Emi Electrical (4th Sem. Notes)
Emi Electrical (4th Sem. Notes)
FOR
4th SEMESTER OF
ELECTRICAL ENGINEERING
(B.TECH PROGRAMME)
UNIT-1
1
SYLLABUS
2
Course Outcome For Electronic Measurement & Instrumentation
3
4
MEASURING INSTRUMENTS
Instrument
An absolute instrument determines the magnitude of the quantity to be measured in terms of the
instrument parameter. This instrument is really used, because each time the value of the measuring
quantities varies. So we have to calculate the magnitude of the measuring quantity, analytically
which is time consuming. These types of instruments are suitable for laboratory use. Example:
Tangent galvanometer.
This instrument determines the value of the quantity to be measured directly. Generally these
instruments are calibrated by comparing with another standard secondary instrument.
Examples of such instruments are voltmeter, ammeter and wattmeter etc. Practically
secondary instruments are suitable for measurement.
Secondary instruments
5
1.3.1 Indicating instrument
This instrument uses a dial and pointer to determine the value of measuring quantity. The pointer
indication gives the magnitude of measuring quantity.
This type of instruments records the magnitude of the quantity to be measured continuously over
a specified period of time.
This type of instrument gives the total amount of the quantity to be measured over a specified
period of time.
1.3.4 Electromechanical indicating instrument
For satisfactory operation electromechanical indicating instrument, three forces are necessary.
They are
(a) Deflecting force
(c)Damping force
When there is no input signal to the instrument, the pointer will be at its zero position. To deflect
the pointer from its zero position, a force is necessary which is known as deflecting force. A system
which produces the deflecting force is known as a deflecting system. Generally a deflecting system
converts an electrical signal to a mechanical force.
6
1.4.1 Magnitude effect
When a current passes through the coil (Fig.1.2), it produces a imaginary bar magnet. When a soft-
iron piece is brought near this coil it is magnetized. Depending upon the current direction the poles
are produced in such a way that there will be a force of attraction between the coil and the soft iron
piece. This principle is used in moving iron attraction type instrument.
Fig. 1.2
If two soft iron pieces are place near a current carrying coil there will be a force of repulsion
between the two soft iron pieces. This principle is utilized in the moving iron repulsion type
instrument.
When a current carrying coil is placed under the influence of magnetic field produced by a
permanent magnet and a force is produced between them. This principle is utilized in the moving
coil type instrument.
Fig. 1.3
When two current carrying coils are placed closer to each other there will be a force of repulsion
between them. If one coil is movable and other is fixed, the movable coil will move away from the
fixed one. This principle is utilized in electrodynamometer type instrument.
7
Fig. 1.4
To make the measurement indicated by the pointer definite (constant) a force is necessary which
will be acting in the opposite direction to the deflecting force. This force is known as controlling
force. A system which produces this force is known as a controlled system. When the external
signal to be measured by the instrument is removed, the pointer should return back to the zero
position. This is possibly due to the controlling force and the pointer will be indicating a steady
value when the deflecting torque is equal to controlling torque.
Td (1.1)
Tc
Two springs are attached on either end of spindle (Fig. 1.5).The spindle is placed in jewelled
bearing, so that the frictional force between the pivot and spindle will be minimum. Two springs
are provided in opposite direction to compensate the temperature error. The spring is made of
phosphorous bronze.
When a current is supply, the pointer deflects due to rotation of the spindle. While spindle is rotate,
the spring attached with the spindle will oppose the movements of the pointer. The torque produced
by the spring is directly proportional to the pointer deflection .
TC (1.2)
The deflecting torque produced Td proportional to ‘I’. When TC Td , the pointer will come to a
8
steady position. Therefore
(1.3)
I
9
Fig. 1.5
Since, and I are directly proportional to the scale of such instrument which uses spring controlled
is uniform.
1.6 Damping force
The deflection torque and controlling torque produced by systems are electro mechanical.
Due to inertia produced by this system, the pointer oscillates about it final steady position before
coming to rest. The time required to take the measurement is more. To damp out the oscillation is
quickly, a damping force is necessary. This force is produced by different systems.
The piston is mechanically connected to a spindle through the connecting rod (Fig. 1.6). The
pointer is fixed to the spindle moves over a calibrated dial. When the pointer oscillates in clockwise
direction, the piston goes inside and the cylinder gets compressed. The air pushes the piston
upwards and the pointer tends to move in anticlockwise direction.
10
Fig. 1.6
If the pointer oscillates in anticlockwise direction the piston moves away and the pressure of the
air inside cylinder gets reduced. The external pressure is more than that of the internal pressure.
Therefore the piston moves down wards. The pointer tends to move in clock wise direction.
An aluminum circular disc is fixed to the spindle (Fig. 1.6). This disc is made to move in the
magnetic field produced by a permanent magnet.
11
When the disc oscillates it cuts the magnetic flux produced by damping magnet. An emf is induced
in the circular disc by faradays law. Eddy currents are established in the disc since it has several
closed paths. By Lenz’s law, the current carrying disc produced a force in a direction opposite to
oscillating force. The damping force can be varied by varying the projection of the magnet over
the circular disc.
12
Fig. 1.7
Principle of operation
When D.C. supply is given to the moving coil, D.C. current flows through it. When the current
carrying coil is kept in the magnetic field, it experiences a force. This force produces a torque and
the former rotates. The pointer is attached with the spindle. When the former rotates, the pointer
moves over the calibrated scale. When the polarity is reversed a torque is produced in the opposite
direction. The mechanical stopper does not allow the deflection in the opposite direction. Therefore
the polarity should be maintained with PMMC instrument.
If A.C. is supplied, a reversing torque is produced. This cannot produce a continuous deflection.
Therefore this instrument cannot be used in A.C.
TC = controlling torque
= angle of
deflection K=spring
constant b=width of
the coil
13
l=height of the coil or length of coil
N=No. of turns
I=current
B=Flux density
A=area of the coil
The force produced in the coil is given by
(1.4)
F BIL sin
When 90
Td NBIL b (1.7)
BINA
Td BANI (1.8)
Td I (1.9)
Advantages
✓ Torque/weight is high
✓ Power consumption is less
✓ Scale is uniform
✓ Damping is very effective
✓ Since operating field is very strong, the effect of stray field is negligible
✓ Range of instrument can be extended
Disadvantages
✓ Use only for D.C.
✓ Cost is high
✓ Error is produced due to ageing effect of PMMC
✓ Friction and temperature error are present
14
1.7.1 Extension of range of PMMC instrument
Case-I: Shunt
A low shunt resistance connected in parrel with the ammeter to extent the range of current. Large
current can be measured using low current rated ammeter by using a shunt.
Fig. 1.8
I= current to be measure
Vm Vsh (1.10)
ImRm IshRsh
I
(1.11)
m
R
Is sh
R
h
m
Eqn (1.12) ÷ by Im
15
I Ish
1 (1.13)
Im Im
16
I
1 (1.14)
Rm
Im Rsh
Rm
I Im 1 R (1.15)
sh
Rm
A large resistance is connected in series with voltmeter is called multiplier (Fig. 1.9). A large
voltage can be measured using a voltmeter of small rating with a multiplier.
Fig. 1.9
Rm =resistance of meter
Let
Rse =resistance of multiplier
Im (1.16)
Ise
17
Vm (1.17)
Vse Rm
Rs
e
Vs (1.18)
e Rse
V Rm
m
18
(1.19)
Apply KVL, V Vm
Vse
V Vse Rse
V 1 V 1R (1.20)
m m m
Rse
V Vm 1 R (1.21)
m
Rse
1 R Multiplication factor
m
Principle of operation
The current to be measured is passed through the fixed coil. As the current is flow through the
fixed coil, a magnetic field is produced. By magnetic induction the moving iron gets magnetized.
The north pole of moving coil is attracted by the south pole of fixed coil. Thus the deflecting force
is produced due to force of attraction. Since the moving iron is attached with the spindle, the
spindle rotates and the pointer moves over the calibrated scale. But the force of attraction depends
on the current flowing through the coil.
20
Fig. 1.10
There is change in inductance since the position of moving iron change w.r.t the fixed
electromagnets.
Let the new inductance value be ‘L+dL’. The current change by ‘di’ is dt seconds.
Let the emf induced in the coil be ‘e’ volt.
d di
e (Li) L i (1.22)
dL
dt dt dt
Multiplying by ‘idt’ in equation (1.22)
di dL
e idt L idt i idt (1.23)
dt dt
Eqn (1.24) gives the energy is used in to two forms. Part of energy is stored in the inductance.
Remaining energy is converted in to mechanical energy which produces deflection.
21
Fig. 1.11
22
Change in energy stored=Final energy-initial energy stored
1 1 2
(L dL)(i di)2 Li
2 2
1
{(L dL)(i2 di2 2idi)
Li2} 2
1
{(L dL)(i2 2idi)
Li 2} 2
1
{Li2 2Lidi i 2dL 2ididL
Li 2} 2
1
{2Lidi
2
i dL} 2 (1.25)
1 2
Lidi i dL
2 (1.26)
Mechanical work to move the pointer by d
Td d
By law of conservation of energy,
Electrical energy supplied=Increase in stored energy+ mechanical work done.
(1.27)
Input energy= Energy stored + Mechanical energy
1 2
Lidi i2dL Lidi i dL Td d
2 (1.28)
1 2
i dL Td d
2
1 2 dL (1.29)
T i
d
2 d
At steady state condition
Td TC
1 2 dL (1.30)
i K
2
d
1 2 dL (1.31)
i
2K d
23
i2 (1.32)
24
Advantages
✓ MI can be used in AC and DC
✓ It is cheap
✓ Supply is given to a fixed coil, not in moving coil.
✓ Simple construction
✓ Less friction error.
Disadvantages
✓ It suffers from eddy current and hysteresis error
✓ Scale is not uniform
✓ It consumed more power
✓ Calibration is different for AC and DC operation
1.8.2 Repulsion type moving iron instrument
Construction:The repulsion type instrument has a hollow fixed iron attached to it (Fig. 1.12). The
moving iron is connected to the spindle. The pointer is also attached to the spindle in supported
with jeweled bearing.
Principle of operation: When the current flows through the coil, a magnetic field is produced by
it. So both fixed iron and moving iron are magnetized with the same polarity, since they are kept
in the same magnetic field. Similar poles of fixed and moving iron get repelled. Thus the deflecting
torque is produced due to magnetic repulsion. Since moving iron is attached to spindle, the spindle
will move. So that pointer moves over the calibrated scale.
Damping: Air friction damping is used to reduce the oscillation.
Control: Spring control is used.
25
Fig. 1.12
Fig. 1.13
26
This instrument can be used for the measurement of voltage, current and power. The difference
between the PMMC and dynamometer type instrument is that the permanent magnet is replaced
by an electromagnet.
Construction:A fixed coil is divided in to two equal half. The moving coil is placed between the
two half of the fixed coil. Both the fixed and moving coils are air cored. So that the hysteresis
effect will be zero. The pointer is attached with the spindle. In a non metallic former the moving
coil is wounded.
Control: Spring control is used.
Damping: Air friction damping is used.
Principle of operation:
When the current flows through the fixed coil, it produced a magnetic field, whose flux density is
proportional to the current through the fixed coil. The moving coil is kept in between the fixed
coil. When the current passes through the moving coil, a magnetic field is produced by this coil.
The magnetic poles are produced in such a way that the torque produced on the moving coil
deflects the pointer over the calibrated scale. This instrument works on AC and DC. When AC
voltage is applied, alternating current flows through the fixed coil and moving coil. When the
current in the fixed coil reverses, the current in the moving coil also reverses. Torque remains in
the same direction. Since the current i1 and i2 reverse simultaneously. This is because the fixed and
moving coils are either connected in series or parallel.
Fig. 1.14
27
Let
L1=Self inductance of fixed coil
L2= Self inductance of moving coil
M=mutual inductance between fixed coil and moving coil
i1=current through fixed coil
i2=current through moving coil
Total inductance of system,
Hence the deflection of pointer is proportional to the current passing through fixed coil and
moving coil.
28
1.9.1 Extension of EMMC instrument
Case-I Ammeter connection
Fixed coil and moving coil are connected in parallel for ammeter connection. The coils are
designed such that the resistance of each branch is same.
Therefore
I1 I2 I
Fig. 1.15
To extend the range of current a shunt may be connected in parallel with the meter. The value
Rsh is designed such that equal current flows through moving coil and fixed coil.
Td I1I2 (1.41)
dM
d
Or Td I2 (1.42)
dM
d
TC (1.43)
K
I2 (1.44)
dM
Kd
Fixed coil and moving coil are connected in series for voltmeter connection. A multiplier may be
connected in series to extent the range of voltmeter.
30
Fig.1.16
Case-III As wattmeter
When the two coils are connected to parallel, the instrument can be used as a wattmeter. Fixed coil
is connected in series with the load. Moving coil is connected in parallel with the load. The moving
coil is known as voltage coil or pressure coil and fixed coil is known as current coil.
Fig. 1.17
31
Assume that the supply voltage is sinusoidal. If the impedance of the coil is neglected in
comparison with the resistance ‘R’. The current,
32
(1.66)
(Td )avg KVI cos
TC (1.67)
VI cos (1.69)
Advantages
✓ It can be used for voltmeter, ammeter and wattmeter
✓ Hysteresis error is nill
✓ Eddy current error is nill
✓ Damping is effective
✓ It can be measure correctively and accurately the rms value of the voltage
Disadvantages
✓ Scale is not uniform
✓ Power consumption is high(because of high resistance )
✓ Cost is more
✓ Error is produced due to frequency, temperature and stray field.
✓ Torque/weight is low.(Because field strength is very low)
Errors in PMMC
✓ The permanent magnet produced error due to ageing effect. By heat treatment, this error
can be eliminated.
✓ The spring produces error due to ageing effect. By heat treating the spring the error can be
eliminated.
✓ When the temperature changes, the resistance of the coil vary and the spring also produces
error in deflection. This error can be minimized by using a spring whose temperature co-
efficient is very low.
1.10 Difference between attraction and repulsion type instrument
An attraction type instrument will usually have a lower inductance, compare to repulsion type
instrument. But in other hand, repulsion type instruments are more suitable for economical
production in manufacture and nearly uniform scale is more easily obtained. They are therefore
much more common than attraction type.
33
INDUCTION TYPE INSTRUMENTS
Consider an aluminum disc placed the between the pole of an electromagnet, as shown in fig. 11.1. Let
the flux produced by flow of current of I Amperes through the coil be F and this flux will lag behind I,
by a small angle β as shown in vector diagram.
Since the aluminum disc act as a short circuited secondary of the transformer, therefore, an e.m.f., (say
e volts) lagging behind the flux F by �radians will be induced in it. As a result of this induced e.m.f.,
the eddy current (I�) starts flowing in the disc. Since the disk is purely resistive therefore the eddy
current will be in phase with induced e.m.f. (e) will lag behind the main flux F by � radians. As the
component of eddy current (I�) along flux F is zero, therefore torque produced is zero. It can be proved
as follows.
