E302 Indicator: P.N. 305669264, Revision L1, August 2013)
E302 Indicator: P.N. 305669264, Revision L1, August 2013)
81320475 en
Easy Reference:
1 INTRODUCTION ........................................................................................ 3
2 SPECIFICATIONS...................................................................................... 5
3 INSTALLATION ......................................................................................... 7
3.1 UNPACKING ...................................................................................... 7
3.2 LOCATING ......................................................................................... 7
3.3 CONNECTION ................................................................................... 8
4 KEYPAD..................................................................................................... 9
5 DISPLAY .................................................................................................. 11
5.1 SYMBOLS AND INDICATORS ......................................................... 11
6 CALIBRATION COUNTER FOR APPROVED INDICATORS ................. 12
7 BATTERY................................................................................................. 14
8 BACKLIGHT ............................................................................................ 14
9 AUTO POWER OFF ................................................................................. 14
10 OPERATION .......................................................................................... 15
10.1 ZEROING ......................................................................................... 15
10.2 TARING ............................................................................................ 15
10.2.1 MANUAL TARE .......................................................................... 15
10.2.2 PRESET TARE (NOT AVAILABLE ON E302-M APPROVED
INDICATORS) ....................................................................................... 16
10.3 WEIGHING ...................................................................................... 17
10.4 PARTS COUNTING......................................................................... 17
10.5 CHECK-WEIGHING ......................................................................... 21
10.5.1 SETTING UP WHILE WEIGHING ............................................... 22
10.5.2 SETTING UP WHILE PARTS COUNTING OR % WEIGHING .... 23
10.6 LIMITS STORING AND RECALLING .............................................. 23
10.7 PERCENT WEIGHING ..................................................................... 25
10.8 ANIMAL (DYNAMIC) WEIGHING .................................................... 28
10.8.1 ANIMAL WEIGHING PROCEDURE .......................................... 29
10.9 ACCUMULATED TOTAL ................................................................. 30
10.9.1 MANUAL ACCUMULATION ....................................................... 30
10.9.2 AUTOMATIC ACCUMULATION ................................................. 32
11 RS-232 SPECIFICATION ...................................................................... 33
11.1 INPUT COMMANDS FORMAT ........................................................ 39
12 CALIBRATION ....................................................................................... 40
13 PARAMETER SETTINGS ...................................................................... 41
13.1 CHECK WEIGHING PARAMETERS ............................................... 41
13.2 RS-232 PARAMETERS ................................................................... 44
13.3 INDICATOR PARAMETERS ........................................................... 48
13.4 PERCENT WEIGHING AND ANIMAL WEIGHING.......................... 50
14 ERROR MESSAGES ............................................................................. 51
15 SERVICE PARAMETERS ...................................................................... 53
1
15.1 ACCESS TO PARAMETERS ........................................................... 53
15.2 USING “0000” TO ENTER THE CALIBRATION PARAMETER ...... 53
15.3 USING “2006” TO ENTER THE SERVICE PARAMETERS ............. 54
15.3.1 F1 -CALIBRATION .................................................................... 56
15.3.2 F2–DECIMAL POINT POSITION ............................................... 57
15.3.3 F3 – CAPACITY......................................................................... 57
15.3.4 F4 –INITIAL ZERO RANGE ....................................................... 58
15.3.5 F5 -RE-ZERO RANGE............................................................... 58
15.3.6 F6 -SUCCESSIVE TARE ........................................................... 59
15.3.7 F7 –ADC COUNTS .................................................................... 59
15.3.8 F8 –ZERO MODE ..................................................................... 60
15.3.9 F9 –LOW VOLTAGE DETECTION ............................................ 60
15.3.10 F10 –CALIBRATION COUNT (E302-M ONLY) ........................ 61
15.3.11 F10 –AUTO ZERO RANGE (NOT AVAILABLE ON E302-M)... 61
16 REPLACEMENT PARTS AND ACCESSORIES .................................... 62
17 SERVICE INFORMATION ...................................................................... 63
18 APPENDIX ............................................................................................. 65
19 DECLARATION OF CONFORMITY ....................................................... 66
2
1 INTRODUCTION
The E302 has LEDs to indicate when a weight is below the low limit,
between the limits or above the high limit next to the display. These
can work in conjunction with an audible alarm for check weighing as
well as LCD showing LO, OK and HI.
The E302 has a sealed keypad with colour coded membrane switches
and a large easy to read liquid crystal type display (LCD) supplied with a
backlight.
OIML Approved models, E302 -M, do not allow pounds units, have
calibration controlled by jumpers or passcodes and other limitations as
noted in the manual.
3
4
2 SPECIFICATIONS
INPUT SECTION
Load Cells Up to 4 , 350 ohm load cells
Minimum 87 ohms, maximum 1120 ohms
Connection 6 wires
2 excitation, 2 sense, 2 signal
Excitation 5Vdc
Sensitivity 0.15uv/d (E302 -M, 1.5uv/e)
Linearity 0.01% FS
Zero Range 0- 10mv
Signal range 0-40mv
ADC Sensitivity Approximately 0.02 µv/ADC count
DIGITAL SECTION
Maximum Range Typically 1kg – 30000kg
Divisions Up to 30,000, (E302 –M, 3000 or less)
Weigh units kg / g / lb for Europe, Asia, Africa
kg / g / lb / oz / lb:oz for USA
kg only for E302-M series
Stabilisation Time 2 Seconds typical
For approved indicators the input specifications is limited to 1.5 µv per division and
the number of divisions is limited to 3000d. Kilograms only.
5
6
3 INSTALLATION
3.1 UNPACKING
This indicator must be connected to a load cell platform and calibrated as necessary
to match the platform and user requirements. See Section 15 for set-up
information.
The users application and the technical specifications of the platform or load cell will
determine the necessary configuration.
3.2 LOCATING
The scales should not be placed in a location that will
reduce the accuracy.
7
3.3 CONNECTION
This indicator must be connected to a load cell platform and calibrated as necessary
to match the platform and user requirements.
The E302 has a connector configured for a 6 wire load cell. Connect the load
cells/platform to the indicator as shown below. The cable length should be as short
as possible, using a large size wire to minimise errors due to resistance in the leads.
E302-M model must use the 6 wire connection and has certain limitations for wire
size and length. Refer to the Approval Test Certificate for details.
8
Figure 1A shows the connections to a 6 wire load cell. Figure 1B shows a preferred
method to attach a 4 wire load cell, using a 6 conductor cable to go from the
indicator to the platform or load cell where it connects to the 4 wires from the load
cells. The Excitation and sense wires are connected together near the load cell.
For less exacting applications you can connect the excitation to the sense at the
connector.
4 KEYPAD
[Zero] Sets the zero point for all subsequent Escape from any setting menus
weighing. The display shows zero.
[Tare] It tares the indicator and stores the current Accept the set values
weight in memory as a tare value, subtracts
the tare value from the weight and shows
the results. This is the net weight.
[Unit] This is used to select the weighing units Allows the weight, unit weight, and
from a preset list of available units. count to be seen when parts
counting or to change from weight
to % in percent weighing
9
[Low It sets the limits for check weighing and None
Limit] allows setting of either the low limit or the
high limit or both.
&
[High
Limit]
10
5 DISPLAY
5.1 SYMBOLS AND INDICATORS
Low battery
bAt LO or
: The colons “:” are used to separate pounds from ounces and
for the real time clock.
11
Next to the LCD are a number of LED’s that show when the weight is below, within
or over the limits during check weighing.
NOTE: The LED’s can be set by the user to off, bar, spot or segment mode. See “F3
LED” in section 13.1
The LED can be set to display as a bar, increasing from Low to OK to High, a single
spot increasing from Low, OK to High, or as a single bar that changes colour as the
weight progresses from Low to OK to High.
In this way if the record of the Calibration or Parameter counters do not agree with
recorded settings the responsible person inspecting the indicator can take
appropriate action.
12
The Counters are incremented any time the calibration section or the Factory
parameters section have been modified.
At power on, the display will show the current software revision number followed
by the message of the Calibration Count “[AL[nt” then a number i.e. “123”. The
number from the counter memory. Then the Parameter Counter message of
“PAr[nt” and probably a different number, i.e. “234”. The counters cannot be
reset to 0, they will increment until the display can no longer hold the values. (1 to
999999). It is expected we will never have more than 1 million calibrations in the
life of the machine.
The indicator will then continue to do the display test and go to normal weighing.
If during the time the counting displays are shown, the user presses the [Tare] key,
the user will be given a message to enter the passcode necessary to calibrate the
indicator, “P - - - - “ Enter the code “P0000” to Enter calibration or “P1000” to
enter the parameters, followed by pressing the [Tare] key.
