10104514E00 - Approved Manual
10104514E00 - Approved Manual
TM
Puritan Bennett
980 Series Ventilator
Copyright Information
COVIDIEN, COVIDIEN with logo, and positive results for life are U.S and internationally registered
trademarks of COVIDIEN AG. All other brands are trademarks of a Covidien company or of their
respective owners.
©2013,2015, 2018 COVIDIEN.
The information contained in this manual is the sole property of Covidien and may not be dupli-
cated without permission. This manual may be revised or replaced by Covidien at any time and
without notice. Ensure this manual is the most current applicable version. If in doubt, contact
Covidien’s technical support department.
While the information set forth herein is believed to be accurate, it is not a substitute for the exer-
cise of professional judgment.
The ventilator should be operated and serviced only by trained professionals.
Covidien’s sole responsibility with respect to the ventilator and software, and its use, is as stated
in the limited warranty provided.
Nothing in this document shall limit or restrict in any way Covidien’s right to revise or otherwise
change or modify the equipment (including its software) described herein, without notice. In the
absence of an express, written agreement to the contrary, Covidien has no obligation to furnish
any such revisions, changes, or modifications to the owner or user of the equipment (including
its software) described herein.
Table of Contents
1 Introduction
2 Product Overview
i
2.6.5 Ventilator Side Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
2.7 Mounting Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2.8 Battery Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2.9 Graphical User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.9.1 Primary Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.10 GUI Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.10.1 Control Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.10.2 GUI Touch Screen Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.10.3 Visual Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.10.4 On-screen Symbols and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
2.10.5 Audible Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.11 Breath Delivery Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.11.1 BDU Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.11.2 Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-37
2.12 Additional Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-38
2.13 Special Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-38
2.14 Color Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-38
2.15 Pneumatic Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-39
3 Installation
ii
3.7.2 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
3.7.3 Stand-By State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
3.7.4 Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
3.8 Product Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
3.8.1 Preparing the Ventilator for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
3.8.2 Configuring the GUI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
3.9 Installation Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-43
3.9.1 SST (Short Self Test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-43
3.9.2 EST (Extended Self Test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-49
3.9.3 EST Test Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-52
3.9.4 EST Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-55
3.10 Operation Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-56
4 Operation
iii
4.9.1 Inspiratory Pause Maneuver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
4.9.2 Expiratory Pause Maneuver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
4.9.3 Other Respiratory Maneuvers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
4.10 Oxygen Sensor Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
4.10.1 Oxygen Sensor Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
4.10.2 Oxygen Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
4.10.3 Oxygen sensor calibration testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
4.11 Ventilator Protection Strategies . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.11.1 Power on Self Test (POST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.11.2 Technical Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.11.3 SST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.11.4 Procedure Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.11.5 Ventilation Assurance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.11.6 Safety Valve Open (SVO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.11.7 Ventilator Inoperative (Vent Inop) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.12 Ventilator Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
6 Performance
iv
6.3 Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
6.4 Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.4.1 Ventilation Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.4.2 Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.4.3 Breath Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6.5 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
6.5.1 Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.5.2 Alarm Reset Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.5.3 Audio Paused Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.5.4 Alarm Volume Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
6.5.5 Alarm Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
6.5.6 Viewing Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
6.5.7 Alarm Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
6.5.8 Alarm Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.5.9 AC POWER LOSS Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-33
6.5.10 APNEA Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-33
6.5.11 CIRCUIT DISCONNECT Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-33
6.5.12 LOSS OF POWER Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
6.5.13 DEVICE ALERT Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
6.5.14 HIGH CIRCUIT PRESSURE Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
6.5.15 HIGH DELIVERED O2% Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
6.5.16 HIGH EXHALED MINUTE VOLUME Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
6.5.17 HIGH EXHALED TIDAL VOLUME Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
6.5.18 HIGH INSPIRED TIDAL VOLUME Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
6.5.19 HIGH RESPIRATORY RATE Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-37
6.5.20 INSPIRATION TOO LONG Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-37
6.5.21 LOW CIRCUIT PRESSURE Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-37
6.5.22 LOW DELIVERED O2% Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-38
6.5.23 LOW EXHALED MANDATORY TIDAL VOLUME Alarm . . . . . . . . . . . . . . . 6-39
6.5.24 LOW EXHALED SPONTANEOUS TIDAL VOLUME Alarm . . . . . . . . . . . . . 6-39
6.5.25 LOW EXHALED TOTAL MINUTE VOLUME Alarm . . . . . . . . . . . . . . . . . . . . 6-39
6.5.26 PROCEDURE ERROR Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-40
6.5.27 SEVERE OCCLUSION Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-40
6.6 Monitored Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-40
6.6.1 Total Exhaled Minute Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-41
6.6.2 Exhaled Spontaneous Minute Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-41
6.6.3 Exhaled Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-41
6.6.4 Proximal Exhaled Minute Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-41
6.6.5 Proximal Exhaled Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
6.6.6 Exhaled Spontaneous Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
6.6.7 Exhaled Mandatory Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
v
6.6.8 Exhaled mL/kg Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
6.6.9 Inspired Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
6.6.10 Proximal Inspired Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
6.6.11 Delivered mL/kg Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-43
6.6.12 I:E Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-43
6.6.13 Mean Circuit Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-43
6.6.14 Peak Circuit Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-43
6.6.15 End Inspiratory Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-43
6.6.16 End Expiratory Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.17 Intrinsic PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.18 PAV-based Intrinsic PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.19 Total PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.20 Plateau Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.21 Total Respiratory Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.22 PAV-based Lung Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-44
6.6.23 PAV-based Patient Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-45
6.6.24 PAV-based Lung Elastance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-45
6.6.25 Spontaneous Rapid Shallow Breathing Index . . . . . . . . . . . . . . . . . . . . . . 6-45
6.6.26 Spontaneous Inspiratory Time Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-45
6.6.27 Spontaneous Inspiratory Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-45
6.6.28 PAV-based Total Airway Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-46
6.6.29 Static Compliance and Static Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-46
6.6.30 Dynamic Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-47
6.6.31 Dynamic Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-48
6.6.32 C20/C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-48
6.6.33 End Expiratory Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-48
6.6.34 Peak Spontaneous Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-48
6.6.35 Displayed O2% . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-48
7 Preventive Maintenance
vi
7.7 Service Personnel Preventive Maintenance . . . . . . . . . . . . . . . . . 7-21
7.8 Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
7.9 Inspection and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
7.10 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
7.11 Storage for Extended Periods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
8 Troubleshooting
9 Accessories
10 Theory of Operations
vii
10.7.3 VC+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-19
10.7.4 Rise time % . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-20
10.7.5 Manual Inspiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-21
10.8 Spontaneous Breath Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-21
10.8.1 Pressure Support (PS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-24
10.8.2 Volume Support (VS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-24
10.8.3 Tube Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-26
10.8.4 Proportional Assist Ventilation (PAV™+) . . . . . . . . . . . . . . . . . . . . . . . . . . .10-30
10.9 A/C Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-30
10.9.1 Changing to A/C Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-33
10.10 SIMV Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-33
10.10.1 Changing to SIMV Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-36
10.11 Spontaneous (SPONT) Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-38
10.11.1 Changing to SPONT Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-39
10.12 Apnea Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-39
10.12.1 Apnea Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-40
10.12.2 Transition to Apnea Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-41
10.12.3 Settings Changes During Apnea Ventilation . . . . . . . . . . . . . . . . . . . . . .10-42
10.12.4 Resetting Apnea Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-42
10.12.5 Apnea Ventilation in SIMV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-42
10.12.6 Phasing in New Apnea Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-43
10.13 Detecting Occlusion and Disconnect . . . . . . . . . . . . . . . . . . . . . . . 10-44
10.13.1 Occlusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-44
10.13.2 Disconnect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-45
10.13.3 Annunciating Occlusion and Disconnect Alarms . . . . . . . . . . . . . . . . . .10-47
10.14 Respiratory Mechanics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-47
10.14.1 Inspiratory Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-49
10.14.2 Expiratory Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-51
10.14.3 Negative Inspiratory Force (NIF) Maneuver . . . . . . . . . . . . . . . . . . . . . . . .10-52
10.14.4 P0.1 Maneuver (Occlusion Pressure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-53
10.14.5 Vital Capacity (VC) Maneuver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-54
10.15 Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-55
10.15.1 Apnea Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-55
10.15.2 Circuit Type and Predicted Body Weight (PBW) . . . . . . . . . . . . . . . . . . .10-56
10.15.3 Vent Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-57
10.15.4 Mode and Breath Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-58
10.15.5 Respiratory Rate (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-59
10.15.6 Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-60
10.15.7 Peak Inspiratory Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-60
10.15.8 Plateau Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-61
10.15.9 Flow Pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-61
viii
10.15.10Flow Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-61
10.15.11Pressure Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-62
10.15.12Inspiratory Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-62
10.15.13Inspiratory Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-63
10.15.14Expiratory Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.15I:E Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.16High Pressure in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.17Low Pressure in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.18High Time in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.19Low Time in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-65
10.15.20TH:TL Ratio in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-65
10.15.21PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-65
10.15.22Pressure Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.23Volume Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.24% Supp in TC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.25% Supp in PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.26Rise Time % . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-67
10.15.27Expiratory Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-67
10.15.28Disconnect Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-68
10.15.29High Spontaneous Inspiratory Time Limit . . . . . . . . . . . . . . . . . . . . . . . . .10-68
10.15.30Humidification Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-69
10.15.31Humidifier Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-69
10.16 Safety Net . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-69
10.16.1 User Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-70
10.16.2 Patient Related Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-71
10.16.3 System Related Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-71
10.16.4 Background Diagnostic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-72
10.17 Power On Self Test (POST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-74
10.18 Short Self Test (SST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-75
10.19 Extended Self Test (EST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-75
11 Specifications
ix
11.8 Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-29
11.9 Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-31
11.10 Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-36
11.11 Essential Performance Requirements . . . . . . . . . . . . . . . . . . . . . . 11-36
A BiLevel 2.0
B Leak Sync
C PAV™+
x
C.2 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
C.3 Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C.4 PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
C.4.1 Setting Up PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
C.4.2 PBW and Tube ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6
C.4.3 Apnea Parameters Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-7
C.4.4 Alarm Settings Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-7
C.4.5 PAV+ Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
C.4.6 PAV+ Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
C.4.7 Monitored Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
C.4.8 PAV+ Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-9
C.5 Ventilator Settings/Guidance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-10
C.5.1 Specified Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-10
C.5.2 Graphics Displays in PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-11
C.5.3 WOB Terms and Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-11
C.5.4 Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-13
C.5.5 Protection Against Hazard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-16
D NeoMode 2.0
E Proximal Flow
Glossary
Index
xi
Page Left Intentionally Blank
xii
List of Tables
xiii
Table 7-5. Component Sterilization Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Table 7-6. Service Preventive Maintenance Frequency . . . . . . . . . . . . . . . . . . . . 7-21
Table 9-1. Accessories and Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Table 10-1. Compliance Volume Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-15
Table 10-2. Maximum Pressure Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-20
Table 10-3. Spontaneous Breath Delivery Characteristics . . . . . . . . . . . . . . . . . .10-22
Table 10-4. Maximum Pressure Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-25
Table 10-5. Inspiratory and Expiratory Pause Events . . . . . . . . . . . . . . . . . . . . . . .10-50
Table 10-6. Values for VT Based on Circuit Type . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-56
Table 10-7. Peak Flow and Circuit Type (Leak Sync Disabled) . . . . . . . . . . . . . .10-57
Table 10-8. Modes and Breath Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-58
Table 10-9. Safety PCV Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-70
Table 10-10. Inspiratory Backup Ventilation Settings . . . . . . . . . . . . . . . . . . . . . . .10-73
Table 11-1. Performance Verification Equipment Uncertainty . . . . . . . . . . . . . . . 11-1
Table 11-2. Physical Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Table 11-3. Pneumatic Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Table 11-4. Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Table 11-5. Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Table 11-6. Interface Pin Designations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Table 11-7. Nurse Call Pin Designations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-8
Table 11-8. Environmental Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-8
Table 11-9. Ventilator Settings Range and Resolution . . . . . . . . . . . . . . . . . . . . . . 11-9
Table 11-10. Alarm Settings Range and Resolution . . . . . . . . . . . . . . . . . . . . . . . . .11-17
Table 11-11. Patient Data Range and Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . .11-20
Table 11-12. Delivery Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-27
Table 11-13. Monitoring (Patient Data) Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . .11-27
Table 11-14. Computed Value Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-28
Table 11-15. Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-32
Table 11-16. Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-32
Table 11-17. Recommended Separation Distances for RF . . . . . . . . . . . . . . . . . .11-35
Table 11-18. Recommended Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-36
Table A-1. Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Table B-1. Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Table B-2. Maximum Leak Compensation Flow Based on Patient Type . . . . . B-3
Table B-3. DSENS settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
Table C-1. Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Table C-2. Absolute limits for PAV+ Monitored Data . . . . . . . . . . . . . . . . . . . . . . . . C-9
Table C-3. PAV+ Work of Breathing terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-12
Table C-4. Default PBW-based Resistance Values . . . . . . . . . . . . . . . . . . . . . . . . . .C-21
Table E-1. Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Table E-2. Proximal Flow Option Data Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-7
xiv
Table E-3. Proximal Flow Option SST test Sequence . . . . . . . . . . . . . . . . . . . . . . . . E-9
Table E-4. Proximal Flow Sensor Volume Accuracy . . . . . . . . . . . . . . . . . . . . . . . . E-16
Table E-5. Proximal Flow Sensor Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-16
Table E-6. Proximal Flow Option and Component Part Numbers . . . . . . . . . . E-16
Table Glossary-1. Glossary of Ventilation Terms . . . . . . . . . . . . . . . . . . . . . . . . . . Glossary-1
Table Glossary-2. Units of Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Glossary-9
Table Glossary-3. Technical Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Glossary-9
xv
Page Left Intentionally Blank
xvi
List of Figures
xvii
Figure 4-13. Respiratory Maneuver Tabs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Figure 4-14. More Settings Screen with O2 Sensor Enabled . . . . . . . . . . . . . . . . . . 4-32
Figure 5-1. Incompatible USB Device Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Figure 5-2. Comm Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Figure 5-3. Port Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Figure 6-1. Alarm Message Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Figure 7-1. EVQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Figure 7-2. EVQ Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Figure 7-3. EVQ Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Figure 7-4. Exhalation Valve Diaphragm Removal . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Figure 7-5. Exhalation Filter Seal Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Figure 7-6. Pressure Sensor Filter Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Figure 7-7. Immersion Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Figure 7-8. EVQ Reprocessing Kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-15
Figure 7-9. Installing the Pressure Sensor Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
Figure 7-10. Installing the Exhalation Filter Seal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
Figure 7-11. Installing the Diaphragm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
Figure 7-12. Installing the EVQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
Figure 8-1. Log Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Figure 9-1. Ventilator with Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Figure 9-2. Additional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Figure 10-1. Inspiration Using Pressure Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Figure 10-2. Inspiration Using Flow Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
Figure 10-3. Breath Activity During Time-triggered Inspiration. . . . . . . . . . . . . . . 10-8
Figure 10-4. Exhalation via the Airway Pressure Method . . . . . . . . . . . . . . . . . . . . . 10-9
Figure 10-5. Exhalation via the Percent Peak Flow Method . . . . . . . . . . . . . . . . .10-10
Figure 10-6. Square Flow Pattern. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-12
Figure 10-7. Descending Ramp Flow Pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-13
Figure 10-8. Ideal Waveform Using Square Flow Pattern. . . . . . . . . . . . . . . . . . . .10-17
Figure 10-9. Ideal Waveform Using Descending Ramp Flow Pattern . . . . . . . .10-18
Figure 10-10. Ideal Waveform Using Pressure Control Ventilation . . . . . . . . . . . .10-19
Figure 10-11. ET Tube Target Pressure vs. Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-29
Figure 10-12. Tracheostomy Tube Target Pressure vs. Flow . . . . . . . . . . . . . . . . . .10-30
Figure 10-13. No Patient Inspiratory Effort Detected . . . . . . . . . . . . . . . . . . . . . . . . .10-31
Figure 10-14. Patient Inspiratory Effort Detected. . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-32
Figure 10-15. Combined VIM and PIM Breaths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-32
Figure 10-16. Mandatory and Spontaneous Intervals . . . . . . . . . . . . . . . . . . . . . . . .10-33
Figure 10-17. PIM Delivered Within Mandatory Interval . . . . . . . . . . . . . . . . . . . . . .10-34
Figure 10-18. PIM Not Delivered Within Mandatory Interval. . . . . . . . . . . . . . . . . .10-34
Figure 10-19. Apnea Interval Equals Breath Period . . . . . . . . . . . . . . . . . . . . . . . . . . .10-40
Figure 10-20. Apnea Interval Greater Than Breath Period . . . . . . . . . . . . . . . . . . . .10-41
xviii
Figure 10-21. Apnea Interval Less Than Breath Period. . . . . . . . . . . . . . . . . . . . . . . .10-41
Figure 10-22. Apnea Ventilation in SIMV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-43
Figure A-1. Spontaneous Breathing at PL. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Figure A-2. BiLevel Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Figure A-3. BiLevel Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Figure A-4. BiLevel with Pressure Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Figure A-5. Spontaneous and Synchronous Intervals . . . . . . . . . . . . . . . . . . . . . . . . A-8
Figure A-6. APRV With Spontaneous Breathing at PH . . . . . . . . . . . . . . . . . . . . . . . . A-9
Figure B-1. Enabling Leak Sync. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Figure B-2. GUI Screen when Leak Sync is Enabled . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
Figure B-3. Leak Sync Monitored Patient Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7
Figure B-4. Circuit Disconnect During VC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-9
Figure C-1. Ventilator Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
Figure C-2. Graphics displays in PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-12
Figure C-3. Use of Default Lung Resistance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-20
Figure E-1. Proximal Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
Figure E-2. Sample GUI screen Showing Proximal Flow Data . . . . . . . . . . . . . . . . E-6
Figure E-3. Message During Autozero and Purge Processes . . . . . . . . . . . . . . . . . E-8
Figure E-4. Attaching Proximal Flow Sensor to Ventilator. . . . . . . . . . . . . . . . . . . E-10
Figure E-5. Enabling/disabling Proximal Flow Sensor. . . . . . . . . . . . . . . . . . . . . . . E-11
Figure E-6. Attaching Proximal Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-13
Figure E-7. Manual Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-14
Figure E-8. Alarm Message — Prox Inoperative . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-15
xix
Page Left Intentionally Blank
xx
1 Introduction
1.1 Overview
This manual contains information for operating the Puritan Bennett™ 980 Series
Ventilators. Before operating the ventilator system, thoroughly read this manual.
To obtain an additional copy of this manual, contact Covidien Customer Service or
your local representative.
Covidien makes available all appropriate information relevant to use and service of
the ventilator. For further assistance, contact your local Covidien representative.
• The Puritan Bennett™ 980 Series Ventilator Operator’s Manual — Provides basic
information on operating the ventilator and troubleshooting errors or malfunctions.
Before using the ventilator, thoroughly read this manual.
• The Puritan Bennett™ 980 Series Ventilator Service Manual — Provides information
to Covidien-trained service technicians for use when testing, troubleshooting, repairing,
and upgrading the ventilator.
1-1
Introduction
Symbol Description
Serial number
Part number
Authorized representative
Manufacturer
This side up
Fragile
Symbol Description
Keep dry
CSA certification mark that signifies the product has been evaluated to the
applicable ANSI/Underwriters Laboratories Inc. (UL) and CSA standards for
use in the US and Canada.
This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
WARNING:
Explosion hazard — Do not use in the presence of flammable gases. An oxygen-rich
environment accelerates combustibility.
WARNING:
To avoid a fire hazard, keep all components of the system away from all sources of
ignition (such as matches, lighted cigarettes, flammable medical gases, and/or
heaters). Oxygen-rich environments accelerate combustibility.
WARNING:
In case of fire or a burning smell, immediately take the following actions if it is safe to
do so: disconnect the patient from the ventilator and disconnect the ventilator from
the oxygen supply, facility power, and all batteries. Provide alternate method of
ventilatory support to the patient, if required.
WARNING:
Replacement of ventilator batteries by inadequately trained personnel could result
in an unacceptable risk, such as excessive temperatures, fire, or explosion.
WARNING:
To minimize fire hazard, inspect and clean or replace, as necessary, any damaged
ventilator parts that come into contact with oxygen.
WARNING:
To prevent electrostatic discharge (ESD) and potential fire hazard, do not use
antistatic or electrically conductive hoses or tubing in or near the ventilator
breathing system.
WARNING:
To ensure proper operation and avoid the possibility of physical injury, only qualified
medical personnel should attempt to set up the ventilator and administer treatment
with the ventilator.
WARNING:
In case of ventilator failure, the lack of immediate access to appropriate alternative
means of ventilation can result in patient death. An alternative source of ventilation,
such as a self-inflating, manually-powered resuscitator (as specified in ISO 10651-4
with mask) should always be available when using the ventilator.
WARNING:
Patients on mechanical ventilation should be monitored by clinicians for proper
patient ventilation.
WARNING:
The ventilator system is not intended to be a comprehensive monitoring device and
does not activate alarms for all types of conditions. For a detailed understanding of
ventilator operations, be sure to thoroughly read this manual before attempting to
use the ventilator system.
WARNING:
To prevent patient injury, do not use the ventilator if it has a known malfunction.
Never attempt to override serious malfunctions. Replace the ventilator and have the
faulty unit repaired by trained service personnel.
WARNING:
To prevent patient injury, do not make unauthorized modifications to the ventilator.
WARNING:
To prevent injury and avoid interfering with ventilator operation, do not insert tools
or any other objects into any ventilator openings.
WARNING:
The audio alarm volume level is adjustable. The operator should set the volume at a
level that allows the operator to distinguish the audio alarm above background
noise levels. Reference To adjust alarm volume, p. 3-39 for instructions on alarm
volume adjustment.
WARNING:
Do not silence, disable, or decrease the volume of the ventilator’s audible alarm if
patient safety could be compromised.
WARNING:
If increased pressures are observed during ventilation, it may indicate a problem
with the ventilator. Check for blocked airway, circuit occlusion,
and/or run SST.
WARNING:
The LCD panel contains toxic chemicals. Do not touch broken LCD panels. Physical
contact with a broken LCD panel can result in transmission or ingestion of toxic
substances.
WARNING:
If the Graphical User Interface (GUI) display/LCD panel is blank or experiences
interference and cannot be read, check the patient, then verify via the status display
that ventilation is continuing as set. Because breath delivery is controlled
independently from the GUI, problems with the display will not, by themselves, affect
ventilation. The ventilator, however, should be replaced as soon as possible and
repaired by qualified service personnel.
WARNING:
The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as
indicated, very limited exposure to trace amounts of phthalates may occur. There is
no clear clinical evidence that this degree of exposure increases clinical risk.
However, in order to minimize risk of phthalate exposure in children and nursing or
pregnant women, this product should only be used as directed.
WARNING:
Even though the 980 Series Ventilator meets the standards listed in Chapter 11, the
internal Lithium-ion battery of the device is considered to be Dangerous Goods (DG)
Class 9 - Miscellaneous, when transported in commerce. The 980 Series Ventilator
and/or the associated Lithium-ion battery are subject to strict transport conditions
under the Dangerous Goods Regulation for air transport (IATA: International Air
Transport Association), International Maritime Dangerous Goods code for sea and
the European Agreement concerning the International Carriage of Dangerous Goods
by Road (ADR) for Europe. Private individuals who transport the device are excluded
from these regulations although for air transport some requirements may apply.
WARNING:
Do not position the ventilator next to anything that blocks or restricts the gas inlet or
cooling air circulation openings, gas exhaust port, fan intake, or alarm speaker, as
this may:
• limit the air circulation around the ventilator, potentially causing overheating;
• limit the ventilator's ability to exhaust patient exhaled gas leading to potential
harm;
WARNING:
To avoid injury, do not position the ventilator in a way that makes it difficult to
disconnect the patient.
WARNING:
To ensure proper operation, do not position the ventilator in a way that makes it
difficult to access the AC power cord.
WARNING:
Do not use the ventilator in a hyperbaric chamber. It has not been validated for use
in this environment.
WARNING:
Do not use the ventilator in the presence of strong magnetic fields. Doing so could
cause a ventilator malfunction.
WARNING:
Do not use the ventilator during radiotherapy (i.e. cancer treatment using ionizing
radiation), as doing so could cause a ventilator malfunction.
WARNING:
To avoid the risk of ventilator malfunction, operate the ventilator in an environment
that meets specifications. Reference Environmental Specifications, p. 11-8.
WARNING:
Do not use the ventilator as an EMS transport ventilator. It has not been approved or
validated for this use.
WARNING:
Before activating any part of the ventilator, be sure to check the equipment for
proper operation and, if appropriate, run SST as described in this manual. Reference
To run SST, p. 3-45.
WARNING:
Check for leaks in the ventilator breathing system by running SST prior to ventilating
a patient.
WARNING:
Lock the ventilator’s casters during use to avoid the possibility of extubation due to
inadvertent ventilator movement.
WARNING:
The ventilator accuracies listed in the Ventilator Settings, Alarm Settings, and Patient
Data tables in Chapter 11 are applicable under specified operating conditions.
Reference Environmental Specifications, p. 11-8. If the ventilator is operated outside
specified ranges, the ventilator may supply incorrect information and the accuracies
listed in the aforementioned tables do not apply. A hospital Biomedical Technician
must verify the ventilator is operated in the environmental conditions specified.
WARNING:
To avoid the risk of electrical shock:
• Use only Covidien-branded batteries, adapters, and cables.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient’s
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient, based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
WARNING:
Avoid nuisance alarms by applying appropriate alarm settings.
WARNING:
To prevent inappropriate ventilation, select the correct Tube Type (ET or
Tracheostomy) and tube inner diameter (ID) for the patient’s ventilatory needs.
Inappropriate ventilatory support leading to over-or under-ventilation could result
if an ET tube or trach tube setting larger or smaller than the actual value is entered.
WARNING:
Setting expiratory volume alarms to OFF increases the risk of not detecting a low
returned volume.
WARNING:
Setting any alarm limits to OFF or extreme high or low values can cause the
associated alarm not to activate during ventilation, which reduces its efficacy for
monitoring the patient and alerting the clinician to situations that may require
intervention.
WARNING:
To prevent electrostatic discharge (ESD) and potential fire hazard, do not use
antistatic or electrically conductive hoses or tubing in or near the ventilator
breathing system.
WARNING:
Adding accessories to the ventilator can change the pressure gradient across the
ventilator breathing system (VBS) and affect ventilator performance. Ensure that any
changes to the ventilator circuit configurations do not exceed the specified values
for circuit compliance and for inspiratory or expiratory limb total resistance.
Reference Technical Specifications, p. 11-4. If adding accessories to the patient circuit,
always run SST to establish circuit compliance and resistance prior to ventilating the
patient.
WARNING:
Use of a nebulizer or humidifier can lead to an increase in the resistance of
inspiratory and exhalation filters. Monitor the filters frequently for increased
resistance or blockage.
WARNING:
During transport, the use of breathing tubing without the appropriate cuffed
connectors may result in the circuit becoming detached from the ventilator.
WARNING:
The added gas from an external pneumatic nebulizer can adversely affect
spirometry, delivered O2%, delivered tidal volumes, and breath triggering.
Additionally, aerosolized particulates in the ventilator circuit can lead to an increase
in exhalation filter resistance.
WARNING:
Carefully route patient tubing and cabling to reduce the possibility of patient
entanglement or strangulation.
WARNING:
Always use filters designed for use with the Puritan Bennett™ 980 Series Ventilator.
Do not use filters designed for use with other ventilators. Reference Accessories and
Options, p. 9-3 for relevant filter part numbers.
WARNING:
To avoid liquid entering the ventilator, empty the expiratory condensate vial before
fluid reaches the maximum fill line.
WARNING:
Accessory equipment connected to the analog and digital interfaces must be
certified according to IEC 60601-1. Furthermore, all configurations shall comply with
the system standard IEC 60601-1-1. Any person who connects additional equipment
to the signal input part or signal output part of the ventilator system configures a
medical system, and is therefore responsible for ensuring the system complies with
the requirements of the system standard IEC 60601-1-1. If in doubt, consult Covidien
Technical Services at 1.800.255.6774 or your local representative.
WARNING:
Do not use HMEs (heat and moisture exchangers) and heated humidifiers together.
This may result in the HME absorbing water and becoming obstructed, resulting in
high airway pressures.
WARNING:
Do not use nitric oxide, helium or mixtures containing helium with the ventilator. It
has not been validated for use with these gas mixtures.
WARNING:
To avoid the risk of ventilator malfunction, do not use the ventilator with anesthetic
gases.
WARNING:
For proper ventilator operation, use only clean, dry, medical grade gases when
ventilating a patient.
WARNING:
Use of only one gas source could lead to loss of ventilation and/or hypoxemia if that
one gas source fails and is not available. Therefore, always connect at least two gas
sources to the ventilator to ensure a constant gas supply is available to the patient in
case one of the gas sources fails. The ventilator has two connections for gas sources:
air inlet and oxygen inlet.
WARNING:
Use of the ventilator in altitudes higher or barometric pressures lower than those
specified could compromise ventilator operation. Reference Environmental
Specifications, p. 11-8 for a complete list of environmental specifications.
WARNING:
The ventilator should be connected to a gas pipeline system compliant to
ISO 7396-1:2007 because:
• Installation of the ventilator on a non-ISO 7396-1:2007 compliant gas pipeline
system may exceed the pipeline design flow capacity.
• The ventilator is a high-flow device and can interfere with the operation of other
equipment using the same gas source if the gas pipeline system is not compliant
to ISO 7396-1:2007.
WARNING:
Patients receiving mechanical ventilation may experience increased vulnerability to
the risk of infection. Dirty or contaminated equipment is a potential source of
infection. It is recognized that cleaning, sterilization, sanitation, and disinfection
practices vary widely among health care institutions. Always follow your hospital
infection control guidelines for handling infectious material. Follow the instructions
in this manual and your institution’s protocol for cleaning and sterilizing the
ventilator and its components. Use all cleaning solutions and products with caution.
Follow manufacturer’s instructions for individual cleaning solutions. Reference
Chapter 7 of this manual.
WARNING:
To prevent infection and contamination, always ensure inspiratory and exhalation
bacteria filters are installed before ventilating the patient.
WARNING:
Never attempt to re-use single patient use components or accessories. Doing so
increases risk of cross-contamination and re-processing of single patient use
components or accessories may compromise functionality leading to possible loss of
ventilation.
WARNING:
To ensure proper operation and avoid the possibility of physical injury, this
ventilator should only be serviced by qualified technicians who have received
appropriate Covidien-provided training for the maintenance of this ventilator.
WARNING:
Follow preventive maintenance according to specified intervals. Reference Operator
Preventive Maintenance Frequency, p. 7-2. Reference Service Preventive Maintenance
Frequency, p. 7-21.
1.3.12 Cautions
Caution:
To prevent possible equipment damage, ensure the casters are locked to prevent
inadvertent movement of the ventilator during routine maintenance, or when the
ventilator is on an incline.
Caution:
Do not use sharp objects to make selections on the display or keyboard.
Caution:
To ensure optimal performance, keep the GUI touch screen and keyboard clean and
free from foreign substances. Reference Surface Cleaning Agents, p. 7-5.
Caution:
To avoid moisture entering the ventilator and possibly causing a malfunction,
Covidien recommends using a wall air water trap when using piped medical air from
a facility-based air compressor.
Caution:
Use only the cleaning agents specified.
Reference Surface Cleaning Agents, p. 7-5. for approved cleaning agents.
Caution:
Clean compressor inlet filter according to the interval listed in Chapter 7. Reference
Operator Preventive Maintenance Frequency, p. 7-2.
Caution:
Do not block cooling vents.
Caution:
Ensure proper connection and engagement of exhalation and inspiratory filters.
Caution:
Follow instructions for proper GUI and BDU (breath delivery unit) mounting as
described in the Puritan Bennett™ 980 Series Ventilator Installation Instructions.
Caution:
Follow proper battery installation instructions as described in this manual.
Caution:
When transferring the ventilator from storage conditions, allow its temperature to
stabilize at ambient conditions prior to use.
Caution:
Remove extended and primary batteries from ventilator prior to transporting in a
vehicle. Failure to do so could result in damage to the ventilator.
1.3.13 Notes
Note:
When using non-invasive ventilation (NIV), the patient’s actual exhaled volume may differ
from the exhaled volume reported by the ventilator due to leaks around the mask.
Note:
When utilizing a closed-suction catheter system, the suctioning procedure can be executed
using existing mode, breath type, and settings. To reduce potential for hypoxemia during
the procedure, elevated delivered oxygen can be enabled using the Elevate O2 control.
Reference To adjust the amount of elevated O2 delivered for two minutes, p. 3-37.
Covidien France SAS Covidien Hong Kong Covidien India Covidien ECE
2 Rue Denis Diderot Unit 12 - 16, 18/F 10th Floor Building No 9B Mariássy u.7.
78990 Elancourt BEA Tower DLF Cyber City Phase III 1095 Budapest
France Millennium City 5 Gurgaon Hungary
Tel +33 (0) 13079 80 4187 Kwun Tong Road Haryana - 122002 Tel + 36 1880 7975
00 Kwum Tong, India Fax + 36 1777 4932
Fax +33 (0) 130 79 80 Kowloon, Hong Kong Tel + 91 1244 709800
Tel + 852 3157 7299 Fax + 91 1244 206850
30
Fax + 852 2838 0749
Covidien Ireland Covidien Israel Covidien Italia S.p.A Covidien Japan Inc.
Block G, Ground Floor, 5, Shasham St. Via Rivoltana 2/D Technical Support Center
Cherrywood Business North Industrial Park I-20090 Segrate (Mi) 83-1, Takashimadaira 1-
Park, POB3069 Italy Chome
Loughlinstown Caesarea, 38900 Tel +39 02 703 173 1 Itabashi-ku, Tokyo 175-0082
County Dublin, Ireland Tel +972 4.627 73 88 Fax +39 02 71740584 Japan
Tel +353 (0) 1.4073173 Fax+972 4.627 76 88 Tel: +81 (0) 3 6859 0120
Fax +353(0) 1.4073174 Fax: +81 (0) 3 6859 0142
Covidien Puerto Rico Covidien Russia Covidien Saglik A.S. Covidien South Africa
Palmas Industrial Park 53 bld. 5 Dubininskaya Maslak Mahallesi Bilim Corporate Park North
Road 869 Km 2.0 Bdlg. StreetMoscow Sokak No: 5, Sun Plaza Kat: 379 Roan Crescent
#1 RUSSIA. 119054 2-3 RandjesparkMidrand, South
Cataño, PR 00962 Tel +70 495 933 64 69 Sisli, Istanbul 34398 Africa
Tel. 787-993-7250 Fax +70 495 933 64 68 Turkey Tel +27 115 429 500
Ext. 7222 & 7221 Tel +90 212 366 20 00 Fax +27 115 429 624
Fax 787-993-7234 Fax +90 212 276 35 25
For online Technical support, visit the SolvITSM Center Knowledge Base at
www.covidien.com. The SolvIT Center provides answers to frequently asked ques-
tions about the ventilator system and other Puritan Bennett products 24 hours a day,
seven days a week.
The ventilator is equipped with an on-screen help system that enables users to
select an item on the screen and display a description of that item. Follow the pro-
cedure below to access and use on-screen help.
Help topics on the ventilator are called tooltips. If a tooltip is available, a glowing blue
outline appears around the item in question.
To access tooltips
1. Touch the item in question for a period of at least 0.5 s, or drag the help icon (the question
mark icon appearing at the lower right of the GUI screen) to the item in question. A tooltip
appears with a short description of the item. Most screen items have tooltips associated with
them, providing the operator with access to a multitude of help topics.
3. Touch “close” to close the dialog, or let it fade away after five (5) s
Note:
• Dragging the help icon causes the tooltip to display in its unexpanded state.
• Dragging the help icon and pausing causes a tooltip to display. Continue dragging to
another item to dismiss the last tooltip and display another tooltip.
Other Resources
Additional resources for information about the ventilator can be found in the Puritan
Bennett™ 980 Series Ventilator Service Manual and appendices in this manual for
BiLevel 2.0, Leak Sync, PAV+, NeoMode 2.0, and Proximal Flow Sensor options.
• Z represents the product code (B= breath delivery unit, G= GUI, C = Compressor, P=
Proximal Flow Monitoring option. The product codes shown here are typically the most
common. There may be other product codes shown in the serial number depending
upon the particular option(s) purchased.
• XXXXX is a sequential number that resets at the beginning of each new year
Serial numbers are located on labels on the back panels of the GUI and BDU, and in
various locations on product options.
1.7 Manufacturer
Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.
Covidien Ireland Limited, IDA Business &Technology Park, Tullamore.
The Puritan Bennett™ 980 ventilator requires special precautions to be taken regard-
ing electromagnetic compatibility (EMC) and must be installed and put into service
according to the EMC information provided in Chapter 11 in this manual.
2.1 Overview
This chapter contains introductory information for the Puritan Bennett™ 980 Series
Ventilator.
Note:
Items shown in bold-italic font are contained as entries in the glossary.
Communication between the ventilator’s graphical user interface (GUI) and the
breath delivery unit (BDU) occurs continuously via independent central process-
ing units (CPUs).
Reference Pneumatic Diagram (Compressor Shown), p. 2-40 and its associated refer-
ence designators when reading the following paragraphs.
Gas delivery starts with the ventilator connected to wall (or bottled) air and oxygen.
Gas travels to the mix module where gas pressures are regulated by their respective
proportional solenoid valves (PSOLs). The PSOLs meter the gases according to the
ventilator settings entered, then the gases flow through individual air and oxygen
flow sensors into the mix manifold and accumulator for mixing. The individual gas
pressures are continuously monitored both before and after they are mixed in the
mix manifold and accumulator assemblies. The mixed gas then flows to the inspira-
tory pneumatic system where it flows through the breath delivery flow sensor and
then the inspiratory PSOL for delivery to the patient.
Before the gas reaches the patient, it passes through an internal inspiratory bacteria
filter, then through an external inspiratory bacteria filter attached to the ventilator’s
gas outlet (the To patient port) where the breathing circuit is attached. When the
gas returns from the patient, it flows through the expiratory limb of the breathing
circuit, to the From patient port on the exhalation bacteria filter (which includes a
condensate vial) before flowing through the exhalation flow sensor and exhalation
valve (EV). A gas exhaust port allows exhaled gas to exit the ventilator and flow to
the room.
The ventilator recognizes the patient’s breathing effort using pressure triggering
(PTRIG) or flow triggering (VTRIG). During pressure triggering, as the patient inhales,
2-1
Product Overview
the airway pressure decreases and the inspiratory pressure transducer (PI) monitors
this pressure decrease. When the pressure drops to at least the value of the pressure
sensitivity (PSENS) setting, the ventilator delivers a breath. During flow triggering, the
difference between inspiratory and expiratory flows is monitored. As the patient
inhales, the exhalation flow sensor measures less flow, while the delivery flow sensor
measurement remains constant. When the difference between the two measure-
ments is at least the value of the operator-set flow sensitivity (VSENS), the ventilator
delivers a breath. If the patient is not inhaling, any difference between delivered flow
and expiratory flow is due to flow sensor inaccuracy or leaks in the ventilator breath-
ing circuit. To compensate for leaks, which can cause autotriggering, the clinician
can increase the VSENS setting or enable Leak Sync, if available.
Note:
Leak Sync is a software option. Details on its operation are provided in the Leak Sync
appendix in this manual.
A backup pressure triggering threshold of 2 cmH2O is also in effect. This provides
enough pressure sensitivity to avoid autotriggering, but will still allow the ventilator
to trigger with acceptable patient effort.
The exhalation valve controls Positive End Expiratory Pressure (PEEP) using feed-
back from the expiratory pressure transducer (PE). the valve controller also cycles
the ventilator into the exhalation phase if the PE measurement equals or exceeds
the operator-set high circuit pressure limit. The PE measurement also controls when
the safety valve (SV) opens. If PE measures 110 cmH2O or more in the ventilator
breathing circuit, the safety valve opens, allowing the patient to breathe room air
through the valve.
• Puritan Bennett™ 980 Neonatal Ventilator — The Neonatal model ventilates neona-
tal patients with predicted body weights from 0.3 kg to 7.0 kg, and with tidal volumes
for mandatory volume-controlled breaths from 2 mL to 320 mL.
• Puritan Bennett™ 980 Universal Ventilator — The Universal model ventilates neona-
tal, pediatric, and adult patients with predicted body weights from 0.3 kg to 150 kg, and
with tidal volumes for mandatory volume-controlled breaths from 2 mL to 2500 mL.
• Type BF
• Mobile
• Internally powered
• IP 21 equipment
• Continuous operation
• Not suitable for use with flammable medical gases (not AP or APG)
Reference BDU Rear Label or Panel Symbols and Descriptions, p. 2-11 for a description
of the meaning of the IP classification.
The ventilator system uses a graphical user interface (GUI) and breath delivery
unit (BDU) for entering patient settings and delivering breaths to the patient. The
GUI contains electronics capable of transferring the clinician’s input (by touching
the screen) to the BDU where pneumatic and electronic systems generate the
breathing parameters.
Note:
Intended typical usage may be defined to include the following for the ventilator system
Hospital Use — Typically covers areas such as general care floors (GCFs), operating rooms,
special procedure areas, intensive and critical care areas within the hospital and in hos-
pital-type facilities. Hospital-type facilities include physician office-base facilities, sleep
labs, skilled nursing facilities, surgicenters, and sub-acute centers.
2.4 Contraindications
Do not operate the ventilator in a magnetic resonance imaging (MRI) environment.
Note:
No parts of the ventilator system contain latex.
Note:
The components in the gas pathway that can become contaminated with bodily fluids or
expired gases during both normal and single fault conditions are:
• External inspiratory filter
The typical ventilator system ships with the following packing list. Depending upon
the ventilator system purchased, your list may vary.
Quantity Item
1 Inspiratory filter
1 Exhalation filter
1 Condensate vial
1 Power cord
1 Operator’s Manual CD
1 Flex arm
1 Drain bag
Reference Common Symbols found on GUI or BDU Labels, p. 2-13 for symbols found
on the GUI or BDU. The “Do Not Push” symbol found on the GUI, only, is shown in
this table.
Symbol Description
To Patient port
Software option labels are applied to the grid located on the back of the ventilator,
as shown below and in the previous image (item 4).
The following table lists the symbols and descriptions found on BDU or base labels.
Symbol Description
User must consult instructions for use. Symbol is also found on “Do not
obstruct” labels on both left and right sides of the ventilator, and on label indi-
cation supply gas connections.
Keep away from fire or flame. Oxygen rich environments accelerate combusti-
bility.
Table 2-3. BDU Rear Label or Panel Symbols and Descriptions (Continued)
Symbol Description
Authorized to bear the CSA certification mark signifying the product has been
evaluated to the applicable ANSI/Underwriters Laboratories Inc. (UL) and CSA
standards for use in the US and Canada.
Table 2-3. BDU Rear Label or Panel Symbols and Descriptions (Continued)
Symbol Description
Symbol Description
Authorized representative.
Serial number.
The GUI incorporates a 15” display that rotates 170° about a vertical axis in either
direction. The GUI can also be tilted up to 45° from vertical.
The clinician enters ventilation parameters via the GUI’s touch screen, also known as
the ventilator’s primary display. The GUI’s keys activate other ventilator functions
including screen brightness, display lock, alarm volume, manual inspiration, inspira-
tory pause, expiratory pause, alarm reset, and audio paused.
The GUI displays the following information depending on the state of the ventilator:
• Ventilator, apnea, and alarm settings
• Patient data
• Waveforms
The GUI bezel has eight off-screen control keys as shown below.
Brightness control key — Adjusts the GUI screen brightness. Press the key and turn the knob
to adjust the brightness.
Display lock key — Actuates a lock to prevent inadvertent settings changes to the ventilator
(including the knob function) while the display is locked. The display lock is useful when
cleaning the touch screen. Press the key again to unlock the display.
Also use the display lock key to reset the GUI touch screen as described in GUI Touch Screen
Reset (2.10.2).
Alarm volume key — Adjusts the alarm volume. The alarm volume cannot be turned OFF.
Manual inspiration key — In A/C, SIMV, and SPONT modes, delivers one manual breath to
the patient in accordance with the current mandatory breath parameters. In BiLevel mode,
transitions from low pressure (PL) to high pressure PH) (or vice versa). To avoid breath stack-
ing, a manual inspiration is not delivered during inspiration or during the restricted phase of
exhalation. Reference Manual Inspiration, p. 10-21 for information on the restricted phase of
exhalation.
The Manual inspiration key can be used to deliver mandatory breaths to the patient or to
run an inspiratory pause maneuver in SPONT mode. The manual inspiration key cannot be
used to run an expiratory pause maneuver in SPONT mode.
Inspiratory pause key — Initiates an inspiratory pause which closes the inspiratory and exha-
lation valves and extends the inspiratory phase of a mandatory breath for the purposes of
measuring end inspiratory pressure (PI END) for calculation of plateau pressure (PPL), static
compliance (CSTAT), and static resistance (RSTAT).
Expiratory pause key — Initiates an expiratory pause which extends the expiratory phase of
the current breath in order to measure total PEEP (PEEPTOT).
Alarm reset key — Clears active alarms or resets high-priority alarms and cancels an active
audio paused condition. An alarm reset is recorded in the alarm log if there is an active
alarm. DEVICE ALERT alarms cannot be reset.
Audio paused key — Pauses alarms for 2 minutes. Cancel the audio paused function by
touching the on-screen Cancel button.
On rare occasions, the GUI touch screen may become unresponsive. If you observe
an unresponsive GUI, inaccurate GUI responses, or unintended GUI responses, reset
the touch screen to restore proper touch screen functionality.
To reset the touch screen:
1. Touch the display lock key on the GUI bezel to lock the screen. The locked padlock icon
appears on the screen and the display lock key illuminates.
2. Touch the display lock key again. Doing so displays a progress bar below the locked
padlock icon, after which time the locked icon will “unlock,” indicating a successful GUI
touch screen reset.
Alternatively, ensure that a patient is not connected to the ventilator and power
cycle the ventilator.
Note:
Do not touch the screen during the unlock period.
Note:
The manual GUI touch screen reset described in this section is different than the automatic
30-second transient reset of the GUI described in Table 2-9.
The table below shows the GUI’s visual indicators. Reference Areas of the GUI, p. 4-3
for area names.
The audio paused function has two visual indicators — the audio paused key on the
GUI bezel glows yellow during an audio paused interval, and a visual countdown
timer appears, showing the amount of time the audio paused interval has remain-
ing.
Symbol Description
Manual Event Touching this text causes the manual event screen to
appear, where a variety of events can be recorded for
viewing in the Trending layout. Reference the Trend-
ing addendum for more information about events
Symbol Description
Symbol Description
Touch an on-screen symbol briefly (0.5 s) to display a tooltip on the GUI screen. The
tooltip contains a definition of the symbol and other descriptive text, available with
either short or long descriptions. The short description expands to show more infor-
mation by touching “more” on the tooltip dialog or collapses by touching less. The
tooltip closes by touching close or fades in five (5) seconds if left alone. Expanding
the tooltip dialog prevents the tooltip from timing out. Touching outside the tooltip
causes the dialog to close.
The table below summarizes the ventilator’s symbols and abbreviations.
TA Apnea interval
TE Expiratory time
TH:TL High pressure time to Low pressure time ratio (in BiLevel)
TI Inspiratory time
PI Inspiratory pressure
VT Tidal Volume
VC Vital Capacity
VS Volume support
A tone sounds when a button on the GUI is touched, and also when settings are
accepted. Audible indicators include pitched tones, beeps, and key clicks. Key clicks
sound whenever a key on the GUI is pressed. Various tones annunciate patient
alarms.
Note:
Pressing the audio paused key mutes alarms for the 2-minute audio paused period.
Caregivers may choose to silence alarms by pressing the audio paused key. A 2-
minute countdown timer appears on the GUI during the audio paused interval.
Cancel the audio paused function by touching Cancel.
Function Description
Medium priority alarm tone A repeating series of three tones. Sounds when a
medium priority alarm occurs.
High priority alarm tone A repeating series of five tones. Sounds when a high
priority alarm occurs.
Soft bound tone One tone. Sounds when a soft bound is reached when
making changes to ventilator settings. A soft bound is
a selected value that exceeds or goes below its limit
and requires acknowledgment to continue.
Hard bound tone (invalid entry) The invalid entry sound occurs when a hard bound is
reached when making changes to ventilator settings.
A hard bound defines the upper or lower limit of the
setting, where the setting cannot be adjusted higher
or lower.
The clinician enters ventilation parameters via the GUI’s touch screen. Reference GUI
Front View, p. 2-6. The keys activate other ventilator functions. Reference GUI Control
Keys, p. 2-16.
BDU Controls
• ON/OFF switch — Lift the switch cover and turn the ventilator ON or OFF.
• Service mode button — Press and release this button when the Covidien splash screen
appears on the status display after powering on the ventilator to enter Service mode.
Note:
The Covidien splash screen shows the Covidien logo and appears momentarily as a banner
on the status display.
BDU AC Indicator
The status display and the AC power indicator are the only visual indicators on the
BDU. The AC indicator illuminates green whenever the ventilator is connected to AC
power. All other visual indicators on the ventilator are on the GUI. Reference Typical
Status Display Indicators and Messages, p. 2-31 for a description of the status display
indicators and symbols. Reference the section below for a summary of the informa-
tion appearing on the status display.
Status Display
The status display is a separate display located on the BDU. Reference BDU Front
View, p. 2-8, item 6. The status display provides the following information according
to the state of the ventilator:
During normal ventilation the status display shows
• Current power source (AC or DC)
• Circuit pressure graph displaying pressure units, 2PPEAK alarm setting and current PPEAK
and PEEP values
Note:
The status display provides a redundant check of ventilator operation. If the GUI stops
operating for any reason, ventilation continues as set.
The figure below shows a sample of the status display during normal ventilation
(compressor option not installed).
The following table lists indicators and messages that appear on the status display.
The continuous tone alarm is the only audible indicator in the BDU, and is described
in Table 2-10.
Indicator Description
Continuous tone alarm A continuous tone annunciated when there is a Ventilator Inoperative (Vent
(Immediate priority) Inop) condition. This alarm lasts for a minimum of two (2) minutes.
2.11.2 Connectors
• Exhalation port (From patient) — The expiratory limb of the patient circuit attaches to
the inlet of the exhalation bacteria filter. This port is compatible with a standard 22mm
(OD) conical connection.
• Proximal Flow sensor — A keyed pneumatic connector for the Proximal Flow Sensor is
provided with a locking feature to prevent inadvertent disconnection. The proximal
flow sensor measures flow and pressure at the patient wye. The Proximal Flow Sensor is
an optional sensor. Details on operation are provided in the appendix in this manual.
Reference Appendix E.
• Standard interface connectors — USB, HDMI, and Ethernet connectors are provided. The
USB connector allows screens to be captured on an external USB storage device and allows
communication with an external patient monitor via serial over USB protocol, and the HDMI
connector allow the GUI image to be displayed on an external video display device. The
Ethernet connector is used by Service Personnel to upload new software and options. Refer-
ence Port Use, p. 5-19 for more information. Reference To configure Comm ports, p. 5-5 for
information on serial over USB data transfer when configuring Comm ports for external
devices.
WARNING:
Use of the compressor in altitudes higher or barometric pressures lower than those
specified could compromise ventilator/compressor operation. Reference
Environmental Specifications, p. 11-8.
Color or Description
Symbol
Atmosphere
Vacuum
Water
Note:
Both the compressor and the Proximal Flow System are hardware options.
2 Solenoid Valve, options supply (SOL2) 25 Restrictor, breath delivery bypass (R2)
3 Pressure sensor, mix accumulator (PMX) 26 Flow sensor, patient gas delivery (FSD)
6 Proportional solenoid valve, patient gas 29 Solenoid valve, compressor unload (SOL7)
delivery (PSOLD)
19 Pressure sensor, exhalation (PE) 42 Pressure sensor, air gas inlet (PAir)
23 Check valve, patient gas delivery (CVD) 46 Filter element, Air (F2)
47 Check valve, wall Air inlet (CVWAir) 51 Flow sensor, Oxygen (FSO2)
50 Pressure sensor, Oxygen gas inlet (PO2) 54 Solenoide Valve, mix accumulator purge
(SOL 1)
5 Humidifier
3.1 Overview
This chapter contains information for the installation and set up of the Puritan Ben-
nett™ 980 Series Ventilator. Before operating the ventilator system, thoroughly read
this Operator’s Manual.
Topics include:
• Safety reminders
• Ventilator setup
• Battery information
WARNING:
Explosion hazard — Do not use in the presence of flammable gases. An oxygen-rich
environment accelerates combustibility.
WARNING:
To ensure proper operation and avoid the possibility of physical injury, only qualified
medical personnel should attempt to set up the ventilator and administer treatment
with the ventilator.
WARNING:
To prevent electrostatic discharge (ESD) and potential fire hazard, do not use
antistatic or electrically conductive hoses or tubing in or near the ventilator
breathing system.
3-1
Installation
WARNING:
Use only gas supply hoses approved by Covidien. Other hoses may be restrictive and
may cause improper ventilator operation.
Caution:
To ensure optimum performance, Covidien recommends preventive maintenance be
performed by factory-trained Biomedical Engineers per the schedule specified. Reference
Service Preventive Maintenance Frequency, p. 7-21.
Ventilator setup should have already been completed by factory-trained service per-
sonnel including successfully passing EST. This manual does not include ventilator
assembly instructions.
Using AC Power
WARNING:
Use only Covidien-branded batteries. Using other manufacturer’s brands could
result in the batteries operating the ventilator for less than the specified amount of
time or could cause a fire hazard.
WARNING:
One primary battery must be installed at all times in the BDU’s primary battery slot
for proper ventilator operation. The ventilator will not complete the startup process
without the primary battery installed. Reference Battery Compartment Locations, p.
3-21 for identification of battery slots.
The ventilator’s primary battery must be installed by qualified service personnel (as
it is shipped separately) before patient use. The ventilator will not complete Power
on Self Test (POST) if the battery is not present, and ventilation is prohibited. Ensure
the battery is fully charged before placing the ventilator into service.
The ventilator employs a battery backup system if AC power becomes unavailable
or drops below approximately 90 volts. A new, fully charged battery provides at least
one hour of power to the ventilator assuming ambient temperature of 20°C (68°F)
to 25°C (77°F), PBW = 70 kg, and at factory default ventilator settings.
The battery back-up systems for the ventilator and compressor contain one primary
battery each. Backup power is supplied to the ventilator in the event of an AC power
loss.
One extended battery slot is available for the ventilator and the compressor. If both
primary and extended ventilator and compressor batteries are present, these batter-
ies can power the ventilator and compressor for two hours (one hour for the primary
battery and one hour for the extended battery) under the environmental conditions
described above. When using battery power, the ventilator and compressor operate
from their extended batteries, if present, first and then switch to the primary batter-
ies. The ventilator and compressor primary and extended batteries are charged
whenever the ventilator is plugged into AC power (the ventilator does not have to
be powered up). If the ventilator or compressor is operating on battery power, the
status display shows which battery is in use and its charge level, and the remaining
time the battery will operate before charging is required again.
Battery Charging
will cause annunciation of the error and battery power will not be available for the
ventilator.
The ventilator status display indicates the charge level of the installed batteries, the
presence of one or more battery faults, and which battery is being charged.
The ventilator operates no differently when its batteries are charging than it does
when the batteries are fully charged.
The ventilator continues operating as set when the ventilator switches from AC
power to battery power and illuminates an indicator on the status display alerting
the operator that the ventilator is now operating on battery power and AC POWER
LOSS alarm annunciates. A medium priority alarm annunciates when the remaining
run-time for the ventilator drops to ten (10) minutes and a high priority alarm annun-
ciates when the remaining time drops to five (5) minutes.
Note:
Power outlet access and power cord position — Ensure that the power outlet used for
the ventilator is easily accessible; disconnection from the outlet is the only way to
completely remove power from the ventilator.
To connect the power cord to AC power
1. Plug the ventilator into a properly grounded power outlet rated for at least 15 A.
2. Verify the connection by checking the AC indicator below the power switch on the front of
the BDU. Reference Ventilator Power Switch and AC Indicator, p. 2-27 for the power switch and
AC indicator locations.
Use the power cord hook located at the back of the ventilator for power cord stor-
age.
WARNING:
For proper ventilator operation, and to avoid the risk of electric shock, connect the
ventilator to a grounded, hospital grade, AC electrical outlet.
The ventilator can be connected to hospital grade wall or bottled air and oxygen.
Reference Connecting the Ventilator to the Gas Supplies, p. 3-9. Both air and O2 supply
pressure ranges must be between 35 and 87 psig (241.3 kPa and 599.8 kPa) and the
average flow requirement for both gases is 60 L/min at 40.61 psi. The transient will
not exceed 200 L/min for ≥ three (3) s.
WARNING:
Due to excessive restriction of the Air Liquide™, SIS, and Dräger™ hose assemblies,
reduced ventilator performance levels may result when oxygen or air supply
pressures < 50 psi (345 kPa) are employed.
Gas cross flow from one high pressure input port of one type of gas to another high
pressure input port of a different gas will not exceed 100 mL/h under normal or
single fault conditions. If, during a single fault condition, cross flow exceeds 100 mL/
h, an audible alarm annunciates.
WARNING:
Use of only one gas source could lead to loss of ventilation and/or hypoxemia if that
one gas source fails and is not available. Therefore, always connect at least two gas
sources to the ventilator to ensure a constant gas supply is available to the patient in
case one of the gas sources fails. The ventilator has two connections for gas sources:
wall air inlet and oxygen inlet. Reference Non-technical Alarm Summary, p. 6-19 for
alarms that occur due to a loss of gas supplies.
To connect the gas sources
1. Connect the oxygen hose to the oxygen inlet fitting (item 1) as shown. Ensure use of a
medical grade oxygen source.
2. Connect the air hose to the air inlet fitting (item 2). Reference Connecting the Ventilator
to the Gas Supplies, p. 3-9.
WARNING:
To prevent a potential fire hazard and possible damage to the ventilator, ensure the
connections to the gas supplies are clean and unlubricated, and there is no water in
the supply gas. If water is suspected, use an external wall air water trap to prevent
damage to the ventilator or its components.
The ventilator system can be purchased with the following gas inlet fittings for both
air and O2: BOC, DISS, DISS female, NIST, Air Liquide, SIS, and Dräger
Reference Accessories and Options, p. 9-3 for part numbers of gas hoses. For countries
outside the USA, contact your local Covidien representative for the proper gas
hoses.
The ventilator is shipped with internal and external inspiratory filters. Reference
Accessories and Options, p. 9-3. To prevent infection and contamination, both inspi-
ratory and exhalation filters must be used with the ventilator.
WARNING:
In order to reduce the risk of infection, always use the ventilator with inspiratory and
exhalation bacteria filters.
WARNING:
Do not attempt to use inspiratory or exhalation filters designed for use with
ventilators other than the Puritan Bennett 980 Series Ventilator. Reference
Accessories and Options, p. 9-3 for relevant part numbers.
WARNING:
Refer to the filter’s instructions for use for details such as cleaning and sterilization
requirements, filtration efficiency, proper filter usage, and maximum filter
resistance, particularly when using aerosolized medications.
WARNING:
Refer to the exhalation filter instructions for use (IFU) for information on reusable
filter cleaning and sterilization and filter efficiency.
WARNING:
Do not re-use disposable inspiratory or exhalation filters, and dispose according to
your institution’s policy for discarding contaminated waste.
Caution:
Ensure both inspiratory and exhalation filters are properly attached to the ventilator.
To install the inspiratory filter
1. Attach the inspiratory filter to the To Patient port.
2. Ensure the direction of flow arrow is pointing outward, toward the patient circuit’s inspi-
ratory limb.
Note:
Refer to the inspiratory filter IFU for information on proper use and handling of the filter.
Note:
Refer to the exhalation filter IFU for information on proper use and handling of the filter and
for emptying the condensate vial for adult and pediatric patients. Reference Appendix D for
information on emptying the condensate vial when using neonatal exhalation filters.
The condensate vial must be assembled to the reusable exhalation filter prior to
installing the assembly to the ventilator.
To assemble the Adult/Pediatric reusable exhalation filter and condensate vial
1. Seat the filter to the condensate vial, ensuring alignment of the condensate vial’s seal
with the mating edge of the exhalation filter.
2. Twist the condensate vial in a counterclockwise direction until the stops on the vial and
exhalation filter meet.
WARNING:
Do not operate the exhalation filter latch during patient ventilation. Opening the
latch during ventilation will result in a patient disconnect condition and
corresponding alarm.
To install the Adult/Pediatric exhalation filter
1. If necessary, remove expiratory limb of patient circuit from exhalation filter.
2. Raise the exhalation filter latch to unlock (item 6). This raises the exhalation valve assem-
bly and allows the filter door to swing away from the ventilator. Reference Adult/Pediat-
ric Filter Installation, p. 3-12.
5. Insert the new filter by sliding the filter along the tracks in the door. Ensure the From
Patient port aligns with the cutout in the door and points away from the ventilator.
2. Lift exhalation filter latch. Reference Installing the Neonatal Filter, p. 3-15 (item 3).
3. Remove existing exhalation filter door by lifting it off of the pivot pins.
2. Attach the drain bag tube to the condensate vial’s drain port.
3. Hang the drain bag on the holder located on the ventilator’s accessory rail, as shown.
Reference Drain Bag, p. 3-14. Reference Accessories and Options, p. 9-3 for part number
of drain bag holder.
Reference Connecting the Adult or Pediatric Patient Circuit, p. 3-16 or Reference Con-
necting the Neonatal Patient Circuit, p. 3-17 to connect the adult, pediatric, and neo-
natal circuits.
WARNING:
Use patient circuits of the lowest compliance possible with the ventilator system to
ensure optimal compliance compensation and to avoid reaching the safety limit of
five times set tidal volume or the compliance compensation limit. Reference the
table below for circuit types corresponding with predicted body weight (PBW).
Note:
Refer to the patient circuit’s instructions for use (IFU) for information on proper use and
handling and care and maintenance of the circuit.
A list of breathing system components and accessories is provided. Reference Acces-
sories and Options, p. 9-3. Use only Covidien- components and accessories in the
patient circuit.
Follow your institution’s protocol for safe disposal of the patient circuit.
Follow the patient circuit’s instructions for use (IFU) for cleaning and disinfection
information for reusable circuits.
Orient the patient circuit by hanging the patient circuit on the circuit management
supports provided with the flex arm.
5 Condensate vial
5 Condensate vial
WARNING:
Do not attempt to sterilize single-patient use circuits.
3.6.1 Batteries
WARNING:
Use only Covidien-branded batteries. Using other manufacturer’s brands or
remanufactured batteries could result in the batteries operating the ventilator for
less than the specified amount of time or could cause a fire hazard.
WARNING:
To reduce the risk of infection due to cross-contamination, using a damp cloth,
disinfect the batteries with one of the solutions listed before installation and
whenever transferring to or from another ventilator. During use, clean external
surfaces of batteries as necessary. Reference Surface Cleaning Agents, p. 7-5. Do not
spray disinfectant directly onto the battery or its connector.
WARNING:
Even though the Puritan Bennett 980 Ventilator meets the standards listed in
Chapter 11, the internal Lithium-ion battery of the device is considered to be
Dangerous Goods (DG) Class 9 - Miscellaneous, when transported in commerce. As
such, the Puritan Bennett 980 Ventilator and/or the associated Lithium-ion battery
are subject to strict transport conditions under the Dangerous Goods Regulation for
air transport (IATA: International Air Transport Association), International Maritime
Dangerous Goods code for sea and the European Agreement concerning the
International Carriage of Dangerous Goods by Road (ADR) for Europe. Private
individuals who transport the device are excluded from these regulations although
for air transport some requirements may apply.
WARNING:
To avoid the risk of fire, explosion, electric shock, or burns, do not short circuit,
puncture, crush, heat above 60°C, incinerate, disassemble the battery, or immerse
the battery in water.
Caution:
Ensure the batteries are oriented properly. Reference Proper Battery Orientation, p. 3-
20.
1 Battery connector
Primary Batteries
The ventilator’s primary battery is located in the rearward battery receptacle on the
right side of the BDU. The compressor’s primary battery is located in the rearward
battery receptacle in the compressor base. Reference Battery Compartment Loca-
tions, p. 3-21. The primary battery may be “hot swapped,” that is it can be replaced
while the ventilator is operating.
2. If the charge level is sufficient, orient the battery as shown, face the front of the ventila-
tor and locate the battery compartments on the right side of the appropriate module.
Reference Battery Compartment Locations, p. 3-21. The receptacle towards the rear of
the ventilator houses the primary battery while the receptacle towards the front of the
ventilator houses the extended battery.
3. The primary battery is fastened in place with a thumbscrew (item 3). Loosen the thumb-
screw approximately four to five turns to allow battery installation.
4. Insert the battery and push into its receptacle all the way until it clicks, indicating it is
latched. The battery will only fit into the slot one way.
5. Tighten the thumbscrew to secure the battery and prevent the primary battery from
being removed.
Note:
Remove the primary battery by reversing the steps. After loosening the thumbscrew, slide
the battery ejector to the left to eject the battery.
1 BDU extended battery receptacle and 3 BDU and compressor primary battery
ejector thumbscrews
2 BDU primary battery receptacle and ejector 4 Compressor primary battery receptacle and
ejector
5 Compressor extended battery receptacle 6 BDU primary battery (positioned for installa-
and ejector tion)
Note:
Remove either primary battery by sliding the battery ejector to the left. The battery ejects
itself from its receptacle.
Extended batteries
The extended battery receptacle is located forward of the primary battery. Like the
primary battery, the extended battery may be hot swapped.
To install or remove an extended battery in either the BDU or compressor
1. Properly orient the battery as shown. Reference Proper Battery Orientation, p. 3-20.
2. Push the battery into the forward receptacle of the appropriate module of the ventilator
all the way until it clicks, indicating the battery is latched. Reference Battery Compart-
ment Locations, p. 3-21.
Note:
Remove the battery by sliding the battery ejector to the left. The battery ejects itself from its
receptacle. There is no thumbscrew for extended batteries.
Note:
Reference Battery Charging, p. 3-3 for battery charging information when batteries are
installed in ventilator.
Battery life is approximately three (3) years. Actual battery life depends on the history
of use and ambient conditions.
Use the flex arm to support the patient circuit between the patient and the ventila-
tor. Reference Flex Arm Installation, which illustrates flex arm installation into the
sockets provided.
3. Hang the patient circuit using the circuit management supports included with the flex
arm.
4. Remove the flex arm by first removing the patient circuit, then un-fastening the flex arm
from the threaded fastener in the handle.
3.6.6 Humidifier
Use the humidifier to add heat and moisture to the inhaled gas. Connect the humid-
ifier to a hospital grade electrical outlet. Choose the humidifier (type and volume
appropriate for the patient). The humidifier may be mounted with the humidifier
bracket as shown. Reference Bracket Installation on Rail, p. 3-26. Reference Accessories
and Options, p. 9-3 for the part number of the humidifier bracket.
WARNING:
Selection of the incorrect humidifier type and/or volume during SST or during
patient ventilation can affect the accuracy of delivered volume to the patient by
allowing the ventilator to incorrectly calculate the compliance correction factor used
during breath delivery. This can be a problem, as the additional volume required for
circuit compressibility compensation could be incorrectly calculated, resulting in
over- or under-delivery of desired volume.
WARNING:
To ensure proper compliance and resistance calculations, perform SST with the
humidifier and all accessories used for patient ventilation installed in the ventilator
breathing system.
WARNING:
Follow the humidifier manufacturer’s Instructions for Use (IFU) when using a
humidifier with patient ventilation.
Caution:
Follow humidifier manufacturer’s instructions for use (IFU) for proper humidifier
operation.
To install the humidifier bracket
1. Attach humidifier bracket to the ventilator’s accessory rail by placing the bracket behind
the railing and fastening the bracket clamp to the bracket with four (4)
5/32 inch hex screws, capturing the railing between the bracket and the clamp. Ensure
the humidifier mounting slots are facing outward from the ventilator.
3. Install the chamber to the humidifier, connect the patient circuit, then run SST.
Note:
Complete instructions for the humidifier bracket and humidifier installation are given in the
Puritan Bennett™ 980 Series Ventilator Humidifier Bracket Installation Instructions, which
includes humidifier bracket part numbers and descriptions.
Normal mode is the default mode used for patient ventilation. The ventilator enters
Normal mode after it has been turned on and POST completes, the ventilator is set
up, and breath delivery parameters have been entered. If the clinician chooses, s/he
can select Quick Start which uses default values or institutionally configured breath
delivery settings after PBW has been entered. Entry into Normal mode is not allowed
if a primary battery is not detected in the ventilator BDU, a major POST fault occurs,
or there is an uncorrected major system fault, or uncorrected Short Self Test (SST)
or Extended Self Test (EST) failures or non-overridden alerts.
During Normal mode, the omni-directional LED on the top of the GUI appears green in color,
in a steadily lit state. If an alarm occurs, the LED flashes in a color corresponding to the priority
of the alarm. Reference Alarm Prioritization, p. 6-16 for details regarding alarm priority. If
another alarm occurs concurrently with an existing alarm, the LED displays the color corre-
sponding to the highest priority level. If the alarm de-escalates, the latched area (located on
either side of the alarm LED indicator) of the alarm LED displays the color of the highest prior-
ity alarm while the center of the LED displays the color of the current alarm’s priority. For more
information on specific alarms, touch the logs icon in the constant access icons area of the
GUI.
Stand-By state can be used when the clinician needs to disconnect the patient for
any reason (prior to a suction procedure, for example). The ventilator enters Stand-
By state if a request is made by the clinician, a patient is disconnected within a fixed
time period determined by the ventilator software, and the clinician confirms the
patient has been disconnected intentionally. If a patient becomes disconnected
from the patient circuit after the time period elapses, an alarm sounds and the
patient-disconnect sequence is initiated. In Stand-by state, gas output is reduced to
ten (10) L/min to limit gas consumption and to allow for detection of patient recon-
nection and O2 concentration becomes 100% for adult and pediatric circuit types
and 40% for neonatal circuit types. Stand-by state is available in all ventilation modes
except during Inspiratory and Expiratory BUV, Occlusion Status Cycling (OSC),
Safety Valve Open (SVO), or Ventilator Inoperative (Vent Inop) conditions.
Note:
Do not block patient circuit wye while in Stand-By state. If the wye is blocked, the ventilator
detects a patient connection and will attempt to resume normal ventilation.
To enter Stand-By state
1. Touch the Menu tab on the left side of the GUI. The menu appears.
2. Touch Stand-By. A Stand-By state pending dialog appears instructing the clinician to dis-
connect the patient circuit. A timer starts which allows 30 s to disconnect the patient.
3. Disconnect the patient circuit and confirm the disconnection by touching Confirm. A
timer starts which allows 30 s for confirmation of disconnect.
• Ventilator settings can be changed, if desired, and will be applied upon patient recon-
nection.
• The ventilator displays an indicator that it is in Stand-by State, and a timer indicating the
elapsed time the ventilator has been in Stand-by state.
WARNING:
Before entering Service Mode, ensure a patient is not connected to the ventilator.
Ventilatory support is not available in Service Mode.
Service Mode is used for Extended Self Test (EST), ventilator calibration, configura-
tion, software upgrades, option installation (all of which must be performed by Covi-
dien factory-trained service personnel), and for making adjustments to institutional
settings. All information stored in the individual logs is available in Service Mode.
Service Mode logs include:
• System diagnostic Log
• Settings Log
• Alarms Log
• Service log
Reference the Puritan Bennett™ 980 Series Ventilator Service Manual for details about
Service Mode logs.
A patient must not be attached to the ventilator when entering Service Mode. Spe-
cific actions must be performed to enter this mode, prior to POST completion.
To access Service Mode
1. Remove the ventilator from patient usage.
3. Press and release the Service Mode button (TEST) at the back of the ventilator, when the
Covidien splash screen appears on the status display after powering on the ventilator.
Reference Service Mode Button (TEST), p. 3-32. Reference Status Display Indicators and
Descriptions, p. 2-31 for an image of the splash screen. The ventilator prompts to confirm
no patient is attached.
Note:
The Covidien splash screen shows the Covidien logo and appears momentarily as a banner
on the status display.
Reference the Puritan Bennett™ 980 Series Ventilator Service Manual for information on
which keys are disabled during EST.
WARNING:
If the ventilator fleet in your institution uses multiple institutionally configured
presets and/or defaults, there can be risks of inappropriate alarm settings.
The ventilator is shipped configured with factory defaults for new patient parame-
ters can be configured to suit institutional preferences. The operator may configure
any desired parameter as long as this option has not been locked out and rendered
unavailable. When configuring the ventilator, it displays the parameters associated
with the operator’s last configuration. The following table lists the factory-config-
ured settings, the institutionally-configurable settings, and the operator-configu-
rable settings.
Vital patient X X X X
data banner
Large font X X X X
patient data
panel
Waveform X X X X
layout
Display bright- X X X
ness (Light set-
tings)
Alarm volume X X X X
Elevate O2 X X X X
control
Date/time X X X X
format
New patient X X X
startup defaults
(including PBW,
vent type,
mode, manda-
tory type,
trigger type,
O2%, elevate
O2)
Opacity X X X X
Caution:
Do not lean on the GUI or use it to move the ventilator. Doing so could break the GUI,
its locking mechanism, or tip the ventilator over.
Prior to ventilating a patient, configure the GUI so it is capable of displaying all the
desired parameters, information, and patient data. This eliminates the necessity for
taking the patient off the ventilator, as configuration of many of the items requires
the unit to be in Service Mode.
To perform institutional configuration
1. Enter Service Mode, and confirm no patient is attached by touching Configuration. Ref-
erence Service Mode, p. 3-31 for instructions on entering Service Mode.
2. Touch Configuration at the top of the screen in Service Mode. A list of buttons appears
allowing configuration of the corresponding parameters.
2. Touch Configuration at the top of the screen in Service Mode. A list of buttons appears
allowing configuration of the corresponding parameters.
3. Select the desired modified setting from the left-hand menu options.
4. Touch Default.
The date and time may be configured to the institution’s preference. The time can
be specified as 12-hour or 24-hour time in HH:MM:SS format with one-hour and one-
minute resolutions, respectively. The date formats are:
• DD-MMM-YYYY where DD is a two-digit day format, MMM is a three-letter abbreviation
for the month, and YYYY is a four-digit representation of the year or
• MM-DD-YYYY where MM is a two digit month format, DD is a two-digit day format, and
YYYY is a four-digit representation of the year
The settable date corresponds to the number of days in the set month and accounts
for leap years.
To institutionally configure the ventilator’s date and time settings
1. Enter Service Mode, and confirm no patient is attached by touching Configuration. Ref-
erence Service Mode, p. 3-32 for instructions on entering Service Mode.
4. Touch Hour and turn the knob to enter the correct hour.
6. Touch the button corresponding to the date format desired (DD-MMM-YYYY or MM-
DD-YYYY).
Pressure Units
2. Touch Light Settings. Sliders appear to adjust the screen brightness and keyboard back-
light.
3. Move the sliders to increase or decrease the brightness and backlight levels. Alternative-
ly, turn the knob to increase or decrease the brightness and backlight levels.
2. Slide the brightness slider or turn the knob to adjust the brightness level.
3. Dismiss the slider by touching anywhere on the GUI screen or allow to time out in five
(5) s.
2. Touch the Vent type, Mode, Mandatory type, and Trigger type buttons corresponding to
the desired parameters.
3. Configure the default PBW and mL/kg ratio, Elevate O2 and O2% by touching its button
and turning the knob.
Elevate O2
Note:
The Elevate O2 control adds a percentage of O2 to the breathing mixture for two minutes.
The additional percentage is shown on the icon in the constant access icon area. The
allowable range is 1% to 100%.
To adjust the amount of elevated O2 delivered for two minutes
1. In the vent setup dialog in Normal mode, touch the Elevate O2 icon in the constant
access icons area of the GUI screen. The icon glows and a dialog appears with a count-
down timer, Elev O2 button highlighted and ready for changes, and Extend, Stop, and
Close buttons.
2. Turn the knob to increase or decrease the amount of oxygen by the amount shown on
the button. The allowable range is +1% to +100% oxygen.
3. Touch Extend to extend the two-minute interval. Touching Extend restarts the two-
minute countdown timer.
4. Touch Stop to stop additional oxygen from being delivered and dismiss the countdown
timer.
• If apnea ventilation occurs during the two-minute interval, the apnea % O2 delivery also
increases by the configured amount.
• During LOSS OF AIR SUPPLY or LOSS OF O2 SUPPLY alarm conditions, the Elevate O2
function is canceled if in progress, and is temporarily disabled until the alarm condition
no longer exists.
• During Safety PCV, the Elevate O2 control has no effect. During circuit disconnect and
stand-by states (when the ventilator is turned on but not ventilating) the Elevate O2
function affects the currently delivered oxygen concentration, not the set oxygen con-
centration.
Alarm Volume
WARNING:
The audio alarm volume level is adjustable. The operator should set the volume at a
level that allows the operator to distinguish the audio alarm above background
noise levels.
To institutionally configure the alarm volume
1. Enter Service Mode, and confirm no patient is attached by touching Configuration. Ref-
erence Service Mode, p. 3-31 for instructions on entering Service Mode.
2. Touch Alarm Volume Defaults. A screen appears allowing configuration of the alarm
volume by circuit type.
3. Slide the alarm slider for each circuit type (adult, pediatric, or neonatal) or turn the knob
to configure the alarm volume. The volume settings range from 1 (minimum) to 10
(maximum).
2. Dismiss the slider by touching anywhere on the GUI screen or allow to time out in five
(5) seconds.
Note:
A sample alarm tone sounds for verification at each volume level change. If necessary, re-
adjust the alarm volume by moving the alarm volume slider to increase or decrease the
volume.
Note:
The alarm volume reverts to the institutionally configured default alarm volume or factory
default if the ventilator’s power is cycled.
Patient data are displayed in the Vital Patient Data banner. The operator can config-
ure the banner for displaying the desired patient data. Reference Areas of the GUI, p.
4-3. A total of 14 values may be configured at one time, with eight (8) values visible,
and six (6) more visible by scrolling the values using the left- and right- pointing
arrows in the patient data area.
Two pages of additional patient data may be viewed by touching or swiping down
on the patient data tab at the top of the GUI. Choose the respective buttons to view
page one or page two. Additional patient data values may not be changed.
Reference Ventilator Settings Range and Resolution, p. 11-9 for default patient data
values.
To institutionally configure patient data displayed on the GUI
1. Enter Service Mode, and confirm no patient is attached by touching Configuration. Ref-
erence Service Mode, p. 3-31 for instructions on entering Service Mode.
2. Touch Patient Data Defaults. Five (5) layout preset buttons appear along with a list of
parameters and descriptions.
3. Touch a preset button and individually select a parameter from the scrollable list below
to appear in that preset’s vital patient data banner. Use the right- and left- pointing
arrows to configure default values for all available parameters. Additionally, touch the
padlock icon above each patient data parameter on the data banner to allow
4. When done configuring the selected preset, touch Accept and select another preset to
configure, if desired.
2. Touch the button corresponding to the replacement parameter. The existing parame-
ter is replaced with the new parameter.
To improve visibility of patient data, a screen is available that appears with a larger
font. Up to 14 data values may be displayed which include:
• Institutional default patient data values (if configured)
• Remaining user selected patient data values (up to 14, including waveforms and loops)
2. Touch Large Font Patient Data Defaults. Five layout presets appear along with a list of
parameters and descriptions.
3. Touch a preset button and individually select a parameter for each of the desired
patient data values.
4. Choose the desired scalar and loop waveforms for the large font patient data display.
Waveform thumbnails only appear in the three right-most cells of the large font data
panel.
5. Touch any of the padlock icons along the right-most edge of the selected layout to
prevent operator configurability of the selected row.
2. Swipe the additional patient data banner’s tab downward or touch the additional
patient data banner’s tab. Patient data appear in a larger font.
3. Swipe the large font patient data panel tab upward or touch the tab to return to the
banner to its normal font size.
The large font patient data parameters are configured in the same way as described
in the patient data configuration section above.
Waveforms
2. Touch Graph Defaults. Five (5) layout presets appear along with a list of parameters and
descriptions.
3. Touch a layout preset button. The parameter(s) button outline glows, signifying that it
can be changed. If more than one parameter can be changed, touch that parameter to
make its outline glow.
4. Select the parameter from the list whose waveform is desired to appear on the wave-
forms screen.
6. Touch the padlock icon above each graphic layout to prevent operator configuration of
the selected layout.
2. Touch the desired waveform(s) icon to display. The selected waveform(s) appear on the
GUI screen and the dialog closes.
2. Turn the knob to change the value. For each axis, turn the knob to the right to decrease
the values, and turn to the left to increase the values.
To pause waveforms
1. Touch the pause icon, located below the waveforms area. The icon glows yellow and
allows the breath to complete. A cursor appears and travels along the waveform while
turning the knob, displaying the x- and y-axis values.
Opacity
4. Touch the padlock icon at the right side of the screen to allow or prevent operator
adjustment of the screen opacity.
Note:
The opacity icon can be found on the vent setup screen and on any of the respiratory
mechanics maneuvers screens.
WARNING:
Always disconnect the patient from the ventilator prior to running SST or EST. If SST
or EST is performed while a patient is connected, patient injury may occur.
WARNING:
Check for circuit occlusion and/or run SST if increased pressures are observed during
ventilation.
WARNING:
When changing any accessories in the patient circuit or changing the patient circuit
itself, run SST to check for leaks and to ensure the correct circuit compliance and
resistance values are used in ventilator calculations.
Note:
When extending ventilator circuits for neonatal patients, the resulting ventilator breathing
system (VBS) compliance may trigger a COMPLIANCE LIMITED VT alarm such that the VC+ or
VS software will not continue to update the pressure target during breath delivery. In this
case the user can change the breath type to pressure control (PC) or pressure support (PS).
When a patient is not attached to the ventilator, run SST to check the patient circuit
for:
• Gas leaks
SST is a five-minute test and must be run under any of the following conditions:
• Prior to ventilating a new patient
No external test equipment is required, and SST requires minimal operator participa-
tion.
Humidification type and volume can be adjusted after running SST, however the
ventilator makes assumptions when calculating resistance and compliance if these
changes are made without re-running SST. For optimal breath delivery, run SST after
changing humidification type and humidifier volume.
SST results are recorded in the SST results log, viewable in Service Mode and in
Normal Mode using the configuration (wrench) icon.
Required Equipment
• Humidifier, if applicable
• Two gas sources (air and oxygen) connected to the ventilator) at a pressure between 35
psi and 87 psi (241.3 kPa and 599.8 kPa)
To run SST
1. Ensure a patient is NOT connected to the ventilator.
2. So that the ventilator does not detect a patient connection, ensure that the breathing
circuit wye is not attached to a test lung or covered in any way that would cause an
increase in pressure at the wye.
3. Turn the ventilator on using the power switch located at the front of the BDU, below the
status display. The ventilator runs POST when the power switch is turned on. Ensure the
ventilator is operating on full AC power. Otherwise, SST test failures may result.
4. Wait at least 15 minutes to allow the ventilator to warm up and stabilize to ensure accu-
rate results.
5. At the ventilator startup screen, touch SST or the Configure icon (wrench) displayed in
the lower right area of the GUI. The SST history log appears along with Patient Setup, Run
Leak Test, and Run All SST buttons.
6. Connect the patient circuit, filters, condensate vial, and all accessories to be used in
patient ventilation. Ensure the patient wye is not blocked.
7. Touch Run All SST to perform all SST tests or touch Run Leak Test to perform the SST Leak
test of the ventilator breathing circuit.
9. After accepting, touch the Circuit Type button corresponding to the patient circuit type
used to perform SST and to ventilate the patient (adult, pediatric, or neonatal).
10. Touch the Humidification Type button corresponding to the humidification type used
for patient ventilation. If no humidifier is used, touch HME. If a humidifier is used, touch
Humidification Volume and turn the knob to enter the volume. See Table 3-4. for adult
and pediatric patients or Table 3-5. for neonatal patients to determine the correct
volume to enter.
12. Follow the prompts. Certain SST tests require operator intervention, and will pause
indefinitely for a response. Reference Individual SST Results, p. 3-48 for a summary of the
SST test sequence and results.
13. After each test, the ventilator displays the results. If a particular test fails, the test result
appears on the screen and a choice to repeat the test or perform the next test is given.
When all of the SST tests are complete, the SST status screen displays the individual test
results.
14. To proceed to patient set up, (if SST did not detect an ALERT or FAILURE) touch EXIT SST,
then touch Accept or cycle the ventilator’s power.
SST Flow Sensor Cross Check Test Tests O2 and Air Flow Sensors
SST Exhalation Valve Performance Calibrates the exhalation valve and creates a table for use during cal-
culations
SST Exhalation Filter Test Checks for exhalation filter occlusion and exhalation compartment
occlusion.
SST circuit Resistance Test Checks for inspiratory and expiratory limb occlusions, and calculates
and stores the inspiratory and expiratory limb resistance parameters.
SST circuit Compliance Test Calculates the attached patient circuit compliance.
SST Prox (if Proximal Flow Option is Verifies functionality of Proximal Flow Subsystem
installed)
Note:
For neonatal patient types, enter the SST humidifier volume listed in Table 3-5. during SST
or when specifying the humidifier volume.
SST Results
SST reports results for each individual test.Three status indicators identify the SST
results and actions to take for each.
• Pass — The individual SST test has met its requirements.
• Alert — Alerts occur when the ventilator detects one or more non-critical faults.
• Fail — The individual SST test did not meet its requirements.
ALERT The test result is not ideal, but is not critical. When the system prompts, touch one of
If SST is in progress, it halts further testing these buttons:
and prompts for decision. • Repeat Test
• Next Test
• Exit SST
FAIL The ventilator has detected a critical Eliminate leaks in the ventilator breathing
problem and SST cannot complete until system and re-run SST. Otherwise, service
the ventilator passes the failed test. the ventilator and re-run SST.
SST Outcomes
WARNING:
Overriding an Alert in SST may result in ventilator performance outside of the stated
specification for accuracy. Choose to override the ALERT status and authorize
ventilation only when absolutely certain this cannot create a patient hazard or add
to risks arising from other hazards.
When SST completes all of the tests, analyze the results.
PASS All SST tests passed. Touch Patient Setup to set up the patient
for ventilation
ALERT The ventilator detected one or more faults. To override the alert, touch Override SST,
Choose to override the ALERT status and then touch Accept.
authorize ventilation only when absolutely To exit SST, touch Exit SST.
certain this cannot create a patient hazard
or add to risks arising from other hazards.
FAIL One or more critical faults were detected. Check the patient circuit to determine the
The ventilator enters the SVO state and problem or restart SST with a different
cannot be used for normal ventilation until patient circuit.
SST passes. Touch Repeat Test, Run All SST, or Exit SST,
then touch Accept.
• there were no changes to the circuit type at the start of the interrupted SST.
During SST, the ventilator displays the current SST status, including the test currently
in progress, results of completed tests. Test data are available in Service Mode where
applicable or are displayed on the screen. The ventilator logs SST results, and that
information is available following a power failure. The audio paused and alarm reset
keys are disabled during SST, as well as the Manual Inspiration, Inspiratory Pause, and
Expiratory Pause keys.
The ventilator’s Extended Self Test (EST) function is designed to verify the ventilator’s
operational subsystem integrity.
All required software support to perform EST is resident on the ventilator. EST
requires approximately 10 minutes to complete.
Note:
SST is not part of the EST test suite. To determine patient circuit resistance and compliance,
run SST.
Follow all identified guidelines when performing the EST self test. Inspect all equip-
ment required for any self test to ensure it is not damaged in any way.
1. Collect all required equipment prior to performing any self test of the ventilator. Suc-
cessful self test is not possible without the use of the listed equipment.
4. Connect the ventilator to AC power using the hospital-grade power cord until comple-
tion of any self test.
6. Ensure both air and oxygen sources register pressure between 35 and 87 psi
(241 to 599 kPa).
To perform Extended Self Test (EST) or to access additional service functions, the
ventilator must be in Service Mode. Reference Service Mode, p. 3-31.
Note:
While in the Service Mode, normal ventilation is not allowed.
WARNING:
Always disconnect the ventilator from the patient before running EST. Running EST
while the ventilator is connected to the patient can injure the patient.
WARNING:
A fault identified during this test indicates the ventilator or an associated component
is defective. Rectify the fault and perform any required repairs prior to releasing the
ventilator for patient use, unless it can be determined with certainty that the defect
cannot create a hazard for the patient, or add to the risks which may arise from other
hazards.
Perform EST during any of the listed conditions.
During EST, the ventilator displays the current EST status, including the test currently
in progress, results of completed tests, and measured data (where applicable). The
ventilator logs EST results, and that information is available following a power failure.
The ventilator disables several offscreen keys located on the bezel of the GUI during
EST.
• Audio paused
• Alarm reset
• Manual inspiration
• Inspiratory pause
• Expiratory pause
Note:
Attempts to run EST with a Neonatal filter can cause some EST tests to fail.
Note:
If using Air Liquide™, Dräger™, and SIS air/oxygen hose assemblies, certain EST tests may fail
when using supply pressures less than50 psi (345 kPa) based on excessive hose restriction.
Note:
If the ventilator has not reached normal operating temperature from recent usage, allow it
to warm up for at least 15 minutes in Service Mode prior to running EST to ensure accurate
testing.
To perform EST
1. Review and perform all self test prerequisites. Reference EST Self Test Prerequisites, p. 3-50.
4. Verify all three CALIBRATION tests under the CALIBRATION tab have passed.
5. Touch the SELF TEST tab from the horizontal banner at the top of the monitoring screen.
7. Touch Run All to run all tests in sequence or select the desired individual test.
8. Choose one of the available options: touch Accept to continue; touch Cancel to go back
to the previous screen; or touch Stop to cancel EST.
9. Follow the prompt to remove the inspiratory filter and connect the gold standard cir-
cuit.
11. Follow prompts to complete EST. The EST tests require operator intervention, and will
pause indefinitely for a response. Reference EST Test Sequence, p. 3-52.
12. At the DISCONNECT O2 prompt, disconnect the high pressure oxygen source.
13. At the CONNECT AIR AND O2 prompt, connect both high pressure air and oxygen sourc-
es.
14. Touch Run All or select the desired individual test. After each test, the ventilator displays
the results.
15. If a particular test fails, either repeat the test or perform the next test.
16. When all of the EST tests complete, review test results by pressing each individual test
listed on the left side of the GUI.
18. Touch Accept. The ventilator reruns POST and then displays the ventilator startup
screen.
Zero Offset Tests inspiratory and expiratory pressure transducers and Follow prompts
flow sensors at ambient pressure.
Mix Accumulator Verifies mix accumulator pressure sensor and overpressure None
switch function.
Exhalation Valve (EV) Verifies exhalation valve.current and loopback current are None
Loopback within range.
Exhalation Valve (EV) Verifies current versus pressure values in flash memory cor- None
Pressure Accuracy respond with actual installed exhalation valve.
Exhalation Valve (EV) Verifies the exhalation valve operates within specifications None
Performance of the last exhalation valve calibration.
Exhalation Valve (EV) Verifies the velocity transducer is sending a signal and the None
Velocity Transducer control circuit recognizes it. It does not verify the quality of
the signal.
Backup Ventilation Verifies backup ventilation systems: mix, inspiratory, and None
exhalation.
Communication Verifies GUI communication ports function, both serial and None
ethernet.
BD Alarms Verifies BD audible alarm is functional. Also verifies power Follow prompts
fail capacitor can operate loss-of-power alarm.
Offscreen Key Test Verifies GUI bezel key function. Follow prompts
Ventilatory Battery Tests ventilator battery and power distribution. Follow prompts
Compressor Battery Tests compressor battery function, as well as compressor Follow prompts
power system and fan function.
ALERT The test result is not ideal, but is not crit- When the system prompts, select:
ical. REPEAT TEST,
If EST is in progress, it halts further NEXT TEST, or
testing and prompts for decision. STOP,
then touch ACCEPT.
NEVER RUN Test still requires successful PASS. Run all EST tests.
PASS All EST tests passed EST successfully completed. Select other
SERVICE MODE functions or prepare for
SST tests prior to returning the ventilator
for patient usage.
ALERT The ventilator detected one or more When the system prompts, select:
faults. The test result is not ideal, but is REPEAT TEST,
not critical. NEXT TEST, or
STOP,
then touch ACCEPT.
FAIL One or more critical faults were detect- When the system prompts, select:
ed. The ventilator enters the SVO state Repeat Test,
and cannot be used for normal ventila- Next Test, or
tion until SST passes. Service is required. Stop,
then touch Accept.
WARNING:
Choose to override the ALERT status and authorize ventilation only when absolutely
certain this cannot create a patient hazard or add to risks arising from other hazards.
4.1 Overview
This chapter describes Puritan Bennett™ 980 Series Ventilator operation and
includes the following sections:
• Setting up the ventilator
4-1
Operation
bacteria filter, through the exhalation valve, which includes the exhalation flow
sensor, and through the exhalation port.
WARNING:
To avoid interrupted ventilator operation or possible damage to the ventilator,
always use the ventilator on a level surface in its proper orientation.
To set up the ventilator
1. Connect the ventilator to the electrical and gas supplies. Reference Power Cord Retainer
on BDU, p. 3-7 and Reference Connecting the Ventilator to the Gas Supplies, p. 3-9.
2. Connect the patient circuit to the ventilator. Reference the figures on p. 3-16 and p. 3-17
to connect the adult/pediatric or neonatal patient circuits, respectively.
3. Turn the ventilator ON using the power switch. Reference Ventilator Power Switch and AC
Indicator, p. 2-27.
4. Before ventilating a patient, run SST to calculate the compliance and resistance with all
items included in the patient circuit. Reference To run SST, p. 3-45.
• Circuit pressure graph displaying PPEAK, PEEP, and pressure-related alarm settings
The GUI is used to interact with the ventilator while it is ventilating a patient or in any
of its operating modes.
Caution:
Do not lean on the GUI or use it to move the ventilator. Doing so could break the GUI,
its locking mechanism, or tip the ventilator over.
The GUI is divided into several areas.
1. Prompt area — Located beneath the waveforms. Any prompts or messages related to
soft or hard bounds display here. A soft bound is a selected value that exceeds its rec-
ommended limit and requires acknowledgment to continue. Hard bounds have
minimum and maximum limits beyond which values cannot be selected, but if the
desired value is equal to a settings hard bound, then it is allowable.
2. Menu tab — Located on the left side of the GUI screen. Swiping the tab to the right and
touching Setup causes the Vent, Apnea, and Alarm tabs to appear. Touching those tabs
opens screens so that changes to ventilator settings, apnea settings and alarm settings
can be made.
3. Waveform area — Located in the center of the GUI screen. Shows various breath wave-
forms. Reference To institutionally configure waveforms and loops, p. 3-41 for information
on how to configure graphics.
4. Breath Phase Indicator — During normal ventilation, the GUI displays a breath indica-
tor in the upper left corner which shows the type of breath [Assist (A), Control (C), or
Spontaneous (S)] currently being delivered to the patient, and whether it is in the inspi-
ratory or expiratory phase. The breath indicator is updated at the beginning of every
inspiration, and persists until the next breath type update. During inspiration, assist (A)
and control (C) breath indicators glow green and spontaneous (S) breath indicators
glow orange, each appearing in inverse video where the indicator appears black sur-
rounded by the colored glow. Reference Areas of the GUI, p. 4-3. During the expiratory
phase the breath indicators appear as solid colors (green during assist or control breaths
and orange during spontaneous breaths).
5. Vital Patient data banner — Located across the top of the GUI screen. The patient data
banner displays monitored patient data and can be configured to show desired patient
data. Reference Vital Patient Data, p. 3-39 for information on configuring patient data for
display.
6. Alarm banners — Located on the right side of the GUI screen. Indicates to the operator
which alarms are active, and shown in a color corresponding to priority
(high is red and flashing, medium is yellow and flashing, low is yellow and steady).
7. Constant access icons — Located at the lower right of the GUI screen. This area allows
access to home (house), configure (wrench), logs (clipboard), elevate oxygen percent-
age (O2), and help (question mark) icon. These icons are always visible regardless of the
function selected on the GUI.
8. Constant access area — Consists of the Current Settings area and the Constant access
icons. This area allows access to any of the patient setup variables shown in these areas.
Touching an icon causes the particular menu for that variable to appear.
9. Current settings area — Located at the lower center of the GUI screen. The ventilator’s
current active settings display here. Touching any of the current settings buttons causes
a dialog to appear, allowing changes using the knob.
10. Vent Setup Button — Located at the lower left of the GUI screen. Touching this button
allows access to the ventilator setup screen.
Reference Status Display, p. 2-29 for information about displayed items during
Service mode.
Screen Opacity
The opacity control enables the operator to adjust the opacity of the displayed informa-
tion between 50% and 100%. At 50%, the displayed image is semi-transparent, and at
100%, the displayed image is opaque. The opacity value remains as set if power is
cycled.Reference To adjust the screen opacity, p. 3-43 for instructions on adjusting this fea-
ture.
Pushpin Feature
The pushpin feature prevents a dialog from closing under certain conditions. Like
the opacity control, the pushpin appears on the settings screen after a new patient
starts ventilation.
1 2
Display Brightness
2. Slide the brightness slider to the right to increase the brightness level or to the left to
decrease the brightness level. Alternatively, turn the knob to increase or decrease the
brightness level. The control disappears from the screen in approximately five (5) sec-
onds.
Display Lock
The primary display provides a display lock key to prevent inadvertent changes to
settings. When active, the display lock disables the touch screen, knob, and off-
screen keys and illuminates an LED on the display bezel. An image of the display lock
icon appears transparently over anything displayed on the GUI, should the operator
attempt to use the GUI. Any new alarm condition disables the display lock and
enables normal use of the GUI.
To lock and unlock the display
1. Press the display lock key on the GUI. The keyboard LED illuminates and a transparent
icon appears on the screen, indicating display lock. The icon shortly disappears, but if
the operator tries to activate any of the touch screen controls, the icon re-appears.
2. To unlock the display, press the display lock key again. The display lock LED turns off.
The GUI incorporates a gesture-based interface where features can be actuated with
the fingers using different motions. The following table explains gestures used with
the GUI.
Swipe Quickly brush the Opening or closing dialogs or Swipe toward the center of the screen to open
screen surface with panels that slide in and out dialogs or panels. Swipe toward the side of the
the fingertip. from the screen sides or top, screen (or upward if viewing the additional
moving waveform data, patient data or large font patient data panels) to
expanding or collapsing toolt- close.
ips, scrolling lists, or alarm ban- To move a paused waveform, swipe in the
ners, maximizing or minimizing desired direction.
waveforms. Swipe upward anywhere on a waveform to max-
imize it, and swipe downward on the maximized
waveform to minimize it.
