Section VII - Schedule of Requirements 95
S.N Item Description Uni Qty Requ Crit Wei Scores Bidder
Name and Model of the product M
PO Box 25-11-276-32-66
Tel: +251-11-276-32-66
Fax: +251-11-275-25-56
Addis Ababa
1. Generic Name: Ultrasound-General Purpose, Eac QBit7
3 40
Color Doppler, Mobile, 3D h
2. GMDN/UMDN Code/ Name:
3. Clinical Purpose/Description:
A robust state of art fully digital color doppler YES
ultrasound system capable of performing imaging
applications in abdominal Ob/Gyn, musculoskeletal, M
cardiovascular, small parts, Urology, Cardiology and
Lung.
4. Technical Specification:
System broad band beam former capable of YES,1.5-
M 18MHz
processing signals from 2-15 MHZ
Mode: B-mode/M-mode, Color doppler mode/Power YES
doppler mode/PW doppler mode, Harmonic mode, M
CW mode, Color Flow imaging.
Height / Tilt / Rotate adjustable monitor or YES
M
equivalent
Functions: Full spectrum imaging, Hybrid full digital YES
beam forming, Pulse Inversion Harmonic (probe
dependent) , Harmonic imaging, Tissue doppler M
imaging, Power doppler imaging, Power pulse
inversion harmonic imaging, Trapezoidal imaging,
Lung assessment, Cardiac measurement
Row data management. M YES
Post processing technology. M YES
Speckle Reduction Filtering M SRA
The capability of analyzing 3D data set for 3D 4D
M
maging
5.00
Dynamic range of 180 db or higher. M R 200dB ≥250dB 290
%
256 shades of gray display M 256
Frame per second: 500 or more. M YES
Facilities for real time and frozen, pan or point zoom. M YES
Cine review for 2500 frames M YES
Doppler Auto trace M YES
Extensive software and automatic and user YES
programmable calculation package for gray scale, M
color Doppler, 3D applications.
Transducer: Active four ports or more M YES
Built in image Management software for off line YES
application when patient has gone after examination, M
such as image manipulation, Multi Planner
reformatting, surface & volume rendering etc.
Section VII - Schedule of Requirements 96
S.N Item Description Uni Qty Requ Crit Wei Scores Bidder
Skin line scaling markers, curved distance Diatance,angl
measurement tool and zoom, pan, rotate and trim e
M
facility to trim panoramic images from start or end of
the panoramic capture.
5.00 HDD
SSD: 500 GB or more M R 700GB ≥1TB
%
Should be DICOM 3.0 Compatible M YES
Should have push handle for pushing. M YES
Integrated trolley with brake M YES
Minimum 19” high resolution medical grade 25.00 21.5 inch
M R 21inch ≥24inch
TFT/LCD screen monitor display %
Transducers:
Curved (Convex) Probe (2-5 MHz ) for Abdomen D3C60L:2-
M 6.8MHz
and OB/GYN, etc.. (Specify model/cat no)
10.00 128
Number of elements:126 M R 150 ≥190
%
5.00
FOV: 50 deg M R 70deg. ≥90deg. 60 deg
%
Physical foot print: approx. 55 x 12mm M 73mm×16mm
Micro-curved Probe (4-9MHz) for neonatal and D6C15L:4-
M 12MHz
pediatrics, etc.. (Specify model/cat no)
10.00 128
Number of elements:126 M R 150 ≥190
%
5.00 ≥120deg 90 deg
FOV: 80 deg. M R
%
90deg.
.
Physical foot print: approx. 22 x 5mm M 26mm×11mm
Linear probe (5-12 MHz) for small parts, vascular, D7L40L:4-
M 15MHz
musculoskeletal, etc.. (Specify model/cat no)
10.00 128
Number of elements:126 M R 150 ≥190
%
FOV:36mm M 40mm
Physical foot print: approx. 42 x 6 mm M 44mm×11mm
Phased Array cardiac transducer (2-6 MHz ) (Specify D3P64L:1.5-
M 5.3MHz
model/cat no)
10.00 64
Number of elements: 64 M R 90 ≥128
%
FOV: 85 deg. M 82 deg
Endo cavity probe Trans-vaginal (4-9MHz) D6C12L:4-
applicable for examining internal organs such as 12MHz
M
vagina, cervix, uterus, fallopian, ovaries, rectum,
etc… (Specify model/cat no)
10.00 128
Number of elements: 126 M R 150 ≥190
%
5.00 ≥170deg 120deg
FOV:120 deg M R
%
150deg.
.
Physical foot print: approx.25 x 5 mm M 41mm×12mm
5. System Configuration Accessories,
Consumables and other components:
UPS with 60 minutes backup M
B/W thermal Printer M YES
Printer paper ---100 rolls M
Gel 250 ml ----- 2 M
Curved (Convex) Probe ---1 M YES
Section VII - Schedule of Requirements 97
S.N Item Description Uni Qty Requ Crit Wei Scores Bidder
Micro-curved Probe ---1 M YES
Linear probe ---1 M YES
Phased Array Cardiac Probe ---- 1 M YES
Endo cavity probe Trans-vaginal---- 1 M YES
All standard accessories and parts required to operate
the equipment and cleaning materials with their M
quantity to be included in the offer.
6. Operating Environment;
Operating Temperature:+10 °C to + 32°C M YES
Relative humidity : < 85% M YES
7. Utility Requirements:
Electrical Power Supply: 220VAC +10% M YES
8. Standards and Safety Requirements:
Should meet IEC-60601(Or Equivalent) General
M
Requirements of Electrical Safety
9. Installation, Training and Commissioning:
The supplier must provide installation, and
M
commissioning of the device at health Facility
The supplier must provide onsite technical and end M
user training
10. Warranty and After Sale service:
The supplier must be provide minimum of two years
warranty including labor and spare part from the date M
of commissioning.
After basic warranty the supplier must agree for
M
providing after sales service on agreement base
11. Documentation:
User and service manual in English M
12. Packaging and Labeling;
Packing of all the goods clearly marked and securely M
packed.
Each goods will be further packed in separate
package with all its standard accessories of distinct M
identification and numbers consecutively.
Additional packing and labeling requirements should
M
bear in each package
Each item with all accessories /spare part shall be
M
configured and packed in one unit.
Federal Democratic Republic of Ethiopia
M
Pharmaceuticals Supply Service (EPSS)
Tender and Purchase Order No. M
Name and Model of the product M
PO Box 25-11-276-32-65
Tel: +251-11-276-32-65
Fax: +251-11-275-25-55
Addis Ababa
4 1. Generic Name: X Ray - Radiography, Digital Eac 50