CUS71D Manual

The document provides operating instructions for the Turbimax CUS71D ultrasonic interface sensor, detailing safety instructions, installation, wiring, device description, maintenance, and troubleshooting. It emphasizes the sensor's designated use in water and wastewater applications, the importance of proper installation by trained personnel, and adherence to safety regulations. Additionally, it includes information on the sensor's dimensions, accessories, and operational safety measures.

Uploaded by

msankararao
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
129 views20 pages

CUS71D Manual

The document provides operating instructions for the Turbimax CUS71D ultrasonic interface sensor, detailing safety instructions, installation, wiring, device description, maintenance, and troubleshooting. It emphasizes the sensor's designated use in water and wastewater applications, the importance of proper installation by trained personnel, and adherence to safety regulations. Additionally, it includes information on the sensor's dimensions, accessories, and operational safety measures.

Uploaded by

msankararao
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 20

Operating Instructions

Turbimax CUS71D
Ultrasonic interface sensor
Immersion sensor for interface measurement

BA490C/07/EN/01.11
71129495
Endress+Hauser
Turbimax CUS71D

Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4 9 Technical Data. . . . . . . . . . . . . . . . . . . 17
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 9.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1.2 Installation, commissioning and operation . . . . . . . . 4 9.2 Performance characteristics . . . . . . . . . . . . . . . . . . 17
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4 9.3 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 9.4 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1.5 Notes on safety icons and symbols . . . . . . . . . . . . . . 5 9.5 Mechanical construction . . . . . . . . . . . . . . . . . . . . 17

2 Identification . . . . . . . . . . . . . . . . . . . . 5 Index . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 5

3 Installation . . . . . . . . . . . . . . . . . . . . . . 6
3.1 Incoming acceptance, transport, storage . . . . . . . . . . 6
3.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . . 7
3.4 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 9
3.5 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 10

4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1 Connecting to the transmitter . . . . . . . . . . . . . . . . 11
4.2 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 11

5 Device description . . . . . . . . . . . . . . . 12
5.1 Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.2 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . 12
5.3 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.4 Sensor monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.6 Cyclic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

6 Maintenance. . . . . . . . . . . . . . . . . . . . 14
6.1 Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . . 14

7 Accessories. . . . . . . . . . . . . . . . . . . . . 15
7.1 Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2 Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.3 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

8 Troubleshooting . . . . . . . . . . . . . . . . . 16
8.1 Troubleshooting instructions . . . . . . . . . . . . . . . . . 16
8.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Endress+Hauser 3
Safety instructions Turbimax CUS71D

1 Safety instructions

1.1 Designated use


CUS71D is an immersion sensor designed for interface measurement in water and wastewater.
The sensor is particularly suited for use in the following applications:
• Wastewater treatment: primary clarifier, sludge thickener
• Water purification: settling basin after flocculant dosage, sludge height in contact sludge process
• Chemical industry: static separation process
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.

1.2 Installation, commissioning and operation


Please note the following items:
• The measurement process in the sensor uses UV light. UV light can damage the eyes. Never look
into the cuvette while the sensor is in operation.
• Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
Trained personnel must be authorized for the specified activities by the system operator.
• Electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood these Operating Instructions and must
adhere to them.
• Before commissioning the entire measuring point, check all the connections. Ensure that
electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional commissioning.
Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorized and specially trained personnel.
• If faults can not be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the
manufacturer’s or by the service organization.

1.3 Operational safety


The sensor has been designed and tested according to the state of the art and left the factory in
perfect functioning order.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.

1.4 Return
If the sensor has to be repaired, please return it cleaned to the sales center responsible.
Please use the original packaging, if possible.
Before returning the device, please clarify all formalities - such as obtaining an identification number
- with your sales center.
Please enclose the completed "Declaration of Hazardous Material and De-Contamination" (copy the
second last page of these Operating Instructions) with the packaging and the transportation
documents. No repair without completed declaration!

4 Endress+Hauser
Turbimax CUS71D Identification

1.5 Notes on safety icons and symbols

Warning!
# This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if
ignored.
Caution!
" This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.

! Note!
This symbol indicates important items of information.

