Ni Stir 5430
Ni Stir 5430
E127-Type Uitrasonic
Reference Blocks
J. A. SlotwinskI
G. V. Blessing
—QC
.056
100
NIST
1994
//5430
NISTIR 5430
J. A. Slotwinski
G. V. Blessing
May 1994
TECHNOLOGY ADMINISTRATION
Mary L Good, Under Secretary for Technology
"
%'t
toV
vL ' .-
K. >
jg#ig,Sia
(w
•
j-
Irl':-
nnaini:';^
,#'S<'«5«?Ii* > -•' -5'>' .
• :
'.V'. ^
iif
% ,
;
••Is
55s.'
k'*.^
"*
V
*
#
'’^
».
...
''>
'
-
-..,
=*1
<*"
J b
fi*
.
VAiWbaC^
-1^ TvV^V:'
^ LdUb-^ bi
'tM‘!juoe.,.-.,wit^M
;.
J.^
—
IkUkliM •
»
oi^ et
>x #^l
TABLE OF CONTENTS
Introduction 1
Measurement Facility 2
Instrumentation 3
Transducer 3
Pulser/Receiver 4
PT-4 Tuning Circuit 5
RC Circuit/Digital Voltmeter 6
Calibration Procedures 6
Preliminary Settings 6
Distance-Amplitude (D-A) Sets 7
Area-Amplitude (A-A) Sets 8
Calibration Report Contents 8
Conclusion 9
Acknowledgments 10
References 11
Appendix I: 13
Procedures for Requesting Ultrasonic Reference Block Calibrations
Appendix II: 15
Data- Acquisition Forms used in NIST Calibrations of ASTM E127-Type Reference Blocks
Appendix III: 19
Sample Distance-Amplitude Calibration Report
Tables 22
Figures 24
1
I" -r 1- ; >
•
„f
tMfbi^iimT ,JM.^
.' '
,
5
'
Dj^, "H
'
^'V-' ''
’
’
'&'
'}.
,
.:.#^ik,'
" ’
^..At
'
-
;
r ^
n \ - •/. ', ^;!
'^''\‘'^,V' i''V
- - ^
-r-V'J ^
« '
'
'
ats
'
:' '
",
"J
' '
’?
,
,
‘'%;i
fioq&M
Sj^ T. (V*J .
*
l'.i/^.!4
a
:!fr
^
^ V a y%$ha!B.td0o-9^
\‘»'. ^pd'
'•
©1
^
'
'T
7S
'-•‘^^
m
A* ^•».^:.V 7 .
ti
ff:-
'•'•IV
€1 j1 iSfeaisqfqA
mi'-
M~ :II jdl>ad<|qA
$'
':. /.
.
.^^
or tin xrbil2>qc|A^
eslctet
7 -
iB?
Infroduction
This document describes the procedures and instrumentation used by the National
Institute of Standards and Technology (NIST) aluminum ultrasonic reference
in calibrating
blocks. The Ultrasonics Group, in the Automated Production Technology Division of the
Manufacturing Engineering Laboratory, performs this calibration service for technological
and industrial customers. The calibration is based on the American Society of Testing and
Materials (ASTM) recognized practice aluminum reference blocks,
for calibrating
designated E127-92a; Standard Practice for Fabricating and Checking Aluminum Alloy
Ultrasonic Standard Reference Blocks [1]^ Instructions for requesting calibrations of
ultrasonic reference blocks are contained in Appendix I of this document.
The NIST ultrasonic reference block calibration service was originated in the late
1970's, and is NBS Technical Note 924 entitled "Procedures for the
described in the
Calibration of ASTM
E127-Type Ultrasonic Reference Blocks." [2]. At that time the
application of one master block for use as a reference standard was considered to be a
temporary or interim measure. Today NIST uses this master block to set the ultrasonic
system sensitivity in all calibrations of ultrasonic reference block sets. In addition, there
are several newly implemented practices employed for checking the calibration system
characteristics, both over a period of years and in preparation for a particular calibration.
This current document updates the NBS Technical Note 924 in describing these practices
^ Steel and Titanium Ultrasonic reference blocks are also calibrated by NIST. While E127-92a does not
apply to these types of blocks, the instrumentation and some of the procedures described in this document
are the same. ASTM E428: Fabrication and Control of Steel Reference Blocks used in Ultrasonic
Inspection covers these types of blocks.
