TORKEL900-Series UG en
TORKEL900-Series UG en
User guide
User guide
Contents
1 Introduction Preparations...................................................... 25
.............................................................. 6 Select "Test Method"........................................ 25
1.1 Product description.............................................. 6 Starting the test................................................. 25
1.2 Features and benefits........................................... 6 Pausing the test................................................. 25
Ending the test.................................................. 25
1.3 Receiving instructions........................................... 7
5.3 Test at constant power....................................... 26
1.4 Warranty............................................................. 7
Preparations...................................................... 26
Warranty repair.................................................... 7
Configuring TORKEL for constant power............ 26
2 Safety Select "Test Method"........................................ 26
.............................................................. 8 Starting the test................................................. 26
2.1 Symbols on the instrument.................................. 8 Pausing the test................................................. 26
Information duty regarding substances on Ending the test.................................................. 27
REACH article 33, SVHC-list ................................ 8
5.4 Test at constant resistance.................................. 27
2.2 Safety instructions................................................ 8
Preparations...................................................... 27
Protection system.............................................. 10
Select "Test Method"........................................ 27
3 Instrument description and Starting the test................................................. 27
Accessories........................................ 12 Pausing the test................................................. 27
3.1 Top panel........................................................... 12 Ending the test.................................................. 27
3.2 Side panel.......................................................... 13 5.5 Testing with a load profile.................................. 28
3.3 Included accessories........................................... 13 Preparations...................................................... 28
TORKEL 910...................................................... 13 Select "Test Method"........................................ 28
TORKEL 930/950............................................... 13 Starting the test................................................. 29
TORKEL Viewer.................................................. 13 Pausing the test................................................. 29
Ground cable..................................................... 14 Ending the test.................................................. 29
3.4 Optional accessories........................................... 14 5.6 V Logger............................................................ 29
TXL extra loads.................................................. 14 Starting "V Logger" directly after a discharge
BVM - Battery Voltage Monitoring..................... 14 test.................................................................... 29
Clamp-on-probe................................................ 14 Starting "V Logger" as a separate test............... 29
Extension cables................................................ 15 Ending the test.................................................. 29
5.7 Viewing results and reporting............................ 30
4 Menu system
Viewing the results............................................ 30
............................................................ 16
REPORT............................................................. 30
4.1 Main menu........................................................ 16
Download tests to PC with Ethernet cable......... 30
Screen buttons.................................................. 16
TORKEL FTP browser.......................................... 31
TEST menu........................................................ 16
Test configuration.............................................. 17 6 External current measurement........... 32
GRAPH.............................................................. 20 6.1 General.............................................................. 32
RESULT.............................................................. 20 6.2 Setting up external current measurement........... 33
CELLS (optional BVM)........................................ 20 DC clamp-on current probe............................... 33
4.2 TORKEL Viewer.................................................. 22 Troubleshooting................................................. 33
5 Making a test 6.3 TORKEL used together with TXL Extra Load........ 34
............................................................ 24 Multiple units used together.............................. 34
5.1 Preparations for testing...................................... 24 Constant current, regular load connected.......... 35
Connecting the current cables to the battery..... 24
5.2 Test at constant current...................................... 25
7 Alarm function
............................................................ 36
7.1 Description........................................................ 36
Resetting the alarm............................................ 36
9 Optional equipment
............................................................ 40
9.1 TXL830/850/865/870/890 (Extra Loads) .............. 40
Panel................................................................. 40
Using the TXL.................................................... 41
Testing............................................................... 41
9.2 BVM - Battery Voltage Monitor.......................... 42
Power & signal connector unit........................... 42
Test with BVM – without TORKEL....................... 45
Optional BVM equipment.................................. 45
10 Troubleshooting
............................................................ 46
11 Calibration
............................................................ 48
11.1 General............................................................ 48
11.2 How to calibrate.............................................. 48
12 Specifications
............................................................ 50
13 Appendix A
............................................................ 54
Accessories for connection of BVM modules to
battery poles..................................................... 54
13.1 Extension lead.................................................. 54
Cable with ring connector for battery poles....... 54
Cable with flat connector for battery poles........ 55
Battery bolt for BVM module connection........... 55
Cable Extension lead with Velcro band............... 56
Index........................................................ 59
1 Introduction
Warranty repair
Equipment returned to the factory for repair must be
shipped prepaid and insured.
Contact your Megger representative for instructions
and a return authorization (RA) number.
Please indicate all pertinent information, including
problem symptoms.
Also specify the serial number and the catalog num-
ber of the unit.
2 Safety
WARNING
Hot, do not cover
1. The electrical voltage and current used in battery
testing is potentially lethal.
Protective conductor terminal. Ensure that the AC supply is disconnected and any
battery under test is disconnected before attempt-
ing any cleaning or maintenance of TORKEL.
2. Connection and disconnection procedures are
WEEE, Waste Electronic Equipment. extremely important. Be sure to follow the
Please utilize your local WEEE collec- instructions carefully..
tion facilities in the disposition of this 3. Do not touch conducting parts of the clamps on
product and otherwise observe all the current cables or the voltage sensing cables
applicable requirements. when they are connected to TORKEL.
The unit can also be returned to Meg- 4. Always connect protective earth (ground).
ger at any time at no charge for the 5. Misuse of TORKEL can be extremely dangerous.
disposal.
6. Inspect cable connections to make sure there is
no short circuit.
Information duty regarding 7. Use an easily accessible power outlet. This will
substances on REACH article 33, ensure that you can disconnect the power
SVHC-list quickly in case of a problem.
This product contains a coin cell battery which con- 8. When a lead acid battery is charged or dis-
tains 1,2- dimethoxyethane (CAS 110-71-4) above charged i.e. when there is a current flow
0.1% by weight. through the battery it is always a risk that the
battery can explode.
If there is a bad connection inside the battery
and there is a current flow - the connection will
burn off and there will be an arc, which will
ignite the oxyhydrogen gas in the battery.
For new open (vented) batteries the risk is me-
dium to low but in old VRLA (sealed) batteries
the risk is medium to high.
9. To minimize the risk for personnel injuries:
Always place TORKEL/TXL as far away from the
battery as possible. Never stand close to a bat-
tery during charge/discharge.
10. Too high discharge current applied on a battery 19. TORKEL connected to grounded battery systems
can cause the battery to explode or get over-
TORKEL connected to positive grounded battery
heated. Be sure to not set too high current.
system.
11. If the external current measurement is inter-
rupted or giving false values during the test, the If connected as the picture below. TORKEL’s max
current will rise to a higher level than the set battery input voltage (voltage between battery
value before the test is shut down. If the battery input poles) is correct. But the voltage between
is too small for this current or in a bad condition TORKEL’s negative battery input and ground is
- it may explode. exceeding the allowed voltage between battery
12. Never use the TORKEL/TXL Extra Load in an explo-
input and ground (500 V).
sive environment. Never put the TORKEL/TXL Extra The below hook up must not be used.
Load where it can be reached by battery gas.
13. Improperly connected cables carrying high cur-
rent can cause fire. Make sure that the cables are
not twisted in such a way that could cause them
to turn and come loose from the connector.
14. Make sure that there is no dust or dirt in or
around the cooling fans before starting a
discharge test. After starting a discharge test,
check that the fans are running properly and
that the airflow is good. Make sure no hair or
clothing is sucked into the fans.
15. Position TORKEL/TXL Extra Load where air flow
is unobstructed and where it does not come into TORKEL connected to negative grounded bat-
contact with any flammable or heat-sensitive tery system.
material. Keep a free distance of 1.5 m (5 ft)
If connected as the picture below. TORKEL’s bat-
to the vertical sides of TORKEL/TXL and 2.0 m
tery input voltage is correct (Voltage between
(6.5 ft) above TORKEL/TXL.
battery input poles). But the voltage between
16. Do not place TORKEL a) near another TORKEL,
TORKEL’s positive battery input and ground is
a) TXL Extra Load or any other heat source or b) exceeding max voltage between battery input
where the cooling airflow can be blocked. TOR- and ground (500 V).
KEL will overheat if there is insufficient cooling.
17.
The below hook up must not be used.
External current shunt may not be used above
300 V DC
18. Do not use any other equipment other than
what is provided or specified.
Important
Protection system
The unit has built-in protection system against over-
heating and malfunctioning cooling fans. A built-in
circuit breaker is an important component in the
safety system. If the temperature will be too high and/
or the cooling fans malfunction, the circuit breaker
will trip. There is also a melting fuse connected in
series with the circuit breaker.
Always follow the safety instructions in section 2.2.
1 2 6 7
5 8
3 4 9
12
10 11 13
14.
Protective conductor terminal TORKEL 930/950
15. MAINS
Connector for mains supply. Cable set
Cable set, 2 x 3 m, 70 mm2 (10 ft), for connecting
16. + TORKEL to the battery.
Connection terminal (+) for the battery (or other
DC source)
17. VOLTAGE SENSE
Input for sensing voltage at the battery terminals.
Impedance to the battery current terminals is
>1 MΩ
18. –
Connection terminal (-) for the battery (or other
DC source).
TORKEL Viewer
(GC-30060)
Sensing leads
For measuring voltage at battery terminals. 2 x 5 m
(16.4 ft).
Clamp-on-probe
Clamp-on probe, 1000 A DC. To measure current in
external circuit (in combination with the TXL instru-
ment).
Clamp-on-probe (XA-12991)
(GA-00545)
Extension cables
Extension for GA-00550, 2 x 3 m, 25 mm2, male / female.
4 Menu system
Screen buttons
Confirm
Run
Stop
Test method
Temperature compensation
A test can be made in six test methods, "Constant I"
Shown when compensation is
is default.
enabled
1] Press the button and select test method.
Quick guide
▪ Constant I (current)
Open test file library ▪ Constant P (power)
▪ Constant R (resistance)
New test
▪ Profile I (current profile)
Test configuration ▪ Profile P (power profile)
Information bar ▪ V Logger (Current with monitoring during charge)
Shows messages and is used for A] For Constant I, P or R you press the numeric
acknowledge alarms field and make the settings for the parameter
Information field using the On-screen keyboard or the control
Shows voltage and current knob.
*) The pause time has limitations regulated by inter- For Profile I or Profile P a new tab will be shown
national standards. If the pause time is too long the on top of the screen together with a window for the
battery may have to be recharged before a new test is settings.
done.
Test configuration
1] Press
ID Labels
Stop limit
When a "Stop" criteria is reached, the "information
bar" turns red and the cause of stop is displayed.
Settings
Temperature compensation
RESULT menu with temp compensation enabled.
Calibration
See section "11.2 How to calibrate" on page 48.
Test Manager
▪ Manage language files for the internal software and select
U or V as the symbol for voltage. The database can store up to 30 tests.
1] Press
About
When an USB stick with a license file is inserted in
one of the USB ports, the “License file” button will be
highlighted in blue.
How to upgrade
1] Press the “License file” button and select the
license file from the menu that appears.
The upgrade will be performed directly. 2] Select a test by marking the checkbox.
3] You can use the buttons at the bottom for
System actions described below:
to open a test
GRAPH
RESULT
After ending a test you can go into the different tabs
to se the result.
Long interconnections
Diagram shows voltage per cell during the test. Numbering Connection change if there is a voltage drop in the intercon-
of cells depends on cell direction (cell 1 "+" or "-"). Informa- nection.
tion on cell numbering and direction are displayed on the
screen. Disconnected cell
▪ The LIVE voltage means the present voltage
▪ The FLOAT voltage means the voltage read prior to
removing the battery charger.
