Met 557
Met 557
Office of Water (MLK 140)   EPA Document No. 815-B-09-012       September 2009   www.epa.gov/safewater
METHOD 557       DETERMINATION OF HALOACETIC ACIDS, BROMATE, AND
                 DALAPON IN DRINKING WATER BY ION CHROMATOGRAPHY
                 ELECTROSPRAY IONIZATION TANDEM MASS SPECTROMETRY
                 (IC-ESI-MS/MS)
                                      Version 1.0
                                    September 2009
                                         557-1
                                             METHOD 557
     1.2   The chromatographic and MRM mass spectrometry conditions described in this method were
           developed using commercially available IC-ESI-MS/MS systems.
     1.3   The single laboratory LCMRL is the lowest spiking concentration such that the probability of
           spike recovery in the 50% to 150% range is at least 99%. Single laboratory LCMRLs for the
           analytes in this method ranged from 0.042 to 0.58 microgram per liter (μg/L), and are listed in
           Table 5. The procedure used to determine the LCMRL is described elsewhere.1
                                                  557-2
     1.4   Laboratories using this method are not required to determine LCMRLs, but they must
           demonstrate that the Minimum Reporting Level (MRL) for each analyte meets the
           requirements described in Section 9.2.4.
     1.5   Detection Limit (DL) is defined as the statistically calculated minimum concentration that can
           be measured with 99% confidence that the reported value is greater than zero.2 The DL is
           dependent on sample matrix, fortification concentration, and instrument performance.
           Determining the DL for analytes in this method is optional (Sect. 9.2.6). DLs for method
           analytes fortified into reagent water ranged from 0.015 to 0.20 µg/L (Table 6).
     1.6   This method is intended for use by analysts skilled in the operation of IC-ESI-MS/MS instru
           mentation and the interpretation of the associated data.
2. SUMMARY OF METHOD
      Residual chlorine present in drinking water samples is reacted with ammonium chloride to form
      chloramines, effectively preventing chlorine-mediated formation of method analytes during
      storage. In addition, the combined chlorine residual prevents microbial degradation in the sample.
      Prior to analysis, isotopically enriched analytes (monochloroacetic acid-2-13C, monobromoacetic
      acid-1-13C, dichloroacetic acid-2-13C, and trichloroacetic acid-2-13C) are added to the samples as
      internal standards. An aliquot of the sample is injected without cleanup or concentration onto an
      ion exchange column specifically designed to separate method analytes from the following
      common anions (matrix components) in drinking water: chloride, carbonate, sulfate, and nitrate.
      The matrix components in the column eluate are monitored via conductivity detection and then
      diverted to waste; the analytes of interest are directed into the ESI-MS/MS system. Acetonitrile is
      added post-column to enhance desolvation of the method analytes in the ESI interface. Each
                                                  557-3
      method analyte is qualitatively identified via a unique mass transition, and the concentration is
      calculated using the integrated peak area and the internal standard technique.
3. DEFINITIONS
     3.1   ANALYSIS BATCH – A sequence of samples, analyzed within a 30-hour period, including
           no more than 20 field samples. Each Analysis Batch must also include all required QC
           samples, which do not contribute to the maximum field sample total of 20. The required QC
           samples include:
     3.2   CALIBRATION STANDARD – An aqueous solution of the method analytes prepared from
           the Primary Dilution Standard (Sect. 3.21) solution. The calibration standards are used to
           calibrate the instrument response with respect to analyte concentration.
     3.4   DETECTION LIMIT (DL) – The minimum concentration of an analyte that can be
           identified, measured, and reported with 99% confidence that the analyte concentration is
           greater than zero. This is a statistical determination (Sect. 9.2.6), and accurate quantitation is
           not expected at this level.
     3.5   DIVERT WINDOW – The period of time during which the column eluate is directed to
           waste for the purpose of diverting matrix components away from the ESI-MS/MS system.
     3.6   ELUTION WINDOW – The period of time during which the column eluate is directed to the
           ESI-MS/MS system for the purpose of measuring the method analytes.
     3.7   INTERNAL STANDARD – A pure compound added to all standard solutions and samples
           in a known amount. Each internal standard is assigned to a specific analyte or multiple
           analytes, and is used to measure relative response.
     3.9   LABORATORY DUPLICATES (LDs) – Two sample aliquots (LD1 and LD2) taken in the
           laboratory from a single sample bottle, and analyzed separately with identical procedures.
           By cancelling variation contributed from sample collection, preservation, and storage
                                                   557-4
       procedures, Laboratory Duplicates provide an estimate of precision associated specifically
       with the analytical determination.
3.10   LABORATORY FORTIFIED BLANK (LFB) – An aliquot of reagent water, containing the
       method preservative, to which known quantities of the method analytes are added. The LFB
       is used during the IDC to verify method performance for precision and accuracy.
3.14   LABORATORY REAGENT BLANK (LRB) – An aliquot of reagent water that contains the
       preservative and internal standards. The LRB is used to determine if the method analytes or
       other interferences are introduced from the laboratory environment, the reagents or
       glassware, and to test for cross contamination.
3.15   LABORATORY SYNTHETIC SAMPLE MATRIX (LSSM) – For this method, the LSSM is
       a solution of common anions prepared at high concentrations relative to their typical
       occurrence in drinking water. Guidance for preparation of the LSSM is provided in Section
       7.2.
3.17   MATERIAL SAFETY DATA SHEETS (MSDS) – Written information provided by vendors
       concerning a chemical’s toxicity, health hazards, physical properties, fire and reactivity data,
       storage instructions, spill response procedures, and handling precautions.
3.18   MINIMUM REPORTING LEVEL (MRL) – The minimum concentration that can be
       reported by a laboratory as a quantified value for the method analyte in a sample following
       analysis. This concentration must meet the criteria defined in Section 9.2.4 and must be no
                                              557-5
            lower than the concentration of the lowest calibration standard for each method analyte. A
            laboratory may be required to demonstrate a specific MRL by a regulatory body if this
            method is being performed for compliance purposes.
     3.20   PRECURSOR ION – The precursor ion is the gas-phase species corresponding to the method
            analyte produced in the ESI interface. In MS/MS, the precursor ion is mass selected and
            fragmented by collision-activated dissociation to produce distinctive product ions of smaller
            mass/charge (m/z) ratio.
     3.21   PRIMARY DILUTION STANDARD (PDS) – An aqueous solution containing the method
            analytes (or internal standards) prepared from Stock Standard Solutions and diluted as
            needed to prepare calibration standards and sample fortification solutions.
     3.23   PRODUCT-ION – For the purpose of this method, a product ion is one of the fragment ions
            produced in MS/MS by collision-activated dissociation of the precursor ion.
     3.24   QUALITY CONTROL SAMPLE (QCS) – A solution containing the method analytes at a
            known concentration, which is obtained from a source external to the laboratory and different
            from the source of calibration standards. The purpose of the QCS is to verify the accuracy of
            the primary calibration standards.
     3.25   REAGENT WATER – Purified water that does not contain any measurable quantity of the
            method analytes or interfering compounds at or above 1/3 the MRL.
     3.26   STOCK STANDARD SOLUTION (SSS) – A concentrated solution containing one or more
            of the method analytes that is prepared in the laboratory using assayed reference materials or
            purchased from a reputable commercial source, so that the concentration and purity of
            analytes are traceable to certificates of analysis.
4. INTERFERENCES
     4.1    GLASSWARE – During method development, no problems with stability of the method
            analytes, interferences, or cross contamination related to glass containers were observed.
            Sample collection bottles and vials containing samples and standards may be reused after
            thorough rinsing with reagent water. Dry glassware in an oven or air dry. Teflon-faced
            septa, if not punctured, may be cleaned and reused. Vials containing PDS solutions must be
                                                  557-6
      rinsed with methanol before reuse and the septa discarded. It is recommended that
      autosampler vials and septa be discarded after a single use.
4.3   MATRIX INTERFERENCES – Matrix interferences are caused by contaminants that are
      present in the sample. The extent of matrix interferences will vary considerably from source
      to source depending upon the nature of the water. Matrix components may directly interfere
      by producing a signal at or near the retention time of an analyte peak. Matrix components
      may also suppress or enhance the signal of the method analytes. (Suppression and
      enhancement effects occur during the ionization process in the electrospray source when a
      co-eluting contaminant influences the ionization of the analyte of interest.) Common anions
      present in drinking water matrices, which would cause matrix suppression, are diverted from
      the MS. In addition, the internal standards recommended in this method performed well in a
      variety of matrices. However, these measures may not compensate for all potential matrix
      effects. The analysis of Laboratory Fortified Sample Matrix (Sect. 9.3.5) provides evidence
      for the presence (or absence) of matrix effects.
                                            557-7
4.7    ION SUPPRESSION IN THE PRESENCE OF CHLORITE – The chlorite anion (ClO2-) may
       be present in drinking water distribution systems employing chlorine dioxide (ClO2) as a
       disinfectant. During method development studies, severe signal suppression of MCAA and
       excessive band broadening of the MCAA chromatographic peak profile were observed in the
       presence of 1 mg/L ClO2-, the maximum contaminant level (MCL). As depicted in Figure 4,
       the ClO2- anion co-eluted with MCAA. Method 557 is not applicable to drinking water
       treated with ClO2 unless the laboratory demonstrates alternate chromatographic conditions
       (other than those used to develop the method, Section 17, Table 1) that eliminate the
       suppression. That is, chlorite must be resolved chromatographically from MCAA and the
       other method analytes. Alternately, the laboratory must demonstrate that the chlorite anion is
       not present in the sample matrix.
4.9    PEAK TAILING – Peak tailing may be observed as the column ages. Peak tailing will limit
       the analyst’s ability to separate matrix components from the method analytes. Peak tailing
       should be minimal with a properly configured ion chromatography system when using a new
       column. (See Figure 2.) If tailing is observed with continued column use, original
       performance can usually be restored by replacing the guard column.
4.11   MANAGING DIVERT WINDOWS – Analyte retention times may slowly shift toward
       lower values as the column ages or becomes fouled. Because this method employs multiple
       divert windows, the analyst must monitor peak locations on a daily basis to ensure that each
       analyte peak elutes entirely within the MS/MS elution windows. Guidance for verifying
       elution/divert windows is provided in Section 10.2.5. To avoid loss of column capacity,
       follow the manufacturer’s instructions for proper operating temperature and for storage
       conditions when the column is not in use.
4.12   BAND BROADENING AND RETENTION TIME (RT) SHIFTS IN HIGH IONIC
       STRENGTH MATRICES – Method performance has not been evaluated for matrix ion
       concentrations exceeding 320 mg/L chloride, 250 mg/L sulfate, 150 mg/L bicarbonate and 20
                                              557-8
           mg/L nitrate. Near these limits, the analyte peaks will widen, peak height will decrease, and
           retention times will decrease slightly. These effects are compound dependent, but affect all
           analytes to some degree. Such effects were minimal in the drinking water matrices
           evaluated, but were more pronounced in Laboratory Synthetic Sample Matrix. [Compare
           Figure 3 (fortified tap water) and Figure 4 (fortified synthetic matrix).] Note that the
           concentrations of common anions in the LSSM are at the limits listed above. This method
           requires the analyst to verify method performance in LSSM during the IDC, and to verify
           elution/divert windows on a daily basis in LFSSM to ensure that these windows are properly
           set to compensate for the potential effects of high ionic strength matrices.
5. SAFETY
     5.1   The toxicity and carcinogenicity of each reagent used in this method has not been precisely
           defined. Each chemical should be treated as a potential health hazard and exposure to these
           chemicals should be minimized. Each laboratory is responsible for maintaining an awareness
           of OSHA regulations regarding safe handling of chemicals used in this method.3 The OSHA
           laboratory standards can be found online at
           http://www.osha.gov/SLTC/laboratories/standards.html. A reference file of MSDSs should be
           made available to all personnel involved in the chemical analysis.
     5.2   Pure standard materials and stock standard solutions of the method compounds should be
           handled with suitable protection for skin, eyes, etc.4
     6.1   SAMPLE CONTAINERS – Amber glass bottles fitted with polytetrafluoroethylene (PTFE)
           -lined screw caps with sufficient volume to allow preparation of all required sample and QC
           aliquots.
     6.2   VIALS FOR SAMPLE PREPARATION – Amber glass vials with PTFE/silicone septa for
           use preparing field samples and QC samples. Forty-milliliter (mL) volatile organic analysis
           (VOA) vials (I-Chem Cat. No. S146–0040 or equivalent) were used during method
           development.
     6.3   AUTOSAMPLER VIALS – Glass vials with PTFE/silicone septa.
6.4 MICRO SYRINGES – Suggested sizes include 50, 100, and 1000 microliters (µL).
     6.5   VOLUMETRIC PIPETTES – Class A, for preparing calibration standards, and for measuring
           aliquots of field samples and QC samples.
     6.6   AUTOMATIC PIPETTE – Electronic, with polypropylene tips (Eppendorf Research Pro or
           equivalent). An automatic pipette is recommended for fortifying samples with internal
           standards.
                                                 557-9
6.7      ANALYTICAL BALANCE – Capable of weighing to the nearest 0.0001 gram (g).
6.9      DESOLVATION GAS – High-purity gas (nitrogen or zero-air) for use in the ESI interface.
         The specific type of gas, purity and pressure requirements will depend on the instrument
         manufacturer’s specifications.
6.10     COLLISION GAS – High-purity gas (nitrogen or argon) for use in the collision cell of the
         mass spectrometer. The specific type of gas, purity, and pressure requirements will depend
         on the instrument manufacturer’s specifications.
       6.12.1.2 ANION TRAP – A continuously regenerated anion trap column (Dionex CR-ATC or
                equivalent).
       6.12.1.3 SAMPLE LOOP – 100-µL size. A 100-µL sample loop was used to generate the data
                presented in this method. Smaller injection volumes may be used as long as the
                Initial Demonstration of Capability (Sect. 9.2), calibration, and sample analyses are
                performed using the same injection volume. The laboratory must be able to meet the
                MRL verification criteria (Section 9.2.4) using the selected injection volume.
                                              557-10
 6.12.1.4 GUARD COLUMN – IonPac® AG24, 2 x 50 millimeters (mm) (Dionex Part
          No. 064151 or equivalent). The guard column is generally packed with the same
          resin as the analytical column.
 6.12.1.5 ANALYTICAL COLUMN – IonPac® AS24, 2 x 250 mm (Dionex Part No. 064153
          or equivalent). Any column that provides on-line separation of common anions
          (chloride, carbonate, sulfate, and nitrate) from the method analytes and symmetrical
          peak shapes may be used. The column must have sufficient capacity to minimize
          retention time shifts in high ionic strength matrices.
6.12.2 AUXILIARY PUMP – Pump capable of precisely delivering flow rates between 0.2 and
       0.3 mL/minute. This pump is used to mix acetonitrile into the suppressed eluent post-
       column. (Dionex high performance metering pump, Model No. AXP-MS or equivalent).
       See Figure 1 for placement of the pump in the sample path.
6.12.3 STATIC MIXING TEE – High pressure, microbore mixing tee. (Upchurch Scientific,
       Oak Harbor, WA, Part No. U-466 or equivalent). The proper placement of the mixing tee
       in the sample path is illustrated in Figure 1.
