Rietveld Refinement
Rietveld Refinement
• lattice parameters
• background parameters
Factors affecting peak intensities
incident beam
absorbing substance
fluorescent
x-rays heat
transmitted electrons
beam
scattered x-rays
(coherent & incoherent)
Scattering considerations
Note: exp(πi) = -1
exp(2πi) = +1
exp(nπi) = exp(-nπi)
Multiplicity (M)
In a cubic structure (a = b = c)
cubic
tetragonal
(100) (110) (111)
Lorentz factor
Thermal vibrations
1. Unit cell expansion causes changes in the 2θ positions.
2. Decrease in the intensities of diffracted lines.
3. Increase in the intensity of background scattering.
Lorentzian
−1
2 4 2
I i ,k = 1 +
πH k H k2
(2θi − 2θk )
psuedo-Voigt
Ii,k = η Li,k + (1- η) Gi,k
Crystallite Size (τ)
• ZrO2
• Space Group Fm-3m (225) site occupancy factor,
• Lattice Parameter a=5.11 sof, is equivalent to N
in the Fhkl equation
Zr 4a 0 0 0 1.14 1
j =1
RF =
∑ I K (' obs' ) 1/ 2 − I K (' calc' ) 1/ 2
R-structure factor
∑I
1/ 2
K (' obs')
RB =
∑ I K (' obs') − I K (' calc')
R-Bragg factor
∑I K (' obs' )
Rp =
∑ yi (obs) − yi (calc)
R-pattern
∑ y (obs)
i
∑ wi ( yi (obs) − yi (calc)) 2
1/ 2
2
Rwp Reduced Chi-squared
χ =
2
Rexp
• Missed peak
• High background
2000
0
4000 Quartz
3000
2000
1000
0
Cristobalite
4000
2000
0
20 30 40 50
Position [°2Theta] (Copper (Cu))
• The three X-ray scattering patterns above were produced by three chemically identical
forms SiO2
• Crystalline materials like quartz and cristobalite produce X-ray diffraction patterns
– Quartz and cristobalite have two different crystal structures
– The Si and O atoms are arranged differently, but both have long-range atomic order
– The difference in their crystal structure is reflected in their different diffraction patterns
• The amorphous glass does not have long-range atomic order and therefore produces
only broad scattering features
Diffraction occurs when light is scattered by a periodic array with
long-range order, producing constructive interference at
specific angles.
• Amorphous materials like glass do not have a periodic array with long-range order, so
they do not produce a diffraction pattern. Their X-ray scattering pattern features broad,
poorly defined amorphous ‘humps’.
Crystalline materials are characterized by the long-
range orderly periodic arrangements of atoms.
• The unit cell is the basic repeating unit that defines the crystal structure.
– The unit cell contains the symmetry elements required to uniquely define the
crystal structure.
– The unit cell might contain more than one molecule:
• for example, the quartz unit cell contains 3 complete molecules of SiO2.
– The crystal system describes the shape of the unit cell Crystal System: hexagonal
– The lattice parameters describe the size of the unit cell Lattice Parameters:
4.9134 x 4.9134 x 5.4052 Å
(90 x 90 x 120°)
• The unit cell repeats in all dimensions to fill space and produce the
macroscopic grains or crystals of the material
The diffraction pattern is a product of the unique
crystal structure of a material
Quartz
8000
6000
Quartz
4000
2000
0
Cristobalite
8000
6000
4000
2000 Cristobalite
0
20 30 40 50 60
Position [°2Theta] (Copper (Cu))
112
Calculated_Profile_00-005-0490
20
110
102
200
111
201
10
003
0
35 40 45 50
Position [°2Theta] (Copper (Cu))
Bragg’s Law
2d hkl sin q
j 1
• The structure factor Fhkl sums the result of scattering from all of the
atoms in the unit cell to form a diffraction peak from the (hkl) planes
of atoms
• The amplitude of scattered light is determined by:
– where the atoms are on the atomic planes
• this is expressed by the fractional coordinates xj yj zj
– what atoms are on the atomic planes
• the scattering factor fj quantifies the efficiency of X-ray scattering at any
angle by the group of electrons in each atom
– The scattering factor is equal to the number of electrons around the atom at 0° θ,
the drops off as θ increases
• Nj is the fraction of every equivalent position that is occupied by atom j
Bragg’s law provides a simplistic model to understand
what conditions are required for diffraction.
s
[hkl]
2d hkl sin q q q
dhkl dhkl
• For parallel planes of atoms, with a space dhkl between the planes, constructive
interference only occurs when Bragg’s law is satisfied.
– In our diffractometers, the X-ray wavelength is fixed.
– A family of planes produces a diffraction peak only at a specific angle 2q.
• Additionally, the plane normal [hkl] must be parallel to the diffraction vector s
– Plane normal [hkl]: the direction perpendicular to a plane of atoms
– Diffraction vector s: the vector that bisects the angle between the incident and
diffracted beam
Many powder diffractometers use the Bragg-Brentano
parafocusing geometry.
Detector
s
X-ray
tube
w
2q
• The incident angle, w, is defined between the X-ray source and the sample.
• The diffraction angle, 2q, is defined between the incident beam and the detector.
• The incident angle w is always ½ of the detector angle 2q .
– In a q:2q instrument (e.g. Rigaku H3R), the tube is fixed, the sample rotates at q °/min and the
detector rotates at 2q °/min.
– In a q:q instrument (e.g. PANalytical X’Pert Pro), the sample is fixed and the tube rotates at a rate -q
°/min and the detector rotates at a rate of q °/min.
• In the Bragg-Brentano geometry, the diffraction vector (s) is always normal to the
surface of the sample.
– The diffraction vector is the vector that bisects the angle between the incident and scattered beam
A single crystal specimen in a Bragg-Brentano diffractometer would
produce only one family of peaks in the diffraction pattern.
2q
At 20.6 °2q, Bragg’s law The (110) planes would diffract at 29.3 The (200) planes are parallel to the (100)
fulfilled for the (100) planes, °2q; however, they are not properly planes. Therefore, they also diffract for this
producing a diffraction peak. aligned to produce a diffraction peak crystal. Since d200 is ½ d100, they appear at
(the perpendicular to those planes does 42 °2q.
not bisect the incident and diffracted
beams). Only background is observed.
A polycrystalline sample should contain thousands of crystallites.
Therefore, all possible diffraction peaks should be observed.
[110] [200]
[100]
s s
s
2q 2q 2q
• For every set of planes, there will be a small percentage of crystallites that are properly
oriented to diffract (the plane perpendicular bisects the incident and diffracted beams).
• Basic assumptions of powder diffraction are that for every set of planes there is an equal
number of crystallites that will diffract and that there is a statistically relevant number of
crystallites, not just one or two.
Powder diffraction is more aptly named polycrystalline
diffraction
• Samples can be powder, sintered pellets, coatings on substrates, engine blocks...
• The ideal “powder” sample contains tens of thousands of randomly oriented
crystallites
– Every diffraction peak is the product of X-rays scattering from an equal
number of crystallites
– Only a small fraction of the crystallites in the specimen actually contribute to
the measured diffraction pattern
• XRPD is a somewhat inefficient measurement technique
• Irradiating a larger volume of material can help ensure that a statistically relevant
number of grains contribute to the diffraction pattern
– Small sample quantities pose a problem because the sample size limits the
number of crystallites that can contribute to the measurement
X-rays are scattered in a sphere around the sample
• Each diffraction peak is actually a Debye diffraction cone produced by the tens of
thousands of randomly oriented crystallites in an ideal sample.
– A cone along the sphere corresponds to a single Bragg angle 2theta
• The linear diffraction pattern is formed as the detector scans along an arc that
intersects each Debye cone at a single point
• Only a small fraction of scattered X-rays are observed by the detector.
X-Ray Powder Diffraction (XRPD) is a somewhat
inefficient measurement technique
• The absolute intensity, i.e. the number of X rays observed in a given peak,
can vary due to instrumental and experimental parameters.
– The relative intensities of the diffraction peaks should be instrument
independent.
• To calculate relative intensity, divide the absolute intensity of every peak by the
absolute intensity of the most intense peak, and then convert to a percentage. The
most intense peak of a phase is therefore always called the “100% peak”.
– Peak areas are much more reliable than peak heights as a measure of
intensity.
Powder diffraction data consists of a record of photon
intensity versus detector angle 2q.
• Diffraction data can be reduced to a list of peak positions and intensities
– Each dhkl corresponds to a family of atomic planes {hkl}
– individual planes cannot be resolved- this is a limitation of powder diffraction versus
single crystal diffraction
Raw Data Reduced dI list
Counts
Position Intensity DEMO08
hkl dhkl (Å) Relative
3600
[°2q] [cts] Intensity
25.2000 372.0000 (%)
25.2400 460.0000
25.2800 576.0000 1600
{012} 3.4935 49.8
25.3200 752.0000 {104} 2.5583 85.8
25.3600 1088.0000
25.4000 1488.0000
{110} 2.3852 36.1
400
25.4400 1892.0000 {006} 2.1701 1.9
25.4800 2104.0000
25.5200 1720.0000 {113} 2.0903 100.0
25.5600 1216.0000 0
25 30 35 40 45 {202} 1.9680 1.4
25.6000 732.0000 Position [°2Theta] (Copper (Cu))
25.6400 456.0000
25.6800 380.0000
25.7200 328.0000
Applications of XRPD
You can use XRD to determine
• Phase Composition of a Sample
– Quantitative Phase Analysis: determine the relative amounts of phases in a
mixture by referencing the relative peak intensities
• Unit cell lattice parameters and Bravais lattice symmetry
– Index peak positions
– Lattice parameters can vary as a function of, and therefore give you
information about, alloying, doping, solid solutions, strains, etc.
• Residual Strain (macrostrain)
• Crystal Structure
– By Rietveld refinement of the entire diffraction pattern
• Epitaxy/Texture/Orientation
• Crystallite Size and Microstrain
– Indicated by peak broadening
– Other defects (stacking faults, etc.) can be measured by analysis of peak
shapes and peak width
• We have in-situ capabilities, too (evaluate all properties above as a
function of time, temperature, and gas environment)
Phase Identification
• The diffraction pattern for every phase is as unique as your fingerprint
– Phases with the same chemical composition can have drastically different
diffraction patterns.
– Use the position and relative intensity of a series of peaks to match
experimental data to the reference patterns in the database
The diffraction pattern of a mixture is a simple sum of
the scattering from each component phase
Databases such as the Powder Diffraction File (PDF) contain dI
lists for thousands of crystalline phases.
• The PDF contains over 300,000 diffraction patterns.
• Modern computer programs can help you determine what phases are
present in your sample by quickly comparing your diffraction data to all of
the patterns in the database.
• The PDF card for an entry contains a lot of useful information, including
literature references.
Quantitative Phase Analysis
..
– must meet the constant volume assumption (see
I(phase a)/I(phase b)
later slides)
50
• The ratio of peak intensities varies linearly as a 40
function of weight fractions for any two phases
in a mixture 30
𝐼α 𝑋
– = K * 𝑋α 20
𝐼β β
• The pattern shown above contains equal amounts of TiO2 and Al2O3
• The TiO2 pattern is more intense because TiO2 diffracts X-rays more efficiently
With proper calibration, you can calculate the amount of each phase present in the sample
Unit Cell Lattice Parameter Refinement
K
B2q
Intensity (a.u.)
L cos q
23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41
2q (deg.)
Preferred Orientation (texture)
8.0
(311)
Intensity(Counts)
6.0 (200)
(220)
4.0
2.0
(222)
(400)
x10 3
40 50 60 70 80 90 100
Two-Theta (deg)
Non-ideal samples: Texture (i.e. preferred
crystallographic orientation)
300
250
Intensity(Counts)
200
150
100
50
0 (111)
(221) JCS#98> CaCO3 - Aragonite
(021) (012)
(102) (112) (220) (041) (132) (113)
(002) (121) (211) (040) (212) (222) (042)
25 30 35 40 45 50 55
Two-Theta (deg)
AC CURRENT
The wavelength of X rays is determined by the anode of
the X-ray source.
• Electrons from the filament strike the target anode, producing characteristic
radiation via the photoelectric effect.
• The anode material determines the wavelengths of characteristic radiation.
• While we would prefer a monochromatic source, the X-ray beam actually
consists of several characteristic wavelengths of X rays.
K
L
M
Spectral Contamination in Diffraction Patterns
Ka1 Ka1
Ka2
Ka2 Ka1
Ka2
W La1
Kb
• The Ka1 & Ka2 doublet will almost always be present
– Very expensive optics can remove the Ka2 line
– Ka1 & Ka2 overlap heavily at low angles and are more separated
at high angles
• W lines form as the tube ages: the W filament contaminates
the target anode and becomes a new X-ray source
• W and Kb lines can be removed with optics
Monochromators remove unwanted wavelengths of radiation
from the incident or diffracted X-ray beam.
Suppression
below the target material on
the periodic table
– For example, when using Cu
radiation
• Cu K-alpha = 1.541 Å
• Cu K-beta= 1.387 Å
• The Ni absorption edge= 1.488 Å
– The Ni absorption of Cu radiation
is:
Cu Ka
Cu Kb
W La
• 50% of Cu K-alpha
• 99% of Cu K-beta Wavelength
H He
Li Be
Fluorescence B C N O F Ne
Na Mg Al Si P S Cl Ar
K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe
Cs Ba L Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
Fr Ra A
• Some atoms absorb incident X-rays and fluoresce them as X-rays of a different
wavelength
– The absorption of X-rays decreases the diffracted signal
– The fluoresced X-rays increase the background noise
• The increased background noise from fluoresced X-rays can be removed by using:
– a diffracted-beam monochromator
– an energy sensitive detector
• The diffracted beam signal can only be increased by using a different wavelength of
radiation
• The most problematic materials are those two and three below the target material:
– For Cu, the elements that fluoresce the most are Fe and Co
The X-ray Shutter is the most important safety device
on a diffractometer
H2O In H2O Out
Cu ANODE
Be
window
SAFETY SHUTTERS
The X-ray beam produced by the X-ray tube is divergent.
Incident-beam optics are used to limit this divergence
2d hkl sin q
• X Rays from an X-ray tube are:
– divergent
– contain multiple characteristic wavelengths as well as Bremmsstrahlung radiation
• neither of these conditions suit our ability to use X rays for analysis
– the divergence means that instead of a single incident angle q, the sample is actually
illuminated by photons with a range of incident angles.
– the spectral contamination means that the smaple does not diffract a single wavelength
of radiation, but rather several wavelengths of radiation.
• Consequently, a single set of crystallographic planes will produce several diffraction peaks
instead of one diffraction peak.
• Optics are used to:
– limit divergence of the X-ray beam
– refocus X rays into parallel paths
– remove unwanted wavelengths
Most of our powder diffractometers use the Bragg-
Brentano parafocusing geometry.
• A point detector and sample are
moved so that the detector is always
at 2q and the sample surface is
always at q to the incident X-ray
beam.
• In the parafocusing arrangement, the
incident- and diffracted-beam slits
move on a circle that is centered on
the sample. Divergent X rays from the
source hit the sample at different
points on its surface. During the
diffraction process the X rays are
refocused at the detector slit. F: the X-ray source
• This arrangement provides the best DS: the incident-beam divergence-limiting slit
SS: the Soller slit assembly
combination of intensity, peak shape, S: the sample
and angular resolution for the widest RS: the diffracted-beam receiving slit
number of samples. C: the monochromator crystal
AS: the anti-scatter slit
Divergence slits are used to limit the divergence of the
incident X-ray beam.
• The slits block X-rays that have too great a
divergence.
• The size of the divergence slit influences
peak intensity and peak shapes.
• Narrow divergence slits:
– reduce the intensity of the X-ray beam
– reduce the length of the X-ray beam hitting
the sample
– produce sharper peaks
• the instrumental resolution is improved so
that closely spaced peaks can be resolved.
One by-product of the beam divergence is that the length of the beam
illuminating the sample becomes smaller as the incident angle
becomes larger.
t
size. e
m
m 10.00
• The width of the beam is d 0.5°DS
)
s Detector
X-ray
tube
w 2q
• Parallel beam optics do NOT require that the incident angle w is always ½
of the detector angle 2q .
• A coupled scan with parallel-beam optics will maintain the diffraction
vector in a constant relationship to the sample.
– If w is always ½ of 2q then the diffraction vector (s) is always normal to the surface of
the sample.
– If w = ½ 2q + τ, then s will be always tilted by τ away from the vertical position.
• That direction will not change as long as both omega and 2theta change in a coupled
relationship so that w is always equal to ½ 2q + τ
Parallel beam optics allow for the possibility of grazing
incidence X-ray diffraction (GIXD)
Detector
s
X-ray
tube
w 2q
• The incident angle, w, is set to a very shallow angle (between 0.2 and 5 deg).
– This causes the X-rays to be focused in the surface of the sample, limiting the penetration depth of the X-
rays
• Only the detector moves during data collection
– The value τ is changing during the scan (where τ = ½*2q - w)
– As a consequence, the diffraction vector (s) is changing its direction during the scan
• Remember the diffraction only comes from crystallites in which dhkl is parallel to s
– Therefore, the direction being probed in the sample changes
– This is perfectly ok for ideal samples with randomly oriented grains; however, for samples with preferred
orientation this will cause a problem.
• Regular GIXD will constrain the X-ray beam in the top few microns of the surface
• IP-GIXD can be configued to constrain diffraction to the top 10-20 nm of the surface.
Other optics:
• limit divergence of the X-ray beam
– Divergence limiting slits Parallel Plate Collimator & Soller
– Parallel plate collimators Slits block divergent X-rays, but
do not restrict beam size like a
– Soller slits divergent slit
• refocus X rays into parallel paths
– “parallel-beam optics”
– parabolic mirrors and capillary lenses
– focusing mirrors and lenses
• remove unwanted wavelengths
– monochromators
– Kb filters
• point detectors
– observe one point of space at a time
• slow, but compatible with most/all optics
– scintillation and gas proportional detectors count all photons, within an
energy window, that hit them
– Si(Li) detectors can electronically analyze or filter wavelengths
• position sensitive detectors
– linear PSDs observe all photons scattered along a line from 2 to 10° long
– 2D area detectors observe all photons scattered along a conic section
– gas proportional (gas on wire; microgap anodes)
• limited resolution, issues with deadtime and saturation
– CCD
• limited in size, expensive
– solid state real-time multiple semiconductor strips
• high speed with high resolution, robust
Area (2D) Diffraction allows us to image complete or
incomplete (spotty) Debye diffraction rings
the area observed by a linear detector the area observed by a linear detector
• a flat plate sample for XRPD should have a smooth flat surface
– if the surface is not smooth and flat, X-ray absorption may reduce the
