0% found this document useful (0 votes)
61 views9 pages

Aeroengine Docs 6

Aeroengine docs 6
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
61 views9 pages

Aeroengine Docs 6

Aeroengine docs 6
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

R TASK 70-38-14-250-003

R HIGH FREQUENCY MHz MANUAL EDDY CURRENT INSPECTION OF FAN BLADE


R LEADING EDGE
R 1. General.
R A. This document describes the general equipment, techniques
R and procedures required to conduct high frequency MHz
R manual eddy current inspection of fan blade leading edge.
R B. The following requirements shall be met by inspection
R personnel:
R (1) Personnel performing this inspection shall be
R certified in accordance with NAS-410 or SNT-TC-1A
R or COSAC, or ATA 105 level 2.
R (2) Personnel performing this inspection should receive
R practical training in the use of this procedure and
R must demonstrate proficiency in the calibration and
R evaluation routines before accept/reject authority is
delegated
R (3) Training per this procedure does not imply personnel
R certification per NAS-410 or SNT-TC-1A or COSAC,
R Procedure EN4179.
R D. The following documents shall form a part of this procedure
R to the extend specified herein. Unless a specific issue
R is specified, the latest revision shall apply.
R (1) Appropriate equipment instruction manuals
R (2) NAS-410 and/or SNT-TC-1A and/or COSAC, EN4179, and/or
R ATA 105.
R (3) Appropriate Manual procedure.
R 2. Equipment.
R A. Instruments.
R (1) Eddy current instruments.
R (a) Nortec Model NDT-19, 19e, 19eII, 24 or -3000
R Eddyscope.

R 70-38-14
R Page 1
R May 31/2005
(b) Rohmann Elostest B1 SDM Eddy current instrument.
B. Tools and Equipments.
(1) Special Tools.
Tool No. Description
R 856A2682G01 Kit containing
- a probe 9970 958 389P01, 3MHz, with a spring
blade
- a probe 9970 958 390P01, 3MHz, with a free
head
- a calibration bloc 9970 958 391G01
- 2 cables, Teflon tape, red and blue pencils

(2) Standard Tools and Equipment.


NOTE: This tooling will be contained in a CFMI
inspection kit. The inspection kit number
shall be identified in the applicable manual
procedure. A complete inspection kit
containing all required tooling can be obtained
from CFMI by contacting the following address:
North and South America Europe, Middle East and
and Asia customers. Africa customers
CFMI Inc. customers CFMI SA customers
CFM International CFM International
Quality Technology Center Snecma-Services
One Neumann Way, MD : QB Manager CFM56 Tooling Programms
Cincinnati, OH 45215 USA Site de Melun-Montereau
Attention : Brenda Tenley BP 1936
Phone : (513) 552 4701 77019 Melun Cedex
Fax : (513) 552 4892 France

70-38-14
Page 2
May 31/2006
R C. Consumable Materials.
R Code No. Description
R CP1040 Methyl Ethyl Ketone
R CP2189 Durite G 422
R CP2191 Durite G 421

R 3. Procedure.
R A. Preparation of the edge.
R (1) Clean the edge of the blade with Methyl Ethyl Ketone
R (CP1040).
R (2) If necessary, lightly remove the small surface defects
R with a thin abrasive cloth (CP2189) or (CP2191).
R B. Calibration.
R (1) Initial calibration.
R Frequency : 3 MHz
R Rotation : to be adjusted to have a vertical
R screen indicator
R Low Pass Filter : 100 Hz
R High Pass Filter : off
R Gain : to be adjusted during calibration
R procedure (ELOTEST –35dB, NORTEC
R -50dB

R (2) Calibration procedure with the 9970 958 389P01 or


R 9970 958 390P01 probe.
R CAUTION: MAKE SURE THAT THE PROBE IS IN GOOD CONDITION.
R LOOK FOR DAMAGED PROBE HEAD OR PROBE WIRE.
R (a) Install the probe on the Eddy current instrument

