Inhalational Anesthesia
Inhalational Anesthesia
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                               60                                                 Ó 2021 Published by Elsevier Ltd.
                                                                 PHARMACOLOGY
  C   Low cost                                     C   Low blood:gas partition coefficient          Central nervous system (CNS)
  C   Easy to manufacture                          C   High oil:gas partition coefficient           C  Only effect the CNS causing anaesthesia
  C   Environmentally friendly                     C   Excreted rapidly and unchanged from the      C  Not epileptogenic
  C   Stable and non-flammable                         lungs                                        C  No change in cerebral blood flow (CBF) or
  C   Liquid at room temperature (boiling point                                                        intracranial pressure (ICP)
      above ambient temperature)                                                                    C  Rapid reversible
  C   Long shelf life                                                                               C  Smooth induction
  C   Non irritant                                                                                  C  Analgesic properties
  C   Low specific heat capacity                                                                    Cardiovascular (CV)
  C   Low latent heat of vaporization                                                               C  No depression
  C   High saturated vapor pressure                                                                 C  Stable heart rate
  C   Preservative free                                                                             C  No effect on coronary blood flow
                                                                                                    Respiratory
                                                                                                    C  Bronchodilation
                                                                                                    C  No respiratory depression, breath holding,
                                                                                                       laryngospasm, or increase in secretions
                                                                                                    Other
                                                                                                    C  No effect on kidneys or liver
                                                                                                    C  Muscle relaxation
                                                                                                    C  Will not cause malignant hyperthermia
                                                                                                    C  No interactions
                                                                                                    C  Safe in pregnancy
                                                                                                    C  Anti-emetic
                                                                                                    C  No interactions with other anaesthetic
                                                                                                       agents and allows use of high FiO2
Table 1
before the partial pressure increases. The molecules readily enter                If the MV is increased, it leads to more anaesthetic agent
the bloodstream from the alveoli, therefore alveolar concentra-                    being delivered to the alveolus, a higher alveolar con-
tion and partial pressure remain low. As a result, the brain partial               centration, and a faster onset.
pressure rises slowly, and anaesthetic onset takes longer. This                 Functional residual capacity (FRC)
means a higher FA/Fi ratio is obtained more slowly.                               A large FRC will dilute the anaesthetic agent, which will
   There are other factors which influence the speed of onset of                   mean that the alveolar concentration will rise more
an agent. They all relate to the balance between uptake of                         slowly and slow the onset of anaesthesia.
anaesthetic into the blood from the alveoli and delivery of drug to             Cardiac output (CO)
the alveoli.                                                                      If the CO is high, it leads to greater pulmonary blood
    Minute ventilation (MV)                                                       flow (if there is no shunt, cardiac output ¼ pulmonary
  Desflurane      168       23.5        89.2       6.6         0.45           29            Pungent           0.02, trifluoroacetic acid
  N20             44        -88         5200       105         0.47           1.4           Odourless         0.01, nitrogen
  Sevoflurane     200.1     58.5        22.7       1.8e2.2     0.7            80            Non-irritant      3.5, inorganic and organic fluorides,
                                                                                                              and compound AeE
  Isoflurane      184.5     48.5        33.2       1.17        1.4            98            Irritant          0.2, trifluoroacetic acid and F
  Enflurane       184.5     56.5        23.3       1.68        1.8            98            Non-irritant      2, Inorganic and organic fluorides
  Halothane       197       50.2        32.3       0.75        2.4            224           Non-irritant      20, trifluoroacetic acid, Cl and Br
  Xenon           131       108        e          71          0.14           1.9           Odourless         Nil
Table 2
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                            61                                               Ó 2021 Published by Elsevier Ltd.
                                                                PHARMACOLOGY
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                             62                                             Ó 2021 Published by Elsevier Ltd.
                                                              PHARMACOLOGY
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                           63                                             Ó 2021 Published by Elsevier Ltd.
                                                              PHARMACOLOGY
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                           64                                             Ó 2021 Published by Elsevier Ltd.
