IHH500 Profile Setup Guide
G F E
H M D
J L C
K A B
Power USB Output Sensor
SENSOR CONNECTIONS
PIN SYMBOL DESCRIPTION PIN SYMBOL DESCRIPTION
A +E +Excitation G GND_OUT Ground
B +S +Signal H 5_OUT 5V Output
C –E –Excitation, TEDS Return J –V –V and –mA Amplified Input
Connections
D –S –Signal
K +V +V and +mA Amplified Input
E TEDS_IO TEDS Data Connections
F 24_OUT 24V Output L PLEAD Leading pulse from sensor
M PLAG Lagging pulse from sensor
NOTES
The IHH500 is set to recognize TEDS automatically upon startup if this Shield should be connected to cable clamp of binder cable assembly when
function is available. connecting on instrument side.
This function must be enabled under the TEDS DATA menu. For more information on TEDS see section 6.8 of the IHH500 manual.
Sensor cable shield connections should be grounded on one end, either More information on setting up and editing a sensor profile can be found
the sensor side or the instrument side, to avoid potential ground loops. in section 6.1 of the IHH500 manual.
Sensor Solution Source
Load · Torque · Pressure · Multi-Axis · Calibration · Instruments · Software
www.futek.com
IHH500 Profile Setup Guide 2
CREATING A SENSOR PROFILE
1. Connect the sensor cable to the IHH500 sensor port and turn on the 9. Using the red left and right arrow keys followed by the red Enter
IHH500 by pressing the POWER button in the lower right corner. button set the sensor direction type.
a. During the power up of the IHH500 an internal auto calibration a. UNI-DIRECTION is used for sensors with one direction of output
will be performed. such as compression only.
b. Next, a screen showing the current selected channel, serial b. BI-DIRECTION is used for sensors with two directions of output
number, sensor type, and sensor sensitivity will briefly be such as compression and tension.
displayed. *For more information on sensor direction see section 6.1.3.2 of
c. The normal operating screen will now be displayed with the the IHH500 manual.
updated peak, tracking, valley, selected channel, gross/tare 10. Using the red down arrow key followed by the red Enter button select
mode, and sampling rate (samples per second). UNIT SELECTION.
2. Press the red MENU button to enter into the IHH500 main menu. 11. Using the red up and down arrow keys followed by the red Enter
a. The menu mode is controlled by the red arrow keys and red button select the type of output to be measured by the sensor.
command words. 12. Using the red left and right arrow keys followed by the red Enter
3. Using the red Enter button select SENSOR PROFILE. button select the engineering units to correspond with the measured
4. Using the red down arrow key followed by the red Enter button select output of the sensor.
NEW CHANNEL. * For more information on selectable engineering units see section
5. Using the red left and right arrow keys followed by the red Enter 6.1.3.3.1 of the IHH500 manual.
button set the number for the sensor channel to be created. 13. Using the red down arrow key followed by the red Enter button select
a. Channel 01 is a calibrated reference channel, set in the IHH500 at SENSOR CAPACITY.
the factory, and cannot beoverwritten or selected during a sensor 14. Using the red left and red right arrow keys select the digit to be
profile setup. changed. Using the red up and down arrow keys set the value for the
6. Using the red Enter button select SENSOR CONFIGURATION. selected digit that corresponds to the sensors full capacity. Use the
red ENTER button to save the value.
a. Each step in this menu must be completed before moving on to
the next step. a. The decimal place can be set by first changing the decimal to a
number and then changing another selected digit into a decimal.
7. Using the red left and right arrow keys followed by the red Enter
button set the sensor type. b. The sensor’s full capacity is usually listed on the sensor calibration
certificate.
a. Sensor type is usually listed on the sensor calibration certificate
by output and can be: 15. Setup SENSITIVITY (+)
FULL BRIDGE for mV/V output sensors. a. Using the red down arrow key followed by the red Enter button
select SENSITIVITY (+).
BRIDGE & PULSE for mV/V output sensors with encoder.
• The sensitivity (+) will be the positive mV/V output for
VOLTAGE OUTPUT for amplified voltage output sensors. bridge type sensors and the maximum positive voltage
VOLTAGE & PULSE for amplified voltage output sensors with output for amplified sensors.
encoder. • The output of the sensor is usually listed on the sensor
CURRENT OUTPUT for amplified current output sensors. calibration certificate.
