Orthoralix 8500 Dde
Orthoralix 8500 Dde
Installation Manual
Panoramic X-ray System
Chapter 3 - Safety
Emergency Stop (not on all models) .................................................................... 3-1
Radiation Protection ........................................................................................... 3-1
Electrical Hazard ................................................................................................. 3-2
Mechanical Hazard .............................................................................................. 3-2
Explosion Hazard ................................................................................................ 3-2
Laser Radiation Hazard ....................................................................................... 3-3
Anatomical Reference Indications ....................................................................... 3-4
Tubehead Mineral Oil Safety .......................................................................................3-4
Proper Disposal of Electronic Equipment ............................................................. 3-5
032-0253-EN Rev 3 -v
Orthoralix 8500 DDE
Intended Use
The Orthorolix 8500 DDE Panoramic Dental X-ray System is to be used as an
extraoral source of X-rays for imaging of the dento-maxillo-facial area.
This manual is intended to assist the user in the safe and efficient operation
of the equipment described. The equipment must be used in accordance with
the procedures contained in the manual and must not be used for purposes
other than those which are described herein.
CAUTION
Damage to the machine or a loss of data is possible.
WARNING
Serious or fatal injury is possible.
WARNING
There is an electrical hazard. Serious or fatal injury is possible.
WARNING
There is a laser radiation hazard. Serious or fatal injury is possible.
WARNING
There is a radiation hazard. Serious or fatal injury is possible.
Major Components
8
2
1
3
7
ATTENTION RAYONS - X
OPERATION SEULEMEMENT PAR
DU PERSONNEL AUTORISE. VOIR
6
MANUEL DE L’OPERATEUR.
CAUTION X-RAYS
TO BE OPERATED ONLY BY
AUTHORIZED PERSONNEL. SEE
OPERATOR’S INSTRUCTIONS.
5
4
In preparation for the panoramic exposure, the motorized rotating imaging assembly is moved
out of the way allowing the patient to move into the initial position. The head support arms,
hand grip, bite guide, and positioning mirror are used to establish the initial patient position.
Lasers aid in establishing the final position of the patient’s head.
1. Hand grip
2. Bite block (can be replaced with the chin rest)
3. Head support arms
REF
E-Stop
Not on all models
ATTENTION RAYONS - X
OPERATION SEULEMEMENT PAR
DU PERSONNEL AUTORISE. VOIR
MANUEL DE L’OPERATEUR.
CAUTION X-RAYS
TO BE OPERATED ONLY BY
AUTHORIZED PERSONNEL. SEE
OPERATOR’S INSTRUCTIONS.
ATTENTION RAYONS - X
Gendex Dental Systems OPERATION SEULEMEMENT PAR
Des Plaines, IL USA DU PERSONNEL AUTORISE. VOIR
MANUEL DE L’OPERATEUR.
0.4 IEC 60336 2.5 Al/85 IEC 60522
CAUTION X-RAYS
Label No. 408-0343
CAUTION
LASER RADIATION
DO NOT STARE INTO BEAM
Label Description
Check Jumper
Position the jumper for the
Position the jumper for the input
Input voltage being used.
This tag must be removed
Before reassembling cover.
J9 J8 J10 J9 J8 J10 voltage being used. This tag must
Placez le pont en fonction
de La tension de réseau utilisée.
120V 230V be removed before reassembling
Cette étiquette doit être retirée avant de replacer le couvercle.
408-0390P1
Poner el jumper en relación con la tensión de red utilizada.
Esta etiqueta debe ser removida ante de tapar.
System ID
Certification: Product complies
with FDA performance standards
for radiation and laser products
except for deviations pursuant to
Laser Notice No. 50, dated 24 June
2007.
Tubehead
Certification: Product complies
with radiation performances
standard under the Federal Food,
Drug, and Cosmetic act.
