BT Series User Manual
BT Series User Manual
Zhuhai DL Biotech Co., Ltd. (hereinafter called DL) owns the intellectual property rights to DL
product and manual. This manual may refer to information protected by copyright or patents and
does not convey any icense under the patent rights or copyright of DL, or of others.
DL intends to maintain the contents of this manual as confidential information. Disclosure of the
information in this manual in any manner whatsoever without the written permission of DL is
strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative
work of this manual in any manner whatsoever without the written permission of DL is strictly
forbidden.
All information contained in this manual is believed to be correct. DL shall not be liable for errors
contained herein nor for incidental or consequential damages in connection with the furnishing,
performance, or use of this manual.
DL is responsible for the effects on safety, reliability and performance of this product, only if:
All installation operations, expansions, changes, modifications and repairs of this product are
conducted by DL authorized personnel.
Electrical installation of the relevant room complies with the applicable national and local
requirements.
Be sure to operate the analyzer under the situation specified in this manual; otherwise, the analyzer
will not work normally and the analysis results will be unreliable, which would damage the analyzer
components and cause personal injury.
1.1 Introduction
This chapter explains how to use your user manual, which is shipped with your BT series of Auto
Blood Culture System and contains reference information about the system and procedures for
operating, troubleshooting and maintaining the system. Read this manual carefully before
operating your system, and operate your system strictly as instructed in this manual.
This manual contains information written for clinical laboratory professionals to:
The following symbols are used to indicate danger and alert information on the instrument.
Symbols Meaning
The following symbols are sticked in or on the instrument; user should comprehend the meaning.
Symbols Meaning
Caution, consult accompanying documents. Note: suggesting the users
consult to accompanying documents to get important safety information.
01
For in vitro diagnostic use
2.2 Model
After inoculating the sample into blood culture bottle, put the bottle into this system and incubate
itat body temperature. As the microbes rapidly grow, metabolic gase is produced. It leads to a change
02
of the color of the bottle sensor. Then, the system recognizes the change and analyzes by multiple
algorithm, and give the result by sound alarm and door flashing and presenting on the operation screen.
2.4 Application
The system should be used with the blood culture bottles produced by Zhuhai DL Biotech. Co.,Ltd..
And it is used for the detection of microbial in human blood or other sterile body fluids under normal
conditions through in vitro culture by medical, health, and scientific research organizations
The device records the patient's names, barcodes and blood test results. All the data is stored in the
database which is set for authority control;
The equipment reserves a wired network interface. The owner of the equipment should ensure that
the equipment is in a safe network condition. It is forbidden to install other software, and refer to the
following inspection items.
03
2.7 Software version
V1.
Remark: 1. The way to differentiate the Host unit and Sub unit is to check if there is a monitor.
The data in Sub unit is designed to export to Host unit and be recorded.
Accuracy 100%
Repeatability 100%
Fuse Requirement F2AL 250V
3 Instrument Installation
The Blood Culture System must be placed on a flat, solid working table.
04
3.1.1 Circuit wiring
Before replace insurance tube inside fuse, please insure that system and power supply are closed.
Fuse specification: 2A/250V/20mm, Type code: F2AL250V.
Display
Constant temperature system for Detection result shown
(Human-PC
blood culture detection cell in the interface
interaction)
Bottle
swinging
3.1.2. Configuration:
Standard Configuration
SN Description Quantity
1 Main Unit 1
SN Description Quantity
1 Power Supply 1
2 User Manual 1
3 Keyboard 1
05
4 Mouse 1
5 Tag 1
6 Quality Certificate 1
3.1.4. Installation
Place the instrument on a stable flat surface, make it more than 20cm away from the wall or other
obstacles;
1.Connect the monitor communication cable to the VGA in the rear of the machine;
06
The Rare view:
HDMI
VGA Instrument
Switch
Network Cable
socket
Output cable to
Sub unit
USB Socket
Dialing code
Power Switch
Dial to identify each Sub detection bin. The code 0001 represents 1 for Sub 1 , the 0010 represents
2 for Sub 2, and so on. The connection and dialing are as below:
07
4 Understanding The System
It contains touch screen, detection system, embedded control system, independent cell heaters,
constant temperature control system and mechanical system.
1
1 Frame
5
2 2 detection module
6 3 Indicator light
3 7
4 Door
8
5 Power supply
4 9 6 Barcode scanner
7 Swing structure
8 Door sensor
9 Display
10 10 Bottom shell
11 Incubator
11
It contains detection system, independent cell heaters, constant temperature control system and
mechanical system.
1 1 Frame
2 2 detection module
5
3 Indicator light
6
7 4 Door
3
5 Power supply
8
6 Barcode scanner
4 9
7 Swing structure
10
8 Door sensor
9 Bottom shell
10 Incubator
08
4.2 Software Introduction
4.2.1. Start up
Turn on the power switch on the instrument rear panel , and press the instrument switch. After the
system boot is completed, the following screen will be displayed (Pic. 1) :
Pic. 1
The menu:
Curve: To check the test value and growth curve of the bottle, and print the curve and data.
Setting: To change the specification of alarm setting, detection time, ability of the cell and gateway,
IP and so on
Debugging: To calibrate the detection racks and cells and Touch screen calibration
About: To see the version of the software and the manufacture information
09
The icon explanations:
Click on the bottle icon to input the patient name and view the information, including barcode and
bottle type, loading time, alarm time, time length of positive sample and the growth curve as shown
in Pic. 2.
Pic. 2
10
4.2.2. Bottling
Open the door, scan the barcode of the bottle and insert it into the unit and close door. the following
screen will be displayed (Pic. 3) :
Pic. 3
Attention:
1.Before bottling, check every tested bottle carefully. When the bottle is damaged, it can’t be used.
2.Barcode must be input when bottling. Otherwise, it’s marked anonymous bottle and presented on
the screen; If insert the bottle without clicking on the button “confirm”, it will be marked as
anonymous bottle, too.
3.When take the anonymous bottle out, its bar code should be recorded, avoiding the lost of its
information.
4.If the bar code is used before, the system should have saved its data. Then this bottle is not allowed
to be loaded again.
5.Please insert the bottle in available incubation unit, otherwise, system won’t record any data or give
any warning.
6.Please read the instruction of blood culture bottle for further reference.
11
4.2.3. Bottle Unloading
Open the door and take out the bottle. Once the bottle is unloaded, its bar code should be shown up
on the screen (Pic.4). Click on “Unload” . If it’s anonymous, input the bar-code by scanning it.
Pic. 4
4.2.4. Debugging
Click on “Debugging” on the home page, Pic. 5will show up. Only authorized engineer is allowed
to operate further.
Pic. 5
12
4.2.5. Print Curve and growth value
Connect the wire between the printer and the instrument, start up the instrument, then press “Esc”
on the keyboard, we can go to the desktop, and search the printer from the System/Administration
/Printer.
Check the website of the printer’s brand, and download the driver for window 10, and install it in
the system.
Input the Barcode (capital letters ) or Cell No., then click Search. It shows the value list of the certain
bottle, shows as Pic 6.
Pic. 6
Click “Curve” icon to view the growth curve, X represents the detection time, Y represents the test
value, shows as Pic 7.
13
Pic. 7
Click “Print” , the dialog box will come out, select the start time and the end time(less than 7 days, or
it will print too much) . Then it shows as Pic.8, and print the report.
Pic. 8
14
4.2.6. Statistics
Click “Details”, it presents the information of the cells with loading time, name, barcode, bottle type,
status, alarm time, and time length of positive sample.
Pic. 9
Click “ Total” , it presents the total name of positive or negative samples during selected period.
Pic. 10
15
4.2.7. Setting
Pic. 11
It presents the system parameter setting, such as test time, time limit for CODE, alarm sound, scanner
light setting, invalid cell, Temperature.
Phrases Explanation:
Test Time ------ To set the maximum test time (day). The system default is 5 days. Users are allowed
to set the time freely.
Time limit for CODE ------ To set the maximum time of scanning the barcode when take out the
anonymous bottle.
16
4.2.8. About
Click “About” to view the version of current system, it indicates the information about the software:
the version, time, property, website link, address of the producer, as in Pic. 12.
Pic. 12
4.2.9. Restart
Click on “Restart” , to choose to reboot the program, as shows in Pic. 13 (Data will be saved every
10 minutes. When the incubation period is less than 10 minutes, there is no data in the system).
Pic. 13
17
Attention & requirements
2.The installation environment should be clean and anti-dust at the temperature below 35℃ , avoiding
that the air conditioner blows directly the instrument. Power supply should be stable. 1000VA is
preferred.
3.Please inspect the blood culture bottle before using it. If broken, please stop using. Otherwise,
it may damage the instrument.
