Blood Fluid Warmer
Blood Fluid Warmer
FW400/FW401/REF FW403
Blood/Fluid Warmer
Service Manual
WARNING
CAUTION
1
Service Manual
FW400 Series Blood/Fluid Warmer
G
H
I
A F
B
C D
E
$ /('7HPSHUDWXUH'LVSOD\ ( %XEEOH7UDS+ROGHU
%OXH*XLGH5DLO/RFDWLRQ ) 3ROH&ODPS
%
,QGLFDWRU * 'HWDFKDEOH3RZHU&RUG
& ([LW3RUW5HOLHI + /LJKWHG2Q2II&LUFXLW%UHDNHU
' :DUPLQJ6HW&DVVHWWH6ORW , 7HVW6ZLWFK$FFHVV6FUHZ
2
Service Manual
FW400 Series Blood/Fluid Warmer
3.2 Disposable Warming Sets Each Warming Set consists of a flexible cassette (for
(see figure 2) insertion into the warmer) and a bubble trap with a manual air
eliminator. They are designed for use below 300 mm Hg
The Medi-Temp II FW400 Series Blood/Fluid Warmer is pressure with hospital standard intravenous (IV)
designed to be used with Gaymar D25000 Series Blood/Fluid administration sets and venipuncture sets utilizing luer type
Warming Sets only. The following Blood/Fluid Warming Sets connectors. Each Warming Set also has input and output
are available for use at this time. Check with your local dealer pinch clamps. See figure 2.
for more details. The Standard Warming Set is designed for use at 300 ml/min
or lower flow rates. The Standard Warming Set with
Extensions provides the user with added patient lead length
(0.6 m; 2 ft). The priming volume of the Standard Warming
Set is 39 ml. The priming volume of the Standard Warming
Set with Extensions is 48 ml.
The Pediatric Warming Set has a lower priming volume of
30 ml. Its patient connection lead utilizes thicker walled
G
tubing to provide added insulation at the lower pediatric
warming flow rates.
The High Flow Warming Set uses leads having a larger inner
A diameter. This reduces restriction during circumstances
requiring rapid infusion. The priming volume of the High Flow
Warming Set is 79 ml. The High Flow Warming Set also
D B incorporates a 200 micron filter and larger bubble trap.
Description Catalog N o.
C
Standard D25340CE
F
Standard with Extensions D25310CE
D
Standard with Extensions (60") D25315CE
E Pediatric D25320CE
Hi g h F l o w D25330CE
3
Service Manual
FW400 Series Blood/Fluid Warmer
3.3 Theory of Operation controller uses this RTD information to control the fluid
(refer to section 10, Drawings/Parts List temperature to a nonadjustable set point of 43.0°C by
applying power to the heaters via an optocoupler and triac on
and section 11, Wiring Diagram) the PC board. The sensed temperature is displayed on the
The Medi-Temp II FW400 series Blood/Fluid Warmer temperature controller's display (which is viewable from the
electrically consists of: front of the Warmer). The controller activates audible alarm
circuitry on the PC board and also provides a visual display
power entry module and detachable power cord of the temperature if the RTD sensed temperature goes above
lighted on/off circuit breaker 45°C or below 34°C.
heater plate assembly
RTD sensor Two bimetallic thermostats mounted on the heater plate
dual heaters assembly will also interrupt power to the heaters in the event
temperature controller of a Warmer failure. See section 5.4.
two bimetallic thermostats
PC board containing audible alarm circuitry and
heater/controller interface circuitry.
The Warmer connects to line voltage via the power inlet and a
detachable power cord. The lighted circuit breaker provides
overcurrent protection as well as on/off function. An RTD
sensor on the inside surface of the heater plate assembly
senses blood/fluid temperature when a Warming Set
Cassette is inserted into the Warmer. The temperature
IPX4
:HLJKW SRXQGV NJ
6L]H [[ PP[PP[PP
4
Service Manual
FW400 Series Blood/Fluid Warmer
Safety Thermostats
5.1 Electronic System Self Check
®
The Medi-Temp II FW400 series Blood/Fluid Warmer Cold fluid through the Warming Set can be used to reset the
noninvasively senses fluid temperature on the surface of the thermostats. The unit must be turned off, then back on to
Warming Set Cassette. At steady state, the Warmer indicates reset the alarm circuits.
approximately 43.0°C on its display. Output fluid temperature
will be 36°C43°C for 10°C input fluid temperature. Fluid °c
5.5 Control Alarm - LO
temperature is affected by fluid flow rate as shown in figure 3 ®
below.
7\SLFDO)OXLG7HPSHUDWXUH If the sensed fluid temperature drops below 34ºC, the
):6HULHV Warmer will activate an audible alarm, and the display will
alternately indicate the word "LO" and the temperature.
7HPSHUDWXUH &
6.0 Cleaning
CAUTION
Do not immerse the Warmer in cleaning or
disinfectant solutions. Do not submerge it in liquids;
)ORZ POPLQ it is fluid-resistant, not fluid-proof.
Figure 3Typical fluid temperature vs. flow values Immersion could result in malfunction or
equipment damage.
5
Service Manual
FW400 Series Blood/Fluid Warmer
7.0 Functional Check and Safety Inspection Clean and dry compressed air source
To assure optimum performance, dependability, and safety, Current Leakage/Ground Resistance Tester
the following should be performed every twelve months or as
specified in the facility's preventive maintenance program.
AC Voltmeter
The Medi-Temp II FW400 series Blood/Fluid Warmer is rated
When performing the following procedures, pay particular IPX4splashproof. To insure against ingress of liquids into
attention to test setups. Any deviation from the setups, the machine, it is important that the labels, covers, and
procedures, or test equipment may result in incorrect or screws be kept in good condition. Replace labels that are
misleading results. The following test procedures will peeling, ripped, or illegible.
verify proper operation of the FW400 series Medi-Temp II
Blood/Fluid Warmer. 7.1 Physical Inspection
WARNING Check that the following items are in good condition and
secure. Replace if necessary.
Always perform the Functional Check and Safety Screws
Inspection after making repairs and before returning
the Warmer to patient use. Labels, if peeling or missing
Failure to perform the Functional Check and Cover on the on/off circuit breaker
Safety Inspection could result in death or serious
injury, equipment damage, or malfunction of the Bubble trap bracket
Warmer. Pole clamp
Power entry module
Inspection Form Power cord and plug
An Inspection Form (figure 5, page 9) is provided to facilitate
and document the inspection process. 7.2 Calibration / Overtemp Check
Equipment or tools required The Medi-Temp II Blood/Fluid Warmer's electronic controller
and RTD sensor system is calibrated at a flow rate of 100 ml
The following test equipment (or equivalent) is required to per minute with a set point of 43.0°C. Verify correct output
perform the Functional Check and Safety Inspection: temperature as follows:
Gaymar Standard Blood/Fluid Warming Set, [CALIBRATION CHECK]
model D25340CE
1. Connect test setup as in figure 4. Use a standard
Temperature Measurement Device (see figure 4, page 7) Blood/Fluid Warming Set (model D25340CE) with
Gaymar FWT1 Temperature Sensor connected to
Temperature sensor (Gaymar catalog FWT1) and an
the output of the bubble trap. An alternate method is
ohmmeter (accuracy,1.5% of reading; maximum
to put a hole in the top of the bubble trap and insert
excitation current of 100 µA) and the resistance/
a Temperature Measurement Device waterproof
temperature chart (see figure 10);
sensor into the bubble trap (to the bottom). This
or,
measurement method will require sealing the
Waterproof temperature sensor/meter with an
bubble trap hole around the sensor. See figure 4.
accuracy of ± 0.3°C across the range 30°C to 60°C
and a thermal time constant of 2 seconds 2. Connect the FWT2 Flowmeter or flow measurement/
or faster. control device to the output of the FWT1
Temperature Sensor; or, if using the sensor in the
Flow Measurement Device (see figure 4)
bubble trap method, connect the FWT2 Flowmeter
Fluid source with calibrated flow meter with a or flow measurement/control device directly to the
minimum accuracy of ± 12 ml/minute (Gaymar output of the D25340CE Warming Set. Insure room
FWT2 Flowmeter); ambient temperature is 20° to 24°C.
or,
3. Flow room temperature water at 100 ml per minute
Fluid source with adjustable flow control. Use a
(cc/min) through the Warmer. Verify bubble trap is
stop watch and graduated cylinder to adjust
maintained 1/2 to 2/3 full.
flow rate.
