Metcalfe3 ch1
Metcalfe3 ch1
3   rd
                                             EDITION
                                                         chapter from
   THE                                                   The Physics of
                                                         Radiotherapy X-Rays
  Physics
     Radiotherapy
                                  OF                     and Electrons©.
    X-RaysElectrons  AND
                                                         This sample chapter is
                                                         copyrighted and made
                                                         available for personal use
                                                         only. No part of this chap-
                    Peter Metcalfe                       ter may be reproduced or
                     Tomas Kron                          distributed in any form
                      Peter Hoban
                       Dean Cutajar
                                                         or by any means without
                        Nicholas Hardcastle              the prior written permis-
                                                         sion of Medical Physics
              MEDI CAL P HYS I C S P U B L I S H I N G
                                                         Publishing.
1            Medical Linear
             Accelerators
                       “A boy surfing on a water wave provides
                          a useful traveling wave analogy.”
                                                —Karzmark and Morton, 2017
1.1      Introduction
         For the treatment of deep-seated tumors, high-energy x-rays with penetrating char-
         acteristics are required. The medical linear accelerator (linac) is currently the most
         popular device for this application. To ensure a full understanding of linac beam
         properties, it is important to review the mechanism of x-ray beam production by
         these types of devices.
              Early historically significant methods of producing x-ray beams for external
         beam radiotherapy (teletherapy) consisted of conventional x-ray tubes (with anode
         and cathode) producing x-rays of energy up to about 300 kV (Johns et al. 1952a).
         These superficial (up to 150 kV) and orthovoltage (up to 300 kV) machines are still
         used very effectively for treatment of skin cancer. They do not, however, provide
         skin-sparing properties (Locke et al. 2001), and their beams are not penetrating
         enough to treat deep-seated tumors effectively.
              The search for a more penetrating beam led to the development of cobalt-60
         machines (Johns et al. 1952b, 1959). These became more popular than other tele-
         therapy sources (e.g., cesium-137) because a large source activity of several thou-
         sand Curies of cobalt-60 could be packed into very small cylinders with diameters
         of about 10 mm (Johns and Cunningham 1983). The cobalt-60 machine is still used
         in radiotherapy applications today. A cobalt-60 beam spectrum has two photo-
         peaks, at 1.17 and 1.33 MeV, giving a mean photon energy of about 1.25 MeV.
         Despite their rather low photon energy compared with that of a linac, they still pro-
         vide reasonable skin sparing properties, as maximum dose is achieved at a depth of
         0.5 cm. Cobalt-60 units are very reliable due to their relatively simple design.
                                            1
       ©Medical Physics Publishing • For Personal Use Only
              The evolution of the linac was a direct result of radar development work that
         culminated in the production of microwave generators in the form of magnetrons
         and the klystrons. The first reported use of a medical linac was in 1953 (Thwaites
         and Tuohy 2006), but due to cost and complexity leading to reliability issues, they
         coexisted with Cobalt-60 units in radiotherapy departments for decades. As reliabil-
         ity improved, linacs gained ascendancy in the 1980s, and they are currently the
         most frequently used modality for radiotherapy applications. By 2013 there were
         only 1880 cobalt-60 units in operation globally, compared with 11,000 linacs
         (Attun et al. 2015).
              Linacs are capable of establishing intense electromagnetic fields in microwave
         cavities. This enables the acceleration of electrons to relativistic velocities when
         incorporated with suitable waveguide structures. In summary, linacs have the fol-
         lowing important features:
         1.   Linacs have multiple electron and photon energies that allow the physician to
              tailor treatment to the required treatment depth. A modern linac is usually
              capable of producing at least two different photon energies and a number of
              different electron energies.
         2.   Linacs have relatively high dose rates (1 to 15 Gy per minute). This enables
              short treatment times.
         3.   Linacs have a sharp dose fall-off at the beam edge in the penumbra.
              Currently the most compelling alternatives to using linac x-ray beams seems to
         be protons. Data from the Particle Therapy Co-operative Group (PTCOG) confirm
         there are currently about 60 clinical proton facilities worldwide with about 30 new
         centers currently under construction (PTCOG 2021). Building a synchrotron or
         cyclotron to produce protons is a high-cost option, even though the beam line may
         be shared by several (usually three) treatment delivery portals. The gantries may be
         on the order of 100 tons, and complex magnetic beam steering is required. The cost
         of these facilities varies, but the capital expenditure for such a facility is typically
         between 5 to 20 times more expensive than for setting up an x-ray linac facility.
         With the recent introduction of superconducting cyclotrons and smaller synchro-
         trons, single-room proton facilities are now available with capital costs being sig-
         nificantly lower than multi-room solutions.
              Synchrotrons can also be used to deliver heavier particles, such as carbon-12.
         There are currently a handful of carbon-12 beam lines with associated gantries
         being used to treat cancer patients. Carbon-12 is particularly interesting, as it seems
         to have a direct ionizing damage effect (like neutrons) that makes it suitable for
         treating hypoxic tumors such as prostate (Ishikawa et al. 2006). Synchrotron
         designs exist that can combine beam lines to provide carbon-12 to one gantry and
         protons to another within the same facility.
              One very favorable property of protons and carbon-12—which x-rays do not
         possess—is their well-defined Bragg peak. Due to this, the beam reaches a peak of
         dose deposition at significant depth and then has a rapid fall-off beyond this depth.