▪ The START voltage is when the discharge tests start.
1] By pressing a cell bar, readings can be seen in
the table as the "Sel" value.
The cell direction can be selected either from
plus to minus or opposite (cell 1 "+" or "-")
4.2 TORKEL Viewer ▪ The “Plot” tab gives you the possibility to check
the battery measurements or the BVM voltages
TORKEL Viewer is delivered together with TORKEL diagram.
(not with TORKEL 910) on a USB stick. By right click and drag on the y-scale of resp. entity
the diagram lines can be modified.
TORKEL Viewer runs under MS windows 7,8 and 10. The time scale can also be adjusted by right click
1] Double click the: and drag the scale.
“TORKEL Viewer Setup.msi” file. By placing the cursor in the diagram and scrolling
with mouse, zooming of the time scale will be done.
2] Follow instructions and the TORKEL Viewer Clicking the “Auto zoom” will reset the diagram to
will be installed on your computer. the default appearance.
Clicking the “Export Plot” button will export the
Note After installation you can find a demo file at diagram in a .png format.
C:\Program Files (x86)\Megger\Torkel Viewer\ The BVM measurements can also be examined by
Reporting. selecting the BVM voltage diagram in the “Choose
plot to display”.
3] Click to open the program.
The diagram can be moved by right click and drag.
Each voltage cell can be checked by clicking on the
cell bar. Float, start and end values are presented.
▪ The “Preview” tab is simply a way of checking the
report after modification, prior saving or print out of
the report.
▪ The "TorkelCalc" tab is used for calculating how
many TORKELs and TXLs are needed.
▪ The "Settings" tab contains, among other things,
language setting.
▪ The cell bar diagrams shows numbering of cells and
whether it has + or - cell direction.
5 Making a test
Connecting the current cables to The sensing cables (dotted lines) are normally not needed.
the battery
Note TORKEL automatically selects the voltage
Important range when voltage is applied to the high-
Connection and disconnection procedures current terminals.
are extremely important. Be sure to fol-
low the instructions. If load polarity is incorrect the load will not
be connected internally. On the display the
”space bar” will be red and an alarm will
Use the cables supplied with TORKEL or other cables sound. This is true for both the battery and
of suitable size. Follow the numbered steps that are voltage sense terminals.
set forth below. Inspect each connection to make sure
it is securely fitted. You can start a test in any of the menus.
While test is running you can go into any tab
1] Connect one end of the first cable to the
and pause the test and make changes and
negative (-) terminal on TORKEL.
continue the test.
2] Connect the other end of the first cable to
the negative (-) pole of the battery.
3] Connect one end of the second cable to the
positive (+) pole of the battery.
4] Connect the other end of the second cable to
the positive (+) terminal on TORKEL.
Tip
You can change the settings for current
and limits at any time in "TEST".
1] Press
Tip
You can change the settings for power
and limits at any time in "TEST".
Preparations
Follow the safety precautions set forth in section "2.2
Safety instructions" on page 8 and the prepara-
tions for testing in section "5.1 Preparations for test-
ing" on page 24.
1] Press
WARNING
Do not connect a discharged battery to a Preparations
battery that has not been discharged. The Follow the safety precautions set forth in section "2.2
batteries must be charged to the same Safety instructions" on page 8 and the prepara-
potential (voltage) before they are con- tions for testing in section "5.1 Preparations for test-
nected together. ing" on page 24.
Tip
You can change the settings for current/
power and limits at any time in "TEST".
1] Press
Starting "V Logger" as a separate
test
Preparations
Follow the safety precautions set forth in section "2.2
Safety instructions" on page 8 and the prepara-
tions for testing in section "5.1 Preparations for test-
ing" on page 24.
2] Press "Yes".
3] Disconnect the cables in reverse order as Select "Test Method"
described in section "Connecting the current 1] Press "V Logger" button.
cables to the battery" on page 24
2] Make settings for warning and stop time, see
section "Warning and stop limit parameters"
WARNING
on page 17.
Do not connect a discharged battery to a
battery that has not been discharged. The 3] Press
batteries must be charged to the same 4] Start charging the batteries.
potential (voltage) before they are con-
nected together.
Ending the test
Tip WARNING
You can select to make the charging test Do not disconnect any cables until the test
immediately after a discharging test and is completed and stopped.
keep all connection as they are.
For viewing results and report see section "5.7 View- 1] Press
ing results and reporting" on page 30.
REPORT
The saved tests can be stored on an USB memory
stick and transferred to a PC or transferred using FTP
communication. The included PC software TORKEL
Viewer can be used to edit and print out reports. See
paragraph "4.2 TORKEL Viewer" on page 30.
Edited reports can be transferred back to TORKEL
i.e. you can design the test and settings in TORKEL
Viewer.
Method 2
Connecting Torkel to PC with an Ethernet
cable(without data switch).
1] Connect the Ethernet cable to Torkel(right
“service” connector)
2] Open up the file manager on TORKEL.
3] Press the button. Transfer menu
opens up with IP address.
4] Setup of IP address on PC, Win 10:
5] From the Control Panel
Network and Internet
Network and sharing centre
6] Click on network “Ethernet”
Properties
20] To save a test file to PC, mark the test file
7] Double click on “Internet Protocol Version and press the “Download” button.
4(TCP/IPv4)”
21] A test can also be transferred to TORKEL by
8] Mark the “Use following IP address:” selecting the “Upload” button.
9] Type in the IP address seen on TORKEL and 22] Tests can be deleted in TORKEL by selecting a
change the last digit to another number. test in the list and click “Delete”.
10] Click on the “Subnet mask:” Values auto-
matically filled.
11] Press "OK" and "OK" to verify change of
settings.
12] Start TORKEL Viewer and select “Open FTP
connection”.
13] See section “TORKEL FTP browser” below.
6 External current
measurement
6.1 General Tip
For tests where it is important to obtain
External current measurement must be used when: the desired current within a few seconds
▪ TORKEL is working together with TXL Loads. or less it is better to use internal current
measurement since it provides faster
▪ Testing without disconnecting the regular load.
regulation.
Since total current is measured, TORKEL can compensate
for changes attributable to the regular load. The total
current from the battery is then kept at a constant value.
This ensures accurate test results.
The external current measurement function enables
TORKEL to measure the total current in an external
path and base regulation on this measurement.
A DC clamp-on current probe (optional accessory) has
to be used for this measurement. It can be applied at
one of the battery terminals or at an inter-cell con-
nector. The DC clamp-on current probe must measure
the total current, including that which passes through
TORKEL.
A current shunt can also be used, but this requires
opening the current path and connecting the shunt
in series. The current shunt must be connected to the
negative side of the battery.
WARNING
External current shunt may not be used
above 300 V DC
Important
Always activate the warning and stop
limit functions when using external cur-
rent measurement. This will protect your
batteries if the DC clamp-on current probe
will malfunction.
Important
If a current shunt is used it must be con-
nected on the negative side of the bat-
tery.
TORKEL 2 Measures all current except the current Now start the TORKEL with the second
through TORKEL 1. highest number, then the third highest, etc.
TORKEL 3 Measures all current except the currents Finally, start TORKEL No. 1.
through TORKEL 2 and TORKEL 1 (and so Starting the TORKELs in this order prevents
forth). the current from being higher than desired
The last Measures only the current through itself at the beginning of the test.
TORKEL and the TXLs.
Constant current, regular load
connected
7 Alarm function
7.1 Description
The TORKEL alarm function is provided by a buzzer
and a relay connected to the <ALARM> - connector.
An external alarm device can be connected to this
connector if desired.
Devices higher than Cat II must not be attached to the
alarm connector.
Relay contact
8 A / 28 V DC
0.28 A / 250 V DC (resistive load only)
8 A / 240 V AC
Connector insulation
Voltage to ground must not exceed 250 V.
2. Current flowing through TXL(s) at final Systems TORKEL 910 and TXL 870 (Pos 1)
voltage 220 V battery (108 cells)
▪ Multiply the total current through the TXL(s) which you Max constant current No. of TORKEL 910 No. of TXL 870
obtained in step 1 above by the final voltage, and then 112 A 1 1
divide by the voltage at the beginning of the test. 156 A 1 2
195 A 1 3
3. Number of TORKELs – for the current 235 A 1 4
regulation 313 A 2 4
The TORKEL or TORKELs in the system must regulate
the current to the desired value and compensate for TORKEL 930
the drop in current through the TXL(s) that occurs at
Systems TORKEL 930 and TXL 830 (Pos 3)
the final voltage.
12V battery (6 cells)
▪ The amount of regulation needed can be obtained by Max constant current No. of TORKEL 930 No. of TXL 830
subtracting the current value (A) obtained in step 2 above 183 A 1 1
from the desired current. 299 A 1 2
414 A 1 3
▪ Calculate the number of TORKELs required for the current
regulation. 24V battery (12 cells)
Max constant current No. of TORKEL 930 No. of TXL 830
4. Are all of the TXLs needed? 384 A 1 1
If the total load-providing capability of the TORKEL(s) 615 A 1 2
846 A 1 3
exceeds the amount of regulation needed by a wide
margin (as set fort in step 3 above), you can perhaps
Systems TORKEL 930 and TXL 850 (Pos 3)
conduct the test with fewer TXLs. If this margin is
wider than the current through one of the TXLs at the 48 V battery (24 cells)
Max constant current No. of TORKEL 930 No. of TXL 850
final voltage, this TXL is not needed.
452 A 1 1
685 A 1 2
Note Discharge is from 2.15 V to 1.8 V per cell
918 A 1 3
1151 A 1 4
TORKEL 910
Systems TORKEL 930 and TXL 870 (Pos 3)
Systems TORKEL 910 and TXL 830 (Pos 2) 110 V battery (54 cells)
12 V battery (6 cells) Max constant current No. of TORKEL 930 No. of TXL 870
Max constant current No. of TORKEL 910 No. of TXL 830 225 A 1 1
144 A 1 1 314 A 1 2
221 A 1 2 392 A 1 3
298 A 1 3 470 A 1 4
376 A 1 4 628 A 2 4
443 A 2 4
24 V battery (12 cells) Systems TORKEL 930 and TXL 870 (Pos 1)
Max constant current(A) No. of TORKEL 910 No. of TXL 830
220 V battery (108 cells)
264 A 1 1
Max constant current No. of TORKEL 930 No. of TXL 870
419 A 1 2
112 A 1 1
574 A 1 3
156 A 1 2
195 A 1 3
Systems TORKEL 910 and TXL 850 (Pos 3) 235 A 1 4
48 V battery (24 cells) 313 A 2 4
Max constant current No. of TORKEL 910 No. of TXL 850
342 A 1 1
575 A 1 2 TORKEL 950
Systems TORKEL 910 and TXL 870 (Pos 3) Systems TORKEL 950 and TXL 890 (Pos 1)
110 V battery (54 cells) 480 V battery (240 cells)
Max constant current No. of TORKEL 910 No. of TXL 870 Max constant current No. of TORKEL 950 No. of TXL 890
188 A 1 1 65 A 1 1
266 A 1 2 70 A 1 2
344 A 1 3 90 A 1 2
422 A 1 4 100 A 1 3
532 A 2 4 120 A 2 2
9 Optional equipment
WARNING
The circuit breaker F1 is an important com-
ponent in the safety system. If the tempera-
ture will be too high and/or the cooling fans
malfunction, the F1 will trip. It is important
to never operate the unit if there is any dam-
age or malfunction to the circuit breaker or
if any damage or malfunction is suspected.
4. Mains inlet
MAINS
Connector used for mains supply, equipped with
1 ON/OFF switch.
5. Connection terminals for the battery
+ (Terminal)
Positive (+) current connection for battery or
2
other DC source, being tested.
– (Terminal)
3 Negative (-) current connection for battery or
other DC source, being tested.
4
Insulation voltage to ground: 2200 V
5
Testing
WARNING
See the chapter "2.2 Safety instructions"
on page 8. for safety precautions.
1
The other parts of the BVM system: AC Adapter, Power and
Signal Connector and cables.
2
3 Power & signal connector unit
The BVM units require external 24 DC power and RS-
1. Control cable – Output socket 485 data communications for operation. These func-
2. Control cable – Input socket tions are provided by a combination of an external DC
3. Battery sense – Dolphin clip power supply and a Power & signal connector unit”.
The BVM is designed in modular form where one BVM The Power & signal connector conversion is performed
device is used for each battery or “jar” in the string within a single moulded plug that connects directly to
to be tested. It measures the voltage for each battery the laptop computer or other data acquisition device.
cell and each unit can handle up to 20 V. The BVMs The Power & signal connector has an RJ-45 connector
connects to the batteries in a “daisy-chain” fashion, that provides a connection to the first BVM unit in the
thereby providing easy and economical expandability chain, and this connection provides RS-485 data and
to meet the testing requirements for small-to-large power to all BVM units in the BVM string.
battery bank systems. Up to 240 BVMs can be daisy-
chained in a single battery bank under test.
The included dolphin-style battery clamp can be easily
removed and exchanged with different styles of stand-
ard banana plug clamps and/or extension cables to
accommodate any battery connection requirement.
There is an Activity LED on the BVM that flashes when-
ever an individual BVM transmits data back to the data
acquisition host computer. This LED activity is useful for
indicating that the device is functioning properly.
WARNING
See the chapter "Safety" for safety pre-
cautions.
Fig.9.2 When the battery bank exceeds 120 cells, this connection with 2 BVM loops shall be used.
Connect the first loop to the BVM1 connector on TORKEL and the second loop to the BVM2 connector on TORKEL.
Fig.9.3 The connection shown above must be used when the battery bank voltage is between 300 V and 500 V.
It includes two opto couplers and two ungrounded power supplies.
Extension cables
The BVM cables (USB) may be extended 2 x 5 m with
an Opto Coupler (e.g. B&B UH401 2 kV) in between.
Cable 5 m Cable 5 m
10 Troubleshooting
Communication lost with BVM Check/replace the Power & Signal con-
Cable/power fault verter.
units
11 Calibration
12 Specifications
VOLTAGE SENSE Impedance to the current terminals is >1 MΩ Voltage (DC) Current Power max.
Outputs max. max.
ALARM TXL830 28 V 300 A 8.3 kW
Relay contact 28 V DC, 8 A, 240 V AC, 8 A TXL850 56 V 300 A 16.4 kW
Devices higher than Cat II must not be at- TXL865 260 V (98 A max) 117 A 25.5 kW
tached
TXL870 280 V (56 A max) 112 A 15.8 kW
TXL STOP
TXL890 480 V (32 A max) 62 A 15.4 kW
Relay contact 250 V DC, 0.28 A, 28 V DC, 8 A, 250 V AC, 8 A
Internal resistance, 3-position selector
9 V DC 9 V DC, ±7% max 100 mA
Position 1 Position 2 Position 3
Communication ports
BVM1 BVM2 USB connection for BVM units TXL830 0.275 Ω 0.138 Ω 0.092 Ω
SPECIFICATIONS BVM
Specifications are valid at an ambient temperature of +25°C,
(77°F). Specifications are subject to change without notice.