X-ray Warnings
ATTENTION RAYONS - X
OPERATION SEULEMEMENT PAR Caution X-rays
DU PERSONNEL AUTORISE. VOIR
MANUEL DE L’OPERATEUR. To be Operated only by Authorized
CAUTION X-RAYS
TO BE OPERATED ONLY BY Personnel. See Operator’s
AUTHORIZED PERSONNEL. SEE
OPERATOR’S INSTRUCTIONS. Instructions.
WARNING - THIS X-RAY UNIT MAY
BE DANGEROUS TO PATIENT AND
OPERATOR UNLESS SAFE Warning - This X-ray Unit may be
EXPOSURE FACTORS,
OPERATING INSTRUCTIONS AND
MAINTENANCE SCHEDULES ARE
Dangerous to Patient and
OBSERVED.
Operator unless Safe Exposure
Factors and Operating Instructions
and Maintenance Schedules are
Observed.
Laser Aperture
CAUTION
LASER RADIATION
DO NOT STARE INTO BEAM
Wavelength: 650nm, output
Power < 1mW
Complies with FDA performance
standards for laser products
except for deviations pursuant to
Laser Notice No. 50, dated June
24, 2007
CLASS II LASER PRODUCT
Symbol Description
IEC Label
Type B: Protection against electrical
shock (IEC 60601-1)
Caution
Consult written instructions in this
Manual.
Warning
This X-ray unit may be dangerous to
patient and operator unless safe
exposure factors and operating
instructions are observed.
Warning
Electrical Hazard, Authorized Personnel
only.
Warning
Laser energy present.
X-ray Generator
Recycle Label
Dispose of in accordance with your
country’s requirements. There is material
in the system that must be separately
collected and recycled in accordance with
the European Waste Electrical and
Electronic Equipment (WEEE) Directive.
Symbol Description
Earth Ground
Emergency Stop
(not on all models)
This system features continuous operation (stand-by mode). X-rays are emitted only when the
exposure handswitch button is pressed.
With correct handling and installation, proper maintenance and servicing will ensure safe and
efficient operation. The equipment must never be used if there are any electrical, mechanical,
or radiation defects whatsoever.
Refer to the Installation Manual for the recommended maintenance activities (for example,
checking wear and tear on cables, belts, and gears). Modifications and additions to the
equipment must be carried out only by personnel or third parties that are expressly authorized
by, and must comply with the applicable legal requirements as well as with the generally
accepted technical regulations.
If an emergency arises where it becomes necessary to interrupt a scan, use the following
procedure.
1. Press the Emergency Stop button.
2. When you have determined that you can safely move the patient away
from the device, make sure the patient’s head will not touch the top of the
device and help the patient move away from the unit.
Radiation Protection
The system implements various built-in measures to prevent excess radiation (leakage,
secondary, and scattered) from reaching the patient, the operator, or other people.
It is the responsibility of the qualified Radiation Protection Inspector to ensure that the proper
measures are taken to prevent undue radiation exposure to personnel and to the public at large.
All personnel in the examination room must exercise radiation safety procedures.
To protect the patient against unnecessary radiation, other accessories, in addition to those
provided on the system, can be used whenever necessary.
Care must be taken that these accessories do not interfere with the proper radiographic
operation of the apparatus. For example, lead shield collars should be avoided as they may
block the useful X-ray beam.
Electrical Hazard
Only qualified service personnel should remove the covers or otherwise obtain access to parts
of the system.
The system may only be used in rooms which comply with the relevant national and/or
international legislation and recommendations (for example, CE and others) concerning
electrical safety in rooms used for medical purposes.
Always disconnect or switch off the equipment before cleaning or disinfecting to avoid the risk
of shock.
Do not allow any liquid to enter the system, as this could cause short-circuits and corrosion.
Mechanical Hazard
When the system is in use, ensure that no body parts or clothing can get caught in any moving
parts.
Explosion Hazard
This system is not suitable for use in the presence of flammable gases or vapors. Certain
disinfectants vaporize and form explosive mixtures. If such disinfectants are used, the vapors
must be left to dissipate before the system is powered up again. For more information, refer to
“Caution with Disinfectants” on page 4-1.
CAUTION
Use of controls or adjustments, or performance of procedures other than
those specified in this manual, may result in hazardous laser radiation
exposure.
This equipment is designed in accordance with IEC 60825-1:2001. Although compliant with
laser safety rules, it is good practice for the operator to avoid exposing anyone’s eyes to laser
beams.
The position of laser sources are identified by means of the following warning labels (black on
yellow).
The laser for the lighting of the midsagittal plane (vertical laser beam) is positioned on the
column above the rotating imaging assembly.
Two lasers for the lighting of the Frankfurt plane (horizontal laser beam) and the laser beam that
is aligned with the lateral incisor roots (cuspid laser beam) are positioned to the left of the
sensor.
CAUTION
LASER RADIATION
DO NOT STARE INTO BEAM
CAUTION
LASER RADIATION
2 lasers:
Frankfurt plane laser
Cuspid laser
Once the Pan procedure has been performed, positioning references (L and R) will be displayed
on the acquired image:
• L (Left) to indicate the left side of the patient’s mouth
• R (Right) to indicate the right side of the patient’s mouth
On symmetrical images, both L and R are displayed. No letter is placed in correspondence of the
center of the mouth.
This symbol on the products and/or accompanying documents means that used
electrical and electronic products should not be mixed with general household waste.
For proper treatment, recovery, and recycling, please take these products to designated
collection points where they will be accepted on a free-of-charge basis. Alternatively, in some
countries, you may be able to return your products to your local retailer upon the purchase of
an equivalent new product.
Disposing of this product correctly will help to save valuable resources and prevent any
potential negative effects on human health and the environment which could otherwise arise
from inappropriate waste handling.
Please contact your local authority for further details of your nearest designated collection
point. Penalties may be applicable for incorrect disposal of this waste in accordance with
national legislation.
NOTE: For Business users in the European Union
If you wish to discard electrical and electronic equipment, please contact
your dealer or supplier for further information.
Patient Protection
All parts coming in direct contact with the patient must be kept clean and
disinfected.
Always replace the disposable hygienic covers on the bite before positioning
a new patient. These covers should be stored in a clean, dry environment and
not exposed to direct sunlight or UV radiation.
The bite block and chin rest can be disinfected by immersion in a cold
sterilizer, and can be steam sterilized in an autoclave up to 125°C.
All other parts used for patient positioning that come in direct contact with the
patient must be kept clean and disinfected. Use an appropriate disinfectant
following the manufacturer's guidelines.
CAUTION
The method of disinfection used should comply with current regulations and
recommendations, including those concerning the prevention of explosive
hazards.
CAUTION
It is not recommended to use spray disinfectants as the disinfectants may
permeate into the system and cause short circuits or corrosion. If an atomizer
is used for disinfection, follow the procedure provided below to protect the
system.
1. Turn the system off and disconnect the system from the mains supply.
2. Cover the system carefully with a plastic sheet.
3. Allow the system to cool down to prevent convection currents from
drawing the vapors into the system.
4. Disperse the disinfectant.
5. Allow the vapors to dissipate before the equipment is returned to use.
6. Remove the plastic sheeting.
7. Reconnect the system to the mains supply and turn the system on.
3. Wait until “dE” is no longer seen in the mA display and press Reset.
If the above mentioned operation is carried out correctly, the system is reset
to the ready state. The Ready LED on the main control panel lights up to
indicate the reset condition. The ready state is a prerequisite for starting any
exposure program.
On the computer, exit the active software. To turn off the computer, choose
Start > Shut Down. In the window that displays, ensure the selection is Shut
down and then click OK.
Turn off the system by pressing the power switch at the base of the column.
Power Switch
Refer to page 5-1 for the location of the power switch.
ATTENTION RAYONS - X
OPERATION SEULEMEMENT PAR
DU PERSONNEL AUTORISE. VOIR
MANUEL DE L’OPERATEUR.
CAUTION X-RAYS
TO BE OPERATED ONLY BY
AUTHORIZED PERSONNEL. SEE
OPERATOR’S INSTRUCTIONS.
88 88
88 88
Reset system
88
ON - System ready
ON Emitting X-rays
ON - Projection selected
Not used
ON- Patient size selected
(child, small, medium, large)
Selecting a Projection
When the system is turned on, the Standard Pan modality is set and the
relevant projection LED lights up.
Reset
The Reset button is used to return the
rotating imaging assembly to the default
starting position.
NOTE: When the rotating imaging
assembly is repositioned, X-rays are
not emitted. The X-ray LED turns off
and there is no audible alarm.
If the above mentioned operation is carried out correctly, the system is reset
to the ready state. The Ready LED on the main control panel lights up to
indicate the reset condition. The ready state is a prerequisite for starting any
exposure program.
Demo Mode
CAUTION
As the demo mode is usually followed by the actual exposure, remove all
metal articles (glasses, removable dentures, earrings, etc.) before positioning
the patient.
If he/she is to be fitted with a lead-lined apron for radiation protection, make
sure that the neck is not covered, because this will cause unexposed areas in
the radiograph.
Press the Reset button to return the rotating imaging assembly to the default
starting position.
ATTENTION RAYONS - X
OPERATION SEULEMEMENT PAR
DU PERSONNEL AUTORISE. VOIR
MANUEL DE L’OPERATEUR.
CAUTION X-RAYS
TO BE OPERATED ONLY BY
AUTHORIZED PERSONNEL. SEE
OPERATOR’S INSTRUCTIONS.
ON - Power on
Reset system
Power-save mode
Press the power-save mode button to toggle the unit in and out of power-save
mode.
Power ON LED
This LED signifies when the system is not in the power-save mode.
Constant pressure for more than 2 seconds will cause either of these buttons
to gradually accelerate until the movement is at the maximum speed.
When aligning the rotating imaging assembly with the patient’s head, use the
left and right buttons to shift the cuspid light beam to the correct position
(move the assembly away from or towards the column).
Lasers Control
Press this button to turn the lasers on and off.
Exposure Handswitch
When the exposure handswitch button is pressed, X-rays are emitted. The
system provides two alerts: the X-ray LED on the main control panel lights up
and an audible alarm is emitted. (When a remote exposure switch is installed,
the same two alerts are provided at the switch.) The actual X-ray emission
starts 5 seconds after pressing the button. Must hold button for entire
exposure.
System Icons
These system icons display in the Windows system tray (near the clock):
• Status icon (the look of the icon changes according to the status)
• GX driver icon (only displays when the driver is installed correctly)
Status Icon
Exposures can only be made when the Status icon is green and blue as shown
in the following graphic:
This green and blue icon indicates the correct operation of the
Orthoralix 8500 DDEelectronics (including the CCD sensor), the connection
between the Orthoralix 8500 DDE and the CCD sensor, and the connection
between the Orthoralix 8500 DDE and the computer or networking device.
Icon Status
Large red X:
- Faulty hardware condition
- Sensor calibration required
GX Driver Icon
When the Gendex driver is installed correctly, the Gx icon is shown in the
system tray.
Right-click this icon to display a menu for viewing and modifying data and
parameters common to all the Gendex imaging system products.
Refer to Appendix B, Technique Factors, for the factory settings for this
projection. For the X-ray projection geometry, refer to Appendix C, X-ray
Projection Geometries.
2. Press Reset. The system is in the ready state when the Ready LED is lit.
5. Insert the supplied bite block (A) in the related seat. Tighten it firmly by
rotating the knob (C) (clockwise=tighten, counter-clockwise = loosen).
NOTE: For an edentulous patient, it is possible to use the supplied
transparent chin rest (B).
6. For hygienic reasons, cover the bite block with a disposable bite-guide
cover. These have been designed and developed for use with the
Orthoralix 8500 DDE system.
2. Ask the patient to step into position in front of the handgrip with their
head between the head support arms.
