Practical Guide On Refraction
Practical Guide On Refraction
National Programme
for
Control of Blindness
A Practical Guide
on
Refraction
A. Equipment
1.Wall chart
2. Near-point VA card
3. Occluder
4. Lamp
B. Set-up
1. Patient wears habitual Rx for the distance being tested (when the clinician wants
to measure the patient's VA both with and without Rx. VA should be measured
without.Rx first)
2. Patient holds occluder (or uses his/her hand)
3. Patient views full chart
C. Step-by-step procedure
D. Recording
E. Example
Pinhole VA
A. Equipment
B. Set-up
C. Step-by-step procedure
1. Patient occludes eye not being tested. If both eyes are to be tested, occlude OS
first.
2. Instruct patient to read the smallest line of letters he/she can through the PH.
3. Urge patient to read the next smallest line, even ifhe has to guess.
D. Recording
E. Example
2
Amplitude of Accommodation: Pushup Method
A. Equipment
B. Set-up
C. Step-by-step procedure
I. Occlude OS.
2. Direct patient's attention to a row of letters one or two lines larger than his best
near VA.
3. Slowly move the chart closer to the patient and instruct the patient to report when
the letters first appear blurry.
4. Measure the distance from the chart to the patient' s spectacle plane in centimeters
and convert to diopters.
5. Repeat with the OD occluded
6. Repeat OU.
D. Recording
E. Example
1. Amp OD 7D
OS 7 D (pushup)
2.Amp OD6D
OS6D
OU 7 D (pushup)
3
Finger Counting Fields (Confrontation Fields)
A. Equipment
1. Occluder
2. Lamp
B. Set-up
C. Step-by-step procedure
1. Instruct patient: 'I will hold up one, two, or five fingers' (demonstrate) in your.
side vision. 'You keep looking right her (point) at my eye'. ' Without looking at
my fingers tell me how you see' .
2. Patient occludes OS/examiner closes OD.
3. Examiners looks at pt's open eye.
4. Examiner places closed fist in side vision equidistant between patient and
examiner, in each of the 8 positions and then exposes 1,2,or 5 fingers. See below.
5. If patient reports number of fingers correctly, move to new position and repeat #3.
If patient fails to count fingers correctly, move hand closer to center of field and
repeat.
6. When FCF for OD has been 'mapped' in 8 positions, patient occludes
OD/Examiner closes OS and then repeats steps 3 through 5.
7. Throughout, Examiner compares his/her presumably normal field to the patient's
field.
D. Recording
4
E. Example
X X
X
X
X
X
X X
5
Colour Vision
A. Equipment
1. Occluder
2. Lamp
3. Color vision test book or other material
B. Set-up
C. Step-by-step procedure
D. Recording
1. Write number of correctly identified plates, a slash mark, and then the number of
plates tested, for each eye separately, ie., a fraction.
2. Record the type of test used.
Notes
6
Cover Test
A. Equipment
....
1. VA chart
2. Near cover test target
3. Occluder
4. Overhead lamp
5. Horizontal and vertical prism bars
B. Set-up
C. Step-by-step procedure
7
Cover right eye observe left eye
UncoJer right eye Uncover left eye Alternatively cover right eye,
Observe both eyes Observe both eyes then left eye, observe
uncovered eye
·I I I I
Both eyes move No movement Both eyes move No movement Movement No movement
The occluder is held in front of the patient's eye for one second and eyes are observed as
indicated in the above chart. Before the occluder is placed in front of the other eye, a few
seconds should be allowed to elapse so the eye can return to their normal relationship.