Swipe a tooltip upward to expand to a long
description and downward to collapse to a short
description. A downward swipe anywhere in the
patient data area opens the additional patient
data panel, and another swipe on the additional
patient data tab displays the large font patient
data panel.
Double-tap Rapidly touch the Maximizing or minimizing the Double-tapping maximizes the viewable wave-
screen surface viewable area of a dialog, con- form area or shows the long description of a
twice with one trol, or waveform, expanding or tooltip. Double-tapping again minimizes the
finger. collapsing tooltips viewable waveform area or shows the short
description of a tooltip. If the control is configu-
rable, double tapping produces the configura-
tion pop-up menu.
Drag Move the fingertip Changing x- and y- axis scales, Touch the axis and drag to the right to increase
over the screen repositioning waveforms, the waveform x-axis scale, and to the left to
surface without moving the waveform cursor, decrease. Touch the axis and drag upward to
losing contact. moving scrollbars, scrolling increase the y-axis scale and downward to
lists. Scrolling speed varies decrease.
depending upon how far To reposition waveforms, touch and drag the
outside the list boundary the graph to the new position.
finger is positioned. To move the waveform cursor, touch the cursor
and drag it right or left. The graph responds sim-
ilarly.
Scroll a list by dragging the scrollbar right or left
or up or down. The list scrolls according to the
direction of the finger movement.
An automatic scrolling feature starts if the finger
is dragged from the inside of a list to outside its
boundary. The farther outside the boundary the
finger is dragged, the faster the list scrolls.
Drag and Touch and drag an Dragging help icon to describe Drag the help icon, located at the lower right of
drop item to another an onscreen item. the GUI screen, to the item in question and drop.
location and lift If a blue glow appears, a tooltip is available and
finger to drop. appears with information about that item (for
example, a control or symbol).
WARNING:
Prior to patient ventilation, select the proper tube type and tube ID.
Caution:
Do not set containers filled with liquids on the ventilator, as spilling may occur.
After turning on the ventilator it will display a “splash screen,” and run Power On Self
Test (POST). After the splash screen appears, the ventilator gives a choice to ventilate
the same patient or a new patient, or run SST.
Ventilation parameters are entered via the graphical user interface (GUI) using the fol-
lowing general steps:
1. Touch the setting displayed on the GUI.
2. Turn the knob to the right to increase or to the left to decrease the value.
3. Touch Accept to apply the setting or Accept ALL to apply several settings at once.
Note:
Quick Start allows for rapid setup and initiation of mechanical ventilation. Review Quick Start
parameters and ensure they are consistent with institutional practice before using this
feature.
To use Quick Start
1. Touch New Patient.
3. Turn the knob to adjust the patient’s PBW or gender and height (if gender is selected,
the height selection becomes available).
5. Connect the circuit wye adapter to the patient's airway or interface connection. The
patient is ventilated with the institutionally configured Quick Start defaults according to
the PBW or gender/height entered, and circuit type used during SST. There is no prompt
to review the settings and the waveforms display appears.
Note:
Connecting the circuit wye adapter to the patient's airway or interface connection prior to
making the ventilation settings causes the ventilator to begin ventilation using Safety
Pressure Control Ventilation (Safety PCV) and annunciate a PROCEDURE ERROR alarm. As
soon as the ventilator receives confirmation of its settings (by touching Accept or Accept
ALL), it transitions out of safety PCV, resets the alarm, and delivers the chosen settings.
Reference Safety PCV Settings, p. 10-70 for a listing of these settings.
To resume ventilating the same patient
1. Touch Same Patient on the GUI screen. The previous ventilator settings are displayed on
the screen for review prior to applying the settings to the patient.
2. If the settings are acceptable, touch Accept to confirm. To change any settings, touch
the setting, turn the knob clockwise to increase the value of the setting or counter-
clockwise to decrease the value of the setting, and touch Accept to confirm. To make
several settings changes at once, make the desired changes, then touch Accept ALL to
confirm. The appearance of the settings changes from white, non-italic font showing
the current setting to yellow italics (noting the pending setting). After the settings are
accepted, the appearance changes back to white non-italic font.
2. Enter the patient’s PBW or gender and height (if gender is selected, the height selection
becomes available).
3. If the default ventilator settings are appropriate for the patient, touch START to confirm
the settings, otherwise, touch a ventilator setting and turn the knob to adjust the
parameter. Continue this process for all parameters needing adjustment.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient's
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
The following ventilator settings appear at the new patient setup screen:
• Predicted Body Weight (PBW) — Adjust the patient’s PBW, or select the patient’s
gender and height. Reference Predicted Body Weight (PBW) Calculation, p. 4-20.
• Mode — Specify the breathing mode (A/C (assist/control), SIMV (synchronous intermit-
tent mandatory ventilation), SPONT (spontaneous ventilation), BiLevel (if the BiLevel
option is installed), or CPAP
• Spontaneous type — If SIMV or BiLevel was selected as the Mode, specify PS (pressure
support) or TC (tube compensation. If SPONT was selected as the Mode, specify PS (Pres-
sure Support), TC (Tube Compensation), or VS (Volume Support) or PAV+ (Proportional
Assist Ventilation) (if the PAV+ software option is installed).
Note:
VS, PAV+, and TC are only available during INVASIVE ventilation.
• Swipe the menu tab on the left side of the GUI and touch Setup
1 Setup button
4. Continue in this manner until all changes are made, then touch Accept or Accept ALL.
5. Touch START. Ventilation does not begin until the breathing circuit is connected to the
patient’s airway. After ventilation begins, waveforms begin plotting on the displayed
waveforms axes. Reference Waveforms, p. 3-41 for information on setting up the graphics
display.
If changes to any settings are required, return to the Vent Setup screen as described
above.
Note:
A yellow triangle icon appears on tabs and buttons displayed on the GUI containing unread
or un-viewed items. When the item containing the icon is touched, the icon disappears.
Note:
To make any settings changes after completing patient setup, touch the Vent tab on the left
side of the Setup dialog and make settings changes as described above. The current setting
appears in white font and changes to yellow italics to note the new value is pending.Touch
Accept or Accept ALL to confirm a single change or a batch of changes. Once the settings are
accepted, their appearance changes to white font.
Note:
Selecting Quick Start, Accept, Accept ALL or Start from the Setup dialog implements all
settings in ALL four Setup tabs (Vent Setup, Apnea, Alarms, and More Settings) and
dismisses the Setup dialog.
Tube Compensation
4. Finish setting up the ventilator as described (reference p. 4-12 for information on enter-
ing ventilator settings.
5. Ensure to select the tube type (either endotracheal or tracheostomy) and set the tube
ID to correspond to patient settings.
6. After making the changes, touch Accept to apply the new settings, or Cancel to cancel
all changes and dismiss the dialog.
WARNING:
To prevent inappropriate ventilation with TC, select the correct tube type ET or
Tracheostomy) and tube inner diameter (ID) for the patient’s ventilatory needs.
5. Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel all changes
and dismiss the dialog.
Note:
The tube type and tube ID indicators flash if TC is a new selection, indicating the need
for entry of the correct tube type and tube ID.
A Humidifier Volume button appears below the selection only if Non-Heated Expiratory
Tube or Heated Expiratory Tube is selected as the humidifier type.
2. Turn the knob to enter a value equal to the dry volume of the humidifier chamber being
used.
3. Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel all changes
and dismiss the dialog.
After making the necessary changes to the ventilator settings touch the Apnea tab
on the left side of the Setup dialog. Although changing the apnea settings is not
required, confirm the default settings are appropriate for the patient. Apnea ventila-
tion allows pressure control or volume control breath types. Parameters in pressure-
controlled apnea breaths include f, PI, TI O2%, and TA. Volume-controlled apnea
breath parameters are f, VT, VMAX, Flow pattern, O2%, and TA.
Note:
If Quick Start is chosen, and the apnea tab on the Vent Setup screen shows a yellow triangle,
indicating the apnea settings have not been reviewed.
2. Enter the desired apnea settings in the same manner as for the ventilator settings.
During apnea pressure ventilation, apnea rise time % is fixed at 50%, and the con-
stant parameter during a respiratory rate change is TI.
After accepting the apnea settings, the display returns once more to show the wave-
forms. Return to the Vent Setup dialog and touch the Alarms tab on the left side of
the GUI screen. The alarms screen appears with the default alarm settings. Reference
Alarms Settings Screen, p. 4-16. Review and adjust the alarm settings appropriately for
the patient.
Note:
If Quick Start is chosen, the alarms tab on the dialog shows a yellow triangle, indicating the
alarm settings have not been reviewed.
Note:
Reference Alarm Settings Range and Resolution, p. 11-17 for new patient default alarm values.
These defaults cannot be changed. The clinician can adjust alarm settings by following the
procedure below. The alarm settings are retained in memory when the ventilator’s power is
cycled, and current settings revert to new patient defaults when a new patient is selected.
2. Turn the knob clockwise to increase the value, or counter-clockwise to decrease the
value.
Note:
There is an additional alarm setting for TC, PAV+, VS, and VC+ breath types: High inspired
tidal volume (2VTI). This alarm condition occurs when the inspired tidal volume is larger than
the setting value. A 1VTI alarm will also cause breath delivery to transition to the exhalation
phase to avoid delivery of excessive inspiratory volumes.
WARNING:
Prior to initiating ventilation and whenever ventilator settings are changed, ensure
the alarm settings are appropriate for the patient.
WARNING:
Setting any alarm limits to OFF or extreme high or low values, can cause the
associated alarm not to activate during ventilation, which reduces its efficacy for
monitoring the patient and alerting the clinician to situations that may require
intervention.
Reference To adjust alarm volume, p. 3-39 to ensure alarm volume is adjusted prop-
erly.
Note:
A sample alarm tone sounds for verification at each volume level change. Re-adjust the
alarm volume by moving the alarm volume slider to increase or decrease the alarm volume.
Note:
Do not block the patient wye while the ventilator is waiting for a patient connection.
Otherwise the blockage could imitate a patient connection.
During ventilator operation, the alarm screen appears with indicators to let the oper-
ator know the current patient data value for each parameter (item 1), the parameter
alarm settings (items 2 and 3), recent range of patient data values for the last 200
breaths (item 4). If an alarm occurs, the slider and corresponding limit button show
a color matching the alarm’s priority. Reference Alarm Screen During Operation, p. 4-
17.
1 Pointers show current value of patient data cor- 3 Low alarm setting (in this case 4VTE SPONT)
responding to the alarm parameter
2 High alarm setting (in this case 2VTE SPONT) 4 Range of patient data values for the particular
parameter during the last 200 breaths
If, during ventilation, settings changes are necessary that don’t involve changes to
PBW, Mode, Breath types, or Trigger types, the current settings area located at the
lower portion of the GUI screen can be used. Reference Areas of the GUI, p. 4-3 for the
location of the current settings area.
2. Touch and turn the knob for any other settings that need to be changed.
The ventilator settings and the alarm settings chosen remain in memory after a
power cycle, as long as the same patient is chosen when the ventilator is set up
again. If a new patient is being ventilated, the ventilator and alarm settings revert to
their default values. If all power is lost (both AC and battery), the ventilator and alarm
settings in effect prior to the power loss are automatically restored if the power loss
duration is five (5) minutes or less. If the power loss lasts longer than five minutes,
ventilation resumes in Safety PCV. Ventilator and alarm settings must be reset for the
patient being ventilated. Reference Safety PCV Settings, p. 10-70 for a list of these set-
tings.
To use the Previous Setup button
1. To return to the previous settings, touch Previous Setup on the GUI screen. The ventilator
restores the main control and breath settings previously used, as well as the alarm and
apnea settings, and prompts a review by highlighting the previous values in yellow. The
ventilator, alarm, and apnea settings tab text is also shown in yellow and the tabs show
a yellow triangle, indicating there are previous settings that have not been reviewed.
The Previous Setup button disappears when the previous settings are confirmed and
re-appears when ventilating with new settings.
A breath timing graph appears at the bottom of the setup screen which illustrates
the relationship between inspiratory time, expiratory time, I:E ratio, respiratory rate,
and the effects on breath timing due to flow pattern, tidal volume, and VMAX during
mandatory PC, VC, BiLevel, or VC+ breaths. With BiLevel, PC and VC+ breaths, three
padlock icons are located underneath the breath timing graph allowing the opera-
tor to select, from left to right, TI I:E ratio, or TE as the constant variable during rate
changes (or TH, TH:TL ratio, or TL in BiLevel). If the ventilation mode is SPONT, the
padlock icons do not appear, and the breath timing graph only displays TI for a
manual inspiration. If the mandatory type is VC, the icons do not appear, but the
breath timing graph displays TI, I:E ratio, and TE.
To choose a constant timing variable for rate changes
1. Touch a padlock icon corresponding to the parameter to make constant during rate
changes (this changes the padlock’s appearance from unlocked to locked). The locked
parameter glows in the settings area.
3. Touch Accept.
respectively. Note that the Fenton tables provided the exclusive information for pre-
mature and infant data between 20 weeks and 50 weeks of fetal and gestational
growth.123
Note:
Any repeated values noted in the tables are the result of decimal rounding.
WARNING:
Use only non-vented patient interfaces with NIV. Leaks associated with vented
interfaces could result in the ventilator’s inability to compensate for those leaks,
even if Leak Sync is employed.
WARNING:
Full-face masks used for non-invasive ventilation should provide visibility of the
patient's nose and mouth to reduce the risk of emesis aspiration.
WARNING:
When using NIV, the patient’s exhaled tidal volume (VTE) could differ from the
ventilator’s monitored patient data VTE reading due to leaks around the mask. To
avoid this, ensure Leak Sync is installed. When NIV is selected, Leak Sync is
automatically enabled. Reference To enable Leak Sync, p. B-3.
Non-invasive ventilation (NIV) is used when the clinician determines a mask or other
non-invasive patient interface rather than an endotracheal tube would result in the
desired patient outcome.
NIV is intended for use by neonatal, pediatric, and adult patients possessing ade-
quate neural-ventilatory coupling and stable, sustainable, respiratory drive.
Covidien has successfully tested the following non-vented interfaces with NIV:
Full-face Mask — Puritan Bennett® Benefit Full Face Mask (large, part number 4-005253-00),
ResMed Mirage™ Non-Vented Full Face Mask (medium)
Infant Nasal Prongs — Sherwood Davis & Geck Argyle® CPAP Nasal Cannula (small),
Hudson RCI® Infant Nasal CPAP System (No. 3)
NIV can be initiated from either the New Patient Setup screen during Vent start-up
or while the patient is being ventilated invasively. Reference the table below for
using NIV patient setup information.
1. Turn the ventilator on. 1. Touch or swipe the menu tab on the left
side of the GUI.
2. Select New Patient.
2. Touch Vent Setup.
3. Enter patient’s PBW or gender and
height. 3. Perform steps 4 through 7 as if setting the
ventilator up for a new patient.
4. Touch NIV vent type.
4. Review the settings, including apnea and
5. Select mode. alarm settings and change if necessary.
WARNING:
For proper ventilation when changing the Vent Type on the same patient, review the
automatic settings changes described.Adjust appropriately based on the relevant
tables.
Some ventilator settings available during INVASIVE ventilation are not available
during NIV. Reference the following table for automatic settings changes when
changing vent type from INVASIVE to NIV.
Breath Mode: SIMV or SPONT High TI SPONT (2TI SPONT limit setting available
Alarm settings: 4PPEAK (if applicable), 4VE TOT, 4VTE Alarm settings: 4PPEAK, 4VE TOT 4VTE MAND, 4VTE
MAND, 4VTE SPONT INSPIRATION TOO LONG (not user- SPONT default to NIV new patient values. Reference
settable) Alarm Settings Range and Resolution, p. 11-17. INSPIRA-
TION TOO LONG alarm not available.
Note:
In any delivered spontaneous breath, either INVASIVE or NIV, if Pressure Support is set to 0
cmH2O, there is always a target inspiratory pressure of 1.5 cmH2O applied.
When in NIV, the Vent Setup button’s appearance changes, letting the operator
know the vent type is NIV.
Figure 4-9. Vent Setup Button “NIV” Indicating NIV vent type
The table below shows automatic settings changes made when changing vent type
from NIV to INVASIVE.
NIV includes a setting in SIMV or SPONT modes for High Spontaneous Inspiratory
Time limit (2TI SPONT). When a patient’s inspiratory time reaches or exceeds the set
limit, the ventilator transitions from inspiration to exhalation, and the 1TI SPONT
symbol appears at the lower left on the GUI screen, indicating the ventilator has
truncated the breath (shown below). The 2TI SPONT) setting does not restrict
changes to PBW; if the PBW is decreased, 2TI SPONT) may decrease automatically to
remain within its allowable limits.
WARNING:
No audible alarm sounds in conjunction with the visual 2TI SPONT indicator, nor does
the indicator appear in any alarm log or alarm message.
It is possible the target inspiratory pressure may not be reached if the 2TI SPONT
setting is not long enough, or if system leaks are so large as to cause the ventilator
to truncate the breath at the maximum allowable 2TI SPONT setting.
Note:
To reduce the potential for not reaching the target pressure, minimize the leaks in the
system and increase the Rise time % and/or decrease the ESENS setting, if appropriate.
Set the patient’s apnea parameters as described. Reference Apnea Settings, p. 4-14. NIV does not
change the way apnea parameters are set.
The system initially sets most alarm settings based on the patient’s PBW. Review all
alarm settings, and change as necessary, but startup does not require confirmation
of the settings. Alarm settings are made in exactly the same way in NIV as for INVA-
SIVE ventilation.
Touch the Alarms tab at any time during ventilation to show the current limits and
the monitored patient value shown in white on the indicating arrows for each alarm.
Reference Default NIV Alarm Settings, p. 4-26. If an alarm is occurring, the indicator
LED color changes based on alarm priority Reference Alarm Prioritization, p. 6-16 for
colors and meanings of alarms and their priorities.
Note:
The upper and lower limits of an alarm cannot conflict with each other.
Note:
The upper limits for the spontaneous exhaled tidal volume and mandatory exhaled tidal
volume alarms are always the same value. Changing the upper limit of one alarm
automatically changes the upper limit of the other.
2. Touch RM.
5. Accept or reject the maneuver results. If the result is accepted, its value is saved.
An inspiratory pause maneuver closes the inspiration and exhalation valves and
extends the inspiratory phase of a single, mandatory breath for the purpose of mea-
suring end inspiratory circuit pressure in order to calculate inspiratory Plateau Pres-
sure (PPL), lung Static Compliance (CSTAT), and Static Resistance (RSTAT) of the
respiratory system. Pressures on either side of the artificial airway are allowed to
equilibrate, which determines the pressure during a no-flow state. A request for an
inspiratory pause is ignored during apnea ventilation, safety PCV, OSC, BUV, and in
Stand-by state. Inspiratory pauses are allowed in A/C, SIMV, BiLevel and SPONT
modes. If an inspiratory pause maneuver has already occurred during the breath, a
second inspiratory pause maneuver is not allowed.
Inspiratory pauses can be classified as automatic or manual. The automatic inspi-
ratory pause lasts at least 0.5 second but no longer than three seconds. A manual
inspiratory pause starts by pressing and holding inspiratory pause key. The pause
lasts for the duration of the key-press (up to seven seconds).
To perform an automatic inspiratory pause
1. Press and release the inspiratory pause key on the GUI bezel or touch and release Start
if performing an inspiratory pause from the GUI screen as shown above. The ventilator
performs the inspiratory pause maneuver and displays PPL, CSTAT, and RSTAT along with
the date and time.
2. Touch the Accept or Reject button to save or dismiss results. If the Accept is touched, the
results are displayed.
2. Touch Accept or Reject to save or dismiss results. If Accept is touched, the results are dis-
played.
Cancel a manual inspiratory pause maneuver by releasing the Inspiratory Pause key.
An expiratory pause extends the expiratory phase of the current breath for the
purpose of measuring end expiratory lung pressure (PEEPTOT) or total PEEP. It has no
effect on the inspiratory phase of a breath, and only one expiratory pause per breath
is allowed. For I:E ratio calculation purposes, the expiratory pause maneuver is con-
sidered part of the exhalation phase.
During an expiratory pause, both inspiratory and exhalation valves are closed, allow-
ing the pressures on both sides of the artificial airway to equilibrate. This allows
intrinsic PEEP (PEEPI) to be calculated. PEEPI is PEEPTOT minus the set PEEP level. An
expiratory pause can be either automatically or manually administered, and is exe-
cuted at the next mandatory breath in A/C, SIMV, or BiLevel modes. In SIMV, the
breath cycle in which the pause becomes active (when the next scheduled ventila-
tor initiated mandatory (VIM) breath occurs) will be extended by the amount of time
the pause is active. For A/C and SIMV, the expiratory pause maneuver is scheduled
for the next end-of-exhalation prior to a mandatory breath. In BiLevel the expiratory
pause maneuver is scheduled for the next end-of-exhalation prior to a transition
from PL to PH. During the expiratory pause maneuver, PEEPI and PEEPTOT equilibra-
tion time values are displayed and regularly updated because stabilization of one of
these values can indicate the pause can be ended. During the expiratory pause, the
Apnea Interval TA is extended by the amount of time the pause is active. Expiratory
pause requests are ignored if the ventilator is in apnea ventilation, safety PCV,OSC,
BUV, and Stand-by state. Additionally, SEVERE OCCLUSION alarms are suspended
during expiratory pause maneuvers. If flow triggering is active, backup pressure sen-
sitivity (PSENS) detects patient breathing effort.
Maximum duration for a manual expiratory pause is 15 seconds and three (3)
seconds for an automatic expiratory pause.
During a manual or automatic expiratory pause, PEEPI and PEEPTOT appear on the
GUI with the next VIM to allow the clinician to view when these values are stabilized,
indicating the maneuver can be ended.
To perform an automatic expiratory pause
1. Press and release the expiratory pause key on the GUI or touch and release Start if per-
forming the expiratory pause from the GUI screen. The ventilator performs the expira-
tory pause maneuver and displays a circuit pressure graph, PEEPTOT, and PEEPI, along
with the date and time.
• Neonatal: 40% O2
• Pediatric/adult: 100% O2
Note:
The oxygen sensor can possess three states: Enabled, Disabled, and Calibrate. The oxygen
sensor is enabled at ventilator startup regardless if New Patient or Same Patient setup is
selected.
2. Touch the More Settings tab. The more settings screen appears.
3. Touch the button corresponding to the desired O2 sensor function (Enable or Disable).
4. Touch Accept.
The O2% setting can range from room air (21% O2) up to a maximum of 100%
oxygen. The sensor reacts with oxygen to produce a voltage proportional to the
partial pressure of the mixed gas. Since ambient atmosphere contains approximate-
ly 21% oxygen, the sensor constantly reacts with oxygen and always produces a
voltage. The useful life of the cell can also be shortened by exposure to elevated
temperatures and pressures. During normal use in the ICU, the oxygen sensor lasts
for approximately one year — the interval for routine preventive maintenance.
Because the oxygen sensor constantly reacts with oxygen, it requires periodic cali-
bration to prevent inaccurate O2% alarm annunciation. Once a calibrated oxygen
sensor and the ventilator reach a steady-state operating temperature, the moni-
tored O2% will be within three percentage points of the actual value for at least 24
hours. To ensure the oxygen sensor remains calibrated, recalibrate the oxygen
sensor at least once every 24 hours.
Typically, the clinician uses an O2 analyzer in conjunction with the information given
by the ventilator. If a NO O2 SUPPLY alarm occurs, compare the O2 analyzer reading
with the ventilator’s O2 reading for troubleshooting purposes. The ventilator auto-
matically switches to 100% air delivery.
The oxygen sensor should be calibrated every 24 hours and before use. The calibra-
tion function provides a single-point O2 sensor calibration.
To calibrate the O2 sensor
1. Touch Vent Setup.
3. Touch Calibrate for the O2 sensor. The oxygen sensor calibrates within two minutes. Ref-
erence More Settings Screen with O2 Sensor Enabled, p. 4-32.
3. Connect the ventilator oxygen hose to another known 100% O2 source (for example, a
second medical-grade oxygen cylinder).
4. Set O2% to each of the following values, and allow one minute after each for the mon-
itored value to stabilize: 21%, 40%, 90%
5. Watch the GUI screen to ensure the value for O2 (delivered O2%) is within 3% of each
setting within one minute of selecting each setting.
The first strategy is to detect potential problems before the ventilator is placed on a
patient. POST checks the integrity of the ventilator's electronics and prevents venti-
lation if a critical fault is found. Reference the Puritan Bennett™ 980 Series Ventilator
Service Manual for a complete description of POST). POST may detect major or minor
system faults which manifest themselves as Device Alerts. Reference DEVICE ALERT
Alarm, p. 6-34 for more information.
A technical fault occurs if a POST or background test has failed. Reference Power On
Self Test (POST), p. 10-74. Based on the test that failed, the ventilator will either venti-
late with current settings, ventilate with modified settings, or enter the Vent Inop
state. A technical fault cannot be cleared by pressing the alarm reset key. It can only
be cleared by correcting the fault that caused it or if alarm reset criteria have been
met.
4.11.3 SST
A procedure error occurs when the ventilator senses a patient connection before
ventilator setup is complete. The ventilator provides ventilatory support using
default Safety Pressure Controlled Ventilation (Safety PCV) settings. Reference Safety
PCV Settings, p. 10-70.
During ventilation, the ventilator performs frequent background checks of its breath
delivery subsystem (Reference Safety Net, p. 10-69). In the event that certain critical com-
ponents in the pneumatics fail, Ventilation Assurance provides for continued ventilatory
support using one of three Backup Ventilation (BUV) strategies, bypassing the fault to
maintain the highest degree of ventilation that can be safely delivered (Reference Back-
ground Diagnostic System, p. 10-72 for a full description of the Backup Ventilation strate-
gies).
Note:
Do not confuse BUV with Safety PCV, which occurs when a patient is connected before
ventilator setup is complete, or with Apnea ventilation, which occurs in response to patient
apnea.
In the event of a serious fault occurring that cannot be safely bypassed, the ventila-
tor, as a last resort, reverts to a Safe State. In Safe State, the ventilator opens the Safety
Valve and the Exhalation Valve, allowing the patient to breathe room air (if able to
do so), provided the patient circuit is not occluded. During SVO, the patient (if con-
nected) can breathe room air through the safety valve after it releases pressure in the
patient circuit. The patient exhales through the exhalation valve with minimal resis-
tance and the exhalation valve also acts like a check valve, limiting gas from being
drawn in through the expiratory filter or expiratory limb of the circuit. SVO condi-
tions are logged into the event and alarm logs as are the events leading to the SVO
condition. If the condition causing SVO clears, the ventilator clears the SVO state.
Patient data do not display on the GUI, but graphics are still plotted. During SVO, the
ventilator ignores circuit occlusions and disconnects. If the condition causing SVO
can only be corrected by servicing the ventilator, the SVO alarm cannot be reset by
pressing the alarm reset key.
Vent Inop occurs when the ventilator detects a catastrophic error which prevents all
other safety states from operating. Vent Inop limits pressure to the patient as the
ventilator enters the SVO state, disables (closes) the inspiratory valves (PSOLs), and
purges the gas mixing system accumulator. The safety valve is opened and a Vent
Inop indicator illuminates and a high priority alarm annunciates from the primary
alarm, and the secondary alarm (continuous tone) is activated. The ventilator can
only exit the Vent Inop state by power cycling and successfully passing EST. The Vent
Inop alarm cannot be reset with the alarm reset key. All detection and annunciation
of patient data alarm conditions is suspended.
5.1 Overview
This chapter describes the features of the Puritan Bennett™ 980 Series Ventilator
designed to provide output to the clinician. This includes language, methods of dis-
playing and transferring data, types of displayed data, and types of external device
ports. Connectivity to an external patient monitoring system is also included.
5.2 Language
The language used on the ventilator is configured at the factory.
• Waveform data: RS-232 port, USB port with USB to serial conversion capability (per
Comm port configuration)
5-1
Product Data Output
• Results from DCI commands: RS-232 port, USB port with USB to serial conversion capa-
bility (per Comm port configuration)
Caution:
The USB interface should be used for saving screen captures and interfacing with an
external patient monitor. It should not provide power to other types of devices
containing a USB interface.
Caution:
Only compatible USB devices should be used, otherwise GUI performance may be
impacted.
A 128 MB flash drive storage device formatted in the 32-bit file format is required for
downloading images from the USB ports. The USB device listed in Table 9-1. is the
ONLY compatible USB device currently available for use on the PB980. To order a
compatible USB device, contact Covidien Technical Services at 800 255 6774 or
contact a local Covidien representative.
To capture GUI screens
1. Navigate to the desired screen from which you wish to capture an image (for example,
the waveforms screen). There is no need to pause the waveform before performing the
screen capture.
2. Touch the screen capture icon in the constant access icons area of the GUI screen. If
desired, navigate to another screen and repeat steps 1 and 2 for up to ten (10) images.
If another image is captured, increasing the queue to eleven images, the newest image
overwrites the oldest image so there are always only ten images available.
Note:
If the camera icon appears dim, it means that the screen capture function is currently
processing images, and is unavailable. When processing is finished, the camera icon is
no longer dim and the screen capture function is available.
2. Touch Screen Capture. A list of screen captures appears, identified by time and date. A
slider also appears indicating more images than shown are present.
3. Insert a passive USB storage device (flash drive) into one of the USB ports at the rear of
the ventilator. The proper orientation of the USB device is with metal contacts facing the
test button. Reference Port Locations, p. 5-18. If more than one USB storage device is
installed in the ventilator, touch the button of the destination USB device where the
image will be copied. If an incompatible device is inserted, the port will be disabled until
the device is removed and removal is confirmed by touching the confirm button. The
message shown in Figure 5-1. appears.
Note:
Removal of the external USB storage device while screenshot files are being written to
it may result in incomplete file transfer and unusable files.
5. Touch Copy. The image is stored on the destination USB storage device.
6. Alternatively, touch Select All, and all images in the list are stored on the USB device and
which can then be viewed and printed from a personal computer.
Note:
The file format of screen captures is .PNG.
2. Touch the Comm Setup tab. The Comm Setup screen appears allowing three (3) ports
that can be configured. These ports can be designated as DCI, Philips, Spacelabs, or
Waveforms.
Note:
Waveforms can be selected on any port, but only on one port at a time.
Configuring the Comm port allows the ventilator to communicate with devices
listed in the Comm Setup screen, or to capture waveform data (in ASCII format) from
the ventilator.
3. Select the desired baud rate. If waveforms was selected, the baud rate automatically
becomes configured to 38400.
Connect the device to the previously configured port. Reference Port Locations, p. 5-
18 for a description and the locations of the Comm ports.
Note:
When a USB port is configured as a Comm port, it is necessary to use a USB-to-serial adapter
cable. This adapter must be based on the chipset manufactured by Prolific. For further
information, contact your Covidien representative.
Selecting waveforms when configuring a Comm port allows the ventilator to con-
tinuously transmit pressure, flow, and sequence numbers in ASCII format from the
selected serial port, at a baud rate of 38400 bits/s, and the operator- selected stop
bits, and parity. A sample of pressure and flow readings is taken every 20 ms. This
sample of readings is transmitted on the selected serial port at the end of each
breath at breath rates of 10/min and higher. For longer duration breaths, at least the
first eight seconds of the breath is transmitted. The format of the data is as follows:
The beginning of inspiration is indicated by “BS, S:nnn,<LF>“ where 'BS’ identifies the
Breath Start, ‘S:nnn’ is a sequence number incremented at every breath, and <LF> is
a line feed character. The fff, and ppp fields show the breath flow and pressure data.
The end of exhalation is indicated by: “BE<LF>“ where ‘BE’ indicates Breath End, and
<LF> is a line feed character.
The ventilator system offers commands that allow communication to and from the
ventilator using a Comm port. Commands to the ventilator from a remote device
include:
• RSET: Reference RSET Command, p. 5-6.
Note:
The ventilator responds only if it receives a carriage return <CR> after the command string.
The RSET command clears data from the ventilator receive buffer. The ventilator
does not send a response to the host system. Enter the RSET command exactly as
shown:
RSET<CR>
The SNDA command instructs the ventilator to send information on ventilator set-
tings and monitored patient data to the host system. Enter the SNDA command
exactly as shown:
SNDA<CR>
When the ventilator receives the command SNDA<CR>, it responds with the code
MISCA, followed by ventilator settings and monitored patient data information.
The MISCA response follows this format:
1 2 3 4 5 6 7
1 Response code to SNDA command 5 Data field, left-justified and padded with
spaces
Fields not available are marked as “Not used.” Underscores represent one or more
spaces that pad each character string.
The table below lists MISCA responses to SNDA commands.
Component Description
Field 6 Ventilator ID to allow external hosts to uniquely identify each Puritan Bennett™ 980 Venti-
lator (18 characters)
Field 9 Mode (CMV___, SIMV__, CPAP__ or BILEVL) (CMV = A/C) setting (6 characters)
Component Description
Field 41 Expiratory component of monitored value of I:E ratio, assuming inspiratory component of
1 (6 characters)
Field 48 High circuit pressure alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Field 51 Low exhaled tidal volume (mandatory or spontaneous) alarm status (NORMAL, ALARM_,
or RESET_) (6 characters)
Field 52 Low exhaled minute volume alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Field 53 High respiratory rate alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Component Description
Field 63 Static compliance (CSTAT) from inspiratory pause maneuver in mL/cmH2O(6 characters)
Field 64 Static resistance (RSTAT) from inspiratory pause maneuver in cmH2O/L/s (6 characters)
Field 96 Inspiratory component of I:E ratio or High component of H:L (BiLevel) setting
(6 characters)
Field 97 Expiratory component of I:E ratio setting or Low component of H:L (BiLevel) (6 characters)
Component Description
Field 100 Constant during rate setting change for pressure control mandatory breaths
(I-TIME or I/E___ or______) (6 characters) (where ______ represents TEor PCV not active)
SNDF is a command sent from an external host device to the ventilator system
instructing it to transmit all ventilator settings data, monitored patient data, and
alarm settings and occurrences. Enter the SNDF command exactly as shown:
SNDF<CR>
When the ventilator receives the command SNDF<CR>, it responds with the code
MISCF, followed by ventilator settings, monitored patient data, and alarm informa-
tion
The MISCF response follows this format:
1 2 3 4 5 6 7
1 Response code to SNDF command 5 Data field, left-justified and padded with
spaces
4 Start of transmission (02 hex) * 1229 if Philips is selected for serial port in
communication setup
Note:
Non-applicable fields will either contain zero or be blank.
Component Description
1225* Number of bytes between <STX> and <CR> (4 characters) *1229 if Phillips is selected for
the Comm port in Communication Setup
Field 6 Ventilator ID to allow external hosts to uniquely identify each Puritan Bennett™ 980 Venti-
lator (18 characters)
Component Description
Field 29 Inspiratory component of Apnea I:E ratio (if apnea mandatory type is PC) (6 characters)
Field 30 Expiratory component of Apnea I:E ratio (if apnea mandatory type is PC) (6 characters)
Field 35 Low inspiratory pressure alarm setting (4PPEAK) in cmH2O or OFF (6 characters)
Field 36 High exhaled minute volume (2VE TOT) alarm setting in L/min or OFF (6 characters)
Field 37 Low exhaled minute volume (4VE TOT) alarm setting in L/min or OFF (6 characters)
Field 38 High exhaled mandatory tidal volume (2VTE MAND) alarm setting in mL or OFF
(6 characters)
Field 39 Low exhaled mandatory tidal volume (4VTE MAND) alarm setting in mL or OFF
(6 characters)
Field 40 High exhaled spontaneous tidal volume (2VTE SPONT alarm setting in mL or OFF
(6 characters)
Field 41 Low exhaled spontaneous tidal volume (4VTE SPONT) alarm setting in mL or OFF
(6 characters)
Field 42 High respiratory rate (2fTOT) alarm setting in breaths/min or OFF (6 characters)
Component Description
Field 48 Inspiratory component of I:E ratio setting or High component of H:L ratio setting
(6 characters)
Field 49 Expiratory component of I:E ratio setting or Low component of H:L ratio setting
(6 characters)
Field 50 Constant during rate change setting (I-time, I/E, or E-time) (6 characters)
Field 53 Humidification type setting (Non-heated exp tube, Heated exp tube, or HME)
(18 characters)
Field 59 Expiratory sensitivity (ESENS) setting in % or L/min for PAV+ breath type (6 characters)
Field 64 High pressure time (TH) setting (in BiLevel) in seconds (6 characters)
Field 65 High spontaneous inspiratory time limit (2TI SPONT) setting in seconds (6 characters)
Field 67 Low pressure time (TL) setting (in BiLevel) in seconds (6 characters)
Component Description
Field 75 Expiratory component of monitored value of I:E ratio, assuming inspiratory component of
1 (6 characters)
Field 90 Intrinsic PEEP (PEEPI) from expiratory pause maneuver in cmH2O (6 characters)
Field 91 Total PEEP (PEEPTOT from expiratory pause maneuver in cmH2O (6 characters)
Field 92 Static compliance (CSTAT) from inspiratory pause maneuver in mL/cmH2O (6 characters)
Field 93 Static resistance (RSTAT) from inspiratory pause maneuver in cmH2O/L/s (6 characters)
Field 94 Plateau pressure (PPL) from inspiratory pause maneuver in cmH2O (6 characters)
Component Description
Field 107 High exhaled minute volume alarm* (1VE TOT) (6 characters)
Field 117 Low exhaled mandatory tidal volume alarm* (3VTE MAND) (6 characters)
Field 118 Low exhaled minute volume alarm* (3VE TOT) (6 characters)
Field 119 Low exhaled spontaneous tidal volume (3VTE SPONT) alarm* (6 characters)
Component Description
Field 129 High inspired spontaneous tidal volume* (3TI SPONT) alarm (6 characters)
Component Description
WARNING:
To avoid possible injury, only connect devices that comply with IEC 60601-1 standard
to any of the ports at the rear of the ventilator, with the exception
of passive memory storage devices (“flash drives”) and serial-to-USB adapter cables.
If a serial-to-USB adapter cable is used, it must be connected to an IEC 60601-1-
compliant device.
WARNING:
To avoid possible injury, do not connect a device that is attached to the patient to
any of the non-clinical ports listed below when the ventilator is ventilating a patient.
RS-232 Port
3. Ensure to specify the baud rate, parity, and data bits in the ventilator communication
setup to correctly match the parameters of the monitoring device.
4. A monitor designed to use this port is required for obtaining data from the ventilator.
Set up the monitoring device to receive ventilator data. These data can include wave-
form data.
5. Program the remote device to send the appropriate RS-232 commands as described in
the next section.
Reference MISCA Response, p. 5-7 and MISCF Response, p. 5-11 for MISCA and MISCF
responses to SNDA and SNDF commands.
Ethernet Port
The Ethernet port is used by Service personnel for accessing various logs and updat-
ing ventilator software.
A remote alarm or nurse call interface is available on the ventilator system which can be
used to remotely annunciate the alarm status of the ventilator. Medium and high priority
alarms are remotely annunciated. The nurse call connector is located at the back of the
ventilator, as shown. Reference Port Locations, p. 5-18.
Reference the remote alarm manufacturer’s instructions for use for information
regarding proper nurse call connection.
USB Ports
The USB ports are used for screen captures, or receiving serial data when a USB port
has been configured as a serial port. This is also known as transferring data via a serial
over USB protocol. Reference Communication Setup, p. 5-4 for Comm setup config-
uration. Screen captures require an external USB memory storage device (“flash
drive”) for screen captures. Instructions for using this port for screen captures are
given. Reference To capture GUI screens, p. 5-2.
HDMI Port
An external display can be used via connection with the HDMI port.
To use the HDMI port with an external display
1. Connect one end of an HDMI cable to the HDMI port at the back of the ventilator (item
6, above).
2. Connect the other end of the cable to the external display. An HDMI to DVI adapter may
be used.
3. Turn the device on. The appearance of the GUI now displays on the external display
device.
Service Port
Note:
Not all patient monitors are compatible with the Puritan Bennett™ 980 Series Ventilator.
6.1 Overview
This chapter contains detailed information about Puritan Bennett™ 980 Series Ven-
tilator performance including:
• Ventilator settings
WARNING:
Use of the ventilator/compressor in altitudes higher or barometric pressures lower
than those specified could compromise ventilator/compressor operation. Reference
Environmental Specifications, p. 11-8 for a complete list of environmental
specifications.
6-1
Performance
The clinician enters the vent type, specifying how the patient will be ventilated; inva-
sively or non-invasively (NIV). The vent type optimizes the alarm limits for NIV
patients, and disables some settings for NIV ventilation.
6.4.2 Mode
• SPONT (Spontaneous) — SPONT mode delivers only spontaneous breaths which are
all patient-initiated.
• BiLevel — BiLevel is also a mixed mode which overlays the patient’s spontaneous
breaths onto the breath structure for PC mandatory breaths. Two levels of pressure, PL
and PH are employed. The breath cycle interval for both SIMV and BiLevel modes is 60/
f where f is the respiratory rate set by the operator.
• CPAP — CPAP is available only when circuit type is neonatal and vent type is NIV. CPAP
mode allows spontaneous breathing with a desired PEEP level. In order to limit inadver-
tent alarms associated with the absence of returned volumes in nasal CPAP breathing,
CPAP does not make available exhaled minute volume and exhaled tidal volume alarm
settings.
Mandatory breath types for A/C and SIMV modes include volume controlled (VC),
pressure controlled (PC), or volume control plus (VC+) breath types, also called Man-
datory Type.
• VC (Volume Control) — The ventilator delivers an operator-set tidal volume.
• VC+ (Volume Control Plus) — Volume control plus (a mandatory, pressure controlled
breath type that does not restrict flow during the inspiratory phase, and automatically
adjusts the inspiratory pressure target from breath to breath to achieve the desired tidal
volume despite changing lung conditions. Reference Mandatory Breath Delivery, p. 10-
16 for more information on VC+.
• Flow Trigger (VTRIG) — Changes in flow in the circuit cause the ventilator to deliver a
breath. The breath delivery and exhalation flow sensors measure gas flow in the venti-
lator breathing system. As the patient inspires, the delivered flow remains constant, and
the exhalation flow sensor measures decreased flow. When the difference between the
two flow measurements is at least the operator-set value for flow sensitivity (VSENS), the
ventilator delivers a breath.
• Time Trigger — The ventilator delivers a breath after a specific amount of time elapses.
• Operator Trigger (OIM) — The operator presses the Manual inspiration key. An opera-
tor initiated mandatory breath is also called an OIM breath. During an OIM breath, the
breath delivered is based on the current settings for a mandatory breath.
Spontaneous breathing modes such as SIMV, BiLevel, and SPONT include the follow-
ing breath types (called Spontaneous Types):
• PS (Pressure Support) — The ventilator delivers an operator-set positive pressure
above PEEP (or PL in BiLevel) during a spontaneous breath. If SIMV is selected as the
mode, PS is automatically selected for spontaneous type.
• PAV+ (Proportional Assist Ventilation) — A software option that allows the ventilator
to reduce the work of breathing (WOB) by assisting the patient’s inspiration by an oper-
ator-set amount proportional to the pressure generated by the patient. Reference
Appendix C for more information on PAV+.
• Operator Trigger (OIM) — Since the operator can only initiate a mandatory breath by
pressing the Manual inspiration key, spontaneous mode allows OIMs, but the breath
delivered is based on the current settings for an apnea breath.
Reference Inspiration — Detection and initiation, p. 10-4 for details on the different
trigger methods.
6.5 Alarms
WARNING:
The ventilator system is not intended to be a comprehensive monitoring device and
does not activate alarms for all types of conditions. For a detailed understanding of
ventilator operations, be sure to thoroughly read this manual before attempting to
use the ventilator system.
WARNING:
Setting any alarm limits to OFF or extreme high or low values can cause the
associated alarm not to activate during ventilation, which reduces its efficacy for
monitoring the patient and alerting the clinician to situations that may require
intervention.
This manual uses the following conventions when discussing alarms:
A description or name of an alarm without specifying the alarm setting is denoted
with an upward or downward pointing arrow (1 or 3) preceding the specific alarm
name. An alarm setting is denoted as an upward or downward pointing arrow with
an additional horizontal limit symbol (2or 4) preceding the specific alarm. Some
alarm conditions actually limit breath delivery such as 1PPEAK and 1VTI by truncating
inspiration and transitioning to the exhalation phase. These alarm conditions are
denoted as alarm limits. Reference Alarm Descriptions and Symbols, p. 6-7.
Alarms are visually annunciated using an indicator on the top of the GUI, which has
a 360° field of view. If an alarm occurs, this indicator flashes at a frequency and color
matching the alarm priority. The alarms also appear as colored banners on the right
side of the GUI screen. If an alarm occurs, this indicator appears in the color match-
ing the alarm priority (yellow for low (!) and medium (!!) priority; red for high (!!!) pri-
ority. For technical alarm and non-technical alarm details, reference the respective
tables on p. 6-18 and p. 6-28.
An alarm is defined as a primary alarm if it is the initial alarm. A dependent alarm
arises as a result of conditions that led to the primary alarm. This is also referred to as
an augmentation. An augmentation strategy is built into the ventilator software to
handle occurrences where the initial cause of the alarm has the potential to precip-
itate one or more additional alarms. When an alarm occurs, any subsequent alarm
related to the cause of this initial alarm augments the initial alarm instead of appear-
ing on the GUI as a new alarm. The initial alarm’s displayed analysis message is
updated with the related alarm’s information, and the Alarm Log Event column
shows the initial alarm as Augmented.
A primary alarm consists of a base message, analysis message, and a remedy mes-
sage. The base message describes the primary alarm. The analysis message
describes the likely cause of the alarm and may include alarm augmentations. The
remedy message provides information on what to do to correct the alarm condition.
Alarm banners, when dragged leftward from the right side of the GUI, display mes-
sages for the indicated active alarms. The figure below, shows the alarm message
format.
2 Analysis message
A latched alarm is one whose visual alarm indicator remains illuminated even if the
alarm condition has autoreset. Latched alarm indicators are located on the sides of
the omni-directional LED. A latched alarm can be manually reset by pressing the
alarm reset key. If no alarms are active, the highest priority latched alarm appears on
the omni-directional LED on the GUI. A lockable alarm is one that does not termi-
nate an active audio paused function (it does not sound an audible alert during an
active audio paused function), while a non-lockable alarm cancels the audio paused
period and sounds an audible alert. All patient data alarms and the CIRCUIT DISCON-
NECT alarm are lockable alarms.
Note:
When a new lockable alarm occurs, the alarm will not start to sound audibly if the previous
lockable alarm was silenced.
The following rules define how alarm messages are displayed:
• Primary alarms precede any dependent alarms.
• The system adds dependent alarms to the analysis messages of each active primary
alarm with which they are associated. If a dependent alarm resets, the system removes
it from the analysis message of the primary alarm.
• The priority level of a primary alarm is equal to or greater than the priority level of any of
its active dependent alarms.
• If a primary alarm resets, any active dependent alarms become primary unless they are
also dependent alarms of another active primary alarm. This is due to different reset cri-
teria for primary and dependent alarms.
• The system applies the new alarm limit to alarm calculations from the moment a
change to an alarm limit is accepted.
• The priority level of a dependent alarm is based solely on its detection conditions (not
the priority of any associated alarms.
• When an alarm causes the ventilator to enter OSC, or safety valve open (SVO), the
patient data display (including waveforms) is blanked. The elapsed time without venti-
latory support (that is, since OSC, or SVO began) appears on the GUI screen. If the alarm
causing OSC, or SVO is autoreset, the ventilator resets all patient data alarm detection
algorithms.
The alarm reset function can be used for any non-technical alarm. Reference Alarm
Handling, p. 6-16 for an explanation of technical vs. non-technical alarms. Alarm
reset reinitializes the algorithm the ventilator uses to initially detect the alarm except
for A/C POWER LOSS, COMPRESSOR INOPERATIVE, LOW BATTERY, NO AIR SUPPLY,
NO O2 SUPPLY, PROCEDURE ERROR alarms and active battery alarms. If the cause of
the alarm still exists after the Alarm Reset key is pressed, the alarm becomes active
again. The ventilator logs all actuations of the alarm reset key.
WARNING:
Do not pause, disable, or decrease the volume of the ventilator's audible alarm if
patient safety could be compromised.
The audio paused feature temporarily mutes the audible portion of an alarm for two
minutes. After the two-minute period, if the alarm condition still exists, the alarm
sounds again. Pressing the audio paused key again re-starts the two minute interval
during which an alarm is muted. An LED within the key illuminates and a count-
down timer appears on the GUI next to an audio paused indicator symbol, indicat-
ing an active audio paused function. The audio paused feature does not allow the
audible alarm to be turned off; the audible portion of the alarm is temporarily muted
for two minutes. The GUI’s omni-directional LED flashes during an active alarm state,
and during an audio paused period and its appearance changes with the priority if
the alarm escalates. Pressing the Alarm Reset key cancels an audio paused condition.
If the condition that caused the alarm still exists, the alarm activates again.
An alarm volume key is available for setting the desired alarm volume. The alarm
volume is automatically set to the factory default setting of 10 (maximum) or to the
institutional default setting based on circuit type if it has been so configured. When
setting the alarm volume, a sample tone is generated, allowing the practitioner to
decide the appropriate alarm volume for the surrounding ambient conditions. If a
high priority alarm occurs, the alarm volume increases one (1) increment from its
current volume level if it is not acknowledged within 30 s. If a high priority alarm is
not acknowledged within 60 s, the audible alarm volume escalates to its maximum
volume.
Reference To adjust alarm volume, p. 3-39 for instructions on adjusting the alarm
volume.
WARNING:
The audio alarm volume level is adjustable. The operator should set the volume at a
level that allows the operator to distinguish the audio alarm above background
noise levels. Reference To adjust alarm volume, p. 3-39.
Testing the alarms requires oxygen and air sources and stable AC power. Test the
alarms at least every six months, using the procedures described.
Required Equipment
• Test lung (P/N 4-000612-00)
If the alarm does not annunciate as indicated, verify the ventilator settings and
repeat the test. The alarm tests check the operation of the following alarms:
• CIRCUIT DISCONNECT
• SEVERE OCCLUSION
• AC POWER LOSS
• APNEA
• NO O2 SUPPLY
1. Disconnect the patient circuit from the ventilator and turn the ventilator off for at least
five minutes.
PBW: 70 kg
Vent type: INVASIVE
Mode: A/C
Mandatory type: VC
Trigger type: VTRIG
5. Set the following new patient settings
f: 6.0 1/min
VT: 500 mL
VMAX: 30 L/min
TPL: 0 s
Flow pattern: SQUARE
VSENS: 3 L/min
O2%: 21%
PEEP: 5 cmH2O
6. Set the following apnea settings
TA: 10 s
f: 6.0 1/min
O2%: 21%
VT: 500 mL
7. Set the following alarm settings
2PPEAK: 70 cmH2O
fTOT: OFF
4VE TOT: 1 L/min
8. Set the graphics display to a volume-time plot (for use in the APNEA alarm test).
9. Connect an Adult patient circuit to the ventilator and attach a test lung to the patient
wye.
Note:
To ensure proper test results, do not touch the test lung or patient circuit during the CIRCUIT
DISCONNECT alarm test.
2. Connect the inspiratory filter to the To Patient port to autoreset the alarm.
VT: 500 mL
VMAX: 30 L/min
2PPEAK: 20 cmH2O
2. After one breath, the ventilator annunciates a HIGH CIRCUIT PRESSURE (1PPEAK) alarm. If
the alarm does not sound, check the patient circuit for leaks.
2PPEAK: 50 cmH2O
4. Disconnect the ventilator breathing circuit from the FROM PATIENT port and block the
gas flow.
5. While maintaining the occlusion, ensure the safety valve open indicator appears on the
status display, the GUI shows the elapsed time without normal ventilation support, and
the test lung inflates and deflates rapidly with small pulses as the ventilator delivers trial
pressure-based breaths.
2. Connect the power cord to AC facility power. The AC POWER LOSS or LOW BATTERY
alarm autoresets.
2PPEAK: 70 cmH2O
Mode: SPONT
Spontaneous type: PS
2. The GUI annunciates an APNEA alarm within 10 s after touching Accept.
3. Squeeze the test lung twice to simulate two subsequent patient-initiated breaths. The
APNEA alarm autoresets.
Note:
To avoid triggering a breath during the apnea interval, do not touch the test lung or patient
circuit.
Note:
For the apnea alarm test, the exhaled tidal volume (VTE) displayed in the patient data area
must be greater than half the delivered volume shown on the volume-time plot in the
graphics display in order for apnea to autoreset. Reference Apnea Ventilation, p. 10-39 for a
technical description of apnea ventilation.
LOW EXHALED SPONTANEOUS TIDAL VOLUME alarm test
1. Make the following patient and alarm settings changes
3. Slowly squeeze the test lung to simulate spontaneous breaths. The ventilator annunci-
ates a LOW EXHALED SPONTANEOUS TIDAL VOLUME (3VTE SPONT) alarm at the start of
the fourth consecutive spontaneous inspiration.
Mode: A/C
4VTE SPONT: OFF
5. Press the alarm reset key to reset the 4VTE SPONT alarm.
2. Connect the oxygen inlet supply. The NO O2 SUPPLY alarm autoresets within two
breaths after oxygen is reconnected.
PSENS: 2 cmH2O
O2%: 100%
2. Make the following apnea settings changes:
TA: 60 s
3. Attach the ventilator’s oxygen gas hose to a known air supply (for example, a medical
grade air cylinder) or a wall air outlet.
4. Attach the ventilator’s air gas hose to a known medical oxygen supply.
5. Observe the GUI screen. The delivered O2% display should decrease, and the ventilator
should annunciate a medium priority 3O2% alarm within 60 s and a high priority 3O2%
alarm within two (2) minutes.
7. Observe the GUI screen. The delivered O2% display should increase, and the ventilator
should annunciate a a medium priority 1O2% alarm within 60 s and a high priority 1O2%
alarm within two (2) minutes.
8. Remove the air gas hose from the oxygen supply and reconnect the hose to a known
medical air supply.
9. Remove the oxygen gas hose from the air supply and reconnect the hose to a known
oxygen supply.
WARNING:
Before returning the ventilator to service, review all settings and set appropriately
for the patient to be ventilated.
When an alarm occurs, the omni-directional LED at the top of the GUI flashes in a
color corresponding to the alarm priority, an audible series of tones sounds, and an
alarm banner displays on the GUI. Reference Areas of the GUI, p. 4-3. When the alarm
banner appears, it displays its base message. Touching the individual alarm causes
an expanded explanation to appear, containing analysis and remedy messages, and
may contain a link to the alarm log or the alarms settings screen. Touch the link to
display requested information. The omni-directional LED remains steadily lit and
may appear multicolored, meaning that multiple alarms with varying priority levels
have occurred. During an event that causes multiple alarms, the ventilator simulta-
neously displays the two highest priority active alarms.
The delay time from the moment the alarm condition first occurs until the alarm is
annunciated is imperceptible.
For alarm conditions relayed via the serial port, the overall delay is dependent upon
the polling rate of the external device. The delay from the time the serial port is
polled by the external device, until the alarm message leaves the serial port does not
exceed three (3) seconds. An example of an external device is a patient monitor.
Current alarm settings are saved in the ventilator’s non-volatile memory (NVRAM).
If the alarm settings are changed by another clinician, those settings become appli-
cable. For example, there are no operator-selectable default alarm settings.
The ventilator system’s alarm handling strategy is intended to
• Detect and call attention to legitimate causes for caregiver concern as quickly as possi-
ble, while minimizing nuisance alarms.
• Identify the potential cause and suggest corrective action for certain types of alarms.
However, the clinician must make the final decision regarding any clinical action.
Ventilator alarms are categorized as high priority, medium priority, or low priority,
and are classified as technical or non-technical.
The ventilator is equipped with two alarms — the primary alarm secondary alarms.
The primary alarm annunciates high, medium, and low priority alarms when they
occur. The secondary alarm (also named “immediate” priority in the table below) is
a continuous tone alarm and annunciates during Vent Inop conditions or complete
loss of power. This alarm is powered by a capacitor and lasts for at least 120 seconds.
The table below lists alarm priority levels and their visual, audible, and autoreset
characteristics. An alarm autoresets when the condition causing the alarm no longer
exists.
High: Immediate atten- Flashing red LED located High-priority audible Visual alarm does not
tion required to ensure on the top of the GUI, red alarm (a sequence of five auto reset. Visual alarm
patient safety. alarm banner on GUI tones that repeats twice, indicators remain steadi-
screen, red bar next to pauses, then repeats ly illuminated following
alarm setting icon on again). an autoreset. The alarm
Alarms screen. reset key must be
pressed to extinguish
visual indicator.
Medium: Prompt atten- Flashing yellow LED Medium-priority audible LED indicator turns off
tion necessary. located on the top of the alarm (a repeating and autoreset is entered
GUI, yellow alarm banner sequence of three into the alarm log.
on GUI screen, and tones).
yellow bar next to alarm
setting icon on Alarms
screen.
Low: A change in the Steadily illuminated Low-priority audible LED indicator turns off
patient-ventilator system yellow LED located on alarm (two tone, non- and autoreset is entered
has occurred. the top of the GUI, yellow repeating). into the alarm log.
alarm banner on GUI
screen, and yellow bar
next to alarm setting
icon on Alarms screen.
A technical alarm is one that is caused by a violation of any of the ventilator’s self
monitoring conditions, such as failure of POST or a fault detected by the ventilator’s
background diagnostic system. This includes faults detected by the ventilator’s
background diagnostic system. Technical alarms cannot be reset by pressing the
Alarm Reset key. (Reference Background Diagnostic System, p. 10-72. Technical alarms
fall into eight categories, shown in the table below.
Reference the table below for a list of ventilator technical alarms, their meaning, and
what to do if they occur.
Reference Alarm Settings Range and Resolution, p. 11-17 for the settings, ranges, reso-
lutions, new patient default values, and accuracies of all the ventilator alarms.
DEVICE ALERT Various. Technical alarm category Follow remedy message displayed
is described. on GUI.
Reference Technical Alarm Catego-
ries, p. 6-18. More information for
the particular technical alarm can
be found in the System diagnostic
log, a link to which is provided on
the expanded alarm banner.
AC POWER LOSS Low Operating on vent main bat- N/A Ventilator automati-
tery. cally switches to
battery power. Power
AC POWER LOSS Low Operating on vent main and N/A switch ON. AC power
compressor battery. not available. Battery
operating indicator
on status display
turns on. Resets when
AC power is restored.
APNEA (patient Medium Apnea ventilation. Breath inter- Check patient & The set apnea interval
data alarm) val > apnea interval. settings. has elapsed without
the ventilator, patient,
High Extended apnea duration or or operator triggering
multiple apnea events. a breath. Resets after
patient initiates a
third consecutive
breath. Possible
dependent alarm:
3VE TOT
1PPEAK (patient Low Last breath ≥ set limit. Check patient, Measured airway
data alarm) circuit & ET tube. pressure ≥ set limit.
Medium Last 3 breaths ≥ set limit. Ventilator truncates
current breath unless
High Last 4 or more breaths ≥ set
already in exhalation.
limit.
Possible dependent
alarms: 3VTE MAND,
3 VE TOT, 1fTOT. Cor-
rective action: Check
patient. Check for
leaks, tube type/ID
setting. Consider
reducing % Supp
setting or increasing
2PPEAK.
1PCOMP (patient Low Last spont breath ≥ set PPEAK In TC: Pressure of sponta-
data alarm) limit - 5 cmH2O. neous breaths ≥ set
• Check for limit. Possible depen-
Medium Last 3 spont breaths ≥ set PPEAK leaks, tube dent alarms: 3VTE
limit - 5 cmH2O. type/ID SPONT,
setting. 3 VE TOT, 1fTOT,
High Last 4 or more spont breaths ≥ 3VTE SPONT
set PPEAK limit- 5 cmH2O. In PAV+:
Corrective action:
Check for leaks. Check
• Check for
for the correct tube
leaks, tube
type.
type/ID
Check that the tube
setting.
inside diameter corre-
sponds to the patient
PBW.
Check the 2PPEAK set-
ting.
3PPEAK (patient Low Last 2 breaths, pressure ≤ set Check for leaks. Peak inspiratory pres-
data alarm) limit. sure ≤ alarm setting.
(Available only when
Medium Last 4 breaths, pressure ≤ set Mandatory Type is
limit. VC+* or when Vent
Type is NIV. Target
High Last10 or more breaths, pres-
pressure = the low
sure ≤ set limit.
limit: PEEP + 3
cmH2O. Ventilator
cannot deliver target
volume. Possible
dependent alarms:
1fTOT.
Corrective action:
Check patient and
settings.
* Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall below PEEP
+ 3 cmH2O, attempting to set the 4PPEAK alarm limit at or below this level will turn the alarm off.
1O2% (patient Medium Measured O2% > set for ≥ 30 s Check patient, The O2% measured
data alarm) but < 2 min. gas sources, O2 during any phase of a
analyzer & venti- breath cycle is 7%
High Measured O2% > set for ≥ 2 lator. (12% during the first
min. hour of operation) or
more above the O2%
setting for at least 30
seconds. (These per-
centages increase by
5% for four minutes
following a decrease
in the O2% setting.)
3O2% (patient High Measured O2% < set O2%. Check patient, The O2% measured
data alarm) gas sources, O2 during any phase of a
analyzer & venti- breath cycle is 7%
lator. (12% during the first
hour of operation) or
more below the O2%
setting for at least 30
seconds, or below
18%. (These percent-
ages increase by 5%
for four minutes fol-
lowing an increase in
the O2% setting.)
1VTE (patient Low Last 2 breaths ≥ set limit. Check settings, Exhaled tidal volume
data alarm changes in ≥ set limit. Alarm
Medium Last 4 breath s≥ set limit. patient's R&C. updated whenever
exhaled tidal volume
High Last 10 or more breaths ≥ set
is recalculated. Possi-
limit.
ble dependent alarm:
1VE TOT
1VE TOT(patient Low VE TOT ≥ set limit for ≤ 30 s. Check patient Expiratory minute
data alarm) and settings. volume ≥ set limit.
Medium VE TOT ≥ set limit for > 30 s. Alarm updated when-
ever an exhaled
High VE TOT ≥ set limit for > 120 s. minute volume is
recalculated. Possible
dependent alarm:
1VTE.
1fTOT (patient Low fTOT ≥ set limit for ≤ 30 s. Check patient & Total respiratory rate
data alarm) settings. ≥ set limit. Alarm
Medium fTOT ≥ set limit for > 30 s. updated at the begin-
ning of each inspira-
High fTOT ≥ set limit for > 120 s. tion. Reset when
measured respiratory
rate falls below the
alarm limit. Possible
dependent alarms:
3VTE MAND,
3VTE SPONT, 3VE TOT
1PVENT (patient Low 1 breath ≥ limit. Check patient, Inspiratory pressure >
data alarm) circuit & ET tube. 110 cmH2O and man-
Medium 2 breaths ≥ limit.
datory type is VC or
High 3 or more breaths ≥ limit. spontaneous type is
TC or PAV+. Ventilator
truncates current
breath unless already
in exhalation. Possible
dependent alarms:
3VTE MAND,
3VE TOT 1fTOT.
Corrective action:
1. Check patient
for agitation.
Agitated
breathing, com-
bined with high
% Supp setting
in PAV+ can
cause over-
assistance. Con-
sider reducing
% Supp setting.
2. Provide alter-
nate ventilation.
Remove ventila-
tor from use and
contact Service.
INSPIRATION Low Last 2 spont breaths = PBW Check patient. Inspiratory time for
TOO LONG based TI limit. Check for leaks. spontaneous breath
(patient data ≥ PBW-based limit.
alarm) Medium Last 4 spont breaths = PBW Ventilator transitions
based TI limit. to exhalation. Resets
when TI falls below
High Last 10 or more spont breaths = PBW-based limit.
PBW based TI limit. Active only when
Vent Type is INVASIVE.
PAV STARTUP Low PAV startup not complete for ≥ Check for leaks, Unable to assess
TOO LONG 45 s. shallow breath- patient’s resistance
(patient data ing, & settings and compliance
alarm) (occurs Medium PAV startup not complete for ≥ for during PAV startup.
only if PAV+ is in 90 s. 1VTI and 1PPEAK. Possible dependent
use) alarms 3VTE SPONT,
High PAV startup not complete for ≥
120 s. 3VE TOT, 1fTOT. Correc-
tive action: Check
patient. (Patient’s
inspiratory times may
be too short to evalu-
ate resistance and
compliance.) Check
that selected humidi-
fication type and
empty humidifier
volume are correct.
PAV R & C NOT Low R and/or C over 15 minutes old. Check for leaks, Unable to assess resis-
ASSESSED shallow breath- tance and/or compli-
(patient data Medium R and/or C over 30 minutes old. ing, & settings ance during PAV+
alarm) (occurs for tube ID, 1VTI steady-state. Startup
only if PAV+ is in and 1PPEAK. was successful, but
use) later assessments
were unsuccessful.
Corrective action:
Check patient.
(Patient’s inspiratory
times may be too
short to evaluate
resistance and com-
pliance.) Check that
selected humidifica-
tion type and empty
humidifier volume are
correct.
LOW BATTERY Medium Vent main battery operational Replace or allow Resets when battery
time < ten minutes. recharge vent has ≥ ten minutes of
main battery. operational time
remaining.
LOW BATTERY Medium Compressor battery operation- Replace or allow
al time < ten minutes. recharge com-
pressor battery.
LOW BATTERY Medium Vent main battery operational Replace or allow Resets when main
time < ten minutes and com- recharge vent battery or compressor
pressor battery operational main battery battery has ≥ ten
time < 10 minutes. and compressor minutes of operation-
battery. al time remaining or
when AC power is
restored.
LOW BATTERY High Vent main battery operational Replace or allow Resets when battery
time < five minutes. recharge vent has ≥ five minutes of
main battery. operational time
remaining or when
LOW BATTERY High Compressor battery operation- Replace or allow AC power is restored.
al time < five minutes. recharge com-
pressor battery.
LOW BATTERY High Vent main battery operational Replace or allow Resets when battery
time < five minutes and com- recharge vent has ≥ five minutes of
pressor battery operational main battery operational time
time < 5 minutes. and compressor remaining or when
battery. AC power is restored.
3VTE MAND Low Last 2 mand breaths ≤ set limit. Check for leaks, Exhaled mandatory
(patient data changes in tidal volume.≤ set
Medium Last 4 mand breaths ≤ set limit. patient's R & C. limit. Alarm updated
alarm)
whenever exhaled
High Last 10 or more mand breaths
mandatory tidal
≤ set limit.
volume is recalculat-
ed. Possible depen-
dent alarms: 3VE TOT,
1fTOT.
3VTE SPONT Low Last 4 spont breaths ≤ set. limit Check patient & Exhaled spontaneous
(patient data settings. tidal volume ≤ set
Medium Last 7 spont breaths ≤ set limit. limit. Alarm updated
alarm)
whenever exhaled
High Last 10 or more spont breaths ≤
spontaneous tidal
set limit.
volume is recalculat-
ed. Possible depen-
dent alarms: 3VE TOT
1fTOT.
1VTI (patient Low Last spont breath ≥ set limit. In TC, VS, or Delivered inspiratory
data alarm) PAV+: volume ≥ inspiratory
Medium Last 3 spont breaths ≥ set limit. Check patient limit. Ventilator transi-
and settings. tions to exhalation.
High Last 4 or more spont breaths ≥
Possible dependent
set limit.
alarms: 3VTE
SPONT,3VE TOT, 1fTOT
Corrective action:
Check for leaks. Check
for the correct tube
type.
Check the VTI or VTI
setting. In PAV+,
check for patient agi-
tation, which can
cause miscalculation
of RPAV and CPAV. Con-
sider reducing %
Supp setting. Check
2VTI.
3VE TOT (patient Low VE TOT ≤ set limit for ≤ 30 s. Check patient & Total minute volume
data alarm) settings. ≤ set limit. Alarm
Medium VE TOT ≤ set limit for > 30 s. updated whenever
exhaled minute
High VE TOT ≤ set limit for > 120 s. volume is recalculat-
ed. Possible depen-
dent alarms3VTE
MAND, 3VTE SPONT,
1fTOT
VOLUME NOT Low Last 2 spont (or mand) breaths Check patient Insp target pressure >
DELIVERED (not pressure > max allowable level. and setting for (PPEAK -
adjustable) 1PPEAK PEEP - 3 cmH2O),
(patient data Medium Last 10 or more spont (or
when spontaneous
alarm) mand) breaths, pressure > max
type is VS or manda-
allowable level.
tory type is VC+ Venti-
lator cannot deliver
target volume. Possi-
ble dependent
alarms: For VC+
breaths: 3VTE MAND,
3VE TOT, 1fTOT. For VS
breaths: 3VTE SPONT,
3VE TOT, 1fTOT
Corrective action:
Check patient and
settings.
NO AIR SUPPLY Low Compressor inoperative. Venti- Check air source. Ventilator delivers
lation continues as set. Only O2 100% O2. Air supply
available. pressure ≤
17 psig. Resets if air
NO AIR SUPPLY High Compressor inoperative. Venti- Check patient supply pressure ≥ 35
lation continues as set, except y air source. psig is connected.
O2% = 100.
INADVERTENT High Ventilator switched OFF with Return power User must acknowl-
POWER OFF patient connected to breathing switch to ON edge turning the
circuit. position and dis- power OFF by touch-
connect patient ing Power Off on the
before turning GUI.
power off.
PROX INOPERA- Low Data from proximal flow sensor Check proximal Data for real time
TIVE is not being used. flow sensor con- waveforms and moni-
nections and tored volumes are
tubes for occlu- obtained from inter-
sions or leaks. nal sensors.
AC POWER LOSS The ventilator/and or compressor is Monitor the battery charge level to
running on battery power. ensure there is enough power
remaining to operate the ventila-
tor/compressor.
APNEA (patient data alarm) The time between patient breaths Check patient and settings.
exceeds the set apnea interval.
CIRCUIT DISCONNECT The patient circuit has become discon- Re-connect the patient circuit, or
nected or there is a large leak in the eliminate the leak.
patient circuit.
Compliance limitedVT Compliance volume required to com- Check patient and circuit type.
(patient data alarm) pensate delivery of a VC, VC+, or VS Inspired volume may be less than
breath exceeds the maximum allowed set.
for three consecutive breaths.
COMPRESSOR INOPERATIVE Air pressure not detected in the com- Service or replace compressor.
pressor’s accumulator. Status display
indicates the compressor is inoperative
1PPEAK (patient data alarm) The measured airway pressure is ≥ set • Check the patient.
limit. Reduced tidal volume likely.
• Check the patient circuit.
3PPEAK (patient data alarm) The peak inspiratory pressure in the Check for leaks in the patient
patient circuit ≤ alarm setting.This alarm circuit and VBS.
is only available when NIV is the selected
Vent Type or when VC+ is the selected
Mandatory type during INVASIVE venti-
lation.*
* Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall below PEEP
+ 3 cmH2O, attempting to set the 4PPEAK alarm setting at or below this level will turn the alarm off.
1O2% (patient data alarm) The O2% measured during any phase of Check the patient, the air and
a breath cycle is 7% (12% during the first oxygen supplies, the oxygen ana-
hour of operation) or more above the lyzer, and the ventilator.
O2% parameter for at least 30 sec-
onds.The percentage window increases
by 5% for four minutes after increasing
the set O2% value.
3O2% (patient data alarm) The O2% measured during any phase of • Check the patient, the air
a breath cycle is 7% (12% during the first and oxygen supplies, the
hour of operation) or more below the oxygen analyzer, and the
O2% parameter for at least 30 seconds. ventilator.
The percentage window increases by
• Calibrate oxygen sensor. Ref-
5% for four minutes after increasing the
erence Oxygen Sensor Calibra-
set O2% value.
tion, p. 4-36 for details
regarding calibrating the
oxygen sensor.
1VTE (patient data alarm) Exhaled tidal volume ≥ alarm setting for • Check patient settings.
the last two breaths.
• Check for changes in
patient’s resistance and
compliance.
1VE TOT (patient data alarm) Minute volume ≥ alarm setting. Check patient settings.
1fTOT (patient data alarm) The breath rate from all breaths is ≥ Check the patient and the ventila-
alarm setting. tor settings.
1PVENT (patient data alarm) The inspiratory pressure transducer has • Check the patient, the
measured a pressure > 110 cmH2O in patient circuit (including fil-
VC, TC, or PAV+. The ventilator transi- ters), and the endotracheal
tions to exhalation. A reduced tidal tube. Ensure the ET tube ID is
volume is likely. the correct size. Check the
ventilator flow and/or
volume settings.
• Re-run SST.
INOPERATIVE BATTERY The battery charge is inadequate after 6 Recharge the battery by plugging
hours of attempted charge time or the the ventilator into AC power or
battery system is non-functional. replace the battery or install an
extended battery.
INSPIRATION TOO LONG The PBW-based inspiratory time for the • Check the patient.
(patient data alarm) last two spontaneous breath exceeds
the ventilator-set limit. Active only when • Check the patient circuit for
Vent Type is INVASIVE. leaks.
LOSS OF POWER The ventilator power switch is ON, but • Check the integrity of the AC
there is insufficient power from the power and battery connec-
mains AC and the battery. There may tions.
not be a visual indicator for this alarm,
but an independent audio alarm (imme- • Obtain alternative ventila-
diate priority) sounds for at least 120 sec- tion if necessary.
onds.
• Install an extended battery.
LOW BATTERY Medium priority alarm indicating < ten Recharge the battery, by plugging
(10) minutes of battery power remaining the ventilator into AC power or
to operate the ventilator or compressor. replace the battery, or install an
High priority alarm indicating < five (5) extended battery.
minutes of battery power remain to
operate the ventilator or compressor.
3VTE MAND (patient data The patient’s exhaled mandatory tidal • Check the patient.
alarm) volume is ≤ alarm setting for the last two
mandatory breaths. • Check for leaks in the patient
circuit.
3VTE SPONT (patient data The patient’s exhaled spontaneous tidal • Check the patient.
alarm) volume is ≤ alarm setting for the last two
spontaneous breaths. • Check the ventilator set-
tings.
3VE TOT (patient data alarm) The minute volume for all breaths is ≤ • Check the patient.
alarm setting.
• Check the ventilator set-
tings.
NO AIR SUPPLY The air supply pressure is less than the • Check the patient.
minimum pressure required for correct
ventilator operation. The ventilator • Check the air and oxygen
delivers 100% O2 if available. If an sources.
oxygen supply is not available, the safety
• Obtain alternative ventila-
valve opens. The ventilator displays the
tion if necessary.
elapsed time without ventilator support.
This alarm cannot be set or disabled.
NO O2 SUPPLY The oxygen supply pressure is less than • Check the patient.
the minimum pressure required for
correct ventilator operation. The ventila- • Check the air and oxygen
tor delivers 100% air if available. If an air sources.
supply is not available, the safety valve
opens. The ventilator displays the • Obtain alternative ventila-
elapsed time without ventilatory sup- tion if necessary.
port. This alarm cannot be set or dis-
abled.
In PAV+:
PROCEDURE ERROR The patient is attached before ventilator • Provide alternate ventilation
startup is complete. Safety ventilation is if necessary.
active.
• Complete ventilator startup
procedure.
SEVERE OCCLUSION The patient circuit is severely occluded. • Check the patient.
The ventilator enters occlusion status
cycling. The elapsed time without venti- • Obtain alternative ventila-
latory support appears. tion if necessary.
1VTI (patient data alarm) Delivered inspiratory volume ≥ high Ventilator transitions to exhalation.
inspiratory volume limit. • Check for leaks and tube
type/ID setting.
VOLUME NOT DELIVERED Insp target pressure > Check patient and 1PPEAK setting.
(patient data alarm (PPEAK - PEEP - 3 cmH2O), when sponta-
neous type is VS or when mandatory
type is VC+.
PAV STARTUP TOO LONG Unable to assess resistance and/or com- Check for leaks, shallow breathing,
(occurs only if PAV+ option is pliance during PAV+ startup. and settings for
in use) 1VTI and 1PPEAK
PAV R & C NOT ASSESSED Unable to assess resistance and/or com- Check for leaks, shallow breathing,
(occurs only if PAV+ option is pliance during PAV+ steady-state. and settings for tube ID, 1VTI and
in use) 1PPEAK
PROX INOPERATIVE A malfunction occurred with the Proxi- Replace the Proximal Flow Sensor
mal Flow Sensor or the pneumatic lines or purge its pneumatic lines. Does
are occluded. not affect data from the ventilator’s
delivery or exhalation flow sensors.
The AC POWER LOSS alarm indicates the ventilator power switch is on and the ven-
tilator is being powered by the battery and an alternate power source may soon be
required to sustain normal ventilator operation. The ventilator annunciates a
medium-priority LOW BATTERY alarm when the ventilator has less than ten minutes
of battery power remaining. The ventilator annunciates a high-priority LOW
BATTERY alarm when less than five minutes of battery power are estimated avail-
able.
The compressor is a DC device, in which AC power is converted to DC power, and it
has its own primary and extended batteries (if the extended battery was purchased).
If AC power is lost, there is no conversion to DC power for the compressor as in
normal operation, but the compressor supplies air, providing the charge level of its
batteries is sufficient.
The APNEA alarm indicates neither the ventilator nor the patient has triggered a
breath for the operator-selected apnea interval (TA). TA is measured from the start of
an inspiration to the start of the next inspiration and is based on the ventilator’s
inspiratory detection criteria. TA can only be set via the apnea ventilation settings.
The APNEA alarm autoresets after the patient initiates two successive breaths, and is
intended to establish the patient's inspiratory drive is reliable enough to resume
normal ventilation. To ensure the breaths are patient-initiated (and not due to auto-
triggering), exhaled volumes must be at least half the VT (this avoids returning to
normal ventilation if there is a disconnect).
The CIRCUIT DISCONNECT alarm indicates the patient circuit is disconnected at the
ventilator or the patient side of the patient wye, or a large leak is present. The
methods by which circuit disconnects are detected vary depending on breath type.
Time, pressure, flow, delivered volume, exhaled volume, and the DSENS setting may
be used in the circuit disconnect detection algorithms. Reference Disconnect, p. 10-
45 for a complete discussion of the CIRCUIT DISCONNECT detection methods.
The CIRCUIT DISCONNECT alarm sensitivity is adjusted via the DSENS setting. During
a CIRCUIT DISCONNECT condition, the ventilator enters an idle state and delivers a
base flow of oxygen to detect a reconnection.
When the ventilator determines the patient circuit is reconnected, the CIRCUIT DIS-
CONNECT alarm autoresets and normal ventilation resumes without having to man-
ually reset the alarm (for example, following suctioning).
A disconnected patient circuit interrupts gas delivery and patient monitoring. Noti-
fication of a patient circuit disconnect is crucial, particularly when the patient cannot
breathe spontaneously. The ventilator does not enter apnea ventilation when a dis-
connect is detected to avoid changing modes during a routine suctioning proce-
dure.
Note:
When utilizing a closed-suction catheter system, the suctioning procedure can be executed
using existing mode, breath type and settings. To reduce potential for hypoxemia during
the procedure, elevate the oxygen concentration using the Elevate O2 control. Reference To
adjust the amount of elevated O2 delivered for two minutes, p. 3-37.
This alarm alerts the operator that there is insufficient battery power and no AC
power to support ventilator or compressor operation. The alarm annunciates as long
as the ventilator’s power switch is in the ON position, and lasts for at least 120 sec-
onds.
A DEVICE ALERT alarm indicates a background test or power on self test (POST) has
failed. Depending on which test failed, the ventilator either declares an alarm and
continues to ventilate according to current settings, or ventilates with modified set-
tings, or enters the ventilator inoperative state. The DEVICE ALERT alarm relies on the
ventilator’s self-testing and notifies the clinician of an abnormal condition requiring
service. Reference Background Diagnostic System, p. 10-72.
The 1PPEAK alarm indicates the currently measured airway pressure is equal to or
greater than the set limit. The 2PPEAK limit is active during all breath types and phases
to provide redundant patient protection (for example, to detect air flow restrictions
downstream of the pressure-sensing device). The 1PPEAK limit is active in all normal
ventilation modes.The 2PPEAK alarm new patient default values are separately con-
figurable for neonatal, pediatric, and adult patients. The2PPEAK limit is not active
during a SEVERE OCCLUSION alarm.
The1PPEAK alarm truncates inspiration and transitions the ventilator into the exhala-
tion phase and the limit cannot be set less than
• PEEP + 7 cmH2O or
• PEEP + PI + 2 cmH2O, or
The 1O2% alarm indicates the measured O2% is at or above the error percentage
above the O2% setting for at least 30 seconds to eliminate transient O2% delivery
variation nuisance alarms. The 1O2% alarm detects malfunctions in ventilator gas
delivery or oxygen monitor. The ventilator declares a 1O2% alarm after 30 seconds.
Although the ventilator automatically sets the 1O2% alarm limits, the oxygen sensor
can be disabled. (The error percentage is 12% above setting for the first hour of ven-
tilator operation, 7% above the setting after the first hour of operation, and an addi-
tional 5% above the setting for the first four minutes following a decrease in the
setting.)
The ventilator automatically adjusts the 1O2% alarm limit when O2% changes due to
100% O2, apnea ventilation, occlusion, circuit disconnect, or a NO AIR/O2 SUPPLY
alarm. The ventilator checks the 1O2% alarm limit against the measured oxygen per-
centage at one-second intervals.
The 1VE TOT alarm indicates the measured exhaled total minute volume for sponta-
neous and mandatory breaths is equal to or greater than the alarm setting. The 1VE
TOT alarm is effective immediately upon changing the setting, to ensure prompt
notification of prolonged high tidal volumes.
The1VE TOT alarm can be used to detect a change in a patient's breathing pattern, or
a change in compliance or resistance. The 1VE TOT alarm can also detect too-large
tidal volumes, which could lead to hyperventilation and hypocarbia.
The 1VTE alarm indicates the measured exhaled tidal volume for spontaneous and
mandatory breaths is equal to or greater than the set 1VTE alarm. The 1VTE alarm is
updated whenever a new measured value is available.
The 1VTE alarm can detect increased exhaled tidal volume (due to greater compli-
ance and lower resistance) and prevent hyperventilation during pressure control
ventilation or pressure support. Turn the 1VTE alarm OFF to avoid nuisance alarms.
(Hyperventilation due to increased compliance is not a concern during volume-
based ventilation, because the tidal volume is fixed by the clinician's choice and the
ventilator’s compliance-compensation algorithm.)
The high inspired tidal volume alarm indicates the patient’s inspired volume
exceeds the set limit. When this condition occurs, the breath terminates and the
alarm sounds. The ventilator displays monitored inspired tidal volume values in the
patient data area on the GUI screen. When Vent Type is NIV, there is no high inspired
tidal volume alarm or setting available, but the monitored inspired tidal volume (VTI)
may appear in the patient data area on the GUI screen.
The 1fTOT alarm indicates the measured breath rate is greater than or equal to the
2fTOT alarm setting. The 1fTOT alarm is updated whenever a new total measured
respiratory rate is available.
The 1fTOT alarm can detect tachypnea, which could indicate the tidal volume is too
low or the patient's work of breathing has increased. The ventilator phases in
changes to the 2fTOT limit immediately to ensure prompt notification of a high respi-
ratory rate condition.
The INSPIRATION TOO LONG alarm, active only when Vent Type is INVASIVE, indi-
cates the inspiratory time of a spontaneous breath exceeds the following time limit:
(1.99 + 0.02 x PBW) seconds (adult and pediatric circuits)
(1.00 + 0.10 x PBW) seconds (neonatal circuits)
where PBW is the current setting for predicted body weight in kg.
When the ventilator declares an INSPIRATION TOO LONG alarm, the ventilator termi-
nates inspiration and transitions to exhalation. The INSPIRATION TOO LONG alarm
applies only to spontaneous breaths and cannot be set or disabled.
Because leaks (in the patient circuit, around the endotracheal tube cuff, or through
chest tubes) and patient-ventilator mismatch can affect accurate exhalation detec-
tion, the INSPIRATION TOO LONG alarm can act as a backup method of safely termi-
nating inspiration. If the INSPIRATION TOO LONG alarm occurs frequently, check for
leaks and ensure ESENS and rise time % are properly set.
WARNING:
Because the VC+ pressure control algorithm does not allow the target inspiratory
pressure to fall below PEEP + 3 cmH2O, attempting to set the 4PPEAK alarm limit at or
below this level will turn the alarm off.
The 3PPEAK alarm indicates the measured maximum airway pressure during the
current breath is less than or equal to the set alarm level during a non-invasive inspi-
ration or during a VC+ inspiration.
The 3PPEAK alarm is active for mandatory and spontaneous breaths, and is present
only when Vent Type is NIV or Mandatory Type is VC+. During VC+, the 3PPEAK alarm
can be turned OFF. The 3PPEAK alarm can always be turned OFF during NIV. The
4PPEAK alarm limit cannot be set to a value greater than or equal to the 2PPEAK alarm
limit.
In VC+, whenever PEEP is changed, 3PPEAK is set automatically to its New Patient
value, PEEP + 4 cmH2O when PEEP ≥ 16 cmH2O, or PEEP + 3.5 cmH2O when PEEP <
16 cmH2O.
There are no alarms dependent upon 3PPEAK, and the 3PPEAK alarm does not depend
on other alarms.
The 3O2% alarm indicates the measured O2% during any phase of a breath is at or
below the error percentage below the O2% setting, or less than or equal to 18%, for
at least 30 seconds. Although the ventilator automatically sets the 3O2% alarm,
replace (if necessary) or disable the oxygen sensor to avoid nuisance alarms. The
error percentage is 12% below setting for the first hour of ventilator operation fol-
lowing a reset, 7% below setting after the first hour of operation, and an additional
5% below setting for the first four minutes following an increase in the setting.
The ventilator automatically adjusts the 3O2% alarm limit when O2% changes due to
apnea ventilation, circuit disconnect, or a NO O2/AIR SUPPLY alarm. The 3O2% alarm
is disabled during a safety valve open (SVO) condition. The ventilator checks the
3O2% alarm against the measured oxygen percentage at one-second intervals.
The 3O2% alarm can detect malfunctions in ventilator gas delivery or the oxygen
monitor, and can ensure the patient is adequately oxygenated. The ventilator
declares a 3O2% alarm after 30 seconds to eliminate nuisance alarms from transient
O2% delivery variations. The O2% measured by the oxygen sensor is shown in the
patient data area. Reference Vital Patient Data, p. 3-39 to include O2% if it is not dis-
played.
The alarm indicates the measured exhaled mandatory tidal volume is less than or
equal to the 3VTE MAND alarm setting. The 3VTE MAND alarm updates when a new mea-
sured value of exhaled mandatory tidal volume is available.
The 3VTE MAND alarm can detect an obstruction, a leak during volume ventilation, or
a change in compliance or resistance during pressure-based ventilation (that is,
when the same pressure is achieved but tidal volume decreases). There are separate
alarms for mandatory and spontaneous exhaled tidal volumes for use during SIMV,
SPONT, and BiLevel. The ventilator phases in a change to the 3VTE MAND alarm imme-
diately to ensure prompt notification of a low exhaled tidal volume condition.
The 3VTE SPONT alarm indicates the measured exhaled spontaneous tidal volume is
less than or equal to the 3VTE SPONT alarm setting. The alarm updates when a new
measured value of exhaled spontaneous tidal volume is available.
The 3VTE SPONT alarm can detect a leak in the patient circuit or a change in the
patient’s respiratory drive during a single breath. The 3VTE SPONT alarm is based on
the current breath rather than on an average to detect changes as quickly as possi-
ble. There are separate alarms for mandatory and spontaneous exhaled tidal
volumes for use during SIMV and BiLevel if this software option is installed. The ven-
tilator phases in a change to the 4VTE SPONT alarm limit immediately to ensure
prompt notification of a low exhaled tidal volume condition.
The 3VE TOT alarm indicates the measured minute volume (for mandatory and spon-
taneous breaths) is less than or equal to the 3VE TOT alarm setting. The 3VE TOT alarm
updates with each new calculation for exhaled minute volume.
The 3VE TOT alarm can detect a leak or obstruction in the patient circuit, a change in
compliance or resistance, or a change in the patient's breathing pattern. The 3VE TOT
alarm can also detect too-small tidal volumes, which could lead to hypoventilation
and hypoxia (oxygen desaturation).
The ventilator phases in changes to the 3VE TOT alarm limit immediately to ensure
prompt notification of prolonged low tidal volumes.
A severe occlusion alarm occurs when gas flow in the ventilator breathing system is
severely restricted. The ventilator enters Occlusion Status Cycling (OSC) where the
ventilator periodically attempts to deliver a pressure-based breath while monitoring
inspiration and exhalation breath phases for a severe occlusion. If an occlusion is not
detected, the ventilator considers the occlusion condition reset, clears the occlusion
alarm, and continues ventilation with the settings in use before the occlusion
occurred. The ventilator indicates an occlusion was detected.
Note:
A blinking patient data value means that the displayed value is greater-than or less-than
either of its absolute limits and has been “clipped” to its limit. A data value that appears in
parentheses means it has questionable accuracy. If no value is displayed, then the ventilator
is in a state where the value cannot be measured.
The following sections contain descriptions of all patient data parameters shown in
the patient data displays.
Note:
All displayed patient volume data represent lung volumes expressed under BTPS
conditions.
Total exhaled minute volume (VE TOT) is the BTPS and compliance compensated sum
of exhaled gas volumes from both mandatory and spontaneous breaths for the pre-
vious one-minute interval. A factory default parameter.
Exhaled spontaneous minute volume (VE SPONT) is the BTPS- and compliance-com-
pensated sum of exhaled spontaneous volumes for the previous minute. A factory
default parameter.
Exhaled tidal volume (VTE) is the volume of the patient’s exhaled gas for the previous
mandatory or spontaneous breath. Displayed VTE is both compliance-and BTPS
compensated, and updates at the next inspiration. A factory default parameter.
Proximal exhaled minute volume (VE TOTY) is the BTPS- and compliance-compensat-
ed sum of exhaled spontaneous volumes for the previous minute.
Proximal exhaled tidal volume (VTEY) is the exhaled tidal volume for the previous
breath measured by the Proximal Flow Sensor (for neonatal patients, only). VTEY is
updated at the beginning of the next inspiration.
Exhaled spontaneous tidal volume (VTE SPONT) is the exhaled volume of the last
spontaneous breath, updated at the beginning of the next inspiration following a
spontaneous breath.
Exhaled mandatory tidal volume (VTE MAND) is the exhaled volume of the last man-
datory breath, updated at the beginning of the next inspiration following a manda-
tory breath. If the mode is SPONT and the ventilator has not delivered mandatory
breaths in a time period of greater than two minutes (for example via a manual inspi-
ration), the VTE MAND patient data indicator becomes hidden.The indicator reappears
when the value updates. A factory default parameter.
Proximal inspired tidal volume (VTIY) is the inspired tidal volume for a mandatory or
spontaneous breath measured by the Proximal Flow Sensor (for neonatal patients,
only). VTIY is updated at the beginning of the following expiratory phase and is dis-
played when data are available.
The ratio of inspiratory time to expiratory time for the previous breath, regardless of
breath type. Updated at the beginning of the next inspiration. When I:E ratio is ≥ 1:1,
it is displayed as XX:1. Otherwise it is displayed as 1:XX. A factory default parameter.
Note:
Due to limitations in setting the I:E ratio in PC ventilation, the monitored data display may
not exactly match the I:E ratio setting.
Mean circuit pressure (PMEAN) is the average circuit pressure for a complete breath
period, including both inspiratory and expiratory phases whether mandatory or
spontaneous. The displayed value can be either positive or negative. A factory
default parameter.
Peak circuit pressure (PPEAK) is the maximum circuit pressure at the patient wye
during the previous breath, including both inspiratory and expiratory phases. A
factory default parameter.
End inspiratory pressure (PI END) is the pressure at the end of the inspiratory phase of
the current breath. A factory default parameter.
End expiratory pressure (PEEP) is the pressure at the end of the expiratory phase of
the previous breath, updated at the beginning of the next inspiration. During an
expiratory pause, the displayed value includes any active lung PEEP. A factory default
parameter.
Intrinsic PEEP (PEEPI) is an estimate of the pressure above the PEEP level at the end
of an exhalation. PEEPI is determined during an Expiratory Pause maneuver.
PAV-based intrinsic PEEP (PEEPI PAV) is an estimate of intrinsic PEEP, updated at the
end of a spontaneous PAV+ breath.
Total PEEP (PEEPTOT) is the estimated pressure at the circuit wye during the Expira-
tory Pause maneuver.
Plateau pressure (PPL) is the pressure measured and displayed during an Inspiratory
Pause maneuver.
Total respiratory rate (fTOT) is the total number of mandatory and spontaneous
breaths per minute delivered to the patient. A factory default parameter.
For a PAV+ breath, PAV-based lung compliance (CPAV) is the change in pulmonary
volume for an applied change in patient airway pressure, measured under zero-flow
conditions and updated upon successful completion of each calculation. CPAV is dis-
played on the waveform screen.
For a PAV+ breath, PAV-based patient resistance (RPAV) is the change in pulmonary
pressure for an applied change in patient lung flow and updated upon successful
completion of each calculation. RPAV is displayed on the waveform screen.
For a PAV+ breath, PAV-based lung elastance (EPAV) is the inverse of CPAV and is
updated upon successful completion of each calculation.
Spontaneous inspiratory time (TI SPONT) is the duration of the inspiratory phase of a
spontaneous breath and updated at the end of each spontaneous breath. TI SPONT is
only calculated when the breathing mode allows spontaneous breaths and the
breaths are patient-initiated. A factory default parameter.
For a PAV+ breath, PAV-based total airway resistance (RTOT) is the change in pulmo-
nary pressure for an applied change in total airway flow and updated upon the suc-
cessful completion of each calculation. If the RPAV value appears in parentheses as
described at the beginning of this section, the RTOT value also appears in parenthe-
ses.
V pt
C STAT = ------------------------------- – C ckt
P ckt – PEEP
C STAT Static compliance P ckt The pressure in the patient circuit mea-
sured at the end of the 100 ms interval
defining the pause-mechanics plateau
Total expiratory volume (patient and PEEP The pressure in the patient circuit mea-
V pt breathing circuit) sured at the end of expiration
RSTAT is calculated using this equation after CSTAT is computed and assuming a VC
breath type with a SQUARE waveform:
C ckt
1 + -------------- - P PEAK – P PL
C STAT
R STAT = --------------------------------------------------------------------
-
V· pt
C ckt Compliance of the breathing circuit during P PL Mean pressure in the patient circuit over
the pause maneuver (derived from SST) the 100 ms interval defining the pause-
mechanics plateau
Flow into the patient during the last 100 ms Peak circuit pressure
P PEAK
V· pt
of the waveform
During the pause, the most recently selected graphics are displayed and frozen, to
determine when inspiratory pressure stabilizes. CSTAT and RSTAT are displayed at the
start of the next inspiration following the inspiratory pause and take this format:
CSTATxxx
or
RSTAT yyy
Special formatting is applied if the software determines variables in the equations or
the resulting CSTAT or RSTAT values are out of bounds:
• Parentheses ( ) signify questionable CSTAT or RSTAT values, derived from questionable
variables.
• RSTAT ------ means resistance could not be computed, because the breath was not of a
mandatory, VC type with square flow waveform.
Dynamic resistance (RDYN) is a dynamic estimate of static resistance for each manda-
tory breath delivered.
6.6.32 C20/C
C20/C is the ratio of compliance of the last 20% of inspiration to the compliance of
the entire inspiration.
End Expiratory Flow (EEF) is a measurement of the end expiratory flow for an appli-
cable breath.
Displayed O2% is the percentage of oxygen in the gas delivered to the patient, mea-
sured at the ventilator’s outlet, upstream of the inspiratory filter. It is intended to
provide a check against the set O2% for alarm determination, and not as a measure-
ment of oxygen delivered to the patient. O2% data can be displayed as long as the
O2 monitor is enabled. If the monitor is disabled, dashes (--) are displayed. If a device
alert occurs related to the O2 monitor, a blinking 0 is displayed.
7.1 Overview
This chapter contains information on maintenance of the Puritan Bennett™ 980
Series Ventilator. It includes
• How to perform routine preventive maintenance procedures, including frequency
WARNING:
To ensure proper ventilator operation, perform preventive maintenance intervals as
specified in the following tables. Reference Operator Preventive Maintenance
Frequency, p. 7-2 and Reference Service Preventive Maintenance Frequency, p. 7-21.
7-1
Preventive Maintenance
Patient circuit: inspiratory and Several times a day or as required • Check both limbs for water
expiratory limbs by the institution's policy. accumulation.
Disposable inspiratory Filter After 15 days of continuous use Discard according to the institu-
(discard) tion’s protocol.
Disposable exhalation Filter After 15 days of continuous use Discard according to the institu-
(discard) tion’s protocol.
Every 100 disinfection cycles. A dis- Replace. Discard used flow sensor
infection event is defined as one according to the institution's pro-
disinfection event as described in tocol. Run exhalation flow sensor
Exhalation Flow Sensor Assembly calibration and SST.
(EVQ) Disinfection (7.5.1) on page 7-
9.
Compressor inlet air filter Every 250 hours Wash in mild soapy water and rinse
thoroughly. Let air dry.
Neonatal door/adapter When visibly soiled or as desired Disinfect per Table 7-3.
Caution:
Use specified cleaning, disinfection, and sterilization agents and procedures for the
appropriate part as instructed.follow cleaning procedures outlined below.
2. Wipe the GUI, BDU, and compressor base, removing any dirt or foreign substances.
4. If necessary, vacuum any cooling vents on the GUI and BDU with an electrostatic dis-
charge (ESD)-safe vacuum to remove any dust.
Ventilator exterior (including touch Wipe clean with a cloth dampened Do not allow liquid or sprays to
screen and flex arm with one of the cleaning agents penetrate the ventilator openings
listed below or equivalent. Use a or cable connections.
damp cloth and water to rinse off Do not attempt to sterilize the ven-
chemical residue as necessary. tilator by exposure to ethylene
oxide (ETO) gas.
Mild dish washing detergent solu-
tion Do not use pressurized air to clean
or dry the ventilator, including the
Isopropyl alcohol (70% solution) GUI cooling vents.
WARNING:
To avoid microbial contamination and potential performance problems, do not
clean, disinfect, or reuse single-patient use (SPU) or disposable components. Discard
per local or institutional regulations.
Risks associated with reuse of single-patient use items include but are not limited to
microbial cross-contamination, leaks, loss of part integrity, and increased pressure
drop. When cleaning reusable components, do not use hard brushes or implements
that could damage surfaces.
• Metricide 28 (2.5%)
• Metricide 28 (2.5%)
Reusable patient circuit tubing Disinfect per manufacturer’s • Inspect for nicks and cuts,
instructions-for-use. and replace if damaged.
Disposable patient circuit tubing Discard Discard per the institution’s proto-
col.
Breathing circuit in-line water traps Disinfect per manufacturer’s • Inspect water traps for cracks
instructions-for-use. and replace if damaged.
Disposable drain bag and tubing Discard when filled to capacity or N/A
(single unit) when changing patient circuit.
Inlet air filter bowl Wash the bowl with mild soap • Avoid exposing the inlet air
solution, if needed. filter bowl to aromatic sol-
vents, especially ketones.
Battery Wipe with a damp cloth using one Do not immerse the battery or get
of the cleaning agents listed. Refer- the contacts wet.
ence Surface Cleaning Agents, p. 7-
5.
2. Thoroughly rinse parts in clean, warm water (tap water is acceptable) and wipe dry.
3. Clean or disinfect parts per the procedures listed for each component. Reference Com-
ponent Cleaning Agents and Disinfection Procedures, p. 7-6. For a list of cleaning and dis-
infection agents.
4. After the components are cleaned or disinfected, inspect them for cracks or other
damage.
Note:
Steps 1 through 3 above do not apply to the EVQ. Reference Exhalation Flow Sensor
Assembly (EVQ) Disinfection, p. 7-9 for disinfection instructions.
Note:
EVQ disinfection is not required on a routine basis but it should be disinfected as desired, or
if SST flow sensor cross check fails. Reference Component Cleaning Agents and Disinfection
Procedures, p. 7-6 for a list of appropriate disinfectants.
Note:
Follow the institution’s infection control protocol for handling, storage, and disposal of
potentially bio-contaminated waste.
Caution:
To avoid damaging the hot film wire, do not insert fingers or objects into the center
port when disinfecting the EVQ.
The EVQ contains the exhalation flow sensor electronics, exhalation valve dia-
phragm, exhalation filter seal, and pressure sensor filter. The exhalation flow sensor
electronics consist of the hot film wire and the thermistor. Since it is protected by
the exhalation filter, it does not require or need replacement or disinfection on a
regular basis. It is, however, removable and can be disinfected as desired, or if SST
flow sensor cross check fails. Expected service life is 100 disinfection cycles.
Caution:
To avoid damage to the exhalation flow sensor element
• Do not touch the hot film wire or thermistor in the center port
• Do not vigorously agitate fluid through the center port while immersed.
• Do not forcefully blow compressed air or any fluid into the center cavity.
WARNING:
Damaging the flow sensor’s hot film wire or thermistor in the center port can cause
the ventilator’s spirometry system to malfunction.
Removal
WARNING:
Prior to cleaning and disinfection, remove and dispose of the disposable
components of the exhalation flow sensor assembly.
2. With thumb inserted into the plastic exhalation port and four (4) fingers under the EVQ,
pull it down until it snaps out. To avoid damaging the flow sensor element, do not
insert fingers into the center port.
2. Dispose of the removed items according to the institution’s protocol. Follow local gov-
erning ordinances regarding disposal of potentially bio-contaminated waste.
Disinfection
WARNING:
Do not steam autoclave the EVQ or sterilize with ethylene oxide gas. Either process
could cause the ventilator’s spirometry system to malfunction when reinstalled in
the ventilator.
WARNING:
Use only the disinfectants described. Reference Component Cleaning Agents and
Disinfection Procedures, p. 7-6. Using disinfectants not recommended by Covidien
may damage the plastic enclosure or electronic sensor components, resulting in
malfunction of the ventilator’s spirometry system.
WARNING:
Follow disinfectant manufacturer’s recommendations for personal protection (such
as gloves, fume hood, etc.) to avoid potential injury.
1. Pre-soak the EVQ in the enzymatic solution. Reference Operator Preventive Maintenance
Frequency, p. 7-2. The purpose for this pre-soak is to break down any bio-film that may
be present. Follow manufacturer’s instructions regarding duration of soak process.
Caution:
Do not use any type of brush to scrub the EVQ, as damage to the flow sensing
element could occur.
4. Immerse in the disinfectant solution, oriented as shown, and rotate to remove trapped
air bubbles in its cavities. Keep immersed for the minimum time period by the manu-
facturer or as noted in the institution’s protocol.
5. At the end of the disinfecting immersion period, remove and drain all disinfectant.
Ensure all cavities are completely drained.
Rinsing
WARNING:
Rinse according to manufacturer’s instructions. Avoid skin contact with disinfecting
agents to prevent possible injury.
1. Rinse the EVQ using clean, de-ionized water in the same manner used for the disinfec-
tion step.
2. Drain and repeat rinsing three times with clean, de-ionized water.
3. After rinsing in de-ionized water, immerse in a clean isopropyl alcohol bath for approx-
imately 15 seconds. Slowly agitate and rotate to empty air pockets.
Drying
1. Dry in a low temperature warm air cabinet designed for this purpose. Covidien recom-
mends a convective drying oven for this process, with temperature not exceeding 60°C
(140°F).
Caution:
Exercise care in placement and handling in a dryer to prevent damage to the
assembly’s flow sensor element.
Inspection
2. Inspect electrical contacts for contaminating film or material. Wipe clean with a soft
cloth if necessary.
3. Inspect the hot film wire and thermistor in the center port for damage and for contam-
ination. DO NOT ATTEMPT TO CLEAN EITHER OF THESE. If contamination exists, rinse
again with de-ionized water. If rinsing is unsuccessful or hot film wire or thermistor is
damaged, replace the EVQ.
1. After drying the EVQ, remove the pressure sensor filter from the reprocessing kit and
install its large diameter into the filter grommet with a twisting motion until flush with
the plastic valve body, as shown. The narrow end faces out.
2. Remove the exhalation filter seal from the kit and turn the assembly so its bottom is
facing up.
3. Install the seal into the recess with the flat side facing outward, away from the recess.
Reference Installing the Exhalation Filter Seal, p. 7-17.
4. Remove the diaphragm from the kit and install it. The outer seal bead rests in the outer
groove.
1. Replace the EVQ any time if cracked or damaged in use, or if a malfunction occurs
during SST or EST.
2. Replace assembly if damage is noted to the hot film wire and thermistor in the center
port.
2. Install the exhalation filter by sliding it onto the tracks in the door, and orienting the fil-
ter’s From Patient port through the hole in the door.
7.5.4 Storage
1. Pre-test the EVQ before storage by installing it into the ventilator and running SST to test
the integrity of the breathing system. Reference To run SST, p. 3-45.
2. After performing SST, remove the assembly and place it into a protective bag or similar
covered container.
2. After the components are sterilized, visually inspect them for cracks or other damage.
Autoclave sterilization
Effective sterilization occurs by steam autoclaving at 132°C (170°F) for 15 minutes for gravity displacement cycles.
Pre-vac sterilization of wrapped goods (132°C for 4 minutes) may also be used. Refer to pre-vac system manu-
facturer’s program parameters or follow the steam sterilizer manufacturer’s instructions.
1. Disassemble the component.
8. Run SST.
*If performing pre-vac sterilization, follow system manufacturer’s instructions for use (IFU).
Reusable exhalation and inspirato- Steam autoclave per manufactur- • Do not chemically disinfect
ry filters er’s instructions-for-use or expose to ETO gas.
Exhalation filter condensate vial Steam autoclave per manufactur- • Inspect the condensate vial
er’s instructions-for-use for cracks after processing.
Reusable drain bag tubing (short Clean and autoclave the reusable N/A
piece of tubing attached to drain tubing; clean the clamp. Reference
bag) and clamp Surface Cleaning Agents, p. 7-5 for
approved cleaning agents.
Primary and extended batteries Perform battery test (as part of EST
and perform stand-alone battery
test in Service mode).
When ventilator location changes Atmospheric pressure transducer Perform atmospheric pressure
by 1000 feet of altitude transducer calibration.
Every 3 years, or when battery test Primary battery Replace primary batteries (ventila-
fails, or when EST indicates battery tor and compressor). Actual
life has been exhausted battery life depends on the history
of use and ambient conditions.
Every 10,000 operational hours Internal inspiratory filter Replace. Do not attempt to auto-
clave or reuse.
BDU 10K hour kit, p/n 10097275 Install. Reference the Puritan Ben-
nett™ 980 Series Ventilator Service
Manual for information on tests
required after installation of 10K
PM Kit.
Compressor 10K hour kit, p/n Install. Reference the Puritan Ben-
10097258 nett™ 980 Series Ventilator Service
Manual for information on tests
required after installation of 10K
PM Kit.
7.10 Documentation
Covidien factory-trained service personnel should manually enter the service date,
time, and nature of repair/preventive maintenance performed into the log using a
keyboard on the GUI.
To manually document a service or preventive maintenance activity
1. Enter Service mode.
4. Select Add Entry, and using the buttons to the right of each line, complete the entry.
2. Recharge batteries prior to patient ventilation. If batteries are older than three (3) years,
use new batteries.
8.1 Overview
This chapter contains information regarding ventilator logs on the Puritan Bennett™
980 Series Ventilator.
WARNING:
To avoid a potential electrical shock, do not attempt to correct any electrical problem
with the ventilator while it is connected to AC power.
8-1
Troubleshooting
• Settings Log — The settings log records changes to ventilator settings for retrospec-
tive analysis of ventilator-patient management. The time and date, old and new set-
tings. and alarm resets are recorded. A maximum of 500 settings changes can be stored
in the log. The settings log is cleared when the ventilator is set up for a new patient. The
settings log is accessible in normal ventilation mode and Service mode.
• Patient Data Log — This log records every minute (up to 4320 patient data entries)
consisting of date and time of the entry, patient data name, and the patient data value
during ventilator operation. It is cleared when the ventilator is set up for a new patient.
Three tabs are contained in the patient data log:
– Vital Patient data — The log contains the same information that the clinician has
configured in the patient data banner at the top of the GUI. If the patient data
parameters in the banner are changed, these changes are reflected the next time
the patient data log is viewed.
– Additional Patient Data – 1 — This log corresponds to the patient data parame-
ters set on page 1 of the additional patient data banner. A total of 15 parameters are
stored here, consisting of date and time of the entry (recorded every minute),
patient data name, and the patient data value during ventilator operation.
– Additional Patient Data – 2 — This log corresponds to the patient data parame-
ters set on page 2of the additional patient data banner. A total of ten (10) parame-
ters are stored here, consisting of date and time of the entry (recorded every
minute), patient data name, and the patient data value during ventilator operation.
• Diagnostic Log — The Diagnostic Log is accessible during normal ventilation and
Service modes and contains tabs for the System Diagnostic Log (default), the System
Communication Log, and the EST/SST Diagnostic Log. The diagnostic log contains tabs
for the following:
– System Diagnostic Log — The System Diagnostic Log contains the date and time
when an event occurred, the type of event, the diagnostic code(s) associated with
each fault or error that occurred, the type of error that occurred, and any notes. Ref-
erence the Puritan Bennett™ 980 Series Ventilator Service Manual (10078090) for spe-
cific information contained in the System Diagnostic Log. The diagnostic log is not
cleared when the ventilator is set up for a new patient.
– EST/SST Diagnostic Log — The EST/SST diagnostic log displays the time, date,
test/event, system code (reference the Puritan Bennett™ 980 Series Ventilator Service
Manual), type, and notes.
• EST/SST status log — The EST and SST status log displays the time, date, test/event,
test status (passed or failed).
• General Event log — The general event log contains ventilator-related information not
found in any other logs. It includes date and time of compressor on and off, changes in
alarm volume, when the ventilator entered and exited Stand-By, GUI key presses, respi-
ratory mechanics maneuvers, O2 calibration, patient connection, elevate O2, and
warning notifications. The General event log can display up to 256 entries and is not
cleared upon new patient setup.
• Service Log — The service log is accessible during normal ventilation and Service
modes and contains the nature and type of the service, reference numbers specific to
the service event (for example, sensor and actuator ID numbers), manual and automatic
serial number input, and the time and date when the service event occurred. It is not
cleared upon new patient setup.
Ventilator logs can be saved by entering Service mode, and downloading them via
the ethernet port. Reference the Puritan Bennett™ 980 Series Ventilator Service Manual
for instructions on downloading ventilator logs.
9.1 Overview
This chapter includes accessories that can be used with the Puritan Bennett™ 980
Series Ventilator. Reference Accessories and Options, p. 9-3 for part numbers of any
items available through Covidien.
The following commonly available accessories from the listed manufacturers can be
used with the ventilator system:
• Filters — DAR/Covidien, Puritan Bennett
WARNING:
The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as
indicated, very limited exposure to trace amounts of phthalates may occur. There is
no clear clinical evidence that this degree of exposure increases clinical risk.
However, in order to minimize risk of phthalate exposure in children and nursing or
pregnant women, this product should only be used as directed.
9-1
Accessories
Reference Ventilator with Accessories, p. 9-2 and the figure above for the parts listed
in the following table.
Note:
Occasionally, part numbers change. If in doubt about a part number, contact your local
Covidien representative.
Note:
The ventilator is designed with a semi-automated short self test (SST) procedure that, in
addition to other tests, measures compliance, resistance, and leak for the ventilator
breathing circuit assembly (inspiratory filter, breathing circuit, humidifier chamber [as
applicable], exhalation filter, and exhalation flow sensor). Reference SST (Short Self Test), page
3-43. When SST is performed according to the instructions provided in SST (Short Self Test)
(3.9.1), a ventilator breathing circuit assembly that passes SST for a particular patient type
(adult, pediatric, or neonatal) will allow the ventilator to operate within specification for that
same patient type. Refer to Table 11-4. for acceptable compliance and resistance ranges.
For countries not identified, contact your local Covidien representative for the proper air and
oxygen hose part numbers.
Assy, patient circuit, adult dual heated wire, disposable, for F&P
MR850 (Medtronic / DAR) 304S14300
Adapter cable: 111/1149
Assy, patient circuit, single heated wire, adult, disposable, for F&P
MR850 (Medtronic / DAR) 304S14291
Adapter cable: 111/1146
Ventilator breathing circuit, adult, dual heated, no water traps, dis- 870-35 KIT
posable (Hudson RCI / Teleflex)2
Assy, patient circuit, with single water trap, heated insp. limb, pedi-
atric, disposable for F&P MR850–(Medtronic / DAR) 306S8987
Adapter cable: 111/1146
Not shown Exhalation valve module reprocessing kit (6/ carton) 10086048
Hardware options
Not shown Gold standard test circuit, 21 inch (for performing EST) 4-018506-00
Software options
10.1 Overview
This chapter provides specific details on breath delivery functions of the Puritan Ben-
nett™ 980 Series Ventilator. The chapter is organized as shown below.
10-1
Theory of Operations
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient's
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
The gas supplies to which the ventilator are connected must be capable of deliver-
ing 200 L/min flow with the proper supply pressure between 35 psig and 87 psig
(241.8 kPa to 599.8 kPa). These supplies may be compressed air from an external
source (wall or bottled) air or oxygen. (An optional compressor is available to be
used as an external air source.)
Air and oxygen hoses connect directly to the rear of the breath delivery unit (BDU).
The flow of each gas is metered by a Proportional Solenoid (PSOL) valve to achieve
the desired mix in the Mix Module. The flow through each PSOL is monitored by sep-
arate flow sensors to ensure the accuracy of the mix. The mixed gases then flow to
the Inspiratory Module.
The blended gas in the Inspiratory Module is metered by the Breath Delivery PSOL
and monitored by the Breath Delivery Flow Sensor to ensure that the gas is delivered
to the patient according to the settings specified by the operator. Delivered tidal
volumes are corrected to standard respiratory conditions (BTPS) to ensure consis-
tent interpretation by the clinician. The Inspiratory Module also incorporates the
Safety Valve, which opens to vent excess pressure and allows the patient to breathe
room air (if able to do so) in the event of a serous malfunction.
An optional compressor, capable of delivering flows of 140 L/min (BTPS) and minute
volumes of up to 40 L/min (BTPS), can be connected to the ventilator. Gas mixing
occurs in the accumulator, protected by a relief valve. A one-way valve allows a
maximum reverse flow into the gas supply system up to
100 mL/min under normal conditions.
Air and O2 gases travel through proportional solenoid valves (PSOLS), flow sensors,
and one-way valves, and are mixed in the mix module (according to the operator-
set O2 concentration), which also has a pressure-relief valve. From here, the gas flows
through another PSOL, to the inspiratory pneumatic system, where it passes by a
safety valve, then through a one-way valve, an internal bacteria filter, an external
bacteria filter, through the humidifier, if used, and then to the patient via the con-
nected breathing circuit.
During exhalation, the gas flows through the expiratory limb of the breathing circuit,
through a condensate vial, a bacteria filter, through the exhalation flow sensor,
through the exhalation valve, and out the exhaust port. The exhalation valve actively
controls PEEP while minimizing pressure overshoots and relieving excess pressures.
Pressure transducers in the inspiratory pneumatic system (PI) and exhalation com-
partment (PE) monitor pressures for accurately controlling breath delivery.
Additional low-pass filters precondition the signals, the signals are then used for
controls and display purposes.
Closed-loop control is used to maintain consistent pressure and flow waveforms in
the face of changing patient/system conditions. This is accomplished by using the
output as a feedback signal that is compared to the operator-set input. The differ-
ence between the two is used to drive the system toward the desired output. For
example, pressure-control modes use airway pressure as the feedback signal to
control gas flow from the ventilator. Reference the figure below. This diagram shows
a schematic drawing of a general feedback control system. The input is a reference
value (e.g., operator preset inspiratory pressure) that is compared to the actual
output value (e.g., instantaneous value of airway pressure). The difference between
those two values is the error signal. The error signal is passed to the controller (e.g.,
the software control algorithm). The controller converts the error signal into a signal
that can drive the actuator (e.g., the hardware drivers and valves) to cause a change
in the manipulated variable (e.g., inspiratory flow).
Disturbances
Feedback Signal
Note:
In the diagram above, the “plant” is the patient and the connected breathing circuit.
10.4 Inspiration
-
— Detection and initiation
When ventilator inspiration occurs, it is called triggering. Breaths are delivered to the
patient based on ventilator settings the practitioner has entered and are determined
by pressure, flow, or time measurements, or operator action. The ventilator uses the
following methods to trigger an inspiration:
• Pressure triggering (PTRIG)
• Time-triggered
• Operator-initiated
If the ventilator detects a drop in pressure at the circuit wye or when there is a
decrease in base flow measured at the exhalation valve, the patient is said to trigger
the breath. Mandatory breaths triggered by the patient are referred to as PIM or
patient-initiated mandatory breaths.
All spontaneous breaths are patient-initiated, and are also triggered by a decrease in
circuit pressure or measured base flow indicating the patient is initiating an inspira-
tion.
Another term, autotriggering, is used to describe a condition where the ventilator trig-
gers a breath in the absence of the patient’s breathing effort. Autotriggering can be
caused by inappropriate ventilator sensitivity settings, water in the patient circuit, or gas
leaks in the patient circuit.
If pressure triggering (PTRIG) is selected, the ventilator transitions into inspiration when
the pressure at the patient circuit wye drops below positive end expiratory pressure
(PEEP) minus the operator-set sensitivity level (PSENS). Reference the figure below. As the
patient begins the inspiratory effort and breathes gas from the circuit (event 5, the A-B
interval in the figure, below), pressure decreases below PEEP. When the pressure drops
below PEEP minus PSENS (event 6), the ventilator delivers a PIM breath. The pressure-
decline time interval between events A and B determines how aggressive the patient’s
inspiratory effort is. A short time interval signifies an aggressive breathing effort. The A-B
interval is also affected by PSENS. A smaller PSENS setting means a shorter A-B time interval.
(The minimum PSENS setting is limited by autotriggering, and the triggering criteria
include filtering algorithms that minimize the probability of autotriggering.)
If flow triggering (VTRIG) is selected the BDU provides a constant gas flow through
the ventilator breathing circuit (called base flow) during exhalation. The base flow is
1.5 L/m greater than the value selected for flow sensitivity (VSENS). Reference Inspira-
tion Using Flow Sensitivity, p. 10-7 where the top graphic represents expiratory flow
and the bottom graphic represents inspiratory flow.]
The ventilator’s breath delivery flow sensor measures the base flow delivered to the
circuit and the exhalation flow sensor measures the flow entering the exhalation
valve. The ventilator monitors patient flow by measuring the difference between the
inspiratory and exhaled flow measurements. If the patient is not inspiring, any differ-
ence in measured flows is due to leaks in the breathing system or flow sensor inac-
curacy. The clinician can compensate for leaks in the breathing system by increasing
VSENS to a value equal to desired VSENS + leak flow.
As the patient begins the inspiratory effort and inspires from the base flow, less
exhaled flow is measured, while the delivered flow remains constant. Reference the
figure below (event A). As the patient continues to inspire, the difference between
the delivery and exhalation flow sensor measurements increases. The ventilator ini-
tiates an inspiration when the difference between the two flow measurements is
greater than or equal to the operator-set flow sensitivity value. Reference Inspiration
Using Flow Sensitivity, (event B).
As with pressure triggering, the time delay between onset of the patient’s effort and
actual gas delivery depends on:
• how quickly the exhaled flow declines (that is, the aggressiveness of the inspiratory
effort). The more aggressive the inspiratory effort, the shorter the interval, and
• the flow sensitivity value. The smaller the value, the shorter the delay.
The ventilator measures the time interval for each breath and breath phase. If the
ventilator is in Assist/Control (A/C) mode (where the ventilator delivers breaths
based on the breath rate setting), a VIM or ventilator initiated mandatory breath is
delivered after the appropriate time interval. The duration of the breath in seconds
(Tb) is 60/f.
If the operator presses the Manual inspiration key, an OIM (operator-initiated man-
datory) breath is delivered. The ventilator will not deliver an OIM under the following
conditions:
• During an active inspiration, whether mandatory or spontaneous
Reference Manual Inspiration, p. 10-21 later in this chapter for information on the
restricted phase of exhalation.
If expiratory sensitivity (ESENS) is set to a value too low for the patient-ventilator com-
bination, a forceful expiratory effort could cause circuit pressure (PPEAK) to rise to its
limit. The ventilator monitors circuit pressure throughout the inspiratory phase, and
initiates an exhalation when the pressure equals the inspiratory pressure (PI) target
value + an incremental value. This transition to exhalation occurs during sponta-
neous pressure-based ventilation and in volume support (VS).
Note:
The allowable incremental value above the target pressure is 1.5 cmH2O once a portion of
inspiration time (Tn) has elapsed. Before Tn, the incremental value is higher to allow for
transient pressure overshoots. For the first 200 ms of inspiration, the incremental pressure is
10% of the target pressure, up or 8 cmH2O, whichever is greater. From 200 ms to Tn, the
incremental pressure decreases in a linear fashion from the initial value to 1.5 cmH2O.
3 Start breath
Note:
PAV+ uses a flow-based cycling method, also called ESENS but it is expressed in L/min rather
than in % of VMAX.
In pressure ventilation, the set inspiratory time (TI) defines the duration of the inspi-
ratory phase. In volume ventilation, TI depends on the tidal volume (VT) setting, peak
flow (VMAX), flow pattern, and plateau time (TPL). The ventilator cycles into exhala-
tion when the set TI (pressure ventilation) or computed TI (volume ventilation)
lapses.
There are four backup methods for preventing excessive duration or pressure during
inspiration
• Time limit — For adult and pediatric patients, the time limit method ends inspiration
and begins exhalation when the duration of a spontaneous inspiration is greater than
or equal to [1.99 s + 0.02 x PBW (kg)] s.
• High circuit pressure limit — During any type of inspiration, inspiration ends and
exhalation begins when the monitored airway pressure (PCIRC) is greater than or equal
to the set high circuit pressure limit.
• High ventilator pressure limit — The ventilator transitions from inspiration to exhala-
tion if the high ventilator pressure (2PVENT) limit of 110 cmH2O is reached.
• High inspired tidal volume limit — The high inspired tidal volume limit terminates
inspiration and commences exhalation during VC+, VS, tube compensated (TC), or pro-
portionally assisted (PAV+) breaths if the delivered volume is greater than or equal to
2VTI.
Note:
The ventilator does not generate subatmospheric airway pressures during exhalation.
Compliance compensation accounts for the gas volume not actually delivered to
the patient during inspiration. This gas is known as the compliance volume, VC. VC
is the gas lost to pressurizing the breathing circuit and includes the volumes of the
patient circuit, any accessories such as a humidifier and water traps, and internal
ventilator gas passages.
3 Set VMAX 6 TI
2 Actual VMAX 6 TI
C pt ckt
Factor = ---------------
-
C pt
V C = C pt ckt P wye – P
C pt ckt Compliance of the patient circuit P Pressure at the end of the current exhalation
volume is determined by the selected patient circuit type and predicted body
weight (PBW), and is summarized by this equation:
where:
Factor = linear interpolation of the values in the following table for adult, pediatric,
and neonatal circuit types. Factor is calculated as:
≤ 10 5 ≤ 10 5
15 4.6 11 3.5
60 2.75 15 2.7
Note:
Compliance compensation calculations are also in effect during exhalation to ensure
spirometry accuracy.
If the patient’s compliance decreases beyond the limits of compliance compensa-
tion, the ventilator relies on the 2PPEAK alarm setting to truncate the breath and
switch to exhalation.
Volumes and flows are BTPS compensated, that is, they are reported by the ventila-
tor at existing barometric pressure, 37°C (98.6°F), and fully saturated with water
vapor.
T b = 60 f
If, during Tb, patient effort is detected, a PIM breath is initiated and a new breath
period starts. If no patient effort is detected before Tb lapses, the next breath deliv-
ered is a VIM, and a new breath period starts.
Reference Ventilator Settings Range and Resolution, p. 11-9 for details on the following
VC+ settings:
• Expiratory time (TE)
• I:E ratio
• Rise time %
Volume Control is the control scheme that controls the flow with for the purpose of
supplying a predetermined volume (set by the practitioner) to the patient. There
are two basic flow wave forms to administer this volume: the “square” that guaran-
tees a constant flow during the inspiration time, or the “descending ramp” whose
slope and initial value are determined to provide the required volume target. Refer-
ence Ideal Waveform Using Square Flow Patternand Reference Ideal Waveform Using
Descending Ramp Flow Pattern. The inspiration time is determined indirectly by the
characteristics of the selected flow wave.
Pressure Control is the control scheme by which the pressure is controlled at the
circuit wye to reach a constant level (set by the practitioner) during inspiration, and
a PEEP level during exhalation. Reference Ideal Waveform Using Pressure Control Ven-
tilation on page 10-19. This level is maintained for a time given by the set inspiration
time, following followed by an exhalation regulated by the exhalation valve until the
PEEP level is reached. As flow is not predetermined, the supplied volume varies
depending on the patient's pulmonary response.
4 Target pressure
10.7.3 VC+
VC+ Startup
During VC+ startup, the ventilator delivers at least one breath (test breath) to deter-
mine the pressure target needed to deliver the desired (set) volume. During the time
the ventilator is delivering the test breaths, the message “VC+ startup” is displayed
in the GUI’s prompt area.
Note:
To allow for optimal function of startup and operation of VC+ in the ventilator it is important
not to block the tubing while the patient is undergoing suctioning or other treatment that
requires disconnection from the ventilator. The ventilator has a disconnect detection
algorithm that suspends ventilation while the patient is disconnected.
After VC+ Startup, the ventilator will make adjustments to the target pressure in
order to deliver the set volume (VT). In order to reach the desired volume promptly,
the maximum allowed pressure adjustments for an Adult or Pediatric patient will be
greatest during the first five breaths following Startup or a change in VT or VT SUPP
The values of the maximum pressure adjustments for each patient type are summa-
rized below.
Less than five breaths ± 10.0 cmH2O ± 6.0 cmH2O ± 3.0 cmH2O
after:
VC+ startup or Change in
VT
Reference Non-technical Alarm Summary, p. 6-19 for details on the following VC+
alarms:
• VOLUME NOT DELIVERED
• COMPLIANCE LIMITED VT
During VC+, inspiratory target pressure cannot be lower than PEEP + 3 cmH2O and
cannot exceed 2PPEAK - 3 cmH2O.
If PC or VC+ is selected as the Mandatory type, adjust rise time % for optimum flow
delivery into lungs.Patients with high impedance (low compliance, and high resis-
tance) may benefit from a lower rise time% whereas patients with low impedance
may better tolerate a more aggressive rise time setting. The rise time % setting spec-
ifies the speed at which the inspiratory pressure reaches 95% of the target pressure.
The rise time setting applies to PS (including a setting of 0 cmH2O), PC, or VC+
breaths. To match the flow demand of an actively breathing patient, observe simul-
taneous pressure-time and flow-time curves, and adjust the rise time % to maintain
a smooth rise of pressure to the target value. A rise time % setting reaching the
target value well before the end of inspiration can cause the ventilator to supply
excess flow to the patient. Whether this oversupply is clinically beneficial must be
evaluated for each patient. Generally, the optimum rise time % for gently breathing
patients is less than or equal to the default (50%), while optimum rise time % for
more aggressively breathing patients can be 50% or higher.
WARNING:
Under certain clinical circumstances (such as stiff lungs, or a small patient with a
weak inspiratory drive), a rise time % setting above 50% could cause a transient
pressure overshoot and premature transition to exhalation, or pressure oscillations
during inspiration. Carefully evaluate the patient’s condition before setting the rise
time % above the default setting of 50%.
When pressed, the Manual Inspiration key delivers one OIM breath to the patient,
using set breath delivery parameters.
The ventilator will not allow a manual inspiration during the restricted phase of
exhalation or when the ventilator is in the process of delivering a breath whether
mandatory or spontaneous). All manual inspiration attempts are logged in the
General Event log.
The restricted phase of exhalation is the time period during the exhalation phase
where an inspiration trigger is not allowed. The restricted phase of exhalation is
defined as the first 200 ms of exhalation OR the time it takes for expiratory flow to
drop to ≤ 50% of the peak expiratory flow, OR the time it takes for the expiratory flow
to drop to ≤ 0.5 L/min (whichever is longest). The restricted phase of exhalation will
end after five (5) s of exhalation have elapsed regardless of the measured expiratory
flow rate.
After selecting the spontaneous breath type, choose the level of pressure support
(PSUPP) for PS, Support volume (VT SUPP) for VS or percent support for TC and PAV+ (if
the PAV+ option is installed) and specify the rise time % and ESENS, where available.
Changes to the spontaneous breath type setting phase in at the start the next inspi-
ration.
Note:
In any delivered spontaneous breath, either INVASIVE or NIV, there is always a target
inspiratory pressure of at least 1.5 cmH2O applied.
During spontaneous breathing, the patient's respiratory control center rhythmically
activates the inspiratory muscles. The support type setting allows selection of pres-
sure-assist to supplement the patient's pressure-generating capability.
Characteristic Implementation
Pressure or flow during inspiration Pressure rises according to the selected rise time %
Spontaneous type = PS and PSUPP < 5 cmH2O and PBW setting, with target pressure equal to the
effective pressure + PEEP:
PSUPP Effective pressure (cmH2O)
0 1.5
1 2.2
2 2.9
3 3.6
4 4.3
Pressure or flow during inspiration Pressure rises according to the selected rise time %
Spontaneous type = PS and PSUPP ≥ 5 cmH2O and PBW setting, and target pressure equals PSUPP +
PEEP.
Pressure or flow during inspiration Pressure rises according to the selected rise time %
Spontaneous type = VS and PBW setting, and target pressure equals the pres-
sure determined during the test breath or pressure
target determined from assessment of delivered
volume from the previous breath. For more informa-
tion on VS,Reference Volume Support (VS), p. 10-24.
Characteristic Implementation
Exhalation valve during inspiration Adjusts to minimize pressure overshoot and maintain
the target pressure.
Inspiratory valves during exhalation For pressure triggering: set to deliver a bias flow of 1 L/
min near the end of expiratory flow.
For flow triggering: set to deliver base flow near the
end of expiratory flow.
Exhalation valve during exhalation Adjusts to maintain the operator-selected value for
PEEP.
Pressure Support is a type of spontaneous breath, similar to PC, by which the pres-
sure is controlled to reach a constant value, preset by the practitioner, once an inspi-
ratory effort is detected. This target value is held until the detection of end of
inspiration. Subsequently, the exhalation valve control initiates the exhalation,
driving the pressure to the PEEP level.
• Rise time %
Technical Description
VS Startup
During startup, the ventilator delivers a breath (test breath) to determine the pres-
sure target needed to deliver the desired (set) volume. During the time the ventila-
tor is delivering the test breath, the message “VS startup” is displayed in the GUI’s
prompt area.
Note:
To allow for optimal function of startup and operation of VS in the ventilator it is important
not to block the tubing while the patient is undergoing suctioning or other treatment that
requires disconnection from the ventilator. The ventilator has a disconnect detection
algorithm that suspends ventilation while the patient is disconnected.
After VS Startup, the ventilator makes adjustments to the target pressure in order to
deliver the set volume (VT SUPP). In order to reach the desired volume promptly, the
maximum allowed pressure adjustments for an Adult or Pediatric patient will be
greatest during the first five breaths following Startup or a change in VT SUPP. The
values of the maximum pressure adjustments for each patient type are summarized
in the table below.
Less than five breaths ± 10.0 cmH2O ± 6.0 cmH2O ± 3.0 cmH2O
after:
VS startup or change in
VT SUPP
• COMPLIANCE LIMITED VT
Reference Ventilator Settings Range and Resolution, p. 11-9 for details on the Sponta-
neous inspired tidal volume patient data parameter available during VS breaths.
Technical Description
Tube Compensation can support all unsupported spontaneous breaths for patients
with predicted body weights ≥ 7.0 kg (15.4 lb), and for endotracheal/tracheostomy
tubes with an inside diameter (ID) of ≥ 4.5 mm. TC can be used within SPONT, BiLevel
(if this option is installed) or SIMV, all of which permit unsupported spontaneous
breaths. With BiLevel selected, TC supports spontaneous breaths at both pressure
levels.
Tube Compensation checks the flow rate every 5 ms, using an internal lookup table
which contains the flow-to-pressure relationship of the selected artificial airway, and
is used to calculate the amount of pressure needed to overcome all or part of the
resistance of the artificial airway. Based the TC setting and the instantaneous flow
measurement, the ventilator’s PSOL valves are continually adjusted, adjusting the
circuit pressure to match the changing tube-pressure compensation requirements.
Reference Non-technical Alarm Summary, p. 6-19 for details of the 1PCOMP, 1PVENT,
and 1VTI alarms associated with TC.
Reference Ventilator Settings Range and Resolution, p. 11-9 for details of the inspired
tidal volume (VTI) monitored patient data parameter a associated with TC.
The ventilator uses “soft bound” values for estimated tube inside diameter (ID) based
on PBW. Soft bounds are ventilator settings that have reached their recommended
high or low limits. When adjusting the tube size, if the inside diameter does not align
with a valid predicted body weight, a Continue button appears. Setting the ventila-
tor beyond these soft bounds requires the operator to acknowledge the prompt by
touching Continue before continuing to adjust the tube size. The limit beyond which
the tube ID cannot be adjusted is called a hard bound, and the ventilator emits an
invalid entry tone when a hard bound is reached.
WARNING:
Greater than expected ventilatory support, leading to unknown harm, can result if
the specified tube type or tube ID is smaller than the actual tube type or tube ID.
Ventilator Settings/Guidelines
Specified Performance
k P E END – P TR V· dt
W = ------------------------------------------------------------------
-
V· dt
The following figures indicate pressures at steady-state flows for ET tubes and tra-
cheostomy tubes, respectively, at 100% support at the wye for sizes between 4.5
mm and 10 mm.
PAV+ is another type of spontaneous breath, which is available only if the PAV+
option is installed. For detailed description of the operating theory, Reference Appen-
dix C in this manual.
When the ventilator is in assist-control (A/C) mode, only mandatory breaths are
delivered. These mandatory breaths can be PC, VC, or VC+ breaths. Reference Man-
datory Breath Delivery, p. 10-16 for a more detailed explanation of VC+ breaths. As for
any mandatory breath, the triggering methods can be
PTRIG, VTRIG, time-triggered, or operator initiated. If the ventilator senses the patient
initiating the breath, a PIM or assist breath is delivered. Otherwise, VIM breaths (con-
trol breaths) are delivered based on the set respiratory rate. The length of the breath
period is defined as:
T b = 60 f
where:
Tb = breath period (s)
f = set respiratory rate (breaths per minute)
The inspiratory phase length is determined by the current breath delivery settings.
At the end of the inspiratory phase, the ventilator enters the expiratory phase as
determined by the following equation:
TE = Tb – TI
where:
TE = length of the expiratory phase (s)
TI = length of inspiratory phase (s) including plateau time, TPL
The figure shown below illustrates A/C breath delivery when there is no patient
inspiratory effort detected (all inspirations are VIMs).
1 VIM 2 Tb
The figure below shows A/C breath delivery when patient inspiratory effort is
detected. The ventilator allows PIM breaths to be delivered at a rate greater than or
equal to the set respiratory rate.
1 PIM 2 Tb set
The figure shown below illustrates A/C breath delivery when there are both PIM and
VIM breaths delivered.
1 VIM 3 Tb set
2 PIM
If changes to the respiratory rate are made, they are phased in during exhalation
only. The new breath period depends on the new respiratory rate, is based on the
start of the current breath, and follows these rules:
• The current breath’s inspiratory time is not changed.
• A new inspiration is not delivered until at least 200 ms of exhalation have elapsed.
• The maximum time t until the first VIM for the new respiratory rate is delivered is 3.5
times the current inspiratory time or the length of the new breath period (whichever is
longer), but t is no longer than the old breath period.
• If the patient generates a PIM after the ventilator recognizes the rate change and before
time t, the new rate begins with the PIM.
Switching to A/C mode from any other mode causes the ventilator to phase in a VIM
and set the start time for the beginning of the next A/C breath period. Following this
VIM, and before the next A/C period begins, the ventilator responds to the patient’s
inspiratory efforts by delivering mandatory breaths.
The first A/C breath (VIM breath) is phased in while following these rules:
• The breath is not delivered during an inspiration.
• The ventilator ensures the apnea interval elapses at least five (5) s after the beginning of
exhalation.
• Any other specially scheduled event (for example, a respiratory mechanics maneuver or
any pause maneuver) is canceled and rescheduled at the next interval.
When the first VIM of the new A/C mode is delivered depends on the mode and
breath type active when the mode change is requested.
1 Tb = SIMV breath period (includes Tm and Ts 3 Ts = Spontaneous interval (VIM delivered if no PIM
delivered during Tm
2 Tm = Mandatory interval (reserved for a PIM breath)
The first part of the period is the mandatory interval (Tm) which is reserved for a PIM.
If a PIM is delivered, the Tm interval ends and the ventilator switches to the second
part of the period, the spontaneous interval (Ts), which is reserved for spontaneous
breathing for the remainder of the breath period. At the end of an SIMV breath
period, the cycle repeats. If a PIM is not delivered during the mandatory interval, the
ventilator delivers a VIM at the end of the mandatory interval, then switches to the
spontaneous interval. The following figure shows an SIMV breath period where a
PIM is delivered within the mandatory interval. Any subsequent trigger efforts
during Its yield spontaneous breaths. As shown, Tm transitions to Ts when a PIM is
delivered.
The following figure shows an SIMV breath period where a PIM is not delivered
within the mandatory interval.
1 VIM 3 Ts
In SIMV, mandatory breaths are identical to those in A/C mode if the ventilator’s
respiratory rate setting is greater than the patient’s natural respiratory rate. Sponta-
neous breaths are identical to those in SPONT mode if the ventilator setting for respi-
ratory rate is significantly below the patient’s natural respiratory rate. Patient
triggering must meet the requirements for pressure and flow sensitivity.
The procedure for setting the respiratory rate in SIMV is the same as in A/C mode.
Once the respiratory rate f is set, the SIMV interval period Tb in seconds is:
T b = 60 f
During the mandatory interval, if the patient triggers a breath according to the
current setting for pressure or flow sensitivity, the ventilator delivers a PIM. Once a
mandatory breath is triggered, Tm ends, Ts begins, and any further trigger efforts
yield spontaneous breaths. During the spontaneous interval, the patient can take as
many spontaneous breaths as allowed. If no PIM or OIM is delivered by the end of
the mandatory interval, the ventilator delivers a VIM and transitions to the sponta-
neous interval at the beginning of the VIM.
The SIMV breathing algorithm delivers one mandatory breath each period interval,
regardless of the patient’s ability to breathe spontaneously. Once a PIM or VIM is
delivered, all successful patient efforts yield spontaneous breaths until the cycle
interval ends. The ventilator delivers one mandatory breath during the mandatory
interval, regardless of the number of successful patient efforts detected during the
spontaneous interval. (An OIM delivered during the mandatory interval satisfies the
mandatory breath requirement, and causes Tm to transition to Ts.)
The maximum mandatory interval for any valid respiratory rate setting in SIMV is
defined as the lesser of:
• 0.6 x the SIMV interval period (Tb), or
• ten s.
remaining constant), breath stacking leads to reduced tidal volumes, which can be
detected by the low tidal volume and minute ventilation alarms.
In SIMV mode it is possible for the respiratory rate to drop temporarily below the f
setting (unlike A/C mode, in which fTOT is always greater than or equal to the f set-
ting). If the patient triggers a breath at the beginning of a breath period, then does
not trigger another breath until the maximum mandatory interval for the following
breath has elapsed, a monitored respiratory rate less than the respiratory rate setting
can result.
If a spontaneous breath occurs toward the end of the spontaneous interval, inspira-
tion or exhalation can still be in progress when the SIMV interval ends. No VIM, PIM,
or OIM is allowed during the restricted phase of exhalation. In the extreme, one or
more expected mandatory breaths could be omitted. When the expiratory phase of
the spontaneous breath ends, the ventilator reverts to its normal criteria for deliver-
ing mandatory breaths.
If an OIM is detected during the mandatory interval, the ventilator delivers the cur-
rently specified mandatory breath then closes Tm and transitions to Ts. If an OIM is
detected during the spontaneous interval, the ventilator delivers the currently spec-
ified mandatory breath, but the SIMV cycle timing does not restart if OIM breaths are
delivered during Ts.
Switching the ventilator to SIMV from any other mode, causes the ventilator to
phase in a VIM and set the start time for the next SIMV period. Following this VIM, but
before the next SIMV period begins, the ventilator responds to successful patient
inspiratory efforts by delivering spontaneous breaths. The first SIMV VIM breath is
phased in according to the following rules:
• The VIM breath is not delivered during an inspiration or during the restricted phase of
exhalation.
• If the current mode is A/C, the first SIMV VIM is delivered after the restricted phase of
exhalation plus the shortest of the following intervals, referenced to the beginning of
the last or current inspiration: 3.5 TI, current TA, or the length of the current breath
period.
• If the current mode is SPONT, and the current or last breath type was spontaneous or
OIM, the first SIMV VIM is delivered after the restricted phase of exhalation plus the short-
est of the following intervals, referenced to the beginning of the last or current inspira-
tion: 3.5x TI or current TA.
• If the current mode is BiLevel in the PH state and the current breath is mandatory, the
PEEP level will be reduced to PL once the exhalation phase is detected
The time t until the first VIM of the new A/C mode is the lesser of:
– PEEP transition time + 2.5 x duration of the active gas delivery phase, or
• If the current mode is BiLevel in the PH state and the current breath is spontaneous, the
PEEP level will be reduced once the exhalation phase is detected.
The time t until the first VIM of the new A/C mode is the lesser of:
– PEEP transition time + 2.5 x duration of the spontaneous inspiration, or
– the start time of the spontaneous breath + the length of the apnea interval (TA).
• If the current mode is BiLevel in the PL state and the current breath is mandatory, the
time t until the first VIM of the new A/C mode is the lesser of:
– PEEP transition time + 2.5 x duration of the active gas delivery phase, or
• If the current mode is BiLevel in the PL state and the current breath is spontaneous and
the spontaneous start time has occurred during PL, the time t until the first VIM of the
new A/C mode is the lesser of:
• If the current mode is BiLevel in the PL state and the current breath is spontaneous and
the spontaneous start time has occurred during PH, the time t until the first VIM of the
new A/C mode is the lesser of:
– the start time of the spontaneous breath + the length of the apnea interval (TA).
If the command to change to SIMV occurs after the restricted phase of exhalation
has ended, and before a next breath or the apnea interval has elapsed, the ventilator
delivers the first SIMV VIM at the moment the command is recognized.
The point at which the new rate is phased in depends on the current phase of the
SIMV interval and when the rate change command is accepted. If the rate change
occurs during the mandatory interval, the maximum mandatory interval is that for
the new or old rate, whichever is less. If the patient generates a successful inspiratory
effort during the spontaneous interval, the ventilator responds by delivering a spon-
taneous breath.
Respiratory rate changes are phased in during exhalation only. The new SIMV inter-
val is determined by the new respiratory rate and is referenced to the start of the
current SIMV period interval, following these rules:
• Inspiratory time (TI) of current breath is neither truncated nor extended.
• The new inspiration is not delivered until 200 ms of exhalation have elapsed.
• VS
• TC
The inspiratory phase begins when the ventilator detects patient effort during the
ventilator’s exhalation phase. Breath delivery during the inspiratory phase is deter-
mined by the settings for pressure support, PEEP, rise time%, and expiratory sensitiv-
ity, unless the breath is an OIM breath.
If Tube Compensation (TC), or Proportional Assist Ventilation (PAV+) (if the PAV+,
option is installed) is selected as the spontaneous type, breath delivery during the
inspiratory phase is determined by the settings for% support (% Supp), expiratory
sensitivity, tube ID, and tube type.
Note:
Given the current ventilator settings, if PAV+ would be an allowable spontaneous type
(except that tube ID < 6 mm) then PAV+ becomes selectable. If selected, tube ID is set to its
New Patient default value based on the PBW entered. An attention icon for tube ID appears.
If Volume Support (VS) is selected as the spontaneous type, breath delivery during
the inspiratory phase is determined by rise time %, volume support level (VT SUPP),
expiratory sensitivity, and PEEP.
Inspiratory pauses are only possible during OIM breaths, and expiratory pauses are
not allowed during SPONT.
Expiratory trigger methods include:
• ESENS (% flow deceleration from peak inspiratory flow)
• 1PPEAK
If the operator changes to SPONT mode during an A/C or SIMV inspiration (manda-
tory or spontaneous), the inspiration is completed, unaffected by the mode change.
Because SPONT mode has no special breath timing requirements, the ventilator
then enters the exhalation phase and waits for the detection of patient inspiratory
effort, a manual inspiration, or apnea detection.
The ventilator declares apnea when no breath has been delivered by the time the
operator-selected apnea interval elapses, plus a small increment of time (350 ms).
This increment allows time for a patient who has begun to initiate a breath to trigger
inspiration and prevent the ventilator from declaring apnea when the apnea interval
is equal to the breath period.
The apnea timer resets whenever an inspiration begins, regardless of whether the
inspiration is patient-triggered, ventilator-triggered, or operator-initiated. The venti-
lator then sets a new apnea interval beginning from the start of the current inspira-
tion. To hold off apnea ventilation, another inspiration must be delivered before (the
current apnea interval + 350 ms) elapses. Apnea detection is suspended during a dis-
connect, occlusion, or safety valve open (SVO) state.
Apnea is not declared when the apnea interval setting equals or exceeds the breath
period. For example, if the respiratory rate setting is 4/min, an apnea interval of 15 s
or more means apnea cannot be detected. The ventilator bases apnea detection on
inspiratory (not expiratory) flow, and allows detection of a disconnect or occlusion
during apnea ventilation. Apnea detection is designed to accommodate interrup-
tions to the typical breathing pattern due to other ventilator features that temporar-
ily extend the inspiratory or expiratory intervals (rate changes, for example), but still
detect a true apnea event.
The figure below shows an apnea breath where TA equals the breath period.
1 Tb0 3 PIM
The figure below shows an apnea breath with TA greater than the breath period.
1 Tb0 4 VIM
3 PIM
The following figure shows an apnea breath with TA less than the breath period.
5 Apnea VIM
When apnea is declared, the ventilator delivers apnea ventilation according to the
current apnea ventilation settings and displays the apnea settings on the graphical
user interface (GUI). Regardless of the apnea interval setting, apnea ventilation
cannot begin until inspiration of the current breath is complete and the restricted
phase of exhalation has elapsed.
All apnea and non-apnea settings remain active on the GUI during apnea ventila-
tion. Both non-apnea and apnea settings changes are phased in according to the
applicable rules. If apnea ventilation is active, new settings are accepted but not
implemented until non-apnea ventilation begins. Allowing key entries after apnea
detection allows adjustment of the apnea interval at setup, regardless of whether
apnea has been detected. During apnea ventilation, the Manual Inspiration key is
active, but Expiratory Pause and Inspiratory Pause keys are not active. The increase
O2 control is active during apnea ventilation, because apnea detection is likely
during suctioning.
The apnea respiratory rate must be ≥ 60/TA Additionally, apnea settings cannot
result in an I:E ratio > 1.00:1.
The following strategy is designed to allow SIMV to avoid triggering apnea ventila-
tion if a VIM breath can be delivered instead:
• If the apnea interval (TA) elapses at any time during the mandatory interval, the ventila-
tor delivers a VIM rather than beginning apnea ventilation.
