Nikon Labophot2 Pol Instructions
Nikon Labophot2 Pol Instructions
CAUTIONS
DAvoid Strong Shocks! (;]When the Lamp Is On
Handle the microscope gently. taking Take care not to touch the lamp housing
care to avoid strong shocks. or bring Inflammable substances such as
gasoline. thinner and alcohol near It, as
fJWhen Carrying the Microscope
some parts of the lamp hOUSing become
When carrying the microscope, support
very hot.
the bottom of the microscope base. The
Instrument weighs about 7. 9kg. Do not mChanging the Lamp-bulb
hold the overhanging portion of the base. Before replaCing the Lamp~bulb, turn off
the main switch and disconnect the power
IiJPurpose
source plug. Before replaCing the
Use the microscope only for microscopIc
halogen lamp (6V~30W), be sure that It is
observations. Do not use It for any other
cool enough Do not touch the glass part
applications.
With your bare hands.
mPlace of Use mJDirt on the Lens
Avoid exposing the microscope to place
Do not leave dust, dirt or finger marks on
where it may be subject to dust,
the lens or bulb surfaces. They will
vibrations, high temperatures, mOisture
prevent you from obserVing the specimen
or direct sunlight.
clearly.
mlnstallation IDFocus Knobs
I nstall the microscope apart from wall In
Never attempt to adjust the tightness of
such a way that the lamp unit warning
the rlght~ and left~hand focus knobs by
label IS vIsible.
turning one while holding the other. It
G)lnput Voltage may cause problems. Avoid forCing the
Confirm that the Input voltage to the mlcro~ coarse focus knobs to turn past their
scope corresponds to your line voltage. limits since that may cause problems.
BLight Source
Use a 6V~30W halogen lamp Do not
use lamps other than the one specified on
p.41 (Electrical Specifications). If a lamp
of more than the suggested wattage IS
used, the light adjusting circuit may be
damaged.
CONTENTS
I .NOMENCLATURE ························1 10. Use of Field Diaphragm .............. ·24
II.ASSEMBLY····································4 1 1 . Optical Path Change· Over ........ · .. ·24
III. MiCROSCOPy··························· 11 12 . Interpupillary DistanceAdjustment·25
N .OPERATIONS IN DETAIL ......... 17 13. Diopter Adjustment ..................... 25
1 . Orientation of the Dia·polarizer ...... 17 14. Oil Immersion Manipulation ........ ·26
2. Focusing and Centering the 15. Brightness and Contrast
Bertrand Lens' ............................. 18 Adjustment ............................... 27
3. SWing Out the Top Lens of the 16. Objectives and Eyepieces ............ 29
Condenser ................................. .. ·19 V .OPTIONAL ACCESSORIES ...... 30
4.1 !4A & Tint Plate ........................ 19 1 . Senarmont Compensator .............. ·30
5. Centering the Objectives ............... 20 2. Quartz Wedge .............................. 31
6. Stage Rotation .... · .. ...................... ·20 3. Pin Hole Eyepiece ........................ 32
7. Focusing .................................... 21 4. Attachable Mechanical Stage" ....... 32
8. Centering the Condenser ............... 22 5. Universal Epi-illuminator ............... 33
9. Use of Condenser Aperture VI.TROUBLESHOOTING ............... 35
Diaphragm ................................. 23 ELECTRICAL SPECIFICATIONS ... 41
..
II
I • NOMENCLATURE
I . NOMENCLATURE
Eyepiece tube
Intermediate tube
clamp screw
Nosepiece centering
Bertrand lens
screw
centering screw
Centering revolving
nosepiece
Analyzer knob
CF Achromat P objective
Intermediate tube
(Stra in -free)
clamp screw
Revolving nosepiece
Specimen clip
clamp screw
Circular graduated
Coarse-focus torque
stage
adjustment ring
Power switch
Brightness adjuster
Fig.1-1
1
II. ASSEMBLY
II. ASSEMBLY
%@
a®---...6
t
~CV
@
~Ptionall Fig.2
Eyepiece
Diopter ring
Bertrand lens
centering screw Bertrand lens turret
Stage rotation
clamp screw
Coarse focus knob
Condenser top lens
swing~out knob
Fine focus knob
Power cord
Condenser focus
knob
(Optional)
Condenser clamp
screw
N D filter knob
Fig.1-2
2
II. ASSEMBLY
Fig.2-1
(2) Lamp and Lamp Housing
I CAUTIONS I
• Use the specified lamp "Nikon Halogen 6V 30W".
e Use the specified Halogen Lamp. See Electrical Specifications
on p.4 1 tor details .
