Re Bath
Re Bath
Instruction Manual
  STU0001 / Version 1.3
    Introduction
    Thank you for purchasing this Stuart product. To get the best performance from the equipment, and for
    your own safety, please read these instructions carefully before use.
Before discarding the packaging check that all parts are present and correct.
1 2
     1 RE400DB or RE400OB
    		(RE400DB shown)
2 UK power lead
5 Instruction manual
        6   Spare fuses x2
                                       6
    		(RE400DB only)
    If the equipment is not used in the manner described in this manual and with accessories other than those
    recommended by the manufacturer, the protection provided may be impaired.
2
General Description
Your Rotary Evaporator Bath has been designed to be used with the Stuart range of Rotary Evaporators. They
are designed for general purpose laboratory use in which samples are gently heated in a Florentine flask to cause
evaporation during rotation.
❖ SHOCK HAZARDS OR UNSAFE PRACTICES ARE DANGEROUS as they can cause severe personal injury, fire
  or death.
❖ DO NOT use combustible substances near hot objects.
❖ DO NOT use the equipment in hazardous atmospheres.
❖ DO NOT operate or handle any part of the product with wet hands or use on surfaces that may become flooded.
❖ NEVER move the product while still connected to the power supply.
❖ HIGH TEMPERATURES ARE DANGEROUS as they can cause serious burns to operators and ignite combustible
  material.
❖ USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS.
❖ NEVER lift or carry the instrument until it has been switched off and allowed to cool.
v DO NOT position the unit so that it is difficult to disconnect from the mains supply using the mains plug.
v The mains outlet socket used should be located close to the equipment and readily identifiable and accessible
  to users.
❖ DO NOT leave equipment switched on and it is not recommended to leave any heating apparatus unattended
  during operation.
❖ The unit should be carried using both hands.
Symbols Defined
Electrical Requirements
Before connection please ensure that the line supply corresponds to the power requirements below:
		      RE400DB          1400 W           230 V a.c. ~ 50/60 Hz
		      RE400OB          1400 W           230 V a.c. ~ 50/60 Hz
The unit is supplied with two mains leads fitted with IEC plugs for connection to the instrument. One lead has a UK
3 pin plug and the other has a 2 pin “Schuko” plug for connection to the mains supply. Choose the lead appropriate
for your electrical installation and discard the other. Should neither lead be suitable take the lead with the UK plug
and replace the plug with a suitable alternative. This involves cutting off the moulded plug, preparing the cable and
connecting to the rewireable plug in accordance with its instructions.
                                                                                                                         3
    NOTE: Refer to the equipment rating plate to ensure that the plug and fusing are suitable for the voltage and
          wattage stated. The wires in the mains cable are as follows:
            BROWN - HOT/LIVE
            BLUE - NEUTRAL
            GREEN/YELLOW – EARTH
    The appropriate mains lead should be connected to the instrument BEFORE connection to the mains supply.
    Should the mains lead require replacement, a cable of 1.25mm² (AWG16) of harmonised code H05VV-F
    connected to an IEC320 plug should be used.
    Before Use
    Place the Rotary Evaporator Bath on a firm, level surface.
              Caution: The Bath may be HOT. As a precaution for the user, take care when handling.
              NOTE: If the unit has been used recently the bath may be too hot for your sample. If this is the case,
              press the stop button and allow the unit to cool before proceeding.
4
Operation
RE400DB Overview
Main View
1 Bath
    2   Bath cover                                                                                             6
    3   Control panel
                                            2
    4   Power switch
5 Bath base
    6   Handle                                                                                                 5
                                            3
Rear View
9 Service connector
7 8 9
RE400DB
			WARNING: DO NOT use the RE400DB with any liquid other than water.
		 May cause fire on failure.
Fill the Teflon coated bowl to the required level with cold water. To avoid scale build up and the necessity for
cleaning it is recommended that the bowl be filled with distilled or deionised water.
The maximum capacity of the bath is 3 litres and it should not be filled to within less than 2cm from the top of the
bowl.
During use water will be lost from the bath by evaporation. For maximum efficiency water should be added
periodically and the level maintained.
                                                                                                                       5
    RE400OB Overview
Main View
1 Bath
2 Bath cover
3 Control panel
4 Power Switch 2
5 Bath Base
                                                                                                                 5
                                                 3
Rear View
8 Service connector
6 7 8
RE400OB
Fill the Teflon coated bowl to the required level with either oil or water.
    The maximum capacity of the bath is 3 litres and it should not be filled to within less than 2cm from the top of the
    bowl.
    If being used with water, during use water will be lost from the bath by evaporation. For maximum efficiency
    water should be added periodically and the level maintained.
6
RE400DB and RE400OB Control Panel
          1                                                                     6
                                                                                7
2 3 4 5
3 Down key - Press to decrease temperature by 1ºC. Press and hold to continually decrease value.
4 Up key - Press to increase temperature by 1ºC. Press and hold to continually increase value.
❖ To set the bath temperature press the Set key once, this enables temperature input mode, during this mode
  the display blinks. Use the       and       keys to set the desired bath temperature. Each press of the
        or      key will increase or decrease the set temperature by 1ºC. Press and hold the       or
  key to continually increase or decrease the value.
❖ Press the Set key to return to measuring mode or after 10 seconds without a key press the display will go back
  to measurement mode, press Set key to re-enter input mode.
❖ Pressing the Run/Stop key starts temperature control at the set temperature.
❖ Be sure to press the Run/Stop key to stop temperature control before turning the power switch off. If the
  power is turned OFF during Run state, the heater will turn on at the same time the unit is next turned on and
  an accident could occur.
                                                                                                                   7
    Setting the alarm function - RE400DB and RE400OB
You can set the upper temperature limit alarm or the power outage recovery function .
    ❖ Hold the Set key for 5 seconds to enter alarm function setting mode,              is displayed. press the Up or
      Down key to switch between the different alarm functions.
❖ When you have set the required alarm function, hold the Set key for 5 seconds to return to the measurement mode.
    ❖ Press the       and       key to set the upper temperature limit. Each press of the       or      key will
      increase or decrease the set temperature by 1ºC. Press and hold the        or       key to continually
      increase or decrease the value. Upper temperature limit setting range +10ºC~200ºC. The factory setting is 200ºC.
    ❖ Press the Set key to return to alarm function setting mode, the setting of the upper temperature limit alarm is
      completed,          will be displayed.
❖ Hold the Set key for 5 seconds to return to the measurement mode.
    You can choose whether temperature control will be resumed when power is turned on again after a power
    outage by setting the power outage recovery function to    or     .
, temperature control will resume after the power is restored in the state immediately before the recovery.
, temperature control will stop irrespective of the state before the recovery.
    ❖ Press the Set key to return to alarm function setting mode, the setting of the power outage recovery
      function is completed,          will be displayed.
❖ Hold the Set key for 5 seconds to return to the measurement mode.
When the overheat protector has been activated the power switch is ON but the display is OFF.
    ❖ Confirm why the overheat protector has been activated (eg bath was heated without oil/water or external
      temperature exceeds 35°C) and make sure that the inside of the bath has sufficiently cooled down and
      press the ‘RESET’ button on the rear of the unit.
❖ Turn the power switch ON. The display comes on and temperature control starts. Run light is on.
If it is not possible to reset the overheat protector switch stop operation and contact a local service representative.
8
Maintenance, Servicing and Cleaning
                     HOT: Before attempting any maintenance, servicing or cleaning, ensure that the unit is cool,
                     and disconnect from the power supply.
                     WARNING: Ensure the unit is disconnected from the power supply before attempting any
                     maintenance, servicing or cleaning.
This product range does not require any routine servicing and there are no servicible parts within the equipment.
The only user maintenance required is the inspection of the power supply unit and mains power lead set.
With proper care and operation, the equipment should give reliable service, however contamination or general
misuse may reduce the effective life of the product and could cause a hazard.
Where used with water, periodically de-scale the bath using a domestic kettle descaler suitable for metal kettles.
Follow the instructions for use supplied with the descaler.
The water bath should be thoroughly rinsed before returning to use. Alternatively use a mild detergent to clean
the bath. Take care not to damage the Teflon coating during cleaning.
WARNING: Observe any cautionary note on the cleaner regarding protective clothing and effluent control.
The exterior of the unit should be cleaned using damp cloth and a mild detergent solution.
Preventative maintenance should include keeping the product clean by protecting it from spillage, contamination
or corrosive environments. If in doubt, please confirm that any intended method of decontamination will not
damage the equipment by contacting Cole-Parmer.
NOTE: Do not use solvents for cleaning any parts of this equipment.
If any part of the equipment other than the bath has been exposed to liquid, it cannot be assumed to meet all
the safety requirements of EN 61010-2-010 until the drying out process has been fully completed and all safety
requirements are met before the unit is used again.
In Case of Contamination
If the equipment has been exposed to contamination, the Responsible Body is responsible for carrying out
appropriate decontamination. If hazardous material has been spilt on or inside the equipment, decontamination
should only be undertaken under the control of the Responsible Body with due recognition of possible hazards.
Before using any cleaning or decontamination method, the Responsible Body should check with the manufacturer
that the proposed method will not damage the equipment. Prior to further use, the Responsible Body shall check
the electrical safety of the unit. Only if all safety requirements are met can the unit be used again.
NOTE: In the event of this equipment or any part of the unit becoming damaged or requiring service, the item(s)
should be returned to the manufacturer for repair accompanied by a decontamination certificate. Copies of the
Certificate are available from the Distributor/Manufacturer.
