QD C-Spine Coil
QD C-Spine Coil
QD C-Spine Coil
Installation and Adjustment Manual
Model MRP-5000
MR-QCS-32
Model MRP-7000
MR-QCS-42
L5E-BMA106-5
1.0 Outline
The QD C-Spine Coil is intended to provide high quality imaging of the cervical spine;
dependent upon patient anatomy, the coil sensitivity may additionally image upper thoracic
and lower cranial structures. The QD C-Spine Coil makes use of a split top coil design. A
releaseable bridge structure is provided for easier patient positioning. Refer to Figures 1A
and 1B below.
Patient Orientation:
Bridge
H A
L
Coil Base
R
Carry Handle P F
(both sides)
Head Immobilizer Pads
Patient Pad
Signal Pins
Alignment Pin
CAUTION
Do not carry the QD C-Spine Coil by the bridge.
Although the bridge latches securely to the coil base,
it contains sensitive electronics, and it is not durable
as a carry handle.
Page 2 L5E-BMA106-5
2.0 Purpose
This document specifies the installation test and adjustment procedure for the QD C-Spine
coils intended for use with the MRP-5000, MRP-7000 and AIRIS. The test results shall be
documented for quality assurance purposes and evaluated to ensure that each production
coil meets the coil performance specifications. The installation test and adjustment method
for the MRP-5000 is described first, followed by the method for the MRP-7000/AIRIS.
5.1.1 Turn on the Vector Impedance Meter, allowing it to warm-up 5 minutes before use.
5.1.2 Set the frequency of the Impedance Meter to the system frequency (f0).
5.1.3 Assemble the QD C-Spine coil. Ensure that the bridge is securely latched to the coil base.
Use pads as necessary such that the coil and phantom are centered vertically in the magnet
bore. Place the QD C-Spine RF coil pad inside the coil. Remove the cover to the No. 3
phantom, as the cover will not be used for the balance of this procedure. Set the No. 3
Phantom on top of the pad, as shown below. Ensure that the phantom bottle cap faces out of
the magnet. Refer to Figure 2 on the following page.
Page 3 L5E-BMA106-5
Coil Landmark Line
#3 Phantom
3/8" RF Coil Pad
Phantom #3
5.1.4 Position the coil on the patient table so the phantom will be centered laterally and vertically
within the magnet. Visually line up the edge of the bottle cap with the edge of the coil base
for longitudinal positioning. Refer to Figure 2 above..
5.1.5 Press set to advance the coil and phantom to magnet isocenter.
5.1.6 For the MRP-5000, plug the coil cables into the connectors on the left inner surface of the
gantry.
5.2.1 Access the COIL DATA ENTRY program and add QD C-Spine as the coil name for coil
#10, or any other unused coil number. Select RF LINE 1 - A only.
Page 4 L5E-BMA106-5
5.2.2 Perform scanogram (coronal) and display the phantom image (as diagrammed in step #1 of
Figure 3).
5.2.3 Adjust Window to about 100, and Level as appropriate, to present an image similar to
phantom diagrammed in step #2 of Figure 3 below.
1 2 3
5.2.4 Locate the center of coil sensitivity (identified for example purposes as ‘A’ in phantom
diagrammed in step #3 of Figure 3).
5.2.5 Move the table as appropriate to align the center of coil sensitivity ‘A’ to coincide with the
magnet isocenter.
5.2.7 Use the SIGNAL GENERATION CHECK program to set the TUNE value to 18.
5.2.8 Use the Impedance Meter to monitor the coil LINE A impedance.
5.2.9 Identify the tunable capacitor labeled “TUNE A” on the coil, to adjust the LINE A
impedance (refer to Figure 4 below).Tune the “TUNE A” capacitor so that the LINE A
impedance, at the system frequency, is specified as :
Page 5 L5E-BMA106-5
Coil Landmark Line
#3 Phantom
3/8" RF Coil Pad A
B
Thin Mattress Pad
Tuning A : Saddle Coil
Ports B : Solenoid Coil
Figure 4. Coil Tuning Ports Locations
5.2.11 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.
5.2.12 Use the SIGNAL GENERATION CHECK program to set the TUNE value to 18.
5.2.13 Access the COIL DATA ENTRY program and select the QD C-Spine coil. Select RF LINE
3 - B only.
5.2.14 Use the Impedance Meter to monitor the coil LINE B impedance. The frequency of the
Impedance Meter should remain at the system frequency (f0) for the measurement.