3. Position the rotating imaging assembly up or down as required.
4. Place the patient close to the bite guide (or chin rest). The bite guide must
be at the level of patient’s mouth (occlusal plane); the chin rest must be
aligned with the patient’s chin.
5. Have the patient firmly grasp the handgrip with both hands as far forward
from the bite guide holder as possible. Ask the patient to move their feet
forward and together. Ensure the patient’s body is straight and slanted,
with the patient hanging slightly from the handgrip (according to the
patient’s physical ability to maintain this uncomfortable position during
the examination). The purpose of this position is to stretch the spine as
6. If using the bite guide, ensure that the patient is not biting too far forward
(beyond the grooves).
NOTE: Edentulous patients should use the chin rest. If the chin rest is
used for dentate patients, he/she should bite so as to align lower
and upper incisors.
CAUTION
Performing procedures other than those specified in this manual, or using the
controls or adjustments other than as specified in this manual, may result in
hazardous laser radiation exposure.
2. With the center of the dentition aligned with the midline of the bite guide
or chin rest, check that the front vertical light beam is centered on the
patient’s mid-sagittal plane. If necessary, gently turn the patient’s head
to adjust.
3. Note the parallel horizontal stripes projected on the side of the patient’s
face. These horizontal stripes must be in alignment with the patient’s
Frankfurt plane (the plane passing through the lower margin of the left
orbit and the tragus at the external auditory meatus).Gently raise or lower
the rotating imaging assembly to tilt the patient’s head backwards or
forwards until the laser beam strips are in alignment with the patient’s
Frankfurt plane.
Patient’s
Frankfurt
plane aligned
with these
parallel laser
beams
NOTE: In step 4, the cuspid light beam indicates the front most position that
the X-ray focal trough will reach during the exposure. Normally, setting
the cuspid light beam at the center of the cuspid aligns the front most
position of the focal trough with the roots of the lateral incisors.
4. While ensuring that the patient doesn’t move, use the forward button or
the backward button to adjust the position of the rotating imaging
assembly until the cuspid light beam is positioned at the center of the
cuspid.
Nominal position is 5 mm
behind the groove of the
bite block
- If the incisor is straight,
adjust to Y < 5 mm
- If the incisor protrudes,
Cuspid light adjust to Y > 5 mm
beam
centered on
the cuspid
Take Exposure
1. When the patient’s head is properly positioned, gently close the head
support arms until they are pressed lightly against the patient’s head.
NOTE: Use care when closing the head support wands. Rapid or abrupt
movement of the lever might cause the arms to snap
uncomfortably against the patient’s temples.
2. Ask the patient to press his/her tongue to the palate and to stay still
during the exposure. Make sure the patient does not shift position during
the exposure process.
3. Verify the correct system status icon displays in the system tray:
NOTE: The exposure handswitch button must be kept firmly pressed during
the entire exposure procedure. Otherwise X-ray emission and rotating
imaging assembly movements will be stopped and an error code is
displayed. If this should happen, the unit must be reset, the patient
must be re-positioned, and the exposure must be repeated.
CAUTION
Repeating the exposure subjects the patient to unnecessary radiation. Keep
the exposure handswitch button firmly pressed during the entire exposure
procedure to prevent repeating the exposure.
5. During X-ray emission, preview images are continuously provided for the
imaging software to display. For imaging software that provides a preview
window, verify the correct image acquisition.
6. When the exposure cycle has been concluded, release the exposure
handswitch button. The final image is transferred to the imaging
software.
7. Open the head support arms. Ask the patient to move out.
To process the final image, refer to the manual for your imaging software.
Press the Function button repeatedly to scroll through the set of projections.
The lamp turns on for the projection selected (shown below with Lateral TMJ
selected).
Refer to Appendix B, Technique Factors, for the factory settings for each
projection. For the X-ray projection geometries, refer to Appendix C, X-ray
Projection Geometries.
Functional Button
Frontal Projection
Lateral Projection
Child Panoramic
NOTE: This standard panoramic projection is ativated when the patient size
selection is pedo (child).
