Publication No.
J32748
2018-11-06
Supersedes J32748
2018-10-26
Site Specifications
for the
VITROS® 3600 Immunodiagnostic System
PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the subject matter
gained by Ortho Clinical Diagnostics (Ortho) prior to publication.
No patent license is granted by this information.
Ortho reserves the right to change this information without notice, and makes no warranty, expressed or
implied, with respect to this information. Ortho shall not be liable for any loss or damage, including
consequential or special damages, resulting from any use of this information, even if loss or damage is
caused by the negligence or other fault of Ortho.
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.
Table of Contents
Description Page
Electrical Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Electrical Specifications for the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electrical Specifications for the Supplied PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . 6
RECEPTACLE Location Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Networking Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Requirements at Customer Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Shipping Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Specifications for the System and the PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Specifications for the 3600 HEAT REJECTION PLENUM. . . . . . . . . . . . . . . . . . . . . 13
2 2018-11-06 – J32748
Electrical Power Requirements
Section 1: Electrical Power Requirements
AC Power Quality
The VITROS 3600 Immunodiagnostic System (hereafter referred to as the System) is designed to perform
satisfactorily with reasonable disturbances and voltage tolerance to the ac power. The quality and reliability of the
power source is often dependent on the geographic location of the site. Some sites, where power quality is
exceptionally poor, may require special power conditioning equipment to ensure that the tolerances are maintained
within acceptable limits. Some models of UNINTERRUPTIBLE POWER SUPPLIES (UPS) are recommended as
optional equipment for use with VITROS Systems, and may be purchased from the Distribution Center. See the UPS
service publication for additional information.
Static Electricity
The System should be adequately grounded to prevent the effects of static charges generated. Static problems can
also be reduced by ensuring that the relative humidity of the room is maintained above 30 percent. Special anti-static
floor pads can be purchased and positioned where the operator is required to make contact with the instrument.
Electromagnetic Interference
Several types of electrical interference may be indigenous to the site location and may require special filtering to
prevent malfunctions in the System. If interference is suspected to be a problem, shielding may be required, or
filtering of the ac line power to the site may be necessary.
Cellular phones, two-way pagers, and other RF transmitting devices should not be used within 1 m of the 3600
System.
2018-11-06 – J32748 3
Section 2: Electrical Specifications for the System
Input Voltage (nominal) 200 to 240 V
Input Voltage (min/max) 180 to 264 V
Input Frequency Limits 47 to 63 Hz
Power Operating 2709 W
(Average)
Maximum Current (does 16A
not include momentary
inrush current)
System tolerance to Tolerant of missing one full cycle at 50 Hz (20 ms)
voltage interruptions Complies with EN61326-2-6 requirements.
Circuit Type (refer to Dedicated with isolated ground
section 7)
POWER CORD
(included with system) IEC 60320-C19 to NEMA L6-20P
Receptacle Type
(No UPS)
North America NEMA L6-20R
OR
Continental Europe CEE 7/7 “Schuko”
OR
Rest of World equivalent to meet local electrical code
Receptacle Type
(with 6 KVA UPS)
North America NEMA L6-30R For sites where a UPS will be
OR installed at a later date,
Continental Europe equivalent to meet local electrical code North America: 1 - 20 Amp
OR and 1- 30 Amp circuits are
Rest of World equivalent to meet local electrical code required.
Continental Europe: 1 - 16
Amp and 1- 32 Amp circuits
are required.
PRINTER Electrical Refer to Section 3: Electrical Specifications for
Requirements the Supplied PRINTER
4 2018-11-06 – J32748
Electrical Specifications for the System
NEMA L6-20R for 200 V ac - 240 V ac
208 V Check between Voltage
L2
60 Hz L1 - L2 190 - 220 V AC
L1
L1 - SAFETY GROUND 113 - 127 V
AC
L2 - SAFETY GROUND 113 - 127 V
AC
240 V Check between Voltage
SAFETY
GROUND 60 Hz L1 - L2 226 - 254 V AC
E002_1332ACA
L1 - SAFETY GROUND 113 - 127 V
AC
L2 - SAFETY GROUND 113 - 137 V
AC
Note
CEE 7/7 “Schuko” plug may be supplied, either straight
or right angle.
2018-11-06 – J32748 5
Section 3: Electrical Specifications for the Supplied PRINTER
Important
The following information is only applicable to PRINTERS supplied by Ortho. If the PRINTER is being supplied by
the customer, refer to the manufacturer’s PRINTER manual for the electrical specifications.
1 RECEPTACLE required 3-wire, 120 V AC, polarized, SAFETY GROUND,
maximum distance of 182.7 cm (72 in.) from each
side of the system.
CIRCUIT Type Dedicated
120 V Check between Voltage
60 Hz N - L1 102 - 132 V AC
15 amperes L1 - SAFETY GROUND 102 - 132 V AC
N - SAFETY GROUND Less than 1 V AC
Detach 1
NEMA 5-15R
SAFETY
GROUND
L1
polarized power
RECEPTACLE
E002_1331ACA
E002_1331AC
6 2018-11-06 – J32748
RECEPTACLE Location Specifications
Section 4: RECEPTACLE Location Specifications
Specifications for PRINTER POWER and Specifications for the System only
NETWORK CONNECTION
Reference Description
Reference Description
Q 30.5 cm (12 in.) minimum
Q 30.5 cm (12 in.) minimum (height from floor) 61.0 cm (24 in.) maximum
(height from floor) 61.0 cm (24 in.) maximum
R 1.98 m (6 ft. 6 in.) maximum
S 2.9 m (8 ft. 10 in.) maximum (system power)
(printer power)
T 5.3 m (17 ft. 6 in.) maximum
(Network)
Detach 2
T
S
Q
Q
Q
Q
Q
Q
2018-11-06 – J32748 7
Section 5: Environmental Specifications
Acoustic Noise Level Idle: 60 dB Thermal Characteristics
Operational: 65 dB On 6130 BTU/hr
Detach 3
Maximum altitude: 2438 m (8000 ft.)
Laboratories operating a VITROS® System at altitudes greater than the Ortho published site specification of 8000 ft
(2438 m) are responsible for validating those systems in their respective laboratories. In addition, Ortho recommends
following good laboratory practice and adhering to all required quality control and maintenance procedures to ensure
result integrity.
This product has been evaluated and is suitable only for Pollution Degree II and Insulation Category II.
Additional Requirements for the 3600 System HEAT REJECTION PLENUM
If the customer chooses to purchase and use the optional Heat Rejection Plenum, the EXHAUST SYSTEM and 2
DUCTS must be installed by the customer.
Specifications per DUCT
Air Capacity
7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute
259 - 366 metres (850 - 1200 ft.) per minute
DUCT (each)
20.5 cm (8 in.) flexible AIR DUCT and attaching hardware
Air Flow - Ventilation System
7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute at 5.1 mm (0.2 in.) of water vacuum
Important
The HEAT REJECTION PLENUM can be installed with an AT CONVERSION KIT up to a temperature limit of 28
degrees C (82.4 degrees F) and a relative humidity level of 30%.
8 2018-11-06 – J32748
Networking Specifications
Section 6: Networking Specifications
Network Connection Specifications
The requirements for e-Connectivity network connection are located in the Network Information Form located
on-line at:
https://techdocs.orthoclinicaldiagnostics.com/notes/contentpages/
techdoc.aspx?view=USOTHERECI&culture=en‐us
The customer must complete the form and return it to Ortho Clinical Diagnostics.
Important
All requirements for lab network connections must be completed before installation of the analyzer(s).
2018-11-06 – J32748 9
Section 7: Requirements at Customer Site
TELEPHONE A TELEPHONE located so that the user can move around the System
must be available.
Air Flow No air flow sources directed towards the system within 3.0 m (10 ft).
Entrance to the room where Minimum width of 81.3 cm (32 in.)
the System will be installed
LAB COMPUTER Interface See the Interface Specification section of the Operator’s Manual.
Network Connection Refer to the Networking Specifications on page 9.
Floor Loading 433.4 kg/m2 (88.9 lb./ft2)
Circuit Breakers 2 circuit breakers capable of handling high in-rush current.
In Europe, the following applies:
1 dedicated line with 16A 'D' type circuit breaker
For systems equipped with a UPS:
1 dedicated line 32A 'D' type circuit breaker
10 2018-11-06 – J32748
Shipping Specifications
Section 8: Shipping Specifications
Shipping Specifications
Reference Description
A 262.9 cm (103.5 in.)
B 115.6 cm (45.5 in.)
C 203.2 cm (80.0 in.)
D 254.0 cm (100 in.)
Weight 948.0 kg (2090 lbs.)
* Minimum overhead clearance required for unpacking.
Detach 4
2018-11-06 – J32748 11
Section 9: Specifications for the System and the PRINTER
System Specifications Supplied PRINTER Specifications ****
Reference Description Reference Description
A 212 cm (83.5 in.) J 40.0 cm (15.7 in.)
B 88.7 cm (34.9 in.) K 51.3 cm (20.2 in.)
C 163.8 cm (64.5 in.) L 34 cm (13.4 in.)
D 213.4 cm (84.0 in.) M 61 cm (24 in.)
Weight 789.2 kg (1740 lb.) Weight 19.7 kg (43.5 lb.)
Port USB Version 2.0
Printer Post Script
Language
Installation Specifications Installation Specifications
for the System only for the Supplied PRINTER ****
Reference Description Reference Description
E 76.2 cm (30.0 in.) N 17.8 cm (7.0 in.)
F 76.2 cm (30.0 in.) O 45.7 cm (18.0 in.)
Required clearance for opening side door to
allow service access.
Does not include PRINTER
It is acceptable to have moveable objects (such as a PRINTER P 61.0 cm (24.0 in.)
STAND) within this area, but no closer than 61 cm (24 in.) to the Requirement for proper system airflow. No
system. objects should be located within this space.
G 45.7 cm (18.0 in.) ** **** If
the PRINTER is being supplied by the
Q 76.2 cm (30.0 in.) customer, refer to the manufacturer’s PRINTER
** 91
manual for any specifications.
cm (36 in.) if SEISMIC ANCHORAGES, AT
Accessory Kit or HEAT REJECTION PLENUM are to
be installed.
B
D
C G
E J
A L
K
P N
AIS
LE M
Q F
12 2018-11-06 – J32748
Specifications for the 3600 HEAT REJECTION PLENUM
Section 10: Specifications for the 3600 HEAT REJECTION PLENUM
HEAT REJECTION PLENUM
Specifications
Reference Description
A 89.1 cm (35.1 in.)
B 74.5 cm (29.3 in.)
C 31.5 cm (12.4 in.)
D 16.0 cm (6.3 in.)
E 14.5 cm (5.7 in.)
Detach 5
Detach 6
2018-11-06 – J32748 13
14 2018-11-06 – J32748
IVD
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed
EC REP
Bridgend
CF35 5PZ
United Kingdom
Ortho Clinical Diagnostics
100 Indigo Creek Drive
Rochester, NY 14626
USA