62 Max/62 Max +

Infrared Thermometer

Calibration Information

Introduction
The Fluke 62 MAX and 62 MAX + Infrared Thermometers (the Product) can determine the surface temperature by
measurement of the infrared energy radiated from the target surface.

Note
Japanese models indicate Celsius only.

Warning
To prevent personal injury, read all safety information before you use the
Product.

Contact Fluke
To contact Fluke, call one of the following telephone numbers:
• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-03-6714-3114
• Singapore: +65-6799-5566
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To download manuals, or to see, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.

June 2013 ©2013 Fluke Corporation. All rights reserved. 1

Downloaded from www.Manualslib.com manuals search engine


62 Max/62 Max +
Calibration Information

Safety Information
A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and
procedures that can cause damage to the Product or the equipment under test.
The symbols used on the Product and in this manual are shown in Table 1.

Warning
To prevent eye damage and personal injury:
• Read all safety Information before you use the Product.
• Do not use the Product if it operates incorrectly.
• Use the Product only as specified, or the protection supplied by the
Product can be compromised.
• Before you use the Product, inspect the case. Do not use the Product if it
appears damaged. Look for cracks or missing plastic.
• See emissivity information for actual temperatures. Reflective objects
result in lower than actual temperature measurements. These objects pose
a burn hazard.
• Do not look directly into the laser with optical tools (for example,
binoculars, telescopes, microscopes). Optical tools can focus the laser and
be dangerous to the eye.
• Do not look into the laser. Do not point laser directly at persons or animals
or indirectly off reflective surfaces.
• Replace the batteries when the low battery indicator shows to prevent
incorrect measurements.
• Do not use the Product around explosive gas, vapor, or in damp or wet
environments.
• Use the Product only as specified or hazardous laser radiation exposure
can occur.

Table 1. Symbols

Symbol Meaning Symbol Meaning

This product complies with the WEEE Directive


(2002/96/EC) marking requirements. The affixed label
indicates that you must not discard this
electrical/electronic product in domestic household
Risk of danger.
waste. Product Category: With reference to the
 Important information.  equipment types in the WEEE Directive Annex I, this
See Manual.
product is classed as category 9 "Monitoring and Control
Instrumentation” product. Do not dispose of this product
as unsorted municipal waste. Go to Fluke’s website for
recycling information.

 Warning. Laser.  Conforms to European Union directives.

 Battery  Conforms to relevant Australian standards.

Conforms to China
 Conforms to relevant South Korean EMC standards.
Metrology Certification

Downloaded from www.Manualslib.com manuals search engine


Infrared Thermometer
Maintenance

Maintenance
Caution
To prevent damage to the Product, do not leave the Product on or near objects
of high temperature.

Clean the Product


Use soap and water on a damp sponge or soft cloth to clean the Product case. Carefully wipe the lens surface with a
moist cotton swab. The swab may be moistened with water. See Figure 1.

hiy002.eps
Figure 1. Clean the Product

Downloaded from www.Manualslib.com manuals search engine


62 Max/62 Max +
Calibration Information

Change the Battery


To install or change the AA IEC LR06 battery, use a flat-head screwdriver to remove the battery compartment door.
Replace the battery as shown in Figure 2. Be sure to observe the correct polarity.

hiy001.eps
Figure 2. Change the Battery

Downloaded from www.Manualslib.com manuals search engine


Infrared Thermometer
Specifications

Specifications
62 MAX 62 MAX +

-30 °C to 500 °C -30 °C to 650 °C


Temperature Range
(-22 °F to 932 °F) (-22 °F to 1202 °F)
≥0 °C: ±1.5 °C or ±1.5 % of reading, ≥0 °C: ±1 °C or ±1 % of reading,
whichever is greater whichever is greater (≥32 °F: ±2 °F or
(≥32 °F: ±3 °F or ±1.5 % of reading, ±1 % of reading, whichever is greater)
Accuracy whichever is greater) ≥ -10 °C to <0 °C: ±2 °C
(Calibration geometry with ambient ≥ -10 °C to <0 °C: ±2 °C (≥14 °F to <32 °F: ±4 °F)
temperature 23 °C ±2 °C)
(≥14 °F to <32 °F: ±4 °F) < -10 °C: ±3 °C
< -10 °C: ±3 °C (<14 °F: ±6 °F)
(<14 °F: ±6 °F)
Response Time (95 %) <500 ms (95 % of reading) <300 ms (95 % of reading)
Spectral Response 8 to 14 microns
Emissivity 0.10 to 1.00
Temperature Coefficient ±0.1 °C/°C or ±0.1 %/°C of reading (whichever is greater)
10:1 12:1
Optical Resolution
(calculated at 90 % energy) (calculated at 90 % energy)
Display Resolution 0.1 °C (0.2 °F)
±0.8 % of reading or ±1.0 °C (2 °F), ±0.5 % of reading or ±0.5 °C (1 °F),
Repeatability (% of reading)
whichever is greater whichever is greater
Power 1 AA IEC LR06 Battery
Battery Life 10 hours with laser and backlight on 8 hours with laser and backlight on
Weight 255 g (8.99 oz)
Size (175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches
Operating Temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature -20 °C to 60 °C (-4 °F to 140 °F), (without battery)
Operating Humidity 10 % to 90 % RH non-condensing @ 30 °C (86 °F)
Operating Altitude 2000 meters above mean sea level
Storage Altitude 12,000 meters above mean sea level
Drop Test 3 meters
IP Rating IEC 60529: IP 54
Implement Standard Q/SXAV37
Laser Safety IEC 60825-1, Class 2

Electromagnetic Environment IEC 61326-1: Portable

Applies to use in Korea only:


Class A Equipment (Industrial Broadcasting & Communication Equipment) [1]
Electromagnetic Compatibility [1] This product meets requirements for industrial (Class A) electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and is not to be used in homes.

Downloaded from www.Manualslib.com manuals search engine


62 Max/62 Max +
Calibration Information

Performance Verification Tests


The subsequent performance tests verify the complete operation of the Product and check the accuracy of each
function against Product specifications. If the Product fails any part of the test, calibration adjustment and/or repair
is necessary. Product calibration can only be adjusted at the Fluke factory. To return your Product for calibration
adjustment, call one of the numbers in the “Contact Fluke” section of this document.
Read the entire Performance Tests section before you make any adjustment.

Required Equipment
Table 2 shows the required equipment necessary for the performance tests.

Note
Fluke recommends plate blackbodies as verification equipment because cavity blackbodies
have small field of view, which could increase uncertainty for a thermometer with low D:S. If
cavity blackbodies are used to verify the Product, make sure that the distance between the
cavity blackbody opening and the Product is equal to the diameter of the cavity blackbody
opening × 2. Where the temperature of the cavity tube near the outside is lower than the
temperature of the cavity, readings will be lower. Moving the 62Max closer to the cavity
opening may be necessary for the 62 Max to only measure the cavity temperature.
If other than Fluke blackbodies are used, the target should be checked for actual radiometric
output with a more accurate device, such as a Heitronic KT19-82.

Table 2. Required Equipment

Item Minimum Use Specifications Calibration Equipment

2.1 Blackbody Range: -10 °C to 100 °C Fluke 4180

2.2 Blackbody Range: 100 °C to 500 °C Fluke 4181

2.3 Blackbody Range: 500 °C to 650 °C Plate Blackbody with 10 cm to 15 cm


target. [1]

[1] Calibrate with an IR thermometer that has low uncertainty such as KT19 and TRT.

Before the Performance Verification Tests


Before you do the performance verification tests:
• Performance verification tests are to be done within the temperature of 23 ±3 °C and between 25 % and 60 %
relative humidity.
• Make sure that all required calibration equipment has a valid calibration certificate and/or label.
• Make sure that any item(s) substituted for the equipment in Table 2 meets or exceeds the stated minimum use
specifications.
• The correct radiometric temperature will need to be determined before blackbodies are used to make
measurements.
• Make sure that the required calibration equipment is allowed a warm-up period as per manufacturer
recommendations. See the documentation for the equipment.

Downloaded from www.Manualslib.com manuals search engine


Infrared Thermometer
Performance Verification Tests

To prepare the Unit Under Test (UUT) for verification:


1. Visually inspect the UUT outer case for cracks or damage.
2. Open the battery compartment and inspect the battery and battery compartment for corrosion or damage.
3. Install a new AA battery prior to verification.
4. Close the UUT battery compartment door.
5. Set the UUT temperature units for degrees Celsius (°C) if necessary.
6. Push SEL until °C alone is shown. Change by pushing SET.

Performance Verification Test Procedures


Note
The subsequent verification test requires the blackbody emissivity to be set to 0.95 at all
temperature points.

Note
For temperatures below ambient, to reduce the chance of condensation or ice buildup on the
emitter plate of the 4180, it is recommended to install a clean dry air or nitrogen gas purge
box on the front of the 4180 Blackbody. Although the purge box is recommended, it is optional.
If temperature measurements are made without a purge box, refer to the 4180 User Manual
for how to make proper temperature measurements below ambient. See Figure 3.

hiy003.eps
Figure 3. Aim the 62 Max + for Verification

Downloaded from www.Manualslib.com manuals search engine


62 Max/62 Max +
Calibration Information

1. Set blackbody emissivity and temperature for the first temperature listed in Table 3. Allow adequate time for the
blackbody to stabilize. See the blackbody documentation.
2. Position the UUT at the correct the distance from the blackbody listed in Table 3. Center the UUT laser(s) on the
blackbody emitter plate. For the 62 MAX +, make sure that the midpoint of the two laser spots is centered on the
blackbody plate. The 62 Max uses one laser.
3. Push and hold the UUT trigger for 5 to 6 seconds.
4. Release the trigger.
5. Record the UUT indication.
6. Move to a location that is not centered on the blackbody.
7. Make sure the UUT reading is within the limits listed in Table 3.
8. Repeat steps 1 through 7 for the remaining temperatures listed in the Temperature column of Table 3. The
62 Max is specified to 500 °C. 461 °C is the last temperature point to verify.
9. Set blackbodies to nominal values.

Table 3. Verification Tests

Distance (cm) [1] Temperature (°C) Emissivity 62 Max Limits (°C) 62 Max + Limits (°C)

Fluke 4180 points

28 -10 0.95 -12.0 to -8.0 -12.0 to -8.0

28 0 0.95 -1.5 to 1.5 -1.0 to 1.0

28 65 0.95 63.5 to 66.5 64.0 to 66

Fluke 4181 point

28 250 0.95 246. 2 to 253.8 247.5 to 252.5

28 461 0.95 454.1 to 467.9 456.3 to 465.6

Fluke 62 Max + only above 500 °C

25.5 580 0.95 574.2 to 585.8

[1] Distance is calculated based on the target size. Typically a spot size that is about 1/2 the target size provides the best results.

Note
Fluke recommends plate blackbodies as verification equipment because cavity blackbodies
have small field of view, which could increase uncertainty for a thermometer with low D:S. If
cavity blackbodies are used to verify the Product, make sure that the distance between the
cavity blackbody opening and the Product is equal to the diameter of the cavity blackbody
opening × 2. Where the temperature of the cavity tube near the outside is lower than the
temperature of the cavity, readings will be lower. Moving the 62Max closer to the cavity
opening may be necessary for the 62 Max to only measure the cavity temperature.
Temperature verification is complete. If the Product does not pass the performance verification tests, calibration
adjustment or repair is necessary. Product calibration can only be adjusted by Fluke. To return your Product for
calibration adjustment, call one of the numbers in the “Contact Fluke” section of this document.

Blackbody Layout
Make sure that the distance between two blackbodies is more than 1 meter. Forced air or HVAC vents should not be
near the surface of the radiation source. Do not have two blackbodies work face to face. For more detail about
blackbody use, see the Fluke 4180 Technical Guide.

Downloaded from www.Manualslib.com manuals search engine


Infrared Thermometer
User-Replaceable Parts

User-Replaceable Parts
To order replacement parts, call one of the numbers in the “Contact Fluke” section of this document. User-
replaceable parts are listed in Table 4.

Table 4. User-Replaceable Parts

Description Fluke Part Number Quantity

Fluke 62 Max Rear Decal 4134155 1

Fluke 62 Max + Rear Decal 4134162 1

Fluke 62 Max Battery Door Assembly 4103199 1

Battery AA Alkaline 1311933 1

Fluke 62 Max + Carabineer Solid Aluminum, 2.3 in with spring hinge


4149787 1
black finish

62 Max/62 Max + Users Manual 4060712 1

LIMITED WARRANTY AND LIMITATION OF LIABILITY


This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty
does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal
conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain
service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization
information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do
not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may
not apply to you.

Fluke Corporation Fluke Europe B.V.


P.O. Box 9090 P.O. Box 1186
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherlands
11/99