Let the instantaneous values of flux and eddy current be given by F = Fmax Sin θ and i = Imax Sin (θ �
α). Where α is the phase angle between the induced eddy current and flux (F).
Instantaneous torque α F i
����������
34
����������
����������
����������
���������� α Fi Cos α
Since in single phase induction type instruments the angle α between main flux F and eddy current I�
is �and Cos �is zero, therefore torque produced is zero. Hence to obtain the resultant torque it is
necessary to produce an eddy current which is either appreciable less than or appreciable more than
�radians, out of phase with the flux which it reacts. Several arrangements are possible but here we will
discuss about the descriptions of the two of these.
As shown in Fig. 11.3, in this method, the working current is passed through the coil of an electromagnet
which has an air gap in one limb. Permanent magnet is used for providing damping torque. The
aluminum disc is mounted on pivots and jewel bearings.
35
Two spiral springs are employed to provide controlling torque, wounded in direction opposite to each
other if the instrument is used as Voltmeter, Ammeter and Wattmeter etc. One half of the pole face is
surrounded by a copper band in order to split the working flux into two different paths. The copper
shading band acts as a single turn short circuited secondary winding of the transformer. The spiral
springs, pointer and scale etc. have been omitted for simplicity.
11.2.1 Theory
Let the total flux produced in the magnetic core be F Weber. Due to shading of pole, this flux will split
up into two fluxes i.e. flux through un-shaded portion and other through the shaded portion. Suppose
the flux F1 be the flux of the shaded portion of the pole. This flux F1 will induce an e.m.f. in the copper
ring, which will lag the flux F1 by 90�, as shown in Fig. 11.4. The induced e.m.f. will force a current
say i to flow in the copper ring which will be lagging behind the flux F1 by 900. The current flowing in
the copper ring will produce its own magnetic field say F�2 in phase with current i. The flux given by
the shaded portion of the pole will be the vector sum of F1 and F�2 which is equal to F2 lagging behind
flux F1 by an angle θ and its value should be 400 to 600 for producing effective deflecting torque.
Let the flux F1 and F2 are the fluxes passing through the shaded and un-shaded portions of the pole
respectively induce e.m.fs. e1 and e2 in the disc, each of which is 900 in phase behind the fluxes
responsible for inducing it. These induced e.m.fs; will induce eddy currents (say i1 and i2) in the disc
lagging by a small angle (say α) behind its voltage due to the inductance of the path in the disc.
From Fig. 11.5, it is obvious that each of the current i1 and i2 has a component in phase with the other
flux such i1′ and i2′. Hence two torques are acting in a directions having angle θ are produced in the
instrument. Resultant of these two torques, provides an operating or deflecting torque.
This method is also known as split phase method. In this method, two laminated magnets A and B are
placed near to each other with aluminum (Al) disc in between and a non inductive resistance R is
connected in series with the magnetizing coil of magnet A and an inductive coil L is connected in series
with the magnetizing coil of magnet B, as shown in Fig. 11.6.
36
Fig. 11.6 Two pole method (split phase)
Hence there will be two fluxes having phase difference of less than 90� with each other, acting on the
disc which will produce a resultant torque in the aluminium disc.
Let the flux produced by the magnet A and B is F1 and F2 respectively. F2 is lagging F1 by an angle θ as
shown in Fig. 11.5. Hence an operating or deflecting torque will be produced as explained above in case
of shaded pole method.
A volt meter is an instrument used to measure the potential difference between the two points in an
electric circuit. In analog voltmeters, the pointer moves over a calibrated scale in proportion to potential
difference across the points where as in case of digital voltmeters, it displays numerical values with the
help of analog to digital converter. The induction type voltmeter operates on the either shaded pole
method or on two pole method�s working principle as explained in Lesson 11.
An ammeter is always connected in series with load current directly or through CT (Current
Transformer). As shown in Fig. 12.2, both the windings on the two laminated electromagnets A and B
are connected in series but winding is shunted by a resistance R with the result of which, the current in
this winding lags with respect to the total current (I). Hence the necessary phase angle (α) required
between two fluxes is produced by the laminated electromagnets A and B.
The operating principle of the induction type instrument is based on the two pole method as discussed
in Lesson 11. Two fluxes produced by laminated magnet A and B are focused upon the aluminum disc,
having a phase angle between them required for producing a resultant torque in the spindle of the moving
system. Being a spring control based controlling torque, the scale is uniform and the deflecting torque is
directly proportional to square the load current. Eddy current damping is used to provide necessary
damping torque by a permanent magnet. Spiral springs, pointer and damping magnets are omitted for
clear understanding of the Fig. 12.2.
Advantages
(a) Damping is very much effective and efficient.
(b) Full scale deflection more than 200� can be obtained.
Disadvantages
(a) Power consumption is large and hence not recommended where continuous monitoring of ac
quantities is required.
38
(b) Variation in temperature and frequency may cause serious errors if necessary compensations
are not provided.
(c) As these instruments are based on principle of induction, they can be used on AC supply
only.
Variation in frequency causes serious errors because deflecting torque is directly proportional to
frequency and also the value of impedance (Z) and Cos α depends upon the supply frequency. The error
is compensated by use of non inductive shunt in case of an Ammeter, when the frequency increases, the
increase in impedance of the winding cause a greater proportion of the total current to flow in the non
inductive shunt (whose impedance remains constant for all frequency) and lesser proportion of the total
current in flow in the winding and to an extent thus compensate the increase in torque (since T α. f).
In case of voltmeter, the impedance of the winding increases with the increase in frequency, hence
smaller current is drawn by the winding, which tends to compensate the increase in torque due to increase
in frequency.
Variation in temperature changes the resistance of the eddy current paths, therefore, may result in serious
errors. The error is compensated in case of an ammeter, employing a shunt of material having a high
temperature coefficient of resistance than the material of the disc. This shunt may be the same one as
used for frequency compensation. When the temperature increases, the resistance of the shunt increases,
hence the greater portion of the current flows through the coil and decreases in torque due to smaller
eddy current in the disc owing to increase in resistance at high temperature is compensated. The
combination of shunt and swapping resistance in series with the instrument is often employed to
compensate the temperature error in case of voltmeters. Since the frequency errors in induction type
instruments are so serious that cannot be compensated satisfactorily. Hence these instruments are used
for only constant frequency supplies or where the fluctuation in frequency is very small.
INDUCTION TYPE WATTMETER, WATT-HOUR METER, AND DYNAMOMETER TYPE
POWER FACTOR METER
These types of watt-meters operate on the same working principle on which the induction type ammeter
and voltmeter operates. These instruments can only be used on ac supply while dynamo-meter type watt
meters can be used on either ac or dc supply system. Induction type watt-meters are useful only when
the supply and frequency remains constant. Since both the coils i.e. current coil and pressure coils are
necessary in such instrument, it is not essential to use shaded pole principle. Because for producing a
39
deflecting torque, two fluxes are essential with suitable phase angle and it would be available from these
two coils.
13.1.1 Construction
A watt-meter has two laminated electromagnet, one of which is excited by load current or definite
fraction of it, and is connected in series with the circuit, known as series magnet and the other is excited
by the current proportional to the applied voltage or fraction of it and is always connected across the
supply, known as shunt magnet. An aluminum disc is so mounted so that it cuts the fluxes produced by
both the magnets. As a result of which, two e.m.f�s are produced which induces two eddy currents in
the disc. C - Magnet is used to provide necessary damping torque to the pointer, to damp out the
oscillations. Deflecting torque is produced due to interaction of these eddy currents and the inducing
flux. Copper shading bands are provided either on central limb or on the outer limb of the shunt magnet,
and can be so adjusted as to make the resultant flux in the shunt magnet lag behind the applied voltage
by 90�. Both the watt-meters are provided with spiral springs A and B, for producing controlling torque
to counter balance the deflecting torque. In Fig. 13.2 the spiral spring and damping magnet is omitted
for simplicity. The scale of such type instruments is quite uniform and extends over an angle of 300�.
Currents up to 100 A can be handled by these watt-meters directly where as beyond this current
transformers are used. Two types of induction type watt meters are available. Line diagrams of both of
the types are detailed in Fig. 13.1 and 13.2.
In the form of the instrument shown in Fig. 13.1, two pressure coils are connected in series in such a way
that both of them send flux through the central limb. The series magnet also carries two small current
coils connected in series and wound so that they magnetized their respective cores in the same direction.
Correct phase displacement between the fluxes produced by series and shunt magnet is obtained by the
adjustment of copper shading band on the central limb.
40
Fig. 13.2 Induction type wattmeter
In Fig. 13.2, there is only one pressure and one current coil. Two projecting poles of shunt magnet are
surrounded by a copper shading band whose position can be adjusted for correcting the phase of the flux
of this magnet with the applied voltage. The pressure coil circuit of induction type instrument is made
as inductive as possible so that the flux of the shunt magnet may lag nearly by 90 degree behind the
applied voltage.
13.1.2 Advantages
The advantages of induction watt meters are the same as those of induction ammeters long scale,
freedom from effects of stray field, and have effective damping torque.
13.1.3 Disadvantages
A watt hour meter is used to sum up the total energy consumed by a consumer during a period so that it
can be charged for the actual energy consumed. The working principle, theory and advantage /
disadvantages are almost similar to single phase watt meter. The construction of single phase watt hour
meter is also almost similar to single phase induction type watt meter as discussed above. The pointer
41
and spiral springs are replaced by wheel-train mechanism for summing up of total energy consumed
where as the damping magnet is replaced by braking magnet. The construction of this type of watt hour
meter is shown in Fig.
The brake magnet and recording wheel-train being omitted for clear understanding of the diagram. The
description of registering mechanism and braking system is detailed below.
The registering or counting system essentially consists of gear train, driven either by worm or pinion
gear on the disc shaft, which turns pointers that indicate on dials the number of times the disc has turned.
The energy meter thus determines and adds together or integrates all the instantaneous power values so
that total energy used over a period is thus known. Therefore, this type of meter is also called
an integrating meter.
Braking of the disk is provided by a small permanent magnet, located diametrically opposite to the
alternating current magnets. The disk moves between the magnets gaps. The movement of rotating disc
through the magnetic field crossing the air gap sets up eddy currents in the disc that reacts with the
magnetic field and exerts a braking torque. By changing the position of the brake magnet or diverting
some of the flux therefore, the speed of the rotating disc can be controlled. Creep error can be rectified
by drilling a small hole in the aluminum disc passing through the magnetic flux of braking magnet.
It is necessary that the energy meter should give correct reading on all power factors, which is only
possible when the field set up by shunt magnet flux lags behind the applied voltage by 90 degree.
42
Ordinarily the flux set up by shunt magnet does not lag behind the applied voltage exactly by 90 degree
because of winding resistance and iron losses. The flux due to shunt magnet is made to lag behind applied
voltage by 90 degree with the help of copper shading band provided on the central limb. An error due to
incorrect adjustment of shading band will be evident when the meter is tested on a load of power factor
less than unity.
An error on the fast side under these conditions can be eliminated by bringing the shading band nearer
to the disc and vice versa. An error in the speed of the meter when tested on non inductive load can be
eliminated by adjustment of the position of the brake magnet. Movement of the brake magnet in the
direction of the spindle will reduce the braking torque and vice versa. Speed of disc is directly
proportional to the distance between the disc and brake magnet.
The two shading bands embrace the flux contained in the two outer limbs of the shunt electromagnet,
and thus eddy current are induced in them which cause a phase displacement between the enclosed flux
and main gap flux. As a result, a small driving torque is exerted on the disc, this torque being adjusted,
by variation of the position of these bands, to compensate for frictional torque in the instrument.
In some energy meter, it is observed that the disc continue to rotate even when the load on the energy
meter is zero and potential coil is in excited condition. This defect is known as creeping and is prevented
by cutting two holes or slots in the disc on opposite sides of the spindle. The disc tends to remain
stationary when one of the holes comes under one of pole of the shunt magnet. In some cases, a small
piece of iron wire is attached to the edge of the disc. The force of attraction of the brake magnet upon
this wire is sufficient to prevent continuous rotation of the disc under no load condition.
The error due to variation in temperature is very small. Since the various effects due to change in
temperature tends to neutralize each other on unity power factor if not on low power factor (lagging).
Since the meters are used normally on fixed frequency and hence these can be adjusted to have a
minimum error at declared supply frequency which is normally 50 cycles / second.
The power factor meter is used to indicate the instantaneous power factor of the consumer. It consist of
two fixed coils CC connected in series carrying the load current (or a definite fraction of it) and two
identical moving coils P1 & P2 wound with a fine copper wire, fixed at right angle to each other and
pivoted on the same spindle. The pressure coils P1 and P2 move together and carry a pointer, which
indicates the power factor of the circuit directly on the scale.
43
Fig. Dynamometer type power factor meter
The pressure coil P1 is connected across the supply through a non inductive resistance R and pressure
P2 is connected across the supply through a highly inductive choke coil of inductance L. The value of
non inductive resistance R and inductance L are so chosen that for the normal frequency, the current in
the two pressure coil P1 and P2 is same. Thus these coils P1 and P2 produce equally strong magnetic field
displaced by 900 in space as well as in the phase. For measurement of power factor on high voltage
system, the current and pressure coils of the instrument may be connected to the main circuit through
current and potential transformer respectively.
13.4.1 Theory
While measuring power factor of an installation, there may be three possibilities of installations power
factor, which are described here:
(a) Power Factor is Unity: When the circuit is switched on, the current in the potential coil P1 will
be in phase with current in coils CC, where as the current in pressure coil P2 will lag 900 behind the
voltage or behind the current in the circuit coli CC. Thus pressure coil P1 will experience a turning
moment so its plane will come in a position parallel to a plane of a current coil CC. The average
torque on coil P2 will be zero but being mechanically coupled to coil P1, it will follow the rotation
of coil P1. Hence the pointer will in the centre of the calibrated scale and it will show the power
factor as unity. The position of coil P1 is shown in Fig. and it will maintain the reading till the load
current is in phase with the voltage.
(b) When Power Factor is Zero (lagging): In this situation, the current flowing in the pressure coil
P2 will be in phase with load current flowing in the fixed current coil CC, both lagging behind the
applied circuit voltage by 900 and current in pressure coil P1 will lead the load current in current coil
CC by 900. Thus only pressure coil P2 will experience a turning moment so its plane will come in a
position parallel to the plane of current coils CC. At this instant, the pointer will indicate zero power
factor lagging.
(c) When Power Factor Zero (leading): When the current flowing in fixed coils CC leads the
applied voltage by 90 degree and, therefore, the field of pressure coils P1 by 90 degree and that of
coil P2 by 180 degree. Hence the polarity of field in current coils is the reverse of that considered
44
above. At this instant, the pointer will indicates the power factor as zero leading on the other half of
the scale.
For an intermediate power factor, the moving system takes up intermediate position and the pointer
makes an angle of (90 degree - F) with the axis of the fixed coils where F the phase angle between load
current is and applied voltage of the load circuit.
Errors in wattmeter and energy meter and their compensation and adjustment
Error
due to connection method
A suitable value capacitor connected in parallel
Error due to pressure coil inductance
with pressure coil.
This error can be reduced by designing pressure
coil circuit such that inductive reactance of the
Error due to Pressure Coil Capacitance
circuit matches exactly with the capacitance
reactance of the circuit i.e. XL=XC.
This error can be reduced by proper design of
pressure coil and current coil system so that they
Error due to mutual inductance effect
always remain in a zero position of mutual
inductance.
To avoid this error, magnetic shield is placed over
Error due to stray magnetic fields
CC & PC.
These are minimized by avoiding solid metal
Error due to eddy currents
parts and using laminated core.
Using zero temperature coefficient materials for
Temperature error
coils and components, this can be minimized.
It is avoided by designing the moving system
Error due to vibration of moving system such that its natural freq is greater than 2 times
the freq of deflecting torque of the wattmeter.
The weight of moving system be reduced to
Error due to friction
minimum possible.
Phantom Loading
45
Definition: Phantom loading is the phenomena in which the appliances consume electricity even
when they turn off. The disc of the energy meter rotates which increases the reading of the meter, but
the devices do not consume power. This type of loading is also known as the vampire or virtual
loading. The phantom loading mainly occurs in the “electronic” appliances.
The phantom loading is used for examining the current rating ability of the energy meter. The actual
loading arrangement will waste a lot of power. The phantom loading consumes very less power as
compared to real loading, and because of this reason, it is used for testing the meter.
In phantom loading, the pressure coil and the current coil are separately excited by the supply source.
The pressure coil is energised from the small supply voltage, and the current energises the current coil
at very small voltages.
The pressure and current coil circuit have low impedance (less obstruction of movement of the
electron) because of which highly rated current is passed through it. The total current supplied for the
phantom loading is the sum of the pressure coil current which is supplied at normal voltage and the
current of the current coil supply at low voltages.
Consider the DC energy meter having rating voltage 220V and current 9 Ampere. The resistance of the
pressure coil and the current coil is 4400Ω and 0.1Ω respectively. The power consumption of the load
by direct and indirect phantom is explained below.
below.
46
The total power consumed by the pressure and current circuit
P = (220)2/4400 = 11watt
The current coil of the phantom loading arrangement is separately excited by the battery of the 9V. The
power of the current coil is measured as
Power = 9 Χ 9 = 81watt
The above example shows that in phantom loading the pressure and the current coil is separately
excited by the meter. Hence the power loss is less in phantom loading as compared to direct loading.
47
Unit -2
Polyphase Metering
Blondel’s Theorem
Blondel's theorem tells about the number of watt-meters required to measure three-phase power.
It states that, in order to measure power in a network with n number of lines. The total number of
watt-meters required is equal to n, and total power is the sum of all the watt-meters readings. It is in
such condition that, if current coils of each wattmeter are connected in each line and corresponding
voltage coils are connected such that, one end to their respective line and other ends of all the voltage
coils are connected together forming a common point.
Suppose, if the common point is to be taken on any one of the lines. Then the other end of the voltage
coils is connected to that common line (i.e., common point). In such conditions, the power can be
measured by (n-1) watt-meters. Thus for measuring 3-phase power, only 2 watt-meters are required,
this is called Two Wattmeter Method.
The pressure coil of all the three wattmeters namely W1, W2 and W3 are connected to a common
terminal known as the neutral point. The product of the phase current and line voltage represents
phase power and is recorded by an individual wattmeter.
The total power in a three wattmeter method of power measurement is given by the algebraic sum
of the readings of three wattmeters. i.e.
Where,
W1 = V1I1
W2 = V2I2
W3 = V3I3
Except for 3 phase, 4 wire unbalanced load, 3 phase power can be measured by using only Two
Wattmeter Method.
have .
Diagram is shown below:
Two switches are given which are marked as 1-3 and 1-2, by closing the switch 1-3 we get reading
of wattmeter as
PF PF W1 W2 W3ph=W1+ Remarks
angle W2
φ (lag) cos φ VLILcos(30- VLILcos(30+ √3VLILcosφ Gen. Case (always
φ) φ) W1≥W2)
00 UPF √3/2 VLIL √3/2 VLIL 2W1 or 2W2 W1=W2
300 0.86 VLIL VLIL/2 1.5W1 or W2=W1/2
6 3W2
600 0.5 √3/2 VLIL ZERO W1 alone W2 reads zero
For taking readings, the PC
or
>600 W1 W2 reads W1+(-W2)
<0.5 CC connection of W2
should
negative be reversed) (LPF case)
Instrument Transformers:
Instrument Transformers
Transformers used in conjunction with measuring instruments for measurement purposes
are called “Instrument Transformers”.
The instrument used for the measurement of current is called a “Current Transformer” or
simply “CT”.
The transformers used for the measurement of voltage are called “Voltage transformer”
or “Potential transformer” or simply “PT”.
Fig 1. indicates the current measurement by a C.T. The current being measured passes
through the primary winding and the secondary winding is connected to an ammeter.
The C.T. steps down the current to the level of ammeter.
Fig 2. shows the connection of P.T. for voltage measurement. The primary winding is
connected to the voltage being measured and the secondary winding to a voltmeter.
The P.T. steps down the voltage to the level of voltmeter.
Merits of Instrument Transformers:
1. Instruments of moderate size are used for metering i.e. 5A for current and 100
to 120 volts for voltage measurements.
2. Instrument and meters can be standardized so that there is saving in costs.
Replacement of damaged instruments is easy.
3. Single range instruments can be used to cover large current or voltage ranges,
when used with suitable multi range instrument transformers.
4. The metering circuit is isolated from the high voltage power circuits. Hence
isolation is not a problem and the safety is assured for the operators
5. There is low power consumption in metering circuit.
6. Several instruments can be operated from a single instrument
Nominal Ratio: It is the ratio of rated primary winding current (voltage) to the rated
secondary winding current (voltage).
The rated burden is the volt ampere loading which is permissible without errors
exceeding the particular class of accuracy.
2.2. Current Transformer equivalent circuit:
2.5. Construction of CT
Construction of Current Transformer:
Current transformers are constructed in various ways. In one method there are two
separate windings on a magnetic steel core. The primary winding consists of a few turns
of heavy wire capable of carrying the full load current while the secondary winding
consist of many turns of smaller wire with a current carrying capacity of between 5/20
amperes, dependent on the design. This is called the wound type due to its wound primary
coil.
Potential transformers are normally connected across two lines of the circuit in which the
voltage is to be measured. Normally they will be connected L-L (line-to-line) or L-G (line-
to-ground). A typical connection is as follows:
2.8. Phasor Diagram of Potential Transformer:
The theory of a potential transformer is the same as that of a power transformer. The
main difference is that the power loading of a P.T. is very small and consequently the
exciting current is of the same order as the secondary winding current while in a power
transformer the exciting current is a very small fraction of secondary winding load
current.
Unit 3
Potentiometers
A potentiometer is an instrument designed to measure an unknown voltage by comparing it
with a known voltage. The known voltage may be supplied by a standard cell or any known
voltage. At null condition, no current flows, so no power consumed. So the measurement is
independent of source resistance.
It can also measure current by measuring voltage drop across a standard resistor.
It is used extensively for calibration of voltmeters & ammeters.
Measurements using comparison methods are capable of a high degree of accuracy because the
result obtained does not depend upon the actual deflection of a pointer, as is the case in
deflection methods, but only upon the accuracy with which the voltage of the reference source
is known.
CONSTRUCTION:-
Modern Lab Potentiometers use calibrated dial resistors and a small circular wire of one or
more turns (instead of previously used long slide wire), thus reducing the size of instrument.
In the figure, there is one dial resistor with 15 steps, each having a precision resistor of 10 Ω.
Total 150 Ω & 1.5 volt.
Slide-Wire:- is of single turn having resistance of 10 Ω & 0.1 Volt. The working current is 10
mA.
So each step of dial-resistor correspond to 10*10* 10-3 =0.1 volt
The slide wire is provided with 200 scale divisions with a total voltage range of 0.1 volt.
So each division=0.1/200=0.005 volt
The Potentiometer is provided with a double throw switch to make connection to either the
standard cell or un-known emf. To operate the Galvanometer at its
maximum sensitivity, provision is made to short the protective resistance near balance
condition.
STEPS TO FOLLOW TO MAKE MEASUREMENT:-
1. The combination of dial-resistor & Slide-wire is set to standard cell voltage (1.0186 volt).
Dial-resistor set at 1 volt, slide- wire set at 0.0186 V.
2. Galvanometer is calibrated with switch S, connected to standard cell. Rheostat is adjusted
foe zero Galvanometer deflection.
3. As the null point is approached, protective resistance is shorted to increase Galvanometer
sensitivity. This completes the process of ‘Standardization’ of Galvanometer.
4. Now switch S is thrown to operate position to find out un-known emf. Now the potentiometer
is balanced using dial- switch & slide-wire.
5. At null, the value of emf is read directly from setting of dial adjust & slide-wire.
6. Standardization is checked again.
VERNIER POTENTIOMETER: -
CONSTRUCTION:-
There are 3 measuring dials.
1 st Dial Measures upto 1.5 v in step of 0.1 v
2 nd Dial has 102 studs & read upto 0.1 V in steps of 0.001 V.
3 rd Dial has 102 studs & reads from -0.0001 v to + 0.0001 v in steps of 0.00001 v (i.e. 10 µv).
There is no Slide-wire.
The 2nd Dial shunts two of the coils of 1st Dial. The moving arm of 2nd -dial carries two arms
spaced two-studs apart.
In practice, the resistance of 2nd -dial is greater than that between two studs in the main dial,
so that voltage drop across 2nd -dial is greater than 0.1 v. If this is not done, voltage drop in
contact resistances & leads would cause 2nd -dial voltage less than 0.1 v.
3 rd -dial is obtained from a shunt ckt which permits a true zero & a small –ve setting is
obtained.
OPERATION:-
The limitations imposed on performance of ordinary potentiometers by slide-wire are
eliminated in a vernier potentiometer. This instrument has two ranges.
1. Normal range of 1.6 v down to 10 µv
2. Lower range of 0.16 v to 1 µv.
The Vernier potentiometer reads to increment of 0.00001 v (10 µv) & has a readability of 1 µv
on 0.1 range. If a 3rd range of * 0.01 is provided, readability becomes 0.1 µv
AC Potentiometer: A potentiometer is an instrument which measures unknown voltage by
balancing it with a known voltage. The known source may be DC or AC. The working
phenomenon of DC potentiometer and AC potentiometer is same. But there is one major
difference between their measurements, DC potentiometer only measures the magnitude of the
unknown voltage. Whereas an AC potentiometer measures both the magnitude and phase of
unknown voltage by comparing it with a known reference.
There are two types of AC potentiometers:
1. Drysdale-Tinsley Potentiometer/ Polar type potentiometer
2. Gall-Tinsley Potentiometer / Coordinate type potentiometer.
Polar type Potentiometer : In such type of instruments, two separate scales are used to
measure magnitude and phase angle on some reference of the unknown e.m.f. There is a
provision on the scale that it could read phase angle up to 3600. It has electrodynamometer
type ammeter along with DC potentiometer and phase-shifting transformer which is operated
by single phase supply.
In a phase-shifting transformer, there is a combination of two ring-shaped laminated steel
stators connected perpendicularly to each other as shown in the figure. One is directly
connected to power supply and the other one is connected in series with variable resistance and
capacitor. The function of the series components is to maintain constant AC supply in the
potentiometer by doing small adjustments in it.
Between the stators, there is laminated rotor having slots and winding which supplies voltage
to the slide-wire circuit of the potentiometer. When current start flowing from stators, the
rotating field is developed around the rotor which induces an e.m.f. in the rotor winding.
Therefore, resultant induced emf in the rotor winding due to two stator winding
In both the cases, the reading of Voltmeter and Ammeter is taken. If the Voltmeter reading is
V and Ammeter reading is I then the measured Resistance will be
Rm = V/I
This measured Resistance Rm will be the true value of the Resistance if and only if the
Resistance of Ammeter is zero and that of Voltmeter is infinite. But actually this is not possible
to achieve zero resistance Ammeter and infinite Resistance Voltmeter. Therefore measured
value of resistance Rm will deviate from the true value R (Say).
So we will discuss both the circuit individually and will calculate the percentage error in the
measurement.
Case1:
We consider first kind of connection as shown in figure 1 above. It is clear from the figure that
Voltmeter is measuring the Voltage drop across the Ammeter as well as resistor.
So V = Va + Vr
Let current measured by Ammeter = I
Therefore, measured Resistance Rm = V/I
So, Rm = (Va+Vr) / I =(IRa+IR) / I = Ra+R
Therefore, the measured Resistance is the sum of Resistance of Ammeter and true Resistance.
Therefore measured value will only represent true value if Ammeter Resistance Ra is Zero.
True value of Resistance R = Rm –Ra
= Rm(1-Ra/Rm)
Relative Error = (Rm-R)/R = Ra/R
Therefore, Relative Error will be less if the true value of Resistance to be measured is high as
compared to the internal Resistance of Ammeter. That’s why this method should be adopted
when measuring high Resistance but it should be under Medium Resistance category.
Case2:
We will consider second connection in which Voltmeter is connected in which Voltmeter is
connected toward Resistance R whose value is to be measured.
It is obvious from figure that Ammeter will read the current flowing through the Voltmeter and
Resistance R. Therefore current measured by Ammeter Ia = Iv+Ir
So, Ia = Iv+Ir
= V/Rv+V/R where Rv is Resistance of Voltmeter and V is Voltmeter reading.
Measured Resistance Rm = V/Ia
= V/(V/Rv+V/R)
= RvR/(R+Rv)
= R/(1+R/Rv) ….Dividing Numerator and Denominator by Rv
Therefore, true value of Resistance R = RmRv/(Rv-Rm)
= Rm(1-Rm/Rv)
Therefore, true value of Resistance will only be equal to measured value if the value
of Voltmeter Resistance Rv is infinite.
If we assume that the value of Voltmeter Resistance Rv is large as compared to the Resistance
to be measured R, then Rv>>>Rm
So, True value R = Rm(1+Rm/Rv)
Thus from the above equation it is clear that the measured value of Resistance is smaller than
the true value.
Relative Error = (Rm-R)/R
= -R/Rv
Therefore, it is clear from the expression of Relative Error that, error in measurement will be
low if the value of Resistance under measurement is very less as compared to the internal
Resistance of Voltmeter.
This is the reason; this method is used for the Contact Resistance Measurement. As the value
of Contact Resistance is of the order of 20 micro Ohm which is very less as compared to the
internal Resistance of Voltmeter.