The Calibration access will allow user calibration (See section 15.1) and the
parameter code will allow access to the following parameters. (see section 15.2).
13
7 BATTERY
The indicators can be operated from the rechargeable battery, if
desired. The battery life is determined by the number and impendence
of the load cells connected. With a single load cell and backlight
disabled the life is approximately 70 hours before needing to be
recharged.
When the battery needs charging a symbol on the display will turn on.
The battery should be charged when the symbol is on. The indicator will
still operate for a period of time after which it will automatically switch
off to protect the battery.
To charge the battery, simply plug into the mains power supply. The
indicator does not need to be turned on.
8 BACKLIGHT
The backlight for the LCD can be set by the user to always off, always on or
automatic (on only when the indicator is in use or a key is pressed). See setting of
the parameter “S2 bL” in section 13.3.
14
10 OPERATION
10.1 ZEROING
You can press the [Zero] key at any time to set the zero point from
which all other weighing and counting is measured. This will usually be
necessary when the platform is empty. When the zero point is obtained
the display will show the zero indicator.
10.2 TARING
10.2.1 MANUAL TARE
Zero the indicator by pressing [Zero]. The zero indicator will be on.
Place a container on the pan and its weight will be displayed.
Press [Tare] when the reading is stable. The weight that was displayed
is stored as the tare value and it is subtracted from the display, leaving
zero on the display. The stable and Net indicator will be on.
As a product is added only the weight of the product will be shown. The
indicator could be tared a second time if another type of product was to
be added to the first one. Again only the weight that is added after
taring will be displayed.
15
NOTE:
If the value to be tared is very large the scale may not allow you to tare the value as
the negative value will not fit on the display area. In this case the scale will beep
twice when the [Tare] key is pressed and then return to normal weighing without
setting tare.
Press [Tare] or [Zero] to remove the tare value and display zero. The Net indicator
will disappear.
When the indicator is at zero with no weight on the platform it is possible to enter a
preset tare.
Zero the indicator by pressing [Zero]. The zero indicator will be on.
Press [Tare] to tare the indicator. The value that was entered is stored
as the tare value and it is subtracted from the display, leaving a
negative number on the display.
16
10.3 WEIGHING
To change the weighing unit press the [Unit] key. The only alternative
weighing unit is pounds. This can be enabled by the user in the parameters
section. See section 13.3.
The indicator can be used to count parts based on the average weight of a
sample weighed. When more parts are added the total number of parts are
displayed.
Press [Cnt] to enter the Parts Counting mode. The display will show the
last sample size used. For example, “10 Pcs”.
17
Either place 10 parts on the platform for determining the average piece
weight or use a different number of parts. For example, place 20 parts on
the platform, press [CE] to clear the last values and then enter the value
20 using the numeric keypad.
If the parts are too light to measure accurately, the count may become
faulty. It is suggested that the samples to be weighed should each weigh
more than the resolution of the indicator.
After the sample has been weighed the indicator will count any other
parts added by applying the average piece weight to the weight of the
parts to be counted.
The [Tare] key works normally during this time, so it is possible to tare
the display with a container on the platform or to enter a preset tare
value as described in section 10.2.2.
During parts counting the display can be changed to show the net
18
weight, unit weight and number of parts by each time pressing the
[Func] key.
19
To count a different sample quantity, press the [Count] key. The display
will show the last used sample size. Either use this sample size with a
different part or enter a new sample size as above.
20
10.5 CHECK-WEIGHING
NOTE:
1. The alarm and the LED bargraph can each be set to OFF (See section 13.1).
The LCD display will indicate whenever the weight is within or exceeds the
limits by showing ‘OK’, ‘HI’ or ‘LO’.
2. The limits can be locked by the manager. A Limit Password must be used to
change the limits or recall other limits from memory.
3. If Limit Password is enabled then enter the password which will allow you
to change the limits or the operation of the beeper or bargraph.
21
10.5.1 SETTING UP WHILE WEIGHING
Press the [Low Limit] key. It will show the current low limit. The “LO”
symbol will appear on the display.
Press the [CE] key to clear the old value and then enter the new low
limit using the numeric keys. The decimal point is fixed at the position
that is used for the current weighing unit. When the desired value is
shown press [Tare] to accept the value. If you want to reset the value to
zero, press [CE] to clear the value.