Repeat the cover-uncover test several times to be sure an alternating tropia or intermittent
tropia is not missed.
a. Cover the patient's right eye, holding the occluder in front of the eye for
approximately one second, then quickly place it in front of the left eye.
b. Alternate the occluder back and forth between the two eyes several times,
maintaining coverage of each eye for one second.
c. Observe the movement of the right eye as the occluder is moved from the right to the
left eye; observe the movement of the left eye as the occluder is moved from the left
to right eye.
d. Always observe the eye that is being uncovered for movement.
e. Responses
1. No movement - ORTHO
2. Eye moves toward the nose as it is uncovered- EXO
3. Eye moves temporally as it is uncovered - ESO
4. Eye moves downward as it is uncovered - HYPER deviation of that eye.
f. Neutralize movement with prism. Alternate cover test is repeated as described in
steps a-d with prism in front of one eye. (The prism bar(s) may be held by the patient
or the examiner). Increase the amount of prisms until no movement is observed as
the eye is uncovered (or until the motion reverses to the opposite directjon).
8
Prism should be added as follows:
Observed motion Prism orientation
EXO Base in
ESO Base out
HYPER Base down
HYPO Base up
4. Patient holds near point target at 40 cm on the same plane as his eyes with overhead
lamp on the target.
5. Repeat step 1-3 at near.
D. Recording
cp ORTHO
-0-
+
ISO
ORTHO,ISO
Note
Lateral phorias reflect the relationship between the two eyes. Therefore they should not
be identified as R, L, or alt. However, R, L, or alt. Must be recorded for all vertical
phorias and all tropias.
E. Examples
CT cc--$-DandN
+
CT sc 20° RXT; 10 XP at N
CT c~ at D; 25° alt ET at N
9
Worth 4 Dot
A. Equipment
I. Red-green glasses
2. Worth 4-dot flashlight for near and/or box for distance.
B. Set-up
C. Step-by-step procedure
1. Turn on the Worth 4-dot stimulus for distance or near (40 cm)
2. Ask patient 'How many spots of light do you see: two, three, four, or five?'
3. If patient reports 2,3 or 4, the test is over.
4. If patient sees 5 spots, ask patient ' Where are the green ones (seen by OS), to the
right, left, above, or below the red ones (seen by OD)?'
D. Recording
E. Examples
10
Stereo psis
A. Equipment
1. Polaroid glasses
2. Stereo target
3. Overhead lamp
B. Set-up
C. Step-by-step procedure
1. Direct pt's attention to the first set of circles. Ask patient 'In target number one,
which of these circles appear closest to you? Which circle appears to be floating
above the· plane of the page?'
2. Continue tes_ting until patient gives two consecutive incorrect answers.
3. If the patient is unable to identify any of the circles correctly, repeat steps 1 and 2
using the animal targets and then repeat using the circle target.
4. If the patient is unable to identify any of the animals correctly, show the patient
the large targets and ask him or her to identify the geometric figures (s) he sees, or
show the patient the fly and ask him or her to pinch the wings of the fly.
D. Recording
E. Examples
11
Near Points of Convergence (NPC)
A. Equipment
1. Penlight
2. Red glass
3. Near accommodative target
4. Overhead lamp
B. Set-up
C. Step-by-step procedure
1. Patient is instructed to look at light (or other target) and report as soon as he sees
the target double.
2. Examiner moves target towards the patient, observing pt' s eyes until he sees one
eye lose fixation on the target. Note distance from patient's eye.
3. Move the target away from patient's eyes and note the distance at which the
patient's deviated eye regains fixation (patient will report single vision at this
distance if he reported diplopia as described below).
4. Patient should report double vision when the examiner sees one of his eyes lose
fixation. If patient reports diplopia, ask the patient to report when the target
becomes single.
If break and recovery are less than 15 cm, record the results. If either is greater
than 15 cm, do step C.5.
5. Repeat steps 1-4 with an accommodative target and a penlight as a target with a
red glass placed over one of the patient's eyes.
