E40 Params and Trouble Shooting
E40 Params and Trouble Shooting
Type E 40
No. 8555-9201-140
Edition 0810
    The authors are always grateful for comments and suggestions for improving
    the documentation. They can be addressed to:
                       Note: • This symbol is not a safety precaution but rather a reference to infor-
                                mation which help to better understand the separator or plant compo-
                                nents and the processes.
1 Timer overview 7
4 Program logic 25
1 Timer overview
                                                                                                                                Active in
 Timer        Designation                     Action                                    Description
                                                                                                                           -0136- -0196- -91-
                                  Operating water valve V 3         The bowl sludge space is partial emptied by actu-
 T00     Partial ejection
                                  opens                             ating the operating water valve V 3. (Option)            X      X      X
                                  Operating water valve V 3         The bowl sludge space is emptied by actuating the
 T04     Total ejection
                                  opens                             operating water valve V 3.                               X      X      X
         Closing water impul-     Operating water valve V 3         Timer for safety water to keep the bowl closed
 T06
         se                       opens                             during separation.                                       X      X      X
                                  After timer is expired, product   After timer is expired, the standard alarms like
 T07     Alarm delay
                                  feed valve V 1 closes.            PAL1/PAH1, TAL/TAH, PAH2 ect. will be initiated.         X      X      X
T09 Waiting after filling Timer to stabilize the system after filling process. --- X X
T11 Alarm Bridging Alarms not activated. Timer to stabilize the process conditions! X X X
T14 Closing water interval Time interval between two closing water pulses. X X X
         Circulation valve
 T15
         close
                                  Circulation valve V 5 closes.     Timer in which the circulation valve is closed!          X     ---    ---
         Circulation valve        Circulation valve V 5 opens.      Timer in which the circulation valve is opened!
 T16
         open                     WMS Sensor is activated.          WMS Sensor Signal is checked.                            X     ---    ---
                                                                                                                               Active in
  Timer       Designation                   Action                                     Description
                                                                                                                          -0136- -0196- -91-
                                program.
          Water impulse by      Operating water valve V 3          Timer for closing the bowl after separator is on
 T19
          start                 opens                              operating speed.                                         X     X      X
 T20      Separation            Product feed valve V 1 opens       Timer for adjusting the separating time.                 X     X      X
                                In case of pressure drop (SMS Timer in which a system pressure drop is eva-
          Min separation time   PT 2) during T21 , triggering of luated as an alarm.
 T21
          with SMS              alarm "Minimum separating                                                                   X     ---    ---
                                time during SMS"                 Timer starts together with timer 20.
                                After delay time of T 25, an       Timer which triggers the alarm "Vibration stage 1"
 T25      Vibration 1 delay
                                ejection is initiated.             only when the signal is activated for the time T 25!     X     X      X
                                After delay time of T 26, the
                                safety water valve V 2 is
                                opened and the pneumatic           Timer which triggers the alarm "Vibration stage 2"
 T26      Vibration 2 delay
                                brake (option) actuated. The       when the signal is activated for the time T 26!          X     X      X
                                separator motor is shut down
                                immediately.
                                Water discharge valve V 6          Timer for sludge space monitoring. Only the SMS
 T33      SMS-Mode              closes.                            sensor PT 2 is activated. The sludge content of         ---    X      ---
                                Circulation valve V 5 closes.      the bowl is checked.
                                                                                                                              Active in
 Timer       Designation                   Action                                    Description
                                                                                                                         -0136- -0196- -91-
         Separator off after   After delay T 39, the separa-     Timer delay for separator motor stop. Function
 T39
         idle run              tor motor is switched off.        must be preselected in the program.                       X      X      X
                                                                 Timer to operate the sludge pump until the sludge
 T40     CU-Sludge pump On Sludge pump activated.
                                                                 tank is completely emptied.                               X      X      X
 T41     CU-Sludge pump Off Sludge pump not activated.           Time interval between two sludge pump starts.             X      X      X
 T42     Free
 T43     Free
                               Circulation valve V 5 closes.     Timer in which the function of the SMS sensor PT
                               Water discharge valve V 6         2 is tested. If T44 is elapsed without SMS Sensor
 T44     SMS-Sensor-Check
                               closes.                           PT 2 Signal, Alarm "SMS sensor failure during             X      X     ---
                               SMS sensor PT 2 is activated.     WMS-SMS check" is triggered.
 T45     Free
 T46     Free
                               When SMS Sensor PT 2 is           Capacity and machine type-dependent timer
                               activated while T 47 is run-      which determines the increse of water in the lube
                               ning, the water discharge         oil during WMS mode. The pressure increase is
 T47     Min.SMS Time
                               valve V 6 is opened for 3         evaluated as a function of T 47 over several             ---     X     ---
                               seconds and Counter 5             successive cycles which are defined via counter 5
                               counts the PT 2 signal.           "SMS time interval water content"
                                                                 Timer to fill filling water into the bowl. This timer
                                                                 runs before timer T61 and T62. Timer is required
 T60     Start Impulse         Filling water valve V 2 opens.
                                                                 for the automatic calculation of timers 01 and 02.        X      X     ---
                                                                 Timer is protected.
                                                                                                                                                                                                                                 2.1
                                                                                                                                                                                                  Separator-Type
                                                                                                                                 Text Display                   Remarks             OSE    OSE     OSE     OSE     OSE    OSE
                                                                                               Full Text
                                                                                                                      (P = Setpoint; A = actual value)                               5      10      20      40      80    120
                                                      Counter
                                                                                                                       C1 Flushing:
                                                     C1         Flushing               Flushing counter                                                                              5      5        5      5       5      5
                                                                                                                       P: 5
                                                                                                                       A: 0
                                                                                                                       C2 Partial Ejection:
                                                     C2         Partial Ejection       Partial ejection counter                                                                      5      5        5      5       5      5
                                                     (Option)                                                          P: 5
                                                                                                                       A: 0
                                                                                                                       C3 SMS Filling:
                                                     C3         SMS-Filling            SMS fillng counter                                                                            5      5        5      5       5      5
                                                                                                                       P: 3
                                                                                                                       A: 0
                                                                                                                       C4 Sensor Check:
                                                     C4         Sensor-Check           Sensor check counter                                                                         10     10       10     10      10     10
                                                     (Option)                                                          P: 10
                                                                                                                       A: 0
                                                                                                                                                                                                                                 Default settings Counter - Time records
                                                                                                                       C5 Water content
                                                     C5         C5 SMS Time interval   SMS Time interval counter
                                                                                                                       P: 5                                                          5      5        5      5       5      5
                                                     (Option)   Water cont             water content Warning
                                                                                                                       A: 0
                                                                                                                       Sludgep.-time
                                                                                       Time Record for                 Max. in      24 h      30 min
                                                     TR2        TR2 Sludgep. time      sludgepump time operation                                                                    30     30       30     40      50     60
                                                                                       in last TR3 (24h) hours.        Act. in      24 h      0 min      (Act. in = actuel 24h)
                                                                                                                       Last. in     24 h      0 min      (Last in = previous 24h)
                                                                                                         Active in:
  No.       Designation:        Status:                  Description:
                                                                                                    -0136- -0196- -91-
                                                                                                                                                                                                                                  2.3
                                                                                              Control Unit E40
                                                                Alarme                                         Display Text                  Remarks                                       Separator-Type
                                                                                       Full Text:
                                                                                                               Control Unit:
                                                     Temperature-                                                                                                                                  OSE
                                                     Monitoring                                                                                                               DO           MDO           LO        HFO
                                                                                                                                    Separation Temperature                 20 °C           40 °C     95 °C         98 °C
                                                                                                              TAH OIL FEED
                                                                                                              Temp. > max
                                                     TAH OIL FEED              Alarm for an increase of the
                                                                                                              Separation-Stop                                              30 °C           50 °C     105 °C        105 °C
                                                     Temp. > max               oil temperature.
                                                                                                              Alarm#06 press
                                                                                                              HELP
                                                                                                              TAL OIL FEED
                                                                                                              Temp. < min
                                                     TAL OIL FEED              Alarm for a decrease of the
                                                                                                              Separation-Stop                                              10 °C           30 °C     85 °C         90 °C
                                                     Temp. < min               oil temperature.
                                                                                                              Alarm#05 press
                                                                                                              HELP
                                                                                                                                                                                                                                  Default settings Temperature - Ejection
                                                     Ejection-
                                                                                                                                                                     OSE 5     OSE 10 OSE 20 OSE 40 OSE 80 OSE 120
                                                     Monitoring
                                                                                                              EJECTION
                                                                                                              MONITORING:
                                                                               Alarm for monitoring the
                                                                                                              Eject. current <       Value depending on the con-
                                                     Ejection monitoring       ejection after separation or
                                                                                                              Offset                 ditions of the power supply      0,5 A        0,5 A   0,5 A         3A   5A            8A
                                                     Eject. current < Offset   after SMS ejection.