To set the high limit press [High Limit], the display will show the high
limit, the “HI” symbol will be on to the left side of the display. Set the
high limit in the same way the low limit was set.
Pressing the [Tare] key to enter the value will return the indicator to
weighing, with the Check-weighing function enabled.
22
10.5.2 SETTING UP WHILE PARTS COUNTING OR % WEIGHING
During parts counting and percent weighing the limits are set in the same way as
above. The limits are displayed in pcs or %.
See Section 10.4 for the description of parts counting and Section 10.7 for percent
weighing.
NOTE:
1. The weight must be greater than 20 scale divisions for the check-
weighing to operate.
2. To disable the check weighing function, enter zero into both the limits as
described above. When the current limits are shown, press [CE] to clear
the settings, then press [Tare] to store the zero values.
The indicator can store up to 10 sets of high and low limits in memory along with the
weighing units in use (including pcs for parts counting and % for percent weighing)
as well as settings for the beeper and bar graph.
During Check weighing the current limits can be stored or previously stored units
can be recalled.
23
Press the [Lim] key. If you are already in
the check weighing mode the display will ask
if you wish to store the current limits by
showing “StOrE” or recall another set of limits
by showing “rECALL”. The [Lim] key can be
used to toggle between “StOrE” and “rECALL”.
24
NOTE:
1. If the recalled limit is for parts counting, the display will show the last sample
value used, ready for a new sample to be counted.
2. If the recalled unit is a percent weighing limit, the display will show the last
sample value used, ready for a new sample to be weighed.
3. If the memory location was empty the indicator will return to weighing.
The indicator can be set to perform percent weighing. See Section 13.1.
The indicator will use a mass on the platform as the 100% reference weight. If the
platform is empty (or the indicator is tared) then the user can input a reference
weight using the keypad.
If using a reference weight (or object) as your 100% reference, add the weight
to the to the platform.
Press [Func]. The first option is “FUnC 1”, press the [Func] key 3 more times to
display “FUnC 4”.
25
Press [Tare] again to enter percent weighing. The indicator will set the sample
mass on the platform as 100% reference weight.
Remove the sample weight. Then any other weight placed on the platform
will be displayed as a percentage of the original sample. For example, if
3500g is placed on the platform and percent weighing is selected, the display
will show 100.00%. Remove the 3500g weight and place a 3000g weight. The
display will show 85.7% as 3000g is 85.7% of 3500g.
The number of decimal points will depend on the weight used. A smaller
weight will show only “100%” while a larger weight might show “100.00%”.
If the indicator was showing zero weight before entering this function, then
the user must manually enter the weight to be set as 100%. When “F4 PCT” is
displayed, enter the weight to be used for the 100% reference, then press
[Tare] to accept the reference weight. The display will show “0.00 %”.
26
If the indicator shows “x x . x x %”, which is the last weight used as a reference,
press [CE] to clear and use the numeric keypad to enter a new value. Press
[Tare] to accept the new reference weight.
NOTE:
The display may jump by large numbers unexpectedly if small weights are used to
set as 100% reference. The indicator checks if the weight is too small and will show
Error 7.
27
10.8 ANIMAL (DYNAMIC) WEIGHING
The indicator can be set to animal (dynamic) weighing for weighing any items that
are unstable or may move. See Section 13.4.
The indicator uses a special filter to minimise the effects of any movement on the
platform.
Press [Func]. The first option is “FUnC 1”, press the [Func] key 3 more times to
display “Func 4”.
Press the [Tare] key. “F4 PCt” will be displayed. Press the [Func] key to
advance to the second function, “F4 AnL”, Animal weighing.
The display will flash “Ani “ 2 times then show the current weight, 0.00. The
indicator is now ready to weigh.
28
10.8.1 ANIMAL WEIGHING PROCEDURE
With the platform empty the display will show zero weight.. Place containers
or blankets onto the platform and press the [Tare] key to zero the display.
The indicator may go into the animal weighing procedure when the items are
placed on the platform but will return to showing zero when the [Tare] key is
When a stable reading is found, the display will show this value, and the
display will be locked until the [Unit] key is pressed. The display will show the
“Hold” symbol while the display is locked. Remove the animal, the display will
hold the weight value.
Press the [Unit] key to unlock the display. The display will flash “Ani” twice,
and be ready for the next animal.