Patient positioning
Operation Variation
The operation is similar to the Standard Panoramic projection except that in
the TMJ program, the longitudinal pre-positioning displacement of the rotating
imaging assembly is inhibited, and the equipment always starts the exposure
cycle from the reset position at Y = 5 mm.
Positioning Instructions
1. Press the Function button until the
Standard Pan is selected.
NOTE: The lasers are timed and they will switch off automatically
after 45 s.
The angle between the major condylar axis and the perpendicular to the
median/sagittal plane is assumed to be 10°.
If the exact angle is known (this can be determined using the submento vertex
exam), it is possible to optimize the orthogonal incidence of the X-ray beam
relative to the major condylar axis, ensuring the highest possible diagnostic
value of the exposure: Position the patient with the cuspid light beam to
coincide with the reference values indicated on the TMJ ruler – for both open
and closed mouth projections.
Frontal TMJ
Positioning Instructions
1. Position the patient with the white chin rest (without the ridge).
2. Align the lateral vertical light with the lower canine.
3. Instruct the patient to keep the mandible as protruded as possible, or
open, so as to displace the condylar head forward, under the eminence.
4. Tilt the patient’s head forward so as to have the Frankfurt plane slightly
(approximately 5°) beyond the horizontal (cervical spine slightly more
extended than with the positioning for a standard radiograph). An
excessive tilt backwards may involve the overlap of the orbital base/
palatal vault on the area of the condyle of diagnostic interest.
Operation Variation
In the TMJ program, the longitudinal pre-positioning displacement of the
rotating imaging assembly is inhibited, and the equipment always starts the
exposure cycle from the reset position at Y = 5 mm.
2. On the following screen, select the “Last Image” tab. On the “Last Image”
screen you will see a “Resend” button. Select the “Resend Button” and
the last image scanned will be resent to the imaging software.
The last scanned image of the patient can be resent to the imaging software
as long as the panoramic machine is not powered down or another patient is
scanned. If the image remains distorted, repeat the above steps. There are no
limitations on the number of times an image can be resent.
If this does not resolve the issue, please contact Gendex Technical Support at
1-800-323-8029.
Problem:
Head turned to
the left
Problem:
Head tilted too
far forward
Problem:
Head tilted too
far backward
Result: The occlusion plane is flat, ascending rami diverging. Often the
hard palate will obscure the upper root apices.
Problem:
Patient standing
too far forward of
cuspid light
beam
Result: The length of the image is shortened and the central incisors are
compressed.
Problem:
Patient standing
too far back of
cuspid light
beam
Result: The length of the image is expanded and the central teeth are
enlarged.
If image results are still not satisfactory after verifying and correcting
positioning or other procedural errors, the system projection geometry
parameters should be checked. Available for the Orthoralix 8500 DDEDEXpan
is a patented image phantom (part no. 4519-124-20471).
A radiograph taken with this phantom will reveal the exact performance of the
system. This can be taken by yourself or the designated service technician.
The image will also provide information necessary for making any required
adjustments. Instructions for the service technician are found in the
Installation Service Manual supplied with the system.
NOTE: In North America, information and assistance can be obtained by
contacting GENDEX Technical Support (phone number is listed on the
back cover). Otherwise contact your local Gendex representative.
Computer
The computer configuration must include a CD-ROM drive along with Windows
Vista, Windows XP Professional, or Windows 2000 and the configuration as
required by the operating system installed.
Hard Disk
The choice of the hard disk depends on the number and size of the images to
be stored. A hard disk with 80GB or greater is usually able to satisfy a standard
user’s requirements. However, for data security, it is recommended to also
back up the stored information and diagnostic images.
Video Card
If not included in the computer, a S-VGA Video card with 1024x768 mode, and
has at least 4 MB RAM must be installed along with the accompanying
software.
Network Card
An Ethernet 10/100 base T network card must be installed.