intensity of low angle peaks
– parallel-beam optics can be used to analyze samples with odd shapes
or rough surfaces
• Densely packed
• Randomly oriented grains/crystallites
• Grain size less than 10 microns
– So that there are tens of thousands of grains irradiated by the X-ray
beam
• ‘Infinitely’ thick
• homogeneous
Preparing a powder specimen
3600
1600
400
0
20 30 40 50
Position [°2Theta] (Copper (Cu))
300
250
Intensity(Counts)
200
150
100
50
0 (111)
(221) JCS#98> CaCO3 - Aragonite
(021) (012)
(102) (112) (220) (041) (132) (113)
(002) (121) (211) (040) (212) (222) (042)
25 30 35 40 45 50 55
Two-Theta (deg)
• the area of your sample that is illuminated by the X-ray beam varies
as a function of:
– incident angle of X rays
– divergence angle of the X rays
• at low angles, the beam might be wider than your sample
– “beam spill-off”
• This will cause problems if you sample is not homogeneous
185mm Radius Goniometer, XRPD
40.00
35.00
L
I
e 30.00
r
n
r
g 25.00
a 2°DS
t
d
h 20.00
i
a 15.00
1°DS
(
t
m
e
m 10.00
d 0.5°DS
)
5.00
0.15°DS
0.00
0 20 40 60 80 100
2Theta (deg)
Penetration Depth of X-Rays
sample
• Poor counting statistics
– The sample is not made up of thousands
of randomly oriented crystallites, as
assumed by most analysis techniques
– The sample might have large grain sizes
• Produces ‘random’ peak intensities and/or
spotty diffraction peaks
• Axial divergence
– Due to divergence of the X-ray beam in plane with the sample
– creates asymmetric broadening of the peak toward low 2theta angles
– Creates peak shift: negative below 90° 2theta and positive above 90°
– Reduced by Soller slits and/or capillary lenses
Counts
0.04rad Soller Slits
0.04rad incident Soller slit and 0.02rad detector Soller Slit
0.02rad Soller Slits
60000
40000
20000
3 4 5 6
Position [°2Theta] (Copper (Cu))
Grazing Incident Angle Diffraction (GIXD)
• Used to determine:
– crystal structure
– orientation
– degree of crystalline perfection/imperfections (twinning, mosaicity,
etc.)
• Sample is illuminated with monochromatic radiation
– The sample axis, phi, and the goniometer axes omega and 2theta are
rotated to capture diffraction spots from at least one hemisphere
– Easier to index and solve the crystal structure because it diffraction
peak is uniquely resolved
Available Free Software
The X-ray diffraction pattern is a fingerprint that lets you figure out what is in your sample.
The diffraction pattern of a mixture is a simple sum of
the diffraction patterns of each individual phase.
Quartz Mixture
Cristobalite
Glass
0
15 20 25 30 35 40
15 20 25 30 35 40
Position [°2Theta] (Cu K-alpha) Position [°2Theta] (Copper (Cu))
K-alpha2
K-alpha1
K-alpha2
• The k-alpha1 and k-alpha2 peak doublets are further apart at higher
angles 2theta
• The k-alpha1 peaks always as twice the intensity of the k-alpha2
• At low angles 2theta, you might not observe a distinct second peak
The experimental data should contain all major peaks
listed in the reference pattern
This is an example of a bad match between the data and the reference pattern
The X-ray diffraction pattern is a sum of the diffraction
patterns produced by each phase in a mixture
Counts
Rutile, syn;
Hematite, syn;
Rutile, syn;
Hematite, syn;
3600
Anatase, syn;
Hematite, syn;
Rutile, syn;
Hematite, syn;
1600
Anatase, syn;
syn;
Rutile, syn;
Anatase, syn;
Anatase, syn;
syn;
Hematite, syn;
Rutile,
Hematite,
400
0
25 30 35 40 45
Position [°2Theta] (Copper (Cu))
• The pattern shown above contains equal amounts of TiO2 and Al2O3
• The TiO2 pattern is more intense because TiO2 diffracts X-rays more efficiently
With proper calibration, you can calculate the amount of each phase present in the sample
Diffraction peak broadening may contain information
about the sample microstructure
• Peak broadening may indicate:
– Smaller crystallite size in nanocrystalline materials
– More stacking faults, microstrain, and other defects in the crystal structure
– An inhomogeneous composition in a solid solution or alloy
• However, different instrument configurations can change the peak width, too
These patterns show the difference between bulk These patterns show the difference between the
ceria (blue) and nanocrystalline ceria (red) exact same sample run on two different instruments.