1
E127-Tvpe Ultrasonic Reference Blocks
The reference artifacts described in E127 are 7075-T6 aluminum alloy, 50.8 mm (2
in)2 diameter cylindrical blocks. Following NIST policy, we give quantities in SI units
with in-lb units in parentheses. The ASTM El 27 standard generally gives the in-lb units
first. The blocks are of varying length and have flat-bottom-holes (FBH) drilled into their
bottom surfaces. A schematic of this type of block is shown in Fig. 1. In describing these
types of blocks the following labeling convention is used in El 27: the hole diameter,
expressed in 64ths of an inch, and the metal path length, expressed in lO-^ of an inch, are
separated by a dash. For example, a block with a 5/64 in diameter hole (also referred to as
a No. 5 hole) and a 0.50 in metal path would be designated as 5-0050. Blocks with this
labeling are commonly used; however, El 27 gives equivalent hole diameters and lengths in
SI units. This labeling convention is used throughout tWs document, but quantities will
generally be given in accord with NIST policy.
Measurement Facility
Distilled water, with a nominal temperature range of 20-24 °C, is used as the coupling
fluid. Previous research has shown that temperature variations in this range do not affect
the calibration results when all components are in thermal equilibrium [3]. The water is
kept chlorinated in order to prevent algae growth, etc., and while the conductivity of the
water is increased as a result, the short-term exposure of the blocks to the water is not
harmful to the physical characteristics of the blocks. In addition, a filtration system is
periodically used to keep the coupling fluid clean of small debris such as dust. Calibration
blocks are placed in the tank on a set of large aluminum steps, machined to a high
which keep all of the blocks' fi-ont-surfaces at the same height. This minimizes
precision,
the amount of vertical transducer manipulation needed during the calibration procedure.
A large moveable stage is attached to the top of the tank, as shown in Fig. 2. The
ultrasonic transducer is connected to by means of a 98 cm long ultra-high-
this stage
fi’equency (UHF) coaxial metal rod. The transducer can be manually manipulated in two
perpendicular planes through the use of an angulator. The transducer also has degrees of
^English units are given first with SI units in parentheses by ASTM in E127 in describing reference
blocks and transducer characteristics. This is contrary to the U.S. national policy of using metric units
only. However, this document both the English and metric equivalent of measured quantities in a
lists
manner consistent with NIST policy to provide coherency for those who are familiar with E127. NIST is
currently working with ASTM to promote the use of metric units only.
2
freedom in the -X, -Y, and -Z directions, with movement in these directions controlled
manually or with motorized linear actuators contained within the stage. This setup is
sufldcient for properly positioning the transducer during all phases of the calibration.
Instrumentation
Transducer
One of the most important components of the calibration system is the ultrasonic
transducer. Ultrasonic transducers convert electrical pulse-excitations into pulses of high-
frequency sound which are above the threshold of human hearing (in this case 5 MHz).
Pulse repetition rates of 200-1000 s*^ are typically used in this calibration. The ultrasonic
pulses are emitted from the transducer and travel through a water medium (i.e. coupling)
to the target of interest. At a material interface or discontinuity some of the ultrasonic
energy is back towards the transducer. The transducer then receives these
reflected
reflected echoes and converts them back into electrical signals for electronic detection.
For a given ultrasonic pulse incident on a target, the amplitude of the return echo depends
on the target size and shape, and on the acoustic properties of both the target and the
coupling medium. The measurement parameter of interest in the calibration of ultrasonic
reference blocks is the amplitude of the return echoes reflected from the flat-bottom-holes
drilled into the bottom of the blocks. In this particular calibration, the interface of interest
is a solid-air boundary between the solid aluminum and the FBH.
An Aerotech 5 MHz quartz immersion-type transducer possessing a 9.525 mm
(0.375 in) effective element size, historically designated "Lab-Standard-4" (LS-4) [4], has
been used in virtually all of NIST's ultrasonic reference block calibrations since the
inception of the calibration service. Quartz transducers, while not commonly used for
generic nondestructive testing due to their insensitivity, do have the advantage of
remaining very stable over long periods of use. In addition, quartz transducers are more
easily constructed to the same or similar performance characteristics, than are ceramic
units [4]. Similarity in performance between like-constructed transducers is important for
maint^ning widespread applicability of the ASTM E127-92a measurement standard when
using different equipment. The pulsed beam-profile characteristics of LS-4 have been
carefully and repeatedly measured over time using pulse-echo techniques with specific
targets. We note that LS-4 meets the requirements of section 11.5 of E127-92a in size,
operating frequency, and lateral and axial beam profiles. The transducer beam profiles,
taken on a 12.7 mm (0.5 in) diameter steel sphere target, are shown in Figs. 4-6. Figure
^Commercial equipment or materials are identified in this paper in order to specify adequately certain
calibration procedures. In no case does such identification imply recommendation or endorsement by the
National Institute of Standards and Technology, nor does it imply that the material or equipment
identified is necessarily the best available for the purpose.
3
4a shows lateral scans ofLS-4 taken at the transducer's near-field (Y^) length ^ The scans
2ire taken at four different rotational orientations of the transducer which are separated by
45° of rotation. Figure 4b shows a comparison of one of the 4a scans with one which was
taken by Chwirut in 1976 [2]. Figure 5 is a lateral scan of LS-4 taken at the transducer's
near-field minimum (1[") point. Figure 6 is an axial scan of LS-4 which clearly shows the
and points in the transducer's sound field. All of the beam profiles taken at the
present time are in good agreement with those obtained by Chwirut in the 1970's [2],
indicating that the transducer's response has not measurably changed with time.
To complement the pulse-echo scans, hydrophone scans have recently been made
of the transducer's beam profile. A thin polymer membrane, made of polyvinyldifluoride
(PVDF) in this case, can measure ultrasonic pulses traveling through it in either direction,
with minimal perturbation to the pulses themselves. Ultrasonic pulses measured by
traditional transducers by the transfer-function characteristics of the
are affected
transducer, resulting in a measured waveform that differs fi'om the actual waveform. In
contrast, the significantly greater bandwidth and closer impedance match to water of a
PVDF hydrophone allows for a high fidelity recording of the actual ultrasonic waveform.
A commercially available hydrophone possessing a circular element size of 0.3 mm^
(0.0005 in2) and a bandwidth > 20 MHz was used to map the ultrasonic beam intensity at
the transducer's near-field (I^) length. The transducer was moved slowly over a 645 mm^
(1 in^) area in a raster-scan fashion. The result, shown in Fig. 7, shows a good symmetric
beam profile which is fi'ee of significant irregularities.
Pulser/Receiver
The pulser/receiver currently used in NIST's calibration service is an Automation
Industries Inc. model M90 Reflectoscope. When this unit is used in conjunction with an
external LC tuning circuit, it provides pulse excitations of sufficient energy to drive the
requisite quartz transducers such as LS-4. The receiver has a coarse-adjustment
bandwidth extending to 10 MHz and a gain range of approximately 0-81 dB. This gain is
transducer's near field length is the natural focus distance in an ultrasonic beam emanating from a
flat transducer. The sound field intensity monotonically decreases beyond this length, and is therefore
generally a better region for target placement. The near field length N of a circular spot source, in a
coupling fluid, can be approximated analytically by the expression:
4A
where D is the transducer's active element diameter and X is the wavelength of the ultrasound in the
coupling fluid. Note that X varies inversely with the frequency of the ultrasound.
4
The receiver linearity of the M90 Reflectoscope is checked periodically, using the
setup shown in Fig. 8. A function generator is used to generate a signal pulse which has a
similar spectral content as a typical echo received by the M90 and LS-4 located 88.9 mm
(3.5 in, its 17 point), from the surface of a reference block target in water. This signal is
attenuated by an independently calibrated attenuation strip, and then detected by the M90
through its receiver port. The attenuation strip is calibrated using a regulated DG source
and has a standard uncertainty of ± 0.1 dB to ± 0.3 dB over the range of 1 dB to 60 dB.
A digital voltmeter measures the peak amplitude of the unattenuated input signal and the
subsequent attenuated signals. The detected peak amplitude is then compared to the value
computed using the unattenuated signal amplitude and the attenuation value. This is
repeated over a range of attenuation values, for each of several different receiver gain
settings on the M90. A typical linearity test for one receiver gain setting is shown in Fig.
9. The receiver portion of the M90 has a measured amplitude linearity of better than 1.5
% of full-scale (FS), over a range of amplitude values that exceed those normally used in
these calibrations. (While the M90 is linear within 1.5 FS up to 6 V, 0.3-5. 0 V has %
been defined to be the instrument's full scale linear range.)
In the calibration, the ultrasonic signal of interest is selected on the M90 screen for
measurement by positioning the M90's receiver gate. The peak amplitude of the signal
marked by this electronic gate is then detected by the M90 and delivered to a digital
voltmeter. An adjustable pulse-length knob on the M90 is kept at its maximum (full
clock-wise) position at all times. The M90 shock-excitation is then tuned with an external
tuning circuit described below.
the M90 is a shock-excitation pulser without the internal tuning capability for the tuned
excitation specified in E127-92a. Together these two units very closely approximate the
performance (when used with the same quartz transducer) of the calibration
pulser/receiver unit used by NIST in the 1970's and 80's.
5
PT-4 tuning circuit, and its corresponding frequency spectrum are shown in Fig. 12. In
the time domain this pulse resembles a damped one-and-a-half period sinusoidal pulse.
