The document provides technical data and safety information for an X-ray tube insert, including specifications such as voltage, focal spot size, and anode material. It emphasizes the importance of safety precautions during installation and use, particularly regarding radiation exposure and environmental hazards. Additionally, it outlines compliance requirements for incident reporting under the Medical Devices Directive 93/42/EEC.

Uploaded by

arashcpi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
37 views13 pages

OX 115 05L Rev.i Eng

The document provides technical data and safety information for an X-ray tube insert, including specifications such as voltage, focal spot size, and anode material. It emphasizes the importance of safety precautions during installation and use, particularly regarding radiation exposure and environmental hazards. Additionally, it outlines compliance requirements for incident reporting under the Medical Devices Directive 93/42/EEC.

Uploaded by

arashcpi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 13

OX/115-05L

TECHNICAL DATA GENERAL INFORMATIONS


When mounting tube inserts adopt proper caution, in order to avoid glass bulb breaking
and fragments projection. Please use protective gloves and glasses.
X-ray tube voltage Tube insert connected to H.V. supply is a radiation source: be sure to take all necessary
nominal 120 kV IEC 60613:2010 safety cautions
overvoltage test 132 kV IEC 60601-2-28:2020
Nominal focal spot 0,5 IEC 60336:2020 ➢ Wash thoroughly with alcohol the external surface of tube insert (care of fire risk).
Remove any labels on the glass bulb. Avoid contact of dirty surfaces with cleaned tube
1,5  3,5 V
Filament characteristics insert.
3,0  4,0 A
Anode material Tungsten ➢ Clamp system inside housing or self-contained units must not mechanically stress the
Target angle 15° tube.
Anode heat storage capacity 30000 J
➢ After installation, check the right working of the tube (no fluctuation of tube current
Continuous anode input power 270 W IEC 60613:2010 nor crackling)
Nominal anode input power
1300,0 kW IEC 60613:2010
0.1 s – DC ➢ Comply with insert thermal parameters, planning and programming the exposure
Permanent filtration 1,1 mm Al IEC 60522-1:2020 parameters and cooling pauses. Housing or self-contained units must be provided with
Maximum diameter 50 mm an adequate thermic protection.
Maximum overall length 150 mm
➢ Voltages indicated in charts are valid for transformer supplied with ground center.
Weight (without lead shield) 370 g
➢ Tube inserts contain environment polluting materials, particularly lead liner tubes.
*With std shank Please apply to qualified operator for waste disposal, according to local regulation
requirements.

INCIDENT REPORT ACCORDING TO 93/42/EEC MEDICAL DEVICES DIRECTIVE


In order to comply with CE marking requirements, end users have to report to local Competent
Authority all the informations about possible incidents involving the device, regarding any deterioration
in its characteristics and performances, as well as any inaccuracies in this documentation, which might
lead to or might have led to the death of patient / user or a deterioration in his state of health. This
information must be promptly reported also to C.E.I. in order to start manufacturer reporting, as per
above mentioned directive.

this mark assures device conformity to


EC Directive 93/42 on Medical Devices Safety.
0051
0051

Rev.I - 23 ott 2023 - Original text: italian Page 1 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
DIMENSIONS STD SHANK
WEIGHT : 370 g CODE: 2050 D0 A0 XX XX

Rev.I - 23 ott 2023 - Original text: italian Page 2 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
DIMENSIONS STD SHANK WITH LEAD SHIELD
WEIGHT: 880 g CODE: 2050 D0 E0 XX XX

Rev.I - 23 ott 2023 - Original text: italian Page 3 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
THERMAL CURVES STD SHANK

Anode heat (J)

Time (s)

Rev.I - 23 ott 2023 - Original text: italian Page 4 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
DIMENSIONS LONG SHANK
WEIGHT : 390 g CODE: 2050 B0 A0 XX XX

Rev.I - 23 ott 2023 - Original text: italian Page 5 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
DIMENSIONS LONG SHANK WITH LEAD SHIELD
WEIGHT: 900 g CODE: 2050 B0 E0 XX XX

Rev.I - 23 ott 2023 - Original text: italian Page 6 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
THERMAL CURVES LONG SHANK

Anode heat (J)

Time (s)

Rev.I - 23 ott 2023 - Original text: italian Page 7 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
DIMENSIONS RADIATOR
WEIGHT : 490 g CODE: 2050 H0 A0 XX XX

Rev.I - 23 ott 2023 - Original text: italian Page 8 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
DIMENSIONS RADIATOR WITH LEAD SHIELD
WEIGHT: 1000 g CODE: 2050 H0 E0 XX XX

Rev.I - 23 ott 2023 - Original text: italian Page 9 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
THERMAL CURVES RADIATOR

Anode heat (J)

Time (s)

Rev.I - 23 ott 2023 - Original text: italian Page 10 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
FILAMENT CHARACTERISTICS EMISSION CHARACTERISTICS DC
Filament voltage (V)

Tube current (mA)


Filament current (A)
Filament current (A)

Rev.I - 23 ott 2023 - Original text: italian Page 11 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
RATING CHARTS DC

Tube current (mA)

Exposure time (s)

Note: This power rating has been verified in a large oil tank. Power rating may vary depending upon the oil capacity as well as the oil circulation in the testing tank.

Rev.I - 23 ott 2023 - Original text: italian Page 12 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/115-05L
INFORMATION:
Even if the tube is shipped after a long seasoning and testing cycle, it is always better to perform a quick seasoning before the first exposure.
Please start the test cycle with the lowest kV possible in your machine, and increase the kV gradually with an incremental step of 5kV. For each
step, perform at least 3 exposures maintaining the current parameter between 1mA and 2mA. A gap of few seconds between each of the
exposures is advised. Increase the kV to the max kV as rated for the respective tube. In case of instability observed, reduce kV by one step.
Once the kV reaches the nominal kV allowed for the respective tube, perform 3 exposures at the nominal power as requested for the concerned
application.

Full seasoning is not required, when installing the machine. Nevertheless, if the tube head has been stored for more than 6 months, it is advised
to perform few exposures at low kV and low power, before sending the machine to the final customer.

Sometimes a gentle metallic noise is audible from inside the tube. This noise is not affecting tube performances and the use of such tubes is
totally safe.

Rev.I - 23 ott 2023 - Original text: italian Page 13 of 13 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy