Oscilloscope
Oscilloscope
The Braun tube, forerunner of the CRT, was known in 1897, and in 1899 Jonathan Zenneck equipped it
with beam-forming plates and a magnetic field for deflecting the trace, and this formed the basis of the
CRT.[4] Early CRTs had been applied experimentally to laboratory measurements as early as the 1920s,
but suffered from poor stability of the vacuum and the cathode emitters. V. K. Zworykin described a
permanently sealed, high-vacuum CRT with a thermionic emitter in 1931. This stable and reproducible
component allowed General Radio to manufacture an oscilloscope that was usable outside a laboratory
setting.[1]
After World War II surplus electronic parts became the basis for the revival of Heathkit Corporation, and
a $50 oscilloscope kit made from such parts proved its premiere market success.
Inputs
The signal to be measured is fed to one of the input connectors, which is usually a coaxial connector such
as a BNC or UHF type. Binding posts or banana plugs may be used for lower frequencies. If the signal
source has its own coaxial connector, then a simple coaxial cable is used; otherwise, a specialized cable
called a "scope probe", supplied with the oscilloscope, is used. In general, for routine use, an open wire
test lead for connecting to the point being observed is not satisfactory, and a probe is generally necessary.
General-purpose oscilloscopes usually present an input impedance of 1 megohm in parallel with a small
but known capacitance such as 20 picofarads.[5] This allows the use of standard oscilloscope probes.[6]
Scopes for use with very high frequencies may have 50 Ω inputs. These must be either connected directly
to a 50 Ω signal source or used with Z0 or active probes.
Less-frequently-used inputs include one (or two) for triggering the sweep, horizontal deflection for X‑Y
mode displays, and trace brightening/darkening, sometimes called z'‑axis inputs.
Probes
Open wire test leads (flying leads) are likely to pick up interference, so they are not suitable for low level
signals. Furthermore, the leads have a high inductance, so they are not suitable for high frequencies.
Using a shielded cable (i.e., coaxial cable) is better for low level signals. Coaxial cable also has lower
inductance, but it has higher capacitance: a typical 50 ohm cable has about 90 pF per meter.
Consequently, a one-meter direct (1×) coaxial probe loads a circuit with a capacitance of about 110 pF
and a resistance of 1 megohm.
To minimize loading, attenuator probes (e.g., 10× probes) are used. A typical probe uses a 9 megohm
series resistor shunted by a low-value capacitor to make an RC compensated divider with the cable
capacitance and scope input. The RC time constants are adjusted to match. For example, the 9 megohm
series resistor is shunted by a 12.2 pF capacitor for a time constant of 110 microseconds. The cable
capacitance of 90 pF in parallel with the scope input of 20 pF and 1 megohm (total capacitance 110 pF)
also gives a time constant of 110 microseconds. In practice, there is an adjustment so the operator can
precisely match the low frequency time constant (called compensating the probe). Matching the time
constants makes the attenuation independent of frequency. At low frequencies (where the resistance of R
is much less than the reactance of C), the circuit looks like a resistive divider; at high frequencies
(resistance much greater than reactance), the circuit looks like a capacitive divider.[7]
The result is a frequency compensated probe for modest frequencies. It presents a load of about
10 megohms shunted by 12 pF. Such a probe is an improvement, but does not work well when the time
scale shrinks to several cable transit times or less (transit time is typically 5 ns). In that time frame, the
cable looks like its characteristic impedance, and reflections from the transmission line mismatch at the
scope input and the probe causes ringing.[8] The modern scope probe uses lossy low capacitance
transmission lines and sophisticated frequency shaping networks to make the 10× probe perform well at
several hundred megahertz. Consequently, there are other adjustments for completing the
compensation.[9][10]
Probes with 10:1 attenuation are by far the most common; for large signals (and slightly-less capacitive
loading), 100:1 probes may be used. There are also probes that contain switches to select 10:1 or direct
(1:1) ratios, but the latter setting has significant capacitance (tens of pF) at the probe tip, because the
whole cable's capacitance is then directly connected.
Most oscilloscopes provide for probe attenuation factors, displaying the effective sensitivity at the probe
tip. Historically, some auto-sensing circuitry used indicator lamps behind translucent windows in the
panel to illuminate different parts of the sensitivity scale. To do so, the probe connectors (modified
BNCs) had an extra contact to define the probe's attenuation. (A certain value of resistor, connected to
ground, "encodes" the attenuation.) Because probes wear out, and because the auto-sensing circuitry is
not compatible between different oscilloscope makes, auto-sensing probe scaling is not foolproof.
Likewise, manually setting the probe attenuation is prone to user error. Setting the probe scaling
incorrectly is a common error, and throws the reading off by a factor of 10.
Special high voltage probes form compensated attenuators with the oscilloscope input. These have a large
probe body, and some require partly filling a canister surrounding the series resistor with volatile liquid
fluorocarbon to displace air. The oscilloscope end has a box with several waveform-trimming
adjustments. For safety, a barrier disc keeps the user's fingers away from the point being examined.
