0% found this document useful (0 votes)
14 views10 pages

Rohleder 2005

This study compares mid-infrared and Raman spectroscopy for the quantitative analysis of serum from 247 blood donors, highlighting their reagent-free nature and speed. Both methods achieved similar accuracies in quantifying analytes such as glucose, cholesterol, and triglycerides, with root mean square errors of prediction for glucose being 14.7 mg/dl for mid-infrared and 17.1 mg/dl for Raman spectroscopy. The research indicates that vibrational spectroscopy can effectively analyze serum, though with limitations in accuracy within the 0.1 mmol/l range.

Uploaded by

GutoGonçalves
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
14 views10 pages

Rohleder 2005

This study compares mid-infrared and Raman spectroscopy for the quantitative analysis of serum from 247 blood donors, highlighting their reagent-free nature and speed. Both methods achieved similar accuracies in quantifying analytes such as glucose, cholesterol, and triglycerides, with root mean square errors of prediction for glucose being 14.7 mg/dl for mid-infrared and 17.1 mg/dl for Raman spectroscopy. The research indicates that vibrational spectroscopy can effectively analyze serum, though with limitations in accuracy within the 0.1 mmol/l range.

Uploaded by

GutoGonçalves
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 10

Journal of Biomedical Optics 10(3), 031108 (May/June 2005)

Comparison of mid-infrared and Raman spectroscopy


in the quantitative analysis of serum

D. Rohleder Abstract. Mid-infrared or Raman spectroscopy together with multi-


Roche Diagnostics GmbH
Mannheim, Germany variate data analysis provides a novel approach to clinical laboratory
and analysis, offering benefits due to its reagent-free nature, the speed of
University of Würzburg the analysis and the possibility of obtaining a variety of information
Institute for Physical Chemistry from one single measurement. We compared mid-infrared and Raman
Würzburg, Germany spectra of the sera obtained from 247 blood donors. Partial least
G. Kocherscheidt squares analysis of the vibrational spectra allowed for the quantifica-
K. Gerber tion of total protein, cholesterol, high and low density lipoproteins,
Dioptic GmbH triglycerides, glucose, urea and uric acid. Glucose (mean concentra-
Weinheim, Germany tion: 154 mg/dl) is frequently used as a benchmark for spectroscopic
analysis and we achieved a root mean square error of prediction of
W. Kiefer 14.7 and 17.1 mg/dl for mid-infrared and Raman spectroscopy, re-
University of Würzburg spectively. Using the same sample set, comparable sample through-
Institute for Physical Chemistry
Würzburg, Germany put, and identical mathematical quantification procedures Raman and
mid-infrared spectroscopy of serum deliver similar accuracies for the
W. Köhler quantification of the analytes under investigation. In our experiments
BaseLine GbR vibrational spectroscopy-based quantification appears to be limited to
Frankfurt, Germany accuracies in the 0.1 mmol/l range. © 2005 Society of Photo-Optical Instrumen-
tation Engineers. [DOI: 10.1117/1.1911847]
J. Möcks
Roche Diagnostics GmbH Keywords: Mid-infrared spectroscopy; Raman spectroscopy; serum.
Mannheim, Germany
Paper SS04144RR received Jul. 28, 2004; revised manuscript received Dec. 23,
2004; accepted for publication Dec. 28, 2004; published online May 24, 2005.
W. Petrich
Roche Diagnostics GmbH
Mannheim, Germany
and
University of Heidelberg
Faculty of Physics and Astronomy
Heidelberg, Germany

1 Introduction In contrast, serum is readily obtained from whole blood if 共in


The spectroscopy of molecular vibrations is experiencing a addition to the cellular components兲 those substances are re-
renaissance due to substantial technical advances in experi- moved which contribute to the coagulation cascade, e.g., fi-
mental methods, increased computational capabilities and brinogen. In daily routine, serum is simply collected from
growing analytical demands. Potential medical applications whole blood by means of centrifugation. Hence, this manu-
are currently being investigated in various fields such as script focuses on the serum which is an easily accessible
sample, needs no addition of anticoagulant, can be obtained
angiology, rheumatology, endocrinology, dentistry, or
reproducibly and is a very frequent sample type in routine
dermatology.1–5 In a routine clinical laboratory setting, body
laboratory diagnostics.
fluids like blood are particularly easy to obtain and are thus
Water forms the basis of body fluids and accounts for 90%
the most frequent primary material of investigation by labo-
of the serum. In the mid-infrared region, where most of the
ratory diagnostics.
basic vibrations of biomolecules occur, the absorption coeffi-
Blood has frequently been subjected to infrared and Ra-
cient of water amounts to 102 – 104 /cm making the direct ob-
man spectroscopy.6 –9 Since cellular components such as
servation of vibrational modes of organic compounds in body
erythrocytes account for ⬃42% of the weight of blood, the
fluids difficult. This difficulty may be mitigated or bypassed,
vibrational spectroscopy of whole blood is modified by Mie
e.g., by using transmission path lengths below 100 ␮m14
scattering. Removal of the solid components leaves the liquid
and/or intense mid-infrared light sources,15,16 by employing
phase, which is called plasma. Although many investigations
attenuated total reflection techniques,10 or by investigating
have shown very promising results for plasma,10–13 the col-
near-infrared overtones instead of the fundamental modes of
lection of plasma unfortunately requires the addition of a
the vibrations.11,17 Moreover, the body fluid could be dried
highly standardized type and concentration of anticoagulant. such that the sample’s water content is substantially reduced
and the remaining components of the body fluid can be di-
Address all correspondence to Dr. Wolfgang Petrich, Roche Diagnostics, San-
hofer Str. 116, Mannheim, 68305 Germany. Tel: +49-621-759-8324. E-mail:
wolfgang.petrich@roche.com 1083-3668/2005/$22.00 © 2005 SPIE

Journal of Biomedical Optics 031108-1 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

Table 1 Mean, maximum and minimum values and standard deviation ␴ ref of the different analyte concentrations as determined by the reference
method. The relative coefficient of variation (%CV) measures the precision of the reference methods upon re-sampling. The concentration of LDL
cholesterol was determined using the Friedewald formula.

Mean Min Max ␴ ref


Analyte [mg/dl] [mg/dl] [mg/dl] [mg/dl] Test method %CV

Total protein 7008 6100 8100 376 Colorimetric assay 0.95

Glucose 154 42 423 103 Enzymatic UV test 1.7

Urea 31 15 56 7 Kinetic UV assay 3.4

Uric acid 5.3 2.5 9.0 1.3 Enzymatic colorimetric test 1.7

Cholesterol 133 37 392 38 CHOD-PAP 1.7

Triglycerides 198 119 338 68 GPO-PAP 1.8

HDL cholesterol 54 14 99 14 Enzymatic colorimetric test 1.85

LDL cholesterol 118 47 249 34 ¯ ¯

rectly investigated by mid-infrared spectroscopy.6 Alterna- were determined by enzymatic tests using a MODULAR® PP
tively, various kinds of Raman spectroscopy may be used to system.§3 The concentration of LDL cholesterol was calcu-
provide access to the fundamental vibrational spectra of body lated by the Friedewald formula. Note that these reference
fluids.8,9,18 –20 Near-infrared Raman spectroscopy appears to concentrations of analytes were determined after the afore-
be particularly favorable, since it takes advantage of the low mentioned steps, particularly after spiking. Furthermore, the
absorption coefficient of water in the near-infrared spectral concentrations of cholesterol, triglycerides, HDL and LDL are
region 共which, e.g., amounts to 0.35/cm at a wavelength of 1 physiologically interrelated. While the square of Pearson’s
␮m兲 and since the fluorescent light background is strongly correlation coefficient indicates a substantial correlation ( r 2
reduced when compared to using visible light for Raman ⫽0.86) between cholesterol and LDL, r 2 is less than 0.3 for
spectroscopy. all other pairs of metabolites. For example, triglycerides are
For the research described in this report, two approaches not completely unrelated to cholesterol ( r 2 ⫽0.19) and HDL
were followed in order to investigate body fluids, namely ( r 2 ⫽0.21) . Qualitatively, these observations also hold true
near-infrared Raman spectroscopy of serum in its native form when calculating r 2 within the various subsets 共e.g., spiked
and mid-infrared spectroscopy of dried films of serum. While samples versus unspiked samples兲 individually.
each of the two spectroscopic techniques was optimized indi- Minimum, mean, and maximum concentrations of each
vidually, identical samples were used for both analyses, the analyte are given in Table 1 together with the standard devia-
samples throughput was required to be identical and the final tion of the concentrations and the reference test method. The
data analysis was performed with the identical software rou- standard measure for precision in a clinical laboratory is the
tines in order to allow for a close comparison of the two coefficient of variation, which is determined by remeasuring
approaches. the concentrations of analytes multiple times. The ratio be-
tween the coefficient of variation and the mean value of the
2 Materials and Methods concentration of the analyte under investigation, i.e., the rela-
Blood samples were collected from 238 healthy donors and tive coefficient of variation %CV, is listed in Table 1 for each
nine patients suffering from diabetes. The samples were cen- of the individual reference methods.
trifuged at 900 g for 30 min using a Heraeus Labofuge GL Since more than 180 million people suffer from diabetes
and the serum was isolated. Serum samples from 80 donors mellitus, a disorder of the glucose metabolism, the quantifica-
were set aside and glucose was added to these samples in the tion of glucose has frequently served as a benchmark for the
form of a glucose solution 共Fresenius Kabi Glucosteril® capabilities of vibrational spectroscopy in the context of clini-
70%兲. Forty of these samples were spiked with 2.2 ␮L glu- cal laboratory analysis. The quantification of glucose has
cose solution per millilitre of serum to increase the glucose therefore been emphasized throughout this manuscript.
concentrations by approximately 150 mg/dl. Glucose concen- We employed an automated pipetting system together with
trations in the other 40 samples were increased by approxi- a BRUKER Matrix HTS-XT spectrometer for the experiments
mately 300 mg/dl upon addition of 4.4 ␮L glucose solution using mid-infrared spectroscopy.21 Ninety-six-well silicon
per millilitre of serum. Subsequently, all of the serum samples sample carriers were used for mid-infrared transmission spec-
were partitioned into multiple aliquots of 3 mL each. All troscopy; 3 ␮L of serum were pipetted onto a sample carrier
samples were frozen at ⫺80 °C for storage purposes. One of in random order and left to dry in ambient air for 30 min.
the aliquots of each donor’s samples was subjected to clinical After drying, the film 共thickness of 2–10 ␮m兲 was subjected
chemistry testing: The concentrations of total protein, glucose,
§
uric acid, urea, cholesterol, triglycerides and HDL cholesterol MODULAR is a trademark of a member of the Roche group.

Journal of Biomedical Optics 031108-2 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

to mid-infrared transmission spectroscopy. Spectroscopy is request a throughput of at least 80 samples per day for both
performed in transmission using a DLaTGS detector, which in spectroscopic methods. Furthermore, liquid nitrogen cooling
contrast to mercury cadmium telluride 共MCT兲 detectors can had to be avoided in the view of possible future laboratory
be operated without liquid nitrogen cooling. Each spectrum application. We used samples from the same study for the
was recorded in the wave number range from 500 to 4000 investigation of both spectroscopic approaches and we split
cm⫺1 and consisted of 3629 data points. Spectra were ac- the spectra into the same calibration and validation sample
quired at a resolution of 4 cm⫺1 and averaged over 32 scans. subsets. Finally, identical multivariate analysis algorithms
Blackmann–Harris three-term apodization was used and the were used for the quantitative analysis of the pre-processed
zero filling factor was 4 共note that the use of a zero filling of spectra.
4 is standard practice in our laboratory but does not contribute After the spectroscopy and the pre-processing of spectra
to the accuracy of our results兲. To improve reproducibility had been completed, all data 共laboratory data and spectra兲 of
each sample was pipetted and measured on three sample car- the 247 donors were divided into a teaching set of 148 donors’
riers and the three absorbance spectra of each sample were data and a set of 99 donors’ data for independent validation.
corrected for sample carrier background and normalized. We Those samples exhibiting the lowest and the highest concen-
used a proprietary algorithm22 to correct for sample carrier trations of the different analytes were always assigned to the
background and to normalize the spectra. 共Note that due to the teaching set. The statistical equivalence of the teaching and
good reproducibility of our measurement method during this validation sets was verified on the basis of two-sample t-tests
study a simple background subtraction and standard vector and two-sample F-tests for the different analytes. For teach-
normalization will give similar results兲. At each wave number ing, partial least squares regression 共PLS兲 was performed us-
the median of the three pre-processed spectra was calculated ing MathWorks’ MatLab™ 6.0 Release 12 together with SIM-
and the resulting spectrum was then subjected to further PLS algorithm implemented in the PLSគToolbox 2.1 by
analysis. Although the actual integration time per spectrum Eigenvector Research, Inc. In order to optimize the training
was only 29 s, the triplicate measurement, the spectroscopic within the teaching data set, the root-mean-square error of
determination of the sample carrier background and the calibration 共RMSEC兲 and the root-mean-square error of
sample handling time resulted in an average processing time leave-one-out cross validation 共RMSECV兲 were calculated

冋 册
of 5 min per sample. The sample carriers were discarded after
N 1/2
use. 兺 i⫽1
teach
共 c pred,i ⫺c ref,i 兲 2
Stability of the mid-infrared system was coarsely assessed RMSEC⫽
N teach⫺LV⫺1
by calculating the area under the spectra before normalizing.

冋 册
Variations of the area under each spectrum are caused prima- N 1/2
rily by variations of the shape and thickness of the dried film 兺 i⫽1
teach
共 c pred,i ⫺c ref,i 兲 2
RMSECV⫽ .
of serum, which finally leads to a variation in optical path N teach
length. We find that the area under curve varied by less than
Here c ref,i and c pred,i denote the concentrations of analytes in
⫾20% among the measurements in this study. However, one
sample i as determined by the reference method and by the
sample was excluded from further analysis since the area un-
spectroscopic measurement, respectively. N teach is the number
der the mid-infrared spectra amounted to less than 50% of the
of teaching samples ( N teach⫽148) and LV is the number of
expected value for all three absorbance spectra originating
latent variables used for the PLS calibration. The optimum
from this sample.
A Kaiser Optical HoloSpec f/1.8i spectrometer was used
LV was chosen by selecting that value for LV, which corre-
sponds to the minimum of RMSECV.
for Raman spectroscopy. Laser radiation 共wavelength 785 nm兲
The validation set remained blinded until the teaching had
interacted with the sample within a quartz cuvette 共power at
been finalized. As a measure for the prediction accuracy of the
the location of the sample: 200 mW兲 and backscattered radia-
system, the root-mean-square error of prediction 共RMSEP兲
tion was collected using an Olympus PL4X lens. Ten quartz
was calculated according to
cuvettes were alternated and, after ten measurements, the cu-

冋 册
vettes were cleaned in 1% Hellmanex II solution 共Hellma N 1/2
GmbH&Co. KG, Müllheim/Baden, Germany兲 at 70 °C, dried 兺 i⫽1
val
共 c pred,i ⫺c ref,i 兲 2
RMSEP⫽ .
and used for the next set of measurements. Spectral resolution N val
was 8 cm⫺1. Spectra were acquired over 5 min during 12
acquisitions of 25 s each. The raw spectra were normalized N val is the number of validation samples ( N val⫽99 for Raman
and a fifth order polynomial background was subtracted in the spectroscopy, N val⫽98 for mid-infrared spectroscopy兲. Rela-
region from 300 to 1870 cm⫺1 using an iterative algorithm.23 tive errors 共%RMSEP兲 are calculated as the ratio between
Further details of the Raman experiments are reported in RMSEP and the mean concentration.
Ref. 19.
The strategy of the comparison was to use an optimum
setting for each spectroscopic method individually, but to re- 3 Results
quire the working conditions from a laboratory standpoint and An example of the mid-infrared spectrum of a dried film of
the data analysis to be as equivalent as possible for both meth- serum is given in Fig. 1. The mid-infrared spectrum is domi-
ods. Differences and similarities between the parameters used nated by the infrared absorption of proteins such as albumin
for the two approaches are listed in Table 2. While many of or globulins, which, after drying, constitute the major compo-
the parameters had become an internal working standard dur- nents of the serum film. Proteins exhibit characteristic vibra-
ing our prior investigations, we paid particular attention to tions of the polypeptide skeleton. The most pronounced peak

Journal of Biomedical Optics 031108-3 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

Table 2 Main characteristics of the parameters used for infrared and Raman spectroscopy. Note that
aliquots of identical serum samples have been used for both approaches.