4. The instrument screen is touchpad. Please don’t use sharp objects to touch the screen. If touch
screen function is lost, users can operate it with a mouse.
5. If the instrument fails, please record the information on the display, and notify the relevant
engineering and technical personnel.
6. Used blood cultures should be dealt with well in order to prevent biological contamination.
7. This product should be used by qualified health care workers in the medical facilities.
Contraindications
None.
Wipe the instrument with a wet towel, then dry it with clean towel.
First, check the inside situation. If the incubation unit is contaminated, take the leaky bottles out of
the instrument immediately. If the shelf is contaminated, uninstall it and immediately disinfect the
bottle surface with disinfectant(containing 5% chlorine) diluted by 20 times. Let the surface and
bleach liquid fuse for 15-30 minutes, then wipe it with a wet towel. At last dry it with clean towel.
18
5.3. If the bottle holder need cleaning, please follow bellowing steps:
1.Insert absorbent material (such as gauze) in the cell to suck liquid. Then throw material into the
biological waste drum to prevent biological contamination.
2.Use gauze to clean the inner wall of the cell with 10% bleach. Then throw it into the biological
waste drum.
3.Insert several layers of gauze, which has been immersed in10% bleach, into the cell for 15-30
minutes’ disinfection.
5.Use another gauze moistened with distilled water to scrub the inner wall of the cell.
7.If the holder is disable, please mark it and notify the relevant engineering and technical personnel.
Description and
Theme Recovery procedure
Error Messages
1.Check the electrical outlet and the power cord connection.
2.Check the fuses by disconnecting the power cord from the
There is no power to the
power entry module, then open the fuse cover.
system when the power
3.For continued protection, replace only with fuses of the type
switch is turned ON.
andrating indicated on the rear panel of the instrument.
Check the power cord plug to see if it is short-circuit.
The start buttoncannot 1.Check the button and its welding connection and its power
start the instrument. connection.
19
2.Check the main board and replace it if it’s broken down.
3.Use a multi meter to check the circuit of main board.
4.If the problem persists, please contact your DL representative.
Remove the adapter plate and check the data line connection
The screen presents
between detection rack and the incubator.
communication failure.
If the problem persists, please contact your DL representative.
Screen touch 1.Check the communication line between the display and main
function failure. board. Re-connect it if necessary.
Such as, the arrow 2.Calibrate the screen.
won’t follow the 3.If the problem persists, it might be the display internal
touch. problem, please contact your DL representative.
21
1.Give no report, if it
excesses maximum 1. Switch the system OFF.
incubation time. 2. Wait 10 to 20 seconds.
2.The system halted. 3. Switch power ON again.
3.Not showing bottle
number, after scanning If the problem persists, it might be the software problem. The
the barcode. software may need to be re-installed or upgraded. please
4Keep presenting contact your DL representative.
wrong warning.
22
6 After-Sales Service
1.If any failures, caused by material or manufacturing, happen within 1 month since the delivery
date, DL is responsible to replace it with a new instrument and take care of all the delivery expense.
2.If any failures, caused by material or manufacturing, happen within 12 months since the delivery
date, the manufacturer provides with free spare parts. Customers bear the cost of freight back to
repair department, while DL bear the freight charge of delivery back to customer.
3When the warranty is expired, DL only charges for favorable price of spare parts.
This warranty did not cover unexpected use, the use in the environment which does not meet the
requirements, and the damage from unauthorized repair or refit.
7.1.The instrument is fixed tight in the carton, anti-rain, anti-moisture and anti-shock. During the
delivery, it is not allowed to shake fiercely and collide, which may seriously damage the instrument.
7.2. If the instrument is not used for a long period of time, it should be packed, anti-rain,
anti-moisture and anti-shock. And it should be stored at the temperature - 10℃~40℃, in the
ventilated indoor with relative humidity≤80% and no corrosive gases.
BT24 1503080007
BT48 1503080008
BT60 1503080009
BT120 1503080010
BT180 1503080011
BT240 1503080012
23
Manufacturing Enterprise
Company Name: Zhuhai DL Biotech Co., Ltd.
Address: 19 Jinhaian Rd, Sanzao Town, Jinwan, Zhuhai, Guangdong, 519041, P.R.China
Tel86-756-6292028 Fax86-756-3391389
Websitewww.medicaldl.com E-maildl@medicaldl.com