[UD-800] Inspection List OSP-4-93-00
Serial Number: Judgement PASS FAIL
(Software Ver.: )
Date: Approval Signature
Temperature & Humidity °C / %
Customer Name:
Distributer Name:
Preparer Signature:
1.Basic Inspection
[Common Inspection] (Mark Yes/No after your inspections)
Yes No Inspection Items Required Tool
□ □ Repair unit was cleaned up / sanitized before starting investigation. -
□ □ Serial number label is attached to the machine. -
□ □ Serial number on the serial number label can be recognized. -
□ □ No breakages or loosen power cable, inlet or fuse holder. -
□ □ No dents, bruises on exterior body. -
□ □ Date and customer settings can be stored correctly. -
□ □ Operation manual is equipped. -
[Operation Inspection] (Mark Yes/No after your inspections)
Yes No Inspection Items Required Tool
□ □ Measurement mode change can be done by operating touch panel. -
□ □ Brightness adjustment function on UTILITY menu works. -
□ □ Sound volume adjustment function on UTILITY menu works. -
□ □ Date & Time setting on UTILITY menu can be done. -
Freeze function on B-Diag mode works by pressing FREEZE button
□ □ -
on monitor or stomping FREEZE pedal of the footswitch.
Data storage to USB memory stick on B-Mode mode works by
□ □ -
stomping a SAVE/PRINT pedal of the footswitch.
□ □ Mode change can be done. [B-Mode], [Axial], [Pachy], [A-Diag] -
□ □ Test piece auto-measurement can be done properly on AXIAL mode. -
Test piece manual-measurement can be done properly on AXIAL
□ □ -
mode.
Measured data on AXIAL mode can be printed out by pressing a PRINT
□ □ -
button.
OSP-0-14-00 - 1/2 -
[UD-800] Inspection List OSP-4-93-00
2. Detailed Inspection
Yes No Inspection Items Required Tool
Software version is the latest in case there’re no restrictions to the
□ □ -
customer’s usage environment.
[Performance Inspection] (Mark Yes/No after your inspections)
Yes No Inspection Items Required Tool
Axial Probe
Measured value of test piece is within the range on a label.
[SN: ]
□ □ Measured value: mm, degree
Pachy Probe
[Range on label: mm to mm]
[SN: ]
□ □ Basic artifact can be seen on the monitor. 10MHz
□ □ No mechanical / electric noise from the probe. B-scan Probe
□ □ No cracks, no bruises on the probe cap. [SN: ]
3. Inspection Results
[Confirmation of Inspection Results] (Mark Yes/No after your inspections)
Yes No Inspection Items Required Tool
□ □ Repair unit was cleaned up / sanitized before packing and shipping. -
[Comments] (Please write down if you have a point of notice)
OSP-0-14-00 - 2/2 -