6.12.5 MS/MS DATA SYSTEM – An interfaced data system is required to acquire, store, and
       output MS data. The computer software must have the capability of processing stored
       data by recognizing a chromatographic peak within a given retention time window. The
       software must allow integration of the ion abundance of any specific ion between
                                        557-11
                   specified time or scan number limits. The software must be able to construct a linear
                   regression or quadratic calibration curve and calculate analyte concentrations using the
                   internal standard technique.
     7.1      REAGENTS AND SOLVENTS – Reagent grade or better chemicals must be used. Unless
              otherwise indicated, it is intended that all reagents will conform to the specifications of the
              Committee on Analytical Reagents of the American Chemical Society (ACS), where such
              specifications are available. Other grades may be used if all the requirements of the IDC are
              met when using these reagents.
           7.1.1   ACETONITRILE (CAS No. 75-05-8) – Post-column organic modifier. High purity,
                   demonstrated to be free of analytes and interferences (Honeywell Burdick & Jackson
                   Brand®, Catalog No. 015 or equivalent).
           7.1.4   SODIUM BICARBONATE (CAS No. 144-55-8) – Laboratory Synthetic Sample Matrix
                   component.
           7.1.5   SODIUM CHLORIDE (CAS No. 7647-14-5) – Laboratory Synthetic Sample Matrix
                   component.
           7.1.6   SODIUM NITRATE (CAS No. 7631-99-4) – Laboratory Synthetic Sample Matrix
                   component.
           7.1.7   SODIUM SULFATE (CAS No. 7757-82-6) – Laboratory Synthetic Sample Matrix
                   component.
     7.2      LABORATORY SYNTHETIC SAMPLE MATRIX (LSSM) – Prepare the LSSM at the
              concentrations listed in the table below. The required concentrations of nitrate (20 mg/L),
              bicarbonate (150 mg/L), chloride (250 mg/L), and sulfate (250 mg/L) are based on the mass
              of the anion, not the sodium salt. The NH4Cl preservative is included in the matrix. LFSSM
              QC samples (Sect. 3.13) can be prepared by diluting the Analyte PDS (Sect. 7.3.2.2) with the
              synthetic matrix solution.
                                                     557-12
                              Empirical    Salt    Anion   Salt Mass            Conc. Stock   Conc. LSSM
             Compound         Formula     (gfw)a   (gfw)     (mg)      H2O, L    (mg/L)b        (mg/L)c
        Ammonium chloride
                               NH4Cl      53.49              500                   1000            100
        (preservative)
                                    -
        Nitrate anion          NO3        84.99    62.00     137                    200             20
                                                                        0.5
        Bicarbonate anion      HCO3-      84.01    61.02     1030                  1500            150
        Chloride anion          Cl       58.44    35.45     2060                  2500            250
        Sulfate anion          SO4-      142.04   96.06     1850                  2500            250
       a
           gfw = gram formula weight of the sodium salt.
       b
           Stock concentration = (salt mass)(gfw anion)/(gfw salt)(0.5 L).
       c
           1:10 dilution of stock (e.g., 50 mL to 500 mL), LSSM = Laboratory Synthetic Sample Matrix.
7.3        STANDARD SOLUTIONS – Solution concentrations listed in this section were used to
           develop this method and are included only as examples. Guidance on the storage stability of
           Primary Dilution Standards and calibration standards is provided in the applicable sections
           below.
           NOTE: When preparing aqueous solutions from MtBE stock solutions, do not add more
           than 0.5% of MtBE relative to the total water volume. MtBE has limited water solubility
           (~5%).
      7.3.1    INTERNAL STANDARDS – This method requires four isotopically enriched internal
               standards. The following table lists the required internal standards and current sources.
       NOTE: TCAA[1-13C] may NOT be substituted for TCAA[2-13C]. TCAA[1-13C] has been
       demonstrated to convert to the native TCAA analyte in the ESI interface, theoretically via gas-
       phase exchange with carbon dioxide-12C in the ionization region of the source. The process is
       temperature dependent (desolvation gas temperature) and was observed on all MS/MS
       platforms evaluated during method development.
                                                      557-13
             glass vial with a PTFE/silicone septum. During method development, addition of 160
             µL of the Internal Standard PDS to each 40-mL field sample, QC sample, or
             calibration standard produced a final concentration of 4.0 µg/L. Analysts are
             permitted to use other PDS concentrations and volumes provided all field samples,
             QC samples, and calibration standards contain the same amount of internal standard,
             the concentration of the internal standard added provides adequate signal to maintain
             precision (as defined in the IDC), and the volume added has a negligible effect on the
             final concentration. The aqueous Internal Standard PDS is stable for 60 days when
             stored at 4 °C.
             NOTE: Storage stability of the Analyte PDS was evaluated during method
             development at a single concentration of 1.0 µg/mL. The aqueous Analyte PDS is
             stable for 60 days when stored at 4 °C. Other PDS concentrations may be selected.
             However, it is recommended that the laboratory independently assess the stability of
             the aqueous PDS to determine safe storage time.
                                            557-14
                        MRL. A constant amount of each internal standard is added to each calibration
                        standard. The calibration standards may also be used as CCCs. An example of the
                        dilutions (starting with the Analyte PDS) necessary to prepare the calibration
                        standards is provided in the table below.
      The preservation requirements for Method 557 are identical to those stipulated in EPA Method
      552.3, Determination of Haloacetic Acids and Dalapon in Drinking Water by Liquid-Liquid
      Microextraction, Derivatization, and Gas Chromatography with Electron Capture Detection.5
           8.1.1    SAMPLE CONTAINERS – Amber glass bottles with PTFE-lined screw caps and
                    sufficient capacity to allow subsequent preparation of all required sample and QC
                    aliquots.
     8.2     SAMPLE COLLECTION – Grab samples must be collected in accordance with conventional
             sampling practices.6 Fill sample bottles taking care not to flush out the ammonium chloride.
             Because the method analytes are not volatile, it is not necessary to ensure that the sample
             bottles are completely headspace-free.
                                                            557-15
           8.2.1   SAMPLING FROM A TAP – When sampling from a cold water tap, remove the aerator,
                   open the tap, and allow the system to flush until the water temperature has stabilized
                   (approximately three to five minutes). Collect a representative sample from the flowing
                   system using a beaker of appropriate size. Use this bulk sample to generate individual
                   samples as needed. Invert the vials several times to mix the sample with the preservation
                   reagent.
     8.3     SAMPLE SHIPMENT AND STORAGE – Samples must be chilled during shipment and
             must not exceed 10 °C during the first 48 hours after collection. Samples must be confirmed
             to be at or below 10 °C when they are received at the laboratory. In the laboratory, samples
             must be stored at or below 6 °C and protected from light until analysis. Samples must not be
             frozen.
     8.4     SAMPLE HOLDING TIMES – Samples must be analyzed within 14 days of collection.
             Chlorinated field samples that are preserved in accordance with the method guidance should
             not exhibit biological degradation of analytes during the allotted 14-day storage time. The
             residency time in the autosampler must be included when calculating the holding time
             from collection until analysis.
9. QUALITY CONTROL
     9.1     QC requirements include the IDC and ongoing QC requirements. This section describes each
             QC parameter, its required frequency, and the performance criteria that must be met in order
             to satisfy EPA quality objectives. The QC criteria discussed in the following sections are
             summarized in Section 17, Tables 11 and 12. These QC requirements are considered the
             minimum acceptable QC program. Laboratories are encouraged to institute additional QC
             practices to meet their specific needs.
                   NOTE: Depending on the source and purity, labeled haloacetic acid internal standards
                   may contain a small percentage of the corresponding native analyte. Therefore, the
                   analyst must demonstrate that the internal standards do not contain the unlabeled analytes
                   at a concentration >1/3 of the MRL when added at the appropriate concentration to
                   samples. An internal standard concentration of 4.0 µg/L was used during method
                   development. Lower concentrations may be used providing the internal standard QC
                   criteria (Sect. 9.3.4) are met.
                                                     557-16
        NOTE: The method must be checked for cross contamination (commonly referred to in
        the environmental laboratory community as “carryover”) by analyzing an LRB
        immediately following the highest calibration standard. If this LRB does not meet the
        criteria outlined in Section 9.3.1, then carryover is present and the cause must be
        identified and eliminated.
9.2.2   DEMONSTRATION OF PRECISION – Prepare and analyze seven replicate LFBs and
        seven replicate LFSSMs. Fortify these samples near the midrange of the initial
        calibration curve. The NH4Cl preservative must be added to the LFBs as described in
        Section 8.1.2. The percent relative standard deviation (RSD) of the concentrations of the
        replicate analyses must be ≤20% for all method analytes.
9.2.3   DEMONSTRATION OF ACCURACY – Using the same sets of replicate data generated
        for Section 9.2.2, calculate the average percent recovery. The average percent recovery
        of the replicate analyses must be within +30% of the true value.
 9.2.4.1   Fortify and analyze seven replicate LFBs at or below the proposed MRL
           concentration. The LFBs must contain the method preservative as specified in
           Section 8.1.2. Calculate the mean (Mean) and standard deviation for these replicates.
           Determine the Half Range for the Prediction Interval of Results (HRPIR) using the
           equation
HRPIR = 3.963S
where S is the standard deviation and 3.963 is a constant value for seven replicates.1
 9.2.4.2   Confirm that the Upper and Lower limits for the Prediction Interval of Results (PIR =
           Mean + HRPIR) meet the upper and lower recovery limits as shown below.
                                          557-17
                                   Mean + HR PIR
                                                       × 100 ≤ 150%
                               Fortified Concentration
                                   Mean − HR PIR
                                                       × 100 ≥ 50%
                               Fortified Concentration
 9.2.4.3   The MRL is validated if both the Upper and Lower PIR Limits meet the criteria
           described above. If these criteria are not met, the MRL has been set too low and must
           be confirmed again at a higher concentration.
NOTE: These equations are only valid for seven replicate samples.
9.2.5   QUALITY CONTROL SAMPLE (QCS) – Analyze a mid-level Quality Control Sample
        (Sect. 9.3.7) to confirm the accuracy of the primary calibration standards.
        Analyses for this procedure must be done over at least three days. Prepare at least seven
        replicate LFBs at a concentration estimated to be near the DL. This concentration may be
        estimated by selecting a concentration at two to five times the noise level. The NH4Cl
        preservative must be added to the samples as described in Section 8.1.2. Process the
        seven replicates through all steps of Section 11.
DL = S x t(n-1,1-α = 0.99)
               where
               t(n-1,1-α = 0.99) = Student's t value for the 99% confidence level with n-1 degrees of
                                   freedom (for seven replicate determinations, the Student’s t value
                                   is 3.143 at a 99% confidence level),
               n = number of replicates, and
               S = standard deviation of replicate analyses.
                                          557-18
9.3      ONGOING QC REQUIREMENTS – This section describes the ongoing QC elements that
         must be included when processing and analyzing field samples. Table 12 summarizes these
         requirements.
      9.3.1   LABORATORY REAGENT BLANK (LRB) – Analyze an LRB during the IDC and
              with each Analysis Batch. The LRB must contain the NH4Cl preservative and the
              internal standards at the same concentration used to fortify all field samples and
              calibration standards. Background from method analytes or contaminants that interfere
              with the measurement of method analytes must be <1/3 the MRL. If method analytes are
              detected in the LRB at concentrations equal to or greater than this level, then all data for
              the problem analyte(s) must be considered invalid for all samples that yielded a positive
              result. Subtracting blank values from sample results is not permitted.
              NOTE: Although quantitative data below the MRL may not be accurate enough for data
              reporting, such data are useful in determining the magnitude of background interference.
              Therefore, blank contamination levels may be estimated by extrapolation when the
              concentration is below the MRL.
      9.3.3   LABORATORY FORTIFIED BLANK (LFB) – Because this method utilizes procedural
              calibration standards, which are fortified reagent waters, there is no difference between
              the LFB and the Continuing Calibration Check standard. Consequently, the analysis of a
              separate LFB is not required as part of the ongoing QC; however, the term “LFB” is used
              for clarity in the IDC.
      9.3.4   INTERNAL STANDARDS (IS) – The analyst must monitor the peak areas of the
              internal standards in all injections of the Analysis Batch. The internal standard responses
              (as indicated by peak areas) for any chromatographic run must not deviate by more than
              ±50% from the average areas measured during the initial calibration for the internal
              standards. If an IS area for a sample does not meet this criterion, check the
              corresponding IS area of the most recent CCC and proceed as follows.
       9.3.4.1   If the IS criterion is met in the CCC but not the sample, reanalyze the sample in a
                 subsequent Analysis Batch. If the IS area fails to meet the acceptance criterion in the
                 repeat analysis, but passes in the most recent CCC, report the sample results as
                 “suspect/matrix.”
       9.3.4.2   If both the original field sample and the CCC fail the IS area criterion, take corrective
                 action (e.g., Sect. 10.4.3). After servicing the instrument, re-inject the sample in a
                 subsequent Analysis Batch. If the IS area fails to meet the acceptance criterion in the
                                                557-19
           repeat analysis, but passes in the most recent CCC, report the sample results as
           “suspect/matrix.”
 9.3.5.1   Prepare the LFSM by fortifying a sample with an appropriate amount of the Analyte
           PDS (Sect. 7.3.2.2). Generally, select a spiking concentration that is greater than or
           equal to the native concentration for most analytes. If the native concentrations of
           method analytes do not allow this criterion to be met without exceeding the
           calibration range, dilution with reagent water containing NH4Cl (100 mg/L) is
           permitted. Selecting a duplicate aliquot of a sample that has already been analyzed
           aids in the selection of an appropriate spiking level. If this is not possible, use
           historical data when selecting a fortifying concentration.
                                   %R =
                                          (A - B) ×100
                                            C
           where
           A = measured concentration in the fortified sample,
           B = measured concentration in the unfortified sample, and
           C = fortification concentration.
 9.3.5.3   Recoveries for samples fortified at concentrations near or at the MRL (within a factor
           of two times the MRL concentration) must be within +50% of the true value.
           Recoveries for samples fortified at all other concentrations must be within +30% of
           the true value. If the accuracy for any analyte falls outside the designated range, and
           the laboratory performance for that analyte is shown to be in control in the CCCs, the
           recovery is judged matrix biased. Report the result for the corresponding analyte in
           the unfortified sample as “suspect/matrix.”
           NOTE: In order to obtain meaningful percent recovery results, correct the measured
           values in the LFSM and LFSMD for the native levels in the unfortified samples, even
           if the native values are less than the MRL. This situation and the LRB are the only
           permitted uses of analyte results below the MRL.
                                          557-20
              NOTE: The variation due to the addition of internal standards must be included in the
              precision estimate. Therefore, first split the original sample and then fortify each aliquot
              with internal standards.
       9.3.6.1   Calculate the relative percent difference (RPD) for duplicate measurements (LD1 and
                 LD2) using the equation:
                                                LD1 − LD 2
                                      RPD =                    ×100
                                              (LD1 + LD2 ) / 2
       9.3.6.2   RPDs for Laboratory Duplicates must be ≤30%. Greater variability may be observed
                 when Laboratory Duplicates have analyte concentrations that are near or at the MRL
                 (within a factor of two times the MRL concentration). At these concentrations,
                 Laboratory Duplicates must have RPDs that are ≤50%. If the RPD of an analyte falls
                 outside the designated range, and the laboratory performance for the analyte is shown
                 to be in control in the CCC, the precision is judged matrix influenced. Report the
                 result for the corresponding analyte in the unfortified sample as “suspect/matrix.”
       9.3.6.3   If an LFSMD is analyzed instead of a Laboratory Duplicate, calculate the RPD for the
                 LFSM and LFSMD using the equation:
                                           LFSM − LFSMD
                                 RPD =                         ×100
                                         (LFSM + LFSMD)/2
       9.3.6.4   RPDs for duplicate LFSMs must be ≤30%. Greater variability may be observed when
                 the matrix is fortified at analyte concentrations near or at the MRL (within a factor of
                 two times the MRL concentration). LFSMs at these concentrations must have RPDs
                 that are ≤50%. If the RPD of an analyte falls outside the designated range, and the
                 laboratory performance for the analyte is shown to be in control in the CCC, the
                 precision is judged matrix influenced. Report the result for the corresponding analyte
                 in the unfortified sample as “suspect/matrix.”
      9.3.7   QUALITY CONTROL SAMPLE (QCS) – A QCS must be analyzed during the IDC, and
              then at least quarterly thereafter. Fortify the QCS near the midpoint of the calibration
              range. The acceptance criteria for the QCS are the same as the mid- and high-level CCCs
              (Sect. 10.4). If the accuracy for any analyte fails the recovery criterion, prepare fresh
              standard dilutions and repeat the QCS evaluation.
                                                 557-21
        9.4.1   Each time method modifications are made, optimize the elution gradient to accomplish
                separation of the method analytes from matrix components, and then verify elution and
                divert windows following the guidance in Section 10.2.5. Establish an acceptable initial
                calibration (Sect. 10.3). Finally, repeat the procedures of the IDC (Sect. 9.2).
        9.4.2   The analyst is also required to evaluate and document method performance for the
                proposed modifications in real matrices that span the range of waters that the laboratory
                analyzes. This additional step is required because modifications that perform acceptably
                in the IDC, which is conducted in reagent water and synthetic matrix, could fail ongoing
                method QC requirements in real matrices. This is particularly important for methods
                subject to matrix effects, such as IC/MS-based methods. For example, a laboratory may
                routinely analyze drinking water from municipal treatment plants that process ground
                water, surface water, or a blend of surface and ground water. In this case, the method
                modification requirement could be accomplished by assessing precision and accuracy
                (Sects. 9.2.2 and 9.2.3) in a surface water with moderate to high total organic carbon
                (e.g., 2 mg/L or greater) and a hard ground water (e.g., 250 mg/L as calcium carbonate
                (CaCO3) equivalent, or greater).
        9.4.3   The results of Sections 9.4.1 and 9.4.2 must be appropriately documented by the analyst
                and independently assessed by the laboratory’s QA officer prior to analyzing field
                samples. When implementing method modifications, it is the responsibility of the
                laboratory to closely review the results of ongoing QC, and in particular, the results
                associated with the LFSM (Sect. 9.3.5), LFSMD (Sect. 9.3.6), CCCs (Sect. 9.3.2), and
                the internal standard area counts (Sect. 9.3.4). If repeated failures are noted, the
                modification must be abandoned.
        10.2.1 MASS CALIBRATION – Method 557 requires the monitoring of low-mass, negatively
               charged ions within nominal m/z range of 35 to 251. Calibrate the mass spectrometer
               with the calibration compounds and procedures specified by the manufacturer. Verify the
               mass assignment accuracy for each precursor ion and each product ion by comparing the
               reported centroid mass to the theoretical mass. (Pay particular attention to the chlorine
               product ion at nominal m/z 35, actual m/z 34.97.) Low signal response could result if
               mass assignments are not centered on the detected mass peaks. If the reported masses
               differ from the calculated masses and low response is observed, the standard calibration
               procedure may be inadequate for this method. Consult the MS/MS manufacturer for an
                                                  557-22
       appropriate low-mass calibration procedure. During method development, accurate mass
       assignments were achieved after calibration via direct infusion of sodium formate
       (10 nanograms per microliter) in 90:10 2-propanol:water, rather than the sodium iodide
       and cesium iodide mixture typically employed for Waters MS/MS systems.
 10.2.2.1 Optimize the ESI-MS/MS at the analytical flow rate (column eluent plus post-column
          acetonitrile addition) via split infusion or flow injection analysis (FIA). Use solutions
          having concentrations between 1 and 5 µg/mL of the method analytes in reagent
          water for split infusion, and solutions having concentrations between 10 and 100 µg/L
          (0.01 and 0.1 μg/mL) in reagent water for FIA. Five- or 10-μL sampling loops are
          recommended for optimizing via FIA. Because the precursor and product ions may
          be identical for some of the haloacetic acids, individual solutions of the method
          analytes are recommended for this step.
 10.2.2.2 Using Table 3 as a guide, select the most abundant precursor ion. Optimize the
          response of the precursor ion for each analyte by infusing the prepared solutions
          (Sect. 10.2.2.1) directly into mobile phase (or via FIA). Vary the ESI parameters
          (probe orientation, capillary extension, source voltages, source and desolvation
          temperatures, gas flows, etc.) and the MS parameters until optimal analyte responses
          are determined. The method analytes may have different optima, thus requiring some
          compromise. ESI-MS/MS systems are subject to drift, which can affect run-to-run
          precision. Accordingly, the optimal conditions may be those that yield adequate
          response while minimizing the potential for drift.
                                         557-23
 10.2.2.3 Using Table 3 as a guide, select the most abundant product ion. Optimize the
          response of the product ion for each analyte by infusing the prepared solutions (Sect.
          10.2.2.1) directly into the mass spectrometer (or via FIA). Vary the MS/MS
          parameters (collision gas pressure, collision energy, etc.) until optimal analyte
          responses are determined.
   10.2.5.1 Stabilize the chromatographic system and analyze an LFSSM at 10 µg/L. Locate
            the first and last analyte peak in each elution window. Display the corresponding
            conductivity trace and note the peak start and peak end times for the chloride,
            carbonate, sulfate, and nitrate matrix anions. Consider a conductivity signal of
            ~5 µS as the peak end (return to baseline) after elution of a matrix anion. [Refer to
            figures 4 and 5 for an example based on use of the Dionex AS24 column. For this
            column, the chloride anion elutes between elution window 1 and elution window 2.
            Carbonate, sulfate and nitrate anions elute between elution windows 2 and 3.]
   10.2.5.2 For the first analyte in each elution window, calculate a maximum time for the
            preceding valve switch: RT – (1.5)(peak width at base of the first analyte in the
            window). For the last analyte in each elution window, calculate a minimum time for
            the following valve switch: RT + (1.5)(peak width at base of the last analyte in the
            window). Verify that the valve switch times set in step 10.2.3 do not overlap the
            calculated times.
                                         557-24
                NOTE: The actual valve switch times are set in accordance with the column
                manufacturer’s guidelines (Section 10.2.3). These could be wider than the start and
                end times for the elution windows calculated using the equations in this section.
       10.2.5.3 By inspection, verify that matrix peaks have returned to baseline before the
                calculated valve switch at the beginning of each elution window. By inspection,
                verify that the valve switch at the beginning of each divert window occurs before a
                matrix component begins to elute. Finally, verify that each analyte elutes entirely
                within the established elution windows. If these conditions are not met, consult the
                column manufacturer’s guidelines for adjusting the elution gradient and reset the
                timing of the divert valve accordingly.
                NOTE: Enough time should be allowed for the baseline to stabilize between the
                valve switch that begins each elution window and the appearance of the subsequent
                analyte signal. If the valve switch is too close to the analyte, the starting point of
                the analyte peak may be difficult to distinguish from the baseline disruption,
                especially for low analyte concentrations.
   10.3.2 CALIBRATION – Calibrate the IC-ESI-MS/MS system using peak areas and the internal
          standard technique. Fit the calibration points with either a linear regression or quadratic
          regression (response vs. concentration). Weighting may be used. Forcing the calibration
          curve through the origin is not recommended. The MS/MS instrument used during
          method development was calibrated using inverse concentration-weighted quadratic
          curves.
          NOTE: Internal standard assignments appropriate for the Dionex AS24 column for each
          method analyte are presented in Table 3, and the mass transitions for the internal
          standards are provided in Table 4. MCAA must always be referenced to MCAA[2-13C].
          MBAA must always be referenced to MBAA[1-13C]. Method 557 was validated with
          bromate referenced to MBAA[1-13C]; however, MCAA[2-13C] could be used if all
          requirements in the IDC are met, as well as all ongoing QC requirements.
                                              557-25
                calculate to be within +30% of their true value. If these criteria cannot be met, the
                analyst will have difficulty meeting ongoing QC criteria. In this case, corrective action is
                recommended such as reanalyzing the calibration standards, restricting the range of
                calibration, or performing instrument maintenance.
      10.4   CONTINUING CALIBRATION CHECKS (CCCs) – Analyze a CCC to verify the initial
             calibration at the beginning of each Analysis Batch, after every tenth field sample, and at the
             end of each Analysis Batch. The beginning CCC for each Analysis Batch must be at or
             below the MRL. This CCC verifies instrument sensitivity prior to the analysis of samples.
             Alternate subsequent CCCs between the remaining calibration levels.
         10.4.1 Verify that the absolute areas of the quantitation ions of each of the internal standards
                have not changed by more than ±50% from the average areas measured during the initial
                calibration. If this limit is exceeded, verify that the background conductivity is in control
                following the guidance in Section 11.3.2. If the background conductivity is normal,
                remedial action may necessary (Sect. 10.4.3).
         10.4.2 Calculate the concentration of each analyte in the CCC. The CCC fortified at ≤MRL
                must calculate to be within +50% of its true value. CCCs fortified at all other levels must
                calculate to be within +30%. If these limits are exceeded, then all data for the failed
                analytes must be considered invalid. Any field samples analyzed since the last acceptable
                CCC that are still within holding time must be reanalyzed after an acceptable calibration
                has been restored.
         10.4.3 REMEDIAL ACTION – Failure to meet CCC QC performance criteria requires remedial
                action. Acceptable method performance may be restored simply by flushing the column
                at the highest eluent concentration in the gradient. Following this and other minor
                remedial action, check the calibration with a mid-level CCC and a CCC at the MRL, or
                alternatively recalibrate according to Section 10.3. If internal standard and calibration
                failures persist, maintenance may be required, such as servicing the ESI-MS/MS system
                and replacing IC columns. These later measures constitute major maintenance, and the
                analyst must return to the initial calibration step (Sect. 10.3) and verify sensitivity by
                analyzing a CCC at or below the MRL.
11. PROCEDURE
      11.1   This section describes the procedures for sample preparation and analysis. Important aspects
             of this analytical procedure include proper sample collection and storage (Sect. 8), ensuring
             that the instrument is properly calibrated (Sect. 10), and that all required QC elements are
             included (Sect. 9).
         11.2.1 All field and QC samples must contain the preservative listed in Section 8.1.2, including
                the LRB. In the laboratory, maintain field samples, QC samples, and calibration
                                                   557-26
          standards at or below 6 °C at all times, including the time these are resident in the
          autosampler awaiting injection.
   11.2.2 Do not filter the samples. Add an appropriate volume of the Internal Standard PDS (Sect.
          7.3.1.2) to a known volume of sample, cap, and mix well. The concentration of the
          internal standards must be the same in the samples as in the calibration standards.
          Transfer an aliquot of each field or QC sample to an autosampler vial. For example,
          weigh 40 mL of sample into a 40-mL VOA vial using a top loading balance and a
          disposable, glass pipette. Add 160 uL of the internal standard PDS (1.0 µg/mL) to
          achieve a concentration of 4.0 µg/L in the sample. Mix well and transfer 1.5 mL to an
          autosampler vial by use a disposable pipette.
11.3.1 Establish IC-ESI-MS/MS operating conditions per the guidance in Section 10.2.
   11.3.2 Flush the column at the highest eluent concentration in the gradient for at least 15
          minutes prior to beginning each analysis sequence. This step is especially important for
          minimizing background conductivity if the column has been installed in the system and
          held at the starting eluent concentration for extended periods.
          NOTE: During method development, the background conductivity ranged from 0.3 to
          2.2 microsiemens (µS). If an internal standard or analyte area response is low due to high
          background conductivity, flush the column at the maximum eluent concentration in the
          gradient and observe the conductivity signal until the background stabilizes at less than
          2.5 µS. Also, ensure that the suppressor is functioning properly or replace the suppressor
          to troubleshoot the problem. Occasional overnight flushing of the column will minimize
          the occurrence of high background conductivity.
   11.3.3 VERIFY ELUTION AND DIVERT WINDOWS - Analyze Laboratory Synthetic Sample
          Matrix fortified at 10 µg/L. Verify the elution and divert windows as specified in Section
          10.2.5. This verification must be done prior to beginning each Analysis Batch.
   11.3.4 THE ANALYSIS BATCH – Establish a valid initial calibration following the procedures
          outlined in Section 10.3 and confirm that the calibration is valid by analyzing a CCC at or
          below the MRL as described in Section 10.4. Alternately, verify that an existing
          calibration, established for a previous Analysis Batch, is still valid by analyzing a CCC at
          or below the MRL. Next, analyze an LRB. Continue the Analysis Batch by analyzing
          aliquots of field and QC samples at appropriate frequencies (Section 9.3), employing the
          optimized conditions used to acquire the initial calibration. Analyze a mid- or high-level
          CCC after every ten field samples and at the end each Analysis Batch.
          NOTE: Each Analysis Batch must begin with the analysis of a CCC at or below the
          MRL for each analyte that the laboratory intends to report, followed by the analysis of an
          LRB. This is true whether or not an initial calibration is analyzed. After 20 field samples
          the low-level CCC and the LRB must be repeated to begin a new Analysis Batch. The
                                             557-27
            acquisition start time of the mid-level CCC at the end of the Analysis Batch must be
            within 30 hours of the acquisition start time of the low-level CCC at the beginning of the
            Analysis Batch. Do not count QC samples (LRBs, LDs, LFSMs, LFSMDs) when
            calculating the frequency of CCCs that are required during an Analysis Batch.
  12.1   Establish an appropriate retention time window for each analyte to identify them in QC and
         field sample chromatograms. Base this assignment on measurements of actual retention time
         variation for each compound in standard solutions over the course of time. The suggested
         variation is plus or minus three times the standard deviation of the retention time for each
         compound for a series of injections. The injections from the initial calibration and from the
         IDC (Sect. 9.2) may be used to calculate the retention time window. However, the
         experience of the analyst should weigh heavily on the determination of an appropriate range.
  12.2   At the conclusion of data acquisition, use the same software settings established during the
         calibration procedure to identify peaks of interest in the predetermined retention time
         windows. Confirm the identify of each analyte by comparison of its retention time with that
         of the corresponding analyte peak in an initial calibration standard or CCC.
  12.3   Calculate analyte concentrations using the multipoint calibration established in Section 10.3.
         Report only those values that fall between the MRL and the highest calibration standard.
         Samples with analyte responses that exceed the highest calibration standard require dilution
         and reanalysis (Sect. 12.7).
         NOTE: In validating this method, concentrations were calculated using the product ions
         listed in Table 3 of Section 17. Other ions may be selected at the discretion of the analyst as
         discussed in Section 10.2.2.
  12.4   Calculations must use all available digits of precision, but final reported concentrations
         should be rounded to an appropriate number of significant figures (one digit of uncertainty),
         typically two, and not more than three significant figures.
  12.5   Prior to reporting the data, the chromatograms must be reviewed for any incorrect peak
         identifications or improper integration.
  12.6   Prior to reporting data, the laboratory is responsible for ensuring that QC requirements have
         been met and that any appropriate qualifier is assigned.
  12.7   EXCEEDING THE CALIBRATION RANGE – The analyst must not extrapolate beyond the
         established calibration range. If an analyte result exceeds the range of the initial calibration
         curve, the sample may be diluted using reagent water containing 100 mg/L NH4Cl with the
         appropriate amount of internal standard added to match the original level. Re-inject the
         diluted sample. Incorporate the dilution factor into final concentration calculations. The
         resulting data must be annotated as a dilution, and the reported MRLs must reflect the
         dilution factor.
                                               557-28
13.    METHOD PERFORMANCE
      13.1   PRECISION, ACCURACY, AND DETECTION LIMITS – The method performance data
             presented in Section 17 were collected using the IC conditions listed in Table 1 and the
             Waters Quattro Premier XE ESI-MS/MS system. ESI-MS/MS conditions for the Waters
             system are presented in Table 2. Tables 3 and 4 list the mass transitions for each analyte and
             internal standard, internal standard assignments, and observed retention times associated with
             the method performance results. LCMRLs and DLs are presented in Tables 5 and 6. Single
             laboratory precision and accuracy data are presented for four water matrices: reagent water
             (Table 7), LSSM (Table 8), chlorinated (finished) ground water (Table 9), and chlorinated
             (finished) surface water (Table 10). Figure 1 depicts the post-column sample path as
             previously cited in this document (Sect. 6.12). Figures 2 through 4 are chromatograms of the
             method analytes in reagent water, drinking water, and LSSM obtained under the conditions
             employed during method development. Figure 4 was acquired with the inclusion of mass
             transitions for the chlorite and chlorate anions to mark their position in the chromatogram
             relative to the method analytes. Figure 5 is a corresponding conductivity trace for the analyte
             chromatogram presented in Figure 4 showing the location of matrix anions, chlorite anion,
             and chlorate anion.
      14.1   For information about pollution prevention that may be applicable to laboratory operations,
             consult “Less is Better: Laboratory Chemical Management for Waste Reduction” available
             from the American Chemical Society’s Department of Government Relations and Science
             Policy, 1155 16th Street N.W., Washington, D.C., 20036, or online at
             http://www.ups.edu/x7432.xml.
      15.1   The analytical procedures described in this method generate relatively small amounts of
             waste since only small amounts of reagents and solvents are used. The matrix of concern is
             finished drinking water. However, the Agency requires that laboratory waste management
             practices be conducted consistent with all applicable rules and regulations, and that
             laboratories protect the air, water, and land by minimizing and controlling all releases from
             fume hoods and bench operations. In addition, compliance is required with any sewage
             discharge permits and regulations, particularly the hazardous waste identification rules and
             land disposal restrictions. For further information on waste management, see the
             publications of the American Chemical Society’s Laboratory Environment, Health & Safety
             Task Force on the Internet at http://membership.acs.org/c/ccs/publications.htm. Additional
                                                   557-29
           waste management information can be found in “Laboratory Waste Minimization and
           Pollution Prevention,” Copyright © 1996 Battelle Seattle Research Center, which can be
           located at http://www.p2pays.org/ref/01/text/00779/ch05.htm.
16. REFERENCES
1. Winslow, S. D.; Pepich, B. V.; Martin, J. J.; Hallberg, G. R.; Munch D. J.; Frebis, C. P.; Hedrick, E.
   J.; Krop, R. A. Statistical Procedures for Determination and Verification of Minimum Reporting
   Levels for Drinking Water Methods. Environ. Sci. Technol. 2006; 40, 281-288.
2. Glaser, J.A.; Foerst, D.L.; McKee, G.D.; Quave, S.A.; Budde, W.L. Trace Analyses for
   Wastewaters. Environ. Sci. Technol. 1981; 15, 1426-1435.
6. Standard Practice for Sampling Water from Closed Conduits; ASTM Annual Book of Standards,
   Section 11, Volume 11.01, D3370-08; American Society for Testing and Materials: Philadelphia,
   PA, 2008.
                                                 557-30
17.      TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
                                                              557-31
TABLE 3.          IC-ESI-MS/MS ANALYTE RETENTION TIMES, PRECURSOR AND PRODUCT
                  IONS
                                        Retention Timea       Internal Standard
Analyte                                      (min)               Assignment            Precursor     Product
Monochloroacetic acid                        12.62              MCAA[2-13C]               93           35
Monobromoacetic acid                         14.05              MBAA[1-13C]               137           79
Bromate                                      14.93              MBAA[1-13C]               127          111
Dalapon                                      23.33              DCAA[2-13C]               141           97
Dichloroacetic acid                          24.26              DCAA[2-13C]               127           83
Bromochloroacetic acid                       26.16              DCAA[2-13C]               173          129
Dibromoacetic acid                           28.89              DCAA[2-13C]               217          173
Trichloroacetic acid                                             TCAA[2-13C]              161          117
                                             41.08
Trichloroacetic acid (alternate)                                 TCAA[2-13C]              163          119
Bromodichloroacetic acid                     42.89               TCAA[2-13C]              163           81
Chlorodibromoacetic acid                     45.50               TCAA[2-13C]              207           79
Tribromoacetic acid                          49.22               TCAA[2-13C]              251           79
a
    Dionex AS24 column (used to collect method performance data).
                                                     557-32
TABLE 5.          IC-ESI-MS/MS LOWEST CONCENTRATION MINIMUM REPORTING LEVEL
                  (LCMRL)
                                                                                                            Calculated LCMRL
             Analyte                                 LCMRL Fortification Levels (µg/L)                            (µg/L)
Monochloroacetic acid                         0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75, 2.0             0.58
                                            0.050, 0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75,
Monobromoacetic acid                                                                                                0.19
                                                                        2.0
                                           0.025, 0.050, 0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5,
Bromate                                                                                                            0.042
                                                                     1.75, 2.0
Dalapon                                          0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75, 2.0                 0.41
                                           0.025, 0.050, 0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5,
Dichloroacetic acid                                                                                                 0.13
                                                                     1.75, 2.0
Bromochloroacetic acid                        0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75, 2.0              0.16
                                           0.025, 0.050, 0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5,
Dibromoacetic acid                                                                                                 0.062
                                                                     1.75, 2.0
Trichloroacetic acid – m/z 163/119           0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75, 2.0               0.25
                                            0.050, 0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75,
Bromodichloroacetic acid                                                                                            0.19
                                                                        2.0
                                            0.050, 0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75,
Chlorodibromoacetic acid                                                                                           0.080
                                                                        2.0
Tribromoacetic acid                           0.10, 0.25, 0.375, 0.50, 0.75, 1.0, 1.25, 1.5, 1.75, 2.0              0.27
                                                           557-33
TABLE 7.         IC-ESI-MS/MS PRECISION AND ACCURACY OF METHOD ANALYTES
                 FORTIFIED AT 1.0 AND 15 µg/L IN REAGENT WATER
                                        Fortified Conc. = 1.0 µg/L           Fortified Conc. = 15 µg/L
                                                   (n=7)                               (n=8)
                                     Mean %         Relative Standard    Mean %        Relative Standard
Analyte                              Recovery           Deviation        Recovery          Deviation
Monochloroacetic acid                  101                 3.5             101                 1.7
Monobromoacetic acid                   97.5                3.7             99.8                1.8
Bromate                                93.3                2.4             104                 7.8
Dalapon                                97.4                4.7             100                 3.3
Dichloroacetic acid                    109                 2.6             97.0                6.5
Bromochloroacetic acid                 103                 2.9             107                 4.4
Dibromoacetic acid                     104                 9.0             111                 6.7
Trichloroacetic acid – m/z 163/119     99.1                2.3             99.8                3.2
Bromodichloroacetic acid               105                 3.7             97.7                2.2
Chlorodibromoacetic acid               90.4                5.9             103                 5.4
Tribromoacetic acid                    101                 5.3             98.9                3.0
                                               557-34
TABLE 9.           IC-ESI-MS/MS PRECISION AND ACCURACY OF METHOD ANALYTES
                   FORTIFIED AT 2.5 AND 10 µg/L IN CHLORINATED GROUND WATERa
                                           Native       Fortified Conc. = 2.5 µg/L           Fortified Conc. = 10 µg/L
                                           Conc.,                  (n=8)                               (n=8)
                                            μg/L      Mean %       Relative Standard      Mean %        Relative Standard
Analyte                                    (n=6)      Recoveryb        Deviation          Recoveryb         Deviation
Monochloroacetic acid                       0.57        95.9               4.7              99.6               5.2
Monobromoacetic acid                        0.41         101               2.1               101               1.8
Bromate                                     0.56         107               7.3               102               4.8
Dalapon                                     0.37        95.5               3.1              98.5               3.0
Dichloroacetic acid                          4.3        99.7               6.2               106               1.7
Bromochloroacetic acid                       3.4        95.2               5.6              93.5               3.1
Dibromoacetic acid                           2.0        95.2               8.1              99.3               6.2
Trichloroacetic acid – m/z 163/119           2.6          101              1.8               102                2.8
                                                             557-35
TABLE 11.    INITIAL DEMONSTRATION OF CAPABILITY (IDC) QUALITY CONTROL
             REQUIREMENTS
Method
            Requirement        Specification and Frequency            Acceptance Criteria
Reference
                                                                      Demonstrate that all method analytes
                                                                      are <1/3 of the Minimum Reporting
            Demonstration of   Analyze a Laboratory Reagent           Level (MRL) and that possible
Section
            low system         Blank (LRB) prior to any other         interferences from reagents and
9.2.1
            background         Initial IDC steps.                     glassware do not prevent the
                                                                      identification and quantitation of
                                                                      method analytes.
                               Analyze an LRB after the high
Section     Test for system                                           Demonstrate that the method analytes
                               calibration standard during the IDC
9.2.1       carryover                                                 are <1/3 of the MRL.
                               calibration.
                               Analyze 7 replicate Laboratory
                               Fortified Blanks (LFBs) and 7
Section     Demonstration of   Laboratory Fortified Synthetic         Percent relative standard deviation
9.2.2       precision          Sample Matrix samples (LFSSMs)         must be ≤20%.
                               fortified near the midrange
                               concentration.
Section     Demonstration of   Calculate average recovery for         Mean recovery within +30% of the true
9.2.3       accuracy           replicates used in Section 9.2.2.      value.
                               Fortify and analyze 7 replicate LFBs
                                                                      Upper PIR ≤ 150%
                               at the proposed MRL concentration.
Section     MRL                Confirm that the Upper Prediction
9.2.4       confirmation       Interval of Results (PIR) and Lower    Lower PIR ≥ 50%
                               PIR (Sect. 9.2.4.2) meet the
                               recovery criteria.
Section     Quality Control                                           Results must be within +30% of the
                               Analyze mid-level QCS.
9.2.5       Sample (QCS)                                              true value.
                                                 557-36
TABLE 12.     ONGOING QUALITY CONTROL REQUIREMENTS
Method
            Requirement              Specification and Frequency            Acceptance Criteria
Reference
                                     Analyze a Laboratory Fortified
Section     Verify divert                                                   See Section 10.2.5 for acceptance
                                     Synthetic Sample Matrix (LFSSM)
11.3.3      windows                                                         criteria.
                                     prior to each Analysis Batch.
                                     Use the internal standard cali        When each calibration standard is
                                     bration technique to generate a        calculated as an unknown using the
                                     linear or quadratic calibration        regression equations, the lowest
Section
            Initial calibration      curve. Use at least five standard      level standard must be within +50%
10.3
                                     concentrations. Validate the cali     of the true value. All other points
                                     bration curve as described in          must be within +30% of the true
                                     Section 10.3.3.                        value.
                                                                            Demonstrate that all method ana
                                                                            lytes are below 1/3 the Minimum
                                                                            Reporting Level (MRL), and that
Section     Laboratory Reagent       Analyze one LRB with each
                                     Analysis Batch.
                                                                            possible interference from reagents
9.3.1       Blank (LRB)
                                                                            and glassware do not prevent identi
                                                                            fication and quantitation of method
                                                                            analytes.
                                                                            The lowest level CCC must be
                                     Verify initial calibration by ana     within +50% of the true value. All
                                     lyzing a low-level CCC at the          other points must be within +30% of
                                     beginning of each Analysis Batch.      the true value.
Section     Continuing Calibra
                                     Subsequent CCCs are required
10.4        tion Check (CCC)
                                     after every 10 field samples, and
                                     after the last field sample in a       Results for field samples that are not
                                     batch.                                 bracketed by acceptable CCCs are
                                                                            invalid.
                                     Isotopically labeled internal          Peak area counts for each IS must be
Section
            Internal standard (IS)   standards are added to all standards   within ±50% of the average peak
9.3.4
                                     and samples.                           areas in the initial calibration.
                                     Analyze one LFSM per Analysis          For LFSMs fortified at concen
                                     Batch. Fortify the LFSM with           trations <2 x MRL, the result must
            Laboratory Fortified
Section                              method analytes at a concentration     be within +50% of the true value.
            Sample Matrix
9.3.5                                greater than the native                At concentrations greater than the 2
            (LFSM)
                                     concentrations of most analytes.       x MRL, the result must be within
                                     Calculate LFSM recoveries.             +30% of the true value.
            Laboratory Fortified
            Sample Matrix Dup                                              For LFSMDs or LDs, relative
Section                              Analyze at least one LFSMD or
            licate (LFSMD) or                                               percent differences must be ≤30%.
9.3.6                                LD with each Analysis Batch.
            Laboratory Duplicate                                            (≤50% if concentration <2 x MRL.)
            (LD)
Section     Quality Control          Analyze mid-level QCS at least         Results must be +30% of the true
9.3.7       Sample (QCS)             quarterly.                             value.
                                                   557-37
Figure 1. Divert valve placement in sample path.
                    557-38
Figure 2. Dionex AS24 column: procedural calibration standard (5 µg/L).
                                557-39
Figure 3. Dionex AS24 column: tap water fortified at 8 µg/L.
                          557-40
Figure 4. Dionex AS24 column: synthetic sample matrix fortified at 5 µg/L plus chlorite (1 mg/L) and chlorate (0.5 mg/L).
                                                         557-41
Figure 5. Dionex AS24 column: conductivity trace, synthetic sample matrix plus chlorite (1 mg/L) and chlorate (0.5 mg/L).
557-42