The document provides detailed specifications and features of a biphasic defibrillator, highlighting its effectiveness in defibrillation with a current-controlled bi-phasic waveform and a wide range of energy levels from 2-300 Joules. It includes advanced functionalities such as voice prompts, metronome support for CPR, and the ability to store critical event ECG information. The device also offers optional enhancements like AED mode, pulse oximetry, and capnography for comprehensive patient monitoring.

Uploaded by

Alex Ethiopia
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
94 views4 pages

Relife 900

The document provides detailed specifications and features of a biphasic defibrillator, highlighting its effectiveness in defibrillation with a current-controlled bi-phasic waveform and a wide range of energy levels from 2-300 Joules. It includes advanced functionalities such as voice prompts, metronome support for CPR, and the ability to store critical event ECG information. The device also offers optional enhancements like AED mode, pulse oximetry, and capnography for comprehensive patient monitoring.

Uploaded by

Alex Ethiopia
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Medical

Technologies

Experience the
Effectiveness
BIPHASIC DEFIBRILLATOR

POWERED BY

www.bplmedicaltechnologies.com
7 inch Colour TFT Screen
Allows the clear and precise display of multiple
waveforms and defibrillation parameters

Bi-phasic waveform (Current - Controlled)


The current controlled bi-phasic waveforms
lowers the electrical threshold for successful
defibrillation, besides adjusting for impedance

High Energy Efficiency


100 charge / discharge of 300J with fully charged
new battery

Wide Energy Levels


2-300 J of energy level can be selected for
Trolley defibrillation by the user

Option
Remote energy selection and charging
Available Remote energy selection, charging and discharging
through paddles facilitate convenient usage

Voice Prompt Support


Continuous voice prompts and on screen messages
to aid in revival and resuscitation process

Metronome Signal for CPR Support


Metronome signal for synchronizing chest
compression during CPR

Synchronized Cardioversion
Synchronized Cardioversion function is available
for providing defibrillation on sync with ECG
waveform

24 Events
24 Memorises always the last 24 Critical Event ECG
information which can be viewed or recorded

Patient-Paddle Contact Indication


Patient-Paddle quality of contact is provided on
screen

Optional Enhancements*

AED Mode with Selectable Energy Protocols Non-invasive Blood Pressure


In Automated External Defibrillator (AED)mode, Accurate NIBP measurements powered by
the shock sequence can be selected based on the
requirement

Pulse oximetry Capnography or External Pacing


O2 Accurate SpO2 measurements powered by
CO2/
PACEMAKER Accurate monitoring of ETCO2 with Masimo Sensor,
Option of Mainstream or sidestream
External Transcutaneous Pacemaker with Fixed or
Demand Mode

*Optional Enhancements at additional cost


Product Specifications
Manual Defibrillator
Defibrillator waveform Current Controlled, Biphasic
Energy selection option 2-300 Joules in steps of 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 250, 300 Joules
Energy selection control Selected using keys located on the front panel or on the sternum paddle
Charge control Control on the front panel and on the apex paddle
Charge ready indication Charge done LED on the front panel CHARGE key and Charge done LED on the apex paddle
Synchroniser Energy is delivered within 50 msec of the detected R-wave peak.
Patient impedance 25 – 200 Ohms
AED
AED mode selection control Control on the front panel
Shock sequence Low-Low-Low / Low-Low-High / Low-High-High (User selectable)

Charge ready indication Onscreen display, voice prompt and blinking LED in the SHOCK key.
Shock delivery control Control on the front panel with blinking LED
ECG analysis time < 15 seconds
ECG
ECG Input modes ECG input through Paddles or ECG Patient cable

ECG Lead selection LEAD I / II / III / PADDLES (3 Lead Patient cable)


LEAD I / II / III / aVR / aVL / aVF / V / PADDLES (5 Lead Patient cable)
CMRR >90 dB at 50 Hz
Frequency response 0.5 to 40 Hz
Heart rate display 10 to 300 bpm; Accuracy ± 2 bpm
User gain settings 8 level setting 2.5, 5, 7.5, 10, 15, 20, 30 and 40 mm/mV

