Astm E1251 24
Astm E1251 24
for the
     Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1251 − 24
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
                                                                                         1
                                                                               E1251 − 24
   E29 Practice for Using Significant Digits in Test Data to                            particular alloy being analyzed. For best results, the reference
     Determine Conformance with Specifications                                          material being used should be within 610 % of the composi-
   E135 Terminology Relating to Analytical Chemistry for                                tion (for each respective element) of the material being
     Metals, Ores, and Related Materials                                                analyzed.
   E305 Practice for Establishing and Controlling Spark
     Atomic Emission Spectrochemical Analytical Curves                                  4. Summary of Test Method
   E406 Practice for Using Controlled Atmospheres in Atomic
                                                                                           4.1 A controlled electrical discharge is produced in an argon
     Emission Spectrometry
                                                                                        atmosphere between the prepared flat surface of a specimen
   E691 Practice for Conducting an Interlaboratory Study to
                                                                                        and the tip of a counter electrode. The energy of the discharge
     Determine the Precision of a Test Method
                                                                                        is sufficient to ablate material from the surface of the specimen,
   E716 Practices for Sampling and Sample Preparation of
                                                                                        break the chemical or physical bonds, and cause the resulting
     Aluminum and Aluminum Alloys for Determination of
                                                                                        atoms or ions to emit radiant energy. The radiant energies of
     Chemical Composition by Spark Atomic Emission Spec-
                                                                                        the selected analytical wavelengths and the internal standard
     trometry
                                                                                        wavelength(s) are converted into electrical signals by either
   E826 Practice for Testing Homogeneity of a Metal Lot or
                                                                                        photomultiplier tubes (PMTs) or a suitable solid state detector.
     Batch in Solid Form by Spark Atomic Emission Spec-
                                                                                        The detector signals are electrically integrated and converted to
     trometry (Withdrawn 2023)3
                                                                                        a digitized value. The signals are ratioed to the proper internal
   E1329 Practice for Verification and Use of Control Charts in
                                                                                        standard signal and converted into mass fractions.
     Spectrochemical Analysis (Withdrawn 2019)3
   E1507 Guide for Describing and Specifying the Spectrom-                                 4.2 Three different methods of calibration defined in 3.2.1,
     eter of an Optical Emission Direct-Reading Instrument                              3.2.2, and 3.2.3 are capable of giving the same precision,
   E2972 Guide for Production, Testing, and Value Assignment                            accuracy, and detection limit.
     of In-House Reference Materials for Metals, Ores, and                                 4.2.1 Binary calibration employs calibration curves that are
     Other Related Materials                                                            determined using a large number of high-purity binary calibra-
   2.2 ANSI Standard:4                                                                  tion materials. This approach is used when there is a need to
   ANSI H35.1/H35.1M American National Standard Alloy                                   analyze almost the entire range of aluminum alloys. Because
     and Temper Designation Systems for Aluminum                                        binary calibration materials may respond differently from alloy
                                                                                        calibration materials, thus the latter are used to improve
3. Terminology                                                                          accuracy by applying a slope or intercept correction, or both, to
  3.1 Definitions—For definitions of terms used in this                                 the observed readings.
Standard, refer to Terminology E135.                                                       4.2.2 Global calibration employs calibration curves that are
                                                                                        determined using many different alloy calibration materials
   3.2 Definitions of Terms Specific to This Standard:                                  with a wide variety of compositions. Mathematical calculations
   3.2.1 alloy-type calibration—calibration curves determined                           correct for both alloy difference and inter-element effects. Like
using calibration materials from alloys with similar composi-                           4.2.1, specific alloy calibration materials may be used to apply
tions.                                                                                  a slope or intercept correction, or both, to the observed
   3.2.2 binary-type calibration—calibration curves deter-                              readings.
mined using binary calibration materials (primary aluminum to                              4.2.3 Alloy calibration employs calibration curves that are
which has been added one specific element).                                             determined using different alloy calibration materials that have
   3.2.3 global-type calibration—calibration curves deter-                              similar compositions. Again, specific alloy calibration materi-
mined using calibration materials from many different alloys                            als may be used to apply a slope or intercept correction, or
with considerable compositional differences.                                            both, to the observed readings.
   3.2.4 two-point drift correction—the practice of analyzing
high and low drift correction materials for each calibration                            5. Significance and Use
curve and adjusting the intensities obtained back to the values                            5.1 The metallurgical properties of aluminum and its alloys
obtained on those particular drift correction materials during                          are highly dependent on chemical composition. Precise and
the collection of the calibration data. The corrections are                             accurate analyses are essential to obtaining desired properties,
calculated mathematically and are applied to both the slope and                         meeting customer specifications, and minimizing scrap due to
intercept. Improved precision may be obtained by using a                                off-grade material.
multi-point drift correction as described in Practice E1329.                               5.2 This test method is applicable to chill cast specimens as
   3.2.5 type standardization—mathematical adjustment of the                            defined in Practices E716 and can also be applied to other types
calibration curve’s slope or intercept using a single reference                         of samples provided that suitable reference materials are
material at or close to the nominal composition for the                                 available. Also, other sample forms can be melted and cast into
                                                                                        a disk, using an appropriate mold, as described in Practices
   3
                                                                                        E716. However, some elements (for example, magnesium)
     The last approved version of this historical standard is referenced on
                                                                                        readily form oxides, while others (for example, sodium,
www.astm.org.
   4
     Available from American National Standards Institute (ANSI), 25 W. 43rd St.,       lithium, calcium, and strontium) are volatile, and may be lost to
4th Floor, New York, NY 10036, http://www.ansi.org.                                     varying degrees during the melting process.
                                                                                    2
                                                                     E1251 − 24
6. Recommended Analytical Wavelengths and Potential                           equipped with an exhaust system that will safely dispose of the
Interferences                                                                 argon gas and the metal dust created during the excitation
  6.1 Table 1 lists the analytical wavelengths commonly used                  cycle. Considering health and cleanliness, the exhausted gas
for aluminum analysis. Other lines may be used if they give                   and dust should not be vented directly into the laboratory.
comparable results. Also listed are recommended mass fraction                 Manufacturers have designed their instruments with some type
ranges, background equivalent mass fractions, detection limits,               of exhaust/filter system to minimize this problem. The exhaust
useful linear ranges, and potential interferences. The values                 can then be vented into an efficient hood system.
given in this table are typical; actual values obtained are                      7.4 Gas Flow System, will be designed so that it can deliver
dependent on instrument design.                                               pure argon gas to the excitation chamber. The purity of the
    NOTE 2—The background equivalent mass fraction and detection limits       argon gas will affect the precision of the results. Argon gas with
listed in Table 1 have been attained with a spectrometer that has a           a minimum purity of 99.995 % has been found acceptable. The
reciprocal dispersion of 54 nm/mm and a working resolution of 3.5 nm,
using an entrance slit width of 25 µm and exit slit widths of 50 µm.
                                                                              gas shall be delivered by a flow system as described in Practice
                                                                              E406. The argon gas source can be from high-purity com-
7. Apparatus                                                                  pressed gas cylinders, a cryogenic-type tank that contains
                                                                              liquid argon or possibly from a central supply (liquid only). It
   7.1 Specimen Preparation Equipment:                                        is essential that only argon gas meeting the minimum purity of
   7.1.1 Sampling Molds, for aluminum, the techniques of                      99.995 % be used. A lower purity grade of argon, such as a
pouring a sample disk are described in Practices E716. Chill                  “welding grade,” should not be used. The delivery system shall
cast samples, poured and cast as described within Practices                   be composed of a two-stage type (high/low pressure) regulator
E716, shall be the recommended form in this test method.                      of all-metal construction with two pressure gages. Delivery
   7.1.2 Lathe, capable of machining a smooth, flat surface on                tubing must not introduce any contamination of the argon
the reference materials and samples. A variable speed cutter, a               stream. Refrigerant-grade copper tubing is recommended. The
cemented carbide or polycrystalline diamond tool bit, and an                  gages on the regulator will allow for the adjustment of the gas
automatic cross feed are highly recommended. Proper depth of                  pressure to the instrument. Delivery pressure specifications
cut and desired surface finish are described in Practices E716.               will vary with instrument manufacturer. Please note that the
   7.1.3 Milling Machine, a milling machine can be used as an                 delivery tube connections should be made with all metal seals
alternative to a lathe.                                                       and the delivery tubing should be kept as short as possible. All
   7.1.4 It is strongly recommended that the same preparation                 metal connections are strongly recommended because the
machinery used to prepare the reference materials is also used                discharge is adversely affected by organic contamination, or by
to prepare the samples. Differences in surface characteristics                as little as 2 ug ⁄g of oxygen or a few ug/g of water vapor.
may influence the analysis.                                                   Argon supply shall be sufficient to support the required flow
   7.2 Excitation Source—In today’s instrumentation, the ex-                  during analysis and bleed during idle periods. All connections
citation source is computer controlled and is normally pro-                   must be leak-free.
grammed to produce: (1) a high-energy pre-spark (of some                         7.5 Measuring and Control System of the instrument con-
preset duration), (2) a spark-type discharge (of some preset                  sists of either photomultiplier and integrating electronics or
duration), (3) an arc-type discharge (of some preset duration),               solid-state photosensitive arrays (CCD or CID) that convert
and (4) a spark-type discharge, during which, time-resolved                   observed light intensities to a digitizable signal. A dedicated
measurements are made for improved detection limits (this                     computer or microprocessor is used to control burn conditions,
may be optional on some instruments).                                         source operation, data acquisition, and the conversion of
   7.2.1 Typical parameters and times are given in Table 2.                   intensity data to mass fractions. Data should be accessible to
Note that the information presented is given as an example                    the operator throughout all steps of the calculation process. The
only and parameters may vary with respect to instrument                       instrument’s control software should include functions for drift
model and manufacturer.                                                       correction (standardization), type standardization, and the ap-
   7.3 Excitation Chamber, shall be designed with an upper                    plication of these functions to subsequent analyses.
plate that is smooth and flat so that it will seal tightly with the
prepared surface of the sample specimen. The seal that is                     8. Materials
formed between the two will prevent atmosphere from entering                     8.1 Counter Electrode—The counter electrode and speci-
the discharge chamber. The excitation chamber will contain a                  men surface are the two terminus points of the spark discharge.