Monitor
To ensure correct viewing of the images, a multisync colored monitor with high
resolution capable of displaying 800x600 min., or 1024x768 pixels, is
recommended. A flat screen monitor is preferred. The resolution is related to
the mode selected by the video card. Contrast and brightness must be
properly adjusted.
Errors
In the event of critical or non-critical conditions that would prevent the proper
operation of the equipment, the exposure will be inhibited and an error
message will be displayed on the main control panel (see page 6-1). When a
fault occurs, a “EC” will be displayed in the mA display. The appropriate error
code number will be displayed in the kV display.
The error codes, type of fault, clearing method, and equipment condition are
listed in the following chart.
Collimator - 112-1143G1
Frequency 50-60 Hz
Max. line current rating 8.5A at 230V, 13.5A at 120V for 80kV and 10mA
High voltage waveform true DC, via square-wave output electronic power
converter (50kHz) and voltage multiplier/rectifier
Leakage radiation from the tubehead: less than 0.25 mGy/ hour at 1 m in any direction
Wavelength 650nm
Pulse Duration 45 s
Modality: Pan
Preset technique
Total exposure Time factors
Movement range of
(seconds)
mA
Standard panoramic 12 24 1.25 0-14 (ref.7) 68 72 76 6
The Gendex Orthoralix 8500 DDE is intended for use in the electromagnetic environment
specified below. The customer or the user of the Gendex Orthoralix 8500 DDEshould assure
that it is used in such an environment.
Harmonic Class A
emissions
IEC 61000-3-2
Voltage Complies
fluctuations/
flicker emissions
IEC 61000-3-3
The Gendex Orthoralix 8500 DDE is intended for use in the electromagnetic environment specified
below. The customer or the user of the Gendex Orthoralix 8500 DDE should assure that it is used in
such an environment.
d 1,2 P
Conducted RF 3 Vrms 3V
80 MHz to 800 MHz
d 1,2 P
IEC 61000-4-6
150 kHz to 80 d 2,3 P 800 MHz to 2,5 GHz
MHz
where P is the maximum output power
rating of the transmitter in watts (W)
Radiated RF 3 V/m 3 V/m according to the transmitter
manufacturer and d is the recommended
IEC 61000-4-3 80 MHz to 2,5 separation distance in meters (m).
GHz Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,a should be less than the compliance
level in each frequency range. b
Interference may occur in the vicinity of
equipment marked with following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and landmobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Gendex Orthoralix 8500 DDE is used exceeds the applicable RF
compliance level above, the Gendex Orthoralix 8500 DDE should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Gendex Orthoralix 8500 DDE.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
The Gendex Orthoralix 8500 DDE is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
Gendex Orthoralix 8500 DDE can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Gendex Orthoralix 8500 DDE as recommended below, according to
the maximum output power of the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Classification
Class and Type of Certified equipment: Orthoralix 8500 DDE belongs to the
Class I Type B equipment per IEC 60601-1 – 1990 / A2 - 1995 and Class Ib per
European MDD 93/42/EEC.
Standards
The Orthoralix 8500 DDE complies with the following standards:
Radiation protection:
• IEC 60601-1-3 Radiation protection/X-Ray Equipment
Electromagnetic Interference:
• IEC 60601-1-2
Lasers:
• IEC 60825-1
The Orthoralix 8500 DDE Tubehead complies with the following standards:
• IEC60601-2-7
• IEC60601-2-28
The certified components of the Orthoralix 8500 DDE system comply with
Radiation Performance Standards 21 CFR, Subchapter J, at time of
manufacture.
This symbol ensures that the product herein described meets the
provisions of UL 60601-1 standard.
In order to fulfil the obligations foreseen by the CE marking, the user is obliged
to report any accident involving the medical device; any alterations to its
features or in its performance – including insufficient user’s instructions –
which could cause death, injuries or health hazards to patient and/or
operator, to the competent Health Authorities. Such reporting must also be
promptly notified to the manufacturer or his agent, in order to permit the
fulfillment of the obligations foreseen for said manufacturer in the a.m.
European Directive.
Installation Manual
Panoramic X-ray System
0413