Sweep speed 25 ± 1 mm/sec


Reset recovery Automatic recovery of waveform within 0.5 seconds after overload or defibrillation
MASIMO SPO2
SpO2 Range 1 – 100%
Accuracy ±2% (adult/child, in non-motion state), ±3% (adult/child, in motion state) within the measurement range of
70%~100%. 1%~69% is not defined.
Averaging time 2-4 / 4-6 / 8 / 10 / 12 / 14 / 16 Seconds (User selectable)
Resolution 1%
Pulse Rate range 25 – 240 bpm
Perfusion Index resolution 0.01%
SunTech NIBP
Method Oscillometric
Operating Modes Manual, Automatic and Stat
Product Specifications
Blood Pressure Range
Systolic Adult: 40 – 250 mmHg | Paediatric: 40 – 230 mmHg

Diastolic Adult: 20 – 200 mmHg | Paediatric: 20 – 160 mmHg

MAP Adult: 26 – 220 mmHg | Paediatric: 26 – 183 mmHg

Pressure Transducer Accuracy ± 3 mmHg or 2% of the reading whichever is greater


Cycle Time in Auto mode 1, 2, 3, 5, 10, 15, 30, 60, 90 Minutes (User selectable)
Cycle Time in Stat mode 5 Minutes
MASIMO ETCO2
Method Option: Mainstream or Sidestream
CO2 Measurement range 0.0 – 25.0 vol% (0mmHg~190mmHg)
Accuracy 0~15%: ±(0.2 vol% + 2% of reading)
15~25%: not defined
Display unit Vol% or mmHg
CO2 resolution 1mmHg, 0.1kPa or 0.1%
Respiration rate (awRR) 0 to 150 ± 1 breaths/min
Warm-up time < 10 seconds
Calibration Sidestream - An automatic zeroing is performed 1 to 3 times per day.
Mainstream - Zeroing recommended when changing Airway adapter.
Dimensions with weight 1.3"" x 3.1"" x 1.9""; <130 g - Sidestream
1.49” x 1.45” x 1.34”; < 25 g - Mainstream
External Pacing
Type External transcutaneous pacemaker
Pacing mode Fixed or Demand
Waveform Rectangular pulse
Pulse duration 40 ± 2 msec
Pacing rate 30 to 180 ppm
Output current 0 to 150mA
Refractory period 340 msec between 30 and 80 ppm. | 240 msec above 80ppm.

Thermal Recorder
Paper speed 25 ± 1 mm/sec
Paper size and type Thermal paper in rolls. 58 mm width, 15 meters

Print channel 1
General
Display 7” Colour TFT, Resolution 800 x 480
Dimensions 370 mm (W) x 300 mm (H) x 290 mm (D)
Weight Less than 8.5 Kg. (including battery, external paddles)
Operating temperature range 0 to 40° C at non condensing RH between 15 and 95%
Storage temperature range -10 to 60° C at non condensing RH between 20 and 80%
AC input 100 – 240 Vac, 50/60 Hz

Power consumption < 240 VA


Battery type Sealed Lead Acid 12V, 4.5 AH, Rechargeable

Battery capacity > 100 discharges of 300 Joules


Battery charging time Approx. 3 hours to 90% of full capacity. 18 hours to 100% capacity
Event information Last 24 critical events are stored along with ECG information. Events include MARK key press, Shock delivery
and HR Alarm violation.
Trend Last 200 NIBP measurement data and last 200 Alarm violation information.
*Technical specification subject to change

CERTIFIED ISO 13485 : 2016 COMPANY

BPL Medical Technologies Private Limited


Regd. Office: 11th KM, Bannerghatta Road,
Arakere, Bangalore - 560076, India.
Toll Free: 1800-4252355
Website: www.bplmedicaltechnologies.com
For Enquiries: sales.medical@bpl.in
CIN: U33110KA2012PTC067282 Follow us on

© 2019 BPL Medical Technologies Private Limited. All rights reserved. BPL Medical Technologies Private Limited reserves the right to make changes in BPL RELIFE900 4PGB 07:05:19
product features, specifications, aesthetics and/or to discontinue the same at any time without notice or obligation

You might also like