                                                                                                              Separation-Stop        system and the clutch shoes!
                                                                               Current is monitored.
                                                                                                              Alarm#07 press
                                                                                                              HELP
                                                                                                              EJECTION
                                                                                                              MONITORING:
                                                                               Alarm for monitoring the
                                                     Ejection monitoring                                      Eject. speed < Off-    Value depending on the con-
                                                                               ejection after separation or
                                                     Eject. speed < Offset                                    set                    ditions of the power supply    100 U/min 100 U/min 100 U/min 100 U/min 100 U/min 100 U/min
                                                                               after SMS ejection.
                                                     (Option)                                                 Separation-Stop        system and the clutch shoes!
                                                                               Speed is monitored.
                                                                                                              Alarm#33 press
                                                                                                              HELP
                                                                                                                                                                                                                                                                            8555-9201-140 / 0810
Maintenance-Interval:
Set: 4000 h 0m 0s
Actual: 22 h 12 m 46 s
                         The factory setting for "Set" is a maintenance interval of 4000 hours as specified
                         in the separator instruction manual. Under "Actual", the operating hours of the
                         separator are counted.
                         When the pre-set number of operating hours is reached, a common alarm is ac-
                         tivated and a corresponding warning appears in the display. A common alarm
                         does not interrupt the separating program and must be acknowledged using the
                         ACK key.
                         Note:
                         Please take the maintenance interval setting from the instruction manual of the
                         separator used.
Maintenance-Interval WMS
Set: 168 h 0m 0s
Actual: 22 h 12 m 46 s
                         The factory setting for "Set" is a maintenance interval of 168 hours (1 week).
                         Under "Actual", the operating hours of the WMS-Sensor water signal is counted.
                         In the case the “Actual” counter came up to the preset operating hours, a main-
                         tenance message will be activated with a corresponding hint that the WMS-
                         Sensor Check must be done! The message does not interrupt the separation
                         process and must be acknowledged with the „ACK“button after the WMS Sen-
                         sor Check. The acknowledgement resets the the “Actual” value.
                         Under “Last water signal WMS” the operating hours of the WMS Sensor since
                         the last water indication are counted. Only a water signal resets the operating
                         hours for “Last water signal WMS”.
In the case the message „Maintenance WMS-Sensor“ occurs in the display, a manual check of the WMS-
Sensor function must be done.
Please follow up the described check!
                                                                         c
          dle connection piece“.
     A:
     Position 1: Closed
     Position 2: Open
5    Check the status of the WMS          Check Digital Input „DI 0.5“ in the control unit.
     Sensor in the digital input list
     (DI 0.5).
     The status of the signal must
     change from “0” to “1”
     In the case that the signal status
     did not change you have to dis-
     assamble the WMS Sensor.
     The surface of the WMS Sensor
     must be cleaned and after as-
     sembling of the WMS Sensor
     the function must be checked
     again.
     If WMS sensor still has malfunc-     Take care to the recommended maintenance intervals in the
     tion it must be exchanged!           instruction manual!
6    Check status of „Maintenance-        Values for „Actual“ and „Last water signal WMS“ must be reset
     Interval WMS“ in the Display of      to “0”!
     the control unit!
7    Disconnect the water feed line
     “1” from the coupler socket “2”
     on top of the Unitrolplus housing
     and close the coupler socket “2”
     with the plug.
                                                                         c
     Reconnect the water feed line
     “1” with the handle connection
     piece.
8    Separation process could be          Push button „F9“or touch key “Prog. ON”
     started again by pushing button
     „F 9 – Program 1 or the similar
     touch key “Prog. ON” on the
     touchpanel in case of E40 con-
     trol unit.
4 Program logic
Water Feed Clean oil discharge Product Feed Dirty oil recirculation
2 6 1 5
PCV1
TAH
      P                                                                                  TI       TT
                                                                                                              TAL
                                V7                        V4
                   PAH
                         PT1
                   PAL
                                                          V1
PI 1
V2 V5 V6
3 8 9
                                                                                   PT2              WMS
                                                                          7
                                                                                                       PI 2
                                                                     10       11
                                                                                                               LSHH
LSH
12 LSL
V1 Feed valve 1
V2 Filling/Displacementwater valve 3
V4 Pre-set valve 1
TT TT Temperature transmitter 1
PI 1 PI 1 Pressure indicator 6
TI TI Temperature indicator 1
1 Product Feed
2 Water Feed
7 Sensing line
11 Sludge discharge
12 Sludge tank
                                           A                                         A
                                      Power supply
                                          ON
                                         Selftest
                                       Control Unit
                                        Start of
                                Sensor Check Sequence
                                    Precondition:
                                Oil temperature > TSL setpoint
                                Oil temperature < TSH setpoint
                                       LSHH not activated
                                      Vibro II not activated
                                             B
                                                                                      B
                                             B
                                          Start of
                                    Separation sequence
                                                                              Process
                                                                             monitoring
                                                                             activated
                                                                                 E
                                      Ejection program
                                     Waiting - Start (T03)
                                     Displacement (T01)
                               Waiting after displacement (T08)
                                     Total ejection (T04)
                               Waiting after first ejection (T12)
                                     Total ejection (T04)
                                    Speed recovery (T05)
                                                                                      C
                                             C
                                Stop of separation sequence
                                      Ejection program
                                     Waiting - Start (T03)
                                     Displacement (T01)
                               Waiting after displacement (T08)
                                     Total ejection (T04)
                               Waiting after first ejection (T12)
                                     Total ejection (T04)
                                    Speed recovery (T05)
                                         Separator
                                     Ready for operation
                                             A
                                                                                 D
                                       D
                                    Process
                                     options
                                    activated
                  Alarm
          Separator OFF activated
                after (T39)
                     A
                                                                                       E
                                             E
                                          Process
                                         monitoring
                                         activated
                                                                CU = Compact Unit
                               Alarm
                             PAL / PAH
                             TAL / TAH
                               LAHH
                                 A
Water Feed Clean oil discharge Product Feed Dirty oil discharge
2 6 1 6
PCV1
TAH
P TI TT
                                                                                                              TAL
                                V7                        V4
                   PAH
PT1
                   PAL
                                                          V1
PI 1
V2 V5 V6
3 8 9
                                                                                   PT2              WMS
                                                                          7
                                                                                                       PI 2
                                                                     10       11
                                                                                                                LSHH
LSH
12 LSL
V1 Feed valve 1
V2 Filling/Displacementwater valve 3
V4 Pre-set valve 1
TT TT Temperature transmitter 1
PI 1 PI 1 Pressure indicator 6
TI TI Temperature indicator 1
1 Product Feed
2 Water Feed
7 Sensing line
11 Sludge discharge
12 Sludge tank
                                           A
                                                                                     A
                                      Power supply
                                          ON
                                         Selftest
                                       Control Unit
                                                                          Process
                                                                         optionens
                                                                         activated
                                                                             D
                                        Start of
                                Sensor Check Sequence
                                           B
                                                                                          B
                                             B
                                          Start of
                                    Separation sequence
                                                                        Process
                                                                       monitoring
                                                                       activated
                                                                           E
                                      Ejection program
                                     Waiting - Start (T03)
                                     Displacement (T01)
                               Waiting after displacement (T08)
                                     Total ejection (T04)
                               Waiting after first ejection (T12)
                                     Total ejection (T04)
                                    Speed recovery (T05)
                                     Separation program
                                 Closing water impulse (T06)
                                      Filling water (T02)
                                  Waiting after filling (T09)
                                       Separation (T20)
                                    Sensor flushing (T38)
                                 Bridging after ejection (T11)
                                                                                    C
                                             C
                                Stop of separation sequence
                                      Ejection program
                                     Waiting - Start (T03)
                                     Displacement (T01)
                               Waiting after displacement (T08)
                                     Total ejection (T04)
                               Waiting after first ejection (T12)
                                     Total ejection (T04)
                                    Speed recovery (T05)
                                         Separator
                                     Ready for operation
                                             A
                                                                                   D
                                           D
                                       Process
                                        options
                                       activated
                     Alarm
             Separator OFF activated
                   after (T39)
                        A
                                                                                         E
                                               E
                                            Process
                                           monitoring
                                           activated