mounting clamp to hold the counter electrode. The excitation                  The counter electrode should be made from tungsten or other
stand assembly will also have some type of clamp or device                    suitable material and have a pointed end. The gap distance
designed to hold the sample firmly against the top plate. Some                between the specimen surface and the tip of the counter
manufacturers may provide liquid cooling for the top plate to                 electrode is specified by the manufacturer. The diameter and
minimize sample heating during the excitation cycle. The                      geometry of the counter electrode is also application and
excitation chamber will also be constructed so that it is flushed             manufacturer dependent. If different designs or configurations
automatically with argon gas during the analytical burn cycle.                are offered, it is recommended that the prospective purchaser
The excitation chamber’s design should allow for a flow of                    test each design to determine which performs best for the
argon gas to prevent the deposition of ablated metal dust on the              intended analytical task. The counter electrode configuration
inner-chamber window(s). The excitation chamber will be                       and auxiliary gap distance must not be altered subsequent to
                                                                          3
                                                                    E1251 − 24
                                                  TABLE 1 Recommended Analytical Lines
                         Wavelength      Recommended        Background            Calculated        High
                                                                                                                           Interferences
          Element          in Air        Mass Fraction      Equivalent,            Detection    Mass Fraction
                                                                                                                      Element, λ(nm) and k, %F
                           (nm)A           Range, %            %B                 Limit, %C,D    Index, %E
Aluminum            256.799   I       70-100
                    266.039   I       70-100
                    237.208   I       70-100
Antimony            231.147   I       0.001-0.5            0.17           0.0002                                Co 231.166           0.6
                    259.806   I       0.001-0.5                           0.0002                                Fe 259.837
                                                                                                                Mn 259.817           0.01
Arsenic             234.984   I       0.005-0.1
Beryllium           234.861   I       0.0001-0.05          0.001          0.00003
                    313.042   II      0.0001-0.05          0.0035         0.00001
                    332.134   I       0.0001-0.05                         0.00001
Bismuth             306.772   I       0.001-0.7            0.04           0.0002
Boron               249.773   I       0.0001-0.05          0.002          0.0001*                               Fe 249.782           0.001
                                                                                                                Mn 249.778           0.007
                    249.678 I         0.0001-0.05
                    208.959 I         0.0001-0.05                                                               Mo 208.952           0.13
Cadmium             228.802 I         0.001-1              0.05           <0.0001                               As 228.812
                    479.992 I         0.005-2              0.15           0.003
Calcium             393.367 IIG       0.001-0.05           0.001          0.00005                               Fe 393.361
Chromium            425.435 I         0.001-1              0.015          <0.0001
                    267.716 II        0.001-1              0.004          0.0005*
                    276.654 IIG       0.005-1
Cobalt              345.351 I         0.0001-2                            <0.0001
Copper              327.396 I         0.001-1.5            0.005          <0.0001               0.7
                    324.754 I         0.001-0.5
                    296.117 I         0.05-20              0.40           0.01*                 >20             Fe 296.128
                    224.700 II        0.01-5               0.03           0.0005*               5
                    510.554 I         0.05-20              0.32           0.01*                 >20
Gallium             294.364 I         0.001-0.05           0.015          <0.0001
                    417.206 IG        0.001-0.05                                                                Fe 417.213
                                                                                                                Ti 417.190
                                                                                                                Cr 417.167
Iron                238.204   II      0.001-1.5            0.015          0.0008                1.0
                    259.940   II      0.001-1.5            0.005          0.0004
                    259.957   I
                    273.955   II      0.01-3.5
                    374.949   IG      0.001-3.5                           0.0001
                    441.512   I       0.01-3.5                            0.0004
                    438.355   I       0.005-3.5
Lead                405.782   I       0.002-0.7            0.04           0.0001                                Mn 405.792           0.01
                                                                                                                Mg 405.763           0.001
                    283.306   I       0.002-0.7            0.07           0.002
Lithium             610.364   I       0.0001-3
                    670.784   I       0.0001-0.02          0.0005         <0.0005
                    323.261   I       0.01-3                                                                    Fe 323.279
                                                                                                                Sb 323.250
Magnesium           279.553   II      0.0005-0.3           0.0006         0.00003               0.04
                    285.213   I       0.0005-0.3           0.008          <0.0001               0.25
                    277.669   I       0.05-11              0.08           0.01                  >11
                    383.231   IG      0.01-11              0.015          0.002*                >11
                    383.826   I       0.1-11
                    518.362   I       0.01-11              0.02           0.002*                >11
Manganese           403.076   IG      0.001-0.1            0.028          0.0001*
                    259.373   II      0.0005-0.5           0.004          0.00005               0.2
                    293.306   II      0.001-2              0.006          0.0002*               >1.1
                    346.033   II      0.01-2
Nickel              341.476   I       0.001-2              0.02           <0.0001               >2.5            Zr 341.466           0.01
                    310.188   I       0.005-4              0.05           0.001*                >5
                    231.604   II      0.001-2              0.015          0.0005*               <2.5
Phosphorus          178.231   IH      0.0001-0.1           0.084          0.0001
Silicon             288.158   I       0.001-1.5            0.01           0.0001                1.5             Cr 288.123
                    251.612   I       0.001-1.5            0.006          0.0001                1.5
                    390.553   IG      0.05-24              0.25           0.01                  >24             Cr 390.566           0.09
                    212.415   I       0.05-24              0.5            0.05                  >24
Silver              328.068   I       0.0005-0.1
                    338.289   I       0.0001-0.1                                                >10
                    466.848   I       0.05-1.5
Sodium              588.995   I       0.0001-0.05          0.0015         <0.0001
Strontium           421.552   IIG     0.0001-0.1           0.0004         0.0001
                    460.733   I       0.0005-0.06
Tin                 317.502   I       0.001-7.5            0.04           0.0001                >10
Titanium            334.904   II      0.0005-0.5           0.004          <0.0001
                    337.280   II      0.001-0.5            0.002          <0.00010
                    363.545   I       0.0005-0.05          0.030          0.003*
                                                                     4
                                                                                  E1251 − 24
                                                                         TABLE 1      Continued
                              Wavelength              Recommended          Background           Calculated             High
                                                                                                                                              Interferences
         Element                in Air                Mass Fraction        Equivalent,           Detection         Mass Fraction
                                                                                                                                         Element, λ(nm) and k, %F
                                (nm)A                   Range, %              %B                Limit, %C,D         Index, %E
Vanadium                318.341   I               0.001-0.15             0.06             0.0003*
                        437.924   I               0.001-0.25
                        310.230   II              0.001-0.15             0.014            <0.0001
Zinc                    213.856   I               0.0005-0.1             0.035            0.0001*                 0.05
                        334.502   I               0.001-10.0             0.065            0.0004                  >8
                        481.053   I               0.01-10                0.07             0.001*                  >10
                        472.216   I               0.01-10                0.26             0.0015                  >10
Zirconium               339.198   II              0.001-1                0.02             0.001*
                        349.621   IIG             0.001-1                0.006            <0.0001
A
  I = atom line, II = ion line.
B
  Background Equivalent Mass Fraction—The mass fraction at which the signal due to the element is equal to the signal due to the background.
C
  In this test method, the calculated detection limit was measured by calculating the standard deviation of ten consecutive burns on a specimen with element mass
fraction(s) at levels below ten times the expected detection limit.
D
  See footnote C. For values marked with an asterisk (*) the available data were for a mass fraction greater than ten (10) times but less than a hundred (100) times the
expected detection limit.
E
  High Mass Fraction Index—The mass fraction at which the slope of the calibration curve drops below 0.75.
F
  Interference Factor, k—The apparent increase in the mass fraction of the element being determined, expressed in percent, due to the presence of 1.0 % of the interfering
element.
G
   Useful analytical wavelengths with improved signal to background ratios due to the complete removal of C-N background by the argon atmosphere.
H
  If phosphorus is determined, the most sensitive wavelength appears to be the 178.231 nm in the second order which requires either a vacuum or a gas filled
spectrometer. The vacuum spectrometer should be operated at a pressure of 25 microtorr or less. The gas filled spectrometer will be charged with nitrogen to a positive
pressure of slightly over one atmosphere (101 k pa). Optimum results are obtained by using a background channel that has been profiled “off peak” of the first order 178.231
nm phosphorus line as the internal standard. The ratio of P 178.231 nm (2nd) / background near the 178.231 nm (1st) is plotted against % phosphorus. Even with this
compensation for variability in background, alloys with highly different compositions of major alloying elements, particularly silicon, require separate reference materials
and analytical curves.
       TABLE 2 Typical Excitation Source Electrical Parameters                           available from the National Institute of Standards and Tech-
           Parameter
                                        High Energy
                                                               Spark        Arc
                                                                                         nology (NIST). Also, there are other commercial sources for
                                         Pre-spark                                       aluminum reference materials.
    Resistance, Ω                           1                   1           15              9.1.2 For trace elements, reference materials that contain
    Inductance, µH                          30                 130          30
    Volts, V                               400                 400         400           variable mass fractions of the trace element in a typical alloy of
    Frequency, Hz                          300                 300         300           constant or nearly constant composition are available. These
    Capacitance, µF                         12                  3           5
                                                                                         reference materials can be used for establishing the analytical
    Time, s                                 10                  5           5
                                                                                         curve, but will not reveal potential interferences from nearby
                                                                                         wavelengths of other elements, or matrix effects that change
spectrometer calibration or calibration adjustments. Electrode                           instrument response or background. For optimum usefulness,
maintenance (frequent brushing of the counter electrode) to                              several of the calibration materials should have mass fractions
maintain its configuration, gap distance, and minimize surface                           for the other elements that vary over the expected ranges in the
contamination are critical to accurate, precise analytical results.                      specimen to be analyzed.
It is recommended that the instrument purchaser order several                               9.1.3 Atomic emission analysis is a comparative technique
spare counter electrodes so that they can be replaced when                               that requires a close match of the metallurgy, structure, and
necessary.                                                                               composition between the reference material and the test
                                                                                         material. Differences in structure, such as result from the
9. Reference Materials                                                                   sodium modification of high silicon alloys, or differences in
   9.1 Calibration Materials—All calibration materials shall                             metallurgical history, due to extruding, rolling, or heat treating,
be homogeneous and free of cracks or porosity. These materials                           induce a variety of effects that can influence the analytical
should also possess a metallurgical condition that is similar to                         results. To ensure analytical accuracy, care must be taken to
the material(s) that are being analyzed. The calibration mate-                           match the characteristics of the reference material to that of the
rials shall be used to produce the analytical curves for the                             test material or suitable corrections to adjust for these influ-
various elements being determined.                                                       ences must be established.
   9.1.1 It is recommended that a calibration curve for any                                 9.2 Drift Correction (Standardization) Materials:
particular element be composed of a minimum of four calibra-                                9.2.1 Materials for Drift Correction—Both high and low
tion materials. The mass fractions of these calibration materials                        mass fraction materials are available from several commercial
should be fairly evenly spaced over the calibrated analytical                            sources or may be developed in-house. The low material is
range so that a mathematically valid calibration curve can be                            usually high-purity aluminum. The high material(s) should
established using all of the points.                                                     have mass fractions near or above the median mass fraction for
   9.1.1.1 The calibration materials used shall be of sufficient                         the calibrated range of each wavelength. The commercially
quality, purchased from a recognized reputable source, and                               available materials are tested for homogeneity and reproduc-
have certified values to the required accuracy for the antici-                           ibility of spectral response but are not necessarily certified for
pated analytical tasks to be performed. A few SRMs are                                   composition of individual elements. Composition certification
                                                                                     5
                                                              E1251 − 24
is not required because these materials are only used to adjust        12. Preparation of Reference Materials and Specimen
intensity ratios back to those obtained during the initial                12.1 Preparation of Reference Materials—All reference
calibration of the instrument. Care should be exercised when           materials shall have their surfaces prepared for analysis ac-
replacing depleted materials with new ones that are from               cording to Practices E716 with the cutting depth usually
different heats or lots, since the actual mass fraction of the         limited to that required to produce a fresh surface (about 0.010
individual element(s) may be different from the material               in. or 250 µm). The surfaces of the reference materials and the
currently in use. Whenever materials are replaced, appropriate         surfaces of the test specimens shall be prepared in the same
procedures must be followed to reference the intensities               manner. See Practices E716 for details.
obtained from the new material to the intensities obtained from
the material being replaced. See 14.3 for details.                       12.2 Preparation of Specimens—For techniques to select
   9.2.2 High-Purity Drift Correction Materials—These shall            and prepare both chill cast samples and other forms of
be homogeneous and shall consist of aluminum with the lowest           aluminum, such as sheet, plate, extrusions, and castings refer to
available mass fraction of the elements being determined.              Practices E716.
These materials are used to establish the background readings            12.3 To achieve the best analytical results, both reference
of the spectrometer for most elements. Their exact composi-            materials and sample specimens should have fresh surfaces.
tions need not be known.                                               Surfaces should not be tested that: (1) are freshly prepared, (2)
   9.2.3 Blank Drift Correction Materials—These materials              appear oxidixed, (3) have porosity, inclusions or other foreign
shall be homogeneous and of similar composition to the alloy           substances, or (4) have been contaminated by handling.
type calibration materials as described in 9.1 but will contain
the lowest available mass fractions of the trace elements being        13. Preparation of Apparatus
determined. They may be used if the lowest mass fraction of              13.1 Prepare the spectrometer for operation in accordance
the element being determined is within ten times the detection         with the manufacturer’s instructions.
limit of that element.
                                                                          NOTE 3—It is not within the scope of this method to prescribe all of the
   9.2.4 Type Standardization Materials—These should be ref-           details that are associated with the correct operation of any spectrometer.
erence materials or equivalent materials that closely match the        The reader is referred to the manufacturer’s manual. Additionally, it is
metallurgical history and composition of the unknown(s).               recommended that the purchaser of the spectrometer determine if training
These usually provide a nominal mass fraction reference point          courses are offered by the manufacturer. Frequently manufacturer’s will
                                                                       offer specific spectrometer training courses several times yearly.
which the instrument’s computer software can use to calculate
a slope or intercept correction to the observed readings to               13.1.1 Instrument Configuration—Instruments are usually
fine-tune the instrument’s calculated response for each element        pre-configured for the analytical program (elements), mass
of interest. This correction is then applied to each subsequent        fraction ranges, and alloy families according to specifications
analysis. When using this approach it is assumed that the              that have been requested by the purchaser. Optionally, the
composition(s) of the unknown(s) will be essentially similar to        purchaser may also choose to specify that the instrument is
the composition of this material.                                      completely pre-calibrated for all alloys and all intended ana-
                                                                       lytical tasks. The purchaser also has the option of completely
10. Hazards                                                            configuring and calibrating the instrument. When this is done,
   10.1 The spark discharge presents a potential electrical            careful attention is required in the selection of the correct
shock hazard. The spark stand or the sample clamping device,           analytical conditions, analytical channels, internal standard
or both, shall be provided with a safety interlock system to           channels, calibration ranges, and calibration materials to meet
prevent energizing the electrode whenever contact can be made          the specific analytical tasks. Whether the manufacturer or the
with the electrode. The instrument should be designed so               end user calibrates an instrument, it is the responsibility of the
access to the power supply is restricted by the use of safety          end user to verify that the instrument is performing according
interlocks.                                                            to the specifications that have been defined in the initial
                                                                       agreement or according to the performance as stated by the
   10.2 Fumes of the fine metallic powder that are exhausted           manufacturer. It is beyond the scope of this test method to
from the excitation chamber can be poisonous if the sample             describe the intricacies of complete instrument configuration.
specimens contain significant levels of hazardous elements.            The user should consult the manufacturer’s hardware and
Therefore, the instrument shall be designed with an internal           software manuals for specific configuration requirements.
exhaust system that is equipped with its own set of filters.              13.1.2 Profiling the Instrument—Profile the instrument ac-
Additionally, the instrument exhaust (after being filtered), may       cording to the manufacturer’s instructions. If the instrument is
be vented directly to a safe area. The filters should be cleaned       newly installed, it is recommended that the profile be checked
or changed according to the manufacturer’s recommendations             several times during the first few weeks of operation to
to enable correct instrument operation.                                determine the stability of the unit. Compare the differences in
                                                                       the settings to the tolerance variability allowed by the manu-
11. Sampling                                                           facturer.
   11.1 Chill Cast Disks and Other Aluminum Forms—For the                 13.1.3 Checking Optical Alignment—Position or test the
techniques used to sample, melt, and cast molten aluminum              position of the spectrometer exit slits, secondary mirrors (if
metal into a chill cast disk suitable for analysis, refer to           used) or refractor plates (if used), and photomultipliers to
Practices E716.                                                        ensure that the peak radiation passes through each slit and
                                                                   6
                                                                E1251 − 24
illuminates the centers of the phototubes. This shall be done by         faster and the gap should be checked more frequently and the
a trained expert initially and as often as necessary thereafter to       electrode may need replaced more frequently.
ensure proper alignment.                                                   13.6 Reference Material/Sample Placement—Reference
   13.1.4 Modern direct reading spectrometers should show                materials and samples should be placed on the spark stand so
little drift in the response channels with time. However, if at          that the hole in the top plate is completely covered. Completely
any time the gain adjustment of any channel drops below 0.5 or           covering the hole will prevent air leaks into the discharge area.
increases above 2, or if the background changes by more than             Air can cause “bad” burns and adversely affect precision and
0.5 to 2 times, that channel should be checked for alignment or          accuracy. The hole should be covered during idle periods for
deterioration of components.                                             the same reason. Samples and reference materials should be
  13.2 Electrical Parameters—Various sets of electrical pa-              sparked approximately 7 mm to 10 mm from their outer edge.
rameters in a rectified-capacitor discharge source produce               This can be best accomplished by placing them so that the
somewhat similar high-frequency oscillatory unidirectional               outer edge of the machined surface just covers the hole in the
waveforms. These have been found to produce comparable                   top plate. Overlapping the burns may adversely affect precision
analytical performance. Refer to 7.2 for typical parameters.             and accuracy.
                                                                            13.7 It is essential that operators learn the difference be-
  13.3 Spark Conditions—These conditions vary with the
                                                                         tween a “good” burn and a “bad” burn. Bad burns can be
manufacturer of the equipment and may require selection by
                                                                         caused by an air leak between the sample and the top plate, a
the user. A longer integration may result in better precision and
                                                                         poor quality sample, poor quality argon, and various other
accuracy with less sample through-put, while a shorter inte-
                                                                         reasons. A “good” burn will have a deeply pitted area in the
gration will increase sample through-put but may decrease
                                                                         center surrounded by a blackish ring. The actual appearance of
precision and accuracy. Typical time ranges are:
                                                                         a burn will vary with source conditions and alloy. A “bad” burn
  Flush period, s                        2   to   7                      will tend to have shallow pits surrounded by a white or silver
  Pre-burn period, s                     2   to   20
  Integration (spark) period, s          2   to   10                     colored ring. Usually the intensity of the aluminum internal
  Integration (arc) period, s            2   to   10                     standard channel for a “bad” burn will be considerably lower
  13.4 Gas Flow—Argon flow rate requirements may vary                    than a good burn. All “bad” burns should be rejected and
between manufacturers and between laboratories. The follow-              replaced.
ing ranges are presented as a guide.                                       13.8 Warm-Up—After any prolonged interval of instrument
  Standby, L/min                  0.03 to 0.5                            non-use, several warm-up burns should be taken. Generally
  During Integration, L/m         3.0 to 10                              two to four burns are sufficient to check for proper gas flow and
   13.4.1 The high-pressure compressed gas cylinder should               consistency of results.
be changed when the pressure falls below 7 kg/cm2 (100 kPa).
If the gas is supplied from a cryogenic tank, caution should be          14. Drift Correction
exercised so that the tank is not allowed to completely empty.              14.1 Need for Drift Correction—Spark Atomic emission
Consult with the gas supplier for their recommendation regard-           spectrometric analyses depend upon relative measurements
ing cryogenic tank changes. See Practice E406 for precautions            that are subject to drift over time. To correct for drift, a suite of
when handling gases.                                                     reference materials that includes both high and low mass
   13.5 Electrode System—The sample specimen serves as one               fractions of the elements is used to drift correct the intensities
electrode, the cathode. The tungsten or other suitable electrode         whenever a correction is required. Failure to routinely correct
serves as the counter electrode. Since the discharge is essen-           for instrument drift will adversely affect analysis results. The
tially unidirectional, the counter electrode is not attacked and         frequency for drift correction should be determined by statis-
therefore can be used for many burns. Because the electrode is           tical process control methods based on periodic measurement
moveable, continual adjustment of the analytical gap is re-              of a control sample.
quired. It is recommended that this gap be checked periodi-                 14.2 Drift Correction—Select a suite of reference materials
cally. The adjustment frequency is dependent on the number of            that will cover the analytical array and anticipated element
burns. Consult the manufacturer to determine the optimum                 mass fraction ranges of the instrument to be drift corrected. It
adjustment frequency for each instrument type. However,                  is highly recommended that the purchaser of a new instrument
material ablated from the sample surface tends to accumulate             specify that the appropriate drift correction materials be
on the tip of some types of electrodes, and can change the gap           included with the purchase of the spectrometer. If the instru-
and may adversely affect results. Therefore the counter elec-            ment comes pre-calibrated, then these materials should auto-
trode should be cleaned (brushed) with a wire brush that is              matically be included with the instrument. It is the responsi-
normally supplied with the instrument. For best performance it           bility of the purchaser to ensure that the correct drift correction
is strongly recommended that the counter electrode be cleaned            materials are included with the instrument. Follow the manu-
after every burn. Also, with continued use the shape of the              facturer’s instructions when drift correcting. The spectrom-
electrode may change due to this buildup of material. Frequent           eter’s software should have a program that will guide the
close inspection of the electrode is recommended. Some                   operator through the drift correction process. If the instrument
instruments use pin type electrodes that are not affected by an          is newly installed, give the unit sufficient time to stabilize in its
accumulation of ablated material. Pin electrodes tend to erode           new environment before proceeding with a drift correction. It
                                                                     7
                                                                   E1251 − 24
is recommended that the spectrometer be allowed to stabilize                when using third and fourth order regressions that enough
under vacuum (if so equipped) and to adapt to the final                     calibration materials are available to adequately cover the
controlled environment surroundings for at least two days                   entire range.
before a drift correction is performed. Note, the instrument                   15.3 Verifying the Accuracy of Calibration—After complet-
must be profiled before performing a drift correction. Refer to             ing a calibration, re-burn several of the calibration materials as
Practice E1329 for further details.                                         unknowns and compare the measured mass fractions for each
  14.3 Number of Burns—It is recommended that four burns                    element with the certified values. Check for clerical errors,
be taken on each of the drift correction materials during the               elemental interferences, or biases if results do not compare
drift correction process.                                                   favorably.
  14.4 Checking Homogeneity of Candidate Drift Correction                      15.3.1 If individual calibration materials give consistently
Materials—If the homogeneity of the material(s) being used is               high readings for an element, check for possible interferences
questionable; the material(s) can be tested for homogeneity.                from other elements. Manually calculate or, using the instru-
                                                                            ment’s software, have the software calculate the appropriate
   14.5 Recording the Drift Correction Readings:                            factors for the interference(s).
   14.5.1 Instruments that come pre-calibrated will have the
initial drift corrected response factors entered into the instru-           16. Procedure for Analyzing Specimens
ment’s software.                                                              16.1 Excitation—Burn the specimens using the parameters
   14.5.2 If the instrument does not come pre-calibrated, then              and conditions in 13.2, 13.3, 13.4, and 13.5.
follow the manufacturer’s instructions regarding establishing
the initial drift correction responses/factors. Initial drift correc-          16.2 Replicate Burns—Burn the specimens from one to
tion responses should be established immediately after calibra-             eight times, depending on the complexity of the alloy, speci-
tion.                                                                       men homogeneity, and the level of confidence required. A
   14.5.3 If one of the drift correction materials must be                  single burn is frequently employed for qualitative analysis of
replaced because it has become unusable (too thin), follow the              primary aluminum to detect major changes in the composition
instructions listed in the instrument’s manual regarding the                of aluminum from individual Hall cells to assess performance.
replacement and recording of the new drift correction materi-               A single burn shall not be used for determining compliance
al’s responses. Failure to properly replace drift correction                with composition specifications. Two to four burns are recom-
materials may adversely affect analysis accuracy.                           mended for most alloys where homogeneity is acceptable and
                                                                            accuracy becomes important. In very complex alloys or in
15. Calibration                                                             alloy systems that are noted for their segregation, additional
   15.1 Obtaining Calibration Data—The following procedure                  burns may be required.
is designed to allow the user to collect accurate data for the                 16.2.1 The determinations from all burns should be aver-
purpose of generating calibration curves. For details on estab-             aged unless a burn produces a very abnormal internal standard
lishing and controlling spectrochemical calibration curves,                 intensity or appears visually to be bad (see 13.7). When a burn
refer to Practice E305. Any recently installed, laboratory-grade            is rejected, it should be replaced in order to maintain the
spectrometer should show minimal drift over an 8-h to 24-h                  normal number of burns to be averaged.
time period when placed in a laboratory with a carefully
                                                                            17. Calculation of Results
controlled environment.
   15.1.1 Select the reference materials for use as the calibra-              17.1 After performing the test material analyses, the instru-
tion materials.                                                             ment software will calculate the mass fractions of the elements
   15.1.2 Follow the manufacturer’s operating manual and use                based on the calibration curves. Further manipulation of the
the instrument’s software to design, and name the analytical                data should not be necessary.
program. Using the software, enter the identities of the selected             17.2 Rounding of test results obtained using this test method
calibration materials and their associated mass fractions for the           shall be performed as directed in Practice E29, Rounding
elements to be calibrated in this calibration.                              Method, unless an alternative rounding method is specified by
   15.1.3 Before starting the collection of calibration data,               the customer or applicable material specification.
thoroughly clean the excitation chamber and check the analyti-
cal gap or replace the electrode as needed. Prepare fresh                   18. Data Reporting
surfaces on the selected calibration materials.                                18.1 Number of Significant Figures—The composition of
   15.1.4 Profile the instrument.                                           alloys shall not be reported with more significant figures or
   15.1.5 Burn the calibration materials and collect the data.              higher precision than that of the materials used to calibrate the
   15.2 Refer to Practice E305 and calibrate the instrument                 spectrometer. As a general guideline, labs should report results
using the instrument’s software, following the instructions in              at least to the number of decimal places given in ANSI
the manufacturer’s manual. Use the appropriate program that                 H35.1/H35.1M if possible. Footnote 3 in Section 2 of ANSI
allows for the calculation of the calibration curves. Take care             H35.1/H35.1M provides the following:
                                                                        8
                                                                 E1251 − 24
                                                 TABLE 3 Summary of Interlaboratory Tests
                               Mass Fraction
    Test Sample                                                     ST                SR        R         2R
                   Certified               Observed
Antimony
  WA1199           0.0013                       0.0012           0.00041           0.00041    0.0011    0.0022
  SS1075           0.0033                       0.0036           0.00024           0.00024    0.00067   0.0013
Arsenic
  L1               0.0014                       0.0012           0.00021           0.00021    0.00059   0.0012
  L5               0.0064                       0.0055           0.00083           0.00083    0.0023    0.0046
Beryllium
  3004-3           0.0004(4)                    0.00046          0.000010          0.000040   0.00011   0.0002
  5056             0.004                        0.0037           0.000011          0.00037    0.0010    0.002
  358              0.24                         0.237            0.0021            0.0034     0.0095    0.02
Bismuth
  1000             0.034                        0.0335           0.0005            0.0010     0.0029    0.006
  2011             0.56                         0.567            0.0048            0.021      0.058     0.12
Boron
  1075             0.0006                       0.00055          0.000024          0.00006    0.00016   0.0003
  18               0.009                        0.0107           0.00026           0.0012     0.0033    0.007
Calcium
  3004-3           0.0002(5)                    0.00027          0.000012          0.000072   0.0002    0.0004A
Chromium
  3003             0.001                        0.00146          0.00009           0.0023     0.0064    0.013
  1000             0.036                        0.0347           0.00025           0.0033     0.0092    0.018
  5056             0.12                         0.1208           0.00093           0.0050     0.0140    0.028
  7075             0.23                         0.2264           0.0036            0.0067     0.0189    0.038
Cobalt
  7091             0.44                         0.4423           0.0054            0.0054     0.015     0.030
Copper
  1075             0.001                        0.00084          0.00014           0.00061    0.0017    0.0034
  1000             0.030                        0.0306           0.00027           0.00123    0.0035    0.007
  3003             0.15                         0.154            0.00074           0.00435    0.0122    0.024
  7075             1.58                         1.564            0.0179            0.0388     0.109     0.22
  2011             5.49                         5.492            0.0231            0.0538     0.151     0.30
Gallium
  1075             0.022                        0.0219           0.00030           0.00053    0.0015    0.0030
Iron
  5056             0.20                         0.195            0.00532           0.01655    0.046     0.09
  2011             0.53                         0.530            0.00264           0.01275    0.036     0.07
Lead
  1000             0.036                        0.035            0.0003            0.0009     0.003     0.005
  2011             0.56                         0.580            0.0060            0.0461     0.129     0.26
Lithium
  3004-3           0.0003(4)                    0.00034          0.000006          0.000025   0.00007   0.00014
  2090             2.14                         2.13             0.023             0.041      0.114     0.23
Magnesium
  1000             0.030                        0.0317           0.00044           0.00158    0.0044    0.009
  356              0.36                         0.354            0.00426           0.00860    0.024     0.05
  7075             2.61                         2.596            0.0219            0.0283     0.079     0.16
  5056             5.36                         5.364            0.0400            0.0441     0.123     0.25
Manganese
  1075             0.001                        0.00102          0.00008           0.00024    0.001     0.013A
  1000             0.032                        0.0316           0.00022           0.00131    0.0037    0.007
  5056             0.10                         0.1006           0.00044           0.0039     0.0109    0.022
  3003             1.21                         1.208            0.00850           0.0139     0.039     0.08
Nickel
  3003             0.005                        0.00474          0.00007           0.00028    0.0008    0.0016
  1000             0.031                        0.0299           0.00028           0.00177    0.0050    0.0099
  850              1.21                         1.219            0.0044            0.0082     0.023     0.048
  336              2.60                         2.596            0.0216            0.0247     0.069     0.139
Phosphorus
  AP-4             0.003                        0.00269          0.00016           0.00028    0.0008    0.0016
Silicon
  1075             0.068                        0.0699           0.0004            0.00266    0.0074    0.015
  2011             0.28                         0.288            0.00194           0.00784    0.022     0.044
  356              7.18                         7.16             0.0414            0.0811     0.23      0.45
  390A            16.36                        16.29             0.13              0.242      0.68      1.35
Sodium
  3004-3           0.0002(9)                    0.00026          0.000011          0.000017   0.00005   0.00010
  18               0.021                        0.024            0.0016            0.0036     0.010     0.020
StrontiumB
  336              0.027                        0.0271           0.0005            0.0008     0.00229   0.00458
  356              0.028                        0.0277           0.0003            0.0016     0.00435   0.00870
Tin
  1000             0.028                        0.0287           0.0002            0.0011     0.003     0.006
Titanium
  1075             0.001                        0.00083          0.000031          0.000341   0.0010    0.0019A
                                                                    9
                                                                                   E1251 − 24
                                                                          TABLE 3       Continued
                                             Mass Fraction
     Test Sample                                                                      ST                       SR                        R                        2R
                                 Certified               Observed
 2011                             0.003                      0.00321               0.000041                 0.000185                  0.0005                    0.0010A
 1000                             0.031                      0.0318                0.000161                 0.000799                  0.0022                    0.0045
 356                              0.12                       0.1176                0.00058                  0.00353                   0.010                     0.020
Vanadium
 1075                             0.002                      0.00195               0.000055                 0.00016                   0.0005                    0.0009
 1000                             0.022                      0.0218                0.000237                 0.00058                   0.0016                    0.0033
Zinc
 1075                             0.002                      0.00229               0.00040                  0.00069                   0.0019                    0.004
 1000                             0.030                      0.0302                0.00065                  0.00071                   0.0020                    0.004
 356                              0.01                       0.1001                0.00059                  0.00111                   0.0031                    0.006
 7075                             5.74                       5.741                 0.0285                   0.0927                    0.26                      0.52
Zirconium
 1075                             0.001                      0.00093               0.00007                  0.00017                   0.0005                    0.0009
 2090                             0.12                       0.120                 0.012                    0.0023                    0.006                     0.013
A
 Values are below minimum quantifiable limit calculated based on the data in this ILS.
B
 Supporting data for strontium have been filed at ASTM International Headquarters and may be obtained by requesting Research Report RR:E01-1113.
Standard limits for alloying elements and impurities are expressed to the                  19. Precision and Bias5
following places:
                                                                                              19.1 Precision—Eight laboratories cooperated in the inter-
Less than 0.001 %                                        0.000X                            laboratory study using either the binary calibration approach or
                                                                                           the alloy-type calibration technique. Since an attempt was
0.001 but less than 0.01 %                               0.00X
                                                                                           made to include all general alloy types, not all laboratories
0.01 but less than 0.10 % unalloyed aluminum             0.0XX                             could analyze all materials or all mass fraction ranges. Testing
made by a refining process
                                                                                           was done in accordance with Practice E691. A summary of the
Alloys and unalloyed aluminum not made by a              0.0X                              inter-laboratory test is shown in Table 3.
refining process
                                                                                              19.2 Bias—There is no evidence of bias since all acceptable
0.10 through 0.55 percent                                0.XX                              individual test results are within one R of the assumed mass
(It is customary to express limits of 0.30 % through                                       fractions.
0.55 % as 0.X0 or 0.X5)
                                                                                              5
                                                                                                Supporting data have been filed at ASTM International Headquarters and may
                                                                                           be obtained by requesting Research Report RR:E01-1113. Contact ASTM Customer
                                                                                           Service at www.astm.org/contact.
                       ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
                   in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
                   of infringement of such rights, are entirely their own responsibility.
                       This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
                   if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
                   and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
                   responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
                   make your views known to the ASTM Committee on Standards, at the address shown below.
                      This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
                   United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
                   address or at 610-832-9585 (phone), or through the ASTM website (www.astm.org/contact). Permission rights to photocopy the
                   standard may also be secured from the Copyright Clearance Center, 222 Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600;
                   http://www.copyright.com/
10