One variable which affects the frequency content of the excitation pulse reaching
the transducer is the length of the BNC cables and UHF rod between the pulser and
transducer. Longer cables result in a higher capacitance between the various pieces of
instrumentation used in the calibration. The higher capacitance affects the system's
of properly tuning the transducer's center frequency. As a result, great care has
capability
been taken to use the same cable lengths for all calibrations to ensure that the spectral
content of the excitation pulses remains constant. The cable lengths used in NIST
calibrations are indicated on the appropriate setups (Figs. 3 and 11.) These lengths, when
used in conjunction with the current calibration instrumentation, allow for a tuned center
frequency of 5.0 MHz.
RC Circuit/Digital Voltmeter
The positive peak amplitude of the detected ultrasonic pulse of interest from the
M90 detector unit is displayed on a digital voltmeter (DVM). This signal is filtered by an
RC circuit possessing a time constant of 0. 1 s, which reduces the effect of spurious noise
on the voltage readings by time averaging. This filtering improves the stability of the
signal reading on the DVM. Due to the long time constant of the circuit, any signal
maximizing (while aligning or normalizing the transducer for example) is usually done by
observing the signal on the M90 screen, and not by reading the DVM.
Calibration Procedures
Preliminary Settings
After the system's excitation pulse and linearity have been checked the calibration
can be performed. The following steps are taken for both Area-Amplitude (A-A) and
Distance-Amplitude (D-A) block calibrations. Appendbc n contains sample data-
acquisition formswhich show the detailed steps of each type of calibration.
is turned on at least one hour before data is taken, in order to
All instrumentation
allow for adequate instrument warm-up and stable voltage readings. Before being placed
in the tanks, all blocks are visually inspected, and any grease or oil removed from their
front surfaces. Nicks, scratches, rust, etc. which are on the front surfaces are noted for
each of the blocks, since these may affect the ultrasonic response of the block. This
includes any surfaces not appearing to meet the surface roughness criteria in section 10.2
of E127-92a [1]. Finally, the hole plugs on the bottom of each block are inspected to
make sure that they appear to be securely sealed and free of corrosion.
As shown in Fig. 2, blocks of varying metal-path-distance are placed in the water
on large aluminum steps. This keeps the blocks' front-surfaces at the same height, thereby
minimizing the need for adjusting the transducer-to-target distance during the calibration.
Once submerged the blocks' entry surfaces are lightly brushed to remove air bubbles which
might otherwise interfere with the ultrasonic beam and therefore affect the calibration
measurement. Back-lighting through the tank's transparent walls provides a means of
easily detecting these air bubbles.
6
A 88.9 mm (3.5 in) long spacer-block is then placed on top of one of the blocks to
be calibrated. The transducer is lowered to the top of this spacer-block, which is
maximum-amplitude reflection from this block's FBH is detected. The system gain is then
adjusted until the reflection amplitude is 80% of full-scale. In this procedure, we note that
a master reference block is used instead of the appropriately sized steel balls which are
called for in E127-92a.^ After setting the initial sensitivity, a gain change which depends
on the hole size of the blocks may be needed. Gain settings for the respective hole sizes
are determined by their relative FBH areas. Thus, for a number 3 size hole D-A set the
gain would be increased by a factor of (25/9), while for a number 8 size hole D-A set the
gain would be decreased by a factor of (25/64) compared with a No. 5 hole set. Previous
research at NIST has shown that these procedural differences yield more precise and
meaningful reference block calibration data, in consonance with E127-92a [6]. In
addition, the gain is increased by (31 to 808.2 %
FS) halfway through a D-A set, in
dB
order to obtain an adequate scope display and remain in the system linearity range for the
echo responses from the longer metal-path blocks. All gain changes are made in the
following manner, where a 3 1 % to 80 % FS gain increase is used as an example: The
transducer is manipulated laterally until the peak amplitude of any appropriate reflection,
usually that from a FBH, decreases to 31 %
FS. The gain controls on the M90 are then
adjusted until the peak amplitude of this signal reads 80 FS. The peak amplitude during %
this gain adjustment is read on the DVM. Throughout this document, including Appendix
n, gain changes are expressed in terms of dB. These are simply unit conversions from the
actual % FS changes used in the gain adjustments. The actual labeled gain values
indicated on the M90 are not used for these gain adjustments.
A check-standard (CS) block is measured for each gmn setting, to verify that the
appropriate gain changes have been made
These check-standard blocks are also
correctly.
used as a means for determining the overall uncertainty of the measurement system, as
discussed below. The physical characteristics of the CS blocks are given in Table 1.