         Proton beams of 250 MeV are used to treat deep-seated tumors, and at about
         90 MeV they play a role in treating optic melanomas. Historical randomized trials
         comparing protons with x-rays are limited (Shipley et al. 1995). Discussions of
         recent clinical trials in progress are outlined by Bekelman et al. (2018).
              In this chapter the basic operational features of linacs are described in Section
         1.2. More details of magnetrons, klystrons, and linac waveguides are described in
         sections 1.3 to 1.5. Beam delivery systems in the treatment head—including bend-
(a)
(b)
Figure 1.1    Medical linear accelerators (linacs). (a) Varian TrueBeam®. (Image courtesy of Varian Medical
              Systems, Inc. All rights reserved.) (b) Elekta Versa HDTM linac. (Image courtesy of Elekta.)
(a)
(b)
Figure 1.2      Linac guide configurations. (a) Linac beam horizontal guide delivery system schematic shown in
                two dimensions, including gun, guide, bending magnet, movable target, flattening filter, carousel
                to house electron foils, and collimation systems. These linacs usually have a removable target
                and are capable of producing multiple x-ray and electron energies up to about 23 MeV. (b)
                Linac beam vertical guide delivery system schematic shown in two dimensions, including gun,
                guide, target, and collimation systems. These linacs have a fixed target and do not have an elec-
                tron mode. Modern designs also do not have a flattening filter and deliver one fixed x-ray
                energy, e.g., 6 MV.
                   Descriptions in this text are intended to give the reader an understanding of the
              principles of operation of the main components of the linac. Detailed descriptions
              of technical design variations of components—such as waveguide structures and
              bending magnets—can be obtained elsewhere (Podgorsak et al. 1999; Karzmark
              and Morton 2017).
Figure 1.3 Block diagram showing linac support and auxiliary system.
         gen-filled thyratron is fired, acts like a switch, discharging the voltage from the
         PFN. The peak current through the thyratron is about 500 amps, hence this is the
         only viable switching device. The resulting current passes through the primary
         windings of a pulse transformer. This is an autotransformer with multiple parallel
         windings; the low-voltage end is connected to earth, and the high-voltage end is
         connected to the cathode of the microwave generator. The voltage across the PFN
         follows a charge–hold–discharge cycle. The discharge is determined by the PFN,
         and the frequency is determined by the pulses applied to the thyratron to make it
         conduct (Greene 1986). The frequency of the thyratron switching is controlled by a
         pulse repetition frequency (PRF) generator. The microwave source is then provided
         with short-duration pulses from the pulse transformer.
(a)
(b)
(c)
Figure 1.4    Microwave cavities and klystrons. (a) The electric and magnetic field orientations at an instant
              of time in a microwave cavity. (b) Schematic of a two-cavity klystron. (c) Timing diagram for
              the E field in the buncher cavity.
                     The low-power microwaves that are fed into the input of the klystron are usu-
                ally provided by a solid state oscillator at about 50 watts. This is amplified by the
                klystron to an output in the order of megawatts, which is used to energize the accel-
                erating waveguide. Klystrons used in linacs usually have five cavities to improve
                high-current bunching and increase amplification. An amplification of microwave
                power in the order of 1011 is typically achieved by this process. Typically 7 mega-
                watts of peak power are produced by klystrons.
                     An alternative way to generate microwaves of sufficient energy is with a mag-
                netron. These are generally used to power lower-energy linacs, but this is not
                always the case. The magnetron is generally physically smaller and lighter than the
                klystron, hence klystrons are normally mounted in or behind the stand, whereas
                magnetrons can be mounted in the gantry. As shown in Figure 1.5, electrons emit-
                ted from the cathode are accelerated by a pulsed electric field, Edc, toward the
                anode. The electrons produce an additional charge distribution that is shown on the
                anode poles and induces an electric field, Edc, of microwave frequency between
                each segment of the anode. The electrons travel in spirals under the combined influ-
                ence of the two E fields and the accompanying H field components. Most of the
                electron path deflection is due to the H fields. In the process, the electrons are
                slowed, and most of the electron beam energy is converted to microwave power. An
                output aerial is inserted into one of the cavities to couple the microwave power from
                the magnetron to the waveguide.
Figure 1.5      A magnetron used to generate microwaves. Electrons move in a coiled path under the influ-
                ence of two electric fields: a static field Edc and a pulsed electric field Erf applied between anode
                and cathode. Because the electrons are moving in coiled paths at varying speeds, bunching
                occurs and microwave energy is given off. Electrons move along the negative direction of the
                field lines, some being accelerated and some retarded due to their phase relative to the RF
                field. The electrons cluster in a spoke formation that rotates around the cathode at the same
                velocity as the RF wave.
Figure 1.6 A grid dispenser type gun commonly used in modern linacs.
         at a lower voltage (nominal 10 kV, versus 45 kV). During gun operation the cathode
         is heated to temperatures of between 1100 and 1200 degrees Celsius. Gun lifetime
         is important because replacing the gun usually requires the linac to be out of opera-
         tion for several hours while the vacuum system is reconditioned. Some designs
         even require the gun and guide to be replaced at the same time.
               Voltages of between –150 and +180 are applied to a small grid in front of the
         cathode to control gun current; the more positive the voltage on the grid with
         respect to the cathode, the higher the gun current. The gun grid controls the flow of
         electrons from the cathode to the anode by altering the amplitude of the grid pulses.
         This is usually called gun current adjustment. If the grid is set at –150 V there is no
         electron flow and, therefore, no gun current. Electrons are injected simultaneously
         with the high-power microwaves into the accelerating guide.
               The electron gun is controlled by a gun driver subsystem. Its functions are to:
             •   supply a source of power to heat the gun filament,
             •   supply a source of high voltage for the gun cathode, and
             •   supply a program pulse of correct phase and wave shape for the gun grid.
Figure 1.7    Traveling wave E field at times t1, t2, and t3. (Reprinted with permission from Karzmark and
              Morton, A Primer on Theory and Operation of Linear Accelerators in Radiation Therapy,
              3rd Ed., Medical Physics Publishing, 2017.)
Figure 1.8      Charge distribution in a traveling wave accelerator in which electrons are accelerated.
                (Reprinted with permission from Karzmark and Morton, A Primer on Theory and Operation
                of Linear Accelerators in Radiation Therapy, 3rd Ed., Medical Physics Publishing, 2017.)
Figure 1.9    Standing waveguide E field distribution at times t1, t2, and t3. (Reprinted with permission from
              Karzmark and Morton, A Primer on Theory and Operation of Linear Accelerators in Radiation
              Therapy, 3rd Ed., Medical Physics Publishing, 2017.)
                   Figure 1.11 shows a sequential look at the E field from times t1 to t9 over a
              complete microwave cycle for a standing wave accelerator. Injected electrons from
              the gun are captured, bunched, and accelerated in the first few cavities. They then
              pass through the next cavities in a negative E field and are accelerated. At this time
              the next cavity has a positive E field and no electrons can be accelerated in it, which
              doesn’t matter because at this time the electrons have not reached this cavity. As the
              electrons cross into the next cavity, the E field in this cavity starts its negative
              excursion, and the electrons are further accelerated.
                   The microwave frequency is about 3 GHz. The magnetron or klystron receives
              5-microsecond pulses from the PFN with a 5-millisecond gap between pulses. Note
              that there are actually thousands of pulses in each sequence. It has been estimated
              that there are 10,000 electrons carried in each 30-picosecond microwave pulse, and
              each pulse has an approximate 300-picosecond separation with about 150,000 elec-
              trons in each 5-microsecond packet of pulses (Krieger and Petzold 1989). Shown in
              Figure 1.12 is a typical pulse sequence, including the pulse-forming network pulse,
              which is visible using an oscilloscope.
Figure 1.10   The evolution of the side-coupled standing waveguide design. (Reprinted with permission from
              Karzmark and Morton, A Primer on Theory and Operation of Linear Accelerators in Radiation
              Therapy, 3rd Ed., Medical Physics Publishing, 2017.)
                        Example 1.1
                        Question: A linac is delivering a dose rate of 2 Gy/minute to a calibra-
                        tion point within a water medium. The radiation is in 5-microsecond
                        pulses (assuming the pulses are continuous), and the pulse period is 5
                        milliseconds. Calculate the instantaneous dose rate during a pulse.
                        Answer: The dose rate is 2 Gy/minute, which is 2/60 Gy/second. The
                        pulses have a 5-millisecond period, which means there are 200 pulses
                        per second.
                                                   2
                        The dose per pulse is            1.67  10 4 Gy. Therefore, the dose rate
                                                60  20
                        in the pulse per minute equals the above value multiplied by 60 to get to
                        dose per minute and divided by 5 × 10–6, as this is the pulse period. This
                        gives an instantaneous dose rate within the pulse of 2000 Gy/minute.
Figure 1.11    Standing wave E field pattern for one full microwave cycle shown at nine discrete time inter-
               vals. (Reprinted with permission from Karzmark and Morton, A Primer on Theory and Opera-
               tion of Linear Accelerators in Radiation Therapy, 3rd Ed., Medical Physics Publishing, 2017.)
Figure 1.12    Linac x-ray and electron beam pulsed beam delivery. Figure shows typical times from the
               pulse- forming network and the approximate number of electrons in each wave packet. (Cour-
               tesy of Saree Alnaghy, University of Wollongong.)
Table 1.1 Features of traveling and standing wave accelerator structures used for linacs
                                                         E = mc2                                (1.1)
               where c is the speed of light.
                    Hence, the kinetic energy, Ek, given to an electron accelerated in a linear accel-
               erator waveguide is
                                                 Ek = mc2 – m0c2                                 (1.2)
           where m0 is the electron rest mass (9.11 × 10–31 kg). By rearranging Equation (1.2)
           then
                                                        E
                                              m = m0 + 2k .                              (1.3)
                                                        c
               The particle’s mass, m, is related to m0 by
                                                      m0
                                             m=                  ,
                                                  ⎛ ν 2 ⎞ 12                              (1.4)
                                                  ⎜1 − 2 ⎟
                                                  ⎝   c ⎠
                    Example 1.2
                    To acquire a feeling for the quantities involved when dealing with linear
                    accelerators, let us apply the above equations to a typical electron
                    energy. By applying equation 1.3 to a 6-MeV electron beam (1 MeV =
                    1.602 × 1013 joules), the result is that the electron acquires a mass m =
                    12.7 m0. By applying Equation (1.5), the velocity of the electron v =
                    0.994 c, which is within 0.006 of the velocity of light.
(a)
(b)
Figure 1.13   (a) Linac bending magnet design for a 270-degree bending magnet and achromatic energy slits.
              (b) Slalom design used to create 90-degree bend at end to enable efficient head-to-patient
              separation.
              at the desired position. This focus of electrons is needed to maintain a small virtual
              beam source size. Most of these magnet designs are fitted with an energy slit that
              removes electrons that are not within 5% of the nominal peak accelerated electron
              energy. The radius of curvature of less or more energetic electrons is different;
              therefore, the energy slit acts as a physical barrier and removes electrons from the
              beam path. In practice, three 90-degree magnets are usually used, connected by
              short drift tubes. This, in effect, reduces the height of the machine isocenter because
              the target can be mounted closer to where the electrons enter the magnet.
              Some accelerators use magnets to deflect electrons by small angles within the
          guide, with the final deflection being about 90 degrees (see Figure 1.12c). This is
          known as a slalom waveguide as employed by some Elekta linacs, and this has
          some perceived advantages:
          1. It reduces the length of the guide.
          2. It acts as an achromatic device, tightening the energy spectrum of the electrons.
          3. It lowers the machine isocenter height.
(a)
(b)
(c)
Figure 1.14   Components within a Varian linear accelerator treatment head. (a) Target from a linear accel-
              erator with two different thicknesses of material for bremsstrahlung production for 6 MV (left)
              and 10 MV (right). The target is water cooled. This target resides inside the guide vacuum,
              hence the seals and bellows. It is pneumatically driven into one of three positions: out (elec-
              trons), 6 MV, and 10 MV. (b) Two flattening filters from a linear accelerator: 18 MV (left) and
              6 MV (right). (c) External view of a monitor unit ionization chamber assembly from a linac.
(a)
(b)
Figure 1.15    Schematics of a monitor unit ionization chamber assembly. Orientation of the ionization cham-
               bers mounted in the treatment head used to monitor dose output, symmetry, and flatness is
               shown. AB and CD ionization chambers are used for integration of charge to give dose output
               in monitor units. (b) The A to B and C to D current ratios are compared for feedback to angu-
               lar beam steering in the radial and transverse planes. The E to F and G to H current ratios are
               compared. The current signal from the monitor ionization chambers to the electronic beam
               steering control circuit in the bending magnet is used to tune the beam for symmetry.
           Kapton®. Mica was previously used, but because of a shortage of this material,
           Kapton is now more common. The bremsstrahlung contamination from Kapton
           chambers in electron mode is also less.
                The mica chambers were usually filled with nitrogen, whereas the Kapton
           chambers are filled with oxygen-enriched air. Both of these types of chambers are
           sealed to avoid the necessity of correcting for changes in gas density within the vol-
           ume due to ambient pressure and temperature variations. Some linacs use chambers
           that are not sealed; these generally have an accurate electronic temperature pressure
           compensation device.
                The monitor unit ionization chamber plates are mounted so that one is at a
           90-degree rotation offset from the other. This enables them to also monitor beam
           symmetry and flatness in the beam’s radial and transverse planes, respectively.
           Beam flatness and symmetry are controlled by feedback circuits that run from the
           ion chambers to the bending magnet’s beam steering coils. Figure 1.13e shows how
           the dose to the chamber plates will alter due to positional and angular deviations in
           beam steering. The prescribed dose needs to be delivered reproducibly for each
           patient treatment. To achieve this routinely, one set of the monitor ionization cham-
           bers’ plates (the inner plates) are used to monitor dose output. The units of dose
           recorded by these are referred to as monitor units (MU); one MU of dose has been
           delivered when the monitor chambers have detected a preset dose. Because these
           chambers are located above the final beam collimation system, an MU setting is
           calibrated to a standard dose for a standard field setting. For example, 1 MU may be
           calibrated to equal 1 cGy at 100 cm source surface distance (SSD) for a 10 ×
           10 cm2 field size at dmax in water. However, 1 MU will equate to a different dose at a
           different field size as the collimator and phantom scatter changes depending on the
           field size.
                The monitor chambers are interlocked to stop beam production at an operator
           preset dose level. As mentioned for the symmetry monitoring, there are two sets of
           monitor ionization chambers. They are linked to two independent calibration cir-
           cuits. One is a backup to ensure that if the primary channel fails, the other channel
           will terminate dose. The backup is set to terminate dose at a small margin beyond
           the primary interlock.
                Further dose regulation, apart from the monitor ionization chambers, includes a
           timer that terminates dose at a preset time. This means the treatment will terminate
           due to the time interlock if both ion chambers fail. Care is taken to ensure that the
           time is set slightly longer than the time needed to deliver the dose, provided the pre-
           set dose rate is achieved. If a significantly lower-than-expected dose rate is being
           given, the treatment may also be terminated first by the time interlock.
                Of course, the operator can terminate treatment at any time by using the beam-
           off button on the control console. This termination usually occurs if patient move-
           ment is detected within the room via the in-room video monitors. If this occurs, the
           treatment is suspended, and the patient is realigned into the correct position prior to
           completing delivery of the prescribed dose. A two-way audio monitor is also pro-
           vided for patient-operator communication.
(a)
(b)
Figure 1.16    Electron scattering foils and spectrum. (a) Two-stage 18-MeV electron scattering foil from a
               Varian linac. (b) Typical spectral energy spread for an electron beam showing how the energy
               changes as the electrons move from the scattering foil to the patient surface. The spectral
               energy at 2.5 cm depth and the depth near the end of the electron practical range is also
               shown. This indicates the rapid decrease in electron energy with depth.
               principle, the dose rate for an electron beam could be much higher for electrons
               than for x-rays.
                    For safety and dosimetric reasons, the electron gun current is typically reduced,
               and the power of the magnetron or klystron may also be reduced so electrons are
               delivered at similar dose rates to x-rays when used in that mode.
                    Most linacs producing electron beams employ a bending magnet to ensure a
               relatively monoenergetic beam. The energy selection slit is typically set to allow
   electrons with a window of some 3% around the nominal energy through the bend-
   ing magnet. Modern linacs typically have five or more electron energies available.
        After the bending magnet, the electron beam has a diameter of only a few mil-
   limeters. In order to create a useful clinical beam, the electron beam must be spread
   out. Two methods have been used in different linacs:
   1. A magnetic field can be used to scan the electron beam across the area to be
        irradiated. This is similar to a television screen and has been used historically
        in some linac systems. This technique produces a superior beam in terms of
        energy definition, i.e., the beam spectrum has a very small spread. Photon con-
        tamination is also reduced as none is introduced by the absent scattering foil.
        However, the beam steering requires additional quality assurance, and there is
        a high instantaneous dose rate in the scanned beam, resulting in dosimetric dif-
        ficulties. Therefore, in the interest of safety, most commercially available lin-
        acs employ a scattering foil.
   2. A scattering foil is employed in most current linac designs. Most linacs use a
        dual scattering foil as shown in Figure 1.16a. The foil device shown is for a
        Varian 18-MeV electron beam. Note the device houses two foils, and there are
        different foils mounted at different locations on the carousel, each optimized
        for different electron energies. The design constitutes a compromise between
        producing a large enough useful beam and limiting both energy degradation
        and bremsstrahlung production.
        Because the electrons have not interacted with a target, the beam spectrum is
   closer to a monoenergetic beam of the peak energy, as shown in Figure 1.16b. Elec-
   trons as charged particles do interact more with air than photons, and this leads to
   some energy degradation as the electrons traverse away from the scattering foil.
   There is also an increase in angular scattering. In order to achieve a relatively well-
   defined beam edge, it is necessary to collimate the beam as closely as possible to
   the patient. This is achieved using an electron applicator, as shown in Figure 1.17a.
   The applicator has several planes in which the beam is collimated. The final colli-
   mation plane allows placement of an insert made from a low-melting alloy (LMA)
   in the applicator. This insert allows customization of the final collimation to indi-
   vidual patients. Figure 1.17b illustrates schematically the role the applicator and
   other collimators play in the setup used on a linear accelerator to generate and
   transfer a useful clinical electron beam to a patient.
(a)
(b)
Figure 1.17    (a) Electron applicator used to improve scattering properties when electrons are employed.
               This applicator is placed on the block tray holder during electron mode and, when used at 100
               cm SSD, the applicator is offset 5 cm from patient skin. An insert is mounted in the applicator
               to create an irregular-shaped electron field. (b) Schematic showing the electron applicator
               design in relation to the linac head components and how the device restricts scattered elec-
               trons from escaping.
1.7        Collimation
           Various devices are used to collimate and modify the intensity of the x-ray beam.
           These devices are reviewed in the following sections.
Figure 1.18    Asymmetric collimators from a beam’s-eye view and side view perspective. Both sets of colli-
               mators can be set asymmetrically or symmetrically about their axis.
               and the other jaw driven farther from the central axis. This creates an asymmetric
               field with the x2 jaw closest to the central axis providing a less-diverging edge,
               while the other jaw, x1, has a more diverging field edge.
1.7.4          Blocks
               The two sets of collimators can only provide rectangular field shapes on the patient
               surface. Therefore, other devices were historically used to create irregular field
               shapes so that organ-at-risk structures—such as heart, lung, or spine—may be
               shielded from the x-rays.
                     Lead blocks (see Figure 1.19) were historically used to modify the x-ray treat-
               ment field dose distribution. These were mounted on trays known as block trays
               that slide into a removable accessory tray mounted at the end of the treatment head.
               Block trays were usually made from acrylic or polycarbonate that was transparent,
               so in combination with the light field this was suitable for field alignment with
               patient skin tattoos. (Typical dimensions were 30 cm × 30 cm × 0.6 cm.) Block
               trays created extra electron contamination in the beam, thus increasing patient skin
               dose. They also attenuated the beam (typically by about 3%), and a block tray cor-
               rection factor for beam output was applied.
                     Block thickness was generally sufficient to provide at least five half-value lay-
               ers (HVLs) of shielding. Each linac used to have a set of lead blocks in standard
               shapes, in particular a set of long, narrow blocks for spinal cord shielding. Low-
               melting-point alloy (LMA) was also used in some radiotherapy departments to
               shape blocks for irregular x-ray fields. The molding of LMA into irregular shapes is
               still common for creating electron inserts for electron treatments. The density of
               lead is 8.3 g cm–3 versus 11.3 g cm–3 for LMA. Blocks now play a very limited role
               in radiotherapy. The introduction of multileaf collimators have rendered blocks vir-
               tually redundant.
Figure 1.19   Blocks that may be placed on the accessory tray mount on a block tray at the exit position of
              the linac treatment head.
(a)
(b)
Figure 1.20    Multileaf collimators (MLCs). (a) Elekta Agility™ multileaf collimator. (Image courtesy of Elekta.)
               (b) Varian Millennium™120-leaf multileaf collimator. (Image courtesy of Varian Medical Sys-
               tems, Inc. All rights reserved.)
               attenuator. Leaf width at isocenter is also 0.5 cm. The Elekta MLCs also employ a
               single-focused design with a rounded leaf end.
                    There is more clearance between the linac treatment head and the patient with
               the Elekta MLC design, which is closer to the target. However, due to beam diver-
               gence this positioning causes greater magnification of MLC width at isocenter, so
               they need to have narrower physical leaf widths in their mounted position because
               they are mounted closer to the target. Most MLCs provide leaf widths that are
               0.5 cm at isocenter.
                    The Varian open gantry linacs retain two sets of jaws, so the MLC is a tertiary
               collimator. A close-up of the Varian Millennium™ MLC is shown in Figure 1.20b.
               This design has 60 pairs of leaves, i.e., 120 leaves in total, which are affixed to the
              machine just beyond the rectangular collimators and in front of the accessory
              holder. Most MLCs employ a single-focus leaf design. This means the leaf sides
              diverge to align with beam divergence in their cross-axis dimension. But because
              the leaves drive horizontally using a threaded screw design, a rounded leaf end is
              used to account for beam divergence in the other dimension. For the millennium,
              MLC leaf width at isocenter is 0.5 cm out to the 20 × 20 cm2 field size, then 1-cm
              leaves from 20 to 40 cm2 field size. Varian also provides a high-definition MLC
              with some central leaf widths being 2.5 mm at isocenter. This option comes with a
              slightly reduced maximum filed size.
                   Some exciting new designs employ dual leaf banks that do not require collima-
              tors. For example, the Varian Halcyon/Ethos linac employs a dual-bank design. The
              leaf banks are offset by half a leaf width, ensuring higher resolution leaf patterns,
              and the upper and lower banks replace the need for collimators. The multileaf colli-
              mators deployed in the ViewRay MRIdian™ MRI-linac has dual banks and a dual-
              focused design to provide alignment with beam divergence in both axes for a very
              sharp beam penumbra.
1.8           Wedges
1.8.1         Physical Wedges
              Wedges are variable-thickness absorbers that are placed in the beam and cause a
              progressive decrease in dose intensity across the beam, resulting in tilted dose pro-
              files. They are designed to give an angled isodose curve at a 10-cm reference depth.
              For example, a set of physical wedges generally allow 15-, 30-, 45-, and 60-degree
              angled isodose curves at a 10-cm depth. Physical wedges are generally made from
              brass, lead, or steel and are used at a distance from the source of about 40 cm. The
              effect of a wedge on the dose profile is shown in Figure 1.21a.
                   The introduction of a wedge produces a reduction in beam transmission (atten-
              uation). This is accounted for by using wedge transmission factors to correct the
              dose delivered when a wedge is introduced into the x-ray beam. Because physical
              wedges attenuate low-energy photons, the beam is slightly harder than the open
Figure 1.21   Different ways of providing a wedge-shaped dose distribution. (a) Physical wedge. (b) Flying
              wedge. (c) Dynamic wedge.
           field. This results in depth dose curves for wedge fields being slightly shallower at
           depth. However, this effect tends to be very small, e.g., less than 2% difference for
           6-MV linacs at a 30-cm depth. This effect does require separate central axis depth
           dose collection and input for modeling in radiation therapy treatment planning
           computers.
Figure 1.22   Electronic portal imaging device (EPID) on a Varian TrueBeam® mounted on a robotic arm so
              that the device can be retracted away during routine treatment or extended out for an MV
              image. (Image courtesy of Varian Medical Systems, Inc. All rights reserved.)
              was to place a film in this position (known as a port film). Most EPIDs consist of an
              amorphous silicon (a-Si) detector array overlaid with a fluorescent layer, usually
              gadolinium oxysulphide (GdOS) or cesium iodide (CsI).
                    Because the EPID image is produced using megavoltage (MV) beam energy,
              the image does not have as much contrast resolution as simulator images (diagnos-
              tic tube). Sufficient bony landmark anatomy may be visible to enable an alignment
              comparison between EPID images and simulator images or, indeed, digitally recon-
              structed radiographs (DRRs), which are images reconstructed from CT data.
Figure 1.23    Cone-beam CT device including kV x-ray tube and flat-panel imager mounted at 90 degrees to
               linac treatment head mounted on an Elekta Versa HD™. (Image courtesy Elekta.)
1.10.2         Halcyon-Ethos
               The Varian Halcyon-Ethos linac consists of a closed-gantry linac, as shown in Fig-
               ure 1.25. It also employs an in-line 6-MV linac in an enclosed cone-beam device.
               While not employing a slip ring, it can rotate several rotations in one direction. Any
               closed-gantry linac has the advantage that it can rotate faster than one revolution
               per minute, which is a safety limit imposed on open-gantry linacs. This linac has a
               fixed target, is flattening filter free, and has a dual-bank multileaf collimator. The
               Ethos version of this machine places an emphasis on being able to replan treatments
               using cone-beam images acquired and then providing an adaptive treatment.