Maximum voltage is in the low tens of kV. (Observing a high voltage ramp can create a staircase
waveform with steps at different points every repetition, until the probe tip is in contact. Until then, a tiny
arc charges the probe tip, and its capacitance holds the voltage (open circuit). As the voltage continues to
climb, another tiny arc charges the tip further.)
There are also current probes, with cores that surround the conductor carrying current to be examined.
One type has a hole for the conductor, and requires that the wire be passed through the hole for semi-
permanent or permanent mounting. However, other types, used for temporary testing, have a two-part
core that can be clamped around a wire. Inside the probe, a coil wound around the core provides a current
into an appropriate load, and the voltage across that load is proportional to current. This type of probe
only senses AC.
A more-sophisticated probe includes a magnetic flux sensor (Hall effect sensor) in the magnetic circuit.
The probe connects to an amplifier, which feeds (low frequency) current into the coil to cancel the sensed
field; the magnitude of the current provides the low-frequency part of the current waveform, right down
to DC. The coil still picks up high frequencies. There is a combining network akin to a loudspeaker
crossover.
Focus control
This control adjusts CRT focus to obtain the sharpest, most-detailed trace. In practice, focus must be
adjusted slightly when observing very different signals, so it must be an external control. The control
varies the voltage applied to a focusing anode within the CRT. Flat-panel displays do not need this
control.
Intensity control
This adjusts trace brightness. Slow traces on CRT oscilloscopes need less, and fast ones, especially if not
often repeated, require more brightness. On flat panels, however, trace brightness is essentially
independent of sweep speed, because the internal signal processing effectively synthesizes the display
from the digitized data.
Astigmatism
This control may instead be called "shape" or "spot shape". It adjusts the voltage on the last CRT anode
(immediately next to the Y deflection plates). For a circular spot, the final anode must be at the same
potential as both of the Y-plates (for a centred spot the Y-plate voltages must be the same). If the anode is
made more positive, the spot becomes elliptical in the X-plane as the more negative Y-plates will repel
the beam. If the anode is made more negative, the spot becomes elliptical in the Y-plane as the more
positive Y-plates will attract the beam. This control may be absent from simpler oscilloscope designs or
may even be an internal control. It is not necessary with flat panel displays.
Beam finder
Modern oscilloscopes have direct-coupled deflection amplifiers, which means the trace could be
deflected off-screen. They also might have their beam blanked without the operator knowing it. To help in
restoring a visible display, the beam finder circuit overrides any blanking and limits the beam deflection
to the visible portion of the screen. Beam-finder circuits often distort the trace while activated.
Graticule
The graticule is a grid of lines that serve as reference marks for measuring the displayed trace. These
markings, whether located directly on the screen or on a removable plastic filter, usually consist of a 1 cm
grid with closer tick marks (often at 2 mm) on the centre vertical and horizontal axis. One expects to see
ten major divisions across the screen; the number of vertical major divisions varies. Comparing the grid
markings with the waveform permits one to measure both voltage (vertical axis) and time (horizontal
axis). Frequency can also be determined by measuring the waveform period and calculating its reciprocal.
On old and lower-cost CRT oscilloscopes the graticule is a sheet of plastic, often with light-diffusing
markings and concealed lamps at the edge of the graticule. The lamps had a brightness control. Higher-
cost instruments have the graticule marked on the inside face of the CRT, to eliminate parallax errors;
better ones also had adjustable edge illumination with diffusing markings. (Diffusing markings appear
bright.) Digital oscilloscopes, however, generate the graticule markings on the display in the same way as
the trace.
External graticules also protect the glass face of the CRT from accidental impact. Some CRT
oscilloscopes with internal graticules have an unmarked tinted sheet plastic light filter to enhance trace
contrast; this also serves to protect the faceplate of the CRT.
Accuracy and resolution of measurements using a graticule is relatively limited; better instruments
sometimes have movable bright markers on the trace. These permit internal circuits to make more refined
measurements.
Both calibrated vertical sensitivity and calibrated horizontal time are set in 1 – 2 – 5 – 10 steps. This
leads, however, to some awkward interpretations of minor divisions.
Digital oscilloscopes generate the graticule digitally. The scale, spacing, etc., of the graticule can
therefore be varied, and accuracy of readings may be improved.
Timebase controls
These select the horizontal speed of the CRT's spot as it creates the trace; this process is commonly
referred to as the sweep. In all but the least-costly modern oscilloscopes, the sweep speed is selectable
and calibrated in units of time per major graticule division. Quite a wide range of sweep speeds is
generally provided, from seconds to as fast as picoseconds (in the fastest) per division. Usually, a
continuously-variable control (often a knob in front of the
calibrated selector knob) offers uncalibrated speeds, typically
slower than calibrated. This control provides a range somewhat
greater than the calibrated steps, making any speed between the
steps available.
