78 Hri R40 R40W

The HRI-R40 and HRI-R40W series are insulation monitoring devices designed for medical rooms, capable of monitoring insulation resistance, temperature, and current overloads. The HRI-R40W model employs advanced measurement techniques to ensure accurate readings in networks with high harmonic and direct-current components. Both models comply with various international standards and offer features such as programmable settings, relay outputs, and optional RS485 communication.

Uploaded by

reza monfared
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
62 views4 pages

78 Hri R40 R40W

The HRI-R40 and HRI-R40W series are insulation monitoring devices designed for medical rooms, capable of monitoring insulation resistance, temperature, and current overloads. The HRI-R40W model employs advanced measurement techniques to ensure accurate readings in networks with high harmonic and direct-current components. Both models comply with various international standards and offer features such as programmable settings, relay outputs, and optional RS485 communication.

Uploaded by

reza monfared
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

HRI-R40 series

MICROENER
HRI-R40W series
INSULATION MONITORING
VERSIONS FOR USE IN MEDICAL ROOMS

GENERAL

HRI-R40 series
This device allow insulation monitoring to earth of supply network and Device has two input of temperature’s measure (one is optional) for tem-
thermal and electric overcharge monitoring of transformer. perature’s probe PT100 or PTC (DIN 44081) for monitoring thermal over-
This works in order to serve everything requested in specific standard charge of insulation’s transformer.
regulation for these applications. There is also a input of current’s measure of current transformer for
Insulation’s resistance monitoring is carried out applying a measure’s monitoring overcharge on network.
signalling between isolated network and earth. Signalling output are apt for coupling with specific panels of signalling
Surveying leakage generated to earth it’s possible to measure insulatio- and remote monitoring PR5 (max 2 panels, on request 4 panels).
n’s level. There is also a output for relay voltage-free with functions adjustable by
Modern and sophisticated measure’s techniques integrated allow correct the user.
measure of insulation’s resistance level also in case of strong obstruc- Optionally it’s available a serial output RS485 for bidirectional commu-
tions, with high harmonic and direct-current components. nication with monitoring system (PLC, PC, and so on). 25
HRI-R40 model uses a monitoring signalling with direct-current compo- Communication’s protocol used is MODUBUS-RTU.
nent. For reducing problems caused by the presence of direct-current Specific characteristics make these devices conform to standard regula-
components on network (rectifiers), device has a digital filter which is tion:
able to divide the majority of direct-current from eventual direct-current EN 61557-8
components in network. IEC 60364-7-710
HRI-R40 could set a large number of programming possibilities with VDE 0100 part 710
frontal button and 3 digit digital display for visualising measuring and CEI 64.8/7-710 V2
programming parameters. UNE 20615.

HRI-R40W series
HRI-R40W version has same fundamental characteristics of previous Actually presence in network of strong distortions with high harmonic
model but it uses particular measuring technique applying a measure’s components (sub harmonic) and direct-current components could cause
signal codified and varying, in order to guarantee a correct measure of measure’s problems to others techniques.
insulation independently from under-control network type. This solution allow the using of HRI-R40W in network with rectifiers,
power’s electronics, variable-speed drive and so on.

MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com
HRI-R40 series
HRI-R40W series MICROENER
INSULATION MONITORING
VERSIONS FOR USE IN MEDICAL ROOMS

FUNCTIONS AND OPERATORS - LEGEND

------------------------------------------------------------------------------------
1 display for visualising under-control parameters’ degree and for
visualising settings
------------------------------------------------------------------------------------
2 green LED SET for indication of programming status
------------------------------------------------------------------------------------
3 yellow LED ALARM for indication alarm for parameter’s
degree out-threshold
2 3 4 5 ------------------------------------------------------------------------------------
4 red LED OUTPUT RELAY for indication status of auxiliary relay output
------------------------------------------------------------------------------------
5 red LED ERROR / LINK FAIL for indication alarm of internal failure,
lack connection to under-control network, temperature’s probe
PT100 open or short circuit
contrel insulation monitor HRI-R40 ------------------------------------------------------------------------------------
1
SET ALARM OUTPUT ERROR /
6 red LED R for indication visualisation of insulation’s resistance
6 RELAY LINK FAIL
parameter; flashing light for out-threshold parameter
R [kΩ]
Z [kΩ] MIN / MAX
------------------------------------------------------------------------------------
7 T1 [°C] 7 red LED Z for indication visualisation of insulation’s impedance
T2 [°C]
+
_ RESET
SET
TEST
ENTER parameter; flashing light for out-threshold parameter.
8 I [A]
------------------------------------------------------------------------------------
9 8 red LED T1 for indication visualisation parameter of transformer’s
temperature; flashing light for out-threshold parameter.
10 ------------------------------------------------------------------------------------
9 red LED T2 for indication visualisation parameter of second sensor
temperature; flashing light for out-threshold parameter.
------------------------------------------------------------------------------------
11 12 13 10 red LED I for indication visualisation current of network parameter;
flashing light for out-threshold parameter.
------------------------------------------------------------------------------------
11 button +/- UP/DOWN for selecting parameter that has to be
visualised, for regulating device’s setting and for visualising
maximum and minimum memorized degrees
------------------------------------------------------------------------------------
12 button RESET / SET for entering device’s programming, for stopping
alarms and memorized degrees resetting
------------------------------------------------------------------------------------
13 button TEST / ENTER for testing device and remote signalling panels
and for confirming SETUP settings
------------------------------------------------------------------------------------

DIMENSIONS
26 74

contrel insulation monitor HRI-R40

SET ALARM OUTPUT ERROR /


RELAY LINK FAIL

90 R [kΩ]
45
Z [kΩ] MIN / MAX
T1 [°C]
T2 [°C]
+
_ RESET
SET
TEST
ENTER
I [A]

44
105 68 HRI-R40 / HRI-R40W

MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com
HRI-R40 series
MICROENER HRI-R40W series
INSULATION MONITORING
VERSIONS FOR USE IN MEDICAL ROOMS

ELECTRIC CHARACTERISTICS

TYPE HRI-R40 HRI-R40W


Auxiliary supply 110 - 230 V ±20% 110 - 230 V ±20%
Frequency 50 ÷ 60 Hz 50 ÷ 60 Hz
Self-consumption 5 VA 5 VA
Unde-control network voltage 24 ÷ 230 V 50 ÷ 60 Hz 24 ÷ 230 V 50 ÷ 60 Hz
Voltage's measure 24 V 24 V
Current's measure 1 mA 1 mA
Internal impedance for resistive measure 200 kohm 200 kohm
Selectable trip threshold 50 ÷ 500 kohm (low insulation)
hysteresis 10 % 20 ÷ 180 °C (over-temperature)
precision 2 % - 1 ÷ 99.9 A (over-current)
precision 2 % - delay 1 ÷ 60 sec
Visualisation insulation and impedance’s degree by three digits display 1 ÷ 999 kohm
temperature’s degree 0 ÷ 200 °C (1st and 2nd probe) by display
current’s value 0 ÷ 99.9 A by display
parameters’ configuration
output status:
- signalling alarms’ led
- led of signalling output of active relays
- led of failed insertion signalling

Output for PR5 (max 5) panel + 1 contact NO-C-NC 5 A - 250 V


low insulation, overload
+ option serial RS485 MODBUS-RTU
Input from isolated network 230 Vac (insulation measure)
1st probe PT100 2 or 3 wires (temperature measure) 30 ÷ 200 °C ±2%
2nd probe PT100 2 or 3 wires (temperature measure) (OPTIONAL) 30 ÷ 200 °C ±2%
CT (overload current’ measure max 5 A precision 2 %
current transformer ratio selectable 1 ÷ 40
Voltage of signal circuit < 24 Vdc < 24 Vdc
Measure’s method signalling dc codified and varying signal
Insulation’s test 2.5 KV 60 sec
Working temperature -10 ÷ 60 °C
Storing temperature -20 ÷ 80 °C
Relative humidity MAX 90 % 27
Standard regulation CEI-EN 61010-1 / CEI-EN 61557-8 / VDE 0413 part.8 / CEI 64.8/7-710 V2
IEC 60364-7-710 / VDE 0100 part.710 / UNE 20615 / CEI-EN 61326-1
Assembling according to DIN 50022 snap on DIN rail 35 mm
Dimensions 6 modules DIN 17.5 mm
Protection’s degree IP50 frontal - IP20 case
Connections by screw terminals max 2.5 mm2

MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com
HRI-R40 series
HRI-R40W series MICROENER
INSULATION MONITORING
VERSIONS FOR USE IN MEDICAL ROOMS