CURRENT & PULSE for amplified current output sensors with • For current output sensors, +/- Sensitivity is not available.
encoder. CALIBRATION must be used for current output sensors. See
* For more information on sensor type see section 6.1.3.1 of the step 19 below.
IHH500 manual. * For more information on sensitivity see section 6.1.3.5 of the
8. Using the red down arrow key followed by the red Enter button select IHH500 manual.
DIRECTION.
IHH500 Profile Setup Guide 3
b. Using the red left and red right arrow keys select the digit to d. With the full load on the sensor, which results in the full positive
be changed. Using the red up and red down arrow keys set output, press the red Enter button to capture the FullScale (+)
the value for the selected digit that corresponds to the sensors reading.
sensitivity (+) and use the red ENTER button to save the value. e. Press the red down arrow key followed by the red Enter button to
• The decimal place can be set by first changing the decimal select Zero load (-) and proceed to the input screen.
to a number and then changing another selected digit into • The direction must be set to Bi-Directional for a – input in
a decimal. CALIBRATION.
16. Setup SENSITIVITY (-) f. With no load on the sensor, press the red Enter button to capture
a. Using the red down arrow key followed by the red Enter button the Zero load (-) reading.
select SENSITIVITY (-). g. Press the red down arrow key followed by the red Enter button to
• SENSITIVITY (-) can only be selected if BI- DIRECTIONAL select FullScale (–) and proceed to the input screen.
was selected earlier. • The direction must be set to Bi-Directional for a – input in
• The sensitivity (-) will be the negative mV/V output for CALIBRATION.
bridge type sensors and the maximum negative voltage h. With the full load on the sensor, which results in the full negative
output for amplified sensors. output, press the red Enter button to capture the FullScale (-)
• The output of the sensor is usually listed on the sensor reading.
calibration certificate. 18. Using the red down arrow key followed by the red Enter button select
• For current output sensors, +/- Sensitivity is not available. PULSE / ROTATE.
Calibration must be used for current output sensors. a. The PULSE / ROTATE menu is selectable only for profiles that
b. Using the red left and red right arrow keys select the digit to specify PULSE as the sensor type.
be changed. Using the red up and red down arrow keys set b. Typical FUTEK settings are 360 and 60.
the value for the selected digit that corresponds to the sensors
sensitivity (-) and use the red ENTER button to save the value. 19. Using the red left and red right arrow keys select the digit to be
changed. Using the red up and red down arrow keys set the value for
• The decimal place can be set by first changing the decimal the selected digit that corresponds to the encoder Pulse counts. Use
to a number and then changing another selected digit into the red ENTER button to save the value.
a decimal.
a. The default value is set to 360.
17. Using the red down arrow key followed by the red Enter button select
CALIBRATION if a live calibration, where a load is to be placed on b. See section 6.1.3.8 of the IHH500 manual for more information
the sensor, is desired. Otherwise, CALIBRATION can be skipped by on PULSE / ROTATE.
pressing the red down arrow key. 20. Using the red down arrow key followed by the red Enter button select
a. Press the red Enter button to select Zero load (+) and proceed to SERIAL NUMBER.
the input screen. 21. Using the red left and red right arrow keys select the digit to be
b. With no load on the sensor, press the red Enter button to capture changed. Using the red up and red down arrow keys set the value for
the Zero load (+) reading. the selected digit that corresponds to the sensor’s serial number. Use
the red ENTER button to save the value.
c. Press the red down arrow key followed by the red Enter button to
select FullScale (+) and proceed to the input screen. 22. Press the red Back button to exit the NEW CHANNEL setup and
return to the SENSOR PROFILE menu.
23. Using the red down arrow key followed by the red ENTER button
select SAVE CHANGES to save and create current sensor profile.
24. Press the red Back button to exit the SENSOR PROFILE menu and
return to the IHH500 main menu.
Drawing Number: SP1187
25. Press the red Back button to exit the IHH500 menu and return to the
Copyright © FUTEK Advanced Sensor Technology, Inc. Neither the whole nor any
normal operation screen.
part of the information contained in, or the product described in this manual, may
be adapted or reproduced in any material or electronic form without the prior 26. The normal operating screen will now be displayed with the current
written consent of the copyright holder. peak reading, sensor output, valley reading, selected channel, gross
or tare mode, and samples per second (sps).
This product and its documentation are supplied on an as-is basis and no warranty
a. Press the Display button to display additional information from