To weigh a second animal press the [Tare] key if necessary to zero the display,
and place the next animal on the platform. It is also possible to simply place
the next animal on the scale without clearing the last value first. The indicator
will detect the new weight and hold it as before.
The indicator will remain in the animal weighing mode until the [Zero] key is
pressed. Then it will return to normal weighing.
29
10.9 ACCUMULATED TOTAL
The indicator can be set to accumulate when a weight is added to the platform
automatically or manually by pressing [Print]. See Section 13.2.
NOTE:
3. If at any time the weighing units are changed, the accumulated data will be
lost.
When the indicator is set to manual accumulation, the weight displayed will be
stored in the memory when the [Print] key is pressed and the weight is stable.
Remove the weight and press [Print] when the indicator is at zero. The
display will show "ACC 1" and then the weight in memory for 2 seconds
before returning to normal. The weight can be output to a printer or PC
using the RS-232 interface.
30
When the indicator is at zero place a second weight on the platform.
When stable press [Print] to accumulate the weight. The display will
show "ACC 2" for 2 seconds and then show the new total.
Continue until all weights have been added. This can continue for up to
99 entries until the capacity of display is exceeded.
To view the total in memory press the [Print] key when the indicator is
at zero. The display will show the total number of accumulation "ACC
xx" and the total weight before returning to zero.
To print the total, press [Print] to recall and then immediately press
[Print] the second time to print the results.
To erase the memory, press [Print] to view the total and then
immediately press [CE] to clear the memory.
31
10.9.2 AUTOMATIC ACCUMULATION
When the indicator has been set to Automatic Accumulation the value is stored in
memory automatically.
Place a weight on the platform. The beeper will sound when the display
is stable indicating the value is accepted. Remove the weight.
The display will show "ACC 1" and then the total in the memory before it
returns to zero. Adding a 2nd weight will repeat the process.
While the weight is on the platform, press the [Print] key to view the
values- first the accumulation number "ACC x" and then the total will be
shown.
NOTE:
1. The indicator will not show the value when a weight is removed.
32
11 RS-232 SPECIFICATION
The E302 indicator is supplied with bi-directional RS-232 interface as standard. The
indicator when connected to a printer or computer outputs the weight with the
selected weighing unit through the RS-232 interface.
Default Specifications:
ASCII code
8 data bits
No Parity
Connector:
Pin 3 Output
Pin 2 Input
The indicator can be set to print text in English, French, German or Spanish. See the
RS-232 parameters section for details.
33
LABEL = On Label Formats are printed
Only weight value along with the weighing unit is printed. If Percent weighing is
used then % is shown in place of weighing units.
<cr><lf>
<cr><lf>
Date 12/09/2006 <cr><lf>
Time 14:56:27 <cr><lf>
<cr><lf>
Scale ID 123456 <cr><lf> If ID is zero, it is left blank
User ID 234567 <cr><lf>
<cr><lf>
Net Wt 1.234 kg <cr><lf> Net Wt. (or Gross Wt.)
<cr><lf>
<cr><lf>
<cr><lf>
<cr><lf>
34
<cr><lf>
<cr><lf>
Date 12/09/2006 <cr><lf>
Time 14:56:27 <cr><lf>
<cr><lf>
Scale ID 123456 <cr><lf>
User ID 234567 <cr><lf>
<cr><lf>
Net Wt. 1.234 kg <cr><lf> Net Wt. (or Gross Wt.)
Unit Wt. 123 g <cr><lf> g for metric and lb for pounds
PCS 10 pcs <cr><lf>
<cr><lf>
<cr><lf>
35
Data Format- Memory Recall Output:
<cr><lf>
Date 12/09/2006 <cr><lf>
Time 14:56:27 <cr><lf>
<cr><lf>
Scale ID 123456 <cr><lf>
User ID 234567 <cr><lf>
<cr><lf>
------------------<cr><lf>
TOTAL
No. 5 <cr><lf>
Wt. 1.234 kg <cr><lf>
PCS 10 pcs <cr><lf>
<cr><lf>
------------------<cr><lf>
<cr><lf>
36
LABEL = Off single line output
1.234 Kg<lf><cr>
1. The accumulated total will not be sent to the RS-232 when the continuous
print is turned on.
2. The continuous print will only be for the current weight and the display data.
3. In other languages the format is the same but the text will be in the language
selected.