Environment
Application field The instrument is intended for use in
medium-voltage substations and industrial
environments.
Altitude <2000 m (6500 ft) above sea level.
Temperature
Operating 0°C to +50°C (32°F to +122°F)
Storage & -20°C to +70°C (-4°F to +158°F)
transport
Humidity 5% – 95% RH, non-condensing
CE-marking
LVD 2014/35/EU
EMC 2014/30/EU
RoHS 2011/65/EU
General
Power supply
Mains voltage 100 – 240 V AC, 50 / 60 Hz
Power 50 VA (max)
consumption
Protection Over voltage, reverse voltage, voltage
transients, ESD
Dimensions
BVM unit 75 x 64 x 25 mm (3” x 2.5” x 1”)
Carrying case 575 x 470 x 205 mm (22.6” x 18.5” x 8.1”)
Weight
BVM unit 0.07 kg (0.15 lbs)
With accessories BVM system of 16 units 7 kg (15 lbs)
and carrying case BVM system of 31 units 8.8 kg (19 lbs)
BVM system of 61 units 12.5 kg (27 lbs)
Measurement section
Maximum number 240 (2x120)
of BVM
Voltage ranges 0-20 V DC
Resolution 1 mV both ranges
Inaccuracy < 0.1% of full scale ±0.01 V DC
Battery string 300 V DC (max)
voltage
Measurement 1 MΩ
input impedance
13 Appendix A
Index
A M
Alarm function................................................ 36 Main menu...................................................... 16
Menu system.................................................. 16
B
Multiple units.................................................. 34
Battery Voltage Monitor.................................. 44
BVM................................................................ 44 O
BVM Cal Kit.................................................... 47 Optional accessories...................................... 14
Optional equipment........................................ 42
C
Calibration...................................................... 50 P
Capacity calculation....................................... 19 Pausing the test.............................................. 25
Clamp-on-probe............................................. 14 Product description........................................... 6
Conducting a test........................................... 24 Protection system........................................... 10
Constant current............................................. 25
R
Constant power.............................................. 26
Regular load................................................... 35
Constant resistance........................................ 27
Regulate current............................................. 38
Current cables................................................ 24
Resetting the alarm........................................ 36
Current, regulate............................................ 38
Current shunt.................................................. 32 S
D Safety............................................................... 8
Safety instructions............................................ 8
Desired current............................................... 38
Screen buttons............................................... 16
E Side panel...................................................... 13
EMERGENCY STOP...................................... 12 Simulator mode.............................................. 19
Emergency stop button.................................. 25 Specifications................................................. 52
Ending the test............................................... 25 Starting the test.............................................. 25
Extension cables............................................ 15 Stop button..................................................... 25
External current measurement....................... 32 Stop limit......................................................... 17
Extra Loads.................................................... 42 Symbols on the instrumentI.............................. 8
F T
FTP browser................................................... 31 Temperature compensation............................ 18
Test at constant current.................................. 25
G
Test configuration........................................... 17
GRAPH........................................................... 20
TEST menu.................................................... 16
I Test method.................................................... 16
ID Labels........................................................ 17 Top panel........................................................ 12
Included accessories...................................... 13 TORKEL Viewer............................................. 22
Introduction....................................................... 6 TORKEL Win software................................... 44
Troubleshooting.............................................. 48
L TXL................................................................. 42
License........................................................... 19 TXL Extra Loads............................................. 38
Load profile..................................................... 28
Load-providing capability................................ 39
58 TORKEL 900-series ZP-CS01E CS033663IE
Index
U
Upgrade License............................................ 19
V
Voltage range................................................. 24
W
Warning limit................................................... 17
Warranty........................................................... 7
Warranty repair................................................. 7
Z
Zoom.............................................................. 20
Manufacturing sites
The company reserves the right to change the specification or design without prior notice.