^The appropriately sized steel balls called for in ASTM-E127-92a are measured after setting system-
sensitivityon URB-1. These values can then be used by the customer as a reference value. Care must be
taken when using the ratio of these values as a "NIST-to-E127 conversion factor," since the different
target geometries have different scattering characteristics. These include differences in the spectral
content of the reflected pulses and thus differences in system linearity for the distinct targets. In addition,
detailed experiments at NIST show that better accuracy is obtained with URB-1 than with steel balls [7].
7
During the calibration, the maximum echo-amplitude from each block's FBH is
read and recorded. The entire measurement scheme is done three times for each block in
the set. These three independent measurements yield an average and a check on
procedural repeatability for each block. A large variation for a particular block may be an
indicationthat one trial measurement was done incorrectly, and that additional
measurements are necessary. The block-response value reported to the customer is the
average value of these three trials.
Calibration Database
Figure 13 shows a comparison of the NIST calibration database, which includes all
calibrated aluminum reference block values, with the minimum and maximum limits
prescribed by E127-92a, for metal paths from 12.7 mm (0.5 in) to 152.4 mm (6.0 in) and
for No. 3, 5, and 8 hole sizes. The average value for each metal-path in the database is
indicated (•), along with their ± two standard deviation (2a) ranges for all blocks with this
metal-path, to indicate their 95% confidence intervals. The minimum and maximum limits
specified in E127
from the NIST database
are also indicated (
fall
— ). For all but the 0575 blocks, the average values
within the specifications called for by E127-92a. For the
larger metal paths there is a trend of the NIST average values to lie closer to the minimum
8
E127-92a values. Table III lists the average and standard deviation of all aluminum
reference blocks calibrated by NIST, for each metal path. The number of blocks
calibrated for each hole size is also indicated. It is worth noting that the large number of
44.5 mm (1.75 in) and 57.2 mm (2.25 in) metal path blocks is artificially large, since these
blocks are measured twice in each D-A calibration.
The ASTM committee E-07-06, which is responsible for maintaining the E127-92a
document, is currently examining the possibility of using the empirical NIST database for
establishing new E127-92a recommended values. If this is to be done a fit of the NIST
valuesmay be used. Here we fit the average NIST values to an empirical relationship
which was previously found by Chwirut to give the best fit for this type of data-set [7].
That empirical relationship is of the following form:
y(x)= (1)
(a + bxy
where x is the material path length and y(x) is the echo-amplitude fi'om the FBH. This
function's form has a sound theoretical justification, since the e"®" term takes into account
the material attenuation a, and the (a + bx)^ term is the predicted relation for the
reflection amplitude fi’om a disc residing in the far-field of the sound beam [7]. The
coefficient a is very small for 5 MHz ultrasound in aluminum, while the a and b
coefficients depend on the transducer characteristics, the water pathlength, transmission
and reflection coefficients, sound velocities of the coupling fluid and material, etc. Using
our database and a least-squares optimization algorithm [8] we have calculated numerical
values for these coefficients. These are presented in Table 11. These values describe the
NIST aluminum block calibration data, using LS-4 at a water path of 88.9 mm (3.5 in).
Figure 14 shows the NIST average values for each metal-path, and a plot of equation (1)
using the calculated values of the three coefficients. The calculated attenuation coefficient
a= 0.047 dB/cm (0.0138 NpAm) is compared with the value a= 0.02 dB/cm for 5 MHz,
as presented by McMaster for AL17ST (0.13 mm grain size) [9].
Check-Standard Database
Figures 15-20 show plots of the check-standard values measured fi’om 1976 to
present. The raw data (o) and ± 2a ( — ) values are indicated. Note that each point is the
average of the three measurements taken for each calibration. The ± 2a values represent
the expanded measurement uncertainties of the calibration system, with a representing the
combined standard uncertainty [10]. The measurement uncertainty of a measured block
value then depends on the particular block's metal-path distance and hole size, as indicated
in Table 1, since different check-standards are used with different-sized blocks.
Conclusion
9
calibration database which contains all No. 3, 5, and 8 hole-size aluminum blocks
calibrated by NIST since 1976.
Acknowledgments
The authors would like to thank Dr. David Gilsirm for his help with the numerical
least-squares algorithm. We would also like to acknowledge the many NIST and ASTM
E-07-06 readers of the draft copy of this document, including Dr. Donald Eitzen, Mr. Bill
Crupe, Mr. Howard Van Valkenburg (New Fairfield, CT.), and Mr. Jerry Green (Alcoa
Aluminum Co., Alcoa Center, PA.). All of their comments and suggestions for this
document are greatly appreciated.
10
References
[1] Standard Practice for Fabricating and Checking Aluminum Alloy Ultrasonic
Standard Reference Blocks, E127-92a, in Annual Book of ASTM Standards Philadelphia, .
[2] Chwirut, D.J., Sushinsky, G.F., and Eitzen, D.G., Procedures for the Calibration of
ASTM E 127-Tvpe Ultrasonic Reference Blocks . Nat Bur. Stand. (U.S.) Tech. Note
924; (September 1976).
[4] Chwirut, D.J., and Boswell, G.D., The Evaluation of Search Units used for
Ultrasonic Reference Block Calibrations . Nat Bur. Stand. (U.S.) Internal Report NBSIR
78-1454; (February 1978).
[5] Eastern NDT Inc., 200 N. Main Street, Suite 212, Hopewell, Virginia 23860
[6] Sushinsky, G.F., et al. . Improved Ultrasonic Standard Reference Blocks . NBSIR
76-984; (November 1976).
[8] Atkinson, K.E., An Introduction to Numerical Analysis . John Wiley and Sons, NY
(1978).
[10] Taylor, B.N., and Kuyatt, C.E., Guidelines for Evaluating and Expressing the
Uncertainty of NIST Measurement Results. NIST Technical Note 1297 (January 1993).
11
THIS PAGE INTENTIONALLY LEFT BLANK
12
Appendix I: Procedures for Requesting Ultrasonic Reference Block
Calibrations
Interested parties should call the technical contact listed below, before sending
block sets to be calibrated. The technical contact will inform the customer of the cost of
the calibration, and any other pertinent information. The customer should then send a
company purchase order and the block set(s) to the address below. The purchase order
should preferably be sent prior to the block set(s) and should clearly state the type of
calibration to be performed, the cost, purchase order number, the bill-to office, and the
name, address, and phone number of the person responsible for the block set(s). No
calibrations can be performed until NIST has received and processed this purchase order.
Therefore, the tum-around time for ultrasonic reference block calibrations can be
shortened if NIST receives the purchase order before the block sets. Normally, customers
can expect a four week tum-around time, although this time can be reduced in special
circumstances. A table listing the various reference block calibration services offered by
Technical Contact:
Mr. John A. Slotwinski
Building 233 / Room AMT
N.I.S.T.
Gaithersburg, Maryland 20899-0001
(301)975-2171
13
NIST Ultrasonics Group Reference Block Calibration Services
These prices are subject to change. Consult the latest edition of the NIST Calibration
Services Users Guide and Fee Schedule (NIST Special Publication 250, January 1993), or
contact the technical contact listed on the previous page for more information.
14
Appendix II: Data-AcquisitioP Fonns used in NIST Calibrations of
ASTM E127-tvpe Reference Blocks
15
COMPANY: DISTANCE-AMPLITUDE CAUB. NO.
S/N: ALUMINUM NO. 3 HOLE TECHNICIAN
DATE
BLOCK Trial 1 Trial 2 Trial 3
URB-1 80% 80% 80%
INCREASE GAIN 8.9 dB
3.175 mm (1/8 in) BALL
CS-1
3-0050
3-0063
3-0075
3-0088
3-0100
3-0125
3-0175
3-0225
INCREASE GAIN 8.2 dB
CS-2
30175
3-0225
30275
3-0325
3-0375
3-0425
3-0475
30525
30575
16
COMPANY: DISTANCE-AMPUTUDE CAUB. NO.:
S/N: ALUMINUM NO. 5 HOLE TECHNICIAN:
DATE:
BLOCK Trial 1 Trial 2 Trial 3
URB-1 80% 80% 80%
7.938 mm (5/16 in) BALL
(DS-S
5-0050
5-0063
50075
5-0088
50100
50125
50175
50225
INCREASE GAIN 8.2 dB
CS-4
50175
50225
50275
5-0325
50375
50425
50475
&0525
50575
17
COMPANY: DISTANCE-AMPUTUDE CAUB. NO.