6
Service Manual
FW400 Series Blood/Fluid Warmer
7
Service Manual
FW400 Series Blood/Fluid Warmer
4. Turn Warmer on and allow to run for 10 minutes. 7.3 Controller LO Alarm Check
5. After 10 minutes, verify output water temperature is 1. Turn Warmer on without a Warming Set in place.
43.0°C ±1.1°C. If using the FWT1 Temperature Allow display to stabilize at approximately 43.0°C.
Sensor and an ohmmeter, verify thermistor
temperature sensor is within the range of 1019 to 2. Using a compressed air nozzle pointed into the
1112 ohms (see resistance/temperature chart, disposable slot at the exit port relief (see figure 1),
figure 10). If not within this range, recheck the test flow compressed air. This will cool the sensor and
setup. See section 8.6. should reduce the display temperature to below
34°C. Insure the audible alarm activates and the
[OVERTEMP CHECK] word LO alternately flashes on the display when
below 34°C. If it does not, see section 8.2.
6. Verify the flow is still 100 ml per minute. Remove
the test switch access screw (item I, figure 1).
Using an electrically nonconductive tool, push and 7.4 Ground Resistance Test
hold switch S2. This will force the heater on. The
display temperature will rise. Use a current leakage/ground resistance tester to
measure the resistance between the ground pin on
7. Verify that the audible alarm sounds and the word the power plug and either of the two stainless
hI is alternately flashed on the display when the screws on the bottom side of the Warmer (item 35,
display temperature surpasses 45°C. If it does not, figure 8B). The resistance value should not exceed
see section 8.2. 0.50 ohms. If it does, see section 8.8.
8. Continue to hold the S2 switch. Using the
temperature measurement equipment, read the 7.5 Current Leakage Test
output resistance/temperature at the instant the
display blanks with the alarm still sounding. This is Measure the maximum earth leakage current
the trip point of the S3 over temperature safety (ground open). Measure at all combinations of
thermostat. The thermostat trip point (alarm
line polarity
sounding, display off) should be 48.6°C to 58.4°C,
neutral open/closed
or within a FWT1 Temperature Sensor range of 856
to 594 ohms. If it is not, replace it. See section 9.4. Measure with the heater on in the neutral closed
test. If the leakage values exceed the values listed
NOTE: The Warmer display value should not be
on the Inspection Form (figure 5, page 9), see
used for this test.
section 8.9.
NOTE: The S4 thermostat is designed to open at
temperatures above the S3 thermostat.
It is a secondary fire protection device.; it
is not a patient safety device. When the
S4 thermostat trips, the entire unit
(display, audible, and heater) turns off.
The S4 thermostat is tested at Gaymar
before each unit is shipped. Gaymar does
not recommend testing the S4 thermostat
in the field.
8
Service Manual
FW400 Series Blood/Fluid Warmer
Item O b s e r v a t io n Results
L ab e ls Yes/No
Over temp Alarm Check Does alarm activate above 45°C? Yes/No
S3 Over Temperature Safety Does output temperature remain within 48.6°C Yes/No
Thermostat Check to 58.4°C?
7.3 Controller "LO" Alarm Check Does alarm activate below 34°C? Yes/No
9
Service Manual
FW400 Series Blood/Fluid Warmer
8.0 Troubleshooting
$XGLEOHLV $XGLEOHLV
QRWVRXQGLQJ VRXQGLQJ
)DXOW\ )DXOW\
7ULDFRU &RQWUROOHURU
)DXOW\ 57'WHVW KHDWHUWHVW )DXOW\
RSWRFRXSOHU FRQWUROOHU 7ULSSHG
LQWHUQDO FRQWLQXLW\ UHVLVWDQFH FLUFXLW
RU3&ERDUG SURJUDPPLQJ WKHUPRVWDW
FRQQHFWLRQ ZLWK ZLWK EUHDNHU
IDXOW\ IDXOW\
RKPPHWHU RKPPHWHU
5HSODFH 5HSODFH
5HSODFH &RUUHFW &RROXQLW
57'KHDWHU FLUFXLW
FRQWUROOHU FRQQHFWLRQ WRUHVHW
DVVHPEO\ EUHDNHU
6HH 6HH WKHUPRVWDWV
6HH 6HH
&DOLEUDWH 3HUIRUPRYHU
:DUPHU WHPSHUDWXUHVDIHW\
RXWSXW WKHUPRVWDWFKHFN
6HH 6HH
5HSODFHLI
QHFHVVDU\
1HLWKHUWKH
%RWKWKHDXGLEOH
DXGLEOH
DQGWKHGLVSOD\
QRUWKHGLVSOD\
WXUQRIIEHORZ
WXUQRIIEHORZ
&
&
6LVIDXOW\
6LVWULSSLQJ 5HSODFH6
WRRORZ EHIRUH WKHUPRVWDW
6GRHV 6HH
10
Service Manual
FW400 Series Blood/Fluid Warmer
)ORZUDWH
)DXOW\ )DXOW\6
JUHDWHUWKDQ
WKHUPRVWDW WKHUPRVWDW
POPLQRI
UHOD\ WRRORZ
&IOXLG
&RROXQLW
WRUHVHW
WKHUPRVWDW )DXOW\
5HGXFH 5HSODFH )DXOW\ )DXOW\
LQWHUQDO
IORZUDWH SRZHUFRUG WULDF KHDWHU
FRQQHFWLRQ
8QSOXJ
7LJKWHQ KHDWHU V
JURXQGLQJ 5HWHVW
VFUHZ FXUUHQW
OHDNDJH
5HSODFH3&
ERDUG
DVVHPEO\
6HH
3HUIRUPFRQWURO
WHPSHUDWXUH
FKHFN
6HH
5HSODFH6
WKHUPRVWDW 3HUIRUPRYHU
6HH WHPSHUDWXUHVDIHW\
WKHUPRVWDWFKHFN
6HH
11
Service Manual
FW400 Series Blood/Fluid Warmer
The Warmer is designed to be fluid- 10. Place a thin bead of RTV sealant around bezel of
resistant by utilizing sealants and gaskets. new controller, then insert controller back into
Inspect gaskets and labels during housing.
reassembly. Replace any that are 11. Reconnect PC board cable and RTD wires to
damaged to maintain this fluid resistance. controller terminal block.
Failure to retain this fluid resistance
could result in Warmer malfunction. NOTE: Route the cable and wires as originally
found.
12
Service Manual
FW400 Series Blood/Fluid Warmer
12. Calibrate the Warmer output. See section 9.6, 11. Replace top cover and 19 screws (use the 17
Calibrating the Warmer Output. black screws around the perimeter). Insure the
frame O ring is in place.
13. Inspect/reinstall controller gasket. Replace if
damaged. 12. Perform Functional Check and Safety Inspection.
See section 7.
14. Replace controller retainer and screws.
13. Clean off any adhesive left on the cover from the
15. Replace top cover and 19 screws (use the 17
old label. Install new top label. Use RTV sealant
black screws around the perimeter). Insure the
on the threads of two screws (item 35, fig. 8B).
frame O ring is in place.
16. Perform Functional Check and Safety Inspection.
See section 7. 9.4 Replacing Thermostats
17. Clean off any adhesive left on the frame from the To replace either of the thermostats (items 16 and 17, figures
old label. Install new front label. 8A/8B), you will need to replace the top label (item 7).
Replace the frame O ring (item 27) if necessary. Use RTV
18. Clean off any adhesive left on the cover from the sealant (item 6) as described.
old label. Install new top label. Use RTV sealant
on the threads of two screws (item 35, fig. 8B). 1. Unplug Warmer.
2. Remove top label.
9.3 Replacing the PC Board Assembly 3. Remove 19 cross-recess screws from top cover.
Do not remove screws from the bottom cover.
To replace the PC board assembly (items 23, 24, or 25;
figures 8A/8B), you will need to replace the top label (item 7) 4. Remove top cover.
and use heat sink compound. Replace the frame O ring (item 5. Remove the two thermostat mounting screws
27) if necessary. Use RTV sealant (item 6) as described. and lockwashers. Remove thermostat.
1. Unplug Warmer. 6. Disconnect thermostat wires from terminal block
2. Remove top label. of PC board assembly.
3. Remove 19 cross-recess screws from top cover. 7. Replace thermostat, two thermostat mounting
Do not remove screws from the bottom cover. screws, and lockwashers. Do not use heat sink
compound.