When evaluating peak broadening, the instrument profile must be considered.
Quantitative Analysis of XRD Data
Diffraction peak positions can be used to calculated
unit cell dimensions
Counts
25.321 deg
d= 3.5145 Å
24.179 deg
1000 d= 3.6779 Å
500
0
23 24 25 26
Position [°2Theta] (Copper (Cu))
500
0
23 24 25 26
Position [°2Theta] (Copper (Cu))
• Nanocrystallite size will produce peak broadening that can be quantified
– Once the crystallite size is larger than a maximum limit, the peak broadening cannot be quantified.
This creates an upper limit to the crystallite size that can be calculated.
– The upper limit depends on the resolution of the diffractometer.
• Non-uniform lattice strain and defects will also cause peak broadening
• Careful evaluation is required to separate all of the different potential causes of peak broadening
The weight fraction of each phase can be calculated if
the calibration constant is known
Counts
1600
400
0
25 30 35 40 45
Position [°2Theta] (Copper (Cu))
• Peak search
– Analysis of the second derivative of diffraction data is used to identify
likely diffraction peaks
– Peak information is extracted by fitting a parabola around a minimum
in the second derivative
– This method is fast but the peak information lacks precision
• Profile fitting
– Each diffraction peak is fit independently with an equation
– The sum of the profile fits recreates the experimental data
– Peak information is extracted from the profile fit equation
– This method provides the most precise peak information
Profile Fitting produces precise peak positions, widths,
heights, and areas with statistically valid estimates
Intensity (a.u.)
• The profile function models the
mixture of Gaussian and
Lorentzian shapes that are typical
of diffraction data
– fit background, usually as a
polynomial function
• this helps to separate intensity in
28.5 29.0 29.5 30.0
peak tails from background 2θ (deg.)
• This peak list itself does not tell you anything about the
sample
– Additional analysis must be done on the peak list to extract
information
• From the peak list you can determine:
– Phase composition: by comparison to a database of reference
patterns
– Semi-quantitative phase composition: calculated from peak
intensities for different phases
– Unit cell lattice parameters: calculated from peak positions
– Crystal system: determined by indexing observed peaks and
systematic absences
– Crystallite size and microstrain: calculated from peak widths and/or
shapes
– A number of engineering indexes are also calculated from peak list
information
Full pattern fitting methods use different models to
produce a calculated pattern
Particle-like properties:
Wave-like properties:
1 h
E = mn v 2 = k BT λ=
2 mn v
Neutron type Energy (meV) Temperature (K) Wavelength (Å)
“Cold” 0.1 – 10 1 – 120 4 - 30
“Thermal” 5 – 100 60 – 1000 1–4
“Hot” 100 – 500 1000 – 6000 0.4 – 1
•For diffraction experiments thermal neutrons are used. Velocity is of the order of ~2000
ms-1 for “room temperature” neutrons (photons 3×108 ms-1).