D. Recording
12
5. Record which eye deviated and in which direction
6. Record diplopia if patient reported seeing two targets
7. Record suppression if patient did not report seeing two targets
8. If examiner was able to move target to the bridge of the patient' s nose without the
patient losing fixation, record 'TTN' (for 'to the nose')
9. Repeat steps 2-8 for each target used
E. Examples
13
Hirschberg (Corneal Reflex)
A. Equipment
1. Penlight
B. Set-up
1. Patient removes Rx
2. Examiner faces patient with 50-70 cm between them
3. Examiner holds penlight
C. Step-by-step procedure
D. Recording
E. Examples
14
Extraocular Muscles
A. Equipment and set up
B. Step-by-step procedure
1. Like Hirschberg:
a. Tum on light/hold directly in front of patient
b. Instruct patient to tell you if,
- He ever sees the light double or become two:
- He feels any pain or discomfort
Move the light
- Move the light to other side, up, down, completing a big H pattern
- Throughout, point light at the patient's eyes, looking for asymmetries of
corneal reflex as the eyes follow the penlight, as in the Hirschberg
- Note: Do not move the penlight too far. Travel of 30-40cm from primary
position is sufficient to discover wake EOM's
- Throughout, observe:
a. Smoothness of movement
b. Accuracy of following the penlight
c. Extent of movement
C. Recording
1. If normal, write 'SAFE' or ' FESA'. You must know what these letters stand for:
S Smooth
A Accurate
F Full
E Extensive
2. If abnormal, record only the letters that apply and describe the abnormality,
whatever applies:
Jerky, unsteady, nystagmoid
failure to follow into (location).
restricted, lagging, non-commitment
3. Identify direction (s) of gaze that result in diplopia (double vision) and /or
discomfort.
4. If only one eye is the abnormal one, be sure to identify it.
5. Identify the possible paretic muscles.
15
D. Examples
1. EOMs: Safe
2. EOMs: Smooth, OS restricted left gaze and diplopia on left gaze.
3. EOM Diplopia on up-right gaze, OD lagging
4. EOM: Fesa, OD pain on left gaze
16
lnterpupillary Distance (IPD)
A. Equipment
B. Set-up
C. Step-by-step procedure
I. Distance P.D.
1. To tell patient to look at your open OS. Close OD.
2. Align zero-point of ruler on a landmark (e.g. nasal pupillary edge or limbus)
of pt's OD. Stabilize ruler by resting hand on patient.
3. Close your OS. Tell the patient to look at your new open OD.
4. Measure the corresponding landmark on patient's OS (e.g. temporal pupillary
edge or limbus).
5. Recheck the entire procedure.
@) OD
OS Examiner
OS- Pt.
OD~
Examiner
OD Pt. OS
OD
OS Examiner.
17
Net result
Although patient never looks at distance, you measure the P.D. with both eyes pointing
straight ahead, as if patient had been looking at distance.
l . Close your OD and tell patient to look at your open OS, tip of nose, or other
near
target, Align zero - point of ruler on a landmark of patient's OD (nasal
limbus)
2. Still using your OS, measure the corresponding landmark on patient's OS
(temporal limbus)
D. Recording
E. Examples
1. 64/61
2. 62/58
3. OD 30/28
4. OS 32/30
1. Bridge
11. Intra Temporal Distance(I.T.D.)
iii. Temple or side
18
Tangen~Screen (Bjerrum's· Screen)
A. Equipment
1. Tangent screen
2. 3 mm white object attached to black, non-glossy stick (wand)
3. Eye patch
4. Bright illumination
B. Set-up
1. Patient wears distance Rx if it is over -4.00 or+1.00 (spherical equivalent)
2. Patch on OS
3. Patient seated 1 meter from screen, eyes level with fixation target.
4. Examiner stands to right of screen to begin
5. Brightly illuminate screen from behind patient
Note
19
c. Move object all the way around the blind spot instructing the patient to report
if the object disappears.
D. Recording
1. Mark locations of your black pins by putting small x' s on a standard VF recording
diagram and connect with straight lines.
2. Cross-hatch areas of non-seeing, incl. blind spot.
10 mm object ~
120
Absolute~
15
195 345
210
315
300
255 270 285
Name : Dinesh
Test Object : 3 mm white
Remarks : Good fixation
Field done with +2. ODS O.U.
20
Amsler Grid
A. Equipment
B. Set up
C. Step-by-step procedure
D. Recording
1. If there are no problems, record 'Amsler' the eye tested followed by 'WNL'
(within normal limits).
2. If there is a problem record the eye, the nature of the problem, and its location on
the grid.
3. It is understood that plate # 1 was used unless otherwise noted.
4. If there are problems, you may also draw what the patient sees on an Amsler
black-on-white recording chart (see below).
21
E. Examples
Note
The Amsler Grid book contains a detailed account of the theory of the test and lists the
questions you should ask. It also explains when and how to use each of the 6 plates
(included in the book).
This procedure sheet is a quick reference to refresh your memory about the Amsler Grid
test. It is not a substitute for Amsler' s book, which should be read in its entirety before
performing an Amsler Grid test on a patient.
22
Lensometry
A. Equipment
1. Lensometer
B. Set up
I. Lensometer ON
2. Glasses held in machine and level, ocular surface away from you
3. Center of mires in center of eyepiece reticle
4. Eyepiece focused for you
5. Always do OD lens first
C. Step-by-step procedure
III. Prism
1. Locate the center of the mires at the center of the eyepiece reticle.
2. Dot this location on the lens (this is the optical center of the lens).
3. Subsequently dot the location of the pt' s line of sight on the lens.
4. Prism is computed by the relationship
P in prism diopters = Dell 0.
23
Where D is the power of the lens in diopters and c is the liner distance
between the patient's line of sight and the optical center of the lens in mm. It
is customary to calculate vertical, and horizontal prism separately.
5. If prism is 'ground' into the lens, it may be impossible to center the mires in
the reticle. In this case, dot the lens with the patient's line of sight & centre
this part of the lens in the reticle.-
6. Measure the amount of prism using the prism scale of the reticle
Note: With high amounts of prism, it is difficult or impossible to locate the
center of the mires. In this case, hand-held prisms must be inserted along with
the lens until the mires can be located. The value of these prisms is added to
the final result.
7. Record'=' anct amount and direction of prism in the glasses.
Note: Prism induced by spectacles is the Net Discrepancy between the prism
induced in the two eyes, not the prism induced by any one lens.
D. Recording
1. Record the Rx for each lens in stand Rx form
E. Examples
I. OD - 2.75DS
OS-2.25DS / -1.0DC x 10
24
Keratometry
A. Equipment
1. Keratometer or ophthalmometer
B. Set-up
1. Patient removes Rx
2. Focus the eyepiece
a. Set the adjustable eyepiece so that it extends as far out as possible
b. Place a white paper in front of the instrument objective to retroilluminate the
reticle
c. Adjust the eyepiece until the reticle is first seen in sharp focus
3. Adjust height of chair and instrument to a comfortable position of both patient
and examiner
4. Unlock instrument controls (B & L keratometer)
5. Instruct patient to place his chin in the chin rest and forehead against the forehead
rest
6. Raise or lower the chin rest until the patient's outer canthus is aligned with the
hash mark on upright support or the pointer on the side of the instrument.
1. From outside the instrument, roughly align the telescope with the patient's eye
2. Instruct the patient to look at the image of his eye in the keratometer
3. Look into the keratometer and align it by moving it P-om side to side or up and
down until you can see the image of the mires(# circulus) on the patient's cornea
(Fig. I)
4. Focus the mires and align them with the reticle in the lower right and hand circle
(Fig II)
5. Lock the instrument (B & L keratometer only)
6. Adjust horizontal and vertical power wheels until mires are in close apposition.
7. Rotate the telescope to align the spurs on the mires to 2 major meridians of the
patient's cornea (Fig III)
8. Adjust the horizontal power wheel until the horizontal mires are co-incident
9. Adjust the vertical power wheel until the vertical mires are co-incident
I 0. Readjust focus and/or re-center reticle as needed.
D. Recording
25
5. Record the type of astigmatism:
WR - with the rule
AR - against the rule
OBL - oblique
6. Record the conditions of the mires e.g. MCAR- mires clear and regular
MIAD - mires irregular and distorted
E. Examples
Keratometry
0
X
+
II. step C 4 IV. Steps C 8-9
0
()+G)
26
Static Retinoscopy with Trial Case
A. Equipment
B. Set-up
1. Patient wears trial frame. It should sit straight and comfortably on the pt's
face
2. Examiner is at same level as the patient, checks pt's OD with his/her OD and
the pt's OS with his/her OS.
3. Examiner holds retinoscope 1 meter or 67 cm from pt's eye. This is the
retinoscopy working distance and is converted to diopters
C. Step-by-step procedure
27
D. Recording
1. Record net static for each eye and the patient's VA through the net static.
Example: Net static
OD - 1.5 sph 6/9
OS - 1.75 sph 6/9
28
Subjective Refraction with Trial Case
A. Refining the Sphere
Jackson crossed cylinder (JCC). Only do this test if the patient does not see 6/6 with the
spherical correction or if patient has symptoms of astigmatism or if there is a large
amount of cylinder (> 1.5 D) and a change of axis. Remember that with hand held JCC,
the white slash marks represent -.50 cylinder and the red slash marks represent +.50
cylinder.
2. Axis Check
a. Place the JCC in front of the trial frame with the red and white marks each at 45°
to the trial lens cylinder axis, so that the handle of the JCC lies along the same
line as the axis of the trial lens cylinder.
b. Instruct the patient 'I will show you 2 views of the line of letters, both may be
blurry. You tell me which view is sharper or less blurry'.
c. The JCC is now twirled so that the white marks (-) and the red marks (+) change
positions.
d. The patient chooses a position, the axis of the cylinder will now be turned 10°
towards the white marks. Plus cylinder 10° to the red marks.
e. Repeat steps a-d. If patient reverses direction, then tum back the axis only 5°, or
select the mean.
f. End point: Both views equally blurry or same.
3. Power check
a. Place the JCC so that the red marks(+) lie along the trial lens cylinder axis.
b. Instruct the patient as above.
c. Twirl the JCC over so now the white marks (-) lie over the trial lens axis.
29
d. The patient chooses which side is clearer. If patient prefers red marks(+) add+
.50 to the cylinder; if patient prefers white marks (-) ass -.50 to the cylinder or
+0.25 DS with-0.50 cylinder. The spherical equivalent /JCC rule).
e. End point: Both views equally blurred or the same.
f. Finish by adding maximum plus spherical lenses for best VA, or do + 1.00 DS
blur check test.
Note: JCC can be confusing to some patients. Do not do this technique if the
patient does not understand/ cooperate.
At the end of the distance refraction you need to check if the patient's accommodation is
relaxed.
1. Tell patient that you will now put a lens in the trail frame which will make his vision
worse. Place a+ 1.00 DS in the frame and ask him to read down the chart.
2. With a pupil size of 4 mm (approx) they will only be able to read down to 6/18 if
their accommodation was relaxed.
- If he can read 6/12 or better this usually indicated that the accommodation
was still active.
- If he can read 6/18 or worse their accommodation must be relaxed so there
is no read to confirm the results with a cycloplegic refraction.
Note
If the pupils are small, 2 mm or less, this test is unreliable. This may happen with elderly
patient's and with patient's using pilocarpine drops for glaucoma. In both these groups
the accommodation is rarely active.
For patients with a best VA of 6/12 or worse, this technique is unreliable. Record the
results in brackets.
30
Binocular Balance with Trial Case
Perform this technique only if the two eyes achieve the same VA separately under
monocular conditions. The purpose of binocular balance is to match and relax
accommodation in the 2 eyes. If you occlude one of the patient's eyes, both eyes will
accommodate up to 1.00 or 2.00 dioptres. By using a technique which encourages both
eyes to relax their accommodation, the full refractive error can be determined.
1. After refracting OD and OS separately, place +0.S0 DS over both eyes. {add
+.50 to the subjective refraction OU).
2. Instruct patient to look at line of letters 3 lines above best VA. If patient is not
6/9 or worse, add more plus to fog patient.
3. Make sure both eyes are able to see target.
4. Alternately occlude OD and then OS asking patient to compare the two eyes
and report which eye is less blurry.
5. Add +0.25 DS to the better eye.
6. Repeat steps 4 + 5 until both eyes equally blurry or same check VA OU
7. Begin to remove plus, +0.25 DS at a time, to both eyes together. Reduce plus
until patient sees 6/6 OU.
Note: It may be difficult to achieve a perfect match between the two eyes. In
this case, leave the slighting dominant eye with the slightly sharper vision.
31
Static Retinoscopy with Phoropter
A. Equipment
I. Retinoscope (streak)
2. Phoropter
3. Fixation target: 6/60 with red/green filter
B. Set-up
C. Step-by-step procedure
D. Recording
1. Record net static for each eye and the patient's VA through the net static
Example: QD + 2.75 sph 6/18
OS+ 1.5-0.50 * 175 6/6
Note
Ask the patient to inform you if your head blocks his view of the fixation target. It may
be necessary to rotate the phoropter slightly to allow patient to see the target and to allow
you to maintain alignment along the visual axis.
32
Subjective Refraction with Phoropter
A. Refining the sphere
6. Think:
a. Do the numbers follow Egger' s chart?
b. Remember MPMVA (Maximum plus for Maximum VA).
c. Patient may not see 6/5 or sharp vision with this procedure since you have
only worked with the sphere, and not yet refined the cyl. You may have to
over-minus the patient (0.25DS) before doing JCC to assure that the circle of
least confusion will be on the retina.
Only do this test if the patient does not see 6/6 with the spherical correction or if the
patient has symptoms of astigmatism or there is a large amount of cylinder (> 1.5 D) and
a change of axis. Remember that in the phoropter, the red marks represent -0.25 cylinder
and the white marks represent +0.25 cylinder. Instruct patient to look at 6/12 line; or one
line above best VA.
I. Axis Check
a. Place JCC before the eye so that the red marks (-0.25 cylinder) and the white marks
(+0.25 cylinder) are at 45° to the axis of the cylinder in the phoropter. It wti.1 click
into place.
b. Instruct the patient: I will show you 2 views of the line ofletters. Both may be blurry.
Tell me which view is sharper or less blurry.
33
c. The JCC is now twirled so that the red marks (-) and white marks (+) change
positions.
d. Move the axis toward the RED marks of the preferred view, about 10°. Report steps
a-c until both views look same or there is a reversal of direction. In this case move
the axis in 5° steps.
e. End point:
- Both views same
- Patient moves back and forth within a narrow range. Select the mean
Remember that the JCC in the phoropter is opposite to the hand held JCC; the red marks
represent-0.25 cylinder and the white marks represent +0.25 cylinder.
34
Binocular Balance with Phoropter
Perform this technique only if the two eyes achieve the same VA separately under
monocular conditions. The purpose of binocular balance is to match and relax
accommodation in the 2 eyes. If you occlude one of the patient's eyes, both eyes will
accommodate up to 1.00 or 2.00 dioptres. By using a technique which encourages both
eyes to relax their accommodation, the full refractive error can be determined.
Methods:
1. Alternate occlusion
2. Prism dissociation
1. Alternate occlusion
2. Prism dissociation
Note
It may be difficult to achieve a perfect match between the two eyes. In this case, leave
the dominant eye slightly sharper vision.
35
Measuring Phorias and Vergences with Phoropter
A. Equipment
I. Phoropter
2. Projector with VA slide/screen
3. Near-point rod
4. Near point card
5. Lamp
B. Set-up
C. The above apply to horizontal and vertical Von Graefe phorias and to horizontal
and vertical vergences
Note
36
4. Instruct patient: "Look at the lower target. But think about the other or upper one.
I will make the upper target move. You tell me when the two images for targets
are vertically lined up like the buttons on a shirt, one directly above the other.
You should always see double, keep the images clear, and continue to look at the
lower target."
5. At about 2 per second, reduce the BI-OD prism until the patient reports vertical
alignment.
6. Continue changing prism in the same direction. i.e. overshoot endpoint until
patient sees two targets, one up and to the left, one down and to the right.
7. Then bringing prism back in the other direction until patient again reports
alignment.
8. Result = average of values in steps 5 and 7, if they are within 3° of each other.
Otherwise take another measurement, emphasizing the instructions to the patient,
and average the 2 closer values.
Note: the endpoint is often variable as the images often move around. Use the
first at which patient reports alignment.
Recording
Recording size (in prism diopters) and type of deviation observed.
Example
DLP: 2 BI or 2 exo
DLP: 4 BO or 4 eso
37
Recording
Record size (in prism diopters) and type of deviation observed. For vertical phorias, you
must also record the eye (as per these instructions) before which at the vertical prism was
placed.
Example
DVP: 2 BD on OS or 2 L hyper
DVP: 1 BU on OS or 1 R hyper (1 L hypo)
Recording
Example
NLP: 6 exo or 6 BI
NLP: 6 BO or 6 eso
Recording
Record size (in prism diopters) and type of deviation observed. For vertical phorias, you
must also record the eye as per these instructions) before which the vertical prism was
placed.
Example
NVP: 2 BD on OS or 2 L hyper
NVP: 3 BU on OS or 3 R hyper (3 L hypo)
Record the two sets of three values (blur, break, recovery for BI and for BO) in
prism diopters. If no blur point is ·observed, a dash (-) is entered.
Note: If values are in the direction opposite to what your expect, record value as
minus diopters. e.g.: Ifrecovery on BO vergences occurs only with the use of BI
prism before the patient record minus the amount of BI prism. If patient reports
that the target is moving to one side, the patient is suppressing one eye.
Determine which eye is suppressing, and record appropriately.
Example
Recording
Record the type of vergence and two sets of values (Break and Recovery for BU
and for BD) and prism diopters
Note: If values are in the direction opposite to what you expect, record value as
minus diopters. If patient reports that the target moving, he is suppressing one
eye. Determine which eye is suppressing and record appropriately.
Example
1. L - infra 4/2, supra 2/-1
2. L - infra 6/4, Supra 3/1
40
J. Vergences (Phoropter) Near Horizontal
Recording
Record the two sets of three values (Blur, Break, Recovery for BI and for BO) in prism
diopters. If no blur point is observed, a dash (-) is entered.
Note
If values are in the direction opposite to what you expect, record value as minus diopters.
e.g.: If recovery on BO vergence occurs only with the use of BI prism before the patient
record minus the amount of BI prism. If patient reports that the target is moving to one
side, the patient is suppressing one eye. Determine which eye is suppressing and record
appropriately.
Examples
1. @N BI: 14/18/10 BO: 18/24/16
2. @N BI: 12/20/14 BO: 4/10/-4
Recording
Record the type of vergence and the two sets of values (break and recovery for BU and
BD) in prism diopters.
Note
If values are in the direction opposite to what you expect, record value as minus diopters.
If patient reports that the target is moving, he is suppressing one eye. Determine which
eye is suppressing and record appropriately.
Example
42
Tentative add as a function of Age and Refractive Status
A. Equipment
1. Phoropter
2. Near point card
3. Near point rod
B. Setup
C. Procedure
D. Recording
1. Record the amount of plus added for the NRA and the amount of minus added
for the PRA
E. Example
NRA/PRA: +2.25/-2.50
43
Maddox Rod Phoria
A. Equipment
B. Set up
C. Step-by-step procedure
1. Tum light-source on
2. Have patient look at the light
3. Refine position of the Maddox rod so that the streak is perfectly vertical (for
OLP and NLP) or horizontal (for DVP and NVP)
4. Have patient tell you where the line is located relative to the light.
D. Recording
E. Example
1. DLP c red MR ortho NLP c red MR 6 exo
2. DVP c red MR iso NVP c red MR O iso
44
Binocular and Accommodative Expecteds
PRA -2.25
45