• Keep the lamp cover on or use gloves when mounting the lamp
bulb, so as not to touch the surface directly with your fingers.
If the bulb surface IS contaminated with finger marks or dirt,
wipe them off with lens tissue. Be sure to remove the lamp
cover after mounting the lamp.
• Remove the lamp housing cover pressing the cover detaching
buttons (Fig. 2-2).
elnsert the lamp Into the socket pin holes until it reaches the limit
(Fig. 2-3).
• Reattach the cover.
Cautions when dismounting
• Turn oft the power switch. (Assure that It is off.) Do not touch the
lamp-bulb Immediately after turning It off because It is very hot.
Wait until it cools enough for replacing.
• Power supply cord should be kept away trom lamp housing while the
lamp hOUSing is hot.
Fig.2-2 Fig.2-3
5
II. ASSEMBLY
Q) Objectives
• Screw in the objectives to the holes of the revolving nosepiece in
such positions that, when viewed from above, their magnifying
power increases as the nosepiece is revolved clockwise (Fig. 2-4).
Cautions when dismounting I
• Remove the NO filter cassette (Ex.)
If attached and lower the stage
by turning the coarse focus
knob.
• Remove any specimen from
the stage.
• Hold the objective so as not to
drop it when unscrewing.
Fig.2-4
® Stage
• Lower the substage by turning the coarse focus knob.
• Loosen the stage clamp screw sufficiently.
• Fit the stage to the circular dovetail of the substage 00. Fasten the
stage with the stage clamp screw ~ (Fig. 2-5).
Fig.2-5 Fig.2-6
@ Specimen Clip
.Insert the two speciment clips into the holes on the stage surface
(Fig. 2-6).
® Condenser
• Raise the substage by turning the coarse focus knob.
• Lower the condenser carrier to its limit by turning the condenser
focus knob .
• Insert the condenser into the condenser carrier 00 with the numerical
aperture plate facing toward the user. Fasten it with the condenser
clamp screw ~ (Fig. 2-7).
• Raise the condenser carrier to its limit by turning the condenser
focus knob.
6
II. ASSEMBLY
Fig.2-8 Fig.2-9
@ Intermediate Tube
• Mount the intermediate tube onto the microscope arm fitting the
notch of the circular devetail of the intermediate tube with the end of
the clamp screw. Tighten the clamp screw holding the mid-stem of
the hexgon wrench. Do not over tighten the screw holding the
L-shaped holder of the wrench. (Fig. 2-9)
7 Fig.2-10 Fig.2-11
II. ASSEMBLY
QJ) Eyepiece
e Use the same magnification eyepieces for both the right and left
eyes.
elnsert the eyepieces into the sleeves of the binocular eyepiece tube
by engaging the three grooves of the eyepiece with the three
protrusions of the sleeve (Fig. 2-12).
e When using eyeguard rubbers, put them onto the eyepieces
(Fig.2-13).
Fig.2-12 Fig.2-13
Fig.2-14 8
II. ASSEMBLY
The rotation clamp ring must be removed when mounting the old
type of eyepiece. When attaching the ring again, note on the
following points.
uCaution when attaching rotation clamp ring
Hold the ring so that the ¢O.8 protrusion on the brim faces up.
There are two pawls on the ring and two grooves on the sleeve to
prevent the ring from failing off. Attach the ring so that these pawls
and grooves fits together.
Binocular eyepiece
O. 8-diameter protrusion Grooves tube sleeve
Pawls
Fig.2-15
Fig.2-16 Fig.2-17
9
II. ASSEMBLY
Fig.2-18 Fig.2-19
@ Power Cord
e Connect the socket of the power cord to the AC inlet on the rear of
the base (Fig. 2-19). Plug in the other end of the cord to an AC line
receptacle with the ground conductor (earth conductor).
For 100-120Varea------------------------------~
e Use only the following power supply cord set.
U L listed, detachable cord set. 3-conductor grounding type SVT,
No. 18 AWG rated a minimum 1 25V, 7 A.
el n case of using the extension cord, use only the power supply
cord including PE wire.
For220-240Varea------------------------------~
e Use only the 3-pole power supply cord type H05VV-F or
H05VVH2-F, which must be approved according to DIN VDE
0625. The plug and the outlet are to be approved according to
DIN VDE 0620 and DIN VDE 0625, respectively.
e Class I equipment should be connected to PE (protective earth)
terminal.
eln case of using the extension cord, use only the power supply
cord including PE wire.
@ Hexagon Wrench
e The hexagon wrench IS stored
in the back of the microscope
sta nd .
Take it out as shown in
Flg.2-20.
Fig.2-20
10
III. MICROSCOPY
III. MICROSCOPY
Fig.3-1
2) Mount the optional ND filter
cassette, if used (Fig. 3-2).
(Refer to II. ASSEMBLY- @
p.9.)
Fig.3-2
-""' ....
...----- ,
'\
3) Turn on the power switch rnto
light the lamp.
"
;,,;
, \, '
:~I
f'MoTe
Move the brightness adjuster
II~/
and align it with the 3rd scale
line from the right ~ (Fig. 3-3).
Fig.3-3
4) Pull the analyzer knob of the
intermediate tube rn to remove
the analyzer from the optical
path. Turn the Bertrand lens
turret to the position "0" ~ to
remove the Bertrand lens from
the optical path (Fig.3-4).
Fig.3-4
11
Ill. MICROSCOPY
Fig.3-5
6) Revolve the revolving nosepiece
to swing in the 10 x objective to
the optical path. Assure that
the nosepiece properly settles in
position (Fig. 3-6).
Fig.3-6
7} Place the specimen (glass slide)
on the stage with its cover glass
facing up and fasten it with the
two specimen clips (Fig. 3-7).
Fig.3-7
8} Raise the condenser uppermost
using the condenser focus knob
(Fig. 3-8).
Fig.3-8
12
1Il. MICROSCOPY
Fig.3-9
10) Bring the object portion of
specimen into the optical path
(Fig. 3-1 0). (Object portion IS
just above the condenser
lens.)
Fig. 3-1 0
11) When the trinocular eyepiece
tube is used, change the
optical path of the eyepiece
Binocular
.....
part: 100%
tube to enter 100% of the light
into the binocular part
(Fig.3-11).
(See p. 24, Operations in
Detail-11.)
Fig.3-11
12) Focus on the specimen
looking in the eyepiece by
manipulating the coarse/fine
focus knob (Fig.3-12).
Fig.3-12
13
Ill. MICROSCOP
Fig .3-13
14) Adjust the interpupillary
distance (Fig. 3-1 4).
(See p. 25, Operations in
Detail-1 2.)
Fig.3-14
15) Center the obj.ective
(Fig.3-15).
(See p. 20, Operations in
Detail-5. )
Fig.3-15
14
III. MICROSCOPY
Eyepiece
viewfield stop
Image of field
diaphragm
15
III. MICROSCOPY
Table 1
16
N. OPERATIONS IN DETAIL (Operation part)
N. OPERATIONS IN DETAIL
Fig.4 Fig.5
® Remove the eyepiece from the sleeve or put the Bertrand lens into the
optical path. Observing the exit pupil of objective inside, rotate the
dia~polarizer so as to form a dark cross image on the exit pupil as
shown in Fig. 5.
eThe analyzer can be rotated by loosening the analyzer clamp screw rn
and rotating the rotation ring ~.
The rotation angle can be read in the range between O· to 180" with
the unit of 0.1· by the vernier.
The analyzer can be removed from the optical path by pulling the
analyzer knob~. (Fig.6)
Note I As the depolarizer is built in the intermediate tube, you don't
need to be concerned about the relation between the
orientation of the polarizing plate and photomicrographic
device.
Fig.6
17
N. OPERATIONS IN DETAIL
Fig.7 Fig.S
18
lV. OPERATIONS IN DETAIL
Fig.9
19
IV. OPERATIONS IN DETAIL
(J) Rotate the stage about 180 and the target is displaced from the
,
center of the cross lines. Move the objective using the centering tools
so that the center of the cross lines moves one half the displacement
of the target. (Fig. 11 )
e Repeat the above procedure two or three times.
e Carry out centeri ng for each objective.
\
Rotate
180 0
----,.-_
d
-----.
~Y·~I I Mid position of
Target displacement
Fig.10 Fig.11
6. Stage Rotation
(Circular graduated stage)
eThe stage can be rotated by
loosening the stage rotation
clamp screw.
The rotation angle of the stage is
readable with the accuracy of
0.10 via a vernier scale. (Fig. 1 2)
Rotation clamp screw
Fig.12
20
N. OPERATIONS IN DETAIL
7. Focusing
(Coarse, Fine focus knobs and Coarse-focus torque adjustment
ring)
e Focusing is carried out by the coarse and fine focus knobs at the left
and right of the microscope stand.
eThe direction of vertical movement of the stage corresponding to the
turning direction of the focus knob is shown in Fig. 13.
e One rotation of the fine focus knob moves the stage 0.1 mm and the
graduation on the fine focus knob is 1 micron. One rotation of the
coarse focus knob moves the stage 1 2mm.
The range of coarse and fine motion is 2mm up and 28mm down
from the standard position.
Never turn the right or left knob while holding the other. It may
cause problems.
Do not turn the coarse and fine focus knobs further than the limit.
eThe rotation of the coarse-focus knob can be tightened when the
torque adjustment ring is turned counter-clockwise.
Fig.13
21
N. OPERATIONS IN DETAIL
Fig.14
(2) If the image decenters from the viewfield of the eyepiece, bring it
roughly to the center of the viewfield by means of the condenser
centering screws (Figs. 14~[Z] and 15).
Q) Change to the 40 x objective. Focus on the specimen by turning the
fine focus knob and form an image of the field diaphragm on the
specimen surface by manipulating the condenser focus knob.
@ When the image decenters from the viewfield of the eyepiece, bring it
to the center of the viewfield by means of the condenser centering
screws. At this time, adjusting the field diaphragm image to be
slightly smaller than the viewfield of the eyepiece may be convenient
for centering (Fig. 14-~).
Fig.15
22
N. OPERATIONS IN DETAIL
Exit pupil of
objective
Aperture
diaphragm
Fig.16
To adjust the size of the condenser aperture diaphragm, manipulate the
diaphragm control ring referring to the condenser's N. A. scale, or
observing the diaphragm Image VISible on the exit pupil inside the
objective. The eXit pupil can be seen by removing the eyepiece from the
eyepiece tube or putting the Bertrand lens into the optical path.
Stopping down the aperture diaphragm too far will lower the resolVing
power. Therefore, it IS recommended not to stop down the aperture to
a size smaller than 60% of the N. A. of the objective in use, except when
observing almost transparent specimens.
I Note I When SWinging out the top lens of the condenser (for
microscopy using 4 x or lower objective), fully open the
condenser aperture diaphragm.
23
N. OPERATIONS IN DETAIL
Fig.17 24
N. OPERATIONS IN DETAIL
Fig.18
1 3. Diopter Adjustment
(Eyepiece)
(Before adjusting the diopter, implement MICROSCOPY procedures 1) to
12) and focus on the specimen with the lOx objective.)
Make diopter adjustments for both the right and left eyepieces.
CD Turn the diopter compensation rings on each eyepiece until the end
surface of the ring coincides with the engraved line.
(This is the position of 0 dioptic compensation.) (Fig. 19~ 1)
® Swing in the 40 x objective by turning the revolving nosepiece and
bring the specimen image Into focus by turning the fine focus knob (or
the coarse focus knob) .
End surface
p"
CFVVN lOx
~
V Engraved line
Fig.19-1 Fig.19-2
25
N. OPERATIONS IN DETAIL
Fig.20
26
N. OPERATIONS IN DETAIL
ND4
ND16
About 50%
About 25%
About 6%
~1/2
~1/4
.-.1/16
~"8 /
1/64 t 1/32
>-'/ '2
8
27
N. OPERATIONS IN DETAIL
Fig.21
eTransparent Cover
Place the transparent cover on the N D filter cassette to protect the
filters from dust (Fig. 22-1). When the ND filter cass.ette is removed,
place the transparent cover on the field lens (Fig. 22-2). However,
remove the cover during observation or photomicrographing with the
viewfield diaphragm stopped down.
Fig.22-1 Fig.22-2
e Other filters
• ¢33mm Lemon skin filter
When the optional achromat condenser is used, insert the ¢33m m
lemon skin filter into the receptacle between the collector lens and
the lamp bulb. (Fig. 23-1)
• ¢45mm heat absorbing filter
When the heat absorbing filter is used, mount it into the diffuser
holder and insert into the bottom of the microscope base.
(Fig.23-2)
28
N. OPERATIONS IN DETAIL
Q - 1:3
¢45mm filterA-D;ff"OO' holder
Fig.23-1 Fig.23-2
29
v . OPTIONAL ACCESSORIES V. OPTIONAL ACCESSORIES
1 . Senarmont Compensator
To be inserted into the compensator slot of the intermediate tube in
place of the 1/4,,1. & tint plate to measure the retardation with the
accuracy of the lambda (A) unit. (Fig. 24)
Fig.24
1) Detecting extinction position
Rotate the stage with the specimen under the crossed Nicols to find
out the direction where the part of the specimen to be measured
appears darkest.
2) Detecting subtraction position
Rotate the stage 45° to bring it from the extinction position to the
diagonal position. Insert the 1/4,1 & tint plate into the optical path,
and confirm that the interference color of the part of the specimen to
be measured changes toward the lower order side. If the color
changes toward the higher order side, rotate the stage another 90°.
3) Measurement
Place the GI F filter on the filter receptacle, and replace the 1/4,1 &
tint plate with the compensator.
Rotate the analyzer so that the part of the specimen to be measured
becomes darkest.
Let the angle of the above analyzer rotation be theta (8) then the
retardation R (nm) will be obtained as follows:
8
R = 180,,1.
Where A: wave length of the light used for the measurement
When the GIF filter is used: ,,1.=546nm
30
v . OPTIONAL ACCESSORIES
2. Quartz Wedge
To be inserted into the compensator slot of the Intermediate tube instead
of the 1/4,1. & tint plate. (Fig.25)
The quartz wedge is engraved with scale and used for the rough
measurement of retardation in the range of 1 A-6A.
Fig.25
1) Detecting extinction position
Detect the position where the part of the specimen to be measured
becomes darkest by rotating the stage under the crossed Nicols.
2) Detecting subtraction position
Rotate the stage 45° to bring it from the extinction position to the
diagonal position. Insert the quartz wedge into the optical path, and
confirm that the interference color of the part of the specimen to be
measured changes toward the lower order side.
If the color changes toward the higher order side, rotate the stage
another 90°.
3) Measurement
Slide the quartz wedge along the slot, and the interference color
changes.
Stop sliding the quartz wedge where the dark stripe covers the part of
the specimen to be measured. In this position, find the part with no
object under the same dark stripe, and compare the interference
color of that part with the Interference Color Chart to assume the
amount of retardation.
If the view field around the part to be measured is entirely filled with
the object, restrict the illumination to the part to be measured by
means of the field diaphragm, remove the specimen from the optical
path and then compare the interference color with the Chart.
31
v . OPTIONAL ACCESSORIES
Fig.26
Fig.27 32
V. OPTIONAL ACCESSORIES
5. Universal Epi-illuminator
Used for episcopic polarizing microscopy, mounted between the
LABOPHOT 2-POL stand and the intermediate tube.
1) Nomenclature and Assembly
Referring to Fig. 28, assemble in the order given.
CD Remove the eyepiece tube and the intermediate tube from the
microscope stand.
(Z) Mount the universal epi-illuminator on the microscope arm,
positioning the illuminator nearly parallel to the arm. Clamp the
screw.
Q) Attach the lamp bulb (1 2-50W Halogen lamp) and socket to the lamp
housing. Release sufficiently the clamp screw on the lamp housing,
and mount it to the universal epi-illuminator. Retighten the clamp
screw.
® Connect the lamp cord to the transformer.
® Remove the accessory ND32 filter slider from the illuminator. Insert
the polarizer slider into the illuminator until the slider clicks twice.
® Place the filters.
(j) Mount the intermediate tube onto the illuminator, fitting the notch of
the circular dovetail with the end of the clamp screw. Fasten the
clamp screw.
® Referring to p. 8, mount the eyepiece tube on the intermediate tube.
I Intermediate tube I
(1" I
~
Lamp housing clamp screw
~ Socket sleeve clamp screw
Aperture diaphragm ring~
Field diaphragm ring ~
II '" r 0 Lamp lateral centering screw
Universal ~ C41
epi-illuminator~ ~ I Transformer I
Lamp vertical centering ring
Optical-path
change-over knob
Dust-tight slider
...~
:5)
-- Polarizer rotation ring
I Polarizer slider
- Intermediate tube clamp screw
I Microscope arm I
Fig.28
33
V. OPTIONAL ACCESSORIES
2) Preparation
(1) Centering the lamp
CD Make certain that the optical-path change-over knob is pushed in to
the limit.
(2) Turn ON the power switch on the transformer and set the voltage to
6V.
Q) Fully open the aperture diaphragm.
@ Place the ND filter on the stage and focus on it using the 10 x
objective.
@ Remove the eyepiece from the sleeve(or put the Bertrand lens into the
optical path), and looking into the exit pupil of objective, move the
lamp housing back and forth to form a sharp image of the lamp
filament on the diffuser of the exit pupil.
@ Manipulate the lamp centering screws to center the filament image on
the exit pupil.
(2) Orientation of polarizer
CD Roughly focus on the N D filter on the stage using the 40 x objective.
(2) Set the analyzer scale to "0".
Q) Bring the Bertrand lens into the optical path. Looking into the exit
pupil of the objective, rotate the polarizer rotation ring to form the
dark cross Image on the exit pupil. (Fig. 29)
Note I Take care not to touch the polarizer rotation ring while
observing the specimen, or the orientation of the polarizer will
have problems.
If it is touched by mistake, readjust the orientation.
Fig.29
3) Objectives
Use the CF M Plan Achromat P series objectives (Strain-free, 210/45).
34
VI. TROUBLESHOOTING
VI. TROUBLESHOOTING
35
VI, TROUBLESHOOTING
36
VI. TROUBLESHOOTING
Place It stably by
Image moves Specimen rises from stage surface. specimen clips on stage.
while being (p 12)
focused.
Condenser lens not correctly
Centering. (p.22)
centered.
Changeover so that
Optical path change-over of
100% of light enters
trlnocular tube not 100% binocular.
binocular part. (p.24)
37
VI. TROUBLESHOOTING
Insufficient
parfocality of Diopter adjustment.
Eyepiece diopter not adjusted.
objective when (p25)
changed-over.
Movement of
image not
smooth when
moving the
Attachable stage not tightly fastened
specimen. Fasten it tightly. (p32)
to stage.
(When optional
attachable
mechanical stage
is used)
Travel of stage
limited to
one-half length
of slide. Improper attachment of attachable Shift the attach ment
(When optional stage. pOSition (p.32)
attachable
mechanical stage
is used)
38
. TING
Adjust interpupillary
Interpupillary distance not adJusted.
distance. (p.25)
Adjust interpupillary
Interpupillary distance not adJusted.
distance. (p.25)
Diopter adjustment
Eyepiece diopter not adJusted.
(p.25)
l
combination of NO
filters. (p.27)
39
VI. TROUBLESHOOTING
ELECTRICAL
Failures Causes Actions
Connect the cord to
No electricity obtained.
socket.
Lamp does not
light even though No bulb attached. Attach lamp (p.5)
switched ON. Bulb blown. Replacement. (p.5)
Bulb immediately
blown. Use 6V 30W halogen
Bulb used not the one specified.
Insufficient lamp.
illumination.
40
ELECTRICAL SPECIFICATIONS
For 100-1 20V a rea - - - - - - - - - - - - - - - - - - - - ,
(1) Power Source: 100-120V AC, 0.8A, 50/60Hz
(2) Light Source
Lamp rating: 6V DC, 30W
Lamp type: Nikon 6V 30W Halogen lamp
(3) Protection Class: Class I
(4) Coforming Standards: This product conforms to UL 1262.
(5) Operating Environmental Conditions
Room temperature: 40°C max.
Relative humidity: 85% max.
REFERENCE
This manual Instructs only how to manipulate the LABOPHOT2-POL
microscope.
For the practical explanation on polarizing microscopy, refer to the
following special works:
• "ORE MICROSCOPY"
- Eugene n. Cameron -
John Wiley & Sons. Inc .
41
• Nikon reserves the right to make such
alterations in design as may be considered
necessary in the light of experience. For
this reason, particulars and illustration in
this handbook may not conform in every
detail to models in current production.
42