At the end of its service life, the product must be accompanied by a Decontamination Certificate.
                                                                                                                     9
 Fuse Replacement
 The mains fuse holder is located at the rear of your product. Refer to Electrical Specification (page 11) for correct
 fuse type and rating. Turn your unit off and disconnect it from the power supply. Always replace fuses with the
 correct type and rating.
     RE400DB                                                                 RE400OB
1 2 3 1 2
 Any repairs or replacement of parts MUST be undertaken by suitably qualified personnel. Only spare parts supplied
 or specified by Cole-Parmer or its agents should be used. Fitting of non-approved parts may affect the performance
 and safety features designed into the instrument. For a comprehensive list of parts required by service engineers
 conducting internal repairs please contact the service department quoting the model and serial number:
 Email: cpservice@coleparmer.com
 Tel: +44 (0)1785 810475
 For any other technical enquiries please contact the Technical Support Department at;
 Email: cptechsupport@coleparmer.com
 Tel: +44 (0)1785 810433
Warranty
 Cole-Parmer Ltd. warrants this instrument to be free from defects in material and workmanship, when used under
 normal laboratory conditions, for a period of 3 years. In the event of a justified claim Cole-Parmer will replace any
 defective component or replace the unit free of charge. This warranty does NOT apply if damage is caused by fire,
 accident, misuse, neglect, incorrect adjustment or repair, damage caused by incorrect installation, adaptation,
 modification, fitting of non-approved parts or repair by unauthorised personnel.
 Cole-Parmer Ltd,
 Beacon Road,
 Stone,
 Staffordshire,
 ST15 0SA,
 United Kingdom
 Email: cpservice@coleparmer.com
 Tel: +44 (0)1785 810475
 Web: www.stuart-equipment.com
10
Spares and Accessories 		
    Part No.                Description                                                      Quantity
    HH179(S)                UK mains lead moulded plug 230 V - 10A BS1363/A                     1
    HH180(S)                European mains lead moulded plug 230 V                              1
Replacement fuses can be ordered by contacting cpspares@coleparmer.com. Please refer to Electrical Specification
(page 11) for correct fuse type and rating.
Technical Specification
  Bath cover                                          PET (with glass fiber)                 PBT (with glass fiber)
  Bath material                                             SUS304                         Aluminium, teflon coating
  Bath capacity                                        Approx. 4.3 litres                       Approx. 5 litres
  Size of bath (internal)                              Ø220 x H120mm                           Ø240 x H120mm
  Max. fill capacity                                                            3 litres
  External dimensions                              W285 x D295 x H244mm                    W282 x D282 x H244mm
  Weight                                                     3.9kg                                   4.5kg
 Display resolution                                                              1°C
                                                                                           ± 3°C (during rotation) Oil
 Accuracy                                            ± 1°C (during rotation)
                                                                                    ± 1.5°C (during rotation) Water
 Ramp rate                                                               Increments of 1°C
 Display                                                                    Digital Display
 Control Sensor                                                                PT sensor
                                                                                                                         11
 Weights and Dimensions
     RE400DB
295mm 285mm
                           244                     244
                           mm                      mm
Weight 3.9kg
RE400OB
282mm 282mm
                          244                      244
                          mm                       mm
Weight 4.5kg
12
Fault Finding
Problem                    Cause                                         Solution
Display does not work      No mains power.                               Check mains power supply.
even if the power is
switched on.                                                             Replace the fuse, if fuse blows again
                           Fuse has blown.                               immediately stop operation and contact a
                                                                         local service representative.
SSR is malfunctioning
                                                                                                                      13
     Lower temperature      Ambient temperature is lower than 0°C
                                                                     Adjust the ambient temperature to
     limit alarm works.
                                                                     between 5°C and 35°C.
     Display shows      .   Measured temperature is lower than 0°C
14
              This product meets the applicable                              and so we cannot guarantee that interference will not
              harmonized standards for radio frequency                       occur in practice. Where there is a possibility that injury,
              interference and may be expected not to                        damage or loss might occur if equipment malfunctions
interfere with, or be affected by, other equipment with                      due to radio frequency interference, or for general
similar qualifications. We cannot be sure that other                          advice before use, contact the manufacturer.
equipment used in its vicinity will meet these standards
                                               EU Declaration of Conformity
                       Product                Laboratory Equipment              File Number            P225
This declaration of conformity is issued under the sole responsibility of the manufacturer
                The object of the declaration described above is in conformity with the relevant Union Harmonisation Legislation:
                                      Low Voltage Directive                2014/35/EU
                                      EMC Directive                        2014/30/EU
                                      RoHS Directive                       2011/65/EC
                References to the relevant harmonised standards used or references to the other technical specifications in relation
                to which conformity is declared: