MICROENER
                                                                                            HRI-R40W series
                                                                                                                      INSULATION MONITORING
                                                                                      VERSIONS FOR USE IN MEDICAL ROOMS
GENERAL
HRI-R40 series
This device allow insulation monitoring to earth of supply network and       Device has two input of temperature’s measure (one is optional) for tem-
thermal and electric overcharge monitoring of transformer.                   perature’s probe PT100 or PTC (DIN 44081) for monitoring thermal over-
This works in order to serve everything requested in specific standard       charge of insulation’s transformer.
regulation for these applications.                                           There is also a input of current’s measure of current transformer for
Insulation’s resistance monitoring is carried out applying a measure’s       monitoring overcharge on network.
signalling between isolated network and earth.                               Signalling output are apt for coupling with specific panels of signalling
Surveying leakage generated to earth it’s possible to measure insulatio-     and remote monitoring PR5 (max 2 panels, on request 4 panels).
n’s level.                                                                   There is also a output for relay voltage-free with functions adjustable by
Modern and sophisticated measure’s techniques integrated allow correct       the user.
measure of insulation’s resistance level also in case of strong obstruc-     Optionally it’s available a serial output RS485 for bidirectional commu-
tions, with high harmonic and direct-current components.                     nication with monitoring system (PLC, PC, and so on).                        25
HRI-R40 model uses a monitoring signalling with direct-current compo-        Communication’s protocol used is MODUBUS-RTU.
nent. For reducing problems caused by the presence of direct-current         Specific characteristics make these devices conform to standard regula-
components on network (rectifiers), device has a digital filter which is     tion:
able to divide the majority of direct-current from eventual direct-current   EN 61557-8
components in network.                                                       IEC 60364-7-710
HRI-R40 could set a large number of programming possibilities with           VDE 0100 part 710
frontal button and 3 digit digital display for visualising measuring and     CEI 64.8/7-710 V2
programming parameters.                                                      UNE 20615.
HRI-R40W series
HRI-R40W version has same fundamental characteristics of previous            Actually presence in network of strong distortions with high harmonic
model but it uses particular measuring technique applying a measure’s        components (sub harmonic) and direct-current components could cause
signal codified and varying, in order to guarantee a correct measure of      measure’s problems to others techniques.
insulation independently from under-control network type.                    This solution allow the using of HRI-R40W in network with rectifiers,
                                                                             power’s electronics, variable-speed drive and so on.
      MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com
         HRI-R40 series
         HRI-R40W series                                                                                                                                              MICROENER
         INSULATION MONITORING
         VERSIONS FOR USE IN MEDICAL ROOMS
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      1    display for visualising under-control parameters’ degree and for
                                                                                                                           visualising settings
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      2    green LED SET for indication of programming status
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      3    yellow LED ALARM for indication alarm for parameter’s
                                                                                                                           degree out-threshold
                                           2                 3        4           5                                        ------------------------------------------------------------------------------------
                                                                                                                      4    red LED OUTPUT RELAY for indication status of auxiliary relay output
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      5    red LED ERROR / LINK FAIL for indication alarm of internal failure,
                                                                                                                           lack connection to under-control network, temperature’s probe
                                                                                                                           PT100 open or short circuit
                  contrel                               insulation monitor    HRI-R40                                      ------------------------------------------------------------------------------------
     1
                                           SET              ALARM    OUTPUT     ERROR /
                                                                                                                      6    red LED R for indication visualisation of insulation’s resistance
     6                                                                RELAY     LINK FAIL
                                                                                                                           parameter; flashing light for out-threshold parameter
                            R    [kΩ]
                            Z    [kΩ]           MIN / MAX
                                                                                                                           ------------------------------------------------------------------------------------
     7                      T1   [°C]                                                                                 7    red LED Z for indication visualisation of insulation’s impedance
                            T2   [°C]
                                           +
                                                            _       RESET
                                                                     SET
                                                                                TEST
                                                                                 ENTER                                     parameter; flashing light for out-threshold parameter.
     8                       I   [A]
                                                                                                                           ------------------------------------------------------------------------------------
     9                                                                                                                8    red LED T1 for indication visualisation parameter of transformer’s
                                                                                                                           temperature; flashing light for out-threshold parameter.
     10                                                                                                                    ------------------------------------------------------------------------------------
                                                                                                                      9    red LED T2 for indication visualisation parameter of second sensor
                                                                                                                           temperature; flashing light for out-threshold parameter.
                                                                                                                           ------------------------------------------------------------------------------------
                                                11                  12           13                                   10   red LED I for indication visualisation current of network parameter;
                                                                                                                           flashing light for out-threshold parameter.
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      11   button +/- UP/DOWN for selecting parameter that has to be
                                                                                                                           visualised, for regulating device’s setting and for visualising
                                                                                                                           maximum and minimum memorized degrees
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      12   button RESET / SET for entering device’s programming, for stopping
                                                                                                                           alarms and memorized degrees resetting
                                                                                                                           ------------------------------------------------------------------------------------
                                                                                                                      13   button TEST / ENTER for testing device and remote signalling panels
                                                                                                                           and for confirming SETUP settings
                                                                                                                           ------------------------------------------------------------------------------------
               DIMENSIONS
26                                                                                                                                                            74
                    90                     R    [kΩ]
                                                                                                                                                                                          45
                                           Z    [kΩ]                       MIN / MAX
                                           T1   [°C]
                                           T2   [°C]
                                                                     +
                                                                                       _       RESET
                                                                                                SET
                                                                                                          TEST
                                                                                                          ENTER
                                            I   [A]
                                                                                                                                                        44
                                                                105                                                                                           68                         HRI-R40 / HRI-R40W
         MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com
                                                                                                           HRI-R40 series
                  MICROENER                                                                               HRI-R40W series
                                                                                                                   INSULATION MONITORING
                                                                                                 VERSIONS FOR USE IN MEDICAL ROOMS
ELECTRIC CHARACTERISTICS
        MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com
      HRI-R40 series
      HRI-R40W series                                                                                                                                                                      MICROENER
      INSULATION MONITORING
      VERSIONS FOR USE IN MEDICAL ROOMS
PT100
                                                                                                 PT100
                        A       B     C                             PT2+A PT2+B      PT2-   PT1+A PT1+B   PT1-   CT-S2 CT-S1        R-NO   R-C   R-NC
19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36
                           auxiliary
                            supply
                         Vaux: 230 Vac
PE
                                                                                                                                                                                                             TO OTHER
                                                                                                                                                                                                             REMOTE
                                                                                                                                                                                                             PANELS PR5
                                 S                   S
                                                                L
             INSULATION                                   PC
           TRANSFORMER                                                                                           INSULATED LINE (IT-M)                                                                              LOAD
                                                                N
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     AUXILIARY SUPPLY - TERMINALS 1-2-3-4
     supply’s section is carried out with double input 115 V nominal
     for device supplying with 230 V, it is required to connect both sections in series
     for device supplying with 115 V, it is required to connect both sections in parallel
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     LINK-FAIL MONITORING - TERMINALS 6-7
     both terminals have to be connected between a isolated network’s phase and unipotential node (PE)
     Maximum applicable voltage is 250 V (see insulation’s monitoring)
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     INSULATION’S MONITORING - TERMINALS 8-9
     both terminals have to be connected between centre tap of secondary transformer or a isolated network’s phase and unipotential node (PE)
     Maximum applicable voltage is 230 Vca so single phase networks could have max voltage of 230 V, three phase networks three wires
     could have max voltage of 230 V phase-phase; but three phase networks four wires could have max voltage of 230 V phase-neutral
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     REMOTE PANEL’S CONNECTIONS PR5 - TERMINALS 10-11-12-13-14-15-16-18
      connections for linking to remote panels PR5, Max voltage on these conductors is 24 V
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     SERIAL PORT RS485 (OPTIONAL) - TERMINALS 19-20-21
28   terminals A-B (19-20) head to serial bus, terminal C (21) is a mass’ reference that could be connected to eventual screen of cable RS485
     Standard protocol used is modbus-rtu, documented in a specific handbook [IM833-U]
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     INPUT TEMPERATURE PROBE 2 (OPTIONAL) - TERMINALS 25-26-27
     connections for linking to a temperature’s sensor - PT100 (EN 60751) or PTC (DIN 44081) sensor could be used.
     In case of PT100 probes three wires, it is required to connect compensation’s conductor to the same terminal of correspondent conductor.
     In case of PTC, it is necessary external resistor of 120 ohm min 1/4 W
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     INPUT TEMPERATURE PROBE 1 - TERMINALS 28-29-30
     connections for linking to a temperature’s sensor - PT100 (EN 60751) or PTC (DIN 44081) sensor could be used.
     In case of PT100 probes three wires, it is required to connect compensation conductor to the same terminal of correspondent conductor.
     In case of PTC, it is necessary external resistor of 120 kohm min 1/4 W
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     INPUT FOR CURRENT MEASURE - TERMINALS 31-32
     connection for external current transformer with secondary 5 A (current transformer ratio is programmable); if there is 3-phase network
     must be used the special adapter type TSA-03 for 3 current transformer to permit to monitor in the insulation relay the highest value
     of the 3-phase currents.
     Only the module of the current value is measured (i.e. it is not important S1-S2 connection sequence).
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
     OUTPUT AUXILIARY RELAY - TERMINALS 34-35-36
     switch contact free from voltage and with programmable functions. Contact’s performance MAX 250 V 5 A resistive load
      -----------------------------------------------------------------------------------------------------------------------------------------------------------
      MICROENER - 49 rue de l'Université - F93160 Noisy le Grand - France - +33(0)148150909 - info@microener.com - www.microener.com