Often the observed signal is offset by a steady component, and only the changes are of interest. An input
coupling switch in the "AC" position connects a capacitor in series with the input that blocks low-
frequency signals and DC. However, when the signal has a fixed offset of interest, or changes slowly, the
user will usually prefer "DC" coupling, which bypasses any such capacitor. Most oscilloscopes offer the
DC input option. For convenience, to see where zero volts input currently shows on the screen, many
oscilloscopes have a third switch position (usually labeled "GND" for ground) that disconnects the input
and grounds it. Often, in this case, the user centers the trace with the vertical position control.
Better oscilloscopes have a polarity selector. Normally, a positive input moves the trace upward; the
polarity selector offers an "inverting" option, in which a positive-going signal deflects the trace
downward.
Multiple-trace oscilloscopes have a switch for each channel to enable or disable display of the channel's
trace.
Delayed-sweep controls
These include controls for the delayed-sweep timebase, which is calibrated, and often also variable. The
slowest speed is several steps faster than the slowest main sweep speed, though the fastest is generally the
same. A calibrated multiturn delay time control offers wide range, high resolution delay settings; it spans
the full duration of the main sweep, and its reading corresponds to graticule divisions (but with much
finer precision). Its accuracy is also superior to that of the display.
A switch selects display modes: Main sweep only, with a brightened region showing when the delayed
sweep is advancing, delayed sweep only, or (on some) a combination mode.
Good CRT oscilloscopes include a delayed-sweep intensity control, to allow for the dimmer trace of a
much-faster delayed sweep which nevertheless occurs only once per main sweep. Such oscilloscopes also
are likely to have a trace separation control for multiplexed display of both the main and delayed sweeps
together.
Triggered sweep
To display events with unchanging or slowly (visibly)
changing waveforms, but occurring at times that may
not be evenly spaced, modern oscilloscopes have
triggered sweeps. Compared to older, simpler
oscilloscopes with continuously-running sweep
oscillators, triggered-sweep oscilloscopes are
markedly more versatile.
A triggered sweep starts at a selected point on the Type 465 Tektronix oscilloscope. This was a
signal, providing a stable display. In this way, popular analog oscilloscope, portable, and is a
triggering allows the display of periodic signals such representative example.
as sine waves and square waves, as well as
nonperiodic signals such as single pulses, or pulses
that do not recur at a fixed rate.
With triggered sweeps, the scope blanks the beam and starts to reset the sweep circuit each time the beam
reaches the extreme right side of the screen. For a period of time, called holdoff, (extendable by a front-
panel control on some better oscilloscopes), the sweep circuit resets completely and ignores triggers.
Once holdoff expires, the next trigger starts a sweep. The trigger event is usually the input waveform
reaching some user-specified threshold voltage (trigger level) in the specified direction (going positive or
going negative—trigger polarity).
In some cases, variable holdoff time can be useful to make the sweep ignore interfering triggers that
occur before the events to be observed. In the case of repetitive, but complex waveforms, variable holdoff
can provide a stable display that could not otherwise be achieved.
Holdoff
Trigger holdoff defines a certain period following a trigger during which the sweep cannot be triggered
again. This makes it easier to establish a stable view of a waveform with multiple edges, which would
otherwise cause additional triggers.[11]
Example
Imagine the following repeating waveform:
The green line is the waveform, the red vertical partial line represents the location of the trigger, and the
yellow line represents the trigger level. If the scope was simply set to trigger on every rising edge, this
waveform would cause three triggers for each cycle:
Assuming the signal is fairly high frequency, the scope display would probably look something like this:
On an actual scope, each trigger would be the same channel, so all would be the same color.
It is desirable for the scope to trigger on only one edge per cycle, so it is necessary to set the holdoff at
slightly less than the period of the waveform. This prevents triggering from occurring more than once per
cycle, but still lets it trigger on the first edge of the next cycle.
Recurrent sweeps
If the input signal is periodic, the sweep repetition rate can be adjusted to display a few cycles of the
waveform. Early (tube) oscilloscopes and lowest-cost oscilloscopes have sweep oscillators that run
continuously, and are uncalibrated. Such oscilloscopes are very simple, comparatively inexpensive, and
were useful in radio servicing and some TV servicing. Measuring voltage or time is possible, but only
with extra equipment, and is quite inconvenient. They are primarily qualitative instruments.
They have a few (widely spaced) frequency ranges, and relatively wide-range continuous frequency
control within a given range. In use, the sweep frequency is set to slightly lower than some submultiple of
the input frequency, to display typically at least two cycles of the input signal (so all details are visible). A
very simple control feeds an adjustable amount of the vertical signal (or possibly, a related external
signal) to the sweep oscillator. The signal triggers beam blanking and a sweep retrace sooner than it
would occur free-running, and the display becomes stable.