                                                             PHARMACOLOGY
  Patient age higher (w10% for every 10 years)                             Decreasing age (peaks at 6 months and slowly declines as child ages)
  Neonatal period                                                          Hyperthermia
  Pregnant patient                                                         Chronic alcohol and opioid use
  Hypothermia                                                              Anxiety and Sympathetic stimulation
  Hypovolaemia
  Hypoxia and Hypocapnia
  Anaemia
  Acute alcohol intoxication
Table 3
It is thought to be due to hepatic hypoxia caused by halothane                  Gases composed of two dissimilar molecules (e.g. carbon
metabolism in the liver leading to this derangement.                              dioxide and nitrous oxide) absorb infrared radiation of a
    Type two halothane hepatitis is less common but leads to                      particular wavelength and convert the energy into molec-
necrosis of hepatocytes and fulminant liver failure. The incidence                ular vibration.
is low, with a reported range between 1 adult patient in every                  Absorption of a wavelength of infrared radiation by a
3500e35,000 cases; however, there is a very high mortality rate                   volatile agent is proportional to the concentration of the
(50e70%). It is caused by an autoimmune response. An inter-                       gas and allows its concentration to be measured.
mediate metabolite called trifluoroacetyl chloride (TFA) cova-                 Using these principles, an infrared gas analyser uses an
lently binds to hepatocytes and forms a complex that induces               infrared light source that passes a light beam through a filter to
antibody formation. Risk factors include repeated exposure,                obtain radiation of a specific wavelength, which is passed
obesity, female gender and hypoxia. It is recommended to avoid             through the sample chamber. The required wavelength of radi-
halothane use if there has been a previous reaction to the drug, if        ation is dependent on the gas being measured. A proportion of
the patient has had the drug within the previous 3 months or if            the light sample is absorbed by the gas during measurement.
they have a history of unexplained liver injury or jaundice.               Once the light has left the sample chamber a detector then
                                                                           measures the intensity of the light signal remaining, and the re-
Seizure activity                                                           sults are then displayed for interpretation. Often, reference
Enflurane is associated with abnormal epileptiform activity on             chambers that do not contain the gas being measured are used
EEG, especially if it is associated with hypocapnia.                       for calibration to improve accuracy. Water vapour can also in-
                                                                           fluence the result, so gases are dried before they are analysed.
Dysrhythmias
                                                                               Other molecules can interfere with the measurement of vol-
Halothane is the most arrhythmogenic inhalational agent. It
                                                                           atile agents at their specific wavelengths. Collision broadening
sensitizes the myocardium to circulating catecholamines, which
                                                                           occurs when carbon dioxide and nitrous oxide interact with one
can induce malignant arrythmias. This is worsened with an
                                                                           another. Normally the absorption peaks of infrared light for
acidosis and hypercapnia. It has also been known to induce
                                                                           carbon dioxide and nitrous oxide are between 4 and 5 mm.
bradycardias, junctional rhythms, and ventricular ectopic beats.
                                                                           However, the energy absorbed by one molecule of carbon diox-
Nephrotoxicity                                                             ide when it has already absorbed light energy is transferred to a
When inhaled anaesthetic agents undergo metabolism via the                 molecule of nitrous oxide when they collide. This means that the
CYP450 system they release halogen ions which are potentially              carbon dioxide molecule can absorb more energy giving a
nephrotoxic, although this does not appear to be clinically sig-           broader absorption spectrum of between 3 and 12 mm. This will
nificant with sevoflurane which undergoes significant                      interfere with and broaden the absorption peaks of volatile
defluorination.                                                            agents which absorb infrared between 8 and 13 mm. However,
                                                                           modern systems compensate for this phenomenon.5
Measuring concentrations of volatile agents                                    Other less common means of measuring vapours include mass
                                                                           spectroscopy, gas chromatography, the piezoelectric effect using
A vapour analyser is essential whenever volatile anaesthesia is            quartz crystals, raman spectroscopy, photoacoustic spectros-
used for anaesthesia. The concentration of volatile anaesthetics           copy, refractometers and katharometers.
can be measured by several means. In theatre, most systems use
infrared spectroscopy.
                                                                           Vaporizers
   Infrared absorption spectroscopy relies on several principles:
     Beer’s law states that absorption of radiation increases as          Some inhalational agents are volatile liquids (for example, des-
       the concentration of a substance increases.                         flurane, sevoflurane, enflurane) which need to be vaporized to
     Lambert’s law states that the intensity of transmitted light         be delivered via inhalation to produce general anaesthesia. A gas
       reduces exponentially as the distance travelled through the         is defined as a substance in its gaseous state in a temperature
       substance increases.                                                above its critical temperature. A vapour is a substance which is
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                          65                                               Ó 2021 Published by Elsevier Ltd.
   ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1
                                                                                                                                                                                            PHARMACOLOGY
                                                                                                [PaCO2                         [PaCO2                        [PaCO2
                                                                                                YY ventilatory response to     YBronchial tone               YBronchial tone
  66
                                                  Table 4
                                                             PHARMACOLOGY
in the gas phase at a temperature below its critical temperature.          aneroid bellows which contract with a reduction in temperature
Nitrous oxide and xenon are gases at ambient temperature and               to obstruct the bypass channel. The vaporizer is also constructed
do not need vaporization prior to delivery.                                of materials of high specific heat capacity and thermal conduc-
    When delivering vapours, a vaporizer is needed to prevent              tivity which act as a heat sink.6
delivery of excessively high concentrations of anaesthetic at                  The way atmospheric pressure affects vaporizer use can be
room temperature and pressure. The saturated vapour pressure               considered when using them at altitude. At varying altitudes, the
(SVP) of a volatile anaesthetic is the partial pressure exerted by         SVP does not change, so the partial pressure leaving the chamber
the vapour when in equilibrium with the liquid phase (equal                is the same. However, the atmospheric pressure does drop at
number of particles leaving and entering each phase). The SVP              higher altitudes, and rises at lower altitudes. Consequently, the
determines the concentration of vapour above the liquid anaes-             SVP becomes a greater proportion of the atmospheric pressure.
thetic agent. Without a vaporizer, at room temperature and                 So, if a vaporizer is set at 1%, at a higher altitude with an at-
pressure, desflurane (SVP of 88.2kPa at 20 C) can achieve a               mospheric pressure of 50.5kPa (compared with sea level atmo-
maximum concentration of 87% and sevoflurane (SVP of                       spheric pressure of 101kPa), it will deliver 2%. However, 1% of
22.7kPa at 20 C) achieves a concentration of 22.4%. Both these            101kPa is the same as 2% of 50.5kPa, therefore it will deliver the
concentrations are far too high for safe anaesthesia and need to           same partial pressure and as a result the settings would not need
be controlled with a vaporizer. Therefore, a higher SVP means a            to be changed.
lower gas flow is needed through the vaporizer to achieve an
appropriate concentration for anaesthesia.                                 Nitrous oxide
    A plenum vaporizer is the most commonly used device on
anaesthetic machines in theatre. It receives gas flow which it             Nitrous oxide is largely used to supplement general anaesthetic
splits into two streams with variable flows. One stream is                 and is commonly used as a potent analgesic in labour using
directed through a chamber with the vapour in and is fully                 Entonox (50:50 mixture of oxygen and nitrous oxide). It is a
saturated with the agent, and the other larger stream is directed          colourless gas that is non-flammable. It is stored often as size E
through a bypass chamber. The proportions of flow through each             cylinders on anaesthetic machines (1800 L) or size J cylinders on
chamber are determined by the dial on the vaporizer. The two               manifolds (18,000 L).
unidirectional streams then join, and an accurate concentration                It is produced by heating ammonium nitrate to 170e240 C;
of anaesthetic is formed to be delivered to the patient. To ensure         however, contaminants are produced when the thermal decom-
a precise concentration of anaesthetic is supplied, the vaporizing         position is not performed accurately. These contaminants include
chamber needs to saturate the gas fully, even at higher gas flows.         ammonia, nitrogen dioxide and nitric oxide which can cause
This is achieved with metal or Teflon wicks which act via                  airway irritation and fibrosis after longer periods of exposure.
capillary action to generate a large surface area for the gas to           These toxic gases are removed by the manufacturers.
ensure full saturation, and baffles which direct the gas onto the              It is stored in cylinders at a gauge pressure of 4400 kPa in
surface of the liquid.                                                     French blue cylinders as a liquid with its vapour phase above. As
    Vaporizers are specific to each agents SVP to enable an ac-            a result, the way the cylinders are filled need to be carefully
curate concentration to be delivered. Varying SVPs of agents at a          managed. The filling ratio is the weight of the fluid in the cylinder
given temperature mean that different concentrations would be              divided by the weight of fluid which would fill the cylinder. In
delivered for a given dialled percentage if volatiles were swapped         the UK, this ratio should equal 0.75. Liquid is less compressible
between vaporizers. Desflurane has an SVP at 20 C of 88kPa and            than gas, so to prevent large rises in pressure with temperature
a boiling point of 23.5 C. This means it needs a particular               rises, the cylinders should not be filled entirely with liquid.
vaporizer to avoid the high concentrations that passive vapor-                 The pressure in the cylinder does not indicate how well filled
ization would cause with even small rises of room temperature.             it is. Its critical temperature (the temperature above which the
Therefore, desflurane uses the Ohmeda TEC6 vaporizers with a               substance cannot be liquified no matter how much pressure has
heating element to maintain a temperature of 39 C. This gives a           been applied) is 36.5 C (above room temperature). When nitrous
stable SVP of 200kPa, which allows an accurate concentration of            oxide is used, the vapour is released first from the cylinder, and
desflurane to be administered to the patient with varying con-             then liquid inside the cylinder vaporizes to replace what has been
ditions in theatre.                                                        used. Therefore, the pressure in the cylinder remains constant
    The latent heat of vaporization is an issue which is compen-           until the liquid has all been used.1
sated for in vaporizers. It is defined as the energy needed to                 Use of nitrous oxide has numerous potential side effects.1
convert a substance from a liquid to a gaseous state. This energy               Nausea due to bowel distention and opioid receptor
is taken from surroundings, including the vaporizer itself. As a                   activation.
result, the vaporizer gets colder as more liquid vaporizes. If this             Diffusion into air filled cavities. It has a high diffusion
was not compensated for, the agent’s SVP would drop, reducing                      capacity which means it quickly enters cavities such as the
the concentration of volatile released from the vaporizer. There                   lungs (causing a pneumothorax to expand) or the middle
are various temperature-controlled valves which alter the ratio of                 ear (which can cause damage after inner ear surgery).
gas split between the bypass and vapour chamber when the                        Bone marrow suppression and neurotoxicity. The cobalt
temperature does change, to prevent any fluctuations in anaes-                     ion in vitamin B12 is oxidized by nitrous oxide and is
thetic delivery. These include bimetallic strips with two metals of                therefore unable to be used as a co-factor for methionine
different expansion coefficients, altering the shape of the strip                  synthetase. This can cause megaloblastic anaemia and
allows changes to the split of gas between the chambers, or                        agranulocytosis by preventing tetrahydrofolate production.
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1                          67                                              Ó 2021 Published by Elsevier Ltd.
                                                             PHARMACOLOGY
       Prolonged exposure causes a neurological condition like                A GWP depends on two factors. The first factor is the way the
       subacute combined degeneration of the cord by prevention            gas changes the solar energy irradiance on the earth’s atmo-
       of myelin production.                                               sphere with a change in its concentration. The second is the at-
    Teratogenic e proven only in rat models.                              mospheric lifetime of the gas, which depends on how quickly it is
    Diffusion hypoxia can occur on waking. This phenomenon                broken down and is determined by its chemical structure. CeF is
       is due to the concentration effect happening in reverse. The        the strongest bond (followed by CeH, CeCl and CeBr), so at-
       volume of nitrous oxide entering the alveolus from the              mospheric OH displaces F less easily. Therefore, the different
       blood is greater than the amount of nitrogen entering the           agents have differing atmospheric lifespans. The IPCC (the
       pulmonary circulation. Therefore, the alveolar gases are            leading authority on climate change) uses the GWP100, which is
       therefore diluted, but using 100% oxygen on waking pre-             the GWP over 100 years to compare the long-term effects of the
       vents the hypoxia occurring.                                        agents.
