inc., www.pstnet.com) by acquiring event
codes sent via parallel port.
Montage files can be loaded by selecting Load Electrode Montage… in the File menu.
The Data Viewer assigns channel names to the acquired analog channels. The Evoked Potentials
Viewer, if enabled, arranges the evoked potential plots in a topographic way according to the 2D
electrode positions in the file.
A recorded data file in HDF5 format can be loaded and reviewed with g.Recorder.
Then select the data file you want to load into g.Recorder. The default path of the recorded files is
My Documents/gRecorder
This displays all HDF data files found in your specific working directory.
Define Datasets enables optionally searching for data sets including Last Name and First Name and if
selected the Date of Birth of a subject. Results are displayed immediately after typing in the first letter.
Found Datasets yields information about the search request.
The desired data set can be selected by highlighting the data set. Double clicking or selecting the OK
button loads the data set for review. Additional short information is given for the data set including the
Last name, first name, Date of Birth, Recording data, Recording length, Filename and if a video
information is available. Data sets can be sorted by clicking on one of the row headers e.g. on the Video
entry.
The entire table can be copied to the clipboard for further use by clicking the Copy Table button.
The Mode is Replay, the Sampling frequency is 256 Hz, the number of recorded channels (Total Chan.)
is 16. The File Name is Record_Session_2016.02.18_10.07.09.hdf and the File Size is 0.413
megabytes.
You can use the scroll bars on the bottom of a window or the orange vertical position marker in any of the
g.Recorder windows to move to an arbitrary position in the file.
To review data on-line from the selected position, click on the Start Data Viewing button.
Visualization Filters can be applied to the recorded analog signals in Replay mode. A progress bar is
displayed for filtering and rendering a large amount of data.
This opens the File Info window yielding information about the data set. To view the file information use
the Select Section pull down menu (the list contains Raw Data, DeviceInfo, Session Description,
Subject Description, Marker Description and the recorded features).
In the Files of type section of the Open dialog select g.MOBIlab+ Files (*.BIN) files and browse to the
folder containing the file to convert. Select the desired file and click Open.
The Save HDF5 File As dialog will open, where the path for the converted HDF5 file can be set.
Scale
The Scale of the printed feature can either be in cm/h (centimeters per hour) or min/page (minutes per
page). When selecting the setting 6 cm/h a dataset of 3 hours would need a total space of 18 cm on the
paper. If the same 3-hour data set is printed with the selection of 60 min/page then 3 pages would be
printed.
The time scale in the printed document can be set with the checkbox Relative Time Annotation. The
time is given in absolute format (e.g. 17:05:01) or relative to the measurement time (e.g. 00:05:14).
Select the Print Document Abstract to print the file and subject information.
Feature Selection
Click Next to go to the Print dialog, select a printer and print the file.
Note: If you are in Replay mode you must close the file first with the Close function in the File menu
before you can exit g.Recorder.
The impedance of electrodes connected to a g.USBamp can be measured in g.Recorder. To start this
tool, click on the Impedance Measurement… button in the Tools menu.
Select the channels for impedance checking individually by clicking on the boxes next to the channel
numbers and click Start. The channel coloring indicates the impedance of the connected electrode.
Optimal values (e.g. values below 5 kilo-ohm for EEG recordings) are color-code green.
Acceptable impedance values are in the range of up to 20 kilo-ohm (color code yellow or red). Non-
connected channels or electrodes with impedance values of more than 20 kilo-ohm are color coded blue.
You can also click on the Select All button to include all electrodes for impedance checking or deselect
the electrodes pressing the Select None button.
In order to accelerate the impedance check, you can check the Measure above green checkbox. This
will only check impedances on electrodes with impedance values greater than 5 kilo-ohm.
Note: For EEG measurements the impedance should be below 10 kilo-ohm for optimal recording quality.
To exit from the Impedance Measurement dialog click on the Close button.
Note: The impedance of electrodes connected to channels 1-16 and the reference channels are
measured against the ground G of each group. To measure the impedance of all 16 channels and 4
references against only one ground connect the ground electrode to ground G of group D.
Press the Apply button to transmit the settings to the g.USBamp and to close the dialog.
The impedance of electrodes connected to a g.HIamp can be measured in g.Recorder. To start this tool,
click on the Impedance Measurement… button in the Tools menu.
Select the type of electrode connector boxes that are currently connected to each of the four groups of
the g.HIamp in the Electrode Types area. Either passive or active electrode connector boxes can be
used. All groups must use the same type of electrodes even if no box is connected to a specific group. At
least channel 1 (Group A) and the ground electrode must be connected.
IMPORTANT: If an active electrode box is connected to group A, a special electrode has to be used at
channel 1. Please see the documentation of the 64-channel active electrode connector box for details on
impedance measurement.
Select or deselect the channels to use for impedance measurement by clicking on the appropriate
channel number. Deselected channels get a dark gray background while selected channels have a white
background.
The channel coloring indicates the impedance of the connected electrode. Channels with passive
electrodes get assigned one of the colors listed in the Passive Electrodes box. Channels with active
electrodes get assigned one of the colors listed in the Active Electrodes box.
Optimal values e.g. for EEG recordings with passive electrodes should be below 5 kilo-ohm (color-code
light green). Acceptable impedance values are in the range of up to 20 kilo-ohm (color code light yellow or
light blue). Bad impedance values above 20 kilo-ohm are color coded light red. Channels whose
impedance couldn’t be measured properly are color coded black.
Optimal values e.g. for EEG recordings with active electrodes should be below 30 kilo-ohm (color code
green). Acceptable impedance values are in the range of up to 100 kilo-ohm (color code yellow or blue).
Bad impedance values above 100 kilo-ohm are color coded red. Channels whose impedance couldn’t be
measured properly are color coded black.
Channel 1 and the ground electrode are used as the reference for impedance measurement of all other
channels. No impedance can be measured for those channels. Hence, channel 1 is always grayed and
no impedance value will be assigned. However, if the reference signal measured on channel 1 is not valid
(which might occur when the skin contact is too poor, or when no electrode is connected on that channel,
or when artifacts mess up the measured signal on that channel), no impedance value can be calculated
and color code black will be assigned to all other channels.
Otherwise, if the reference signal measured on a specific channel is not valid (which might occur when
the skin contact is too poor, or when no electrode is connected on that channel, or when artifacts mess up
the measured signal on that channel), no impedance value can be calculated and color code black will be
assigned to that specific channel.
Click Stop to stop impedance measurement (even if impedance measurement has aborted with a
connection error, for example if the device was disconnected during measurement) and close the dialog
by clicking Close.
If Acoustic feedback is checked, a short beep is produced on the PC speaker when the impedance of
any of the electrodes changes from red to any better impedance value. The better the impedance value
the electrode changes to, the higher the frequency of the beep. If the impedance of any electrode
changes to red, a long beep sounds. To turn off acoustic feedback, uncheck the Acoustic feedback
checkbox.
Check Show channel names if the names of the channels should be displayed instead of their numbers.
This only works for channels that got a name assigned before in the Channel Settings dialog. Nothing
additional is displayed for channels that got no name assigned that way.
g.Recorder has a built-in function to calibrate a g.HIamp device. Select Calibration… from the Tools
menu.
Calculation of new calibration values lasts about 10 seconds. A dialog will inform you after calculation has
been completed. The actual calibration values are then visualized for each channel, but not applied yet.
If at least one channel could not be calibrate due to broken channels or a broken calibration set, for
example, a dialog will inform you about the involved channels and their calibration value entries are
displayed as NaN (“not a number”).
To restore the initial factory calibration values, click Restore Factory Settings.
Press the Apply button to transmit the newly calculated or restored factory calibration values to the
g.HIamp and apply them on the device. Press the Restore button to restore the old values on the
amplifier. To edit the calibration values manually, double-click the desired cell and enter your value for
scaling and offset (the settings will actually be transmitted and applied after the Apply button is pressed).
