LVM Screen Instructions for MGL & MGY

This document provides step-by-step instructions for calibrating the Candela MGL & MGY laser systems using a 15mm handpiece. It details the necessary settings for fluence, rate, and time, and guides the user through the calibration process to reach the Laser Variable Mode (LVM) screen. The final step emphasizes the importance of taking a picture of the LVM screen for technician reference.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
55 views4 pages

LVM Screen Instructions for MGL & MGY

This document provides step-by-step instructions for calibrating the Candela MGL & MGY laser systems using a 15mm handpiece. It details the necessary settings for fluence, rate, and time, and guides the user through the calibration process to reach the Laser Variable Mode (LVM) screen. The final step emphasizes the importance of taking a picture of the LVM screen for technician reference.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

INSTRUCTIONS TO GET TO LVM SCREEN

System: Candela MGL & MGY

GentleLase Mini (MGL) - GentleYag Mini (MGY or VPYAG)

The MGL & MGY must be calibrated with a 15mm handpiece. If you
know how to calibrate your machine, proceed to Step Seven.

Step One: Turn key to turn the machine on

Step Two: Allow machine to warm up for around 10 minutes. Insert


15mm handpiece into the calport. Your screen will confirm, on the
lower left corner, that the handpiece is the correct size.

Fluence

Rate Time

This is your main screen.

Step Three: Adjust the energy, speed & timing by pressing the Fluence,
Rate & Time buttons and using the up & down arrows in the center of
the screen to adjust (see above picture).
MGL Settings- FLUENCE: 30 J/cm2 RATE: 1.0 Hz TIME: 3 ms
MGY Settings- FLUENCE: 44 J/cm2 RATE: 1.0 Hz TIME: 3 ms

L as e r S e rv ic e S o lu tio n s  6 5 0 G r o ve Ro ad , S u ite 1 0 4  Pau ls b o r o , N J 08066


w w w . las e r s e r v ic e so lu tio n s . co m  P: 856 - 8 5 3 - 7 555  F : 856 - 8 5 3 - 7 5 7 5
Step Four: Press down arrow button and your next screen will ask you
to confirm the 15mm handpiece. Press the Check button on the left.

Step Five: Your next screen will say “Press”. Press and hold the blank
center button in the middle of screen until your laser begins to fire and
calibrate. Continue to press & hold until screen changes.

L as e r S e rv ic e S o lu tio n s  650 G r o ve Ro ad , S u ite 104  Pau ls b o r o , N J 08066


w w w . las e r s e r v ic e so lu tio n s . co m  P: 8 5 6 - 8 5 3 - 8555  F : 856 - 8 5 3 - 7 5 7 5
Step Six: When your laser is done calibrating, it will tell you to remove
the handpiece from the calport. When removed, it will return you to
the main screen.

Step Seven: Press the arrow button on bottom right of screen.

L as e r S e rv ic e S o lu tio n s  650 G r o ve Ro ad , S u ite 104  Pau ls b o r o , N J 08066


w w w . las e r s e r v ic e so lu tio n s . co m  P: 8 5 6 - 8 5 3 - 8555  F : 856 - 8 5 3 - 7 5 7 5
Step Eight: The next screen will look like this:

Press and hold the


arrow button on the
bottom right of this
screen.

Step Nine: You are now at the Laser Variable Mode (LVM) screen.

Take a picture of this screen.


The technician will need the Last Faults and the HV@60JHD