The figure below shows an illustration of how SIMV is designed to deliver a VIM
rather than trigger apnea ventilation, when possible.
1 Tb 5 TA
2 Last breath (PIM) 6 Tm (If TA elapses during Tm, ventilator delivers a VIM rather than
beginning apnea ventilation
3 VIM 7 Ts
4 Tm max
How a new apnea interval is phased in depends on whether or not apnea ventilation
is active. If apnea ventilation is active, the ventilator accepts and implements the
new setting immediately. During normal ventilation (that is, apnea ventilation is not
active), these rules apply:
• If the new apnea interval setting is shorter than the current (or temporarily extended)
apnea interval, the new value is implemented at the next inspiration.
• If the new apnea interval setting is longer than the current (or temporarily extended)
apnea interval, the old interval is extended to match the new interval immediately.
10.13.1 Occlusion
The ventilator detects severe patient circuit occlusions in order to protect the
patient from excessive airway pressures, or from receiving little or no gas. Occlusions
require immediate attention to remedy.
The ventilator detects a severe occlusion if:
• The inspiratory or expiratory limb of the breathing circuit is partially or completely
occluded (condensate or secretions collected in a gravity-dependent loop, kinked or
crimped tubing, etc.).
• The ventilator EXHAUST port is blocked or resistance through the port is too high.
• The exhalation valve fails in the closed position (occlusion detection at the From patient
port begins after 195 ms of exhalation has passed.)
• The exhalation valve fails in the closed position and the pressure in the exhalation limb
is less than 2 cmH2O.
• Silicone tubing is attached to the EXHAUST port of the ventilator (i.e. for metabolic mon-
itoring purposes).
The ventilator checks the patient circuit for occlusions during all modes of breathing
(except Stand-by state and Safety Valve Open) at delivery of every breath. Once the
circuit check begins, the ventilator detects a severe occlusion of the patient circuit
within 200 ms. The ventilator checks the EXHAUST port for occlusions during the
expiratory phase of every breath (except during disconnect and safety valve open).
Once the EXHAUST port check begins, the ventilator detects a severe occlusion
within 100 ms following the first 200 ms of exhalation. All occlusion checking is dis-
abled during pressure sensor autozeroing.
When an occlusion is detected, an alarm sounds, the ventilator enters the OSC
(Occlusion Status Cycling) state and displays a message indicating the length of
time the patient has gone without ventilation (how long the ventilator has been in
OSC). This alarm has the capability to autoreset, since occlusions such as those due
to patient activity (for example, crimped, or kinked tubing) can correct themselves.
Once a severe occlusion is detected, the ventilator acts to minimize airway pressure.
Because any severe occlusion places the patient at risk, the ventilator minimizes the
risk while displaying the length of time the patient has been without ventilatory sup-
port. Severe occlusion is detected regardless of what mode or triggering strategy is
in effect. When a severe occlusion is detected, the ventilator terminates normal ven-
tilation, terminates any active alarm silence, annunciates an occlusion alarm, and
enters the safe state (exhalation and inspiratory valve de-energized and safety valve
open) for 15 s or until inspiratory pressure drops to 5 cmH2O or less, whichever
comes first.
During a severe occlusion, the ventilator enters OSC, in which it periodically
attempts to deliver a pressure-based breath while monitoring the inspiration and
expiration phases for the existence of a severe occlusion. If the severe occlusion is
corrected, the ventilator detects the corrected condition after two complete OSC
breath periods during which no occlusion is detected. When the ventilator delivers
an OSC breath, it closes the safety valve and waits 500 ms for the safety valve to close
completely, delivers a breath with a target pressure of 15 cmH2O for 2000 ms, then
cycles to exhalation. This breath is followed by a mandatory breath according to the
current settings, but with PEEP = 0 and O2% equal to 100% for adult/pediatric circuit
types or 40% for neonatal circuits. During OSC (and only during OSC), the 2PPEAK
(high circuit pressure) alarm limit is disabled to ensure it does not interfere with the
ability of the ventilator to detect a corrected occlusion. When the ventilator does not
detect a severe occlusion, it resets the occlusion alarm, re-establishes PEEP, and rein-
states breath delivery according to current settings.
Inspiratory and expiratory pause, and manual inspirations are suspended during a
severe occlusion. Pause maneuvers are canceled by a severe occlusion. During a
severe occlusion, ventilator settings changes are possible. Severe occlusions are not
detected when the ventilator is in the Safety Valve Open (SVO) state.
A corrected occlusion is detected within 15 s.
10.13.2 Disconnect
A circuit disconnect condition is detected when the ventilator cannot ensure that a
patient is receiving sufficient tidal volume (due to a large leak or disconnected
patient circuit). This discussion applies when Leak Sync is disabled.
When a disconnect is detected, an alarm sounds, the ventilator indicates that a dis-
connect has been detected, and displays a message indicating the length of time
the patient has gone without ventilation.
Patient data are not displayed during a circuit disconnect condition.
The ventilator monitors the expiratory pressure and flow, delivered volume, and
exhaled volume to declare a disconnect using any of these methods:
• The ventilator detects a disconnect when the expiratory pressure transducer measures
no circuit pressure and no exhaled flow during the first 200 ms of exhalation. The venti-
lator postpones declaring a disconnect for another 100 ms to allow an occlusion (if
detected) to be declared first, because it is possible for an occlusion to match the dis-
connect detection criteria.
• Despite many possible variations of circuit disconnections and/or large leaks, it is possi-
ble for a patient to generate some exhaled flow and pressure. The ventilator then uses
the disconnect sensitivity (DSENS, the percentage of delivered volume lost during the
exhalation phase of the same breath to declare a disconnect) setting to detect a discon-
nect.
• If the disconnect occurs at the endotracheal tube, the exhaled volume will be much less
than the delivered volume for the previous inspiration. The ventilator declares a discon-
nect if the exhaled volume is lower than the DSENS setting for three consecutive breaths.
The DSENS setting helps avoid false detections due to leaks in the circuit or the patient’s
lungs, and the three-consecutive-breaths requirement helps avoid false detections due
to a patient out-drawing the ventilator during volume control (VC) breaths.
• Flow less than a value determined using the DSENS setting and pressure less than 0.5
cmH2O detected for ten (10) consecutive seconds during exhalation.
WARNING:
When vent type is NIV, and DSENS setting is turned OFF, the system may not sound an
alarm for leaks and some disconnect conditions.
Once the ventilator detects a patient circuit disconnect, the ventilator declares a
high-priority alarm and discontinues breath delivery, regardless of what mode
(including apnea) was active when the disconnect was detected. If there is an active
alarm silence when the disconnect occurs, the alarm silence is NOT canceled. The
ventilator displays the length of time the patient has been without ventilatory sup-
port. During the disconnect, the exhalation valve closes, idle flow (10 L/min flow at
100% O2 or 40% O2 in NeoMode, if available with Leak Sync disabled and 20 L/min
with Leak Sync enabled) begins, and breath triggering is disabled. A message
appears identifying how long the patient has gone without ventilatory support.
The ventilator monitors both expiratory flow and circuit pressures to detect recon-
nection. The ventilator declares a reconnect if any of the following criteria are met
for the applicable time interval:
• Exhaled idle flow within the reconnect threshold is detected.
• Inspiratory and expiratory pressures are both above or both below reconnect threshold
levels or,
If the disconnect condition is corrected, the ventilator detects the corrected condi-
tion within one second.
Ventilator triggering, apnea detection, expiratory and inspiratory pause, manual
inspirations, and programmed maneuvers or one-time events are suspended during
a patient circuit disconnect condition. Spirometry is not monitored during a discon-
nect, and all alarms based on spirometry values are disabled. During a disconnect
condition, ventilator settings changes are possible.
If the disconnect alarm is autoreset or manually reset, the ventilator re-establishes
PEEP. Once PEEP is reestablished, the ventilator reinstates breath delivery according
to settings in effect before the disconnect was detected.
Occlusion and disconnection cannot be declared at the same time. Therefore, the
ventilator annunciates only the first event to be declared.
Circuit disconnect detection is not active during OSC, SVO, or prior to patient con-
nection.
• Safety PCV
• SVO
• When any other respiratory maneuver has already taken place during the same breath
The GUI also displays any maneuver request, distinguishing between requests that
are accepted or rejected, and any maneuver that has begun, ended, or has been
canceled.
When a maneuver is selected, a GUI information panel is opened, displaying the
maneuver name, user prompts and controls, and recent calculated results.
Any maneuver is canceled automatically upon declaration of any of the following
alarms:
• 1PPEAK alarm
• 1PVENT alarm
• 1VTI
The following Respiratory Mechanics maneuvers are not available in BiLevel ventila-
tion:
• P0.1 – Occlusion Pressure
• VC – Vital Capacity
Note:
Inspiratory pause and expiratory pause maneuvers can be performed directly by pressing
the respective keys on the GUI or by swiping the Menu tab on the left side of the GUI. For
more information on how to perform Respiratory Mechanics Maneuvers from the Menu tab,
Reference Respiratory Mechanics Maneuvers, p. 4-27.
An inspiratory pause extends the inspiratory phase of a single mandatory breath for
the purpose of measuring end inspiratory circuit pressure which is used to calculate
static compliance of the patient’s lungs and thorax (CSTAT), static resistance of the
respiratory system (RSTAT), and inspiratory plateau pressure (PPL). To calculate these
pressures, the inspiratory and exhalation valves are closed, allowing pressures on
both sides of the artificial airway to equalize, revealing the actual lung inflation pres-
sure during a no-flow condition. An inspiratory pause can be either automatically or
manually administered, and is only available during the next mandatory breath in
A/C, SIMV, BiLevel or SPONT modes. In BiLevel, an inspiratory pause maneuver is
scheduled for the next inspiration prior to a transition from PH to PL. Only one inspi-
ratory pause is allowed per breath. An inspiratory pause cannot occur during apnea
ventilation, safety PCV, Stand-by state, Occlusion, and SVO.
An automatic inspiratory pause begins when the inspiratory pause key is pressed
momentarily or the maneuver is started from the GUI screen. Reference To access
respiratory mechanics maneuvers, p. 4-27 for more information on performing respi-
ratory mechanics maneuvers from the Menu tab on the GUI rather than using the
keys on the GUI. The pause lasts at least 0.5 s but no longer than three (3) s. A manual
inspiratory pause starts by pressing and holding the inspiratory pause key. The
pause lasts for the duration of the keypress (up to seven (7) s).
An active manual inspiratory pause is considered complete if any of the following
occur:
• The inspiratory pause key is released and at least two (2) s of inspiratory pause have
elapsed or pressure stability conditions have been detected for not less than 0.5 s.
A manual inspiratory pause maneuver request (if the maneuver is not yet active) will
be canceled if any of events 1- 10 occur. Reference Inspiratory and Expiratory Pause
Events.
4 A disconnect is detected
5 Occlusion is detected
6 Apnea is detected
14 BUV is entered
An automatic inspiratory pause maneuver request (if the maneuver is not yet
active) will be canceled if events 1-9, 11,12,14, or 15 occur. Reference Inspiratory and
Expiratory Pause Events.
Other characteristics of inspiratory pause include:
• During an inspiratory pause, the apnea interval (TA) is extended by the duration of the
inspiratory pause.
• If the ventilator is in SIMV, the breath period during which the next scheduled VIM
occurs will also be extended by the amount of time the inspiratory pause is active.
• All activations of the inspiratory pause control are logged in the Patient Data Log.
• When calculating I:E ratio, inspiratory pause is considered part of the inspiration phase.
• The expiratory time remains unchanged, and will result in a change in the I:E ratio for
the breath that includes the inspiratory phase.
Once the inspiratory Pause maneuver is completed the operator can review the
quality of the maneuver waveform and accept or reject the maneuver data.
at least 0.5 s, but no longer than 3.0 s. A manual expiratory pause starts by pressing
and holding the expiratory pause key and lasts for the duration of the key-press (up
to 15 s).
An active manual expiratory pause is terminated if any of events 1-12 occur. Refer-
ence Inspiratory and Expiratory Pause Events, p. 10-50.
An active manual expiratory pause is complete if the expiratory pause key is
released and at least three (3) s of expiratory pause have elapsed, pressure stability
conditions have been detected for ≥ 0.5 s, or pause duration lasts 15 s.
An active automatic expiratory pause is terminated if any of events 1, 3, or 11-13
occur.Reference Inspiratory and Expiratory Pause Events, p. 10-50.
An active automatic expiratory pause is complete if pause duration reaches three
(3) s or pressure stability conditions have been detected for ≥ 0.5 s, or pause duration
lasts 15 s.
The automatic expiratory pause maneuver request (the maneuver is not yet active)
is canceled if events 1-9, 11, 12, or 15 occur:
The automatic expiratory pause maneuver is terminated and inspiration begun if
any of events 1, 3, or 11-13 occur. Reference Inspiratory and Expiratory Pause Events,
p. 10-50.
Other characteristics of expiratory pause include:
• During an active manual expiratory pause, severe occlusion detection is suspended.
• When calculating I:E ratio, the expiratory pause is considered part of the exhalation
phase.
• During the expiratory pause, the inspiratory time remains unchanged, so the I:E ratio is
changed for the breath that includes the expiratory pause.
• All activations of the expiratory pause control are logged in the Patient Data log.
Once the expiratory pause maneuver is completed the operator can review the
quality of the maneuver waveform and accept or reject the maneuver data.
The Negative Inspiratory Force (NIF) maneuver is a coached maneuver where the
patient is prompted to draw a maximum inspiration against an occluded airway (the
inspiratory and exhalation valves are fully closed).
A NIF maneuver is canceled if:
• Disconnect is detected
• Occlusion is detected
• SVO is detected
• The maneuver has been active for 30 s and an inspiration is not detected
P0.1 is the negative airway pressure (delta pressure change) generated during the
first 100 ms of an occluded inspiration. It is an estimate of the neuromuscular drive
to breathe.
When a P0.1 maneuver ends successfully, the calculated airway pressure displays on
the waveforms screen and on the maneuver panel. A P0.1 maneuver is terminated if
seven (7) s elapse and a trigger has not been detected to activate the maneuver.
A P0.1 maneuver is canceled if:
• Disconnect is detected
• Occlusion is detected
• SVO is detected
The Vital Capacity (VC) maneuver is a coached maneuver where the patient is
prompted to draw a maximum inspiration (regardless of the current settings) and
then to slowly and fully exhale.
When the Vital Capacity maneuver becomes active, the ventilator delivers a sponta-
neous inspiration in response to patient effort (with
PSUPP = 0, Rise time % = 50, and ESENS = 0), and then allows for a full exhalation effort.
When a Vital Capacity maneuver is requested, a single Volume-Time waveform grid
is automatically displayed. A Vital Capacity maneuver is canceled if:
• Disconnect is detected
• Occlusion is detected
• SVO is detected
• Cancel is touched
Apnea ventilation is a backup mode and starts if the patient fails to breathe within
the apnea interval (TA) set by the operator. TA defines the maximum allowable
length of time between the start of inspiration and the start of the next inspiration.
Available settings include mandatory type (PC or VC). For PC breaths the allowable
settings are
• Apnea interval (TA)
• Flow pattern
• O2%
During apnea ventilation with PC selected as the mandatory type, rise time % is fixed
at 50%, and the constant parameter during a rate change is inspiratory time (TI).
If apnea is possible (that is, if (60/f) > TA) increasing the non-apnea O2% setting auto-
matically changes apnea ventilation O2% if it is not already set higher than the new
non-apnea O2%. Apnea ventilation O2% does not automatically change by decreas-
ing the non-apnea O2%. Whenever there is an automatic change to an apnea set-
ting, a message appears on the GUI, and the apnea settings screen appears.
During apnea ventilation, changes to all non-apnea ventilation settings are allowed,
but the new settings do not take effect until the ventilator resumes normal ventila-
tion. Being able to change TA during apnea ventilation can avoid immediately re-
entering apnea ventilation once normal ventilation resumes.
Because the minimum value forTA is 10 s, apnea ventilation cannot take place when
non-apnea f is greater than or equal to 5.8 1/min.The ventilator does not enter
apnea ventilation if TA is equal to the breath period interval. Set TA to a value less
than the expected or current breath period interval as a way of allowing the patient
to initiate breaths while protecting the patient from the consequences of apnea.
Together, circuit type and PBW (displayed in lb or kg) provide the basis for new
patient values and absolute limits on various ventilator settings such as tidal volume
(VT) and Peak flow (VMAX). Run SST in order to change the circuit type.The table
below gives the minimum, maximum, and new patient default values for VT based
on circuit type.
Reference Ventilator Settings Range and Resolution, p. 11-9, VT setting, for more infor-
mation on VT calculations based on PBW and circuit type.
Table 10-7. Peak Flow and Circuit Type (Leak Sync Disabled)
Neonatal 30 L/min
Pediatric 60 L/min
There are two Vent Type choices: INVASIVE and NIV (non-invasive). INVASIVE ventila-
tion is conventional ventilation used with endotracheal or tracheostomy tubes. All
installed software options, breathing modes, breath types, and trigger types are
available during INVASIVE ventilation.
NIV interfaces include non-vented full-faced or nasal masks or nasal prongs. Refer-
ence NIV Breathing Interfaces, p. 4-22 for a list of interfaces that have been success-
fully tested with NIV).
NIV enables the ventilator to handle large system leaks associated with these inter-
faces by providing pressure-based disconnect alarms, minimizing false disconnect
alarms, and replacing the INSPIRATION TOO LONG alarm with a High Spontaneous
Inspiratory Time limit (2TI SPONT) setting and visual indicator.
The following list shows the subset of INVASIVE settings active during NIV:
• Mode — A/C, SIMV, SPONT. (BiLevel is not available during NIV.)
During NIV alarm setup, the clinician may set alarms to OFF and must determine if
doing so is appropriate for the patient’s condition.
Specifying the mode defines the types and sequences of breaths allowed for both
INVASIVE and NIV Vent Types.
SIMV INVASIVE: PC, VC, or VC+ Pressure supported (PS) or TC Each new breath begins
NIV: VC or PC with a mandatory inter-
val, during which a
patient effort yields a
synchronized mandatory
breath. If no patient
effort is detected during
the mandatory interval,
the ventilator delivers a
mandatory breath. Sub-
sequent patient efforts
before the end of the
breath yield sponta-
neous breaths.
SPONT Not allowed (PC or VC INVASIVE: Pressure supported All spontaneous (except
allowed only for manual (PS), Tube compensated (TC), for manual inspirations).
inspirations). Volume supported (VS), Propor-
tionally assisted (PAV+)
NIV: PS
Breath types must be defined before settings can be specified. There are only two
categories of breath type: mandatory and spontaneous. Mandatory breaths are
volume controlled (VC) or pressure controlled (PC or VC+). The ventilator currently
offers spontaneous breaths that are pressure supported (PS) volume supported (VS),
tube compensated (TC), or proportionally assisted (PAV+), if the PAV+ option is
installed. The figure below shows the modes and breath types available on the ven-
tilator.
A/C,
SIMV,
SPONT,
BiLevel
Mandatory Spontaneous
PC VC VC+ PS TC VS PAV+
The mode setting defines the interaction between the ventilator and the patient.
• Assist/control (A/C) mode allows the ventilator to control ventilation within boundaries
specified by the practitioner. All breaths are mandatory, and can be PC, VC, or VC+.
• Spontaneous (SPONT) mode allows the patient to control ventilation. The patient must
be able to breathe independently, and exert the effort to trigger ventilator support.
• BiLevel is a mixed mode that combines both mandatory and spontaneous breath types.
Breaths are delivered in a manner similar to SIMV mode with PC selected, but providing
two levels of pressure. The patient is free to initiate spontaneous breaths at either pres-
sure level during BiLevel.
Changes to the mode are phased in at the start of inspiration. Mandatory and spon-
taneous breaths can be flow- or pressure-triggered.
The ventilator automatically links the mandatory type setting to the mode setting.
During A/C or SIMV modes, once the operator has specified volume or pressure, the
ventilator displays the appropriate breath parameters. Changes in the mandatory
type are phased in at the start of inspiration.
The f setting determines the minimum number of mandatory breaths per minute for
ventilator-initiated mandatory breaths in A/C, SIMV, and BiLevel modes.If the mode
is A/C or SIMV and VC is the breath type, specifying VMAX and flow pattern deter-
mines TI, TE, and I:E. In PC breaths, specifying TI automatically determines the other
timing variables. Reference Inspiratory Time, p. 10-63 for an explanation of the inter-
dependencies of f, TI, TE and I:E. Changes to the f setting are phased in at the start of
inspiration.
The ventilator does not accept a proposed f setting if it would cause the new TI or
TE to be less than 0.2 second, the TI to be greater than eight(8) s, or I:E ratio greater
than 4.00:1. (The ventilator also applies these restrictions to a proposed change to
the apnea respiratory rate, except that apnea I:E cannot exceed 1.00:1. An exception
to this rule occurs in BiLevel ventilation where the proposed f setting will allow the
I:E ratio to be greater than 4.00:1 only until the minimum TL is reached.
The tidal volume (VT) setting determines the volume of gas delivered to the patient
during a VC mandatory breath. The delivered VT is compensated for BTPS and
patient circuit compliance. Changes to the VT setting are phased in at the start of
inspiration. The VT setting only affects the delivery of mandatory breaths.
When proposing a change to the VT setting, the ventilator compares the new value
with the settings for f, VMAX, flow pattern, and TPL. If the proposed setting would
result in an I:E ratio that exceeds 4.00:1 or a TI greater than eight(8) s or less than 0.2
s, or a TE less than 0.2 s, the ventilator disallows the change.
The peak inspiratory flow (VMAX) setting determines the maximum rate of delivery
of tidal volume to the patient during mandatory VC breaths, only. Changes to VMAX
are phased in at the start of inspiration. Mandatory breaths are compliance compen-
sated, even at the maximum VMAX setting. Circuit compliance compensation does
not cause the ventilator to exceed the ventilator’s maximum flow capability.
When proposing a change to the VMAX setting, the ventilator compares the new
value with the settings for VT, f, flow pattern, and TPL. It is impossible to set a new
VMAX that would result in an I:E ratio that exceeds 4.00:1, or a TI greater than 8.0 s or
less than 0.2 s, or a TE less than 0.2 s.
The plateau time (TPL) setting determines the amount of time inspiration is held in
the patient's airway after inspiratory flow has ceased.TPL is available only during VC
mandatory breaths (for A/C and SIMV mode, and operator-initiated mandatory
breaths). TPL is not available for PC mandatory breaths. Changes to the TPL setting
are phased in at the start of inspiration.
When proposing a change to the TPL setting, the ventilator computes the new I:E
ratio and TI, given the current settings for VT, f, VMAX, and flow pattern. It is impossi-
ble to set a new TPL that would result in an I:E ratio that exceeds 4.00:1, or a TI greater
than eight s or less than 0.2 s, or a TE less than 0.2 s. For the I:E ratio calculation, TPL is
considered part of the inspiration phase.
The flow pattern setting defines the gas flow pattern of volume-controlled (VC)
mandatory breaths only. The selected values for VT and VMAX apply to both the
square or descending ramp flow patterns. If VT and VMAX and are held constant, TI
approximately halves when the flow pattern changes from descending ramp to
square (and approximately doubles when flow pattern changes from square to
descending ramp), and corresponding changes to the I:E ratio also occur. Changes
in flow pattern are phased in at the start of inspiration.
The settings for flow pattern, VT,f, TPL, and VMAX are interrelated. If any setting
change would cause any of the following, the ventilator does not allow that change
• I:E ratio > 4:1
• TE < 0.2 s
The flow sensitivity (VSENS) setting defines the rate of flow inspired by a patient that
triggers the ventilator to deliver a mandatory or spontaneous breath. When VTRIG is
selected, a base flow of gas (1.5 L/min) travels through the patient circuit during the
ventilator’s expiratory phase. Once a value for flow sensitivity is selected, the venti-
lator delivers a base flow equal to
VSENS + 1.5 L/min (base flow is not user- selectable). When the patient inhales and
their inspiratory flow exceeds the VSENS setting, a trigger occurs and the ventilator
delivers a breath. Reductions to VSENS are phased in immediately, while increases are
phased in at the start of exhalation.
When VSENS is active, it replaces pressure sensitivity (PSENS). The VSENS setting has no
effect on the PSENS setting. VSENS can be active in any ventilation mode (including
pressure supported, volume controlled, pressure controlled, and apnea ventilation).
When VSENS is active, a backup PSENS setting of 2 cmH2O is in effect to detect the
patient's inspiratory effort, even if the flow sensors do not detect flow.
Although the minimum VSENS setting of 0.2 L/min (adult/pediatric circuit types) or
0.1 L/min (neonatal circuit type) can result in autotriggering, it can be appropriate
for very weak patients. The maximum setting of 20 L/min (adult/pediatric circuit
types) or 10 L/min (neonatal circuit type) is intended to avoid autotriggering when
there are significant leaks in the patient circuit.
The pressure sensitivity (PSENS) setting selects the pressure drop below baseline
(PEEP) required to begin a patient-initiated breath (either mandatory or sponta-
neous). Changes to PSENS are phased in immediately. The PSENS setting has no effect
on the VSENS setting and is active only if the trigger type is PTRIG.
Lower PSENS settings provide greater patient comfort and require less patient effort
to initiate a breath. However, fluctuations in system pressure can cause autotrigger-
ing at very low settings. The maximum PSENS setting avoids autotriggering under
worst-case conditions if patient circuit leakage is within specified limits.
The inspiratory pressure (PI) setting determines the pressure at which the ventilator
delivers gas to the patient during a PC mandatory breath. The PI setting only affects
the delivery of PC mandatory breaths. The selected PI is the pressure above PEEP.
(For example, if PEEP is set to five cmH2O, and PI is 20 cmH2O, the ventilator delivers
gas to the patient at 25 cmH2O.) Changes to the PI setting are phased in at the start
of inspiration.
The sum of PEEP + PI+ 2 cmH2O cannot exceed the high circuit pressure (2PPEAK)
limit. To increase this sum of pressures, first raise the 2PPEAK limit before increasing
the settings for PEEP or PI. The minimum value for PI is
5 cmH2O and the maximum value is 90 cmH2O.
The inspiratory time (TI setting) determines the time during which an inspiration is
delivered to the patient for PC mandatory breaths. The ventilator accepts a setting
as long as the resulting I:E ratio and TE settings are valid. Changes to TI phase in at
the start of inspiration. Directly setting TI in VC mandatory breaths is not allowed.
The ventilator rejects settings that result in an I:E ratio greater than 4.00:1, a TI greater
than eight (8) s or less than 0.2 s, or a TE less than 0.2 s to ensure the patient has ade-
quate time for exhalation.
Setting f and TI automatically determines the value for I:E and TE.
60 f – T I = T E
This equation summarizes the relationship between TI, I:E, TE, and breath period
time:
60
T I = ------ I :E 1 + I :E
f
If the f setting remains constant, any one of the three variables (TI, I:E, or TE) can
define the inspiratory and expiratory intervals. If the f setting is low (and additional
spontaneous patient efforts are expected), TI can be a more useful variable to set
than I:E. As the f setting increases (and the fewer patient-triggered breaths are
expected), the I:E setting becomes more relevant. Regardless of which variable is
chosen, a breath timing bar always shows the interrelationship between TI, I:E, TE
and f.
The expiratory time (TE) setting defines the duration of exhalation for PC and VC+
mandatory breaths, only. Changes to the TE setting are phased in at the start of exha-
lation. Setting f and TE automatically determines the value for I:E ratio and TI. Refer-
ence Inspiratory Time, p. 10-63 for an explanation of the interdependencies of f, TI, TE,
and I:E.
The I:E ratio setting is available when I:E is selected as the constant during rate
change. The I:E setting determines the ratio of inspiratory time to expiratory time for
mandatory PC breaths. The ventilator accepts the specified range of direct I:E ratio
settings as long as the resulting TI and TE settings are within the ranges established
for mandatory breaths. Changes to the I:E ratio phase in at the start of inspiration.
Directly setting the I:E ratio in VC mandatory breaths is not allowed. Reference Inspi-
ratory Time, p. 10-63 for an explanation of the interdependencies of f, TI, TE, and I:E.
Setting f and I:E automatically determine the values for TI and TE. The maximum I:E
ratio setting of 4.00:1 is the maximum that allows adequate time for exhalation and
is intended for inverse ratio pressure control ventilation.
The high pressure level (PH) setting is the pressure level entered by the operator for
the inspiratory phase of the mandatory breath in BiLevel ventilation.
The low pressure level (PL) setting is the pressure level entered by the operator for
the expiratory phase of the mandatory breath in BiLevel ventilation.
The high time (TH) setting is the duration of time (in seconds) the ventilator main-
tains the set high pressure level in BiLevel ventilation.
The low time (TL) setting is the duration of time (in seconds) the ventilator maintains
the set low pressure level in BiLevel ventilation.
10.15.21 PEEP
This setting defines the positive end-expiratory pressure (PEEP), also called baseline
airway pressure. PEEP is the positive pressure maintained in the patient circuit
during exhalation. Changes to the PEEP setting are phased in at the start of exhala-
tion.
The sum of:
• PEEP + 7 cmH2O, or
cannot exceed the 2PPEAK limit. To increase the sum of pressures, first raise the 2PPEAK limit
before increasing the settings for PEEP, PI, or PSUPP.
If there is a loss of PEEP from occlusion, disconnect, Safety Valve Open, or loss of
power conditions, PEEP is re-established (when the condition is corrected) by the
ventilator delivering a PEEP restoration breath. The PEEP restoration breath is a 1.5
cmH2O pressure-supported breath with exhalation sensitivity of 25%, and rise time
% of 50%. A PEEP restoration breath is also delivered at the conclusion of Vent Start-
up. After PEEP is restored, the ventilator resumes breath delivery at the current set-
tings.
Note:
PEEP restoration breath parameters are not user adjustable.
The pressure support (PSUPP) setting determines the level of positive pressure above
PEEP applied to the patient's airway during a spontaneous breath. PSUPP is only avail-
able in SIMV, SPONT, and BiLevel, in which spontaneous breaths are allowed. The
PSUPP setting is maintained as long as the patient inspires, and patient demand
determines the flow rate. Changes to the 2PSUPP setting are phased in at the start of
inspiration. The pressure support setting affects only spontaneous breaths.
The sum of PEEP + PSUPP + 2 cmH2O cannot exceed the 2PPEAK limit. To increase the
sum of pressures, first raise the 2PPEAK limit before increasing the settings for PEEP
or PSUPP. Since the 2PPEAK limit is the highest pressure considered safe for the
patient, a PSUPP setting that would cause a 1PPEAK alarm requires re-evaluating the
maximum safe circuit pressure.
Volume support (VT SUPP) is defined as the volume of gas delivered to the patient
during spontaneous VS breaths. Changes to the to the VT SUPP setting are phased in
at the start of inspiration.
10.15.24 % Supp in TC
In TC, the% Supp setting represents the amount of the imposed resistance of the
artificial airway the TC breath type will eliminate by applying added pressure at the
patient circuit wye. For example, if the % Supp setting is 100%, TC eliminates 100%
of the extra work imposed the by the airway. At 50%, TC eliminates 50% of the added
work from the airway. TC is also used with BiLevel, and is available during both PH
and PL phases.
In PAV+, the % Supp setting represents the percentage of the total work of breath-
ing provided (WOB) by the ventilator. Higher inspiratory demand yields greater
support from the ventilator. The patient performs the remaining work. If the total
WOB changes (resulting from a change to resistance or compliance) the percent
support remains constant.
The rise time % setting allows adjustment of the speed at which the inspiratory pres-
sure reaches 95% of the target pressure. Rise time settings apply to PS (including a
setting of 0 cmH2O), VS, PC, or VC+ breaths. The higher the value of rise time %, the
more aggressive (and hence, the more rapid) the rise of inspiratory pressure to the
target (which equals PEEP + PI (or PSUPP)). The rise time % setting only appears when
pressure-based breaths are available. The range of rise time % is 1% to 100%. A
setting of 50% takes approximately half the time to reach 95% of the target pressure
as a setting of 1.
• For mandatory PC, VC+, or BiLevel breaths, a rise time setting of 1 produces a pressure
trajectory reaching 95% of the inspiratory target pressure (PEEP + PI in two (2) s or 2/3
of the TI, whichever is shortest.
• For spontaneous breaths (VS, or PS), a rise time setting of 1 produces a pressure trajec-
tory reaching 95% of the inspiratory target (PEEP + PSUPP) in
(0.4 x PBW-based TI TOO LONG x 2/3) s.
• When both PC and PS breaths are active, the slopes and thus the pressure trajectories
can appear to be different. Changes to TI and PI cause PC pressure trajectories to
change. Changes in rise time % are phased in at the start of inspiration.
• When PSUPP = 0, the rise time % setting determines how quickly the ventilator drives
circuit pressure to PEEP + 1.5 cmH2O.
The expiratory sensitivity (ESENS) setting defines the percentage of the measured
peak inspiratory flow at which the ventilator cycles from inspiration to exhalation in
all spontaneous breath types. When inspiratory flow falls to the level defined by
ESENS, exhalation begins. ESENS is a primary setting and is accessible from the GUI
screen. Changes to ESENS are phased in at the next patient-initiated spontaneous
inspiration.
ESENS complements rise time %. Rise time % should be adjusted first to match the
patient's inspiratory drive, and then the ESENS setting should cause ventilator exha-
lation to occur at a point most appropriate for the patient. The higher the ESENS set-
ting, the shorter the inspiratory time. Generally, the most appropriate ESENS is
compatible with the patient's condition, neither extending nor shortening the
patient's intrinsic inspiratory phase.
ESENS in a PAV+ breath is expressed in L/min instead of percent.
Note:
If DSENS is set to OFF during NIV, the ventilator is still capable of declaring a CIRCUIT
DISCONNECT alarm.
Note:
DSENS cannot be turned OFF if Leak Sync is enabled.
Changes to DSENS are phased in at the start of inspiration.
Refer to Table 11-4. for acceptable compliance and resistance ranges.
The high spontaneous inspiratory time limit setting (2TI SPONT) is available only in
SIMV or SPONT modes during NIV, and provides a means for setting a maximum
inspiratory time after which the ventilator automatically transitions to exhalation.
The default 2TI SPONT setting is based upon circuit type and PBW.
For pediatric/adult circuit types, the new patient default value is
(1.99 + (0.02 x PBW)) s
The humidification type setting sets the type of humidification system (heated expi-
ratory tube, non-heated expiratory tube, or heat-moisture exchanger (HME) used on
the ventilator and can be changed during normal ventilation or short self test (SST).
Changes in humidification type phase in at the start of inspiration.
SST calibrates spirometry partly based on the humidification type. Changing the
humidification type without rerunning SST can affect the accuracy of spirometry
and delivery.
The accuracy of the exhalation flow sensor varies depending on the water vapor
content of the expiratory gas, which depends on the type of humidification system
in use. Because the temperature and humidity of gas entering the exhalation filter
differ based on the humidification type being used, spirometry calculations also
differ according to humidification type. For optimum accuracy, rerun SST to change
the humidification type.
The dry, compressible volume in mL of the humidification chamber for the humidi-
fication type entered during SST. Only applies if a humidifier is used.
egies to minimize the impact of system faults on patient safety. In these scenarios,
The ventilator is designed to alarm and to provide the highest level of ventilation
support possible. in case of ventilator malfunction. If the ventilator is not capable of
ventilatory support, it opens the patient circuit and allows the patient to breathe
from room air if able to do so (this emergency state is called safety valve open
(SVO). Safety mechanisms are designed to be verified periodically or to have redun-
dancy. The ventilator is designed to ensure that a single-point failure does not cause
a safety hazard or affect its ability to annunciate a high-priority audible alarm.
The ventilator is designed to prevent the operator from implementing settings that
are clearly inappropriate for the patient's predicted body weight (PBW). Each setting
has either soft bounds (can be overridden) or hard bounds (no override allowed)
that alert the operator to the fact that the settings may be inappropriate for the
patient. In the event that the patient is connected without any parameters being
specified, the ventilator enters Safety PCV, a safe mode of ventilation regardless of
the circuit type in use (neonatal, pediatric, or adult) or patient's PBW. Safety PCV is
entered after POST, if a patient connection is made prior to settings confirmation.
Safety PCV uses New Patient default settings with exceptions shown in the following
table:
PBW Neonatal: 3 kg
Pediatric: 15 kg
Adult: 50 kg
mode A/C
mandatory Type PC
TI Neonatal: 0.3 s
Pediatric: 0.7 s
Adult: 1 s
PI 15 cmH2O
O2 % Neonatal: 40%
Pediatric: 100%
Adult: 100%
PEEP 3 cmH2O
PSENS 2 cmH2O
1PPEAK 20 cmH2O
Note:
In Safety PCV, expiratory pauses are not allowed.
In case of patient problems, the ventilator remains fully operative and annunciates
the appropriate alarm. The detection, response, and priority of each patient-related
alarm is determined by the actual patient problem. Reference Alarms, p. 6-4 for a
comprehensive description of the patient alarm system.
The ventilator is designed to prevent system faults. Its modular design allows the
breath delivery unit (BDU) to operate independently of the graphical user interface
(GUI) and several modules of the breath delivery sub-system have redundancy that,
if certain faults occur, provides for ventilatory support using settings that do not
depend on the suspect hardware. System faults include the following:
• Hardware faults (those that originate inside the ventilator and affect its performance)
• Soft faults (faults momentarily introduced into the ventilator that interfere with normal
operation
• Boundary checks performed at every analog measurement. These checks verify mea-
surement circuitry, including sensors.
or flow sensors. During MIX BUV, the normal mix controller is bypassed and ventila-
tion continues as set, except that the gas mix reverts to 100% Oxygen or Air,
depending on where the fault indication was detected. Backup circuits then control
the pressure in the accumulator to keep it in the proper range for the Inspiratory
Module.
Inspiratory BUV is invoked if Background Diagnostics detect a problem in the inspi-
ratory module (PSOL and/or flow sensor signal out of range). In inspiratory BUV, ven-
tilation continues with the following settings:
Mode A/C
Mandatory type PC
f Neonatal: 25 1/min
Pediatric: 16 1/min
Adult: 16 1/min
TI Neonatal: 0.3 s
Pediatric: 0.7 s
Adult: 1 s
PI 15 cmH2O
PEEP 3 cmH2O
During inspiratory BUV, the delivery PSOL is disabled, but gas delivery is achieved via
an inspiratory BUV solenoid valve, the gas flow being created by pressure in the mix
accumulator.
Exhalation BUV is invoked if problems with the Exhalation Valve driver are detected.
A backup analog circuit is enabled to control the exhalation valve though the more
advanced control features (active exhalation valve control) are not functional.
Note:
During Mix and Inspiratory BUV, gas supply to installed options is disabled.
Entry into BUV is logged in the alarm log and system diagnostic log, and the status
display provides an indicator that the ventilator is in BUV and which subsystem is
affected.
When in BUV, a high priority alarm is annunciated, and the GUI displays an alarm
banner indicating BUV, blanks patient data, and a displays a pressure waveform.
If the ventilator cannot provide any degree of reliable ventilatory support and fault
monitoring, then the ventilator alarms and enters the safety valve open (SVO) emer-
gency state. During SVO, the ventilator de-energizes the safety, expiratory, and inspi-
ratory valves, annunciates a high-priority alarm, and turns on the SVO indicator.
During SVO, a patient can spontaneously inspire room air (if able to do so) and
exhale. Check valves on the inspiratory and expiratory sides minimize rebreathing of
exhaled gas during SVO. During SVO the ventilator:
• Displays the elapsed time without ventilatory support
Visible indicators on the ventilator's GUI and status display illuminate when the ven-
tilator is in the SVO state. Other safeguards built into the ventilator include a one-
way valve (check valve) in the inspiratory pneumatic circuit allowing the patient to
inhale through the safety valve (if able to do so) with limited resistance. This check
valve also limits exhaled flow from entering the inspiratory limb to reduce the pos-
sibility of re-breathing exhaled CO2 gas.
WARNING:
Do not enter Service mode with a patient attached to the ventilator. Serious injury
could result.
11.1 Overview
This chapter contains the following specifications for the Puritan Bennett™ 980
Series Ventilator:
• Physical
• Electrical
• Interface
• Environmental
• Performance (Ranges, resolution, and accuracies for ventilator settings, alarm settings,
and patient data)
• Regulatory Compliance
WARNING:
Due to excessive restriction of the Air Liquide™, SIS, and Dräger™ hose assemblies,
reduced ventilator performance levels may result when oxygen or air supply
pressures < 50 psi (345 kPa) are employed.
11-1
Specifications
During breath delivery performance verification for flow and pressure based mea-
surements, the equipment inaccuracy is subtracted from the acceptance specifica-
tion as follows:
• Net Acceptance Gain = Requirement Specification Gain - Measurement Uncertainty
Gain
For derived parameters, such as volume, compliance, etc., the individual sensor
uncertainties are combined and applied as applicable to determine the acceptance
limits.
Weight Ventilator: 113 lb (51.26 kg) including BDU, GUI, standard base, and
primary battery
BDU only: 69 lb (31.3 kg)
Ventilator and compressor: 157 lb (71.2 kg) including GDU, GUI, ventilator
and compressor primary batteries, base assembly, and compressor
Compressor: 89 lb (40.4 kg) including base assembly
BDU only: 69 lb (31.3 kg)
A-weighted sound pressure level, At a distance of one (1) meter does not exceed 45 dBA
ventilator (average) below 500 mL/min
A-weighted sound pressure level, At a distance of one (1) meter does not exceed 49 dBA below
ventilator and compressor 500 mL/min
A-weighted sound power level, Does not exceed 58 dBA below 500 mL/min
ventilator
A-weighted sound power level, Does not exceed 63 dBA below 500 mL/min
ventilator and compressor
Connectors Inspiratory and expiratory limb connectors are 22mm OD conical fittings
compliant with ISO 5356-1
Inspiratory/ exhalation filters Refer to filter Instructions For Use for complete specifications
Oxygen and air inlet supplies Pressure: 241 to 600 kPa (35 psi to 87 psi)
Flow: Maximum of 200 L/min
Oxygen sensor life Up to one year. Operating life varies depending on oxygen usage and
ambient temperature.
Maximum limited pressure (PLIM max) A fixed pressure limit to the safety valve limits circuit
pressure to < 123 hPa (125 cmH2O) at the patient wye.
Maximum working pressure (PW max) PW max is ensured by the high pressure limit (2PPEAK)
when PI is < 100 cmH2O (98.07 hPa)
Response time to change in FiO2 setting from 21% O2 < 18 s for volumes > 150 mL
to 90% O2 (measured at the patient wye) < 19 s for volumes ≥ 30 mL but ≤ 150 mL
< 50 s for volumes ≥ 2 mL but < 30 mL
Minute volume (VE TOT) capability, compressor Up to 40 L/min BTPS, including compliance compen-
sation
Results of ventilator testing using circuits identified for use with the ventilator system
Internal Inspiratory filter particle filtration efficiency > 99.97% retention of particles 0.3 mm nominal at 100
L/min flow
Internal Inspiratory filter resistance 0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min flow
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min flow
External Inspiratory filter resistance 0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min flow
0.2 cmH2O < resistance < 2.2 cmH2O at 15 L/min flow
Combined inspiratory limb resistance 0.2 cmH2O < resistance < 5.5 cmH2O at 30 LL/min flow
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min flow
External Inspiratory filter resistance 0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min flow
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min flow
External Inspiratory filter particle filtration efficiency > 99.97% retention of particles 0.3 mm nominal at 100
L/min flow
External Inspiratory filter resistance (reusable inspira- 0.2 cmH2O < resistance < 4.2 cmH2O at 60 L/min
tory filter) 0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min
External Inspiratory filter particle filtration efficiency, > 99.97% retention of particles 0.3 mm nominal at 100
disposable inspiratory filter) L/minflow
Exhalation filter particle filtration efficiency, reusable > 99.97% retention of particles 0.3 mm nominal at 100
L/min flow
Exhalation filter resistance (pediatric/adult, reusable < 2.5 cmH2O at 100 L/min when new
and disposable) < 1.7 cmH2O at 15 L/min
Exhalation filter particle filtration efficiency, dispos- > 99.97% retention of particles 0.3 mm nominal at 100
able L/min flow
Exhalation filter resistance, disposable < 2.5 cmH2O at 100 L/min when new
Exhalation filter particle filtration efficiency > 99.97% retention of particles 0.3 m
nominal at 100 L/min flow
Exhalation filter particle filtration efficiency (neonatal, > 99.97% retention of particles 0.3 mm nominal at 100
disposable) L/min flow
Exhalation filter resistance (neonatal, disposable) < 0.58 cmH2O at 2.5 L/min
Circuit compliance (acceptable ranges of VBS compli- ADULT: 1.3 mL/cmH2O to 4.2 mL/cmH2O
ance for each patient type) PEDIATRIC: 0.9 mL/cmH2O to 3.0 mL/cmH2O
NEONATAL: 0.4 mL/cmH2O to 1.5 mL/cmH2O
Inspiratory limb circuit resistance (acceptable ranges ADULT (at 60L/min): 1.15 cmH2O to 11.0 cmH2O
of VBS inspiratory limb circuit resistance for each PEDIATRIC (at 30L/min): 0.46 cmH2O to 4.5 cmH2O
patient type) NEONATAL (at 10L/min): 0.37 cmH2O to 4.5 cmH2O
(6.0 cmH2O for Prox)
Expiratory limb circuit resistance (acceptable ranges of ADULT (at 60L/min): 1.15 ccmH2O to 11.0 cmH2O
VBS expiratory limb circuit resistance for each patient PEDIATRIC (at 30L/min): 0.46 cmH2O to 4.5 cmH2O
type) NEONATAL (at 10L/min): 0.37 cmH2O to 4.5 cmH2O
(6.0 cmH2O for Prox)
Audio alarm volume (primary) Range: High priority alarm volume range (dBA): 58
Measurement uncertainty: ± 3 dBA (volume setting 1) to 86 (volume setting 10)
Medium priority alarm volume range (dBA): 52
(volume setting 1) to 78 (volume setting 10)
Low priority alarm volume range (dBA): 50 (volume
setting 1) to 76 (volume setting 10)
Measured 1 m from front, rear, and sides of ventilator
Reference Alarm Volume Key, p. 6-9 for alarm volume
behavior during an alarm condition.
Resolution: 1
Audio alarm volume (secondary) Minimum 64 dBA measured 1 m from front, rear, and
Measurement uncertainty: ± 3 dBA sides of ventilator.
5 GND Ground
Pin Configuration
2 Relay common
4 Not connected
Operation Storage
Atmospheric Pressure 70kPa to 106 kPa (10.15 psi to (15.37 50 kPa to 106 kPa (7.25 psi to 15.37
psi) psi)
Note:
The limits marked on the device label represent out-of-box storage conditions as follows:
• Temperature: (10°C to 40°C (50°F to 104°F)
Reference Ventilator Settings Range and Resolution, p. 11-9 for ranges and resolutions
for ventilator settings. Reference Alarm Settings Range and Resolution, p. 11-17 for
alarm settings, and Reference Patient Data Range and Resolution, p. 11-20 for dis-
played patient data parameters.
Apnea ventilation A safety mode of ventilation that See individual apnea settings.
starts if the patient does not
receive a breath for an elapsed
time exceeding the apnea interval.
Apnea expiratory time (TE) For mandatory PC apnea breaths, Range: 0.20 s to 59.8 s
the time interval between the end Resolution: 0.01 s
of inspiration and the beginning of
the next inspiration.
Apnea I:E ratio In PC breath types, specifies the Range: I:E ≤ 1.00:1
ratio of apnea inspiratory time to Resolution:
apnea expiratory time. 0.01 for values > 1:10.0;
0.1 for values ≤ 1:10 and > 1:100;
1 for values ≤ 1:100
Apnea flow pattern The flow shape of the delivered Range: SQUARE, descending ramp
mandatory volume-based (VC)
apnea breath.
Apnea inspiratory pressure The pressure above PEEP at which Range: 5 cmH2O to 90-PEEP cmH2O
(PI) gas is delivered to the patient Resolution:1 cmH2O
during mandatory PC apnea
breaths.
Apnea inspiratory time (TI) Same as inspiratory time for non- Range: 0.20 s to 8 s
apnea ventilation Resolution:
0.01 s in PC or VC+, 0.02 s in VC
Apnea interval (TA) The time after which the ventilator Range: 10 s to 60 s or OFF in CPAP
transitions to apnea ventilation Resolution: 1 s
TA≥ 60/fA
Apnea peak inspiratory flow The maximum rate of tidal volume Range: When mandatory type is VC:
(VMAX) delivery during mandatory NEONATAL: 1L/min to 30 L/min
volume-based apnea breaths. PEDIATRIC: 3.0 L/min to 60 L/min
ADULT: 3.0 L/min to 150 L/min
Resolution:
0.1 L/min for flows
< 20 L/min (BTPS);
1L/min for flows ≥ 20 L/min (BTPS)
Apnea respiratory rate (fA) Sets the number of volume- or Range: 2.0 1/min to 40 1/min Resolution:
pressure-based breaths per 0.1 1/min for 2.0 to 9.9;
minute for ventilator initiated 1 1/min for 10 1/min to 40 1/min
mandatory (VIM) apnea breaths.
Apnea tidal volume (VT) Sets the volume of gas delivered to Range:
the patient’s lungs during a man- NEONATAL: 3 mL to 315 mL
datory, volume-controlled apnea PEDIATRIC/ADULT:
breath. Apnea tidal volume is com- ≥ 25 mL to 2500 mL
pensated for body temperature
and pressure, saturated (BTPS) and
the compliance of the patient cir-
cuit.