4. Remove top cover.
8. Reconnect thermostat wires to PC board
5. Disconnect all wires from terminal block of PC assembly terminal block. Refer to figure 8B and
board. Disconnect wires from circuit breaker. wiring diagram (figure 9) for correct wiring. Be
6. Remove the four screws which mount the PC sure to tape the wires as shown in figure 8B.
board assembly to the lower plate. Remove the 9. Replace top cover and 19 screws (use the 17
screw and lockwasher which mounts the triac. black screws around the perimeter). Insure the
7. Remove the PC board. Do not remove the PC frame O ring is in place.
board insulation from the unit. 10. Perform Functional Check and Safety Inspection.
8. Apply heat sink compound to the new triac heat See section 7.
sink surface. Install triac using screw and 11. Clean off any adhesive left on the cover from the
lockwasher. Install new PC board assembly using old label. Install new top label. Use RTV sealant
four screws. on the threads of two screws (item 35, fig. 8B).
9. Reconnect all wires to PC board assembly
terminal block. Refer to figure 8B and the wiring
9.5 Replacing the RTD/Heater Assembly
diagram (figure 9) for correct wiring.
If the RTD/Heater Assembly requires repair or replacement,
10. Reconnect wires to circuit breaker (brown wire
please return the Warmer to your local dealer or Gaymar
on top).
Industries. See Return Authorization (section 2.4, page 1).
13
Service Manual
FW400 Series Blood/Fluid Warmer
&DOLEUDWLRQ
2XWSXW 5HTXLUHG
2IIVHW
7HPSHUDWXUH 2XWSXW &DOFXODWH
Figure 6Installing Push-button keys WR8VH
PHDVXUHG 7HPSHUDWXUH
LQ&RQWUROOHU
& & >@
& & >@
9.6 Calibrating the Warmer Output & & >@
To calibrate the Warmer output, you will need push-button & & >@
keys (Gaymar P/N 10109-000; see figure 6 above).
Figure 7Sample Offset Calculations
1. Connect Warmer test setup as shown in figure 4.
Use a Standard Blood/Fluid Warming Set (model 8. With the front label, controller retainer, and controller
D25340CE) with an FWT1 Temperature Sensor gasket removed from the Warmer, insert the
connected to the output of the Warming Set. An push-button keys onto the face of the controller
alternate method is to put a hole in the top of the (set button on the left).
bubble trap and insert a temperature measurement
device sensor into the bubble trap (to the bottom). NOTE: The following three steps allow access to the
This measurement method will require sealing the calibration offset menu location. These steps may be
bubble trap hole around the sensor. repeated at any time to view the current offset
condition or make an adjustment in it.
2. Connect the FWT2 Flowmeter or flow measurement/
control device to the output of the FWT1 9. Depress both the up arrow and down arrow
Temperature Sensor; or, if using the sensor in the buttons simultaneously until the letters Aut appear
bubble trap method, connect the FWT2 Flowmeter on the controller display. Depress the up arrow or
or flow measurement/control device directly to the down arrow key until the letters Pid appear.
output of the D25340CE Warming Set. Insure 10. Hold down the set button and depress either the
ambient room temperature is 20° to 24°C. up arrow or down arrow button until yes
3. Power up Warmer. appears in the display, then release the set button.
11. Depress the down arrow button to index the display
DANGER to indicate CAL.
12. Press the set button and hold it while pressing the
Risk of electrical shock when parts are
up arrow or down arrow button to enter the offset
electrified.
value calculated from step 7.
14
Service Manual
FW400 Series Blood/Fluid Warmer
NOTE: The next step is used to return to the current 15. Perform Functional Check and Safety Inspection.
measured temperature from any place in the menu. See section 7.
13. Depress both the up arrow and down arrow 16. Clean off any adhesive left on the cover from the old
buttons for three seconds until the measured label. Install new top label. Use RTV sealant on the
temperature appears in the display. threads of two screws (item 35, fig. 8B).
14. Repeat step 6 to verify correct output temperature.
Repeat steps 7 through 13 as needed. 9.8 Replacing the Circuit Breaker
15. Inspect/reinstall controller gasket. Replace if To replace the circuit breaker (items 1, 2, or 3, figures
damaged. 8A/8B), you will need to replace the top label (item 7), on/off
label (item 8), and use RTV sealant (item 6). Replace circuit
16. Replace controller retainer and screws.
breaker gasket (item 4), circuit breaker cover (item 5), and
17. Perform Functional Check and Safety Inspection. frame O ring (item 27) if necessary.
See section 7.
1. Unplug Warmer.
18. Clean off any adhesive left on the frame from the
old label. Install new front label. 2. Remove top label.
3. Remove 19 cross-recess screws from top cover.
9.7 Replacing the Power Inlet Assembly
4. Remove top cover.
To replace the power inlet assembly (item 12, figures 8A/8B),
5. Disconnect the wires from circuit breaker.
you will need to replace the top label (item 7) and use RTV
sealant (item 6). Replace the frame O ring (item 27) as 6. Remove on/off label.
necessary.
7. Remove two cross-recess retainer mounting
1. Unplug Warmer. screws and remove retainer.
2. Remove top label. 8. Remove cover.
3. Remove 19 cross-recess screws from top cover. 9. Remove gasket.
4. Remove top cover. 10. Remove circuit breaker from frame.
5. Disconnect blue and brown wires from circuit 11. Apply RTV to underside of circuit breaker lip and
breaker. install new circuit breaker into frame. Insure circuit
6. Remove the nut and lockwasher from the ground breaker is positioned so that the off position is
terminal and disconnect the ground wire. closest to corner of frame.
7. Remove the two cross-recess mounting screws 12. Reinstall gasket, then cover.
holding the power inlet assembly.
13. Replace retainer and screws.
8. Remove assembly from Warmer.
14. Reconnect wires to circuit breaker, so that the
9. Scrape away old RTV sealing material from frame. brown wires are on top, blue wires are on bottom,
10. Insert new power inlet assembly through gasket into and the wires from the power inlet are connected to
frame after applying RTV to frame hole. the terminals on the circuit breaker labelled line.
11. Replace screws. 15. Replace top cover and 19 screws (use the 17 black
screws around the perimeter). Insure the frame O
12. Reconnect ground wire to ground terminal. Place ring is in place.
lockwasher on terminal first, then ground wire, then
second lockwasher, and then the nut. Insure nut is 16. Perform Functional Check and Safety Inspection.
tight. See section 7.
13. Reconnect wires to circuit breaker (brown wire on 17. Clean off any adhesive left on the cover from the old
top). label. Install new top label. Use RTV sealant on the
threads of two screws (item 35, fig. 8B).
14. Replace top cover and 19 screws (use the 17 black
screws around the perimeter). Insure frame O ring 18. Clean off any adhesive left on the cover from the old
is in place. label. Install new on/off label.
15
Service Manual
FW400 Series Blood/Fluid Warmer
10.0 Drawings/Parts List NOTE: Most replacement tasks will require more than one of the parts
in figures 8A/8B. Refer to section 9 before ordering parts.
1 Circuit breaker (FW400, 120V) 90961- 013 28 Top cover kit 78130- 000
2 Circuit breaker (FW401, 100V) 90961- 013 Bottom cover kit (FW400) 78166- 000
3 Circuit breaker (FW403, 220V–240V) 90961- 014 29 Bottom cover kit (FW401) 78169- 000
4 Circuit breaker gasket 10086- 000 Bottom cover kit (FW403) 78167- 000
9 Bubble trap bracket 10044- 000 Power cord (U nited Kingdom) PC002
24 PC board assembly (FW401, 100V) 1 0 1 8 5 - 0 01 42 Push-button keys for controller (see fig. 6) 1 0 1 0 9 - 0 00
27 Frame O-ring 09975- 000 45 Ver tical mounting kit (not shown) 7 8 2 6 4 - 0 00
* Do not attempt to replace the RTD/heater assembly or heater plate O ring. See 9.5, page 13.
16
Service Manual
FW400 Series Blood/Fluid Warmer
Power Cords
(Item 32)
17
Service Manual
FW400 Series Blood/Fluid Warmer
11.0 Wiring
Diagram
ATTENTION
Static-Sensitive
Devices
18
Service Manual
FW400 Series Blood/Fluid Warmer
FW303
&
FW403
FW300A/
FW400
&
FW301/
FW401
19
Service Manual
FW400 Series Blood/Fluid Warmer
& RKPV & RKPV & RKPV & RKPV & RKPV & RKPV & RKPV
Figure 10
Resistance/
Temperature
Chart
20
Service Manual
FW400 Series Blood/Fluid Warmer
21
Service Manual
FW400 Series Blood/Fluid Warmer