The Voltmeter Ammeter Method for Cases1 and Case2 are simple method but it is not accurate
method. The error in the value of Resistance depends on the accuracy of Ammeter as well as
Voltmeter. If the accuracy of both the instrument are supposed 0.5% then when both the
instrument read near full scale, the error in measurement of Resistance may vary from 0 to 1%
while if both the instrument read near half scale then error may double and so on.
However this method is very useful where high accuracy is not required. The suitability of
Case1 or Case2 depends on the value of Resistance to be measured. The division point between
the two methods is at the Resistance for which both the method give same Relative Error.
So, Ra/R = R/Rv
R=
4.1.2 Measurement of Medium Resistance by Substitution Method
In Substitution Method, the Resistance whose value is to be measured is compared with the
Standard Resistance by some technique which is described in this section. The connection
diagram for Substitution Method is given below.
Here, R is the unknown Resistance, S the Standard variable Resistance, A is Ammeter and r is
Regulating Resistance.
When we put the Switch at position 1 then R is connected in the circuit. The Regulating
Resistance r is adjusted till the reading of Ammeter is at a chosen scale mark. Now the Switch
is thrown to position 2 putting the Standard variable Resistance S in the circuit. Now the
variable Resistor S is adjusted till the reading of Ammeter is same as when R was in the circuit.
The setting of dial of S is read. Since the substitution of one resistance for another has left
current unaltered, and provided that EMF of battery and position of Regulating Resistance r
remain unaltered, the two Resistance R and S must be equal. Thus the value of unknown
Resistance R is equal to the dial setting of Standard Resistance S.
This method of measurement is more accurate as compared to the Ammeter Voltmeter Method
as in this method measurement is not affected by the accuracy of Ammeter. However, the
accuracy of this method is greatly affected if there is any change in the Battery EMF during
the time when the reading in two settings is taken. Therefore to avoid the error because of
change of EMF of Battery, a Battery of enough capacity is used so that it remains constant
during the entire period of testing.
The accuracy of this method also depend on resistance of circuit excluding R and S, upon the
sensitivity of instrument and upon the accuracy with which the Standard Resistance S is known.
This method is not widely used for simple Resistance measurement and is used in modified
form for the measurement of High Resistance. The Substitution Method is however very
important as it finds its use in application of bridge method and in high accuracy A.C
measurement.
4.1.3 WHEATSTONE BRIDGE
For measuring accurately any electrical resistance Wheatstone bridge is widely used. There are
two known resistors, one variable resistor and one unknown resistor connected in bridge form
as shown below. By adjusting the variable resistor the current through the Galvanometer is
made zero. When the current through the galvanometer becomes zero, the ratio of two known
resistors is exactly equal to the ratio of adjusted value of variable resistance and the value of
unknown resistance. In this way the value of unknown electrical resistance can easily be
measured by using a Wheatstone Bridge. THEORY The general arrangement of Wheatstone
bridge circuit is shown in the figure below. It is a four arms bridge circuit where arm AB, BC,
CD and AD are consisting of electrical resistances P, Q, S and R respectively. Among these
resistances P and Q are known fixed electrical resistances and these two arms are referred as
ratio arms. An accurate and sensitive Galvanometer is connected between the terminals B and
D through a switch S2. The voltage source of this Wheatstone bridge is connected to the
terminals A and C via a switch S1 as shown. A variable resistor S is connected between point
C and D.
The potential at point D can be varied by adjusting the value of variable resistor. Suppose
current I1 and current I2 are flowing through the paths ABC and ADC respectively. If we vary
the electrical resistance value of arm CD the value of current I2 will also be varied as the
voltage across A and C is fixed. If we continue to adjust the variable resistance one situation
may comes when voltage drop across the resistor S that is I2.S is becomes exactly equal to
voltage drop across resistor Q that is I1.Q. Thus the potential at point B becomes equal to the
potential at point D hence potential difference between these two points is zero hence current
through galvanometer is nil. Then the deflection in the galvanometer is nil when the switch S2
is closed.
Now, from Wheatstone bridge circuit
And
Now potential of point B in respect of point C is nothing but the voltage drop across the resistor
Q and this is
Again potential of point D in respect of point C is nothing but the voltage drop across the
resistor S and this is
A Wheatstone bridge is a fairly convenient and accurate method for measuring resistance.
1. Discrepancies between the true and marked values of resistances of the three known arms
can introduce errors in measurement.
2. Inaccuracy of the balance point due to insufficient sensitivity of the galvanometer may result
in false null points.
3. Bridge resistances may change due to self-heating (I 2R) resulting in error in measurement
calculations.
4. Thermal emfs generated in the bridge circuit or in the galvanometer in the connection points
may lead to error in measurement.
5. Errors may creep into measurement due to resistances of leads and contacts. This effect is
however, negligible unless the unknown resistance is of very low value.
6. There may also be personal errors in finding the proper null point, taking readings, or during
calculations.
Thermal emfs in the bridge arms may cause serious trouble, particularly while measuring low
resistances. Thermal emf in galvanometer circuit may be serious in some cases, so care must
be taken to minimize those effects for precision measurements. Some sensitive galvanometers
employ all-copper systems (i.e., copper coils as well as copper suspensions), so that there is no
junction of dissimilar metals to produce thermal emf. The effect of thermal emf can be balanced
out in practice by adding a reversing switch in the circuit between the battery and the bridge,
then making the bridge balance for each polarity and averaging the two results.
Figure Guard circuit for measurement of high resistance: (a) Circuit without guard (b) Circuit
with guard
3. Due to electrostatic effects, stray charges may be induced in the measuring circuit.
Flow of these stray charges can constitute a current that can be comparable in magnitude with
the low value current under measurement in high resistance circuits.
This may thus, cause errors in measurement. External alternating electromagnetic fields can
also affect the measurement considerably. Therefore, the measuring circuit needs to be
carefully screened to protect it against such external electrostatic or electromagnetic effects.
4. While measuring insulation resistance, the test object often has considerable amount of
capacitance as well. On switching on the dc power supply, a large charging current may flow
initially through the circuit, which gradually decays down. This initial transient current may
introduce errors in measurement unless considerable time is provided between application of
the voltage supply and reading the measurement, so that the charging current gets sufficient
time to die down.
5. High resistance measurement results are also affected by changes in temperature, humidity
and applied voltage inaccuracies.
6. Reasonably high voltages are used for measurement of high resistances in order to raise the
current to substantial values in order to be measured, which are otherwise extremely low. So,
the associated sensitive galvanometers and micro-ammeters need to be adequately protected
against such high voltages.
Taking these factors into account, the most well-known methods of high resistance
measurements are (i) direct deflection method, (ii) loss of charge method, and (iii)
megohmmeter or meggar.
4.3.2 Direct Deflection Method for High Resistance Measurement
The direct deflection method for measuring high resistances is based on the circuit described
in Figure, which in effect is the voltmeter-ammeter method. For measurement of high
resistances, a sensitive galvanometer is used instead of a micro ammeter as shown in Figure.
A schematic diagram for describing the direct deflection method for measurement of insulation
resistance of a metal sheathed cable is given in Figure.
Figure.
Figure: Loss of charge method for measurement of high resistance
In Above Figure, the unknown resistance R to be measured is connected across the capacitor
C and their parallel combination is connected to the dc voltage source.
Let the capacitor is initially charged up to a voltage of V while the switch is kept ON.
Once the switch is turned OFF, the capacitor starts to discharge through the resistance R.