    It can induce respiratory depression by reducing tidal                   Carbon dioxide has a GWP100 of 1, and carbon dioxide
       volume and reduce cardiac contractility with a mild direct          equivalency (CO2e) is used as the reference for other gases
       effect.                                                             relative to CO2. Anaesthetic gases have 100e1000 times higher
    Release of nitrous oxide contributes to global warming.               warming potential than carbon dioxide. The GWP100 of des-
    Increases pulmonary vascular resistance-caution in pul-               flurane is 2540, sevoflurane is 130, isoflurane is 510, and nitrous
       monary hypertension.                                                oxide is 265. Obtaining a MAC of 1.0 over a period of 1 hour at
    Increases cerebral metabolic rate, cerebral blood flow and            1 L/min fresh gas flow is equivalent to driving 4 miles in a car
       consequently intracranial pressure.                                 using sevoflurane, 8 miles with isoflurane and a massive 190
   The concentration and second gas effect have been described             miles with desflurane. It is also worth noting that using N2O with
earlier in the article.                                                    sevoflurane leads to a significant worsening of the effect on the
                                                                           carbon footprint with your anaesthetic. A MAC of 0.6 N2O, with a
Xenon                                                                      0.4 MAC of sevoflurane, over 1 hour with fresh gas flows of
                                                                           1 L/min is equivalent to driving 62 miles.7
Xenon is an inert and odourless gas which has a very fast onset
                                                                              Therefore, the most important measures to reduce the carbon
and offset of anaesthesia due to its low blood:gas partition co-
                                                                           footprint of an anaesthetic is to avoid desflurane and nitrous
efficient. It is extremely expensive to produce by fractional
                                                                           oxide, use low flows on the anaesthetic machine and use regional
distillation of air and anaesthetic equipment is rarely calibrated
for xenon, which both limit its usage. It is not metabolized and is
                                                                           or total intravenous anaesthetic if you are able.                A
excreted unchanged via the lungs. It is non-toxic, not flammable,
and is non-irritant to the airway.
                                                                           REFERENCES
    Xenon has analgesic properties and can induce a degree of
                                                                           1 Peck T, Hill S. Pharmacology for anaesthesia and intensive care.
muscle relaxation at higher concentrations. It has minimal car-
                                                                             Cambridge: Cambridge University Press, 2017; 106e21.
diovascular effects (small decrease in heart rate only) and min-
                                                                           2 Khan KS, Hayes I, Buggy DJ. Pharmacology of anaesthetic agents
imal respiratory effects (slows respiratory rate slightly, but
                                                                             II: inhalation anaesthetic agents. Cont Educ Anaesth Crit Care Pain
increase tidal volume to compensate). It does not cause malig-
                                                                             2014; 14: 106e11.
nant hyperthermia.1
                                                                           3 Mccombe K, Wijayasiri L. the primary FRCA structured oral exam.
                                                                             Study guide 2. Boca Raton: CRC Press, 2016; 53e9.
Environment and volatile anaesthesia
                                                                           4 Leslie RA, Johnson EK, Goodwin APL. Dr Podcast scripts for the
The crisis of climate change is a huge threat to the health of the           primary FRCA. Cambridge: Cambridge University Press, 2014;
world, with greenhouse gas emissions causing a gradual increase              175e85. Repr.
in temperature with detrimental consequences. Inhalational                 5 Langton JA, Hutton A. Respiratory gas analysis. Cont Educ
anaesthetic agent use has been targeted, as it is known to                   Anaesth Crit Care Pain 2009; 9: 19e23 [online]. Available at: https://
generate by-products which have a direct effect on global                    academic.oup.com/bjaed/article/9/1/19/465989.
warming. Anaesthetic gases represent 5% of the carbon footprint            6 Leslie RA, Johnson EK, Goodwin APL. Dr Podcast scripts for the
of the whole NHS.8                                                           primary FRCA. Cambridge: Cambridge University Press, 2014;
   Production of ‘infrared active’ molecules is key to global                304e9. Repr.
warming. ‘Infrared active’ means that the gas molecules vibrate            7 McGain F, Muret J, Lawson C, Sherman JD. Environmental sus-
and change their shape in response to infrared radiation. This               tainability within anaesthesia and critical care continuing education
happens to all heteronuclear gases (gases with two or more                   in anaesthesia critical care & pain. 2020 [online]. Available at:
atoms e.g. CO2) and some homonuclear molecules (e.g. O3                      https://bjanaesthesia.org/article/S0007-0912(20)30547-X/pdf.
ozone). These molecules absorb and emit infrared radiation                 8 Charlesworth M, Swinton F. Anaesthetic gases, climate change,
leading to the temperature changes. These changes caused by a                and sustainable practice. The Lancet Planetary Health 2017; 1:
gas are described in terms of its ‘global warming potential’                 e216e7 [online] Available at: https://www.thelancet.com/journals/
(GWP).                                                                       lanplh/article/PIIS2542-5196(17)30040-2/fulltext.
ANAESTHESIA AND INTENSIVE CARE MEDICINE 23:1 68 Ó 2021 Published by Elsevier Ltd.