Single sweeps
Some oscilloscopes offer these. The user manually arms the sweep circuit (typically by a pushbutton or
equivalent). "Armed" means it is ready to respond to a trigger. Once the sweep completes, it resets, and
does not sweep again until re-armed. This mode, combined with an oscilloscope camera, captures single-
shot events.
external trigger, a pulse from an external source connected to a dedicated input on the
scope.
edge trigger, an edge detector that generates a pulse when the input signal crosses a
specified threshold voltage in a specified direction. These are the most common types of
triggers; the level control sets the threshold voltage, and the slope control selects the
direction (negative or positive-going). (The first sentence of the description also applies to
the inputs to some digital logic circuits; those inputs have fixed threshold and polarity
response.)
video trigger, also known as TV trigger, a circuit that extracts synchronizing pulses from
video formats such as PAL and NTSC and triggers the timebase on every line, a specified
line, every field, or every frame. This circuit is typically found in a waveform monitor device,
though some better oscilloscopes include this function.
delayed trigger, which waits a specified time after an edge trigger before starting the sweep.
As described under delayed sweeps, a trigger delay circuit (typically the main sweep)
extends this delay to a known and adjustable interval. In this way, the operator can examine
a particular pulse in a long train of pulses.
Some recent designs of oscilloscopes include more sophisticated triggering schemes; these are described
toward the end of this article.
Delayed sweeps
More sophisticated analog oscilloscopes contain a second timebase for a delayed sweep. A delayed sweep
provides a very detailed look at some small selected portion of the main timebase. The main timebase
serves as a controllable delay, after which the delayed timebase starts. This can start when the delay
expires, or can be triggered (only) after the delay expires. Ordinarily, the delayed timebase is set for a
faster sweep, sometimes much faster, such as 1000:1. At extreme ratios, jitter in the delays on consecutive
main sweeps degrades the display, but delayed-sweep triggers can overcome this.
The display shows the vertical signal in one of several modes: the main timebase, or the delayed timebase
only, or a combination thereof. When the delayed sweep is active, the main sweep trace brightens while
the delayed sweep is advancing. In one combination mode, provided only on some oscilloscopes, the
trace changes from the main sweep to the delayed sweep once the delayed sweep starts, though less of the
delayed fast sweep is visible for longer delays. Another combination mode multiplexes (alternates) the
main and delayed sweeps so that both appear at once; a trace separation control displaces them. DSOs can
display waveforms this way, without offering a delayed timebase as such.
Switching channels can be asynchronous, i.e. free-running, with respect to the sweep frequency; or it can
be done after each horizontal sweep is complete. Asynchronous switching is usually designated
"Chopped", while sweep-synchronized is designated "Alt[ernate]". A given channel is alternately
connected and disconnected, leading to the term "chopped". Multi-trace oscilloscopes also switch
channels either in chopped or alternate modes.
In general, chopped mode is better for slower sweeps. It is possible for the internal chopping rate to be a
multiple of the sweep repetition rate, creating blanks in the traces, but in practice this is rarely a problem.
The gaps in one trace are overwritten by traces of the following sweep. A few oscilloscopes had a
modulated chopping rate to avoid this occasional problem. Alternate mode, however, is better for faster
sweeps.
True dual-beam CRT oscilloscopes did exist, but were not common. One type (Cossor, U.K.) had a beam-
splitter plate in its CRT, and single-ended deflection following the splitter. Others had two complete
electron guns, requiring tight control of axial (rotational) mechanical alignment in manufacturing the
CRT. Beam-splitter types had horizontal deflection common to both vertical channels, but dual-gun
oscilloscopes could have separate time bases, or use one time base for both channels. Multiple-gun CRTs
(up to ten guns) were made in past decades. With ten guns, the envelope (bulb) was cylindrical
throughout its length. (Also see "CRT Invention" in Oscilloscope history.)
In dual and multiple-trace oscilloscopes, an internal electronic switch selects the relatively low-level
output of one channel's early-stage amplifier and sends it to the following stages of the vertical amplifier.
In free-running ("chopped") mode, the oscillator (which may be simply a different operating mode of the
switch driver) blanks the beam before switching, and unblanks it only after the switching transients have
settled.
Part way through the amplifier is a feed to the sweep trigger circuits, for internal triggering from the
signal. This feed would be from an individual channel's amplifier in a dual or multi-trace oscilloscope, the
channel depending upon the setting of the trigger source selector.
This feed precedes the delay (if there is one), which allows the sweep circuit to unblank the CRT and start
the forward sweep, so the CRT can show the triggering event. High-quality analog delays add a modest
cost to an oscilloscope, and are omitted in cost-sensitive oscilloscopes.
The delay, itself, comes from a special cable with a pair of conductors wound around a flexible,
magnetically soft core. The coiling provides distributed inductance, while a conductive layer close to the
wires provides distributed capacitance. The combination is a wideband transmission line with
considerable delay per unit length. Both ends of the delay cable require matched impedances to avoid
reflections.
X-Y mode
Most modern oscilloscopes have several inputs for voltages, and
thus can be used to plot one varying voltage versus another. This is
especially useful for graphing I-V curves (current versus voltage
characteristics) for components such as diodes, as well as
Lissajous figures. Lissajous figures are an example of how an
oscilloscope can be used to track phase differences between
multiple input signals. This is very frequently used in broadcast
engineering to plot the left and right stereophonic channels, to A 24-hour clock displayed on a CRT
ensure that the stereo generator is calibrated properly. Historically, oscilloscope configured in X-Y mode
stable Lissajous figures were used to show that two sine waves as a vector monitor with dual R–2R
had a relatively simple frequency relationship, a numerically-small DACs to generate the analog
ratio. They also indicated phase difference between two sine voltages
waves of the same frequency.