Interactions of neutrons and X-rays with matter
www.ncnr.nist.gov
Elastic scattering of X-rays from electrons
Ratio of radiated electric field magnitude to incident electric field magnitude is:
Erad ( R, t ) ⎛ e2 ⎞ e i kR
= −⎜⎜ ⎟
2 ⎟
cosψ
E in ⎝ 4πε 0 mc ⎠ R Thomson scattering length r0 = 2.82 × 10-5 Å
(the “ability” of an electron to scatter X-rays)
Minus sign indicates that incident and radiated fields are 180°out of phase
Elastic scattering of neutrons from nuclei
www.ncnr.nist.gov
•Neutron-nucleus interaction involves very short-range forces (on the order of 10-15 m). A
metastable nucleus + neutron state is formed which then decays, re-emitting the neutron as a
spherical wave with a phase change of 180°
•Radius of nucleus is ~10-17 m – much smaller than wavelength of thermal neutrons (10-10 m),
thus can be considered “point-like”
Basics of diffraction
k k'
2θ rj
unit cell
F crystal
(Q ) = ∑ f j e iQ⋅r j
∑ e iQ ⋅ R n integer multiple of
2π when Bragg’s
rj Rn law is satisfied
lattice vectors -Rn
unit cell structure factor lattice sum
Neutron v X-ray diffraction: atomic number
•For neutrons there is no systematic trend in scattering length with atomic number- it depends
on the nucleus (isotope, nuclear spin). Scattering length is negative for some nuclei!
•Adjacent atoms in the periodic table often have very different neutron scattering lengths,
allowing them to be distinguished easily.
Absorption for thermal neutrons and 8keV X-rays
• Neutrons are
absorbed by nuclear
processes that destroy
the neutrons, emitting
secondary radiation
(α, β, or γ) as a result.
www.ncnr.nist.gov
Neutron diffraction – magnetic structure
F 2
hkl =F 2
Nuc ( hkl ) +q F 2 2
Mag ( hkl )
q = 1 − (ε ⋅ κ )
2 2
magnetic
⎛ eγ ⎞ 2
form factor
Magnetic scattering amplitude: pn = ⎜⎜ ⎟ gSf m
2 ⎟
(tabulated)
⎝ 2mc ⎠
electronic spin
electron mass
Landé g factor, usually ~2
Neutron diffraction – magnetic structure
Neutrons only ever see the components of the magnetization that are perpendicular to the scattering vector!
Introduction to magnetic symmetry
Paramagnet (T > TC)
n
intensity
n n
n
2θ
Ferromagnet (T < TC)
n n+m
intensity
2θ
Introduction to magnetic symmetry
Paramagnet (T > TN)
n
intensity
n n
n
2θ
Antiferromagnet (T < TN)
n n+m
intensity
m
n+m n
m Unit cell is larger
m
n New peaks appear
2θ
Example: multiferroic TbMn2O5
Neutron scattering facilities
www.veqter.co.uk
Neutron sources – nuclear reactor
•A steady supply of neutrons is produced by the 235U fission chain reaction (2.5 neutrons
per fission event, 1.5 are reabsorbed by the fuel).
•Neutrons are extracted from the core by neutron guide tubes and slowed down by a
moderator. A particular wavelength can then be selected using a crystal monochromator.
Broad spectrum
www.euronuclear.org
Wavelength selected by monochromator
Neutron sources – spallation (pulsed) source
•Protons are accelerated in a synchrotron ring and then collide with a heavy metal target,
which emits many subatomic particles including neutrons (more than 10 per proton).
http://www.isis.stfc.ac.uk
Spallation sources – time-of-flight diffraction technique
arrival time of neutron
at detector
λ ⎛ h ⎞⎛ 1 ⎞ ht hkl
d hkl = =⎜ ⎟⎜ ⎟=
2 sin θ ⎝ mv ⎠⎝ 2 sin θ ⎠ 2mL sin θ
+ There is often strong contrast in scattering between neighbouring elements (eg. can
distinguish Mn from Fe).
+ Light elements can give strong scattering eg. 2D, 12C, 14N, 16O.
+ Strong interaction with magnetic moments- can determine magnetic structures routinely.
- Neutron sources are much weaker than X-ray sources – in general large samples are needed.
- Some nuclei strongly absorb neutrons and cannot be probed eg. 10B, 113Cd, 157Gd.
- Some nuclei are almost transparent to neutrons and cannot easily be probed eg. 51V. Some
nuclei have strong incoherent scattering giving high background eg. 1H.
Neutron diffraction: summary