%RH, Wet bulb, Dew point, Temp.
, type K/J Temp
PRECISION
HYGROMETER
Model : HT-3017 ISO-9001, CE, IEC1010
LUTRON ELECTRONIC
%RH, Wet bulb, Dew point, Temp., type K/J Temp.
PRECISION HYGROMETER
Model : HT-3017
FEATURES
* Precision hygrometer to measure the Humidity ( %RH ), Optional Type K thermocouple probe.
Wet bulb Temp., Dew point Temp., Temperature value at Accessories AC to DC 9V adapter.
the same time and show all the 4 kind reading value in the USB power to DC 9V converter.
same LCD. USB cable, USB-01.
* Type K, Type J thermocouple thermometer. RS232 cable, UPCB-02.
* Fast humidity measuring response time. Data acquisition software, SW-U801-WIN.
* Separate probe, easy for operation of different Excel data acquisition software, SW-E802.
measurement environment. Hard carrying case, CA-06.
* LCD with backlight, easy reading.
* Can default auto power off or manual power off. ELECTRICAL SPECIFICATIONS (23± 5 ℃)
* Data hold, record max. and min. reading.
* Microcomputer circuit, high accuracy. Humidity/ Temperature
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter.
* RS232/USB PC COMPUTER interface. Range 5 % to 95 % R.H.
Humidity Resolution 0.01 % R.H.
GENERAL SPECIFICATIONS Accuracy ± 2% R.H.
Circuit Custom one-chip of microprocessor LSI @ 20 %RH to 80 %RH
circuit. @ 10 ℃ to 40 ℃
Display LCD size : 52 mm x 38 mm
LCD with green backlight ( ON/OFF ). Range 0 ℃ to 50 ℃,32 ℉ to 122 ℉.
Measurement Humidity/Temp. : %RH/℃ or ℉. Temperature Resolution 0.01 degree
Unit Dew point ( Humidity ) : ℃ or ℉. Accuracy ℃ ± 0.5 ℃.
Wet bulb ( Humidity ) : ℃ or ℉. @ 5 ℃ to 35 ℃. ℉ ± 1.0 ℉.
Type K/J thermometer : ℃ or ℉.
Advanced * Auto power OFF management Dew Point ( Humidity )
setting * Set beep Sound ON/OFF
* Set thermometer type to Type K or Type J ℃ Range -25.3 ℃ to 48.9 ℃
* Set temperature unit to ℃ or ℉ Resolution 0.01 ℃
Temperature Automatic temp. compensation for the ℉ Range -13.5 ℉ to 120.1 ℉.
Compensation humidity function and the type K/J Resolution 0.01 ℉.
thermometer. Remark :
Data Hold Freeze the display reading. * Dew Point display value is calculated from the
Memory Recall Maximum & Minimum value. Humidity/Temp. measurement automatically.
Sampling Time Approx. 1 second. * The Dew Point accuracy is sum accuracy value of
of Display Humidity & Temperature measurement..
Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable Wet bulb ( Humidity )
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable ℃ Range -21.6 ℃ to 50.0 ℃
USB-01 will get the USB plug. Resolution 0.01 ℃
Operating 0 to 50 ℃. ℉ Range -6.9 ℉ to 122.0 ℉.
Temperature Resolution 0.01 ℉.
Operating Less than 85% R.H. Remark :
Humidity @ Meter only. * Wet bulb display value is calculated from the
Power Supply * AAlkaline or heavy duty DC 1.5 V battery Humidity/Temp. measurement automatically.
( UM3, AA ) x 6 PCs, or equivalent. * The Dew Point accuracy is sum accuracy value of
* ADC 9V adapter input. ( AC/DC power Humidity & Temperature measurement..
adapter is optional ).
Power Current Normal operation ( LCD Backlight is OFF) : Type K/J thermometer
Approx. DC 3.5 mA.
* AIf LCD backlight on, the power Sensor Resolution Range Accuracy
consumption will increase approx. Type
12 mA. Type K 0.1 ℃ -50.0 to 1300.0 ℃ ± ( 0.2 % + 0.5 ℃ )
Weight 489 g/1.08 LB. -50.1 to -100.0 ℃ ± ( 0.2 % + 1℃)
Dimension 177 x 68 x 45 mm 0.1 ℉ -58.0 to 2372.0 ℉ ± ( 0.2 % + 1℉)
(7.0 x 2.7x 1.9 inch) -58.1 to -148.0 ℉ ± ( 0.2 % + 1.8 ℉ )
Accessories * Instruction manual...........................1 PC Type J 0.1 ℃ -50.0 to 1200.0 ℃ ± ( 0.2 % + 0.5 ℃ )
Included * Humidity/Temp. probe..................... 1 PC -50.1 to -100.0 ℃ ± ( 0.2 % + 1℃)
* Soft carrying case............................1 PC 0.1 ℉ -58.0 to 2192.0 ℉ ± ( 0.2 % + 1℉)
-58.1 to -148.0 ℉ ± ( 0.2 % + 1.8 ℉ )
* Appearance and specifications listed in this brochure are subject to change without notice. 1506-HT3017