S/N: ALUMINUM NO. 8 HOLE TECHNICIAN
DATE
BLOCK Trial 1 2
Trial Trial 3
URB-1 31.3% 31.3% 31.3%
17.463 mm (11/16 in) BAU
CS-5
8-0050
8-0063
8-0075
8-0088
8-0100
8-0125
Br0]75
8-0225
INCREASE GAIN 8.2 dB
CS^
8-0175
8-0225
8-0275
8-0325
8-0375
8-0425
8-0475
8-0525
8-0575
18
Appendix HI: Sample Distance-Amplitude Calibration Report
TN-OOOOOO-93
J.A. Slotwinski/bam
September 23, 1993
REPORT OF CALIBRATION
OF
Requested by
Any Company
P.O. Box 0001
Anywhere, USA
The set of fifteen aluminum reference blocks was calibrated according to the procedures
of the NIST document entitled "NIST Calibration of ASTM E127-Type Ultrasonic
Reference Blocks." In these procedures, the ultrasonic system sensitivity is based on the
echo amplitude from the National Institute of Standards and Technology (NIST) master
flat-bottom-hole (FBH) block of size 5-0050, rather than the steel reference balls called
for in ASTM E127-92a. Gain settings for the respective hole sizes are in turn determined
by FBH areas. Research at NIST has indicated that these procedural
their relative
differences yield more and meaningful reference block calibration data in
precise
consonance with ASTM E127-92a. (Using these procedures, the echo amplitude response
value of the 7.9 mm (5/16 in) steel ball was 70.1 FS.) %
The calibration results are given in Fig. For each reference block, the percent of full
1.
scale (% FS) average of three readings is graphed and tabulated. Note that the system
gain was increased from 31 to 80% FS, after the first measurement of the 57.2 mm (2.25
in) metal-path block. After taking into account NIST procedural differences and the
combined standard uncertainty of the system, the echo amplitude responses of all blocks
except 5-0575 fell within the range of the E127-92a established values. The 5-0575 block
was slightly lower than the established values of E127-92a.
19
We also report NIST system combined standard uncertainties using a NIST master set of
aluminum FBH blocks as check standards. These (total) system combined standard
uncertainty values are determined for each receiver gain setting that is used for each range
of blocks. They represent random and some systematic errors associated with the NIST
system. (We note that other systematic errors or changes may produce differences greater
than the uncertainties reported here.) Based on check standard data accumulated since
NIST mm mm
1976, the
of the block response readings, and for the long blocks [44.5
V64.4y
^ 2,4
mm to 152.4 mm (1.75 to 6.00 in)] after a gain increase is of the block response
readings. (The stated expanded uncertainty is twice the combined uncertainty (2a value),
corresponding to a 95% confidence interval.)
We note that the values reported herein apply to the blocks calibrated only with the
ultrasonic system and instrument settings used at NIST. The National Institute of
Standards and Technology cannot assure that the customer's blocks will have these same
values when measured in the customer's facility on another day. The dependence of the
echo amplitude on the particular ultrasonic system used is well documented, and is
20
I 1
m
D
CM
Amplitude
S
(%FS) M 00 ^ lO te <0 U CM o to 04 CO CO
(5 O i 8 oo.*»
48.8
40.6
S CO
S85
35.4
a •o
CM 8 CD
Echo
s
«o
I1 11
i
u> 0 U) c > «o to
8 !5 to
E
fl
§8 c
r>.
5 o o O i o c5S 8 8 S S 8 8
BlocklO
3 li
1
C
1z 5
CALIBRATION
BLOCK
V) Number
iP >* o
C
E o
D a
C E
E o ID
D o
< o
REFERENCE o Block
OQ
o o
o o
ULTRASONIC
O
CO
O Q
"O
Q
io
O
^ O
CO
O
CM
O
Table 11; Calculated values of numerical coefficients in Eq. (1), using NIST database.
22
Table HI: Average, standard deviation, and number of blocks calibrated for each metal
path in the NIST database, since 1976.
23
Ultrasonic Wave Entry Surface
Flat-Bottom-Hole
24
Figure 2: Photograph of ultrasonic calibration system, immersion tank, and block holder.
25
Figure 3: Ultrasonic Instrumentation Setup for calibration of ASTM E127-92a type
aluminum reference blocks, with cable lengths indicated.
26
90
(a)
0 Degrees
(b)
1993
• 1976
Figure 4: (a) Lateral scans of LS-4 taken using a 12.7 mm (0.5 in) steel sphere target at
89.9 mm (I^ point) water path distance. The scans are taken at four different rotational
orientations of the transducer, which are separated by 45° of rotation, (b) Comparison of
one of the 4a scans with one which was taken by Chwirut [2] in 1976. The good
agreement between the 1993 continuous scan data and the 1976 discrete data is one
indication of LS'4's stability over time.
27
o
o
oo
o s s o o
CO
o
CM
O O
(SJ 7o) spn«|duiv 0403
28
target,
sphere
steel
in)
(0.5
mm scan.
12.7
the
a on
using
indicated
taken
are
points
LS-4
of
Yf
scan
and
Axial
6:
transducer's
Figure
O
O Q
^ Q
lO Q
^ O
CO
O
CN
O O
29
10
map
its by
at This
separated
field
mV.
sound 40
are
irregularities.
to
LS-4 contours
the
of dark
corresponding
significant
in^)]
of
(1 heavy
contour free
The
mm^
is
645
dark
which
hydrophone.
(central) profile
[approximately
first
PVDF beam
a the
map
by with symmetric
Intensity measured
good
7: increments,
as a
Figure
point,
shows
mV
30
Figure 8: Arrangement used to measure the M90‘s receiver amplitude linearity.
31
JD
2
13
o o
0
•D
8 peak
ideal
8 and
measured
8
the
4>
8 D
3 showing
id •i—
a
E
amplitude.
8
< measurement
o peak
<D
CL
T3
8 O
predicted
linearity
O
T3 of
0)
o. M90
8 typical
function
a
A as
9:
8
amplitudes
Figure
8.00 88888888
l<'dlO‘^COCN»—
(A) epruiiduiv ^loed pejnsDe|/\|
o
8
32
"Output" Jack "Receive" Jack
"Transmit" Jack
Figure 10: PT-4 tuning circuit schematic. This circuit is used to tune the M90 shock-
excitation.
33
Figure 11: Setup for checking the finequency
content of the M90 shock-excitation.
34
(a)
Time (fis)
Frequency (MHz)
Figure 12: (a) Typical shock excitation pulse (tuned) and its (b) corresponding frequency
spectrum on a linear vertical scale.
35
)
two
NIST {± and
2a
by ±
minimum
the
calibrated
the
represent
blocks
represent
all
(X)
the
contains
(
and
The
value,
which
database.
amplitude
database
the
in
block echo
length E127.92a.
average
path
reference
ASTM
the metal
in
ultrasonic
represent
each
specified
for
NIST (•)
as
The
The deviation)
limits
13: 1976.
maximum
standard
Figure
since
o o o o o o O o
0*001
g g R g g g o
CO
C)
CM
c5 o
36
best-fit
II.
corresponding
Table
the
from
and
(o)
coefficients
values
calculated
average
the
database
and
1
NIST
Eq.
the using
of
plot made
A is
14:
which
),
Figure —
8 S O
CO
O S S o o
CO
O
(N
37
O)
1.
Cd
Check-Standard
o
o>
indicated.
from
00
00 1
are
Inc. CO measured
Tabulation
(D limits
154-4
CO
00
> ) —
Industries,
CO amplitude (
TO
s/n c
00 (D deviation
CO echo
Data 3-0225,
CM
o the
Automation
00 0 of
standard
E
o
00
i- charting
±
two
CS-1 and
(©)
00 Historical
h- points
15:
CO data
Figure
The
o o
o o)
o
(30
o
r^
o
in
o o
CsJ Check-Standard
o>
O indicated.
from
abulation Inc.
CO are
00 measured
limits
154-4 )
CD
—
T
Industries,
00 amplitude (
s/n
deviation
00 echo
3-0175,
Data
Automation the
CVJ standard
00 of
o
00
charting
±
two
CS-2
and
Historical
(o)
00
points
16:
CO data
Figure
The
ooooooooooo
O0)00N-CDLOM-CQC\J'i-
(Sd %) 0 pni!idijuv OM03
39
3.
a>
eg Check-Standard
O)
o
o>
indicated.
from
are
abulation 00 measured
CO limits
Year)
122
)
Inc. CO
(
T s/n
00 amplitude
(Calendar deviation
Trienco, 00 echo
5-0075,
Data
the
CM of
standard
00
Time
two
charting
±
CS-3 and
Historical
(o)
points
17:
data
CO
Figure
The
O)
CM Check-Standard
o>
o from
indicated.
are
abulation
00 measured
00 limits
122 ) —
CD
Inc.
00 (
T s/n
amplitude
deviation
Trienco,
00 echo
5-0375,
Data the
standard
CM of
00
two
o
charting
±
CS-4 00 and
Historical
(o)
00 points
r^
18:
data
CD
Figure
The
ooooooooooo
OG)CX)r^CDLO^COCM^
(Sd %) 0pni!|dijuv 0^03
41
O)
OJ
CD
O
O)
abulatlon
00
00
123
Inc.
CO
T s/n
00
Trienco,
00
Data 8-0175
CM
00
CS-5
CO
"M"
O)
CVJ Check-Standard
O)
o
O)
indicated.
from
are
00 measured
00 limits
Tabulation
123 ) —
Inc. CO
00 co amplitude (
s/n "D
c
o deviation
00 CO
echo
Trienco,
Data
8-0325 the
CM
00
0 of
standard
e
o H charting
±
two
CS-6 00 and
Historical
(o)
00
points
r--
20:
data
co
Figure
The
ooooooooooo
oocDh-com^cocvj^
(Sd %) apniiidoiv 0403
43