The X-Y mode also lets the oscilloscope serve as a vector monitor to display images or user interfaces.
Many early games, such as Tennis for Two, used an oscilloscope as an output device.[12]
Complete loss of signal in an X-Y CRT display means that the beam is stationary, striking a small spot.
This risks burning the phosphor if the brightness is too high. Such damage was more common in older
scopes as the phosphors previously used burned more easily. Some dedicated X-Y displays reduce beam
current greatly, or blank the display entirely, if there are no inputs present.
Z input
Some analogue oscilloscopes feature a Z input. This is generally an input terminal that connects directly
to the CRT grid (usually via a coupling capacitor). This allows an external signal to either increase (if
positive) or decrease (if negative) the brightness of the trace, even allowing it to be totally blanked. The
voltage range to achieve cut-off to a brightened display is of the order of 10–20 volts depending on the
CRT characteristics.
An example of a practical application is if a pair of sine waves of known frequency are used to generate a
circular Lissajous figure and a higher unknown frequency is applied to the Z input. This turns the
continuous circle into a circle of dots. The number of dots multiplied by the X-Y frequency gives the Z
frequency. This technique only works if the Z frequency is an integer ratio of the X-Y frequency and only
if it is not so large that the dots become so numerous that they are difficult to count.
Bandwidth
As with all practical instruments, oscilloscopes do not respond equally to all possible input frequencies.
The range of sinusoid frequencies an oscilloscope can usefully display is referred to as its bandwidth.
Bandwidth applies primarily to the Y-axis, though the X-axis sweeps must be fast enough to show the
highest-frequency waveforms.
The bandwidth is defined as the frequency at which the sensitivity is 0.707 of the sensitivity at DC or the
lowest AC frequency (a drop of 3 dB).[13] The oscilloscope's response drops off rapidly as the input
frequency rises above that point. Within the stated bandwidth the response is not necessarily exactly
uniform (or "flat"), but should always fall within a +0 to −3 dB range. One source[13] says there is a
noticeable effect on the accuracy of voltage measurements at only 20 percent of the stated bandwidth.
Some oscilloscopes' specifications do include a narrower tolerance range within the stated bandwidth.
Probes also have bandwidth limits and must be chosen and used to handle the frequencies of interest
properly. To achieve the flattest response, most probes must be "compensated" (an adjustment performed
using a test signal from the oscilloscope) to allow for the reactance of the probe's cable.
Another related specification is rise time. This is the time taken between 10% and 90% of the maximum
amplitude response at the leading edge of a pulse. It is related to the bandwidth approximately by:
For example, an oscilloscope with a rise time of 1 nanosecond would have a bandwidth of 350 MHz.
In analog instruments, the bandwidth of the oscilloscope is limited by the vertical amplifiers and the CRT
or other display subsystem. In digital instruments, the sampling rate of the analog-to-digital converter
(ADC) is a factor, but the stated analog bandwidth (and therefore the overall bandwidth of the instrument)
is usually less than the ADC's Nyquist frequency. This is due to limitations in the analog signal amplifier,
deliberate design of the anti-aliasing filter that precedes the ADC, or both.
For a digital oscilloscope, a rule of thumb is that the continuous sampling rate should be ten times the
highest frequency desired to resolve; for example a 20 megasample/second rate would be applicable for
measuring signals up to about 2 MHz. This lets the anti-aliasing filter be designed with a 3 dB down point
of 2 MHz and an effective cutoff at 10 MHz (the Nyquist frequency), avoiding the artifacts of a very
steep ("brick-wall") filter.
A sampling oscilloscope can display signals of considerably higher frequency than the sampling rate if
the signals are exactly, or nearly, repetitive. It does this by taking one sample from each successive
repetition of the input waveform, each sample being at an increased time interval from the trigger event.
The waveform is then displayed from these collected samples. This mechanism is referred to as
"equivalent-time sampling".[15] Some oscilloscopes can operate in either this mode or in the more
traditional "real-time" mode at the operator's choice.
Other features
Some oscilloscopes have cursors. These are lines that can be
moved about the screen to measure the time interval between two
points, or the difference between two voltages. A few older
oscilloscopes simply brightened the trace at movable locations.
These cursors are more accurate than visual estimates referring to
graticule lines.[17][18]
Better quality general purpose oscilloscopes include a calibration A computer model of the sweep of
signal for setting up the compensation of test probes; this is (often) the oscilloscope
a 1 kHz square-wave signal of a definite peak-to-peak voltage
available at a test terminal on the front panel. Some better
oscilloscopes also have a squared-off loop for checking and adjusting current probes.
Sometimes a user wants to see an event that happens only occasionally. To catch these events, some
oscilloscopes—called storage scopes—preserve the most recent sweep on the screen. This was originally
achieved with a special CRT, a storage tube, which retained the image of even a very brief event for a
long time.