WIRING DIAGRAMS - LEGENDA HRI-R40AND HRI-R40W

PT100

PT100
A B C PT2+A PT2+B PT2- PT1+A PT1+B PT1- CT-S2 CT-S1 R-NO R-C R-NC

19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36

RS485 input 2 input 1 input 2


serial port probe probe current relay aux
(optional) PT100 PT100 TA .../5A output
remote panel
(o PTC) (o PTC)
HRI-R40 PR5

Vaux: * link fail insulation


115V 115V check check remote check panel
1 2 3 4 5 6 7 8 9 10 11
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 COM-S F-S V-P ACUS ACK OVER FAULT COM-P TEST- TEST-S TEST+
AUX-L1 AUX-01 AUX-L2 AUX-02 LF-PE LF-L VC-L VC-PE TEST+ TEST- ACK ACUS OVER COM-P V-P FAULT

auxiliary
supply
Vaux: 230 Vac

PE

TO OTHER
REMOTE
PANELS PR5
S S

L
INSULATION PC
TRANSFORMER INSULATED LINE (IT-M) LOAD
N

-----------------------------------------------------------------------------------------------------------------------------------------------------------
AUXILIARY SUPPLY - TERMINALS 1-2-3-4
supply’s section is carried out with double input 115 V nominal
for device supplying with 230 V, it is required to connect both sections in series
for device supplying with 115 V, it is required to connect both sections in parallel
-----------------------------------------------------------------------------------------------------------------------------------------------------------
LINK-FAIL MONITORING - TERMINALS 6-7
both terminals have to be connected between a isolated network’s phase and unipotential node (PE)
Maximum applicable voltage is 250 V (see insulation’s monitoring)
-----------------------------------------------------------------------------------------------------------------------------------------------------------
INSULATION’S MONITORING - TERMINALS 8-9
both terminals have to be connected between centre tap of secondary transformer or a isolated network’s phase and unipotential node (PE)
Maximum applicable voltage is 230 Vca so single phase networks could have max voltage of 230 V, three phase networks three wires
could have max voltage of 230 V phase-phase; but three phase networks four wires could have max voltage of 230 V phase-neutral
-----------------------------------------------------------------------------------------------------------------------------------------------------------
REMOTE PANEL’S CONNECTIONS PR5 - TERMINALS 10-11-12-13-14-15-16-18
connections for linking to remote panels PR5, Max voltage on these conductors is 24 V
-----------------------------------------------------------------------------------------------------------------------------------------------------------
SERIAL PORT RS485 (OPTIONAL) - TERMINALS 19-20-21
28 terminals A-B (19-20) head to serial bus, terminal C (21) is a mass’ reference that could be connected to eventual screen of cable RS485
Standard protocol used is modbus-rtu, documented in a specific handbook [IM833-U]
-----------------------------------------------------------------------------------------------------------------------------------------------------------
INPUT TEMPERATURE PROBE 2 (OPTIONAL) - TERMINALS 25-26-27
connections for linking to a temperature’s sensor - PT100 (EN 60751) or PTC (DIN 44081) sensor could be used.
In case of PT100 probes three wires, it is required to connect compensation’s conductor to the same terminal of correspondent conductor.
In case of PTC, it is necessary external resistor of 120 ohm min 1/4 W
-----------------------------------------------------------------------------------------------------------------------------------------------------------
INPUT TEMPERATURE PROBE 1 - TERMINALS 28-29-30
connections for linking to a temperature’s sensor - PT100 (EN 60751) or PTC (DIN 44081) sensor could be used.
In case of PT100 probes three wires, it is required to connect compensation conductor to the same terminal of correspondent conductor.
In case of PTC, it is necessary external resistor of 120 kohm min 1/4 W
-----------------------------------------------------------------------------------------------------------------------------------------------------------
INPUT FOR CURRENT MEASURE - TERMINALS 31-32
connection for external current transformer with secondary 5 A (current transformer ratio is programmable); if there is 3-phase network
must be used the special adapter type TSA-03 for 3 current transformer to permit to monitor in the insulation relay the highest value
of the 3-phase currents.
Only the module of the current value is measured (i.e. it is not important S1-S2 connection sequence).
-----------------------------------------------------------------------------------------------------------------------------------------------------------
OUTPUT AUXILIARY RELAY - TERMINALS 34-35-36
switch contact free from voltage and with programmable functions. Contact’s performance MAX 250 V 5 A resistive load
-----------------------------------------------------------------------------------------------------------------------------------------------------------
MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com