4. When the scale is in the Lb:oz weighing unit the RS-232 output will only show
pounds. 10lb:8oz will be printed as 10.5 lb.
37
Description ENGLISH FRENCH GERMAN SPANISH
38
11.1 INPUT COMMANDS FORMAT
The indicator can be controlled with the following commands. Press the [Enter] key
of the PC after each command.
T<cr><lf> Tares the indicator to display the net weight. This is the same as
pressing [Tare].
Z<cr><lf> Sets the zero point for all subsequent weighing. The display shows
zero.
39
12 CALIBRATION
The E302 indicator can be calibrated using kilogram weights or using
pounds weights, depending on the weighing unit selected at the time of
calibration.
To start the calibration, either get into the calibration section through
the Indicator Settings (“FUnC 3”- see Section 13.3) or turn the indicator
off and switch on again and then press [Tare] during the self-test. Enter
code number 0000 and press [Tare]. This will take you directly to the
calibration section.
Remove all weight from the platform and then press the [Tare] key
when the display is stable. After the Zero point is set, the display will
show “Ld xx”. Place the suggested calibration mass on the platform. It is
best to use a weight close to the full capacity of the indicator. If the
mass is different from the displayed value, enter the value of the mass
in whole numbers. The kg or the lb symbol will be on to show the active
unit.
The indicator will calibrate to the mass. When complete, it will display
“PASS” and then either display “S8 CAL” (if entered the calibration
section through the Indicator Settings as per section 13.3) or return to
normal weighing (if entered directly). Remove the calibration mass.
40
13 PARAMETER SETTINGS
Pressing the [Func] key allows the user to access the parameters for customising the
indicator. The parameters are split into 4 groups-
2. RS-232 parameters
3. Indicator parameters
When [Func] is pressed, display will first show “FUnC 1” for Check weighing
parameters.
Enter [2] for RS-232 parameters or [3] for Indicator parameters or [4] for
percent weighing and animal weighing, or press the [Func] key to advance
through the groups “FUnC 1”, “FUnC 2” , “FUnC 3” and “FUnC 4”. Press [Tare] to
enter the desired group of parameters.
Press [Zero] to return to the group “FUnC 1”. If you press [Zero] again, the
indicator will exit the User Parameter section and return to normal weighing.
Shortcut to enter this group is to press and hold the [Unit] key for 4 seconds.
The display will go directly to “FUnC 1”.
Press [Func] to scroll through the parameters and press [Tare] to enter a
parameter setting.
41
Press [Func] to view the options for setting.
Press [Tare] to confirm the change and then advance to the next parameter
by pressing the [Func] key.
F1 LLk This parameter prevents the With LLK set to Off (oFF), oFF
normal user from changing the user is allowed to
the limits with the help of a change limits at any
Limit Lock. time.
42
F2 LEd This parameter sets the LED bAr - Bar type bAr
indicator to off or on and the Spot - Spot type
LED type (whether LED’s are
Seg - Segment
on in the form of a
continuous bar or a spot LED oFF - Off
or a segment of colour).
NOTE:
1. The Check weighing password is separate from the indicator password, see
section 13.3.
2. If the password is other than 0000, user must enter the password to gain
access to “F3 LLk”, “F4 LEd”, “F5 bEP”, “F6 CPS” and “F7 nCK”.
43
13.2 RS-232 PARAMETERS
Shortcut to enter this group is to press and hold the [Print] key for 4 seconds.
The display will go directly to “C1 on”.
Press [Tare] to enter a parameter. Press [Func] to view the options for the
parameter settings.
Press [Tare] to confirm the change and then advance to the next parameter
by pressing the [Func] key.
Press [Zero] to return to the group “FUnC 2”. If you press [Zero] again, the
indicator will exit the User Parameter section and return to weighing.
This group of parameters can be set by the user for setting the language, baud rate,
printing mode, etc. The user can also set a Scale ID number and a User ID number.
Parameter Description Options Default Values or
setting
44
C3 PrM Printing Mode- Manual as mA StA mA Sta
soon as stable,
45
The indicator will perform the following, depending on the Accumulation and Print
Settings:
ACCUMULATION AC on AC Off
SETTINGS
PRINT SETTINGS
46
Accumulate and print when Print after the scale is stable and the
the scale has reached weight has been removed.
stability after the weight is
Au End removed.
If the [Print] key is pressed it is
ignored.
If the [Print] key is pressed it
Do not accumulate.
is ignored.
Note: Manual Printing is allowed at zero only when the scale is stable. If set to Auto
the scale will not automatically print when it returns to zero.
47
13.3 INDICATOR PARAMETERS
Shortcut to enter this group is to press and hold the [Count] Key for 4
seconds. The display will go directly to “S1 Un ”.
Press [Func] to view the list of parameters.
Press [Tare] to enter a parameter. Press [Func] to view the options for the
parameter settings.
Press [Tare] to confirm the change and then advance to the next parameter
by pressing the [Func] key.
Press [Zero] to return to the group “FUnC 3”. If you press [Zero] again, the
indicator will exit the User Parameter section and return to normal weighing.
This group of parameters are used to control the operation of the indicator.
48
S3 AoF Auto Off- Disable or set SLP 0 SLP 0
time increment to turn off SLP 1
the indicator
SLP 5
SLP 10
S4 dt Set Time and Date format Enter the time 00:00:00
and settings manually
mm:dd:yy
Enter the date
manually
49
13.4 PERCENT WEIGHING AND ANIMAL WEIGHING
See section 10.7 and 10.8 for details of these special weighing modes.
F4 AnL Enter the Animal Weighing Set the filter value. Enabled
mode of operation, See Always
section 10.8
50
14 ERROR MESSAGES
During the initial power-on testing or during operation, the indicator may show an
error message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your dealer for support.
Err 4 Initial Zero is greater than Weight on the platform when turning
allowed (4% of maximum the indicator on.
capacity) when power is
Excessive weight on the platform
turned on or when the
when zeroing the indicator.
[Zero/Enter] key is pressed.
Platform is not installed.
Damaged Electronics.
Err 8 High limit input error Low limit is set first, then the high
limit is set lower than the low limit
51
and high limit not equal to zero.
Err 9 Low limit input error High limit is set first, then the low
limit is set higher than the high limit
and low limit not equal to zero.
52
15 SERVICE PARAMETERS
15.1 ACCESS TO PARAMETERS
APPROVED INDICATORS
Or if the Calibration and Parameters have been enabled (see 15.2.10) the user must
enter the correct password to have access. See Section 6.0.
Entering passcode 0000 will allow calibration as shown in 15.1, entering 1000 will
allow access to a limited set of parameters as described in section 6.2 and entering
the passcode 2006 will allow access to all parameters as shown in section 15.2.
NON-APPROVED INDICATORS
Non EC Approved indicators will allow entry to the parameters if the Tare key is
pressed during the power on cycle. The passwords work as above.
“LoAd” “6” “KiLoS” Load the requested calibration weight and press
53
[Tare]
Or, “SPAn” “FAiLEd ” This means calibration has failed. Remove the
calibration weight and repeat the process.
For the non approved indicator press the [Tare] key during the display counting
when turned on,
Approved Indicators: For the Approved version a jumper can be installed to allow
access or the Calibration and Parameter Counters must be enabled (see 15.2.10).
54
Apply power to the indicator. If the jumper has been used the display will ask for a
code number, “Pn “ on the Weight Display immediately. Or press the [Tare] key
during the time the calibration counters are being displayed.
Enter the number 2006 when “Pn “ is displayed and then press [Tare].
The displays will show the first parameters, called “F1” “CAL”.
To select another parameter press the [Func] key to advance through the
parameters.
When a parameter is entered by pressing the [Tare] key, the displays will guide you
through the parameter selected and the options available.
To enter the calibration parameter, press the [Tare] key when “F1” is displayed. The
indicator will be calibrated using 2 masses of approximately 1/3Maximum and
Maximum. If the indicator has been calibrated once the values will be stored. If this
is the first time the indictor is calibrated the user must enter the values for the
calibration masses.
It is necessary to set the decimal point location and the capacity before calibration is
possible.
The display will instruct you to remove any weight from the platform, “UnloAd”.
Press [Tare].
The display will tell you to add the first weight to the platform: “Ld 1” “ 10 kg” If
necessary change the value shown to match the weight to be used. Press [CE] to
clear the old value and then enter the new value. All values entered are in whole
numbers only.
Add the weight shown, wait for stability then press the [Tare] key.
The display will tell you to add the second weight to the platform: “Ld 2 “ “ 30 kg”
Add the weight shown, wait for stability then press the [Tare] key.
For the approved indicator the display will then show “JP” “On” indicating the jumper
is still in place if the jumper within the indicator was used to access the parameters..