5.2.15 Identify the tunable capacitor labeled “TUNE B” on the coil, to adjust the LINE B
impedance (refer to Figure 4 above). Tune the “TUNE B” capacitor so that the LINE B
impedance, at the system frequency, is specified as :
5.2.17 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.
5.3.1 Access the COIL DATA ENTRY program and select the QD C-Spine coil. Select RF LINE
2 - A and B. Set the ATT value to 2.
5.3.2 Perform scanogram (coronal) to identify the center of the coil sensitivity. Move the table as
appropriate to align the center of coil sensitivity with the magnet isocenter.
Page 6 L5E-BMA106-5
5.3.3 Perform scanogram (transverse) to check vertical centering of the phantom. Adjust the
phantom height as necessary.
5.3.5 Perform the RF GAIN (automatic) function on the coil to find the transmit power level for
TRS and COR planes. Calculate the average of the TRS and COR values, and use the
‘Manual’ function to enter the average as the power for this coil number.
MRP-5000
Sequence SE
TR 500
TE 40
Slice Thickness 10
AVE 2
Projection 256
SQ No. 3
FOV 260
6.2 Verify that the SN Ratio falls within the specified range according to the following table;
contact HMSA Technical Support if it does not:
6.3 Inspect the SNR image for asymmetrical shading or image artifacts; contact HMSA
Technical Support if any are noted.
Page 7 L5E-BMA106-5
8.0 MRP-7000/AIRIS Materials
8.1 Equipment List
● QD C-Spine Coil
● MRP-7000 or AIRIS MRI system;
● Vector impedance meter, HP 4193A or equivalent (qty. 1);
● Test cable (qty. 1): RG58 BNC male/BNC female, half wave length, with…
● Integral fine-tuning adjustable capacitor;
● Hitachi RF coil loading phantom no. 2;
● Non-magnetic screw drivers;
● Non-metallic screwdriver tuning tool.
9.1.1 Turn on the Vector Impedance Meter, allowing it to warm-up 5 minutes before use.
9.1.2 Set the frequency of the Impedance Meter to the system frequency (f0). Calibrate the half
wave length cable for the particular system frequency by fine tuning the capacitor of the
cable assembly. The cable should be calibrated as:
9.1.3 Assemble the QD C-Spine coil. Ensure that the bridge is securely latched to the coil base.
Use pads as necessary such that the coil and phantom are centered vertically in the magnet
bore. Place the QD C-Spine RF coil pad inside the coil, and set the No. 2 Phantom (130mm)
on top of the pad, as shown below. Ensure that the label on the end of the phantom faces out
of the magnet. For the newer phantoms, ensure that the phantom bottle cap faces out of the
magnet. Refer to Figure 5 on the following page.
Page 8 L5E-BMA106-5
Coil Landmark Line
#2 Phantom (130mm)
3/8" RF Coil Pad
Foam Pad(s)
(as needed)
Phantom #2
Phantom #2
Page 9 L5E-BMA106-5
9.1.4 Plug in the coil and position the coil in the center of the magnet on the patient table.
9.1.5 Press set to advance the coil and phantom to magnet isocenter.
9.1.6 For the MRP-7000, plug the coil cable into the connector on the left inner surface of the
front facade, aligning the D-shaped connector before inserting.
9.1.7 For the AIRIS, remove the pre-amp assembly cover. Plug the coil cable into the connector
on the table, aligning the D-shaped connector before inserting.
9.2.1 Access the COIL DATA ENTRY program and add QD C-Spine as the coil name for coil
#23. Select RF LINE 1 - A only.
9.2.2 Perform scanogram (coronal) and display the phantom image (as diagrammed in step #1 of
Figure 6).
9.2.3 Adjust Window to about 100, and Level as appropriate, to present an image similar to
phantom diagrammed in step #2 of Figure 6 below.
1 2 3
9.2.4 Locate the center of coil sensitivity (identified for example purposes as ‘A’ in phantom
diagrammed in step #3 of Figure 6).
9.2.5 Move the table as appropriate to align the center of coil sensitivity ‘A’ to coincide with the
magnet isocenter.
9.2.7 Use the SIGNAL GENERATION CHECK program to set the TUNE value to 18.
9.2.8 Use the Impedance Meter and the half wave length cable to monitor the coil LINE A
impedance.
Page 10 L5E-BMA106-5
9.2.9 Identify the tunable capacitor labeled “TUNE A” on the coil, to adjust the LINE A
impedance (refer to Figure 7 below).Tune the “TUNE A” capacitor so that the LINE A
impedance, at the system frequency, is specified as :
#2 Phantom (130mm)
QD C-Spine RFCoil Pad A
B
Foam Pad(s)
(as needed)
Tuning A : Saddle Coil
Ports B : Solenoid Coil
Figure 7. Coil Tuning Ports Locations
9.2.11 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.
9.2.12 Use the SIGNAL GENERATION CHECK program to set the TUNE value to 18.
9.2.13 Access the COIL DATA ENTRY program and select the QD C-Spine coil. Select RF LINE
3 - B only.
9.2.14 Use the Impedance Meter and the half wave length cable to monitor the coil LINE B
impedance. The frequency of the Impedance Meter should remain at the system frequency
(f0) for the measurement.
9.2.15 Identify the tunable capacitor labeled “TUNE B” on the coil, to adjust the LINE B
impedance (refer to Figure 7 above). Tune the “TUNE B” capacitor so that the LINE B
impedance, at the system frequency, is specified as :
9.2.17 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.
Page 11 L5E-BMA106-5
9.3 Registration of Coil Data
9.3.1 Access the COIL DATA ENTRY program and select the QD C-Spine coil. Select RF LINE
2 - A and B. Set the ATT value to 2. For systems with Operating System Software Ver. 4.0
or later, set the RF GAIN Adjust to Y.
9.3.2 Perform scanogram (coronal) to identify the center of the coil sensitivity. Move the table as
appropriate to align the center of coil sensitivity with the magnet isocenter.
9.3.3 Perform scanogram (transverse) to check vertical centering of the phantom. Adjust the
phantom height as necessary.
9.3.5 Perform the RF GAIN (automatic) function on the coil to find the transmit power level for
TRS and COR planes. Calculate the average of the TRS and COR values, and use the
‘Manual’ function to enter the average as the power for this coil number.
MRP-7000/AIRIS
Sequence SE
TR 500
TE 40
Slice Thickness 5
AVE 2
Projection 256
SQ No. 3
FOV 260
10.2 Verify that the SN Ratio falls within the specified range according to the following table;
contact HMSA Technical Support if it does not:
10.3 Inspect the SNR image for asymmetrical shading or image artifacts; contact HMSA
Technical Support if any are noted.
Page 12 L5E-BMA106-5
12.0 AIRIS II Materials
12.1 Equipment List
● QD C-Spine Coil
● AIRIS II MRI system;
● Vector impedance meter, HP 4193A or equivalent (qty. 1);
● Test cable (qty. 1): RG58 BNC male/BNC female, half wave length, with…
● Integral fine-tuning adjustable capacitor;
● Hitachi RF coil loading phantom no. 2;
● Non-magnetic screw drivers;
● Non-metallic screwdriver tuning tool;
● BNC female to SMB jack adapter.
13.1.1 Turn on the Vector Impedance Meter, allowing it to warm-up 5 minutes before use.
13.1.2 Set the frequency of the Impedance Meter to the system frequency (f0). Calibrate the half
wave length cable for the particular system frequency by fine tuning the capacitor of the
cable assembly. The cable should be calibrated as:
13.1.3 Assemble the QD C-Spine coil. Ensure that the bridge is securely latched to the coil base.
Use pads as necessary such that the coil and phantom are centered vertically in the magnet
bore. Place the QD C-Spine RF coil pad inside the coil, and set the No. 2 Phantom (130mm)
on top of the pad, as shown below. Ensure that the label on the end of the phantom faces out
of the magnet. For the newer phantoms, ensure that the phantom bottle cap faces out of the
magnet. Refer to Figure 8 on the following page.
Page 13 L5E-BMA106-5
Coil Landmark Line
#2 Phantom (130mm)
3/8" RF Coil Pad
Foam Pad(s)
(as needed)
Phantom #2
13.1.4 Plug in the coil and position the coil in the center of the magnet on the patient table.
13.1.5 Press set to advance the coil and phantom to magnet isocenter.
13.1.6 For the AIRIS II, remove the pre-amp assembly cover. Plug the coil cable into the
connector on the table, aligning the D-shaped connector before inserting.