FTIR RAMAN

Process parameters
Samples throughput 80/day
Need for liquid nitrogen cooling no
Sample carrier type Silicon plate Quartz cuvette
Sample carrier reuse No Yes
Background measurement Yes No
Sample volume used 100 ␮L 1 mL
Sample handling Automated Manual
Sample drying Yes No
Multiplicity of measurement Triplicate Single
Spectroscopy parameters
Light source Globar Semiconductor laser
Detector type DLaTGS CCD
Acquisition time for a single spectrum 30 s 5 min.
Detected wave number range 500–4000 cm−1 300–3500 cm−1
−1
Spectral resolution 4 cm 8 cm−1
Zero filling 4 ⬃2
Analysis parameters
Data pre-treatment Background correction, Subtraction of 5th order
normalization polynomial,
normalization
Wave number range used for PLS analysis 1220–1690 cm−1 300–1500 cm−1
of proteins
Wave number range used for PLS analysis 500–1800 and 300–1500 cm−1
of all other analytes 2500–3300 cm−1
Teaching set 148 serum samples
Teaching algorithm SIMPLS
Determination of optimum No. LV minimum of RMSECV
Independent validation set 99 serum samples
Measure of quality of quantification RMSEP

at 1653 cm⫺1 is caused by the Amide I vibration of the pep-


tide chain. Similarly, the peaks at 1545 cm⫺1 and around 1270
cm⫺1 can be assigned to the Amide II and Amide III vibra-
tions. The O–H stretch vibration is reflected as a broad fea-
ture around 3300 cm⫺1, which also contains a triplet structure
arising from stretch vibrations of ⫺C–H in ⫺CH2 and
⫺CH3 .
Figure 2 shows the Stokes-shifted Raman signal of serum
after background subtraction. While the ⫺C–Hx stretch vi-
brations around 2900 cm⫺1 appear to be similar to the mid-
infrared case, the spectrum substantially differs at lower wave
numbers. The Amide I band is strongly decreased and the
Amide III band is part of the most prominent feature of the
spectrum. Furthermore, the essential amino acids phenylala-
nine and tyrosine can be clearly identified at 1003 cm⫺1 and at
the 829/851 cm⫺1 duplet, respectively.
Quantification of the concentration of the analytes was per-
formed by training a PLS algorithm individually for each ana-
lyte within the spectral regions shown in Table 2. As an ex- Fig. 1 Mid-infrared spectrum of a film of dried serum after subtraction
ample, the root-mean-square errors resulting from the PLS of the background signal caused by the silicon sample carrier (␯:
analysis of infrared spectra are illustrated in Fig. 3 for the case stretch vibration. ␦: bending vibration).

Journal of Biomedical Optics 031108-4 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

ferent from that calculated for any value of LV between 9 and


31 共F-test, ␣⫽0.05兲. Similarly, different ranges of statistically
equivalent values for LV are given in Table 3 for the eight
parameters under investigation.
After the teaching of the algorithms had been finalized and
LV min had been determined, the spectra of the validation set
were subjected to blinded validation. The analyte concentra-
tions of the 98 validation samples investigated by mid-
infrared spectroscopy were predicted based on the PLS model
and subsequently compared to the concentration derived by
the laboratory method 共see Fig. 4; the predictions for the case
of the 99 Raman spectra are illustrated in Ref. 19兲. The results
of the quantitative analysis are summarized in Table 4 for
both, mid-infrared and Raman spectroscopy. For complete-
ness and since the particular values of LV min are subject to
some randomness 共as outlined above兲 we have also calculated
the RMSEP for all those PLS models, in which the number of
latent variables was within the discussed ranges of statistically
Fig. 2 Raman signal of serum after background subtraction. The equivalent values of LV. The minimum ( RMSEPmin) and
wavelength scale represents the wavelength of the Raman scattered maximum ( RMSEPmax) prediction error observed within the
light during illumination of the sample with laser light at 785 nm (␯: given range of LV are listed in Table 4 for each analyte.
stretch vibration. ␦: bending vibration). The energy difference allows
The predictions obtained by using mid-infrared spectros-
for the calculation of the Raman shift, which is expressed in terms of
wave numbers in this graph. copy may be compared to the results of Raman spectroscopy.
As an example, the difference between the predicted concen-
tration and its actual concentration as determined by labora-
tory analysis is illustrated for the case of glucose in Fig. 5 for
of glucose. Here, the minimum of RMSECV occurs at
both technologies. Beyond the qualitative impression of the
LV min⫽15. For glucose as well as the other seven parameters
scatter of the data, the calculation of RMSEP allows for a
LV min is given in Table 3 together with the corresponding
more quantitative comparison. A RMSEP of 14.7 and 17.1
values for RMSEC and RMSECV.
mg/dl was achieved for mid-infrared and Raman spectros-
Since the values of RMSECV are very similar in the vi-
copy, respectively. No significant differences between the
cinity of LV min , it is interesting to ask in how far RMSECV
mean values of the shown residue 共paired t-test; P⫽0.03) as
becomes minimal at the given values of LV min at random. In
well as their spread 共F-test; P⫽0.04兲 could be detected be-
other words, will the minimum of RMSECV be found at ex-
tween the two spectroscopic methods.
actly the same values of LV min as listed in Table 3 if one were
to repeat the whole experiment? We find that, for instance, in
the case of glucose the distribution of residue of the leave- 4 Discussion
one-out cross validation for LV min⫽15 is not significantly dif-
Our comparative study was dedicated to the quantitative
analysis of Raman spectra of native serum and mid-infrared
spectra of films formed from serum upon drying. To the best
of our knowledge the investigations presented in this manu-
script constitute the most comprehensive comparison between
mid-infrared and Raman spectroscopy with regard to deter-
mining the concentration of analytes in serum.
Particular attention was paid to requesting identical oper-
ating parameters for both methods from a clinical laboratory
viewpoint, namely equal throughput and the avoidance of liq-
uid nitrogen cooling. In addition, the data analysis procedures
were identical once the pre-treatment of the raw spectra of
each method was finalized. Furthermore, it was important that
the identical samples were used for both methods, including
the identical splitting and sorting into teaching and validation
data sets.
For an unbiased analysis it was important to rigorously
train the PLS algorithm using the teaching data only and to
perform an independent validation thereafter. As part of this
Fig. 3 Root-mean-square error of calibration (RMSEC), leave-one-out clear separation between the teaching and the validation data
cross validation (RMSECV) and prediction (RMSEP) as a function of
set, the search for the optimum dimensionality of the problem
the number of latent variables using the quantification of glucose
based on infrared spectra as an example. The minimum of RMSECV 共i.e., identifying ‘‘optimum’’ number of latent variables LV min
determines the ‘‘optimum’’ number of latent variables (15 in the case for the PLS algorithm兲 was based on the teaching set only.
of glucose). After all parameters of the PLS model had been defined, in-

Journal of Biomedical Optics 031108-5 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

Table 3 Results of the teaching procedure. LV min denotes that number of latent variables, for which the
root-mean-square error of leave-one-out cross validation (RMSECV) becomes minimal. RMSEC is the
root-mean-square error of calibration. Since RMSECV does not statistically differ from its minimum value
within a range of latent variables, the range of statistically equivalent values of LV is noted in brackets.

LV min RMSEC [mg/dl] RMSECV [mg/dl]

Analyte IR Raman IR Raman IR Raman

Total protein 3 10 300 118 318 157


(1–20) (7–15)

Glucose 15 10 9.5 17.1 15.6 22.6


(9–31) (8–15)

Urea 18 12 2.3 2.5 4.0 3.9


(11–50) (11–19)

Uric acid 10 12 0.9 0.6 1.2 1.2


(1–19) (1–28)

Cholesterol 12 12 10.9 7.4 15.1 11.8


(8–25) (10–50)

Triglycerides 19 15 9.8 7.3 16.7 19.3


(12–28) (10–50)

HDL cholesterol 15 10 6.6 7.2 10.9 9.8


(4–50) (7–26)

LDL cholesterol 12 14 13.9 5.9 19.5 15.3


(6–33) (10–50)

dependent validation was performed and the root-mean- highest concentration and they can be quantified within a rela-
square error of prediction 共RMSEP兲 was calculated. In retro- tive error as low as 2.5%. This tendency holds true for both
spect we find that those values of LV which provide the mid-infrared and Raman spectroscopy. In order to relate our
minimum value for RMSEP within the independent validation findings with present day clinical chemical analyzers it is also
set 共see Table 4兲 are frequently very close to our estimates instructive to understand the measurement accuracy in terms
LV min , which were derived from the teaching set only. Thus, of the number of molecules rather than their mass-related con-
we conclude that the method we used for estimating the opti- centration: Considering the molar weights of the analytes in-
mum number of latent variables provides a reasonable ap- vestigated, vibrational-spectroscopy based quantification ap-
proach to the problem of dimensionality. pears to be limited to accuracies in the 0.1 mmol/L range,
In our measurement setup, Raman spectroscopy required regardless of the particular choice of the spectroscopic tech-
larger sample volumes than the infrared spectroscopy. Al- nique. This finding is also supported by prior publications of
though we envisage that the volume used here 共1 mL兲 can be our and other groups as listed in Table 5.6,8,18 –19,24 –27
reduced to 200 ␮L by means of automation, it would still be High signal-to-noise ratios are considered a fundamental
twice the volume used in infrared spectroscopy. For infrared strength of infrared spectroscopy when compared to Raman
spectroscopy, the volume may even be reduced further: in spectroscopy. However, we find that this advantage does not
fact, we designed our system such that it can operate with result in a superior prediction accuracy when compared to
volumes as low as 70 ␮L and even lower volumes are con- Raman spectroscopy. This result supports our prior finding,
ceivable. that reproducibility rather than signal-to-noise ratio imposes a
Given the fact that vibrational changes in dipole moment lower limit on the prediction errors in mid-infrared spectros-
共or polarizability in the case of Raman spectroscopy兲 are of a copy, even if particular attention is paid to the reproducibility
similar order of magnitude for most biomolecules, the by virtue of automation, triplicate measurement, standardiza-
sample-specific detection capabilities mainly depend on the tion, and computational efforts.21 A small supplementary in-
concentration. The RMSEP values of the eight analytes under vestigation also points at the importance of reproducibility:
investigation are shown for both spectroscopic techniques as a five randomly chosen samples from the above study were re-
function of mean concentration in Fig. 6. RMSEP appears to measured over the course of the above experiments using
increase with analyte concentration. However, the ratio be- mid-infrared spectroscopy and the concentrations of analytes
tween RMSEP and mean value decreases with increasing con- were predicted on the basis of the PLS algorithm described
centration 共dashed lines in Fig. 6兲: Uric acid exhibits the low- above. For each sample and each analyte, the predicted con-
est concentration of all of the analytes investigated and centrations vary from measurement to measurement. In anal-
pertains a relative error of up to 26% upon quantification. In ogy to the clinical laboratory guidelines, the relative coeffi-
contrast, proteins constitute the molecular group with the cient of variation 共%CV兲 can be calculated as a measure for

Journal of Biomedical Optics 031108-6 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

Fig. 4 Concentrations of analytes in the validation samples as predicted by mid-infrared spectroscopy ( c pred) as compared to the concentrations
determined by the laboratory methods ( c ref). Corresponding Raman data have been published in Ref. 19.

the precision of the system. We find that, on average, %CV remeasuring the sample still substantially contributes to the
ranges from 4% 共protein兲 to 16% 共LDL兲. These numbers have overall error in the case of infrared spectroscopy of serum.
to be compared to 4.7% and 16.4% for the %RMSEP of pro- The challenge in reproducibility might be caused by the high
tein and LDL, respectively. Thus, we find that for these, as susceptibility of mid-infrared spectroscopy to changes in en-
well as most of the other analytes, the error observed upon vironmental conditions 共in particular water vapor and tem-

Journal of Biomedical Optics 031108-7 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

Table 4 Results of the independent validation. RMSEP兵 LV min其 denotes the root-mean-square error of
prediction at that number of latent variables, for which the root-mean-square error of leave-one-out cross
validation (RMSECV) became minimal (see Table 3). RMSEPmin and RMSEPmax denote the minimum and
maximum values of RMSEP observed when using PLS calibration models on the basis of different values
of LV (in braces) within a range statistically equivalent to LV min .

RMSEP兵 LV min其 关mg/dl兴 RMSEPmin兵LV其 关mg/dl兴 RMSEPmax兵LV其 关mg/dl兴

Analyte IR Raman IR Raman IR Raman

Total protein 328 兵3其 176 兵10其 323 兵4其 169 兵8其 434 兵20其 198 兵7其

Glucose 14.7 兵15其 17.1 兵10其 13.4 兵24其 16.9 兵9其 17.6 兵9其 21.1 兵14其

Urea 3.3 兵18其 4.4 兵12其 3.3 兵21其 4.4 兵12其 5.6 兵50其 4.9 兵17其

Uric acid 1.4 兵10其 1.1 兵12其 1.3 兵7其 1.1 兵11其 1.6 兵19其 1.3 兵1其

Cholesterol 16.1 兵12其 11.5 兵12其 15.0 兵11其 11.1 兵11其 18.0 兵24其 14.1 兵29其

Triglycerides 18.1 兵19其 20.7 兵15其 17.5 兵17其 19.8 兵12其 21.4 兵27其 23.9 兵50其

HDL cholesterol 11.9 兵15其 11.0 兵10其 11.8 兵14其 10.0 兵12其 21.1 兵44其 13.7 兵25其

LDL cholesterol 19.4 兵12其 15.7 兵14其 18.6 兵18其 14.6 兵11其 25.3 兵33其 19.1 兵50其

perature兲 which affect both the spectroscopy and the drying protein in our study population. Similar conclusions may be
process. In turn, the lower signal-to-noise ratio generally ob- drawn for HDL and uric acid, for which the relative prediction
served during Raman spectroscopy does not prevent the quan- errors exceed the biological variation among the donors of
tification of analytes in serum if a measurement time of 5 min our study population by only 30% or less. In contrast, the
per sample is acceptable. %RMSEP values for cholesterol, triglycerides, LDL and urea
In the light of a routine clinical laboratory application, the are up to four times smaller than the biological spread of
relative prediction errors 共%RMSEP兲 may be compared to the concentrations showing that for those parameters mid-infrared
standard deviations of reference concentrations, which prima- and Raman spectroscopy might supply a valuable tool for
rily reflect the physiological variations within the population quantification. It may be speculated that—similar to the case
under investigation. For example, the standard deviation of of glucose, where we have artificially spiked the samples to
the reference values amounts to only 5.4% of their mean con- deliver concentrations of glucose outside the normal, but well
centration for total protein. On the other hand, the concentra- within the possible physiological range—the quantification
tion of proteins can be predicted with a relative prediction accuracy of protein, HDL, and uric acid may appear more
error of 4.7% for mid-infrared spectroscopy. Ignoring any favorable in future studies, using samples which originate
non-Gaussian contribution to the distribution of concentra- from diseased people suffering, e.g., from dyslipidemia or
tions, it appears reasonable to conclude that the relative pre- gout.
diction error of the infrared spectroscopic approach is compa-
rable to the biological variations of the concentration of total

Fig. 5 Difference between the concentration c pred as predicted by Ra-


man (circles) or mid-infrared (triangles) spectroscopy and the concen- Fig. 6 Root-mean-square error of prediction as a function of mean
tration c ref as determined by the laboratory analysis. The dashed (dot- analyte concentration in the case of Raman (circles) and mid-infrared
ted) lines indicate the values of ⫾RMSEP for the mid-infrared (Raman) (triangles) spectroscopy. The dotted lines indicate relative errors of
spectroscopic data. prediction (%RMSEP) of 5%, 10% and 20%.

Journal of Biomedical Optics 031108-8 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

Table 5 Results of the multivariate analysis of mid-infrared and Raman spectra. The given values are the root-mean-square errors of prediction
(RMSEP) of a validation using N val independent samples. As an exception the results reported in Refs. 6 and 8 were obtained using leave-one-out
(LOO) validation and the values are therefore marked with an asterisk. For the study reported in this manuscript N val⫽99 samples were subjected
to the validation process for both methods. (Note that in the case of infrared spectroscopy one sample was excluded from the evaluation due to
unusually low absorbances in all three repetitions of the pipetting.) N tot⫽247 is the total number of samples used in the study. HDL and LDL denote
the high and low density lipoprotein fraction of cholesterol, respectively. All concentrations are given in mg/dl.

Reference N tot N val Total protein Triglycerides Cholesterol HDL LDL Glucose Urea Uric acid

Mid-infrared

This study 247 99 328 18.1 16.1 11.9 19.4 14.7 3.3 1.4
a
300 100 280 20.1 10.8 ¯ ¯ 7.4 6.6 2.4
b
300 100 310 23.6 11.2 ¯ ¯ 27 7.2 ¯
c
90 30 ¯ 30.6 14.7 12.0 13.5 ¯ ¯ ¯
d
122 24 ¯ 13 15 ¯ ¯ 16 ¯ ¯
e
306 (LOO) 240* 16.6* 11.3* ¯ ¯ 9.5* 2.0* ¯

Raman

This study 247 99 176 20.7 11.5 11.0 15.7 17.1 4.4 1.1
f
60 18–24 71 ¯ 10.4 ¯ ¯ ¯ ¯ ¯
g
66 (LOO) 190* 29* 12* ¯ ¯ 26* 3.8* ¯
a
Reference 24.
b
Reference 25.
c
Reference 26.
d
Reference 27.
e
Reference 6.
f
Reference 18.
g
Reference 8.

The accuracy of present-day laboratory testing for the pa- References


rameters investigated is still significantly better than the spec- 1. A. Mahadevan-Jansen, M. G. Sowa, G. J. Puppels, Z. Gyczynski, T.
troscopic results. Even more so, the presented results of spec- Vo-Dinh, and J. R. Lakowicz, 共Eds.兲 ‘‘Biomedical vibrational spec-
troscopy may be perceived as overoptimistic since, e.g., long- troscopy and biohazard detection technologies,’’ Proc. SPIE 5321,
term drifts and instrument-to-instrument variations were Bellingham, WA 共2004兲.
2. H.-U. Gremlich and B. Yan 共Eds.兲, Infrared and Raman Spectroscopy
purposely avoided in our study. Thus, one might be tempted of Biological Materials, Dekker, New York 共2001兲.
to conclude that the quantitative analysis of serum based on 3. J. M. Chalmers and P. R. Griffiths 共Eds.兲, Handbook of Vibrational
vibrational spectroscopy cannot compete with present day Spectroscopy, Vol. 5, Wiley, Chichester 共2002兲.
laboratory diagnostics. However, spectroscopy has the advan- 4. D. Naumann, ‘‘FT-infrared and FT-Raman spectroscopy in biomedi-
cal research,’’ Appl. Spectrosc. Rev. 36, 238 –298 共2001兲.
tage that only one measurement is needed in order to quantify 5. W. Petrich, ‘‘Mid-infrared and Raman spectroscopy for medical di-
all the shown parameters simultaneously. Furthermore, no re- agnostics,’’ Appl. Spectrosc. Rev. 36, 181–237 共2001兲.
agents are needed for the analysis, thus eliminating reagent 6. G. Werner, D. Boecker, H.-P. Haar, H. J. Kuhr, and R. Mischler,
costs and reducing logistic efforts. In cases for which moder- ‘‘Multicomponent assay for blood substrates in human sera and
haemolysed blood by mid-infrared spectroscopy,’’ in Infrared Spec-
ate accuracy is permissible, vibrational spectroscopy might troscopy: New Tool in Medicine, H. Mantsch and M. Jackson 共Eds.兲,
open the path towards a less expensive and more rapid analy- Proc. SPIE 3257, 91–100 共1998兲.
sis with the additional benefit of requiring small sample 7. S. Low-Ying, R. A. Shaw, M. Leroux, and H. H. Mantsch, ‘‘Quanti-
volumes. fication of glucose and urea in whole blood by mid-infrared spectros-
copy of dried films,’’ Vib. Spectrosc. 28, 111–116 共2002兲.
8. A. J. Berger, T.-W. Koo, I. Itzkan, G. Horowitz, and M. S. Feld,
Acknowledgments ‘‘Multicomponent blood analysis by near-infrared Raman spectros-
copy,’’ Appl. Opt. 38, 2916 –2926 共1999兲.
The work presented in this manuscript has benefited substan- 9. A. M. K. Enejder, T.-W. Koo, J. Oh, M. Hunter, S. Sasic, and M. S.
tially from prior work on mid-infrared spectroscopy at Roche Feld, ‘‘Blood analysis by Raman spectroscopy,’’ Opt. Lett. 27, 2004 –
Diagnostics and we thank R. Mischler and G. Werner for 2006 共2002兲.
10. H. M. Heise, R. Marbach, T. Koschinsky, and F. A. Gries, ‘‘Multi-
sharing their experience. This work would not have been pos- component assay for blood substrates in human plasma by mid-
sible without the enthusiasm and expertise of A. Orosz and F. infrared spectroscopy and its evaluation for clinical analysis,’’ Appl.
Reichert. Spectrosc. 48, 85–95 共1994兲.

Journal of Biomedical Optics 031108-9 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms


Rohleder et al.: Comparison of mid-infrared and Raman spectroscopy . . .

11. H. M. Heise and A. Bittner, ‘‘Multivariate calibration for near- 19. D. Rohleder, W. Kiefer, and W. Petrich, ‘‘Raman spectroscopy of
infrared spectroscopic assays of blood substrates in human plasma serum and serum ultrafiltrate,’’ Analyst (Cambridge, U.K.) 129, 906 –
based on variable selection using PLS-regression vector choices,’’ 991 共2004兲.
Fresenius’ J. Anal. Chem. 362, 141–147 共1998兲. 20. C. R. Yonzon, C. L. Haynes, X. Zhang, J. T. Walsh, Jr., and R. P. Van
12. E. Diessel, S. Willmann, P. Kamphaus, R. Kurte, U. Damm, and H. Duyne, ‘‘A glucose biosensor based on surface-enhanced Raman
M. Heise, ‘‘Glucose quantification in dried-down nanoliter samples scattering: improved partition layer, temporal stability, reversibility
using mid-infrared attenuated total reflection spectroscopy,’’ Appl. and resistance to serum protein interference,’’ Anal. Chem. 76, 78 – 85
Spectrosc. 58, 442– 450 共2004兲. 共2004兲.
13. G. Deleris and C. Petibois, ‘‘Application of FT-IR spectroscopy to 21. J. Moecks, G. Kocherscheidt, W. Köhler, and W. Petrich, ‘‘Progress
plasma contents analysis and monitoring,’’ Vib. Spectrosc. 32, 129– in diagnostic pattern recognition,’’ Proc. SPIE 5321, 117–123 共2004兲.
136 共2003兲. 22. J. Moecks, D. Rohleder, and W. Petrich, German patent application
14. R. Vonach, J. Buschmann, R. Falkowski, R. Schindler, B. Lendl, and 共pending兲.
R. Kellner, ‘‘Application of mid-infrared transmission spectrometry 23. C. A. Lieber and A. Mahadevan-Jansen, ‘‘Automated method for sub-
to the direct determination of glucose in whole blood,’’ Appl. Spec- traction of fluorescence from biological Raman spectra,’’ Appl. Spec-
trosc. 52, 820– 822 共1998兲. trosc. 57, 1363–1367 共2003兲.
15. K. Hebestreit, T. Beyer, A. Lambrecht, R. Mischler, M. Schoemaker, 24. R. A. Shaw, S. Kotowich, M. Leroux, and H. H. Mantsch, ‘‘Multi-
and W. Petrich, ‘‘Infrared spectroscopy of glucose solutions using analyte serum analysis using mid-infrared spectroscopy,’’ Ann. Clin.
quantum cascade lasers,’’ in SPIE Technical Summary Digest (BiOS Biochem. 35, 624 – 632 共1998兲.
2004, 5321–31), p. 116, SPIE, Bellingham, WA 共2004兲. 25. R. A. Shaw and H. H. Mantsch, ‘‘Multianalyte serum assay from
16. S. Schaden, M. Haberkorn, J. Frank, J. R. Baena, and B. Lendl, mid-infrared spectra of dry film on glass slides,’’ Appl. Spectrosc. 54,
‘‘Direct determination of carbon dioxide in aqueous solution using 885– 889 共2000兲.
mid-infrared quantum cascade lasers,’’ Appl. Spectrosc. 58, 667– 670 26. K.-Z. Liu, R. A. Shaw, A. Man, T. C. Dembinski, and H. H. Mantsch,
共2004兲. ‘‘Reagent-free, simultaneous determination of serum cholesterol in
17. K. H. Hazen, M. A. Arnold, and G. W. Small, ‘‘Measurement of HDL and LDL by infrared spectroscopy,’’ Cin. Chem. 48, 499–506
glucose and other analytes in undiluted human serum with near- 共2002兲.
infrared transmission spectroscopy,’’ Anal. Chim. Acta 371, 255–267 27. W. Petrich, B. Dolenko, J. Früh, M. Ganz, H. Greger, S. Jacob, F.
共1998兲. Keller, A. E. Nikulin, M. Otto, O. Quarder, R. L. Somorjai, A. Staib,
18. J. Y. Qu, B. C. Wilson, and D. Suria, ‘‘Concentration measurements G. Werner, and H. Wielinger, ‘‘Disease pattern recognition in infrared
of multiple analytes in human sera by near-infrared Raman spectros- spectra of human sera with diabetes mellitus as an example,’’ Appl.
copy,’’ Appl. Opt. 38, 5491–5497 共1999兲. Opt. 39, 3372–3379 共2000兲.

Journal of Biomedical Optics 031108-10 May/June 2005 䊉


Vol. 10(3)

Downloaded From: http://biomedicaloptics.spiedigitallibrary.org/ on 05/18/2015 Terms of Use: http://spiedl.org/terms

You might also like