Selenia Dimensions ® ®
3Dimensions TM
Digital Mammography System
Digital Tomosynthesis System
Customer Release Notes
For Software Versions 1.10 and 2.1
Part Number MAN-05762
Revision 002
August 2019
Technical Support
USA: +1.877.371.4372 Asia: +852 37487700
Europe: +32 2 711 4690 Australia: +1 800 264 073
All Other: +1 781 999 7750 Email: BreastHealth.Support@hologic.com
© 2019 Hologic, Inc. Printed in the USA. This manual was originally written in English.
Hologic, Selenia, Dimensions, 3Dimensions, Affirm, C-View, Cenova, I-View, ImageChecker, Quantra, and associated logos
are trademarks and/or registered trademarks of Hologic, Inc., and/or its subsidiaries in the United States and/or other
countries. All other trademarks, registered trademarks, and product names are the property of their respective owners.
This product may be protected by one or more U.S. or foreign patents as identified at www.Hologic.com/patents.
Hologic Inc. Brazilian Contact:
36 Apple Ridge Road Imex Medical Group do Brasil
Danbury, CT 06810 USA Rua das Embaúbas, 601- Fazenda Santo Antônio
1.800.447.1856 São José /SC - Brasil - 88104-561
+55 48 3251-8800
www.imexmedicalgroup.com.br
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
Table of Contents
Table of Contents
1: Quality Control Requirements (US Sites Only) ___________________________________1
1.1 Introduction ............................................................................................................................................................. 1
1.2 Radiologic Technologist ......................................................................................................................................... 1
1.3 Medical Physicist .................................................................................................................................................... 1
1.4 Applications Support ............................................................................................................................................. 2
2: Selenia Dimensions Software 1.10 and 3Dimensions Software 2.1 Release Notes _____3
2.1 Introduction ............................................................................................................................................................. 3
2.2 Optional Licensed Features ................................................................................................................................... 3
2.3 ImageChecker Computer Aided Detection (CAD) Software ............................................................................ 4
2.4 Quantra Breast Density Assessment Software .................................................................................................... 4
3: Windows 10 Operating System _________________________________________________5
4: Software Enhancements _______________________________________________________6
4.1 Image Presentation Enhancements....................................................................................................................... 6
4.2 User Interface Enhancements ................................................................................................................................ 7
4.3 Biopsy Enhancements ............................................................................................................................................ 8
4.4 Cybersecurity Enhancements ................................................................................................................................ 8
4.5 Quality Control Enhancements ............................................................................................................................. 9
MAN-05762 Revision 002 iii
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
1: Quality Control Requirements (US Sites Only)
1: Quality Control Requirements (US Sites Only)
1:
1.1 Introduction
This document is provided as an overview of Selenia® Dimensions® software version 1.10
and 3Dimensions™ software version 2.1.
Note
This document is not meant to replace the Selenia Dimensions system or 3Dimensions
system User Guide. Changes described in these release notes may not be reflected in the
current revision of the User Guide.
In order to have your system upgraded to Selenia Dimensions software 1.10 or
3Dimensions software 2.1, you may have to have your Acquisition Workstation
hardware upgraded to a level that accommodates this new software, as some features
have specific hardware requirements. Please check with your Hologic sales
representative about whether a hardware upgrade is necessary.
1.2 Radiologic Technologist
Note
If the system was upgraded from software versions 1.9 or 2.0, this section applies.
This software upgrade requires that the radiologic technologist performs the following
QC tests by following the corresponding tests in the technologist section of the Selenia
Dimensions/3Dimensions system Quality Control Manual, part number MAN-03706:
• Phantom image evaluation
• Signal-to-noise and contrast-to-noise measurements
• DICOM printer quality control
The preceding tests shall be conducted on each individual Selenia Dimensions system
and 3Dimensions system that was upgraded to this software release.
1.3 Medical Physicist
The software upgrade does not require any testing by a medical physicist. However, the
tests described in the preceding section, which the technologist performs, are considered
to be conducted under the oversight of the medical physicist retained by the facility. The
medical physicist needs to be made aware of and provided the opportunity to review the
results of the tests.
The medical physicist should check that the dose reported on the Mammography QC
phantom after the software upgrade is similar to the dose reported during the last
phantom image quality evaluation test performed by the technologist prior to the
upgrade.
MAN-05762 Revision 002 1
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
1: Quality Control Requirements (US Sites Only)
1.4 Applications Support
You can contact Hologic with any questions about this software version.
• In the United States: call the Hologic Applications Hotline at 877-371-4372.
• In Europe and the Middle East: email to BE-Applications@hologic.com.
• In Asia-Pacific: email to AP-AppsSupport@hologic.com.
• In Australia/New Zealand: email to AU-ApplicationsSupport@hologic.com.
MAN-05762 Revision 002 2
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
2: Selenia Dimensions Software 1.10 and 3Dimensions Software 2.1 Release Notes
2: Selenia Dimensions Software 1.10 and 3Dimensions
2:
Software 2.1 Release Notes
2.1 Introduction
This chapter provides an overview of the enhancements associated with the 1.10 Selenia
Dimensions software upgrade and the 2.1 3Dimensions software upgrade. This upgrade
can affect daily workflow or other tasks. Carefully review these customer release notes to
understand the new software enhancements and software changes introduced with this
upgrade.
Note
This document is not meant to replace the Selenia Dimensions system and 3Dimensions
system User Guide. Changes described in these customer release notes may not be
reflected in the current revision of the User Guide.
2.2 Optional Licensed Features
This release of the Selenia Dimensions software and 3Dimensions software introduces two
optional features, sold separately. The following licensed features are available:
• ImageChecker® Computer Aided Detection (CAD) Software
• Quantra™ 2.2 Breast Density Assessment Software
In order to have your system upgraded, you may have to have your Acquisition
Workstation hardware upgraded to a level that accommodates this new software, as some
features have specific hardware requirements. Again, please check with your Hologic
sales representative about whether a hardware upgrade is necessary.
Note
Availability of these optional features may depend on local regulations – please check
with your Hologic sales representative to find out if these features are available in your
region.
MAN-05762 Revision 002 3
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
2: Selenia Dimensions Software 1.10 and 3Dimensions Software 2.1 Release Notes
2.3 ImageChecker Computer Aided Detection (CAD) Software
2.3.1 Feature Description
ImageChecker 10.0 computer aided detection (CAD) that is currently offered in the
Hologic Cenova™ image analytics server is now also available in the Selenia Dimensions
1.10 and 3Dimensions 2.1 software.
2.3.2 Prerequisites
None
2.4 Quantra Breast Density Assessment Software
2.4.1 Feature Description
Quantra 2.2 breast density assessment software that is currently offered in the Hologic
Cenova image analytics server is now also available in the Selenia Dimensions 1.10 and
3Dimensions 2.1 software.
MAN-05762 Revision 002 4
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
3: Windows 10 Operating System
3: Windows 10 Operating System
3:
The Selenia Dimensions 1.10 and 3Dimensions 2.1 software runs exclusively on a
customized version of Windows 10, which has mainstream support from Microsoft until
October 2025.
MAN-05762 Revision 002 5
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
4: Software Enhancements
4: Software Enhancements
4:
4.1 Image Presentation Enhancements
4.1.1 Implant Brightness Option
There is a new optional image processing setting for 2D imaging to reduce the implant
brightness to improve visibility of the implant/breast boundary region. To enable this
option, please contact your Hologic Field Service Engineer.
4.1.2 Skin Line Brightness Reduction Option
There is a new optional image processing setting for 2D imaging to reduce the brightness
and overshadow of the skin on screening views. To enable this option, please contact your
Hologic Field Service Engineer.
MAN-05762 Revision 002 6
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
4: Software Enhancements
4.2 User Interface Enhancements
4.2.1 Support for Siemens Grayscale Presentation State (GSPS) Objects
Grayscale Presentation State (GSPS) overlays created on Siemens Workstations can now
be viewed on the Acquisition Workstation.
4.2.2 Cine Tab Automatically Selected when Projection Image Is Selected
To better facilitate reviewing projection images for motion, when the Projection Image
view button is selected, the Tools menu will automatically select the Cine tool tab.
4.2.3 Eject USB
There is now a button to safely eject a USB device from within the Admin menu.
4.2.4 Auto Pair Override Now Automatically Moves to 1-Up Display
When the Auto-Pair override button is selected, the image display will now automatically
revert to 1-Up mode.
MAN-05762 Revision 002 7
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
4: Software Enhancements
4.3 Biopsy Enhancements
4.3.1 Resend Targets Button Now Disabled when No Active Targets on
View
To better differentiate the Resend Targets button from the Project Targets button, when
the current biopsy view has no active targets, the Resend Targets button will be disabled.
4.3.2 Warning Displayed when a Target Is Created with a Non-clinical
Needle
A warning dialog will now be displayed to the user when a biopsy target is created on a
patient image when a non-clinical needle, such as the QAS needle, is selected.
4.4 Cybersecurity Enhancements
4.4.1 Cybersecurity Hardened Windows 10 Operating System
Hologic’s team of Certified Information Systems Security Professionals (CISSP) and
Certified Secure Software Lifecycle Professionals (CSSLP), utilizing guidance from the
NIST Cybersecurity framework, have designed a custom version of Windows 10 that is
hardened against cybersecurity threats.
4.4.2 User Management Via Windows 10
All user management and authentication, including password policies, is now handled by
the Windows 10 operating system (for local authentication) or domain level (if Active
Directory is used.)
4.4.3 All ePHI Encrypted at Rest
All ePHI on disk, both within the DICOM image files and within the database is
encrypted using AES-256 strong encryption.
4.4.4 Windows Firewall Enabled
Windows Firewall is now enabled by default with the appropriate port exclusions
configured.
4.4.5 Clinical User Accounts Function on Standard Windows Accounts
All clinical user accounts operate on non-administrator Windows user accounts for better
system security.
4.4.6 OS Patches
All necessary OS patches released before the final release of this software are installed on
the system.
MAN-05762 Revision 002 8
Selenia Dimensions 1.10 and 3Dimensions 2.1 Customer Release Notes
4: Software Enhancements
4.5 Quality Control Enhancements
4.5.1 QC Phantom CEDM and CEDM Combo Views Added
QC Phantom CEDM and CEDM Combo Views have been added to the software.
4.5.2 Additional Test Patterns Included
The following patterns have been natively added to the software and are available on the
Test Patterns screen: TG18-LPH10, TG18-LPH50, TG18-LPH89, TG18-LPV10, TG18-
LPV50, TG18-LPV89.
MAN-05762 Revision 002 9