2 Identification

2.1 Nameplate
Compare the order code indicated on the nameplate to the product structure and your order.
The nameplate contains the following information:
• Manufacturer data
• Order code (device version)
• Extended order code
• Serial number

! Note!
To discover what sensor version you have, enter the order code on the nameplate into the search
screen at the following address:
www.products.endress.com/order-ident

2.2 Scope of delivery


The scope of delivery comprises:
• 1 sensor Turbimax CUS71D in the ordered version
• 1 Operating Instructions BA490C/07/EN
If you have any questions, please contact your supplier or your local sales center.

2.3 Certificates and approvals


Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with
the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.

Endress+Hauser 5
Installation Turbimax CUS71D

3 Installation

3.1 Incoming acceptance, transport, storage


• Make sure the packaging is undamaged!
Inform the supplier about any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the contents. Keep the damaged products until the matter
has been settled.
• Check that the order is complete and agrees with your shipping documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your local sales center.

3.2 Dimensions

mm (inch)
20 (0.79)

mm (inch) NPT 3/4"


20 (0.79)

46 (1.81)
G1

NPT 3/4"
12 (0.47)
46 (1.81)

G1 34.7 (1.37)
12 (0.47)

34.7 (1.37)
110 (4.33)
135.5 (5.33)
68.8 (2.71)
43.3 (1.70)

19 (0.75)

74.3 (2.93)
74.3 (2.93)

a0014922 a0014921

Fig. 1: Dimensions of standard sensor Fig. 2: Dimensions of sensor with wiper

6 Endress+Hauser
Turbimax CUS71D Installation

3.3 Installation instructions

3.3.1 Measuring system


A complete measuring system comprises:
• Ultrasonic sensor Turbimax CUS71D
• Multi-channel transmitter Liquiline CM44x
and optional:
• Weather protection roof CYY101
• Holder system Flexdip CYH11
• Fixed or rotatable immersion pipe Flexdip CYA112

3
1
2

a0014923

Fig. 3: Ultrasonic sensor with holder system and multi-channel transmitter

1 Holder system Flexdip CYH112 4 Assembly Flexdip CYA112


2 Multi-channel transmitter Liquiline CM44x 5 Ultrasonic sensor Turbimax CUS71D
3 Weather protection roof

Endress+Hauser 7
Installation Turbimax CUS71D

3.3.2 Measuring system with pendulum holder

1 2 3

a0015156

Abb. 4: Measuring system with pendulum holder

1 Holder system Flexdip CYH112 (cross clamp) 3 Assembly Flexdip CYA112 with CUS71D
2 Holder system Flexdip CYH112 (pendulum holder) 4 PVC wiper (C-UA 100826-61)

The PVC wiper (C-UA 100826-61) protects the ultrasonic sensor from getting damaged by the
surface skimmer.

8 Endress+Hauser
Turbimax CUS71D Installation

3.4 Installation conditions


Basin configuration

A B

E
F

a0002181

Fig. 5: Basin configuration


A Sensor
B Minimum distance of sensor to basin wall = 45 cm (1.48 ft.)
C Reference point e.g. water surface
D Zero point
E Basin depth
F Opening angle of ultrasonic cone, 6°

Installation instructions
Look at the construction drawing of the basin for a suitable position for the sensor. In doing so, you
must take the following factors into account:
• The minimum distance between the basin wall and the sensor is 45 cm (1.48 ft.) (sensor emits
ultrasound in conical form).
• There should not be any basin wall protrusions or piping in the measuring range below the sensor.
Scrapers that are only temporarily in this area are permitted.
• Do not install the sensor in zones in which air bubbles, turbulence, high levels of turbid material
or suspended matter or foam formation occur (e.g. inlet).
• Using an immersion tube, install the sensor 20 cm ( 0.66 ft.) beneath the surface of the water.
• The transmitter may not be installed in a second enclosure (heat accumulation).
• If possible, do not install the transmitter near high voltage sources. In addition, also avoid sources
of magnetic fields, e.g. large transformers or frequency converters.
• The system can only detect a separation zone if there is a clear transition between the zones.
Unclear transition from the liquid to the solid phase cannot be detected.

Endress+Hauser 9
Installation Turbimax CUS71D

Circular Clarifier

2
1/3 2/3
4

12
13
1/3 2/3
11

14

a0015232

Fig. 6: Basin configuration in circular clarifier

A View from top B Cross section


1 Surface skimmer 11 Sensor
2 Walk way 12 Hand rail
3 Sensor mounting 13 Surface skimmer
4 Bottom rake 14 Bottom rake
5 Rake direction

3.5 Post-installation check


• Sensor and cable undamaged?
• Cap undamaged?
• Compliance with permissible sensor installation position?
• Is the sensor installed in an assembly and is not suspended from the cable?
• Avoid moisture by rain by putting the protective cap on the assembly?

10 Endress+Hauser
Turbimax CUS71D Wiring

4 Wiring
# Warning!
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.

4.1 Connecting to the transmitter


The sensor will be connected to the transmitter as follows:

a0012460

Fig. 7: Sensor connection

The maximum cable length is 100 m (328 ft).

4.2 Post-connection check

Instrument status and specifications Remarks


Are the sensor, assembly, junction box or cable damaged? Visual inspection

Electrical connection Remarks

Does the supply voltage of the transmitter match the specifications on the
nameplate?

Are the installed cables strain-relieved and not twisted ?

Is the cable type route completely isolated ? Power cable/weak current cable

Are the power supply and signal cable correctly connected to the Use the connection diagram of the
transmitter ? transmitter.

Long enough length of cable core stripped and correct in terminal? Check seating (pull slightly)

Are all the screws terminals properly tightened ? Tighten

Are all the cable entries installed, tightened and sealed ? For cable entries lateral: cable loops
downwards for water to be able to drip off.
Are all the cable entries installed downwards or lateral ?

Endress+Hauser 11
Device description Turbimax CUS71D

5 Device description

5.1 Sensor design


The sensor is designed for continuous in-situ determination of interfaces.
The sensor includes all necessary modules:
• Power supply
• Ultrasonic source sends the measurement signals.
• Ultrasonic receiver receives the measurement signals, digitalizes and converts the signals to a
measurement value.
• The microcontroller of the sensor controls the internal operations and the data transmission.
All data - including calibration data - are stored in the sensor. That means:
• A pre-calibrated sensor can be used at the measuring point.
• The sensor can be calibrated externally.
• The sensor can be used at multiple measuring points with different calibration data.

5.2 Measuring principle


A piezoelectric crystal is integrated in a flat cylindrical plastic housing. When the crystal is excited
by an electrical voltage, it generates a sonar signal. The ultrasonic waves are transmitted at a
frequency of 657 kHz at an angle of 6° to scan the separation zones.
The parameter measured is the time it takes for the transmitted ultrasonic signal to reach the solid
particles in the separation zone and return to the receiver.
A sensor version with wiper avoids film formation at the sensor membrane. If a turbidity signal is
also needed a sensor version with wiper and integrated turbidity measurement is available.

5.3 Function
The speed of the sound varies according to the physical properties of the measuring medium and is
affected by temperature and air pressure. The liquid zones and solids content of the medium also
vary.
To obtain precise measurement results, it is therefore vital to adapt system variables to the process,
e. g. pulselength and the speed of the sound.
The CM44x offers the following possibilities for signal evaluation:
• Mask out regions where the separation zone is not expected.
• Evaluate received signal strengths differently.
• Select leading or trailing signal edges in the evaluation.
• Amplify sensor signals at different rates, e. g. for floating sludge.
• Define a region (gate) above and below the separation zone. Signal evaluation only takes place in
the defined region. The gate wanders with the separation zone. This makes smoothing algorithms
unnecessary.
• Arrow indicator for basin floor.

5.4 Sensor monitoring


The optical signals are continuously monitored und checked for plausibility.
Discrepancies are reported via error messages by the transmitter.
The sensor check system of the Liquiline CM44x reports the following failure conditions:
• Implausible high or low measuring values
• Disturbed controlling due to erroneous measuring values

12 Endress+Hauser
Turbimax CUS71D Device description

5.5 Calibration

5.5.1 Factory calibration


The sensor is precalibrated when leaving the factory. After adaption of the basin parameters, the
sensor can be used in a wide range of applications without additional calibration.
The factory calibration cannot be deleted and can be retrieved at any time. All other calibrations -
performed as customized calibrations - are referenced to this factory calibration.

5.5.2 Calibration in the process


Sensor calibration in the process is performed by inputs at Liquiline CM44x (see BA451C/07/EN).

5.6 Cyclic cleaning


For cyclic cleaning a ultrasonic sensor with integrated wiper is available. The time interval is
selectable via software.

Endress+Hauser 13
Maintenance Turbimax CUS71D

6 Maintenance
You have to perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the system, the installation conditions and the
medium in which measurement takes place.

6.1 Cleaning the sensor


Sensor without wiper
Sensor fouling can affect the measurement results and even cause a malfunction.
The sensor must be cleaned at regular intervals to ensure reliable measurement results. The
frequency and intensity of the cleaning process depends on the medium.
Clean the sensor:
• As specified in the maintenance schedule
• Before every calibration
• Before returning the sensor for repair

! Note!
You must rinse the sensor thoroughly with water after cleaning.
Sensor with wiper
The time interval is selectable via software. The cleaning time interval depends on the medium. It
is recommended to replace the wiper once a year.

14 Endress+Hauser
Turbimax CUS71D Accessories

7 Accessories

7.1 Assemblies
Wastewater assembly Flexdip CYA112
• Modular assembly system for sensors in open basins, channels and tanks
• Versions in stainless steel or PVC
• Ordering acc. to product structure (Technical Information TI432C/07/en)

7.2 Holder
Holder system Flexdip CYH112 for water and wastewater assembly Flexdip CYA112
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• The holder system CYH112 works for nearly any type of fixing - fixing on the floor, wall or
directly on a rail.
• Material: stainless steel
• Ordering acc. to product structure (Technical Information TI430C/07/en)

7.3 Transmitter
Liquiline CM44x
• Multiple-channel transmitter for the connection of digital sensors with Memosens technology
• Power supply: 85 to 265 V AC, 18 to 36 V DC or 20 to 28 V AC (not CM448)
• Universally upgradeable
• SD card slot
• Alarm relay
• IP 66
• Ordering acc. to product structure (Technical Information TI444C/07/en)

Endress+Hauser 15
Troubleshooting Turbimax CUS71D

8 Troubleshooting

8.1 Troubleshooting instructions


You must take the entire measuring point into account when troubleshooting:
• Transmitter
• Electrical connections and cables
• Assembly
• Sensor
The possible causes of error indicated in the table below primarily refer to the sensor.

Problem Check Remedial measures

Nothing displayed, no Power supplied to the transmitter? Connect the mains voltage
reaction from the sensor Sensor connected correctly? Connect sensor correctly
Buildup on sensor membrane? Clean sensor

Display value too high or Buildup on sensor membrane? Clean


too low Check basin configuration. Adjust

Display value fluctuates a Check mounting location. Select other mounting location
lot

! Note!
Please observe the troubleshooting instructions provided in the transmitter operating manual.
Examine the transmitter if necessary.

8.2 Return
If the sensor has to be repaired, please return it cleaned to the sales center responsible.
Please use the original packaging, if possible.
Before returning the device, please clarify all formalities - such as obtaining an identification number
- with your sales center.
Please enclose the completed "Declaration of Hazardous Material and De-Contamination" (copy the
second last page of these Operating Instructions) with the packaging and the transportation
documents. No repair without completed declaration!

8.3 Disposal
The device contains electronic components and must therefore be disposed of in accordance with
regulations on the disposal of electronic waste.
Please observe local regulations.

16 Endress+Hauser
Turbimax CUS71D Technical Data

9 Technical Data

9.1 Input

Measured variable Standard sensor Interface

Sensor with wiper Interface

Sensor with wiper and turbidity Interface


measurement Turbidity

Measuring ranges Standard sensor 0.3 to 10.0 m (1.0 to 32 ft)

Sensor with wiper 0.3 to 10.0 m (1.0 to 32 ft)


Sensor with wiper and turbidity 0.3 to 10.0 m (1.0 to 32 ft)
measurement 0 to 50 (200) NTU

9.2 Performance characteristics


Measured error Interface 35 mm at 3.0 m

Turbidity 1 % of the measuring range at 50 NTU

Wavelength Interface 3 mm at 3.0 m

Turbidity 1 NTU
Measuring interval Adjustable

Calibration The sensor is factory calibrated delivered.


The "speed of sound" is adjustable.

9.3 Environment

Storage temperature -20 to 52 °C (-4 to 125 °F)

Degree of protection IP 68 (test conditions: 1 m (3.3 ft) water column during 60 days, 1 mol/l KCl)

9.4 Process

Process temperature 1 to 52 °C (34 to 125 °F)


range
Process pressure 0.0 to 6 bar (0 to 87 psi) absolute

9.5 Mechanical construction

Dimensions See "Installation conditions"

Weight Standard sensor 1.02 kg (2.25 lb)

Sensor with wiper 1.25 kg (2.75 lb)

Sensor with wiper and turbidity 1.25 kg (2.75 lb)


measurement

Materials Sensor ABS and epoxy plastic

Wiper Rubber

Optical window Sapphire

Process connections G1 and NPT ¾''

Endress+Hauser 17
Turbimax CUS71D

Index
A S
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Sensor
C Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Factory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
In the process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Checking
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 T
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Compressed air cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . 13 Transmitter connection. . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Connection Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Troubleshooting instructions . . . . . . . . . . . . . . . . . . . . . . . 16
D
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 5 U
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 W
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
F
Factory calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

H
Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

I
Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 6–7
Installation conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Measuring system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

O
Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operational safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

P
Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

R
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 16

18 Endress+Hauser
a
0
0
0
2
5
8
7

Declaration of Hazardous Material and De-Contamination


Erklärung zur Kontamination und Reinigung
Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA#
clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
RA No. Bitte geben Sie die von E+H mitgeteilte Rücklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie diese
auch außen auf der Verpackung. Nichtbeachtung dieser Anweisung führt zur Ablehnung ihrer Lieferung.

Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Material
and De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of the
packaging.
Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die unterschriebene
"Erklärung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt außen an der
Verpackung an.

Type of instrument / sensor Serial number


Geräte-/Sensortyp ____________________________________________ Seriennummer ________________________

Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gerät in Schutzeinrichtungen

Process data/ Prozessdaten Temperature / Temperatur_____ [°F] _____ [°C] Pressure / Druck _____ [psi] _____ [ Pa ]
Conductivity / Leitfähigkeit ________ [µS/cm] Viscosity /Viskosität _____ [cp] _____ [mm2/s]

Medium and warnings


Warnhinweise zum Medium

harmful/
Medium /concentration Identification flammable toxic corrosive irritant other * harmless
Medium /Konzentration CAS No. entzündlich giftig ätzend gesundheits- sonstiges* unbedenklich
schädlich/
reizend
Process
medium
Medium im
Prozess
Medium for
process cleaning
Medium zur
Prozessreinigung
Returned part
cleaned with
Medium zur
Endreinigung
* explosive; oxidising; dangerous for the environment; biological risk; radioactive
* explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv
Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions.
Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.

Description of failure / Fehlerbeschreibung __________________________________________________________________________


______________________________________________________________________________________________________________
______________________________________________________________________________________________________________

Company data / Angaben zum Absender

Company /Firma ___________________________________ Phone number of contact person /Telefon-Nr. Ansprechpartner:


_________________________________________________ ____________________________________________
Address / Adresse Fax / E-Mail ____________________________________________
_________________________________________________
_________________________________________________ Your order No. / Ihre Auftragsnr. ____________________________

“We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.We further certify that the returned
parts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities.”
“Wir bestätigen, die vorliegende Erklärung nach unserem besten Wissen wahrheitsgetreu und vollständig ausgefüllt zu haben. Wir bestätigen
weiter, dass die zurückgesandten Teile sorgfältig gereinigt wurden und nach unserem besten Wissen frei von Rückständen in gefahrbringen-
der Menge sind.”
nt a X IV
www.endress.com/worldwide

BA490C/07/EN/01.11
71129495
FM+SGML 6.0 71129495