Figure 1.24   Radixact® Tomotherapy linac mounted on a closed gantry using helical delivery and pneu-
              matic- driven binary multileaf collimators. (Image used with permission from Accuray Incorpo-
              rated.)
Figure 1.25   Varian Ethos linac mounted on a closed gantry using dual-bank multileaf collimators. (Image
              courtesy of Varian Medical Systems, Inc. All rights reserved.)
Figure 1.26    External view of a CyberKnife®, which is a linac mounted on a robotic arm that incorporates
               2D stereoscopic x-ray imaging and motion-adaptive treatments. (Image used with permission
               from Accuray Incorporated.)
1.10.3         CyberKnife®
               The CyberKnife® is a linac mounted on a robotic arm that can produce radiotherapy
               x-ray beams from multiple noncoplanar directions (see Figure 1.26). The device
               can provide frameless stereotactic radiosurgery (SRS) using an image correlation
               algorithm for target localization (Adler et al. 1997). It is also used in combination
               with diagnostic x-ray tubes and flat-panel images in order to obtain frameless ste-
               reotactic images. This device was originally designed to compete with the Gamma
               Knife®. The Gamma Knife is an extremely precise multiple-source cobalt-60
               machine that was first introduced in the 1960s. It continues to provide sub-millime-
               ter cranial radiotherapy precision treatment for brain tumors, metastases, and, in
               particular, arteriovenous malformations (AVMs).
                    Due to the mobility provided by its robotic arm, CyberKnife is also being used
               in extra-cranial parts of the body due to its image-guidance capability. It has been
               used for tracking radiotherapy lung treatment by a combination of lung position
               prediction and tracking of in situ markers (Whyte et al. 2003). A review comparing
               three stereotactic radiosurgery systems—including Gamma Knife, CyberKnife, and
               the Novalis radiosurgery linac—has been reported by Andrews et al. (2006).
Figure 1.27   Elekta Unity MRI-linac combining a 7-MV linac with 1.5-T MRI for image guidance. An extra
              mounted body coil is shown which is employed in combination with a gantry-mounted body
              coil to enhance image quality. (Image courtesy of Elekta.)
              essence, MR image guidance enables high soft tissue contrast, no increased second-
              ary cancer risk from imaging dose, and intra-fraction images. There may also be
              potential for biological analysis of tumor response using on-line perfusion imaging.
                   The ViewRay MRIdian 0.35-T MRI was originally combined with cobalt-60
              and treated the first patients in 2014. These devices have been modified to MRI-lin-
              acs and have provided patient treatments since 2017 (Mutic and Demsey 2014).
              Another MRI-linac, the Elekta Unity 1.5-T 7-MV linac, provides a high magnetic
              field strength (Lagendijk et al. 2008; Langendijk et al. 2014). The Elekta Unity
              device also started patient treatments in 2017 and is shown in Figure 1.27. Both lin-
              acs rotate around a fixed MRI gantry, so the x-ray treatment beam delivery is trans-
              verse to the magnetic field. This introduces some interactions between the Lorentz
              force of the B field and Compton-generated electrons, leading to interesting elec-
              tron return effects (Oborn et al. 2010). There are currently about 70 MRI-linacs
              installed, with the number of machines increasing rapidly. A recent review of MRI-
              linacs and their real-time image guidance capabilities is given by Lagendijk et al.
              (2020).
                   This chapter outlined the basic physics principles of linac operational design.
              This chapter has also provided a taste of the exciting proliferation of innovative
              design for new linac beam delivery systems. Subsequent chapters will identify and
              explain these key innovations in more detail.
References
          Adler, J. R. Jr., S. D. Chang, M. J. Murphy, J. Doty, P. Geis, and S. L. Hancock (1997). “The
                CyberKnife: a frameless robotic system for radiotherapy.” Stereotact. Funct. Neurosurg.
                69(1–4 Pt2):124–28.
          Andrews, D. W., G. Bednarz, J. J.Evans, and B. Downes (2006). “A review of 3 current radiosur-
                gery systems.” Surg. Neurol. 66(6):559–64.
          Attun, R., D. A. Jaffray, M.B. Barton, F. Bray, et al. (2015). “Expanding global access to radio-
                therapy.” Lancet Oncol. 16:1193–1224.
          Bekelman, J. E., A. Denicoff, and J. Buchsbaum (2018). “Randomized Trials of Proton Therapy:
                Why they are at risk, proposed solutions, and implications for evaluating advanced technolo-
                gies to diagnose and treat cancer.” J. Clin. Oncol. 38(24):2461–66.
          Einstein, A. (1905). “On the electrodynamics of moving bodies.” Ann. Physik. 17:891.
          Greene, D. (1986). “Linear Accelerators for Radiation Therapy.” Medical Physics Handbook 17.
                Bristol, UK: Adam Hilger.
          Ishikawa, H., H. Tsuji, T. Kamada, N. Hirasawa, T. Yanagi, J. E. Mizoe, K. Akakura, H. Suzuki,
                J. Shimazaki, and H. Tsuijii (2006). “Risk factors of late rectal bleeding after carbon ion
                therapy for prostate cancer.” Int. J. Radiat. Oncol. Biol. Phys. 66(4):1084–91.