Apnea constant during rate Specifies which of the three opera- Range: TI
change tor-adjustable breath timing vari-
ables remains constant when
respiratory rate is changed during
apnea ventilation.
Circuit type Specifies the circuit for which com- Range: NEONATAL, PEDIATRIC, ADULT
pliance and resistance values
during SST have been calculated.
Constant during rate change Specifies which of the three opera- Range: I:E ratio, TI, TE for PC or VC+
tor-adjustable breath timing vari- breaths; TH:TL ratio, TH,TL in BiLevel
ables remains constant when
respiratory rate is changed.
Disconnect sensitivity Leak Sync disabled: The percent- Range (Leak Sync disabled): 20% to 95%
(DSENS) age of returned volume lost, above or OFF
which the ventilator declares a Range (Leak Sync enabled:
circuit disconnect alarm. NEONATAL:
Leak Sync enabled: The leak at Invasive: 1 L/min to 15 L/min
PEEP value in L/min, above which NIV: 1 L/min to 40 L/min
the ventilator declares a CIRCUIT PEDIATRIC: 1 L/min to 40 L/min
DISCONNECT alarm. ADULT: 1 L/min to 65 L/min
Resolution (Leak Sync disabled: 1%
Resolution (Leak Sync enabled):
0.5 L/min for values<10 L/min;
1 L/min for values ≥10 L/min
Expiratory sensitivity The percentage ofVMAX that, when Range: 1% to 80% when Spontaneous
(ESENS) reached, causes the ventilator to Type is PS, or VS
cycle from inspiration to exhala- 1 L/min to 10 L/min when Spontaneous
tion during spontaneous, pres- Type is PAV+.
sure-based breaths. Resolution:1% when Spontaneous Type
is PS, TC, or VS; 1 L/min when Sponta-
neous Type is PAV+.
NOTE: Default value is not expected to
need adjustment. Only adjust after
becoming experienced with PAV+ and
only if it is suspected that the ventilator
is not cycling at the patient’s end-of-
inspiration.
Expiratory time (TE) For PC or VC+ breaths, the time Range: ≥ 0.20 s
interval between the end of inspi- Resolution: 0.01 s
ration and the beginning of the
next inspiration. The end of the
exhalation phase is considered to
be when the flow rate at the
patient wye remains less than 0.5
L/min above the base flow.
Flow pattern The flow shape of the delivered Range: SQUARE, descending ramp
mandatory or VC breath
High spontaneous inspirato- Active in NIV only, allows the oper- Range:
ry time limit (2TI SPONT) ator to select the maximum spon- NEONATAL: 0.2 s to 1.7 s
taneous inspiratory time. PEDIATRIC/ADULT: 0.4 s to 5 s
Resolution: 0.1 s
Humidification type The type of humidification system Range: HME, non-heated expiratory
used on the ventilator. tube, heated expiratory tube
Humidifier volume The empty fluid volume of the cur- Range: 100 mL to 1000 mL
rently installed humidifier. Resolution: 10 mL
I:E ratio In PC and VC+ breath types, speci- Range: 1:299 to 149:1
fies the ratio of inspiratory time to Resolution: 0.01 for values > 1:10; 0.1 for
expiratory time. values ≤ 1:10.0 and
> 1:100.0;
1 for values ≤ 1:100
Displayed as XX:1 when I:E ≥ 1; displayed
as 1:XX when I:E < 1
Inspiratory pressure (PI) The pressure above PEEP at which Range:5 cmH2O to 90 cmH2O
gas is delivered to the patient Resolution:1 cmH2O
during mandatory PC breaths.
Inspiratory time (TI) The time during which an inspira- Range: 0.2 s to 8 s for mandatory PC and
tion is delivered to the patient VC+ breaths
during mandatory PC or VC+ TPL+ 0.2 s to 8 s in VC
breaths. Resolution: 0.01 s for PC or VC+ breaths;
0.02 s for VC breaths
Leak Sync (leak compensa- Compensates for leaks during Range: Enabled or Disabled
tion) INVASIVE or non-invasive (NIV)
ventilation.
Mandatory type The type of mandatory breath Range: PC, VC, VC+
delivered in A/C, SPONT or SIMV
modes. SPONT mode allows man-
datory type selection for operator
initiated mandatory (OIM) breaths.
mL/kg ratio The default tidal volume/PBW ratio Range: 5.0 mL/kg to 10 mL/kg
(only adjustable in Service mode) Resolution: 0.5 mL/kg
Mode The ventilation mode.The mode Range: A/C, SPONT, SIMV, BiLevel (if
determines the allowable breath option installed but not available when
types: vent type is NIV); CPAP (only available
A/C – (assist/control) – a mandato- when circuit type is NEONATAL and vent
ry mode allowing volume con- type is NIV)
trolled (VC), pressure controlled
(PC), or VC+breath types.
SPONT – allows the patient to initi-
ate the breath. Applicable SPONT
breath types are pressure support
(PS), volume support (VS), tube
compensated (TC) or PAV+ if the
PAV+ option is installed.
SIMV – Synchronized Intermittent
Mandatory Ventilation – a mixed
ventilatory mode providing man-
datory breaths and allowing a
patient spontaneous breaths
during the breath cycle.
BiLevel – a mixed ventilatory mode
combining the attributes of both
mandatory and spontaneous
breaths incorporating two pres-
sure levels, PH and PL.
Peak inspiratory flow (VMAX) The maximum rate of tidal volume Range: When mandatory type is VC:
delivery during mandatory NEONATAL: 1 L/min to 30 L/min
volume-based breaths. PEDIATRIC: 3.0 L/min to 60 L/min
ADULT: 3.0 L/min to 150 L/min
Resolution:
0.1 L/min for values <
20 L/min (BTPS);
1 L/min for values ≥ 20 L/min (BTPS)
Pressure sensitivity (PSENS) For pressure triggered breaths, Range: 0.1 cmH2O to 20.0 cmH2O
determines the amount of pres- Resolution: 0.1 cmH2O
sure below PEEP required to begin
a mandatory or spontaneous
patient initiated breath.
Pressure support (PSUPP) or The positive pressure above PEEP Range: 0 cmH2O to 70 cmH2O
PS (or PL in BiLevel) during a sponta- Resolution: 1 cmH2O
neous breath.
Spontaneous type The breath type for patient initiat- Range: PS, TC, PAV+, or VS
ed spontaneous breaths in SIMV,
SPONT, and BiLevel modes.
TH:TL ratio In BiLevel, specifies the ratio of Range: 1:299 to 4:1; in BiLevel TH:TL
insufflation time to expiratory time. Resolution: 0.01 for < 10.00:1 and >
1:10.00; 0.1for [< 100.0:1 and ≥ 10.0:1] or
[≤ 1:10.0 and > 1:100.0]; 1 for < 1:100.0 or
≥ 100:1
Volume support (VT SUPP) or The volume of gas delivered to the Range:
VS patient during spontaneous, NEONATAL: 2 mL to 310 mL
volume supported breaths. PEDIATRIC: 25 mL to 1590 mL
ADULT: 25 mL to 2500 mL
Resolution: 0.1 mL for ≥ 5 mL;
1 mL for 5 mL to 100 mL; 5 mL for 100 mL
≥ 400 mL
Tube type The type of artificial airway used to Range: Endotracheal (ET), tracheal (Trach
ventilate the patient.
Apnea interval (TA) The Apnea alarm condition indi- Range: 10 s to 60 s or OFF in CPAP
cates that neither the ventilator Resolution: 1
nor the patient has triggered a
breath for the operator-selected
Apnea Interval (TA). When the
Apnea alarm condition is true, the
ventilator invokes mandatory ven-
tilation as specified by the opera-
tor.
High circuit pressure setting The 1PPEAK alarm indicates the Range: 7 cmH2O to 100 cmH2O
(2PPEAK) patient’s airway pressure ≥ the set Resolution: 1 cmH2O
alarm level.
Low circuit pressure setting The 3PPEAK alarm indicates the Range:
(4PPEAK) measured airway pressure ≤ the NIV: OFF or ≥ 0.5 cmH2O to < 100
set alarm limit during an NIV or cmH2O
VC+ inspiration. Resolution:
0.5 cmH2O for values
< 20.0 cmH2O;
1 cmH2O for values
≥ 20 cmH2O
High exhaled minute volume The 1VE TOT alarm indicates the Range: OFF and
alarm setting (2VE TOT) measured total minute volume ≥ NEONATAL: 0.1 L/min to 10 L/min
the set alarm limit. PEDIATRIC:
0.1 L/min to 30 L/min
ADULT:
0.1 L/min to 100 L/min
Resolution:
0.005 L/min for values <
0.50 L/min;
0.05 L/min for values ≥
0.5 L/min to < 5.0 L/min;
0.5 L/min for values ≥ 5.0L/min
High exhaled tidal volume alarm The 1VTE alarm indicates that the Range: OFF and
setting (2VTE) measured exhaled tidal volume ≥ NEONATAL: 5 mL to 500 mL
the set alarm limit for spontaneous PEDIATRIC: 25 mL to 1500 mL
and mandatory breaths. ADULT: 25 mL to 3000 mL
Resolution:
1 mL for values < 100 mL;
5 mL for values ≥ 100 mL and < 400
mL;
10 mL for values ≥ 400 mL
High inspired tidal volume alarm The 1VTI alarm indicates the deliv- Range: 6 mL to 6000 mL
limit (2VTI ered volume of any breath ≥ the Resolution: 1 mL for values < 100
set alarm limit. mL;
5 mL for values ≥ 100 mL to < 400
mL; 10 mL for values ≥ 400 mL
High respiratory rate alarm setting The 1fTOT alarm indicates the mea- Range: OFF or
(2fTOT) sured breath rate ≥ the set alarm NEONATAL: 10 1/min to 170
limit. 1/ min
PEDIATRIC/ADULT: 10 1/min to
110 1/min
Resolution: 1 1/min
High spontaneous inspiratory time The 1TI SPONT indicator allows the Range:
limit (2TI SPONT) operator to select the maximum NEONATAL: 0.2 to ≤ the value of
spontaneous inspiratory time of an the NIV inspiratory time limit
NIV breath. No alarm is annunciat- trigger for the patient’s PBW and
ed; only the symbol 2TI SPONT circuit type s
appears on the screen near the NIV PEDIATRIC/ADULT: 0.4 s to ≤ the
indicator when inspiration time value of the NIV inspiratory time
exceeds the setting. If 2TI SPONT is limit trigger for the patient’s PBW
and circuit type s
exceeded, the ventilator transi-
Resolution: 0.1 s
tions from inspiration to exhala-
tion.
Low exhaled mandatory tidal The 3VTE MAND alarm indicates the Range: OFF and
volume alarm setting (4VTE MAND) measured mandatory tidal volume NEONATAL: 1 mL to 300 mL
≤ the set alarm limit. PEDIATRIC: 1 mL to 1000 mL
ADULT: 1 mL to 2500 mL
Resolution: 1.0 mL for values
< 100 mL;
5 mL for values ≥ 100 mL and
< 400 mL;
10 mL for values ≥ 400 mL
Low exhaled minute volume alarm The3VE TOT alarm indicates the Range: OFF when vent type = NIV
setting (4VE TOT) measured exhaled minute volume and
≤ the set alarm limit for mandatory NEONATAL: OFF, 0.01 L/min to 10
and spontaneous breaths. L/min
PEDIATRIC:
0.05 L/min to 30 L/min
ADULT:
0.05 L/min to 60 L/min
Resolution:
0.005 L/min for values
< 0.50 L/min;
0.05 L/min for values
≥ 0.50 L/min and < 5.0 L/min;
0.5 L/min for values > 5.0L/min
Low exhaled spontaneous tidal The 3VTE SPONT alarm indicates the Range: OFF and
volume alarm setting (4VTE SPONT) measured spontaneous tidal NEONATAL: 1 mL to 300 mL
volume ≤ the set alarm limit. PEDIATRIC: 1 to 1000 mL
ADULT: 1 to 2500 mL
Resolution: 1 mL for values
< 100 mL;
5 mL from 100 mL to < 400 mL; 10
mL for values ≥ 400 mL
Breath phase The breath phase indicator dis- Range: Control (C), Assist (A), Sponta-
plays the breath delivery phase neous (S)
(inspiration or exhalation) currently
being delivered to the patient.
Inspired tidal volume (VTI) The volume inspired for a pressure- Range:0 mL to 6000 mL
based breath. Resolution:
0.1 mL for 0 mL to 9.9 mL;1
mL for values 10 mL to 6000 mL
Inspired tidal volume (VTL) The volume inspired for each Range: 0 mL to 6000 mL
during Leak Sync breath when Leak Sync is enabled. Resolution:
1 mL for values < 10 mL;
1 mL for values 10 mL to 6000 mL
Dynamic resistance (RDYN) The change in pressure per unit Range: 0.0 cmH2O/L/s to 100
change in flow. cmH2O/L/s
Resolution:
0.1 cmH2O/L/ for values
< 10 cmH2O/L/s;
1 cmH2O/L/s for values
≥ 10 cmH2O/L/s
End expiratory flow (EEF) The rate of expiratory flow occur- Range: 0 to 150 L/min
ring at the end of exhalation. Resolution:
0.1 L/min for values < 20 L/min
1 L/min for values ≥ 20 L/min
Exhaled mandatory tidal The exhaled volume of the last Range: 0 mL to 6000 mL
volume (VTE MAND) mandatory breath. When the Resolution:
mode is SPONT, and no mandatory 0.1 mL for 0 mL to 9.9 mL;
breaths have occurred for a time 1 mL for 10 mL to 6000 mL
period ≥ 2 minutes, the VTE MAND
indicator is hidden. Mandatory
breaths can occurs during SPONT
mode via manual inspiration.
Exhaled minute volume (VE A calculated sum of the volumes Range: 0.00 L/min to 99.9 L/min
TOT)
exhaled by the patient for manda- Resolution: 0.01 L/min for 0.00 to 9.99
tory and spontaneous breaths for L/min; 0.1 L/min for 10.0 to 99.9 L/min
the previous one-minute interval
(also applies in BiLevel).
Exhaled spontaneous The sum of exhaled spontaneous Range: 0 L/min to 99.9 L/min
minute volume (VE SPONT) volumes per minute (also applies Resolution:
in BiLevel). 0.01 L/min for 0.00 to 9.99 L/min; 0.1
L/min for 10.0 to 99.9 L/min
Exhaled spontaneous tidal The exhaled volume of the last Range: 0 mL to 6000 mL
volume (VTE SPONT) spontaneous breath. Resolution: 0.1mL for 0 mL to 9.9 mL;
1 mL for 10 mL to 6000 mL
Exhaled tidal volume (VTE) The volume exhaled by the patient Range: 0 mL to 6000 mL
for the previous mandatory or Resolution:
spontaneous breath (also applies 0.1mL for 0 mL to 9.9 mL;
in BiLevel). 1 mL for 10 mL to 6000 mL
Leak Sync exhaled tidal The volume exhaled by the patient Range: 0 mL to 6000 mL
volume (VTE) for the previous mandatory or Resolution:
spontaneous breath during Leak 0.1mL for 0 mL to 9.9 mL;
Sync (also applies in BiLevel). 1 mL for 10 mL to 6000 mL
I:E ratio The ratio of the inspiratory time to Range: 1:599 to 149:1
expiratory time for the previous Resolution: 0.1 for 9.9:1 to 1:9.9; 1 for
breath. 149:1 to 10:1 and 1:10 to 1:599
Negative inspiratory force The negative pressure generated Range: ≤ 0 cmH2O to ≥ -50 cmH2O
(NIF) during a maximally forced inspira- Resolution: 1 cmH2O for values
tory effort against an obstruction ≤ -10 cmH2O;
to flow.
0.1 cmH2O for values
> -10 cmH2O
PAV based intrinsic PEEP The estimated intrinsic PEEP Range: 0 to 130 cmH2O
(PEEPI PAV) during a PAV+ breath. Intrinsic Resolution: 0.1 cmH2O for values < 10
PEEP is an estimate of the pressure cmH2O; 1 cmH2O for values
above PEEP at the end of every
≥ 10 cmH2O
pause exhalation.
PAV-based lung compliance The calculated change in pulmo- Range: 2.5 mL/cmH2O to 200 mL/
(CPAV)1 nary volume for an applied change cmH2O
in patient airway pressure when Resolution: 0.1 mL/cmH2O for values
measured under conditions of zero
< 10 mL/cmH2O;
flow during a PAV+ plateau
maneuver. When PAV+ is selected, 1 cmH2O for values
the ventilator displays the current ≥ 10 mL/cmH2O
filtered value for patient compli-
ance, and updates the display at
the successful completion of each
estimation. CPAV can be displayed
in the vital patient data banner.
Reference Vital Patient Data, p. 3-
39.
PAV-based lung elastance For a PAV+ breath, EPAV is calculat- Range: 5.0 cmH2O/L to 400 cmH2O/L
(EPAV)1 ed as the inverse of CPAV (see Resolution: 0.1 cmH2O/L for values <
above). EPAV can be displayed in 10 cmH2O/L;
the vital patient data banner. Refer- 1 cmH2O/L ≥ 10 cmH2O/L
ence Vital Patient Data, p. 3-39.
PAV-based total airway resis- RTOT is an estimated value cap- Range: 1.0 cmH2O/L/s to 80 cmH2O/
tance (RTOT)1 tured just past peak expiratory flow L/s
and is equal to the pressure loss Resolution:
across the patient plus respiratory 0.1 cmH2O/L/s for values
system (patient + ET tube + expira- < 10 cmH2O/L/s; 1 cmH2O/L/s for
tory limb of the VBS)/expiratory values ≥ 10 cmH2O/L/s
flow. This pressure loss is divided
by the expiratory flow estimated at
the same moment, yielding the
estimate for RTOT.The complete
operation is orchestrated and
monitored by a software algo-
rithm. When PAV+ is selected, the
ventilator displays the current fil-
tered value for total resistance, and
updates the display at the success-
ful completion of each calculation.
RTOT can be displayed in the vital
patient data banner. Reference
Vital Patient Data, p. 3-39.
PAV-based work of breath- The estimated effort needed for Range: 1.0 J/L to10.0 J/L
ing (WOBTOT) patient inspiration including both Resolution: 0.1 J/L
patient and ventilator.
Peak expiratory flow (PEF) The maximum speed of exhala- Range:0 to 150 L/min
tion. Resolution:
0.1 L/min for PEF < 20 L/min;
1 L/min for PEF ≥ 20 L/min
Peak circuit pressure (PPEAK) The maximum pressure during the Range:
previous breath, relative to the -20.0 cmH2O to 130 cmH2O
patient wye, including inspiratory Resolution:
and expiratory phases. 0.1 cmH2O for values
-20.0 to 9.9 cmH2O;
1.0 cmH2O for values
10 cmH2O to 130 cmH2O
Peak spontaneous flow (PSF) The maximum flow rate sampled Range:0 to 200 L/min
during a spontaneous inspiration. Resolution:
0.1 L/min for values
< 20 L/min;
1L/min for values ≥ 20 L/min
Proximal exhaled tidal For neonatal patients, the exhaled Range: 0 mL to 500 mL
volume (VTEY) volume of the previous breath Resolution: 0.1mL for values0
measured by the Proximal Flow mL to 9.9 mL; 1 mL for values
Sensor) (if installed). 10 mL to 500 mL
Proximal exhaled total For neonatal patients, the exhaled Range: 0.00 to 99.9 L/min
minute volume (VE TOTY) minute volume measured by the Resolution: 0.01 L/min for 0.00 to 9.99
Proximal Flow Sensor). L/min;
0.1 L/min for 10.0 to 99.9 L/min
Proximal inspired tidal For neonatal patients, the exhaled Range: 0 mL to 500 mL
volume (VTIY) volume of the previous breath Resolution: 1 mL
measured by the Proximal Flow
Sensor) (if installed).
Spontaneous rapid shallow A calculated value using exhaled Range: 0.1 1/min-L to 600 1/min-L
breathing index (f/VT) spontaneous tidal volume. High Resolution: 0.1 1/min-L for values < 10
values indicate the patient is 1/min-L; 1 1/min-L; for values ≥ 10 1/
breathing rapidly, but with little min-L
volume/breath. Low values indi-
cate the inverse scenario.
Total respiratory rate (fTOT) The number of mandatory or Range: 1 to 200 1/min
spontaneous breaths/min deliv- Resolution: 0.1 1/min for values < 10
ered to the patient. 1/min; 1 1/min for 10 1/min to 200 1/
min
Vital capacity (VC) The maximum amount of air that Range: 0 mL to 6000 mL
can be exhaled after a maximum Resolution:
inhalation. 0.1 mL for values < 10 mL;
1 mL for values ≥ 10 mL
Tidal volume (VT) For adult and pediatric circuit type For adult and pediatric circuit type
settings: settings:
For TI < 600ms: ± (10 + 10% of 25 mL to 2500 mL
setting x 600 ms/TI ms) mL For neonatal circuit type settings:
For TI ≥ 600 ms 2 mL to 310 mL
± (10 + 10% of setting) mL
For neonatal circuit type settings:
For setting of 2 mL (VC+ only):
± (1 + 10% of setting) mL
For setting of 3 mL to 4 mL: ± (2 +
10% of setting) mL (delivered
volume shall be ≥ 1 mL
For setting of 5 mL to 20 mL ± (3 +
15% of setting)
For setting of ≥ 20 mL: ± (4+10% of
setting) mL
Inspired tidal volume during Leak For adult and pediatric circuit type For adult and pediatric circuit type
Sync settings: settings:
For TI ≤ 600ms: ± (10 + 20% x 600 25 mL to 2500 mL
ms/TI ms of reading) mL For neonatal circuit type settings:
For TI > 600 ms: (10 + 20% of read- 2 mL to 310 mL
ing) mL
For neonatal circuit type setting: ±
(10 + 20% of reading) mL
For readings < 100 mL, the accura-
cy shall apply when the percent-
age of inspiratory leak volume is
less than 80%
Exhaled tidal volume (VTE) during For adult and pediatric circuit type For adult and pediatric circuit type
Leak Sync settings: settings:
For TE ≤ 600 ms: ± (10 + 20% x 600 25 mL to 2500 mL
ms/TEms of reading) mL For neonatal circuit type settings:
For TE > 600 ms: ± (10 + 20% of 2 mL to 310 mL
reading) mL
For neonatal circuit type settings: ±
(10+20% of reading) mL
For readings < 100 mL, the accura-
cy shall apply when the percent-
age of inspiratory leak volume is
less than 80%
PAV-based lung compliance (CPAV) ± (1+20% of measured value) mL/ 10 to 100 mL/cmH2O
cmH2O
PAV based total airway resistance ± (3 + 20% of measured) cmH2O/ 5.0 to 50 cmH2O/L/s
(RTOT) L/s
PAV based work of breathing ± (0.5 + 10% of measured work) J/ 0.7 J/L to 4 J/L
(WOBTOT) L with a percent support setting of
75%
WARNING:
The ventilator accuracies listed in this chapter are applicable under the operating
conditions identified in the table, Environmental Specifications on page 11-8.
Operation outside specified ranges cannot guarantee the accuracies listed in the
tables above, and may supply incorrect information.
• CSA C22.2 No. 60601-1:2008 Medical Electrical Equipment, Part 1: General Requirements
for Basic safety and essential performance
• IEC 60601-1-8:2006, Medical electrical equipment - Part 1-8: General requirements for
basic safety and essential performance
• IEC 60601-1-1:2000, Medical electrical equipment -- Part 1-1: General requirements for
safety
• IEC 60601-2-12:2001, Medical electrical equipment Part 1-2: General requirements for
basic safety and essential performance
• ISO 15223-1:2012, Medical devices - Symbols to be used with medical device labels,
labeling and information to be supplied - Part 2: symbol development, selection and
validation
• ISO 7000:2004, Graphical symbols for use on equipment- Registered symbols - Fourth
edition
• ISO 80601-2-55:2011 and EN ISO 80601-2-55: 2012, Medical electrical equipment - Part
2-55: Particular requirements for the basic safety and essential performance of respira-
tory gas monitors - First Edition
• IEC 60068-2-31:2008, Environmental testing - Part 2-31: Tests - Test Ec: Rough handling
shocks, primarily for equipment-type specimens - Edition 2.0
• EN 60068-2-31:2009, Environmental testing - Part 2-31: Tests - Test Ec: Rough handling
shocks, primarily for equipment-type specimens - Edition 2.0
• ISO 3744:2010, Acoustics - Determination of sound power levels and sound energy
levels of noise sources using sound pressure - Engineering methods for an essentially
free field over a reflecting plane - Third Edition
• IEC 60601-1:1988, Medical Electrical Equipment, Part 1: General Requirements for Safety
• IEC 60601-1-4:2000, Medical Electrical Equipment - Part 1-4: General Requirements for
Safety - Collateral Standard: Programmable Electrical Medical Systems
• IEC 60601-1-6:2010, Medical electrical equipment - Part 1-6: General requirements for
basic safety and essential performance - Collateral Standard: Usability
• EU 2002/96/EC, Directive of the European Parliament and of the Council on waste elec-
trical and electronic equipment (WEEE)
WARNING:
Portable and mobile RF communications equipment can affect the performance of
the ventilator system. Install and use this device according to the information
contained in this manual.
WARNING:
The ventilator system should not be used adjacent to or stacked with other
equipment, except as may be specified elsewhere in this manual. If adjacent or
stacked used is necessary, the ventilator system should be observed to verify normal
operation in the configurations in which it will be used.
Caution:
This equipment is not intended for use in residential environments and may not
provide adequate protection to radio communication services in such environments.
The ventilator is intended for use in the electromagnetic environment specified below. The customer of the oper-
ator of the ventilator should assure that it is used in such an environment.
Radiated RF emissions Group 1 The ventilator uses RF energy only for its internal
CISPR 11 Class A functions.
The ventilator is intended to be used only in hos-
pitals and not be connected to the public mains
network.
Harmonic emissions IEC 61000-3-2 Class A The ventilator is intended to be used only in hos-
pitals and not be connected to the public mains
Voltage fluctuations/flicker Complies network.
IEC 61000-3-3
The ventilator is intended for use in the electromagnetic environment specified below. The customer of
the operator of the ventilator should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic envi-
ronment – guidance
Electrical fast transient/ ± 2 kV for power supply ± 2 kV for power supply Mains power quality
burst IEC 61000-4-4 lines lines should be that of a
± 1 kV for input/output ± 1 kV for input/output typical hospital environ-
lines lines ment.
The ventilator is intended for use in the electromagnetic environment specified below. The customer of
the operator of the ventilator should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic envi-
ronment – guidance
d = 3.5 P
10 Vrms in ISM bands1 1 Vrms in ISM bands1 Recommended separa-
tion distance
d = 12 P
The ventilator is intended for use in the electromagnetic environment specified below. The customer of
the operator of the ventilator should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic envi-
ronment – guidance
d = 2.3 P
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter man-
ufacturer and d is the separation distance in meters (m)2. Field strengths from fixed transmitters, as determined
by an electromagnetic site survey3, should be less than the compliance level in each frequency range4. Interfer-
ence may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 these guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
1. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 to 6,795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz; and 40.66 MHz to 40.70 MHz. The compliance levels in the ISM frequency bands between 150 kHz
and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood mobile/portable communications equip-
ment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calcu-
lating the separation distance for transmitters in these frequency ranges
2. The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended
to decrease the likelihood mobile/portable communications equipment could cause interference if it is inadvertently brought into patient
areas. For this reason, an additional factor of 10/3 is used in calculating the separation distance for transmitters in these frequency ranges.
3. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
980 Series Ventilator is used exceeds the applicable RF compliance level above, the 980 Series Ventilator should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the ventilator.
4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
The ventilator is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the operator of the ventilator can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
ventilator as recommended below, according to the maximum output power of the communications equipment.
Rated 150 kHz to 80 MHz 150 kHz to 80 MHz 80MHz to 800 MHz 800 MHz to 2.5 GHz
maximum outside of ISM bands inside of ISM bands
output power
of transmitter d = 1.2 P d = 2.3 P
(W) d = 3.5 P
d = 12 P
10 11 38 3.8 7.3
100 35 120 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec-
tion from structures, objects and people.
WARNING:
The use of accessories and cables other than those specified with the exception of
parts sold by Covidien as replacements for internal components, may result in
increased emissions or decreased immunity of the ventilator system.
A.1 Overview
This appendix describes the operation of the BiLevel 2.0 ventilation mode on the
Puritan Bennett™ 980 Series Ventilator.
BiLevel is a mixed mode of ventilation that combines attributes of mandatory and
spontaneous breathing, with the breath timing settings determining which breath
type is favored. In BiLevel Mode, mandatory breaths are always pressure-controlled,
and spontaneous breaths can be pressure-supported (PS) or tube compensated
(TC).
1 PCIRC (cmH2O) 4 PH
2 TH 5 PL
3 TL 6 Spontaneous breaths
BiLevel resembles SIMV mode, except that BiLevel establishes two levels of positive
airway pressure. Cycling between the two levels can be triggered by BiLevel timing
settings or by patient effort.
A-1
BiLevel 2.0
2 PL 6 Pressure support
3 PH 7 Time-based transitions
4 Spontaneous breath
The two pressure levels are called Low Pressure (PL) and High Pressure (PH). At either
pressure level, patients can breathe spontaneously, and spontaneous breaths can
be assisted with tube compensation or pressure support. BiLevel monitors manda-
tory and spontaneous tidal volumes separately.
Inspiratory time and expiratory time in BiLevel become Time high (TH) and Time low
(TL), respectively. During these inspiratory and expiratory times, PH is maintained
during TH and PL is maintained during TL.
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient's
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
2. Touch BiLevel. After selecting BiLevel mode, the ventilator uses the PC mandatory
breath type, which cannot be changed.
5. Select desired ventilator settings. The default settings for BiLevel mode appear. To
change a setting, touch its button and turn the knob to set its value. PH must always be
at least 5 cmH2O greater than PL.
6. Set TL, TH, or the ratio of TH to TL. To select settings that would result in a TH:TL ratio
greater than 1:1 or 4:1, you must touch Continue to confirm after reaching the 1:1 and
4:1 limits.
7. Touch Start.
8. Set apnea and alarm settings by touching their respective tabs at the side of the venti-
lator settings screen and changing settings appropriately.
Note:
The rise time % setting determines the rise time to reach target pressure for transitions from
PL to PH and for spontaneous breaths, even when pressure support (PSUPP) = 0. Expiratory
sensitivity (ESENS applies to all spontaneous breaths.
• Spontaneous patient efforts at PH are not pressure supported unless PSUPP > (PH - PL).
All spontaneous breaths (whether or not they are pressure supported) are assisted by a
pressure of 1.5 cmH2O.
• If PSUPP + PL is greater than PH + 1.5 cmH2O, all spontaneous breaths at PL are assisted
by the PSUPP setting, and all spontaneous breaths at PH are assisted by PSUPP - (PH - PL).
• All spontaneous breaths not supported by PS or TC (for example, a classic CPAP breath)
are assisted with an inspiratory pressure of 1.5 cmH2O.
1 Pressure (y-axis) 4 PH
During spontaneous breaths, the pressure target is calculated with respect to PL.
To avoid breath stacking, the ventilator does not cycle from one pressure level to
another during the earliest stage of exhalation.
A.8 Specifications
Reference the table, Ventilator Settings Range and Resolution, in Chapter 11 of this
manual for the following specifications:
• Low pressure (PL)
• TH:TL ratio
• Rise time %
variables TLand TH. Transitions between the two pressure levels, PL and PH, are anal-
ogous to breath phase transitions in PC.
At the extreme ranges of TL and TH, BiLevel can resemble the single breath type
mode A/C - PC, or the more complex breath type mode, an “inverted-like” IMV. If TH
and TL assume “normal” values with respect to PBW (for example TH:TL > 1:2 or 1:3),
then BiLevel assumes a breathing pattern similar to, if not qualitatively identical to
A/C - PC. However, as TL begins to shorten with the TH:TL ratio extending beyond 4:1,
the breathing pattern assumes a distinctly different shape. In the extreme, the exag-
gerated time at PH and abrupt release to PL would match the pattern patented by
John Downs* and defined as APRV.
In between the A/C-PC-like pattern and the APRV-like pattern, there would be pat-
terns with moderately long TH and TL intervals, allowing the patient sufficient time
to breathe spontaneously at both PH and PL In these types of breathing patterns,
(but less so with APRV) BiLevel, like SIMV, can be thought of as providing both man-
datory and spontaneous breath types. In this sense, BiLevel and SIMV are classified
as mixed modes.
Direct access to any of the three breath timing parameters in BiLevel is accom-
plished by touching the Padlock icon associated with the TH period, TL period or the
TH:TL ratio displayed on the breath timing bar in the setup screen.
While in BiLevel mode, spontaneously triggered breaths at either pressure level can
be augmented with higher inspiratory pressures using Pressure Support (PS) or
Tube Compensation (TC) breath types.
*. Downs, JB, Stock MC. Airway pressure release ventilation: A new concept in ventilatory support. Crit Care Med 1987;15:459-461
1 Pressure (y-axis) 5 TL
2 TH 6 Synchronous interval
3 PH 7 Spontaneous interval
4 PL
If the patient breathes spontaneously at either pressure level, BiLevel monitors and
displays the total respiratory rate, including mandatory and spontaneous breaths.
BiLevel also displays the exhaled tidal volume and total exhaled minute volume for
both mandatory and spontaneous breaths.
Lengthening the TH period and shortening the TL period to only allow incomplete
exhalation of the mandatory breath volume, results in an inverse TH:TL ratio. In this
breath timing configuration with TH:TL ratios of greater than 4:1, BiLevel becomes
Airway Pressure Release Ventilation (APRV).
APRV is characterized as longer TH periods, short TL periods (usually less than one
second), and inverse TH:TL ratios. Since, at these breath timing settings, all of the
patient-triggered spontaneous breaths occur during the TH period, APRV resembles
CPAP ventilation with occasional, short periods of incomplete exhalation referred to
as “releases“ which are controlled by the f setting.
mandatory breaths to help manage CO2 and alveolar ventilation, and the f setting
controls the number of releases per minute which are used to help manage the
patient's CO2 levels. The f setting also impacts the mean airway pressure.
In APRV the operator can configure the BiLevel settings to allow direct control of TL
to assure that changes in the f setting will not inadvertently lengthen the TL period
resulting in destabilization of end-expiratory alveolar volume. With the TL period
locked, changes in set f will change the TH period to accommodate the new f setting
while maintaining the set TL period.
In BiLevel, the ventilator establishes two levels of baseline pressure. One level is
essentially the same as the standard PEEP level set for all common modes of venti-
lation. The second pressure level is the level established at TH. Both pressure levels
permit CPAP, TC and PS breaths. The breath timing settings determine whether the
patient can initiate any of these breath types.
B.1 Overview
This appendix describes the operation of the Puritan Bennett™ 980 Series Ventilator
Leak Sync option. The Leak Sync option enables the ventilator to compensate for
leaks in the breathing circuit while accurately detecting the patient’s effort to trigger
and cycle a breath. Because Leak Sync allows the ventilator to differentiate between
flow due to leaks and flow due to patient respiratory effort, it provides dynamic com-
pensation and enhances patient-ventilator synchrony. Reference Chapter 4 in this
manual for general parameter and operational information.
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
B-1
Leak Sync
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient's
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
WARNING:
With significant leaks, pressure targets may not be reached due to flow limitations.
Note:
The default value for Leak Sync is Disabled when the circuit type is Pediatric or Adult and
the Vent Type is Invasive. Otherwise the default value for Leak Sync is Enabled.
Note:
Leak Sync is not allowed for tube compensated (TC) and Proportional Assist Ventilation
(PAV+) breath types.
• If the ventilator detects a leak during a respiratory mechanics maneuver, the message
Leak Detected is displayed.
• A new leak or change in leak rate is typically quantified and compensated within three
breaths. Monitored patient data stabilizes within a few breaths.
• Select inspiratory sensitivity settings as usual. if the ventilator auto-triggers, try increas-
ing flow sensitivity (VSENS).
Note:
The absence of the Leak Detected message does not mean there is no leak.
Note:
Leak Sync is automatically enabled when Vent Type is NIV or if New patient is selected and
circuit type is neonatal, regardless of the Vent Type. If Leak Sync is disabled while the Vent
Type is invasive but the Vent Type is changed to NIV, it remains disabled. Leak Sync becomes
disabled when Vent Type is set to INVASIVE and circuit type is Adult or Pediatric.
1 LS appears on Vent Setup button notifying the operator that Leak Sync is enabled
When Leak Sync is enabled, the Circuit Disconnect alarm becomes active based on
the DSENS setting, which is the maximum allowable leak rate at set PEEP.
WARNING:
When Vent Type = NIV and Leak Sync is disabled, DSENS is automatically set to OFF.
Reference the table below for a summary of DSENS settings when Leak Sync is
enabled. Note that it is possible to set DSENS below maximum Leak Sync flow.
WARNING:
Setting DSENS higher than necessary may prevent timely detection of inadvertent
extubation.
When Leak Sync is enabled, three additional parameters are displayed on the More
Patient Data screen and updated for each breath. Display the More Patient Data
screen by swiping the tab on the patient data banner. These leak parameters may
also be configured on the patient data banner and the large font patient data panel.
Reference the table Patient Data Range and Resolution in Chapter 11 of this manual
for information regarding the following monitored patient data parameters:
• VLEAK
• % LEAK
• LEAK
Displayed values for Exhaled Tidal Volume (VTE) and Inspired Tidal Volume (VTL) are
leak-compensated, and indicate the estimated inspired or exhaled lung volume. The
accuracies for VTE and VTL also change when Leak Sync is enabled (see Technical Dis-
cussion for more information). Graphic displays of flow during Leak Sync indicate
estimated lung flows.
Reference Patient Data Range and Resolution, p. 11-20, VTL parameter, for VTL accura-
cy.
For readings < 100 mL, accuracy ranges apply when the percentage of inspiratory
leak volume is < 80%, where the percentage of leak volume is:
(Leak volume during inspiration / total delivered inspiratory volume) x 100
Note:
Inspired tidal volume is labeled as VTL when Leak Sync is enabled, and as VTI when Leak Sync
is disabled.
Reference Patient Data Range and Resolution, p. 11-20, VTE parameter, for accuracy
when Leak Sync is enabled.
where TE = time to exhale 90% of volume actually exhaled by the patient.
For readings < 100 mL, accuracy ranges apply when the percentage of inspiratory
leak volume is < 80%, where the percentage of leak volume is:
(Leak volume during inspiration/total delivered inspiratory volume) x 100
Reference Patient Data Range and Resolution, p. 11-20, % LEAK parameter, for speci-
fications.
The Circuit Disconnect alarm is activated if the overall leak volume during the whole
breath exceeds the maximum leak volume derived from the DSENS setting. During
VC, the Circuit Disconnect alarm is also activated if the end-inspiratory pressure falls
below (set PEEP + 1 cmH2O) for three consecutive breaths. The screen shows this
alarm message:
If the compressor is in use and the DSENS setting > 25 L/min, a DSENS of
25 L/min is used to determine Circuit Disconnect. If LEAK > 25 L/min, the alarm
banner shows the following message:
Check patient. Reconnect circuit. Leak may exceed maximum compensation value for
compressor.
Normal operation resumes if the ventilator detects a patient connection.
C.1 Overview
This appendix describes the operation of the PAV™*+ software option for the Puritan
Bennett™ 980 Ventilator.
Proportional Assist™* Ventilation (PAV+) is designed to improve the work of breath-
ing of a spontaneously breathing patient by reducing the patient’s increased work
of breathing when pulmonary mechanics are compromised.
The PAV+ breath type differs from the pressure support (PS) breath type in the fol-
lowing way:
PAV+ acts as an inspiratory amplifier; the degree of amplification is set by the %
Support setting (% Supp). PAV+ software continuously monitors the patient’s
instantaneous inspiratory flow and instantaneous lung volume, which are indicators
of the patient’s inspiratory effort. These signals, together with ongoing estimates of
the patient’s resistance and compliance, allow the software to instantaneously
compute the necessary pressure at the patient wye to assist the patient’s inspiratory
muscles to the degree selected by the % Supp setting. Higher inspiratory demand
yields greater support from the ventilator.
PAV+ software reduces the risk of inadvertent entry of incompatible settings, such
as small predicted body weight (PBW) paired with a large airway.
*. Proportional Assist and PAV are registered trademarks of The University of Manitoba, Canada. Used under license.
C-1
PAV™+
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient's
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
WARNING:
PAV+ is not an available breath type in non-invasive ventilation (NIV). Do not use
non-invasive patient interfaces such as masks, nasal prongs, uncuffed ET tubes, etc.
as leaks associated with these interfaces may result in over-assist and patient
discomfort.
WARNING:
Breathing circuit and artificial airway must be free from leaks. Leaks may result in
ventilator over-assist and patient discomfort.
WARNING:
Ensure high and low tidal volume alarm thresholds are set appropriately because an
overestimation of lung compliance could result in an under-support condition
resulting in the delivery of smaller than optimal tidal volumes.
C.4 PAV+
WARNING:
Ensure that there are no significant leaks in the breathing circuit or around the
artificial airway cuff. Significant leaks can affect the performance of the PAV+ option
and the accuracy of resistance (R) and elastance (E) estimates.
WARNING:
Do not use silicone breathing circuits with the PAV+ option: the elastic behavior of a
silicone circuit at the beginning of exhalation can cause pressure-flow oscillations
that result in underestimates of patient resistance.
The act of inspiration requires the patient’s inspiratory muscles to develop a pressure
gradient between the mouth and the alveoli sufficient to draw in breathing gas and
inflate the lungs. Some of this pressure gradient is dissipated as gas travels through
the artificial airway and the patient’s conducting airways, and some of the pressure
gradient is dissipated in the inflation of the lungs and thorax. Each element of pres-
sure dissipation is characterized by a measurable property: the resistance of the arti-
ficial and patient airways, and the compliance (or elastance) of the lung and thorax.
PAV+ software uses specific information, including resistance of the artificial airway,
resistance of the patient’s airways, lung-thorax compliance, instantaneous inspirato-
ry flow and lung volume, and the % Supp setting to compute the instantaneous
pressure to be applied at the patient connection port (patient wye). PAV+ software
randomly estimates patient resistance and compliance approximately every four to
ten breaths. Every five (5) ms, the software estimates lung flow, based on an estimate
of flow at the patient wye, and lung volume, based on the integral of the value of
estimated lung flow.
PAV+ begins to assist an inspiration when flow (generated by the patient’s inspira-
tory muscles) appears at the patient wye. If the patient ceases inspiration, the assist
also ceases. Once inspiratory flow begins, PAV+ software monitors instantaneous
flow and volume every 5 ms and applies the pressure calculated to overcome a pro-
portion (determined by the % Supp setting) of the pressure losses dissipated across
the resistances of the artificial and patient airways and lung/thorax compliance.
Because the PAV+ algorithm does not know the patient’s mechanics when the
PAV+ breath type is selected, the software performs a startup routine to obtain initial
data. At startup, PAV+ software delivers four consecutive PAV+ breaths, each of
which includes an end-inspiratory pause maneuver that yields estimates of the
patient’s resistance and compliance. The first breath, however, is delivered using the
predicted resistance for the artificial airway and conservative estimates for patient
resistance and compliance, based on the patient’s PBW.
Each of the next three PAV+ breaths averages stepwise decreased physiologic
values with the estimated resistance and compliance values from the previous
breath, weighting earlier estimates less with each successive breath, and yielding
more reliable estimates for resistance and compliance. The fifth PAV+ breath (the
first non-startup breath) is delivered using the final estimates with the clinician-set
% Supp setting. Once startup is complete, the PAV+ software randomly applies a
maneuver breath every four to ten breaths after the last maneuver breath to re-esti-
mate patient resistance and compliance. New values are always averaged with
former values.
The PAV+ option graphically displays estimates of patient lung pressure (intrinsic
PEEP), patient compliance, patient resistance, total resistance, total work of inspira-
tion, patient work of inspiration, inspiratory elastic work (an indicator of lung-thorax
work), and inspiratory resistive work.
The % Supp setting ranges from a minimum of 5% (the ventilator performs 5% of the
work of inspiration and the patient performs 95%) to a maximum of 95% (the venti-
lator performs 95% of the work and the patient performs 5%), adjustable in 5% incre-
ments.
The PAV+ option also includes alarm limits, safety checks, and logic checks that
reject non-physiologic values for patient resistance and compliance as well as inap-
propriate data.
Humidification type and volume can be adjusted after running SST, however the
ventilator makes assumptions when calculating resistance and compliance if these
changes are made without re-running SST. For optimal breath delivery, run SST after
changing humidification type and humidifier volume.
To set up PAV+
1. At the ventilator setup screen, enter the patient’s gender and height or the patient’s
PBW.
7. Select the tube ID. Initially, a default value is shown based on the PBW entered at venti-
lator startup. If this ID is not correct for the airway in use, turn the knob to adjust the ID
setting.
Note:
If the operator selects an internal diameter that does not correspond to the PBW/tube ID
range pairs listed in the following table, touch Continue to override the tube ID setting. If
attempts are made to choose a tube ID less than 6.0 mm or greater than 10 mm, a hard
bound limit is reached, as PAV+ is not intended for use with tubes smaller than 6.0 mm or
larger than 10.0 mm. When touching Dismiss, the setting remains at the last tube ID
selected. Touch Accept or Accept ALL to accept changes, or touch Cancel to cancel changes.
Note:
If Leak Sync is currently enabled, it becomes disabled when PAV+ is selected.
Note:
When the ventilator is used on the same patient previously ventilated using PAV+, the GUI
displays an attention icon and the tube type and tube ID previously used, as a reminder to
the clinician to review those settings during ventilator setup.
The ventilator uses “soft bound” and “hard bound” values for estimated tube inside
diameters based upon PBW. Soft bounds are ventilator settings that have reached
their recommended high or low limits. When adjusting the tube size, if the inside
diameter does not align with a valid predicted body weight, a Continue button
appears. Setting the ventilator beyond these soft bounds requires the operator to
acknowledge the prompt by touching Continue before continuing to adjust the
tube size. The limit beyond which the tube ID cannot be adjusted is called a hard
bound, and the ventilator emits an invalid entry tone when a hard bound is reached.
WARNING:
Ensure that the correct artificial airway ID size is entered. Because PAV+ amplifies
flow, entering a smaller-than-actual airway ID causes the flow-based pressure
assistance to over-support the patient and could lead to transient over-assist at high
values of % Supp. Conversely, entering a larger-than-actual ID results in under-
support. PAV+ software monitors the settings for the PBW and artificial airway. If the
PBW and tube ID settings do not correspond to the above PBW/tube ID range pairs,
confirm or correct the settings. Confirming or correcting the actual ID size minimizes
the likelihood that PAV+ will over-support or under-support.
To apply new settings for the artificial airway follow these steps
1. Touch Vent Setup at the lower left of the GUI screen.
2. Touch Tube Type and turn the knob to select Trach or ET to set the tube type.
3. Touch tube ID and turn the knob to set the tube ID.
4. Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel.
3. For non-HME humidification types, touch Humidifier Volume, then turn the knob to
adjust the (empty) humidifier volume.
WARNING:
To ensure the accuracy of PAV+ breaths and spirometry measurements, run SST
following any change to the humidification type or humidification volume settings.
Ensure that the intended circuit is used with the SST.
After accepting the PAV+ settings, touch the Apnea Setup screen. Adjust the Apnea
parameters as required.
PAV+ includes the high inspired tidal volume 2VTI) and low exhaled spontaneous
tidal volume alarm (4VTE SPONT) alarm limit settings. Reference PAV+ Alarms, p. C-9.
Note:
Because of the breathing variability that PAV+ allows, the 4VTE SPONT alarm, by default, is
turned OFF to minimize nuisance alarms. To monitor adequate ventilation, use the 3VE TOT
alarm condition instead.
3. Turn the knob to adjust the value of the alarm limit. Proposed values are highlighted.
You can change more than one alarm limit before applying the changes.
Reference the table Ventilator Settings Range and Resolution, in Chapter 11 of this
manual for a summary of PAV+ ventilator settings for the following parameters:
• % Supp
• Tube type
• Tube ID
• Trigger type
Reference the table Alarm Settings Range and Resolution, in Chapter 11 of this manual
for a summary of the following alarm settings available when PAV+ is active:
• High inspired tidal volume limit (2VTI)
Reference Patient Data Range and Resolution in Chapter 11 of this manual for the fol-
lowing monitored data associated with PAV+:
• PAV-based lung compliance (CPAV)
25 0 to 50 2.5 to 29 34 to 400
35 0 to 44 3.5 to 41 24 to 286
45 0 to 31 4.5 to 52 19 to 222
65 0 to 20 6.4 to 75 13 to 156
75 0 to 18 7.4 to 87 11 to 135
85 0 to 17 8.4 to 98 10 to 119
• 1VTI
WARNING:
For optimal performance of PAV+, it is important to select the humidification type,
tube type, and tube size that match those in use on the patient.
The instantaneous pressure generated at the patient wye during inspiration is a
function of the patient effort, % Supp setting, tube type and size, patient resistance
and elastance, and the instantaneously measured gas flow and lung volume. Set
2PPEAK to a safe circuit pressure, above which truncation and alarm annunciation are
appropriate.