                                                                  CU = Compact Unit
                                 Alarm
                               PAL / PAH
                               TAL / TAH
                                 LAHH
                                   A
Water Feed Clean oil discharge Product Feed Dirty oil recirculation
2 6 1 5
PCV1
TAH
P TI TT
                                                                                                         TAL
                                 V7                       V4
                   PAH
PT1
                   PAL
                                                          V1
PI 1
V2
3 9
                                                                                                  PT2
                                                                          7
PI 2
10 11 LSHH
LSH
12 LSL
V1 Feed valve 1
V2 Filling/Displacementwater valve 3
PI 1 PI 1 Pressure indicator 6
TI TI Temperature indicator 1
1 Product Feed
2 Water Feed
7 Sensing line
11 Sludge discharge
12 Sludge tank
                                                                                       A
                                         A
                                    Power supply
                                        ON
                                       Selftest
                                     Control Unit
                                                                            Process
                                                                           optionens
                                                                           activated
                                                                               D
                                      Start of
                              Sensor Check Sequence
                                         B
                                                                                        B
                                           B
                                        Start of
                                  Separation sequence
                                                                           Process
                                                                          monitoring
                                                                          activated
                                                                              E
                                    Ejection program
                                   Waiting - Start (T03)
                                   Displacement (T01)
                             Waiting after displacement (T08)
                                   Total ejection (T04)
                             Waiting after first ejection (T12)
                                   Total ejection (T04)
                                  Speed recovery (T05)
                                   Separation program
                               Closing water impulse (T06)
                                    Filling water (T02)
                                Waiting after filling (T09)
                                     Separation (T20)
                               Bridging after ejection (T11)
                                                                                    C
                                           C
                              Stop of Separation sequence
                                    Ejection program
                                   Waiting - Start (T03)
                                   Displacement (T01)
                             Waiting after displacement (T08)
                                   Total ejection (T04)
                             Waiting after first ejection (T12)
                                   Total ejection (T04)
                                  Speed recovery (T05)
                                       Separator
                                   Ready for operation
                                           A
                                                                                   D
                                           D
                                       Process
                                        options
                                       activated
                     Alarm
             Separator OFF activated
                   after (T39)
                        A
                                                                                       E
                                             E
                                          Process
                                         monitoring
                                         activated
                               Alarm
                             PAL / PAH                          CU = Compact Unit
                             TAL / TAH
                               LAHH
                                 A
                                                KNOW-HOW PROTECTION
                                                No GEA-specific device
                                                No warranty by operation!
                                                Alarm#00 Please check:
                                                Contact Westfalia
                                                customer service !
                                                   MOTOR PROTECTION
                                                   Alarm#01 Please check:
                                                   Motor protection function / Brake
                                                   Thermistor relay / Operating water
                                                   Valve V2 / No. of clutch shoes
                                                   Drive blocked / Feed capacity
                                                   Press ESC then ACK
                                                    EJECTION MONITORING
                                                    Alarm#07 Please check:
                                                    Function of bowl / Hydraulics with V3
                                                    Connector valve V3 / Current offset setting/
                                                    Water quantity / TIMER 28
                                                    Current converter / Pre-strainer
                                                    Press ESC then ACK
                                                   VIBRATION STAGE 1
                                                   Alarm#11 Please check:
                                                   TIMER 25 / Solids holding space
                                                   Polyamide sealing ring / Sensor
                                                   Fixing / Limit value 1 / Correct
                                                   bowl parts / Foundation fixing
                                                   Press ESC then ACK
                              or
                                        Step 2     Alarm acknowledgement
                                                   VIBRATION STAGE 2
                                                   Alarm #12 Please check:
                                                   Limit value 2 / Solids holding space
                                                   Polyamide sealing ring / Sensor
                                                   Fixing / Correct bowl parts
                                                   Foundation fixing
                                                   Press ESC then ACK
                                                BOWL MONITORING
                                                Alarm#18 Please check:
                                                TIMER T06 / Current offset
                                                Frequency power supply
                                                Water pressure for hydraulic
                                                Function V3
                                                Press ESC then ACK
                         TR1 =    Time record of WMS water sensor during HFO operation, TR1 will
                                  be deleted when TR3 has expired.
                         TR2 =    Time record of sludge pump operation, TR2 will be deleted when
                                  TR3 has expired.
                         TR3 =    Time counter factory setting 24 h
                           TR1 =       Time record of WMS water sensor during HFO operation, TR1 will
                                       be deleted when TR3 has expired.
                           TR2 =       Time record of sludge pump operation, TR2 will be deleted when
                                       TR3.has expired.
                           TR3 =       Time counter factory setting 24 h
                                                 MAINTENANCE required
                                                 Alarm#27 Please check:
                                                 Operating hours have reached
                                                 4000; 8000; 12000; 16000 hours
                                                 or multiple / Carry out maintenance
                                                 according manual
                                                 Press ESC then ACK
                                                 EJECTION MONITORING
                                                 Alarm#33 Please check:
                                                 Function of bowl / Hydraulics with V3
                                                 Connector valve V3 / Speed offset eject.
                                                 Water quantity / TIMER 28 / Speed pick- up
                                                 defective / Pre-strainer
                                                 Press ESC then ACK
                                                 BOWL MONITORING
                                                 Alarm#34 Please check:
                                                 TIMER T06 / Water pressure
                                                 for hydraulic / Speed offset
                                                 bowl monitoring / Function V3
                                                 Speed pick- up defective
                                                 Press ESC then ACK
                                                 SEPARATOR- START
                                                 Alarm#35 Please check:
                                                 TIMER T10 / Brake unfastened / Clutch
                                                 shoes oily / Drive belt oily / Speed setting
                                                 of operation speed / Speed pick-up defective
                                                 Press ESC then ACK
                                                 BELT MONITORING
                                                 Alarm#36 Please check:
                                                 Drive belt / Motor defect electrically
                                                 Clutch shoes worn or oily
                                                 Speed pick- up defective
                                                 Press ESC then ACK
                                                  PT 1 MONITORING
                                                  Alarm#39 Please check:
                                                  Switching function PT 1
                                                  PT 1 dirty / Setpoint PAL 1
                                                  Bowl hydraulics / Ejection
                                                  monitoring / Discharge pressure
                                                  Press ESC then ACK
                                                 NO WATER SEAL
                                                 Alarm#40 Please check:
                                                 Clean oil discharge pressure
                                                 Function V2 / Filling water quantity
                                                 Valve V2 and filter / Water pressure
                                                 TIMER T06 / Setting TAL / Setting PT 2
                                                 Press ESC then ACK
                                                  EMERGENY STOP
                                                  Alarm#46 Please check:
                                                  Emergency Button
                                                  Press ESC then ACK
                                                 WATER DISCHARGE
                                                 Alarm#48 Please check:
                                                 Clean oil valve V7 / Water cont. in Oil
                                                 Filling water quantity / oil
                                                 Temperature / Regulating ring
                                                 Polyamide ring / Press. sensor PT2
                                                 Press ESC then ACK
Alarm#02
Free
Alarm#03
Free
Alarm#04
 TAL PRODUCT FEED          Separation   Not possible                Set setting higher than the     • Separation
 Temp. < min                                                        current temperature               stop
 Separation - Stop                                                                                  • V1 in circula-
 Alarm#05                                                                                             tion
                                                                                                      Common alarm
 TAH PRODUCT FEED          Separation   Not possible                Set setting lower than the      • Separation
 Temp. > max                                                        current temperature               stop
 Separation - Stop                                                                                  • V1 in circula-
 Alarm#06 press ENTER                                                                                 tion
                                                                                                      Common alarm
 EJECTION MONITORING       Ejection     Close manual ball valve     Disconnect cable head of V3     • Separation
 Current< Offset                        which is in front of V3     or increase offset Ejection       stop
 Separation – Stop                      shortly before and during   current during discharge        • V1 in circula-
 ALARM#07                               discharge cycle T04.        cycle T04.                        tion
                                                                                                      Common alarm
 LAHH SLUDGE TANK          Control      Immerse or dampen           Disconnect cable head of        • Separation
 Sludge tank - full        Active       sensor LSHH.                LSHH Sensor                       stop
 Separation - Stop                                                                                  • V1 in circula-
 Alarm#08                                                                                             tion
                                                                                                      Common alarm
 FEED PUMP FAULT           Control      Not possible                Trigger motor protection        • Separation
 Motor protection          Active                                   switch                            stop
 Separation - Stop                                                                                  • V1 in circula-
 Alarm#09 press ENTER                                                                                 tion
                                                                                                      Common alarm
 SLUDGE PUMP FAULT         Control      Not possible                Trigger motor protection        • Separation
 Program - stop            Active                                   switch                            stop
                                                                                                    • V1 in circula-
 Alarm#10                                                                                             tion
                                                                                                      C          l
Alarm#14
 WMS SYSTEM FAULT          Separation    Open valve V2 to add    Bridge WMS Sensor and set     • Separation
 Too much water                          water into bowl until   pressure sensor PT2 switch-     stop
 Separation - Stop                       Timer 22 is elapsed.    ing point to 0 or bridge.     • V1 in circula-
 Alarm#15                                                                                        tion
                                                                                               • Common alarm
                                                                                               • Ejection
 Free
Alarm#16
Free
Alarm#17
 BOWL MONITORING           Separation    Not possible.           Extend T06 during separa-     • Common alarm
 Current too high                                                tion and shorten T14 to       •
 during TIMER 06                                                 300sec.! Wait until T14 is
 Alarm#18                                                        elapsed and Alarm appears.