          Johns, H. E. and J. R. Cunningham (1983). The Physics of Radiology, 4th Ed. Springfield, Illinois:
                Charles C. Thomas.
          Johns, H. E. and J. R. Cunningham (1959). “A precision cobalt-60 unit for fixed and rotational
                therapy.” Am. J. Roentgenol. 81:4–6.
          Johns, H. E., S. O. Fedoru, R. O. Kornelson, E. R. Epp, and E. K. Darby (1952a). “Depth dose
                data, 150 kVp to 400 kVp.” Br. J. Radiol. 25:542–49.
          Johns, H. E., E. R. Epp, D. V. Cormack, and S. O. Fedoruk (1952b). “1000 Curie cobalt units for
                radiation therapy.” Br. J. Radiol. 25:296.
          Karzmark, C. J. and R. J. Morton (2017). A Primer on the Theory and Operation of Linear Accel-
                erators in Radiation Therapy, 3rd Ed. Madison, WI: Medical Physics Publishing.
          Kijewski, P. K., L. M. Chin, and B. E. Bjarngard (1978). “Wedge-shaped dose distributions by
                computer-controlled collimator motion.” Med. Phys. 5:426–29.
          Krieger, H. and W. Petzold (1989). Strahlenphysik, Dosimetrie und Strahlenschutz. Stuttgart, Ger-
                many: B.G. Teubner.
          Lagendijk, J. and B. Raaymakers (2008). A MRI-linac integration. Radiother. Oncol. 86:25–29.
          Lagendijk J., B. Raaymakers, and M. van Vulpen (2014). “The Magnetic Resonance Imaging-
                Linac.” Semin. Radiat. Oncol. 24:207–09.
          Lagendijk, J., B. Raaymakers, R. Tijssen, and B. van Asselen (2020). “Real-Time Image Guidance
                with Magnetic Resonance.” In The Modern Technology of Radiation Oncology: A Compen-
                dium for Medical Physicists and Radiation Oncologists, Vol 4. J. Van Dyk, Ed. Madison, WI:
                Medical Physics Publishing.
          Levitt, D. D., M. Martin, J. H. Moeller, and W. L. Lee (1990). “Dynamic wedge field technique
                through computer-controlled collimator motion and dose delivery.” Med. Phys. 17:87–91.
          Locke, J., S. Karimpour, G. Young, M. A. Locket, and C. A. Perez (2001). “Radiotherapy for Epi-
                thelial Skin Cancer.” Int. J. Radiol. Oncol. Biol. Phys. 51(3):748–55.
          Mackie, T. R., T. Holmes, S. Swerdloff, P. Reckwerdt, J. O. Deasy, J. Yang, B. Paliwal, and
                T. Kinsella (1993). “Tomotherapy: a new concept for the delivery of dynamic conformal
                radiotherapy.” Med. Phys. 20(6):1709–19.
          Mutic, S. and J. F. Dempsey (2014). “The ViewRay system: magnetic resonance-guided and con-
                trolled radiotherapy.” Semin. Radiat. Oncol. 196–99.
          Oborn, B., P. E. Metcalfe, M. J. Butson, and A. B. Rosenfeld (2010). “Monte Carlo characteriza-
                tion of skin doses in 6MV transverse field MRI-linac systems: Effect of field size, surface
                orientation, magnetic field strength, and exit bolus.” Med. Phys. 5208–17.
          Particle Therapy Co-operative Group (PTCOG) (2021). Available at https://www.ptcog.ch/
                index.php/facilities-in-operation.
          Podgorsak, E. B., J. Van Dyk, and P. Metcalfe (1999). “Linear accelerators.” In The Modern Tech-
                nology of Radiation Oncology. J. Van Dyk, Ed. Madison, WI: Medical Physics Publishing.
          Shipley, W. U., L. J. Verhey, J. E. Munzenrider, H. D. Suit, M. M. Urie, et al. (1995). “Advanced
                prostate cancer: the results of a randomized comparative trial of high dose irradiation boost-
                ing with conformal protons compared with conventional dose irradiation using photons
                alone.” Int. J. Radiat. Oncol. Biol. Phys. 32(1):3–12.
          Thwaites, D. I. and J. B. Tuohy (2007). “Back to the future: the history and development of the
                first clinical linear accelerator.” Phys. Med Biol. 51(13):R343–62.
          Whyte, R. I., R. Crownover, M. J. Murphy, D. P. Martin, T. W. Rice, et al. (2003). “Stereotactic
                radiosurgery for lung tumors: preliminary report of a phase I trial.” Ann. Thorac. Surg.
                75(4):1097–101.
Questions
      1.1   Explain the main difference between traveling and standing waveguide
            accelerators. (See Section 1.5.)
      1.2   Explain the main difference between a magnetron and a Klystron. (See Sec-
            tion 1.5.)
      1.3   If a 6-MeV electron beam needs to travel in a radius of curvature of 15 cm to
            traverse around the bend magnet, what magnitude of magnetic field is
            required if one assumes a constant magnetic field? (See Section 1.6.)
      1.4   Describe the main efficiency gain from multileaf collimators (MLCs) that
            lead to them replacing shielding blocks. (See Section 1.8.)
      1.5   What are the practical advantages of using dynamic wedges instead of con-
            ventional wedges in combination with MLCs. Describe another way of using
            MLCs that negates the requirement for wedges. (See Section 1.8.)