56
Switch off the indicator, and switch it on again to continue with the other Service
parameters.
To set the value for the decimal point location. The options are 0, 0.0, 0.00,
0.000, 0.0000
To enter this parameter, press the [Tare] key when “F2 dEC” is shown.
Press the [Func] key to change the value. Select from 0, 0.0, 0.00, 0.000, 0.0000
15.3.3 F3 – CAPACITY
To enter this parameter, press the [Tare] key when “F3 CAP” is shown.
Enter the numeric values using the keypad. The indicator will check the number of
divisions n = maximum/increment is less than 30,000 divisions. (3000 divisions for
EC approved versions)
On non-approved versions the display then lets you select the increment, “InC 2”
For example 100kg x 0.01kg the increment is 10grams. but the last digit increments
by 1.
57
The display will show the current increment value as used with the current decimal
point position.
Press the [Func] key to change the value. Select from 1,2,5,10, 20 or 50
Not all increments may be available for the capacity you selected.
For EC Approved versions the indicator will determine the increment that maintains
the number of divisions to be 3000 or less.
To enter this parameter, press the [Tare] key when “F4 int” is shown.
Press the [Func] key to change the value and press [Tare] to accept the value.
To enter this parameter, press the [Tare] key when “F5 rEZ” is shown.
To enter this parameter, press the [Tare] key when “F6 SCS” is shown.
To enter this parameter, press the [Tare] key when “F7 Cnt” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter.
This can be an aid to service.
59
15.3.8 F8 –ZERO MODE
To enter this parameter, press the [Tare] key when “F8 ZEm” is shown.
Select the Zero mode desired. In all but special cases Zero Mode 1 is used. The
other 2 zero modes are for unique locations in the world and effect the +/- range of
the zero.
This parameter allows detection of low voltage when the battery wears down.
To enter this parameter, press the [Tare] key when “F9 LVd” is shown.
60
15.3.10 F10 –CALIBRATION COUNT (E302-M ONLY)
This parameter allows the calibration and parameter counting function to be active.
To enter this parameter, press the [Tare] key when “F10 Cn” is shown.
The display will show if the Calibration Counting Mode is set to on or oFF. If On the
Calibration count and Parameter count will be seen at power on as described in
section 6.0. If set to Off the only method that can be used for access to the
calibration or parameters is to place the jumper on pins 1-2 of J1, See section 15.2.
To enter this parameter, press the [Tare] key when “F10 Az” is shown.
The display will show the current setting, 0.5d, 1d, 2d, or 5d.
61
16 REPLACEMENT PARTS AND ACCESSORIES
If you need to order any spare parts and accessories, contact your supplier or
Cooperativa Bilanciai. A partial list of such items is as follows-
62
17 SERVICE INFORMATION
This manual covers the details of operation. If you have a problem with the indicator
that is not directly addressed by this manual then contact your supplier for
assistance. In order to provide further assistance, the supplier will need the
following information which should be kept ready:
63
A. Details of your company
(This part of information should always be available for any future correspondence.
We suggest you to fill in this form as soon as the unit is received and keep a print-out in
your record for ready reference.)
Date of Purchase:
64
18 APPENDIX
PARAMETER LAYOUT for E302 / E302-M SCALES
Key functions while in this section [Tare] enter a parameter or accept the changes
Press the [Func] key to [Func] move to next parameter or option
enter Functions mode. [Zero] return to previous parameter or return to weighing
65
This symbol on the weighing instrument you have purchased means that:
• this electric or electronic appliance cannot be disposed of as solid municipal waste;
• it must be collected separately;
• it can be returned to the dealer when a new one is purchased;
• improper use or disposal of this appliance may pollute the environment or damage the health of human
beings;
• failure to comply with the aforementioned instructions is punished by law.
In particular:
• the external casing and the mechanical components are made of plastic and/or metal materials;
• the casing contains printed circuits with electronic components;
• there is an Ni-Cd or Li battery on the CPU board;
• the electrical connections are made with insulated copper conductors;
• the rechargeable batteries are the Ni-MH or Pb type.
66
19 DECLARATION OF CONFORMITY
We
SOCIETÀ COOPERATIVA BILANCIAI CAMPOGALLIANO
Via S. Ferrari, 16
41011 Campogalliano (MO) - Italia
Franco Brighenti
Management Division
67