Optia Exchange Kit

The document provides a detailed description of the Spectra Optia Exchange Set, including its components and their functions. Each part is numbered and described, covering lines for blood, saline, anticoagulants, and various sensors. The document serves as a reference for understanding the setup and operation of the exchange set in a medical context.

Uploaded by

anupa000
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
175 views3 pages

Optia Exchange Kit

The document provides a detailed description of the Spectra Optia Exchange Set, including its components and their functions. Each part is numbered and described, covering lines for blood, saline, anticoagulants, and various sensors. The document serves as a reference for understanding the setup and operation of the exchange set in a medical context.

Uploaded by

anupa000
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Put cursor on number

or click for description


®
Spectra Optia Exchange Set
Catalog No. 12220

20

19

21

9 23

22 18

14 17
15

13
10

1 2 11
3 4

5
12
6
16

1 Inlet line 9 Saline line (green) 17 Channel


2 Inlet line manifold 10 Inlet line trap 18 Connector
3 Inlet saline line clamp (red) 11 Inlet pressure sensor diaphragm 19 Vent bag
4 AC check valve 12 Centrifuge pressure sensor diaphragm 20 Remove bag
5 Return line 13 Reservoir 21 Remove line
6 Return line manifold 14 Return pressure sensor diaphragm 22 Extra remove line
7 Return saline line clamp (blue) 15 Plasma pressure sensor diaphragm 23 Replace line
8 Anticoagulant (AC) line (orange) 16 Centrifuge loop

©2015 Terumo BCT, Inc. Part No. 306611-010


Spectra Optia® Exchange Set
Catalog No. 12220
Part Descriptions

1. Inlet line: Used to carry anticoagulated blood from the patient to the channel. Includes:
 Luer connector (male): Connects the inlet line to the patient inlet access.
 Inlet line clamp (red): Used to clamp the inlet line.

2. Inlet line manifold: Provides a connection point for the inlet line, the AC line, and the inlet saline line.
Includes an injection port.

3. Inlet saline line clamp (red): Used to close the inlet saline line or to control the flow of saline to the
patient.

4. AC check valve: Allows the system to pump AC through the tubing but prevents the free flow of AC.

5. Return line: Used to carry fluid from the reservoir to the patient. Includes:
 Luer connector (male): Connects the return line to the patient return access.
 Return line clamp (blue): Used to clamp the return line.

6. Return line manifold: Provides a connection point for the return line and the return saline line.
Includes:
 Injection port (needleless): Provides an access site for the infusion of medication or the drawing
of samples.

7. Return saline line clamp (blue): Used to close the return saline line or to control the flow of saline to
the patient.

8. Anticoagulant (AC) line (orange): Used to carry AC from the AC container to the inlet line manifold.
Includes:
 Luer connector: Unique connector used to connect an ACD-A bag that has a compatible luer
connector.
 Sterile barrier filter (0.2 microns): Prevents air from entering the line.

9. Saline line (green): Used to carry saline from the saline container to the inlet line manifold and to
the return line manifold. Includes:
 Spike (green): Used to spike the saline container.

10. Inlet line trap: Prevents larger solid cellular matter from entering the channel.

11. Inlet pressure sensor diaphragm: Attaches to the inlet pressure sensor on the front panel to monitor
pressure in the inlet line.

12. Centrifuge pressure sensor diaphragm: Attaches to the centrifuge pressure sensor on the front
panel to monitor pressure in the channel and the lines in the centrifuge.

©2015 Terumo BCT, Inc. Part No. 306611-010


Spectra Optia Exchange Set Part Descriptions

13. Reservoir: Used to hold the fluid that is returned to the patient. Includes a 200-micron filter.

14. Return pressure sensor diaphragm: Attaches to the return pressure sensor on the front panel to
monitor pressure in the return line.

15. Plasma pressure sensor diaphragm: Attaches to the plasma pressure sensor on the front panel to
monitor pressure in the plasma line.

16. Centrifuge loop: Consists of lines that are loaded in the centrifuge. Includes:
 Four-lumen tubing: Used to carry fluid into and out of the channel.
 Sleeves: Used to reinforce the tubing.
 Collars: Used to secure the ends of the loop in the centrifuge.
 Bearings: Used to connect the loop to the centrifuge.

17. Channel: Used during centrifugation to separate the patient’s blood into cellular components.

18. Connector: Secures the lines that carry the separated components out of the channel.

19. Vent bag: Holds air that is displaced from the system.

20. Remove bag: Holds removed blood components.

21. Remove line: Used to carry removed blood components to the remove bag.

22. Extra remove line: Used to carry removed blood components to an extra remove bag or to an
extension line, if connected. Includes:
 Luer connector (female): Connects the extra remove line to an extra remove bag or to an
extension line.
 Clamp (yellow): Used to clamp the extra remove line.

23. Replace line: Used to carry replacement fluid from the replacement fluid container to the reservoir.
Includes:
 Spikes (white): Used to spike the replacement fluid container.
 Clamps (white): Used to clamp the lines to the spikes for the replacement fluid.
 Luer connector (male/female): Connects the replace line to a filter or blood warmer tubing, if
used.

©2015 Terumo BCT, Inc. Part No. 306611-010