Some digital oscilloscopes can sweep at speeds as slow as once per hour, emulating a strip chart recorder.
That is, the signal scrolls across the screen from right to left. Most oscilloscopes with this facility switch
from a sweep to a strip-chart mode at about one sweep per ten seconds. This is because otherwise, the
scope looks broken: it is collecting data, but the dot cannot be seen.
All but the simplest models of current oscilloscopes more often use digital signal sampling. Samples feed
fast analog-to-digital converters, following which all signal processing (and storage) is digital.
Many oscilloscopes accommodate plug-in modules for different purposes, e.g., high-sensitivity amplifiers
of relatively narrow bandwidth, differential amplifiers, amplifiers with four or more channels, sampling
plugins for repetitive signals of very high frequency, and special-purpose plugins, including
audio/ultrasonic spectrum analyzers, and stable-offset-voltage direct-coupled channels with relatively
high gain.
Examples of use
One of the most frequent uses of scopes is
troubleshooting malfunctioning electronic equipment.
For example, where a voltmeter may show a totally
unexpected voltage, a scope may reveal that the
circuit is oscillating. In other cases the precise shape
or timing of a pulse is important.
Once the faulty stage is found, further probing can usually tell a skilled technician exactly which
component has failed. Once the component is replaced, the unit can be restored to service, or at least the
next fault can be isolated. This sort of troubleshooting is typical of radio and TV receivers, as well as
audio amplifiers, but can apply to quite different devices such as electronic motor drives.
Another use is to check newly designed circuitry. Often, a newly designed circuit misbehaves because of
design errors, bad voltage levels, electrical noise etc. Digital electronics usually operate from a clock, so a
dual-trace scope showing both the clock signal and a test signal dependent upon the clock is useful.
Storage scopes are helpful for "capturing" rare electronic events that cause defective operation.
Oscilloscopes are often used during real-time software development to check, among other things, missed
deadlines and worst-case latencies.[19]
Pictures of use
Automotive use
First appearing in the 1970s for ignition system analysis, automotive oscilloscopes are becoming an
important workshop tool for testing sensors and output signals on electronic engine management systems,
braking and stability systems. Some oscilloscopes can trigger and decode serial bus messages, such as the
CAN bus commonly used in automotive applications.
Software
Many oscilloscopes today provide one or more external interfaces to allow remote instrument control by
external software. These interfaces (or buses) include GPIB, Ethernet, serial port, USB and Wi-Fi.
Dual-beam oscilloscope
The dual-beam analog oscilloscope can display two signals
simultaneously. A special dual-beam CRT generates and deflects
two separate beams. Multi-trace analog oscilloscopes can simulate For analog television, an analog
a dual-beam display with chop and alternate sweeps—but those oscilloscope can be used as a
features do not provide simultaneous displays. (Real-time digital vectorscope to analyze complex
oscilloscopes offer the same benefits of a dual-beam oscilloscope, signal properties, such as this
but they do not require a dual-beam display.) The disadvantages of display of SMPTE color bars.
the dual trace oscilloscope are that it cannot switch quickly
between traces, and cannot capture two fast transient events. A
dual beam oscilloscope avoids those problems.
Digital oscilloscopes
While analog devices use continually varying voltages, digital devices use numbers that correspond to
samples of the voltage. In the case of digital oscilloscopes, an analog-to-digital converter (ADC) changes
the measured voltages into digital information.
The digital storage oscilloscope, or DSO for short, is the standard
type of oscilloscope today for the majority of industrial
applications, and thanks to the low costs of entry-level
oscilloscopes even for hobbyists. It replaces the electrostatic
storage method in analog storage scopes with digital memory,
which stores sample data as long as required without degradation
and displays it without the brightness issues of storage-type CRTs.
It also allows complex processing of the signal by high-speed
digital signal processing circuits.[1]
Mixed-signal oscilloscopes
A logic analyzer is similar to an oscilloscope, but for each input signal only provides the logic level
without the shape of its analog waveform. A mixed-signal oscilloscope (or MSO) meanwhile has two
kinds of inputs: a small number of analog channels (typically two or four), and a larger number of logic
channels (typically sixteen). It provides the ability to accurately time-correlate analog and logic signals,
thus offering a distinct advantage over a separate oscilloscope and logic analyzer. Typically, logic
channels may be grouped and displayed as a bus with each bus value displayed at the bottom of the
display in hexadecimal or binary. On most MSOs, the trigger can be set across both analog and logic
channels.
Mixed-domain oscilloscopes
A mixed-domain oscilloscope (MDO) is an oscilloscope that comes with an additional RF input which is
solely used for dedicated FFT-based spectrum analyzer functionality. Often, this RF input offers a higher
bandwidth than the conventional analog input channels. This is in contrast to the FFT functionality of
conventional digital oscilloscopes, which use the normal analog inputs. Some MDOs allow time-
correlation of events in the time domain (like a specific serial data package) with events happening in the
frequency domain (like RF transmissions).