R 70-38-14
R Page 3
R May 31/2005
R CAUTION: TAKE CARE TO LET FREE THE MOBILE PART OF THE
R PROBE
R CAUTION: MAKE SURE THAT THE TEFLON IS CHANGED AT
R REGULAR TIME.
R (b) Apply a new small piece of Teflon tape on the
R coil of each probe.
R (c) Calibration.
R CAUTION: YOU MUST DO THE CALIBRATION AT THE
R BEGINNING AND THE END OF THE INSPECTION,
R EVERY FIFTH BLADE, AND AT ANY CHANGE OF
R ANY COMPONENT. YOU MUST PERFORM THE
R CALIBRATION OF THE INSTRUMENTATION HAS
R BEEN TURNED OFF.
R 1 Put the coil away from the notch of the Electro
R Discharge Machining (EDM) on the “not blended”
R area.
R 2 Set the EDM to zero.
R NOTE: The point is in the middle of the screen.
R 3 Examine surface along the “not blended” area
R with the probe and adjust the gain and rotation
R to have a peak to peak on full screen with a
R vertical indication when the coil touches the
R EDM notch. Refer to figure 1.
R 4 Keep the same parameters and put the coil away
R from the notch of the EDM.
R 5 Set the EDM to zero on the blended area.
R 6 Examine the bottom of the blended area with the
R EDM notch.
R 7 If the indication is equal or less than a half
R of full screen, adjust these following
R parameters :
R hand position, movement of examination, gain
R and rotation to have a full screen indication
R on the EDM notch of the “not blended” area and
R more than a half full screen in the “blended”
R area. Refer to figure 1.

R 70-38-14
R Page 4
R May 31/2005
R

R Indications (Typical)
R Figure 1

R 70-38-14
R Page 5
R May 31/2005
R C. Inspection Procedure.
R CAUTION: YOU MUST PERFORM TWO INSPECTIONS WITH TWO
R DIFFERENT PROBES BECAUSE OF THE DESIGN OF THE
R BLADE (SHAPE AND DIFFERENT THICKNESSES)
R ESPECIALLY IN THE BLENDED AREAS.
R (1) Do a first inspection of the leading edge of the blade
R with 9970 958 389P01 probe.
R (a) Put the probe in position with its coil in contact
R with the pressure face with the same adjustment
R than for calibration. Refer to figure 2.
R CAUTION: MAKE SURE THAT THE COIL IS ALWAYS IN CONTACT
R WITH THE INSPECTED AREA.
R (b) Move the probe to examine the surface along the
R leading edge.
R NOTE: You can perform this inspection a second
R time if it is not sufficient.
R (c) Put the probe in position with its coil in contact
R with the suction face with the same adjustment
R than for calibration. Refer to figure 2.
R CAUTION: MAKE SURE THAT THE COIL IS ALWAYS IN CONTACT
R WITH THE INSPECTED AREA.
R (d) Move the probe to examine the surface along the
R leading edge.
R NOTE: You can perform this inspection a second
R time if it is not sufficient.
R NOTE: You can move the probe three times or more
R across the blade section to make sure that
R the indication is not a malfunction.
R (e) Set to zero each time the probe touches a
R blended area.
R (2) Do a second inspection of the leading edge of the blade
R with 9970 958 390P01 probe.
R (a) Put the probe in position with its coil in contact
R with the pressure face with the same adjustment
R than for calibration. Refer to figure 2.

R 70-38-14
R Page 6
R May 31/2005
R

R Positioning of the Probes


R Figure 2

R 70-38-14
R Page 7
R May 31/2005
R CAUTION: MAKE SURE THAT THE COIL IS ALWAYS IN CONTACT
R WITH THE INSPECTED AREA.
R (b) Move the probe to examine the surface along the
R leading edge.
R NOTE: You can perform this inspection a second
R time if it is not sufficient.
R (c) Put the probe in position with its coil in contact
R with the suction face with the same adjustment
R than for calibration. Refer to figure 2.
R CAUTION: MAKE SURE THAT THE COIL IS ALWAYS IN CONTACT
R WITH THE INSPECTED AREA.
R (d) Move the probe to examine the surface along the
R leading edge.
R NOTE: You can perform this inspection a second
R time if it is not sufficient.
R NOTE: You can move the probe three times or more
R across the blade section to make sure that
R the indication is not a malfunction.
R (e) Set to zero each time the probe touches a
R blended area.
R D. Inspection Criteria.
R (1) If you measure in a blended or non-blended area an
R indication equal to or more than 25 percent of full
R screen, you must do the evaluation of the default.
R Refer to figure 1.
R (2) If you measure in a blended or non-blended area an
R indication equal to or more than 50 percent of full
R screen, you must discard the part. Refer to figure 1.
R D. Evaluation of the Default.
R (1) Make sure that the calibration procedure was done
R correctly.
R (2) Make sure that two check were done with the two probes,
R on suction and pressure faces around the area you
R examine.
R (3) If the indication equal to or more than 25 percent of
R full screen, discard the part. Refer to figure 1.

R 70-38-14
R Page 8
R May 31/2005
R (4) If you have no indication or an indication less than
R 25 percent of full screen, the part is acceptable.
R Refer to figure 1.

R 70-38-14
R Page 9/10
R May 31/2005

You might also like