Note:
PAV+ has a built-in high pressure compensation (1PCOMP) limit that is determined by
the2PPEAK setting minus 5 cmH2O or 35 cmH2O, whichever is less. If the inspiratory pressure
at the patient wye (PIwye) reaches the 1PCOMP limit, the inspiration is truncated, and the
ventilator transitions to exhalation. Reference p.C-18 for more details regarding 1PCOMP and
1PPEAK.
k P V· dt
i i
W = -----------------------------------
· i dt
V
When PAV+ is active (the mode is SPONT and the spontaneous breath type is PAV+),
a work of breathing (WOB) graphic is automatically displayed (Reference Graphics
displays in PAV+, p. C-12) which shows:
• an indicator showing the proportion of patient inspiratory work to overcome the elas-
tance (E) of the lung-thorax and the combined resistance (R) of the artificial airway and
the patient.
– the estimated total work of breathing (in Joules/L) of the patient and ventilator
during inspiration (WOBTOT)
• PAV-based patient data estimates, including patient resistance (RPAV), lung compliance
(CPAV), and intrinsic PEEP (PEEPI PAV).
Note:
Graphic displays of lung pressure and patient work of breathing are not actual
measurements, and are derived from equations using filtered estimates of pressure and
flow.
The WOB graphic is only available when SPONT mode and the PAV+ breath type are
selected. The shadow trace can be enabled or disabled when selecting the graphic
display, or after a display is paused.
The act of pausing does not affect the WOB graphic, but does store the shadow
trace. Once paused, the operator can enable or disable the shadow trace, then view
the paused waveform again with or without the shadow trace.
The following table provides a definition and description of each of the Work of
Breathing (WOB) terms.
WOBTOT Total Work of Inspiration With the PAV+ breath type active, the patient and the venti-
lator always share the in the work of breathing. The percent
WOBTOT performed by the ventilator always equals the %
Supp setting and the percent WOBTOT performed by the
patient always equals (100 minus the % Supp setting).
WOBTOT is the sum of the work to move the breathing gas
through the artificial airway and the patient's own airways
plus the work to inflate the patient's elastic lung-thorax.
WOBPT Patient Work of Breathing That part ofWOBTOT performed by the patient.
WOBPT ELASTIC Inspiratory Elastic Work That part of WOBPT attributed to inflating the patient’s elastic
lung-thorax.
WOBPT RESISTIVE Inspiratory Resistive Work That part of the WOBPT attributed to moving breathing gas
through resistive elements in the gas path.
EQUATION 1
PMUS Pressure generating capability of R Resistance elements (artificial plus patient airways)
patient’s inspiratory muscles
VL Flow through the resistance ele- ELUNG-THORAX Elastance of the lung and thorax (1/CLUNG-THORAX)
ments and into the lungs
If the PAV+ software estimates of patient resistance and elastance (RPAV and (EPAV)
remain stable, this equation could be rewritten as:
EQUATION 2
i i i i i
P MUS = V L R airway + V L K 1 + V L K2
K2 EPAV
PiMUS could then be estimated at every control period if ViL, Riairway, and ViL were also
known.
Throughout any inspiration, the individual pressure elements that make up PMUS
can be expressed as:
EQUATION 3
PMUS Pressure generating capability of PFLOWPATIENT Flow based pressure drop across the
patient’s inspiratory muscles patient
PFLOWARTIFICIAL Flow based pressure drop across the PVOLUMEPATIENT Volume based pressure to overcome the
artificial airway lung-thorax elastance
AIRWAY
Equations 2 and 3 provide the structure to explain how PAV+ operates. The clinician
enters the type and size of artificial airway in use, and the software uses this informa-
tion to estimate the resistance of the artificial airway at any lung flow.
Applying a special pause maneuver at the end of selected inspirations provides the
information the software needs to estimate patient resistance (RPAV) and compli-
ance (CPAV, which is converted to elastance, EPAV). Immediately following the end of
the pause event, software captures simultaneous values for PLUNG, Pwye, and VE
which yield an estimate for RTOT at the estimated flow.
All raw data are subjected to logic checks, and the estimates of RPAV and CPAV are
further subjected to physiologic checks. The estimates of RPAV and CPAV are discard-
ed if any of the logic or physiologic checks fail. If CPAV is rejected, RPAV is also rejected.
Valid estimates of RPAV and CPAV are required for breath delivery, and are constantly
updated by averaging new values with previous values. This averaging process
smooths data and avoids abrupt changes to breath delivery. If new values for RPAV
and CPAV are rejected, the previous values remain active until valid new values are
obtained. PAV+ software monitors the update process and generates an escalating
alarm condition if the old values do not refresh.
During PAV+, maneuver breaths are randomly performed every four to ten breaths
after the last maneuver breath. A maneuver breath is a normal PAV+ inspiration with
a pause at end inspiration. Because muscle activity is delayed for at least 300 ms fol-
lowing the end of neural inspiration, the patient’s respiratory control center does not
detect the pause. With this approach, maneuver breaths are delivered randomly so
that their occurrence is neither consciously recognized nor predictable.
A PAV+ breath begins, after the recognition of a trigger signal, with flow detection
at the patient wye. The sample and control cycle of the ventilator (the value of i in
Equation 2) is frequent enough to yield essentially constant tracking of patient inspi-
ration. At every ith interval, the software identifies instantaneous lung flow (ViL,
which is impeded by the resistances of the artificial airway and patient airways) and
integrates this flow to yield an estimate of instantaneous lung volume, (ViL), which is
impeded by the elastic recoil of the lung and thorax).
Using the values for instantaneous lung flow and lung volume, PAV+ software cal-
culates each of the pressure elements in Equation 2, which gives the value of PMUS
at each ith interval.
At this point, Equation 2 and the subsequent analysis identifies that an appropriate
patient, supported by PAV+ and with an active PMUS (an absolute requirement) will,
within a few breaths, enable the algorithm to obtain reasonable estimates of RPAV
and EPAV. Once these physiologic data are captured (and over a relatively brief time
they are improved and stabilized), the PAV+ algorithm mirrors the patient's respira-
tory mechanics, which then allows the ventilator to harmoniously amplify PMUS. The
key point to recognize is that patient's continuous breathing effort “drives“ the PAV+
support — no effort, no support.
The % Supp setting specifies the amount of resistance- and elastic-based pressure
to be applied at each ith interval at the patient wye.
By taking all of the above information into consideration, EQUATION 2 can be rewrit-
ten to include the % Supp setting recognizing that ViL and ViL are driven by the
patient, not by the ventilator. (It is important to note that the ventilator is not ampli-
fying its own flow — only the flow generated by PMUS.)
EQUATION 4
P wye = S V· L R airway + S V· L K 1 + S V L K 2
i i i i i
Piwye Pressure generated by the ventilator in S % Supp setting/100 (ranges from 0.05 to
response to the instantaneous values of 0.95
lung flow and lung volume at the wye. This
value is the sum of the three individual
pressure elements (in parentheses) in Equa-
tion 4
The pressure gradient driving breathing gas into the patient’s lungs is given by the
sum of Piwye and the patient’s inspiratory effort, therefore:
EQUATION 5
i i i
P GRADIENT = P wye + P MUS
PAV+ software is designed to reduce the risk that hyperinflation may occur. The
potential for hyperinflation could arise if the software were to overestimate actual
patient resistance or underestimate actual patient lung-thorax compliance (that is,
to overestimate actual elastance). If the software cannot generate valid estimates of
RPAV and CPAV, PAV+ cannot start. If, after startup, the values of RPAV and CPAV cannot
be updated with valid new values, the previous values become less reliable.
The stability of PAV+ is primarily determined by the relationship between the true
lung elastance [EL (true)] and the true lung volume [VL(true)]. Although Pi wye (resis-
tive) also plays a part, the following discussion focuses on the elastic component.
At all lung volumes, the true state of the lung and thorax is expressed by:
i i
P L recoil = V L true E L true
PiL recoil True lung recoil pressure EL (true) True lung elastance
Over-inflation will not occur as long as Pi wye (elastic) < PiL recoil, which is equivalent
to the inequality:
where:
K2 = EPAV1
As long as EPAV (estimated) = EPAV (true) and ViL (estimated) = ViL (true) then Pirecoil>
Pi wye even at high values of % Supp (i.e. between 85% and 95%).
This means that if the pressure applied to the lung-thorax is never greater than EL
(true) x VL, lung volume will collapse if wye flow vanishes. As long as EPAV (estimated)
≤ EL (true), ViL (estimated) ≤ ViL (true), and RPAV (estimated) ≤ RL(true), PMUS is the
modulator of Pi wye.
Hyperinflation could occur if the estimated EPAV were greater than the true value of
EL. At a high % Supp setting, Pi wye (elastic) could exceed PiL recoil, causing a self-gen-
erating flow at the patient wye, which in turn would cause a self-generating inflation
of the lungs. This is part of the reason that the % Supp setting is limited to 95%.
Likewise, if the estimated RPAV were to exceed the true value of RL at a high % Supp
setting, PIwye (resistive) could exceed the value necessary to compensate for pres-
sure dissipation across the artificial and patient airways, resulting in early hyperinfla-
tion of the lungs. As flow declines after the first third of inspiration, however, the
hyperinflating effect would most likely disappear.
PAV+ software includes these strategies to minimize the possibility of hyperinflation
of the lungs:
1. The maximum % Supp setting is limited to 95%.
2. The raw data for RPAV and CPAV are checked for graph/math logic, and estimated
mechanics values are checked against PBW-based physiologic boundaries. These
checks reduce the possibility of overestimating patient resistance or underestimating
patient compliance, which could lead to potential over-inflation.
3. The high inspiratory tidal volume limit (2VTI) places an absolute limit on the integral of
lung flow (including leak flow), which equals lung volume. If the value of VTI reaches this
limit, the ventilator truncates inspiration and immediately transitions to exhalation.
4. The 2VTIsetting places an upper limit on the value of the PVOLUMEPATIENT component of
Piwye (see Equations 3 and 4). At the beginning of each new inspiration, PAV+ software
calculates a value for PVOLUMEPATIENT as follows:
where P*wye is the unique value for the elastic threshold limit of Piwye that will cause
the lung volume to expand to 75% of 2VTI. When Piwye (elastic) = P*wye (elastic thresh-
old limit), the software stops increasing Piwye (elastic). This means that any further
increase in lung volume must be accomplished by the patient, which tends to hasten
the conclusion of inspiratory effort and avoid truncation due to lung volume reaching
the 2VTI limit.
5. The high inspiratory pressure limit (2PPEAK) applies to all breaths, and is used by PAV.+
software to detect the high compensation pressure condition (1PCOMP):
If the user-adjustable 2PPEAK limit is reached, the ventilator truncates inspiration and
immediately transitions to exhalation. If Piwye (the targeted wye pressure calculated in
Equation 4) equals the 1PCOMP for 500 ms, the inspiration is truncated and exhalation
begins. Further, when Piwye = 1PCOMP, Piwye is limited to 1PCOMP. Although this freezes
the value of Piwye , patient activity such as coughing could drive Piwye to 2PPEAK,
causing inspiration to end.
The rapid rise of Piwye to the 1PCOMP limit would likely occur in the first third of inspira-
tion, and only if RPAV were overestimated and % Supp were set above 85%. The 1PCOMP
condition guards against over-inflation due to overestimation of RPAV.
6. The% Supp setting ranges from 5 to 95% in 5% increments. Reducing the level of
support decreases the possibility of over-inflation. A significant decrease could produce
a sensation of inadequate support, and the patient would absorb the additional work of
inspiration or require an increase in the level of support.
A significant increase could cause a surge in the ventilator generated value for
Pwye, which in turn could cause Piwye to reach 2PCOMP and lead to temporary patient-
ventilator disharmony. To minimize this possibility, PAV+ software limits the actual
increase in support to increments of 10% every other breath until the new setting is
reached.
7. Spirometry remains active during PAV+ operation. 2VTI can be set high enough to allow
spontaneous sigh breaths, while 4VE TOT and 2VE TOT remain active to reveal changes
in minute ventilation.
Because PAV+ cannot operate without valid estimates of RPAV and CPAV, and because those
values are unknown when PAV+ starts, a startup routine obtains these values during four
maneuver breaths that include an end inspiratory pause that provides raw data for RPAV and
CPAV, and both estimated values must be valid. If either value is invalid during any of the four
startup breaths, the software schedules a substitute maneuver breath at the next breath. Ref-
erence PAV+, p. C-3.
A low-priority alarm becomes active if a 45-second interval elapses without valid esti-
mates for RPAV and CPAV. If the condition persists for 90 seconds, the alarm escalates to
medium-priority. If the condition persists for 120 seconds, the alarm escalates to high
priority. The 1VE TOT and1fTOT alarms are also associated with this condition.
Similarly, if RPAV and CPAV cannot be updated with valid values after a successful PAV+
startup, a low-priority alarm is activated if the condition persists for 15 minutes. If the
values still cannot be updated with valid values after 30 minutes, the alarm escalates to
medium priority.
8. If PAV+ estimates a high lung resistance following a sharp spike in the expiratory flow
waveform, then a PBW-based resistance value is used. Reference the waveform and
table below.
D-1
NeoMode 2.0
E.1 Overview
This appendix describes the operation of the Proximal Flow Option for the Puritan
Bennett™ 980 Series Ventilator. The Proximal Flow Option is solely used for monitor-
ing flows, pressures, and tidal volumes and does not control these parameters in any
way.
The Proximal Flow Sensor is designed to measure the lower flows, pressures and
tidal volumes at the patient wye typically associated with invasively ventilated neo-
natal patients.
For general parameter and general ventilator setup information, reference Chapter
4 in this manual.
E-1
Proximal Flow
Proximal Flow Sensor — The Puritan Bennett Proximal Flow Sensor is required for use with
the Proximal Flow Option. The sensor is installed near the patient circuit wye. The other
end of the sensor connects to the ventilator’s front panel behind a clear door designed
to protect the connection point from exposure to spills or from sprayed liquids during
cleaning and disinfection.
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
WARNING:
The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as
indicated, very limited exposure to trace amounts of phthalates may occur. There is
no clear clinical evidence that this degree of exposure increases clinical risk.
However, in order to minimize risk of phthalate exposure in children and nursing or
pregnant women, this product should only be used as directed.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the patient's
treatment, the clinician should carefully select the ventilation mode and settings to
use for that patient based on clinical judgment, the condition and needs of the
patient, and the benefits, limitations and characteristics of the breath delivery
options. As the patient's condition changes over time, periodically assess the chosen
modes and settings to determine whether or not those are best for the patient's
current needs.
WARNING:
Inspect the Proximal Flow Sensor prior to use, and do not use it if the sensor body,
tubing, or connector are damaged, occluded, or broken.
WARNING:
Do not use the Proximal Flow Sensor if there are kinks in the tubing.
WARNING:
Prior to patient ventilation with the Proximal Flow Option, run SST with the exact
configuration that will be used on the patient. This includes a neonatal patient
circuit, Proximal Flow Sensor, and all accessories used with the patient circuit. If SST
fails any Proximal Flow Sensor test, check the patient circuit and the Proximal Flow
sensor for leaks or occlusions and replace the flow sensor, if necessary. If SST
continues to fail, it may indicate a malfunction or a leak within the Proximal Flow
hardware which could compromise accuracy or increase the likelihood of cross-
contamination; thus, replace the Proximal Flow hardware.
WARNING:
Changing ventilator accessories can change the system resistance and compliance.
Do not add or remove accessories after running SST.
WARNING:
If the Proximal Flow Option fails to respond as described in this appendix,
discontinue use until correct operation is verified by qualified personnel.
WARNING:
The Proximal Flow Sensor measures gas flow at the patient wye. The actual volume
of gas delivered to the patient may be affected by system leaks between the patient
and the Proximal Flow Sensor, such as a leak that could occur from the use of an
uncuffed endotracheal tube.
WARNING:
Position the Proximal Flow Sensor exactly as described in this appendix or the
Instructions for Use (IFU) provided with the sensor.
WARNING:
Do not position the Proximal Flow Sensor cables or tubing in any manner that may
cause entanglement, strangulation or extubation which could lead to hypercarbia or
hypoxemia. Use the cable management clips supplied to mitigate this risk.
WARNING:
To reduce the risk of extubation or disconnection, do not apply tension to or rotate
the Proximal Flow Sensor by pulling on the Proximal Flow Sensor’s tubing.
WARNING:
Do not install the Proximal Flow Sensor in the patient circuit if the sensor is not also
connected to the BDU.
WARNING:
Excessive moisture in the Proximal Flow Sensor tubing may affect the accuracy of the
measurements. Periodically check the sensor and tubing for excessive moisture or
secretion build-up.
WARNING:
The Proximal Flow Sensor is intended for single use only. Do not re-use the sensor.
Attempts to clean or sterilize the sensor may result in bioincompatibility, infection,
or product failure risks to the patient.
WARNING:
Install the Proximal Flow Sensor as shown. Reference Attaching Proximal Flow Sensor,
p. E-13. Improper orientation of the flow sensor could lead to misinterpretation of
data or incorrect ventilator settings.
Caution:
Do not use aerosolized medications with the Proximal Flow Sensor. Such
medications may damage the sensor.
Caution:
To prevent damage to pneumatic lines, use supplied cable management clips.
Caution:
Use only Covidien-branded Proximal Flow Sensors with the Proximal Flow Option.
Inspired and exhaled flows and volumes at the patient wye are measured and iden-
tified by the symbols shown below, and correspond to their non-proximal flow
equivalents. These values appear in the patient data panel if so configured. Refer-
ence Vital Patient Data, p. 3-39 and the figure above.
VTE SPONTY Exhaled spontaneous tidal volume (at patient circuit wye)
VTE MANDY Exhaled mandatory tidal volume (at patient circuit wye)
VCIRC Y Flow throughout the breath cycle (at patient circuit wye)
VTLY Inspired tidal volume (at patient circuit wye with Leak Sync enabled)
Note:
In the patient data symbols shown above, the “Y” appears in inverse video, as shown.
Reference Sample GUI screen Showing Proximal Flow Data, p. E-6.
Note:
When the Proximal Flow and Leak Sync options are enabled, the following parameters are
available for display:
• VTLY and VTL
When only the Proximal Flow option is enabled, VTIY and VTI are available for display.
When a “Y” appears in the symbol, the data are measured with the proximal flow sensor.
When a “Y” is absent from the symbol, the data are measured by the ventilator’s internal flow
sensors.
lines. Autozero and purge functions are only active during exhalation which limits
the effect of the purge gas on delivered oxygen concentration.
During the autozero or automatic purge processes, the measurement and display of
proximal flow data is not shown in real time and a brief message appears on the GUI
indicating the purge process is occurring.
During autozero or automatic purge processes, the pressure waveforms, when
shown display the current PEEP value and the flow waveform, when shown, displays
a value of 0.
Note:
Failure of the Proximal Flow Option to pass SST does not prevent ventilation, but will
prevent measurement with the Proximal Flow Option. The ventilator will use its internal flow
sensors for measurement instead of the Proximal Flow Option.
SST Flow Sensor Cross Check Tests O2 and Air Flow Sensors N/A
SST Exhalation Filter Checks for exhalation filter occlu- Ventilator prompts the user to
sion and exhalation compartment block the proximal flow sensor
occlusion. outlet during Leak test. When
prompted to reconnect the
patient to the exhalation filter
during the Exhalation Filter test,
resume blocking the proximal flow
sensor outlet.
During SST the ventilator prompts to attach the Proximal Flow Sensor.
To attach the Proximal Flow Sensor to the patient circuit
1. Verify the Proximal Flow Sensor, pneumatic lines, and connector are not damaged.
2. Open the connector panel door and firmly attach the sensor connector to the recepta-
cle in the BDU’s front connector port labeled Prox.
4. When prompted to attach the Proximal Flow Sensor, unblock the circuit wye and insert
the smaller end of the sensor into the wye.
5. When prompted, cap or seal the larger end of the sensor (marked with “UP” and an
arrow).
If SST fails, check the patient circuit and flow sensor connections for leaks or occlu-
sions and replace the Proximal Flow Sensor, if necessary. Replace the Proximal Flow
Option hardware if SST continues to fail, then repeat SST to determine circuit com-
pliance and resistance. Reference the Puritan Bennett™ 980 Series Ventilator Hardware
Options Installation Instructions, p/n 10084704 for instructions on replacing the Prox-
imal Flow Option hardware.
2. Touch the Options tab. A screen appears containing the Installed Options and Prox tabs.
Note:
If the Proximal Flow Option has been disabled or enabled, SST does not have to be re-run
unless the breathing circuit or other breathing system accessories have been changed,
removed, or added.
2. Open the connector panel door and firmly attach the sensor connector to the right-
most receptacle in the BDU’s front connector port labeled Prox. Reference Attaching
Proximal Flow Sensor to Ventilator, p. E-10.
To attach the Proximal flow sensor between the endotracheal tube and patient
circuit
1. Connect the larger end of the sensor (marked with “UP” and an arrow) to the endotra-
cheal tube. Reference the figure below. Do not force the connection; when the sensor
is oriented correctly, insertion requires little effort.
Note:
If using a Heat-Moisture Exchanger (HME) on the endotracheal tube, place the Proximal
Flow Sensor between the HME and the breathing circuit wye.
2. Connect the smaller end of the sensor to the breathing circuit wye.
3. Ensure the sensor tubing is positioned in an upward direction, as shown in the figure
above. If the sensor needs repositioning, DO NOT rotate it by pulling on the tubing.
Reposition as follows:
a. Grasp the sensor’s plastic body with one hand and the breathing circuit wye with
the other hand.
b. Rotate the sensor body and wye towards each other until the sensor tubing is
upright.
c. Confirm a tight connection between the sensor and breathing circuit wye.
4. Use the three cable management clips provided with the sensor to attach the sensor
tubing to the breathing circuit tubing. Space the clips evenly along the length of the
sensor tubing. Twist the ends of each clip to close.
Note:
When the ventilator is set up for Proximal Flow Option operation, the Proximal Flow Sensor
can be switched as necessary. There is no need to run SST after switching sensors unless the
breathing circuit or other ventilator accessories have been changed.
A manual purge may be performed any time the sensor lines contain excessive con-
densation, moisture, or secretions.
To perform a manual purge:
1. Touch the Configure icon on the in the constant access icons area of the GUI.
2. Touch the Options tab. A screen appears containing the Installed Options and Prox tabs.
4. Touch Start that appears next to the text “Prox Manual Purge: To begin touch the Start
button”. During the purge, a message appears in the GUI prompt area stating the purge
process is being performed.Reference Message During Autozero and Purge Processes, p.
E-8.
E.12 Alarms
If the Proximal Flow Option becomes inoperable during ventilation, the ventilator
annunciates an alarm and flow sensing reverts to the ventilator’s internal delivery
and exhalation flow sensors. This switch over may be triggered by any of the follow-
ing events:
• The Proximal Flow Sensor is not detected
• There is a communication failure between the ventilator and the Proximal Flow option
If any of these conditions occur, the GUI displays an alarm message similar the one
shown below.Follow the information contained in the remedy message to trouble-
shoot the alarm.
Measurement Accuracy1
Parameter Specification
Weight 6.6 g
Includes:
Installation hardware and accessories
analysis message A message displayed on the GUI screen during an alarm condition, identifying
the root cause of the alarm.
assist breath A mandatory breath triggered by patient inspiratory effort in A/C and SIMV
modes.
assist-control A/C mode A ventilation mode where only mandatory VC, PC, or VC+breaths are delivered
to the patient.
audio paused (alarm Used interchangeably with the term alarm silence, the 2-minute period that
silence) begins after the audio paused (alarm silence) key is pressed, where the audible
portion of an alarm is muted.
augmented alarm The initial cause of an alarm has precipitated one or more related alarms. When
an alarm occurs, any subsequent alarm related to the cause of this initial alarm
“augments” the initial alarm.
background checks Continuously running tests during ventilation that assess the ventilator’s elec-
tronics and pneumatics hardware.
backup ventilation (BUV) A safety net feature which is invoked if a system fault in the mix subsystem, inspi-
ratory subsystem, or expiratory subsystem occurs compromising the ventilator’s
ability to ventilate the patient as set.
base flow A constant flow of gas through the patient circuit during the latter part of exha-
lation during flow triggering (VTRIG). The value of this base flow is 1.5 L/min
greater than the operator selected value for flow sensitivity.
base message A message given by the ventilator during an alarm condition, identifying the
alarm.
batch changes Changes to multiple settings that go into effect at the same time.
battery back-up system The system for supplying battery back-up power to a device. The ventilator's
battery back-up system consists of a single primary battery to provide up to one
(1) hour of battery power to the ventilator. An optional extended battery with the
same characteristics as the primary battery is available.
BD, BDU Breath delivery or breath delivery unit. The ventilator component that includes
inspiratory and expiratory pneumatics and electronics.
Glossary-1
Table Glossary-1. Glossary of Ventilation Terms (Continued)
BiLevel mode A mixed ventilation mode combining mandatory and spontaneous breaths,
where two levels of pressure are delivered (PL and PH) corresponding to expira-
tory and inspiratory times TLand TH.
BOC British Oxygen Company. A standard for high pressure gas inlet fittings.
breath stacking The delivery of a second inspiration before the first exhalation is complete.
BTPS Body temperature and pressure, saturated, 37°C, at ambient barometric pres-
sure, at 100% relative humidity.
compliance volume The volume of gas that remains in the patient circuit and does not enter the
patient's respiratory system.
compressor The compressor provides compressed air, which can be used in place of wall or
bottled air.
constant during rate One of three breath timing variables (inspiratory time, I:E ratio, or expiratory time)
change the operator can hold constant when the respiratory rate setting changes.
Applies only to the pressure control (PC) mandatory breath type (including VC+
and BiLevel).
CPU Central processing unit. The electronic components of the ventilator (BD and
GUI) responsible for interpreting and executing instructions entered by the oper-
ator.
dependent alarm An alarm that arises as a result of another primary alarm (also referred to as an
augmentation).
DSENS Disconnect sensitivity. A setting that specifies the allowable loss (percentage) of
delivered tidal volume, which if equaled or exceeded, causes the ventilator to
declare a DISCONNECT alarm. The greater the setting, the more returned volume
must be lost before DISCONNECT is detected. If the Leak Sync option is in use,
DSENSis the maximum allowable leak rate and is expressed in terms of L/min.
DISS Diameter index safety standard. A standard for high pressure gas inlet fittings.
ESENS Expiratory sensitivity. A setting that determines the percent of peak inspiratory
flow (or flow rate expressed in L/min in a PAV breath) at which the ventilator
cycles from inspiration to exhalation for spontaneous breaths. Low settings will
result in longer spontaneous inspirations.
expiratory pause an operator-initiated maneuver that closes the inspiration (proportional sole-
noid) and exhalation valves during the exhalation phase of a mandatory breath.
The maneuver can be used to determine intrinsic (auto) PEEP (PEEPI).
exhalation valve (EV) The valve in the expiratory limb of the ventilator breathing system that controls
PEEP.
f, fTOT Respiratory rate, as a setting (f) in A/C, SIMV, and BiLevel the minimum number
of mandatory breaths the patient receives per minute. As a monitored value
(fTOT), the average total number of breaths delivered to the patient.
FAILURE A category of condition detected during SST or EST that causes the ventilator to
enter the safety valve open state. A ventilator experiencing a FAILURE requires
removal from clinical use and immediate service.
flow pattern A setting that determines the gas flow pattern of mandatory volume-controlled
breaths.
gold standard test circuit Test circuit designed for use with EST.
GUI Graphical user interface. The ventilator’s touch screen used to enter patient set-
tings. and alarm settings, including off-screen keys, soft keys, and knobs.
hard bound A ventilator setting that has reached its minimum or maximum limit.
humidification type A setting for the type of humidification system (HME, non-heated expiratory
tube, or heated expiratory tubing) in use on the ventilator.
I:E ratio The ratio of inspiratory time to expiratory time. Also, the operator- set timing vari-
able that applies to PC and VC+ mandatory breaths.
inspiratory pause An operator-initiated maneuver that closes the inspiration (proportional sole-
noid) and exhalation valves at the end of the inspiratory phase of a mandatory
breath. The maneuver can be used to determine static compliance (CSTAT) and
static resistance (RSTAT).
invasive ventilation Patient ventilation while intubated with an endotracheal (or tracheostomy) tube.
latched alarm An alarm whose visual alarm indicator remains illuminated after the alarm has
autoreset.
lockable alarm An alarm that does not terminate an active alarm silence function.
maintenance All actions necessary to keep equipment in, or restore it to, serviceable condition.
Includes cleaning, servicing, repair, modification, overhaul, inspection, and per-
formance verification.
mandatory breath A breath whose settings and timing are preset; can be triggered by the ventila-
tor, patient, or operator.
mandatory type The type of mandatory breath: volume control (VC), VC+ or pressure control (PC).
mode Ventilatory mode. The algorithm that determines type and sequence of breath
delivery.
NIST Non-interchangeable screw thread. A standard for high pressure gas inlet fit-
tings.
non-invasive ventilation Patient ventilation without the use of an endotracheal tube; instead using inter-
(NIV) faces such as masks, nasal prongs, or uncuffed endotracheal tubes.
non-technical alarm An alarm caused due to a fault in the patient-ventilator interaction or a fault in
the electrical or gas supplies that the practitioner may be able to alleviate.
normal ventilation The state of the ventilator when breathing is in progress and no alarms are active.
O2 % Both a ventilator setting and a monitored variable. The O2% setting determines
the percentage of oxygen in the delivered gas. The O2% monitored data is the
percentage of oxygen in the gas delivered to the patient, measured at the ven-
tilator outlet upstream of the inspiratory filter.
ongoing background Continuously running tests during ventilation that assess the ventilator's elec-
checks tronics and pneumatics hardware.
OSC Occlusion status cycling. A ventilation mode in effect during a severe occlusion.
In this mode, the ventilator periodically attempts to deliver a pressure-based
breath while monitoring the inspiration and expiration phases for the continuing
existence of the occlusion.
OVERRIDDEN The final status of an SST or EST run in which the operator used the override fea-
ture. (The ventilator must have ended the test with an ALERT condition.)
patient circuit The entire inspiratory-expiratory conduit, including tubing, humidifier, and water
traps.
patient data alarm An alarm condition associated with an abnormal condition of the patient's respi-
ratory status.
patient problems A definition used by the ventilator's safety net. Patient problems are declared
when patient data are measured equal to or outside of alarm thresholds and are
usually self-correcting or can be corrected by a practitioner. The alarm monitor-
ing system detects and announces patient problems. Patient problems do not
compromise the ventilator's performance.
PBW Predicted body weight, a ventilator setting that specifies the patient's body
weight assuming normal fat and fluid levels. Determines absolute limits on tidal
volume and peak flow, and allows appropriate matching of ventilator settings to
the patient.
PEEP Positive end expiratory pressure. The measured circuit pressure (referenced to
the patient wye) at the end of the expiratory phase of a breath. If expiratory
pause is active, the displayed value reflects the level of any active lung PEEP.
PEEPI Intrinsic PEEP. Indicates a calculated estimate of the pressure above the PEEP
level at the end of exhalation. Determined during an expiratory pause maneuver.
PI END End inspiratory pressure. The pressure at the end of the inspiration phase of the
current breath. If plateau is active, the displayed value reflects the level of end-
plateau pressure.
PMEAN Mean circuit pressure, a calculation of the measured average patient circuit pres-
sure over an entire respiratory cycle.
PPEAK Maximum circuit pressure, the maximum pressure during the inspiratory and
expiratory phases of a breath.
PSENS Pressure sensitivity. The operator-set pressure drop below PEEP (derived from
the patient's inspiratory flow) required to begin a patient-initiated breath when
pressure triggering is selected.
PSUPP Pressure support. A setting of the level of inspiratory assist pressure (above PEEP)
at the patient wye during a spontaneous breath (when spontaneous breath type
is PS).
remedy message A message displayed on the GUI during an alarm condition suggesting ways to
resolve the alarm.
restricted phase of exhala- The time period during the exhalation phase where an inspiration trigger is not
tion allowed. The restricted phase of exhalation is defined as the first 200 ms of exha-
lation, OR the time it takes for expiratory flow to drop to ≤ 50% of the peak expi-
ratory flow, OR the time it takes for the expiratory flow to drop to ≤ 0.5 O2%
(whichever is longest). The restricted phase of exhalation will end after five (5)
seconds of exhalation have elapsed regardless of the measured expiratory flow
rate.
rise time % A setting that determines the rise time to achieve the set inspiratory pressure in
pressure-controlled (PC), VC+ BiLevel, or pressure-supported (PS) breaths. The
larger the value, the more aggressive the rise of pressure.
safety net The ventilator's strategy for responding to patient problems and system faults.
safety valve (SV) A valve residing in the ventilator’s inspiratory module designed to limit pressure
in the patient circuit. When open, it allows the patient to breathe room air if able
to do so.
safety ventilation A mode of ventilation active if the patient circuit is connected before ventilator
startup is complete, or when power is restored after a loss of five (5) minutes or
more.
service mode A ventilator mode providing a set of services tailored to the needs of testing and
maintenance personnel. When in the service mode, the ventilator does not
provide ventilation.
SIS Sleeved index system. A standard for high pressure gas inlet fittings.
soft bound A ventilator setting that has reached its recommended high or low limit, accom-
panied by an audible tone. Setting the ventilator beyond this limit requires the
operator to acknowledge a visual prompt to continue.
SPONT Spontaneous. A ventilatory mode in which the ventilator delivers only sponta-
neous breaths. In SPONT mode, the patient triggers all breaths delivered by the
ventilator with no set mandatory respiratory rate. The patient controls the breath
variables, potentially augmented by support pressure.
spontaneous type A setting that determines whether spontaneous breaths are pressure-supported
(PS), tube-compensated (TC), volume-supported (VS), or proportionally assisted
(PAV).
SST Short self test. A test that checks circuit integrity, calculates circuit compliance
and filter resistance, and checks ventilator function. Operator should run SST at
specified intervals and with any replacement or alteration of the patient circuit.
STPD Standard temperature and pressure, dry. Defined as dry gas at a standard atmo-
sphere (760 mmHg, 101.333 kPa, approximately 1.0 bar) and 0°C.
SVO Safety valve open. An emergency state in which the ventilator opens the safety
valve so the patient can breathe room air unassisted by the ventilator if able to
do so. An SVO state does not necessarily indicate a ventilator inoperative condi-
tion. The ventilator enters an SVO state if a hardware or software failure occurs
that could compromise safe ventilation, with the loss of the air and oxygen sup-
plies, or if the system detects an occlusion.
system fault A definition used by the ventilator's safety net. System faults include hardware
faults (those that originate inside the ventilator and affect its performance), soft
faults (faults momentarily introduced into the ventilator that interfere with
normal operation), inadequate supply (AC power or external gas pressure), and
patient circuit integrity (blocked or disconnected circuit).
TA Apnea interval, the operator-set variable that defines the breath-to-breath inter-
val which, if exceeded, causes the ventilator to declare apnea and enter apnea
ventilation.
TE Expiratory time. The expiratory interval of a breath. Also the operator-set timing
variable that determines the expiratory period for pressure-controlled (PC) or
VC+ mandatory breaths.
technical alarm An alarm occurring due to a violation of any of the ventilator's self monitoring
conditions, or detected by background checks.
TI Inspiratory time, the inspiratory interval of a breath. Also, the operator-set timing
variable that determines the inspiratory interval for pressure-controlled (PC) or
VC+ mandatory breaths.
TPL Plateau time. The amount of time the inspiration phase of a mandatory breath is
extended after inspiratory flow has ceased and exhalation is blocked. Increases
the residence time of gas in the patient's lungs.
VE TOT Minute volume, the expiratory tidal volume normalized to unit time
(L/min). The displayed value is compliance- and BTPS-compensated.
VBS Ventilator breathing system. Includes the gas delivery components of the venti-
lator the patient circuit with tubing, filters, humidifier, and other accessories; and
the ventilator's expiratory metering and measurement components.
Ventilation Assurance A safety net feature which is invoked if a system fault in the mix subsystem, inspi-
ratory subsystem, or expiratory subsystem occurs compromising the ventilator’s
ability to ventilate the patient as set.
Ventilator Inoperative An emergency state the ventilator enters if it detects a hardware failure or a crit-
(vent inop) ical software error which could compromise safe ventilation. During a ventilator
inoperative condition, the safety valve opens to allow the patient to breathe
room air unassisted by the ventilator if able to do so. Qualified service personnel
must power up the ventilator and run EST before normal ventilation can resume.
VMAX Peak flow. A setting of the peak (maximum) flow of gas delivered during a VC
mandatory breath. (Combined with tidal volume, flow pattern, and plateau, con-
stant peak flow defines the inspiratory time.) To correct for compliance volume,
the ventilator automatically increases the peak flow.
VSENS Flow sensitivity. A setting that determines the rate of flow inspired by the patient
that triggers the ventilator to deliver a mandatory or spontaneous breath (when
flow triggering is selected).
VT Tidal volume. A setting that determines the volume inspired and expired with
each breath. The VT delivered by some Puritan Bennett ventilators is an operator-
set variable that determines the volume delivered to the patient during a man-
datory, volume-based breath. VT is compliance-compensated and corrected to
body temperature and pressure, saturated (BTPS).
VTRIG Flow triggering. A method of recognizing patient inspiratory effort in which the
ventilator monitors the difference between inspiratory and expiratory flow mea-
surements. The ventilator triggers a breath when the difference between inspi-
ratory and expiratory flows increases to a value that is at least the value selected
for flow sensitivity (VSENS).
AC, also ac Alternating current. The movement of electrical charge that periodically
reverses direction.
DC, also dc Direct current. The movement of electrical charge flowing in a single direction.
NVRAM, also NovRam Non-volatile random access memory. Memory that is kept active across resets
and power cycles and is not normally initialized at startup.
Index-1
Index
Exhalation — Detection and Initiation . . . .10-8–10-11 icons
expiratory module configure ............................................................................... 2-20
Exhalation flow sensor assembly elevate O2 ................................................................................ 2-20
removal, disinfection, reassembly ............7-9–7-19 grid lines ................................................................................. 2-20
Expiratory Pause . . . . . . . . . . . . . . . . . . . . . . . 10-51–10-52 help ........................................................................................... 2-20
Expiratory Pause Maneuvers . . . . . . . . . . . . . . . . . . . .4-30 high priority alarm ............................................................ 2-21
Expiratory Sensitivity (ESENS) . . . . . . . . . . . . . . . . . . 10-67 logs ........................................................................................... 2-19
Expiratory Time (TE) . . . . . . . . . . . . . . . . . . . . . . . . . . 10-64 low priority alarm .............................................................. 2-21
extended battery installation . . . . . . . . . . . . . . . . . . .3-22 maximize waveform ........................................................ 2-21
Extended Self Test (EST) . . . . . . . . . . . . . . . . . . . . . . 10-75 medium priority alarm ................................................... 2-21
pause ....................................................................................... 2-20
F restore waveform .............................................................. 2-21
Filter Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10 screen capture .................................................................... 2-20
Flow Pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-61 unread items ........................................................................ 2-20
Flow Sensitivity (VSENS) . . . . . . . . . . . . . . . . . . . . . . . 10-61 ventilator setup .................................................................. 2-19
waveform layout ............................................................... 2-20
IEC classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
G Inspiration — Detection and Initiation . . . . 10-4–10-8
gas failure cross flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8 Inspiratory Pause . . . . . . . . . . . . . . . . . . . . . . 10-49–10-51
Gestures Inspiratory pause maneuvers . . . . . . . . . . . . 4-29, 10-49
..............................................................................................4-6–4-7 Inspiratory Pressure (PI) . . . . . . . . . . . . . . . . . . . . . . . . 10-62
double-tap ...............................................................................4-7
Inspiratory Time (TI) . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-63
drag and drop ........................................................................4-7
touch and hold ......................................................................4-7 Installation Testing (testing prior to ventilating a
gestures patient) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-43–3-48
Drag .............................................................................................4-7
swipe ...........................................................................................4-6 L
Graphical User Interface (GUI) . . . . . . . . . . . . . . . . . . .2-16 Low Pressure (PL) in BiLevel . . . . . . . . . . . . . . . . . . . . 10-64
GUI Control Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16 Low Time (TL) in BiLevel . . . . . . . . . . . . . . . . . . . . . . . 10-65
GUI control keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
GUI indicators M
audible ....................................................................................2-25 Mandatory Breath Delivery . . . . . . . . . . . . . . . . . . . . 10-16
visual ...........................................................................2-18–2-21 Manual Inspiration . . . . . . . . . . . . . . . . . . . . . . 4-27, 10-21
GUI screen capture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . 11-31
H Measurement Uncertainty . . . . . . . . . . . . . . . . . . . . . 11-1
hard bound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 MISCA response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
High Pressure (PH) in BiLevel . . . . . . . . . . . . . . . . . . 10-64 MISCF response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
High Spontaneous Inspiratory Time Mode and Breath Type . . . . . . . . . . . . . . . . . . . . . . . . 10-58
Limit (2TI SPONT ) . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-68 Monitored Patient Data . . . . . . . . . . . . . . . . . . . . . . . . 6-40
High Time (TH) in BiLevel . . . . . . . . . . . . . . . . . . . . . 10-64 Monitoring accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . 11-27
How to Connect the Gas Supplies . . . . . . . . . . . . . . . . 3-7
N
How to Connect the Patient Circuit . . . . . . . . . . . . .3-14
How to connect the Ventilator to AC Power . . . . . . 3-5 NIV
how to enter Service mode . . . . . . . . . . . . . . . . . . . . .3-31 alarm settings ...................................................................... 4-26
How to Install Accessories . . . . . . . . . . . . . . . . 3-18–3-28 apnea settings .................................................................... 4-26
How to Store the Ventilator for an Extended Time high spontaneous inspiratory time limit setting ..4-
Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-23 25
how to use the ventilator system . . . . . . . . . . . 4-7–4-20 setup ........................................................................................ 4-22
how to use the ventilator’s user interface . . . . 4-3–4-7 Non-invasive ventilation (NIV) . . . . . . . . . . . . 4-21–4-27
how to view ventilator logs . . . . . . . . . . . . . . . . . . . . . . 8-4
Humidification Type . . . . . . . . . . . . . . . . . . . . . . . . . . 10-69 O
humidifier installation . . . . . . . . . . . . . . . . . . . . . . . . . .3-25 Occlusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-44
Humidifier Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-69 Omni-directional LED . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
On-screen Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
I On-screen Symbols and Abbreviations . . . . 2-21–2-25
I:E ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-64 Operation Verification . . . . . . . . . . . . . . . . . . . . . . . . . . 3-56
Index-2
Index
oxygen sensor return to previous ............................................................. 4-19
calibration .............................................................................4-33 ventilator ................................................................................ 4-10
calibration test ....................................................................4-33 Short Self Test (SST) . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-75
function ..................................................................................4-31 SIMV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-33
SIMV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-38
P soft bound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
P0.1 maneuver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-53 specifications
Patient Data Parameters . . . . . . . . . . . . . . . . . . 6-41–6-48 electrical ................................................................................. 11-7
patient data range and resolution . . . . . . 11-20–11-26 environmental .................................................................... 11-8
Patient Related Problems . . . . . . . . . . . . . . . . . . . . . 10-71 performance ........................................................................ 11-9
Peak Inspiratory Flow (VMAX) . . . . . . . . . . . . . . . . . 10-60 physical characteristics .................................................. 11-3
pneumatic ............................................................................. 11-4
PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-65 technical ................................................................................. 11-4
Percent Support in PAV+ . . . . . . . . . . . . . . . . . . . . . . 10-66 Spontaneous (SPONT) Mode . . . . . . . . . . . 10-38–10-39
Percent Support in TC . . . . . . . . . . . . . . . . . . . . . . . . 10-66 Spontaneous Breath Delivery . . . . . . . . . . . . . . . . . . 10-21
physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . .11-3 SST
Plateau Pressure (PPL) . . . . . . . . . . . . . . . . . . . . . . . . . . .6-44 how to run ............................................................................ 3-45
Plateau Time (TPL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-61 outcomes .............................................................................. 3-48
Pneumatic Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-39 results ...................................................................................... 3-48
Power On Self Test (POST) . . . . . . . . . . . . . . . . . . . . 10-74 test sequence ...................................................................... 3-45
Preparing the Ventilator for Use . . . . . . . . . . . . . . . . .3-34 Status Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29–2-37
Pressure Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-62 Surface Cleaning of Exterior Surfaces . . . . . . . . . . . . 7-4
Pressure Support (PSUPP) . . . . . . . . . . . . . . . . . . . . . . 10-66 symbols
primary battery installation . . . . . . . . . . . . . . . . . . . . .3-20 BDU rear panel label symbols
Primary display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16 and descriptions .......................................................... 2-11
Printing data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-21 safety symbol definitions .................................................1-3
Product Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 shipping label symbols and descriptions ..............1-2
pushpin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 System Related Problems . . . . . . . . . . . . . . . . . . . . . 10-71
Q T
quick start use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 TC
alarms ................................................................................... 10-27
R monitored patient data .............................................. 10-27
respiratory maneuvers PBW and tube ID ............................................................ 10-27
expiratory pause maneuver ..................................... 10-51 technical description ................................................... 10-26
inspiratory pause maneuver .................................... 10-49 tube type, tube ID, humidification .......................... 4-13
NIF maneuver ................................................................... 10-52 technical assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
respiratory mechanics maneuvers Technical Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Negative Inspiratory Force maneuver (NIF) ..... 10-52 Tidal Volume (V T ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-60
P0.1 maneuver .................................................................. 10-53
vital capacity maneuver (VC) ................................... 10-54 U
Respiratory Rate (f ) . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-59 Used Part Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Rise Time % . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-67 Using Battery Power . . . . . . . . . . . . . . . . . . . . . . . . 3-2–3-3
RS-232 commands
RSET .............................................................................................5-6 V
SNDA ...........................................................................................5-6 VC+
SNDF .........................................................................................5-10 maximum pressure adjustments .......................... 10-20
startup ...................................................................10-19–10-20
S Vent Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-57
Safety Net . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-69 ventilating a new patient . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-36 ventilating the same patient . . . . . . . . . . . . . . . . . . . . 4-8
screen captures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 ventilator
Serial commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5 alarm log ...................................................................................8-2
serial number interpretation . . . . . . . . . . . . . . . . . . . .1-19 available languages ............................................................5-1
settings BDU controls and indicators ....................................... 2-26
alarm ........................................................................................4-15 BDU front view ......................................................................2-8
apnea .......................................................................................4-14
Index-3
Index
BDU rear label symbols flow pattern ....................................................................... 10-61
and descriptions .............................................2-11–2-13 flow sensitivity (VSENS) ................................................ 10-61
BDU rear view ......................................................................2-10 high pressure (PH) .......................................................... 10-64
BDU right side view .............................................2-14, 2-15
high time (TH) .................................................................. 10-64
Components List ..................................................................2-4
connectors ............................................................................2-37 humidification type ...................................................... 10-69
Description ..............................................................................2-2 humidifier volume ......................................................... 10-69
EST/SST status log ................................................................8-3 I:E ratio .................................................................................. 10-64
function .....................................................................................4-1 inspiratory pressure (PI) .............................................. 10-62
gas flow overview .............................................................10-2 inspiratory time (TI) ....................................................... 10-63
general event log .................................................................8-3 low pressure (PL) ............................................................. 10-64
GUI front view ........................................................................2-6 low time (TL) ..................................................................... 10-65
GUI rear view ..........................................................................2-7 mode and breath type ................................................ 10-58
Indications For Use ..............................................................2-3 peak inspiratory flow (VMAX) ................................... 10-60
Operation ....................................................................4-7–4-21
patient data log .....................................................................8-2 PEEP ....................................................................................... 10-65
service log ................................................................................8-3 percent support in TC .................................................. 10-66
settings log ..............................................................................8-2 plateau time (TPL) .......................................................... 10-61
system diagnostic log ........................................................8-3 pressure sensitivity (PSENS) ........................................ 10-62
Ventilator Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 pressure support (PSUPP) ............................................ 10-66
Ventilator Operating Modes . . . . . . . . . . . . . . . 3-28–3-33 range and resolution ....................................... 11-9–11-16
ventilator operating modes respiratory rate (f ) .......................................................... 10-59
Normal mode ......................................................................3-28 rise time % ......................................................................... 10-67
Quick Start mode ..............................................................3-28 tidal volume (V T ) ............................................................ 10-60
Service mode ......................................................................3-31 vent type ............................................................................. 10-57
Stand-By state .....................................................................3-29 volume support (V T SUPP) .......................................... 10-66
Ventilator Protection Strategies . . . . . . . . . . . 4-34–4-36 Ventilator Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
ventilator settings Volume Support (V T SUPP) . . . . . . . . . . . . . . . . . . . . . 10-66
2TI SPONT ................................................................................10-68 VS
apnea ventilation ........................................................... 10-55 maximum pressure adjustments .......................... 10-25
circuit type and PBW .................................................... 10-56 startup .................................................................................. 10-25
configuration ..........................................................3-35–3-43
disconnect sensitivity (DSENS) ................................. 10-68 W
expiratory sensitivity (ESENS) ..................................... 10-67 Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
expiratory time (TE) ....................................................... 10-64 waveform axis scaling . . . . . . . . . . . . . . . . . . . . . . . . . . 3-42
Index-4
Part No. 10104514 E 2018-02
© 2013 Covidien.
Covidien llc
15 Hampshire Street, Mansfield, MA 02048
Covidien Ireland Limited, IDA Business and Technology Park,Tullamore.
www.covidien.com