During the discharge process, the voltage v across the capacitor at any instant of time t is given
by Thus, the insulation resistance can be calculated as With known value of C and recorded
values of t, V and v, the unknown resistance R can be estimated using (24).
The pattern of variation of voltage v with time is shown in Figure.
Capacitor discharge pattern Great care must be taken to record the voltages V and v and also
the time t very precisely, otherwise large errors may creep in to the calculation results.
This method, though simple in principle, require careful choice of the capacitor. The capacitor
C itself must have sufficiently high value of its own leakage resistance, at least in the same
range as the unknown resistance under measurement. The resistance of the voltmeter also needs
to be very high to have more accurate results.
4.3.4 Megohmmeter, or Meggar, for High Resistance Measurement
One of the most popular portable type insulation resistance measuring instruments is the
megohmmeter or in short, meggar. The meggar is used very commonly for measurement of
insulation resistance of electrical machines, insulators, bushings, etc. Internal diagram of a
meggar is shown in Figure.
The traditional analog deflecting-type meggar is essentially a permanent magnet crossed-coil
shunt type ohmmeter.
The instrument has a small permanent magnet dc generator developing 500 V dc (some other
models also have 100 V, 250 V, 1000 or 2500 V generators). The generator is hand driven,
through gear arrangements, and through a centrifugally controlled clutch switch which slips at
a predefined speed so that a constant voltage can be developed. Some meggars also have
rectified ac as power supply.
At balance condition,
I1 z2
=
I2 z1
I3 z4
=
I4 z3
Summary
For balance condition,
The choke for which R1 and L1 have to measure connected between the points ‘A’ and ‘B’.
In this method the unknown inductance is measured by comparing it with the standard inductance.
Disadvantages
Disadvantages
For measuring chokes with higher value of Q-factor, the value of C4 and R4 should be
higher. Higher values of standard resistance are very expensive. Therefore this bridge cannot be
used for higher value of Q-factor measurements.
5.1.2 Hay’s bridge
Disadvantages
Such that the magnetic fields are additive, the resultant inductor of these two can be calculated as
Now if the connections of any one of the coils is reversed then we have
Thus the mutual inductor of the two coils connected in series is given by one-fourth of the difference
between the measured value of self inductor when taking the direction of field in the same direction and
value of self inductor when the direction of field is reversed.
However, one needs to have the two series coils on the same axis in order to get most accurate result.
Let us consider the circuit of Heaviside mutual inductor bridge, given below,
Main application of this bridge in industries is to measure the mutual inductor in terms of self inductance.
Circuit of this bridge consists of four non inductive resistors r1, r2, r3 and r4 connected on arms 1-2, 2-3,
3-4 and 4-1 respectively. In series of this bridge circuit an unknown mutual inductor is connected. A
voltage is applied to across terminals 1 and 3. At balance point electric current flows through 2-4 is zero
hence the voltage drop across 2-3 is equal to voltage drop across 4-3. So by equating the voltage drops
of 2-4 and 4-3 we have,
Also we have,
Now let us consider the circuit of Campbell’s Heaviside bridge given below:
This is the modified Heaviside bridge. This bridge is used to measure the unknown value of self inductor
in terms of mutual inductance.The modification is due to addition of balancing coil l, and R in arm 1 –
4 and also electrical resistance r is included in arm 1-2. Short circuit switching is connected across r2 and
l2 in order to have two sets of readings one while short circuiting r2 and l2 and other while open circuiting
r2 and l2.
Now let us derive the expression for self inductor for this modified Heaviside bridge. Also let us assume
that the value of M and r with switch open be M1 and r1, M2 and r2 with switch closed.
For open switch, we have at balance point,
Battery is applied between terminals marked as 1 and 4. The arm 1-2 consists of capacitor c1 (whose
value is unknown) which carries current i1 as shown, arm 2-4 consists of pure resistor (here pure
resistor means we assuming it non inductive in nature), arm 3-4 also consists of pure resistor and arm
4-1 consists of standard capacitor whose value is already known to us.
Let us derive the expression for capacitor c1 in terms of standard capacitor and resistors.
At balance condition we have,
In order to obtain the balance point we must adjust the values of either r3 or r4 without disturbing any
other element of the bridge. This is the most efficient method of comparing the two values of capacitor
if all the dielectric losses are neglected from the circuit.
Now let us draw and study the phasor diagram of this bridge. Phasor diagram of De Sauty bridge is
shown below:
Let us mark the current drop across unknown capacitor as e1, voltage drop across the resistor r3 be e3,
voltage drop across arm 3-4 be e4 and voltage drop across arm 4-1 be e2. At balance condition the
current flows through 2-4 path will be zero and also voltage drops e1 and e3 be equal to voltage drops
e2 and e4 respectively.
In order to draw the phasor diagram we have taken e3 (or e4) reference axis, e1 and e2 are shown at
right angle to e1 (or e2). Why they are at right angle to each other? Answer to this question is very
simple as capacitor is connected there, therefore phase difference angle obtained is 90o.
Now instead of some advantages like bridge is quite simple and provides easy calculations, there are
some disadvantages of this bridge because this bridge give inaccurate results for imperfect capacitor
(here imperfect means capacitors which not free from dielectric losses). Hence we can use this bridge
only for comparing perfect capacitors.
Here we interested in modify the De Sauty’s bridge, we want to have such a kind of bridge that will
gives us accurate results for imperfect capacitors also. This modification is done by Grover. The
modified circuit diagram is shown below:
Here Grover has introduced electrical resistances r1 and r2 as shown in above on arms 1-2 and 4-1
respectively, in order to include the dielectric losses. Also he has connected resistances R1 and R2
respectively in the arms 1-2 and 4-1. Let us derive the expression capacitor c1 whose value is unknown
to us. Again we connected standard capacitor on the same arm 1-4 as we have done in De Sauty’s
bridge. At balance point on equating the voltage drops we have:
Let us mark δ1 and δ2 be phase angles of the capacitors c1 and c2 capacitors respectively. From the
phasor diagram we have tan(δ1) = dissipation factor = ωc1r1 and similarly we have tan(δ2) = ωc2r2.
From equation (1) we have
Hence if dissipation factor for one capacitor is known. However this method is gives quite inaccurate
results for dissipation factor.
2.5.1 Wein’s bridge
Wein’s bridge is popularly used for measurements of frequency of frequency. In this bridge, the
value of all parameters are known. The source whose frequency has to measure is connected as
shown in the figure.
Fig Wein’s bridge
Wagner earthing consists of ‘R’ and ‘C’ in series. The stray capacitance at node ‘B’ and ‘D’ are
CB, CD respectively. These Stray capacitances produced error in the measurements of ‘L’ and ‘C’.
These error will predominant at high frequency. The error due to this capacitance can be eliminated
using wagner earthing arm.
Close the change over switch to the position (1) and obtained balanced. Now change the switch to
position (2) and obtained balance. This process has to repeat until balance is achievedin both the
position. In this condition the potential difference across each capacitor is zero. Current drawn by
this is zero. Therefore they do not have any effect on the measurements.
Precaution
✓ The load inductance is eliminated by twisting the connecting the connecting lead.
AÎ0Î r
✓ In the case of capacitive bridge, the connecting lead are kept apart.(QC = )
d
✓ In the case of inductive bridge, the various arm are magnetically screen.
✓ In the case of capacitive bridge, the various arm are electro statically screen to reduced
the stray capacitance between various arm.
✓ To avoid the problem of spike, an inter bridge transformer is used in between the source
and bridge.
✓ The stray capacitance between the ends of detector to the ground, cause difficulty in
balancing as well as error in measurements. To avoid this problem, we use wagner earthing
device.