Handheld oscilloscopes
Handheld oscilloscopes are useful for many test and field service applications. Today, a handheld
oscilloscope is usually a digital sampling oscilloscope, using a liquid crystal display.
Many handheld and bench oscilloscopes have the ground reference voltage common to all input channels.
If more than one measurement channel is used at the same time, all the input signals must have the same
voltage reference, and the shared default reference is the "earth". If there is no differential preamplifier or
external signal isolator, this traditional desktop oscilloscope is not suitable for floating measurements.
(Occasionally an oscilloscope user breaks the ground pin in the power supply cord of a bench-top
oscilloscope in an attempt to isolate the signal common from the earth ground. This practice is unreliable
since the entire stray capacitance of the instrument cabinet connects into the circuit. It is also a hazard to
break a safety ground connection, and instruction manuals strongly advise against it.)
Some models of oscilloscope have isolated inputs, where the signal reference level terminals are not
connected together. Each input channel can be used to make a "floating" measurement with an
independent signal reference level. Measurements can be made without tying one side of the oscilloscope
input to the circuit signal common or ground reference.
PC-based oscilloscopes
Some digital oscilloscope rely on a PC platform for display and
control of the instrument. This can be in the form of a standalone
oscilloscope with internal PC platform (PC mainboard), or as
external oscilloscope which connects through USB or LAN to a
separate PC or laptop.
Related instruments
A large number of instruments used in a variety of technical fields
are really oscilloscopes with inputs, calibration, controls, display PicoScope 6000 digital PC-based
calibration, etc., specialized and optimized for a particular oscilloscope using a laptop
computer for display and processing
application. Examples of such oscilloscope-based instruments
include waveform monitors for analyzing video levels in
television productions and medical devices such as vital function monitors and electrocardiogram and
electroencephalogram instruments. In automobile repair, an ignition analyzer is used to show the spark
waveforms for each cylinder. All of these are essentially oscilloscopes, performing the basic task of
showing the changes in one or more input signals over time in an X‑Y display.
Other instruments convert the results of their measurements to a repetitive electrical signal, and
incorporate an oscilloscope as a display element. Such complex measurement systems include spectrum
analyzers, transistor analyzers, and time domain reflectometers (TDRs). Unlike an oscilloscope, these
instruments automatically generate stimulus or sweep a measurement parameter.
See also
Eye pattern
Phonodeik
Tennis for Two, an oscilloscope game
Time-domain reflectometry
Vectorscope
Waveform monitor
References
1. Kularatna, Nihal (2003), "Fundamentals of Oscilloscopes", Digital and Analogue
Instrumentation: Testing and Measurement, Institution of Engineering and Technology,
pp. 165–208, ISBN 978-0-85296-999-1
2. How the Cathode Ray Oscillograph Is Used in Radio Servicing (http://www.cfp-radio.com/do
cumentations/how%20the%20cathode%20ray%20oscillograph%20is%20used%20in%20ra
dio%20servicing.pdf) Archived (https://web.archive.org/web/20130524122111/http://www.cfp
-radio.com/documentations/how%20the%20cathode%20ray%20oscillograph%20is%20use
d%20in%20radio%20servicing.pdf) 2013-05-24 at the Wayback Machine, National Radio
Institute (1943)
3. "Cathode-Ray Oscillograph 274A Equipment DuMont Labs, Allen B" (http://www.radiomuseu
m.org/r/dumont_la_cathode_ray_oscillograph_7.html) (in German). Radiomuseum.org.
Archived (https://web.archive.org/web/20140203122536/http://www.radiomuseum.org/r/dum
ont_la_cathode_ray_oscillograph_7.html) from the original on 2014-02-03. Retrieved
2014-03-15.
4. Marton, L. (1980). "Ferdinand Braun: Forgotten Forefather" (https://books.google.com/book
s?id=27c1WOjCBX4C&pg=PA252). In Suesskind, Charles (ed.). Advances in electronics
and electron physics. Vol. 50. Academic Press. p. CRT. ISBN 978-0-12-014650-5. Archived
(https://web.archive.org/web/20140503063825/http://books.google.com/books?id=27c1WOj
CBX4C&lpg=PA252&pg=PA252) from the original on 2014-05-03. "occurs first in a pair of
later papers by Zenneck (1899a,b)"
5. The 20 picofarad value is typical for scope bandwidths around 100 MHz; for example, a
200 MHz Tektronix 7A26 input impedance is 1 MΩ and 22 pF. (Tektronix (1983, p. 271); see
also Tektronix (1998, p. 503), "typical high Z 10× passive probe model".) Lower bandwidth
scopes used higher capacitances; the 1 MHz Tektronix 7A22 input impedance is 1 MΩ and
47 pF. (Tektronix 1983, pp. 272–273) Higher bandwidth scopes use smaller capacitances.
The 500 MHz Tektronix TDS510A input impedance is 1 MΩ and 10 pF. (Tektronix 1998,
p. 78)
6. Probes are designed for a specific input impedance. They have compensation adjustments
with a limited range, so they often cannot be used on different input impedances.
7. Wedlock & Roberge (1969)
8. Kobbe & Polits (1959)
9. Tektronix (1983, p. 426); Tek claims 300 MHz resistive coax at 30 pF per meter; schematic
has 5 adjustments.
10. Zeidlhack & White (1970)
11. Jones, David. "Oscilloscope Trigger Holdoff Tutorial" (http://www.eevblog.com/2011/03/30/e
evblog-159-oscilloscope-trigger-holdoff-tutorial/). EEVblog. Archived (https://web.archive.or
g/web/20130128160235/http://www.eevblog.com/2011/03/30/eevblog-159-oscilloscope-trigg
er-holdoff-tutorial/) from the original on 28 January 2013. Retrieved 30 December 2012.
12. Nosowitz, Dan (2008-11-08). " 'Tennis for Two', the World's First Graphical Videogame" (http
s://gizmodo.com/5080541/retromodo-tennis-for-two-the-worlds-first-graphical-videogame).
Retromodo. Gizmodo. Archived (https://web.archive.org/web/20081207052713/http://gizmod
o.com/5080541/retromodo-tennis-for-two-the-worlds-first-graphical-videogame) from the
original on 2008-12-07. Retrieved 2008-11-09.
13. Webster, John G. (1999). The Measurement, Instrumentation and Sensors Handbook
(illustrated ed.). Springer. pp. 37–24. ISBN 978-3540648307.
14. Spitzer, Frank; Howarth, Barry (1972), Principles of modern Instrumentation (https://archive.
org/details/principlesofmode00spit/page/119), New York: Holt, Rinehart and Winston, p. 119,
ISBN 0-03-080208-3
15. "Equivalent Time Sampling Oscilloscope vs. Real-Time Oscilloscope" (http://literature.cdn.ke
ysight.com/litweb/pdf/5989-8794EN.pdf) (PDF). Archived (https://web.archive.org/web/2015
0402113819/http://literature.cdn.keysight.com/litweb/pdf/5989-8794EN.pdf) (PDF) from the
original on 2015-04-02. Retrieved 2015-03-20.
16. Horowitz, Paul; Hill, Winfield (2015). "O.2.1 What's different?". The Art of Electronics (3rd
and Kindle ed.). Cambridge University Press. ISBN 978-0-521-80926-9.
17. Hickman, Ian (2001). Oscilloscopes (https://books.google.com/books?id=O2oj04vbQqgC).
Newnes. pp. 4, 20. ISBN 0-7506-4757-4. Retrieved 2022-01-15.
18. Herres, David (2020). Oscilloscopes: A Manual for Students, Engineers, and Scientists.
Springer Science+Business Media. pp. 120–121. doi:10.1007/978-3-030-53885-9 (https://do
i.org/10.1007%2F978-3-030-53885-9). ISBN 978-3-030-53885-9. S2CID 226749445 (http
s://api.semanticscholar.org/CorpusID:226749445).
19. Marchesotti, M.; Migliardi, M.; Podesta, R. (2006). "A measurement-based analysis of the
responsiveness of the Linux kernel" (https://www.researchgate.net/publication/4230105).
13th Annual IEEE International Symposium and Workshop on Engineering of Computer-
Based Systems (ECBS'06). pp. 10 pp.-408. doi:10.1109/ECBS.2006.9 (https://doi.org/10.11
09%2FECBS.2006.9). ISBN 0-7695-2546-6. S2CID 15440587 (https://api.semanticscholar.o
rg/CorpusID:15440587) – via ResearchGate.
20. Green, Leslie (June 21, 2001), "The alias theorems: practical undersampling for expert
engineers" (http://www.edn.com/design/analog/4346463/The-alias-theorems-practical-under
sampling-for-expert-engineers), EDN, archived (https://web.archive.org/web/201306201735
36/http://www.edn.com/design/analog/4346463/The-alias-theorems-practical-undersampling
-for-expert-engineers) from the original on 20 June 2013, retrieved 11 October 2012
US 2883619 (https://worldwide.espacenet.com/textdoc?DB=EPODOC&IDX=US2883619),
Kobbe, John R. & Polits, William J., "Electrical Probe", published 1959-04-21
Tektronix (1983), Tek Products, Tektronix
Tektronix (1998), Measurement Products Catalog 1998/1999, Tektronix
Wedlock, Bruce D.; Roberge, James K. (1969), Electronic Components and Measurements,
Prentice-Hall, pp. 150–152, Bibcode:1969ecm..book.....W (https://ui.adsabs.harvard.edu/ab
s/1969ecm..book.....W), ISBN 0-13-250464-2
US 3532982 (https://worldwide.espacenet.com/textdoc?DB=EPODOC&IDX=US3532982),
Zeidlhack, Donald F. & White, Richard K., "Transmission Line Termination Circuit", published
1970-10-06
External links
The Cathode Ray Tube site (https://web.archive.org/web/20140701141404/http://www.crtsit
e.com/page3.html)
Virtual Oscilloscope Museum (http://www.oscilloscopemuseum.com)