Page 14 L5E-BMA106-5
13.2 Coil Tuning
13.2.1 Access the COIL DATA ENTRY program from the System Table Editor. Add QD C-Spine
as the coil name for coil #23. Select RECEPTION LINE 2 - A only:
Parameter Value
Coil Name QD C-Spine
Physical Coil No. 12
Attenuation Value 00
Tuning Measure Points 22
RF Gain Start Value 0e0
RF Gain Adjustment Step 16
Adjustment Average 1
Sensitivity Correction Coefficient 64
First Modulate Decoupling 64
Second Modulate Decoupling 64
Third Modulate Decoupling 2f
Fourth Modulate Decoupling 00
Selector 1
Number of Channels 1
Shading Correction No
Bed Movement No
Reception Line 2
QD Attenuation Value -2
Method A
13.2.2 Perform scanogram (coronal) and display the phantom image (as diagrammed in step #1 of
Figure 9).
13.2.3 Adjust Window to about 100, and Level as appropriate, to present an image similar to
phantom diagrammed in step #2 of Figure 9 below.
1 2 3
13.2.5 Move the table as appropriate to align the center of coil sensitivity ‘A’ to coincide with the
magnet isocenter.
13.2.7 Use the RF UNIT CONTROLLER to set the TUNE value to 18.
13.2.8 Use the Impedance Meter and the half wave length cable to monitor the coil LINE A
impedance.
13.2.9 Identify the tunable capacitor labeled “TUNE A” on the coil, to adjust the LINE A
impedance (refer to Figure 7 below).Tune the “TUNE A” capacitor so that the LINE A
impedance, at the system frequency, is specified as :
#2 Phantom (130mm)
QD C-Spine RFCoil Pad A
B
Foam Pad(s)
(as needed)
Tuning A : Saddle Coil
Ports B : Solenoid Coil
Figure 10. Coil Tuning Ports Locations
13.2.11 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.
13.2.12 Use the RF UNIT CONTROLLER program to set the TUNE value to 18.
Page 16 L5E-BMA106-5
13.2.13 Access the COIL DATA ENTRY program from the System Table Editor and select the QD
C-Spine coil. Select RECEPTION LINE 1 - B only:
Parameter Value
Coil Name QD C-Spine
Physical Coil No. 12
Attenuation Value 00
Tuning Measure Points 22
RF Gain Start Value 0e0
RF Gain Adjustment Step 16
Adjustment Average 1
Sensitivity Correction Coefficient 64
First Modulate Decoupling 64
Second Modulate Decoupling 64
Third Modulate Decoupling 2f
Fourth Modulate Decoupling 00
Selector 1
Number of Channels 1
Shading Correction No
Bed Movement No
Reception Line 1
QD Attenuation Value -2
Method A
13.2.14 Use the Impedance Meter and the half wave length cable to monitor the coil LINE B
impedance. The frequency of the Impedance Meter should remain at the system frequency
(f0) for the measurement.
13.2.15 Identify the tunable capacitor labeled “TUNE B” on the coil, to adjust the LINE B
impedance (refer to Figure 10 above). Tune the “TUNE B” capacitor so that the LINE B
impedance, at the system frequency, is specified as :
13.2.17 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.
Page 17 L5E-BMA106-5
13.3 Registration of Coil Data
13.3.1 Access the COIL DATA ENTRY program from the System Table Editor and select the QD
C-Spine coil. Set the ATT value to 2. Select RECEPTION LINE 3 - A+B:
Parameter Value
Coil Name QD C-Spine
Physical Coil No. 12
Attenuation Value 00
Tuning Measure Points 22
RF Gain Start Value 0e0
RF Gain Adjustment Step 16
Adjustment Average 1
Sensitivity Correction Coefficient 64
First Modulate Decoupling 64
Second Modulate Decoupling 64
Third Modulate Decoupling 2f
Fourth Modulate Decoupling 00
Selector 1
Number of Channels 1
Shading Correction No
Bed Movement No
Reception Line 3
QD Attenuation Value -2
Method A
13.3.2 Perform scanogram (coronal) to identify the center of the coil sensitivity. Move the table as
appropriate to align the center of coil sensitivity with the magnet isocenter.
13.3.3 Perform scanogram (transverse) to check vertical centering of the phantom. Adjust the
phantom height as necessary.
Page 18 L5E-BMA106-5
14.0 AIRIS II SNR Measurement
14.1 Calculate the SN ratio with the following sequence:
AIRIS II
Sequence SE
TR 500
TE 40
Slice Thickness 5
AVE 1
Projection 128
FOV 260
Bandwidth 7.2k
AMI 25%
14.2 Verify that the SN Ratio falls within the specified range according to the following table;
contact HMSA Technical Support if it does not:
14.3 Inspect the SNR image for asymmetrical shading or image artifacts; contact HMSA
Technical Support if any are noted.
Page 19 L5E-BMA106-5