 SEPARATOR MOTOR OFF       Circulation   Not possible.           Set TIMER 39 to 10 seconds    • Common alarm
 Sep. in circulation                                             and push program stop         • Motor off
 T 39 expired M. off                                             button.
 Alarm#19
Free
Alarm#20
Free
Alarm#21
 EXCESSIVE RUNNING         Control      Set TR2 to 1 minute and     Set TR2 to 1 minute              • Common alarm
 TIME OF SLUDGE PUMP       Active       immerse or dampen LSH       Trigger testfunction with
 In TR3 hours                           Sensor to start sludge      magnet at testpoint of LSH
 Alarm#22                               pump until Alarm ap-        sensor until Alarm appears.
                                        pears.
 WMS ACTIVATED             Separation   Set T22 to 300sec. and      Set T22 to 300sec., TR1 to 1     • Common alarm
 TOO OFTEN                              TR1 to 1 minute and         minute and bridge WMS
                                        open valve V2 to add        sensor until Alarm appears.
 Alarm#23                               water into bowl until
                                        Alarm appears.
 Free
Alarm#24
Free
Alarm#25
Free
Alarm#26
 Maintenance               Separation   Not possible.               Set value to actual value        • Message in the
 required see manual                                                plus 1hour and wait 1hour.         display
Alarm#27
 WARNING                   Separation   Set T17 to 1s and           Set T17 to 1s and                • Message in the
 WATER CONTENT                          set counter 5 to 1 and      set counter 5 to 1 and             display
                                        T47 to 60sec.than open      bridge pressure sensor           • V6 opens
 Alarm#28                               valve V2 for 10sec. to      PSL2 or set optional PT2 to
                                        add water into bowl until   0 and 1 in short intervals
                                        warning appears.            during T47 for minimum 2
                                                                    times, so that warning ap-
                                                                    pears.
 Free
Alarm#29
Free
Alarm#30
Free
Alarm#31
Free
Alarm#32
 EJECTION MONITORING       Ejection      Close manual ball valve     Disconnect cable head of V3   • Separation
 Eject. speed < Offset                   which is in front of V3     or increase offset Ejection     stop
 Separation – Stop                       shortly before and during   speed during discharge        • V1 in circula-
 Alarm#33                                discharge cycle T04.        cycle T04.                      tion
                                                                                                   • Common alarm
 BOWL MONITORING           Separation    Not possible.               Extend T06 during separa-     • Common alarm
 Speed drop too high                                                 tion and shorten T14 to       •
 during TIMER 06                                                     300sec.! Wait until T14 is
 Alarm#34                                                            elapsed and Alarm appears.
 SEPARATOR-START           Separator     Set Timer 10 to 10 sec.     Set setting for min. speed    • Motor off
 monitoring                Motor Start   and press motor start       higher than the normal        • Message in the
 during T10                              button.                     operating speed.                display
 Alarm#35                                                                                          • Common alarm
 BELT MONITORING           Separator     Not possible.               Disconnect speed pick-up      • Motor off
 Drive belt broken         Motor Start                               sensor and press motor        • Message in the
 during start up                                                     start button.                   display
 Alarm#36                                                                                          • Common alarm
 MAX.- SPEED               Separator     Not possible.               Set setting for Max. Speed    • Motor off
 MONITORING                Motor Start                               lower than normal operating   • Message in the
 during operation          and Separa-                               speed, during start up or       display
 Alarm#37                  tion                                      during operation.             • Common alarm
                                                                                                   •
 MIN.- SPEED               Separator     Not possible.               Set setting for Min. Speed    • Motor off
 MONITORING                Motor Start                               higher than normal operat-    • Message in the
 during operation          and Separa-                               ing speed,                      display
 Alarm#38                  tion                                      during start up or during     • Common alarm
                                                                     operation.                    •
 PT 1 MONITORING           Ejection      Not possible.               Set pressure sensor PT1       • Separation
 clean oil discharge                                                 switching point to 0 while      stop
 Separation - Stop                                                   ejection sequence is run-     • V1 in circula-
 Alarm#39                                                            ning.                           tion
                                                                                                   • Common alarm
                                                                                                   •
 NO WATER SEAL             Separation    Reduce T34 and T35 to 5     Set pressure sensor PT2 to    • Separation
 during TIMER 35                         sec. and close hand         0 and disconnect cable of       stop
 Separation - Stop                       valve before V2 during      WMS sensor.                   • V1 in circula-
 Alarm#40                                T02.                                                        tion
                                                                                                   • Common alarm
                                                                                                   • Ejection
 MIN. SEPARATING TIME           Separation    Not possible.                  Disconnect cable head on          • Separation
        < 10 Min                                                             pressure sensor PT2 during          stop
 Separation - Stop                                                           T21.                              • V1 in circula-
 Alarm#41                                                                                                        tion
                                                                                                               • Common alarm
                                                                                                               • Ejection
 SMS SENSOR FAULT               Sensor        Close hand valve at V2         Disconnect cable head on          • Common alarm
 during SMS Check               Check         while test is running.         pressure sensor PT2 while         • Ejection
 during TIMER 44                                                             test is running.
 Alarm#42
 PT2 SENSOR ACTIVE              Circulation   Open valve V2 until pres-      Set pressure sensor PT2 to        • Message in the
 in Idle run                                  sure increase on pres-         0 during idle run.                  display
                                              sure sensor PT2 during                                           • Common alarm
 Alarm#43                                     idle run.                                                        • Ejection
 PAL1 CLEAN OIL DISCHARGE       Separation    Open clean oil discharge       Disconnect cable head on          • Separation
 Pressure < setting                           valve V7                       pressure sensor PT1                 stop
 Separation - Stop                                                                                             • V1 in circula-
 Alarm#44                                                                                                        tion
                                                                                                               • Common alarm
                                                                                                               • Ejection
 PAH 1 CLEAN OIL DISCHARGE      Separation    Close clean oil discharge      Set pressure sensor PT1           • Separation
 Pressure > setting                           valve V7                       switching point to a lower          stop
 Separation - Stop                                                           value than the current value.     • V1 in circula-
 Alarm#45                                                                                                        tion
                                                                                                               • Common alarm
                                                                                                               • Ejection
 EMERGENCY STOP                 Control       Not possible.                  Push emergeny button.             • All electric
 Button pushed                  Active                                                                           components
                                                                                                                 are shut down
 ALARM#46                                                                                                        like Motors,
                                                                                                                 Control Units
                                                                                                                 ect!
 FREE
 ALARM#47
 WATER DISCHARGE:               Separation    Close clean oil discharge      Set pressure sensor PT1           • Separation
 Overflow                                     valve V7. Set optional PAH1    switching point to 0 while          stop
                                              (PT1) in clean oil discharge
 Separation - Stop
                                              to 10 bar.
                                                                             separation sequence is            • V1 in circula-
 Alarm#48                                                                    running.                            tion
                                                                                                               • Common alarm
                                                                                                               • Ejection
 MAINTENANCE required           Circulation   Not possible.                  Set value to actual value         • Message in the
 WMS Sensor                                                                  plus 1minute and wait               display
                                                                             1minute. Add extra water to
 Alarm#49                                                                    get WMS Signal if neces-
                                                                             sary.
 Free
Alarm#50
For a simulation, the function must be activated and must be valid for the product.
             ATTENTION:
             After the simulation, all changes must be reset to the initial conditions!!!!
8.1 Bowl-Volumes
Separator type
                                                           Volumes in dm3
Bowl volume            1,00   1,00   2,00    2,00   3,60    3,60 10,00 10,00 23,00 23,00 40,00 40,00
Sludgespace volume     0,50   0,50   0,90    0,90   1,60    1,60   4,00     4,00 10,00 10,00 14,60 14,60
Fillingwater volume    0,20   0,50   0,30    0,90   0,40    1,60   0,50     4,00   1,00 10,00 3,00 14,60
Displacementwater
volume
                       0,50   0,50   1,10    1,10   2,00    2,00   6,00     6,00 13,00 13,00 25,40 25,40
                                                                 Separator type
                              OSE 5 / OSE 10 / OSE 20              OSE 40           OSE 80          OSE 120
                                                               Speed in U/min
 Max. Speed                            12 500                       8 700            7 300            6 400
 Min. Speed                            11 000                       7 200            5 800            4 900
 Rated Speed                           12 000                       8 200            6 800            5 900
 Speed -Offset
                                            100                     100               100              100
 Ejection monitoring
 Speed -Offset
                                            100                     100               100              100
 Bowl monitoring