For channels that received NaN values as calibration result, 1.0 will be set for scaling and 0.0 for offset on
pressing the Apply button.
The impedance of electrodes connected to a g.HIamp can be measured in g.Recorder. To start this tool,
click on the Impedance Measurement… button in the Tools menu.
Select or deselect the channels to use for impedance measurement by clicking on the appropriate
channel number. Deselected channels have a dark gray background, while selected channels have a
white background.
Click Start to start impedance measurement. Impedance measurement is performed periodically for all
channels simultaneously, and lasts a few seconds for each run.
The channel coloring indicates the impedance of the connected electrode. The color code for the
impedance measurement is listed in the Active Electrodes box.
Optimal values e.g. for EEG recordings with active electrodes should be below 30 kilo-ohm (color code
green). Acceptable impedance values are in the range of up to 100 kilo-ohm (color code yellow or blue).
The Cz electrode and the ground electrode are used as the reference for impedance measurement of all
other channels. If the reference signal measured on channel Cz is not valid (which might occur when the
skin contact is too poor, or when no electrode is connected on that channel, or when artifacts affect the
measured signal on that channel), no impedance value can be calculated and color code black will be
assigned to all other channels.
Otherwise, if the reference signal measured on a specific channel is not valid (which might occur when
the skin contact is too poor, or when no electrode is connected on that channel, or when artifacts affect
the measured signal on that channel), no impedance value can be calculated and color code black will be
assigned to that specific channel.
Click Stop to stop impedance measurement (even if impedance measurement has aborted with a
connection error, for example if the device was disconnected during measurement) and close the dialog
by clicking Close.
If Acoustic feedback is checked, the PC speaker presents a short beep when the impedance of any of
the electrodes changes from red to any better impedance value. The greater the improvement in the
impedance value, the higher the frequency of the beep. If the impedance of any electrode changes to red,
a long beep sounds. To turn off acoustic feedback, uncheck the Acoustic feedback checkbox.
Check Show channel names if the names of the channels should be displayed instead of their numbers.
This only works for channels that got a name assigned before in the Channel Settings dialog. Nothing
additional is displayed for channels that got no name assigned that way.
Check Show impedance values in kΩ to display the measured impedance in kilo-ohm. If the measured
impedance value is very low, LOW is displayed instead of a value in kilo-ohm. If the impedance value is
very high, HIGH is displayed instead of a value in kilo-ohm.
This opens the g.Nautilus calibration dialog. Click the Calibrate Amplifier button to measure offset and
gain of each channel. Confirm the following dialog reminding you to apply the calibration set by clicking
OK.
A dialog will inform you after calculation has been completed. The actual calibration values are then
visualized for each channel, but not applied yet.
To activate the Administrator mode just click OK in the Administrator Authentication dialog. If a
password is set, enter the password and click OK.
To change the password, click on Change Password and enter the Old Password. Then, press the
Enter key and enter the New Password, then enter the password again in the Retype new Password
field and click OK.
Select the desired g.HIamp device that should be used for measurement from Available Hardware and
click on the arrow button. This g.HIamp and its serial number are displayed in the Selected Hardware
box.
The configuration dialog opens with the amplifier’s serial number (here e.g. HA-2016.02.01):
OPTIONS
Master set the amplifier to master mode.
Enable Trigger indicates if the digital trigger input line is acquired as well. This checkbox will
be enabled automatically if at least one trigger line is selected to acquire (see
the digital IN settings section for further details). If this checkbox gets disabled
by the user, the trigger input line will not be acquired.
Enable Counter a sample counter is applied on the first acquired channel, which overruns at
1000000.
Enable Hold enable/disable the signal hold function.
TEST MODE
Activate Test Signal enable/disable the internal test signal generator, which generates a test signal
using the specified settings on the GND (ground) connector.
Wave Shape specify the wave shape of the test signal.
Frequency [Hz] specify the frequency of the test signal.
Amplitude [µV] specify the amplitude of the test signal in microvolts.
Offset [µV] specify the offset of the test signal in microvolts.
CHANNEL SETTINGS
Bipolar perform a bipolar derivation between 2 input channels.
Bandpass perform a digital bandpass filtering of the input channels.
Notch perform a bandstop filtering to suppress the power line frequency of 50 Hz or
60 Hz.
1. Highlight the channels in the list box that should be edited. Use the Ctrl key or the Shift key to
highlight multiple channels. To select all channels, use Ctrl+A. To highlight all channels of one
group, click the group header. Highlighted channels are represented with an orange background
color, as illustrated for channel 3.
2. Check or uncheck the checkbox of one of the selected channels in the Acquire column to include
the highlighted channels in data acquisition or exclude them.
3. If no common reference has been set before, or you want to change the reference/bipolar
channel for a specific channel, select the desired bipolar channel from the Bipolar list and press
the corresponding Apply to highlighted button right below.
5. Select a notch filter from the Notch list and press the corresponding Apply to highlighted button
right below.
If no channel is highlighted when one of the Apply to highlighted buttons is clicked, the system instead
informs the user of this error and the settings do not change.
Click OK to finish the configuration of the g.HIamp and apply it to the device.
In the Settings menu, click on g.HIamp Digital IN… to open the configuration dialog for the digital I/O
settings.
The Trigger inputs are sampled synchronously with the analog input channels.
If the event occurs, a marker is generated and visualized with the biosignal data.
If Combined Triggers are selected, all individual trigger channels are enabled and the individual channel
selection becomes disabled. Combined triggers are always configured to detect rising edges. It is
possible to change the Color and Description for combined triggers. The displayed marker name and
color for a combined trigger are always the same. The marker's value will be the binary coded decimal
made up of the current HIGH and LOW values of all individual trigger channels with trigger channel 1 as
least significant bit and trigger channel 16 as most significant bit.
From Available Hardware, select the g.USBamp device that should be used for the measurements and
click on the arrow button, as shown in the figure above. This g.USBamp with the serial number is now
displayed in the Selected Hardware box. Repeat these steps If multiple g.USBamps are connected to
the PC.
The configuration window opens with the amplifier serial number (here e.g. UA-2007.10.15):
AMPLIFIER SETTINGS
Common Ground check the ground that should be connected to common ground
Common Reference check the reference that should be connected to common reference
Channel Selection check the analog input channels that should be acquired
Options
Mode
Amplitude specify the amplitude of the test signal (max: 244 mV)
CHANNEL SETTINGS
1. Highlight the channels in the listbox that should be edited. Use the Ctrl key or the Shift key to
highlight multiple input channels.
2. Select the Bandpass filter HP: 0.5 / LP: 30 and press the Apply button to assign the bandpass to
the highlighted channel(s). The selection is shown in the listbox.
3. Select the 50 Hz Notch filter to suppress the power line interference at 50 Hz and then press the
Apply button
4. To perform a bipolar derivation between channels 1 and 2, select the first channel in the listbox.
Then select channel 2 under Bipolar and press the Apply button. The settings appear in the
listbox. This configuration subtracts channel 2 from channel 1 and the bipolar derivation will be
visible on channel 1.
NOTE: Select 0 under Bipolar and assign it to the channel if no bipolar derivation should be
performed
g.USBamp 2.0 has 3 digital inputs and 2 digital outputs. In the Settings menu click on g.USBamp Digital
IO… to open the dialog.
The Trigger input is sampled together with the analog input channels and therefore gives a synchronous
response. The inputs and outputs D1, D2, D3 and D4 are sampled asynchronously.
Check the box of the desired channels, enter a Description, select a color (Color Sel.) and define the
edge (Edge Detection).
If the event occurs, a marker is generated and visualized with the biosignal data.
If Combined Triggers are selected, all individual trigger channels are enabled and the individual channel
selection becomes disabled. Combined triggers are always configured to detect rising edges. It is
possible to change the Color and Description for combined triggers. The displayed marker name and
color for a combined trigger are always the same. The marker's value will be the binary coded decimal
made up of the current HIGH and LOW values of all individual trigger channels with trigger channel 1 as
the only bit in that value.
NOTE: The Trigger input must be used if very accurate timing is needed.
g.USBamp 3.0 has 8 digital inputs and 4 digital outputs. In the Settings menu, click on g.USBamp
Digital IO… to open the dialog.
The trigger inputs Trigger 1 to Trigger 8 are sampled synchronously with the analog input channels. The
outputs Out 1 to Out 4 are sampled asynchronously.
If the corresponding trigger event occurs, a marker is generated and visualized along with the biosignal
data.
If Combined Triggers are selected, all individual trigger channels are enabled and the individual channel
selection becomes disabled. Combined triggers are always configured to detect rising edges. It is
possible to change the Color and Description for combined triggers. The displayed marker name and
color for a combined trigger are always the same. The marker's value will be the binary coded decimal
made up of the current HIGH and LOW values of all individual trigger channels with trigger channel 1 as
least significant bit and trigger channel 8 as most significant bit.
From Available Hardware, select the g.MOBIlab+ that should be used for the measurements and click
on the arrow button as shown in the figure above. Now this g.MOBIlab+ with the serial number and the
COM port number is displayed in the Selected Hardware box. The COM port of your g.MOBIlab+ can be
found in Bluetooth Devices in the Control Panel.
Info
In this section, the COM port where g.MOBIlab+ is connected and the serial number of g.MOBIlab+ are
shown.
The channels to be recorded can be selected in the Analog Channels section. The Sensitivity, the
highpass HP, the lowpass LP and the Polarity are displayed. Additionally a 50 or 60 Hz Notch filter can
be applied.
Sampling Frequency
SD card
If a SD Card is inserted into g.MOBIlab+, then the SD card status is displayed in the SD Card section
including the space left on the SD Card.
Operation Mode
The checkboxes Generate test signal and Enable streaming are located in the Operation Mode
section on the bottom of the dialog.
With the Generate test signal checkbox selected, g.MOBIlab+ will output a test signal consisting of
ramps with different slopes on the eight analog channels and the digital channels will toggle at different
rates. This function is implemented to test the connection between g.MOBIlab+ and PC or laptop. It is
recommended that the markers are turned off for this function.
The Enable streaming to SD card checkbox is only enabled if a SD Card is inserted into g.MOBIlab+. If
the checkbox is selected then it is possible to disconnect g.MOBIlab+ from the g.Recorder environment
and the device will continue to stream data to the built-in SD Card. If a connected g.MOBIlab+ is already
streaming data then this Status is also displayed in the Operation Mode section.
g.MOBIlab+ provides eight digital channels. To configure the digital inputs and outputs click on the
g.MOBIlab+ Digital IO… button in the Settings menu.
Perform the settings as shown in the figure above and close the dialog with OK. Then the following
controls will be visible in g.Recorder:
Click on the buttons to set the markers. All markers are stored together with the biosignal data. The digital
inputs and outputs are sampled together with the biosignal data.
If Enable streaming in the g.MOBIlab+ dialog is selected and the data acquisition is started by clicking
on the Display button the g.MOBIlab+ Start Streaming to SD Card dialog will appear.
If g.MOBIlab+ is already streaming, the Status in the Device section will be set to Streaming. The SD
Card section will be disabled.
The Start Streaming to SD Card dialog asks the user to provide a Filename for the data recorded to the
SD card.
To confirm streaming to SD card click Yes. Then data are displayed in g.Recorder for control purposes
and data are streamed to the built-in SD card on g.MOBIlab+
Click No to simply display data in g.Recorder.
When the data acquisition is stopped, then g.MOBIlab+ Stop Streaming To SD card dialog will appear.
To stop the data transfer to the PC and data acquisition to the SD Card in g.MOBIlab+ click Yes. To
continue streaming data to the SD Card click No.
If g.MOBIlab+ is already streaming data to the SD card, then you can reconnect the device to
g.Recorder. An information dialog will appear. Click OK to display data again from g.MOBIlab+ in
g.Recorder.
Select the g.Nautilus that should be used for the measurements from Available Hardware and click on
the arrow button as shown in the figure above. Now this g.Nautilus with the serial is displayed in the
Selected Hardware box.
AMPLIFIER SETTINGS
Input Source specify the input source for data acquisition.
Electrode: record EEG data
Shortcut: record with shortcut of inputs
Testsignal: enables the internal test signal generator, which generates a
square wave test signal on each channel.
Sample Rate specify the sampling frequency in Hz.
Channel SETTINGS
Sensitivity set channel sensitivity of analog input channels.
Bipolar perform a bipolar derivation between 2 input channels.
Bandpass perform a digital bandpass filtering of the input channels.
Notch perform a bandstop filtering to suppress the power line frequency of 50 Hz or
60 Hz.
CAR select channels for common average reference calculation (CAR checkbox
must be enabled to active the filter).
Noise Reduction select the channels that should be used for calculating the DRL signal for
noise reduction (Noise reduction checkbox must be enabled to active the DRL
function).
1. Highlight the channels in the list box that should be edited. Use the Ctrl key or the Shift key to
highlight multiple channels. To select all channels, use Ctrl+A. To highlight all channels of one
group, click the group header. Highlighted channels are represented with an orange background
color, as illustrated for channel 3.
2. Check or uncheck the checkbox of one of the selected channels in the Acquire column to include
the highlighted channels in data acquisition or exclude them.
3. If you want to change the reference/bipolar channel for a specific channel, select the desired
bipolar channel from the Bipolar list and press the corresponding Apply to highlighted button
right below.
5. Select a notch filter from the Notch list and press the corresponding Apply to highlighted button
right below.
6. If the CAR checkbox is checked and you want to include the highlighted channels in CAR
calculation, select Yes from the CAR list and press the corresponding Apply to highlighted
button right below.
7. If the Noise reduction checkbox is enabled and you want to consider the highlighted channels
for noise reduction, select Yes from the Use for Noise Reduction list and press the
corresponding Apply to highlighted button right below.
If no channel is highlighted when one of the Apply to highlighted buttons is clicked, the system instead
informs the user of this error and the settings do not change.
Click OK to finish the configuration of the g.HIamp and apply it to the device.
g.Nautilus has 8 digital inputs. In the Settings menu, click on g.Nautilus Digital IN… to open the dialog.
The trigger inputs Trigger 1 to Trigger 8 are sampled synchronously with the analog input channels.
Check the box of the desired channels, enter a Description, select a color (Color Sel.) and define the
edge (Edge Detection).
If Combined Triggers are selected, all individual trigger channels are enabled and the individual channel
selection becomes disabled. Combined triggers are always configured to detect rising edges. It is
possible to change the Color and Description for combined triggers. The displayed marker name and
color for a combined trigger are always the same. The marker's value will be the binary coded decimal
made up of the current HIGH and LOW values of all individual trigger channels with trigger channel 1 as
least significant bit and trigger channel 8 as most significant bit.
Select Serial Number 1 from the Available Hardware list and click on the arrow button to add it to the
Selected Hardware list. If more than 16 channels should be simulated, add as many DAQ Simulators as
needed.
To configure the DAQ Simulator, select DAQ Simulator from the Settings… menu.
Sampling Rate [Hz] specify the sampling frequency of the DAQ Simulator in Hz
Time Interval [ms] set the interval between the imports of the samples. (e.g. set the Timer
Interval to 50 ms at a sampling rate of 1 kHz then every 50 ms 50 samples
are drawn)
If multiple simulators are configured use the Apply to all Amplifiers button to perform the settings for all
amplifiers.
Warning: High sampling rates, many channels or extensive feature calculation can overload the PC's
hardware resources, resulting in data loss. This might occur immediately or even after some time after
starting acquisition. Please plan feature calculation meaningfully and carefully.
In the Feature Definition dialog select the Cerebral Function Monitor (CFM)
The channel number and the selected feature are displayed on the right side.
Calculate the HR from the ECG channel 5 and the CSA also from EEG channel 1.
To configure the CFM feature select it and click on the Advanced button. This opens the CFM
Configuration dialog below.
Filter Band
Lowest Frequency [Hz] – select the lower cutoff frequency for the CFM calculation
Highest Frequency [Hz] – select the upper cutoff frequency for the CFM calculation
Classification
Display
Time on Screen [min] – enter the minutes shown on one page in the Feature Viewer
Then enter a Name, the Lower level, the Upper level and define a Color for the specific pattern.
Click the Add button to add the pattern to the list. With Remove, a selected pattern can be removed from
the list. With the Apply changes button, the settings for a selected pattern in the list can be applied.
The figure above shows the standard classification used for neonatal recordings.
Select the HR feature and click on the Advanced button to open the following dialog:
General
Calculation Interval [min] – specify the interval that is used to calculate the HR
Time on Screen [min] - enter the minutes shown on one page in the Feature Viewer
Show Digital Display – if selected, the actual information for heart rate or heart rate variability is
displayed right next to the feature channel.
Output
Select the CSA feature and click on the Advanced button to open the following dialog:
Calculation
Display
Time on Screen [min] - enter the minutes shown on one page in the Feature Viewer
Click the Colormap… button in the Calculation section to define the colormap used for the CSA. This
opens the following dialog:
Define one marker with the key m and name Movement, select the green color. Do not check the Toggle
checkbox to have a marker that indicates a specific point in time. Use the Add button to define the
marker.
Then define key e as a region marker by checking the Toggle checkbox. Set the description to Event.
Filtering:
• Bandpass filter – contains a list of previous defined filters to preprocess incoming raw data. The
filter is applied before the data is framed by the occurrence of a trigger. The filter is only applied
for EP calculation, not to recorded or displayed raw data in the Data Viewer.
Trigger Channels:
• Enable/disable calculation of evoked potentials for target and/or non-target EPs.
• Select a trigger source from the list of acquired digital inputs or defined keyboard markers. In the
example above, keyboard marker a (Keyboard) is selected as the trigger for calculation of target
EPs and keyboard marker s (Keyboard) is selected as the trigger for calculation of non-target
EPs.
Baseline Correction:
• Enable/Disable Correct baseline of frames before averaging to either remove or allow offsets
for evoked potential calculations. If enabled, the offset of the pre-trigger period is removed from
the whole frame for each channel and each trial individually.
Averaging:
• Select Average all frames to average up each triggered frame of the same class as long as the
session lasts. Select Average a limited number of most recent frames only to only include the
selected number of most recently triggered frames of the corresponding class in the averaged
result.
Significance:
• Calculate significance – Enable/Disable highlighting of statistical significant differences between
the target and non-target evoked potential classes. This option will only be available if two
classes of evoked potentials are calculated.
• The Significance level is the maximum acceptable probability of error for significance calculation
in percent.
• Select the number of Minimum frames required for calculation and display of significance.
Trigger Timing:
• Set the Pre-Trigger period duration to define the acquired time before the trigger occurrence.
• Set the Post-Trigger period offset to define the gap to leave blank after the trigger occurrence.
This can be useful to suppress artifacts that accompany the stimulus (e.g. audio trigger artifacts)
in display only. The Post-trigger period offset gap is not excluded from calculation.
• Set the Post-Trigger Period duration to define the acquired time after Post-Trigger period
offset.
If the configuration file cannot be found at startup, then the last used configuration will be loaded. This
option is useful for providing specific study setups.