0% found this document useful (0 votes)
20 views1 page

Microsoft Word - 180. VC20

The MTN/VC20 Vibration Calibrator is designed for the rapid calibration of vibration transducers and measuring systems, featuring a load-independent vibration level for test objects up to 600 grams. It operates at a frequency of 159.15 Hz with a maximum acceleration of 10 m/s² and includes a factory calibration certificate traceable to national standards. The device is rugged, battery-operated, and comes with various accessories for ease of use in both laboratory and field environments.

Uploaded by

Manoj Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
20 views1 page

Microsoft Word - 180. VC20

The MTN/VC20 Vibration Calibrator is designed for the rapid calibration of vibration transducers and measuring systems, featuring a load-independent vibration level for test objects up to 600 grams. It operates at a frequency of 159.15 Hz with a maximum acceleration of 10 m/s² and includes a factory calibration certificate traceable to national standards. The device is rugged, battery-operated, and comes with various accessories for ease of use in both laboratory and field environments.

Uploaded by

Manoj Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

TECHNICAL DATA

Datasheet DS0180 Issue 1 1/1

MTN/VC20 VIBRATION CALIBRATOR


Application
• Rapid and easy calibration of vibration transducers for acceleration,
velocity and displacement
• Calibration and fault detection of vibration measuring systems

Features
• Load independent vibration level for test objects up to 600 grams
• Vibration magnitude of 10 m/s² (VC20)
• Quartz controlled vibration frequency of 159.15 Hz (VC20)
• Display shows frequency, magnitude, error in percent and calibration
date
• Traceable to national PTB standard, factory calibration certificate
included
• Rugged design
• Mains buffered battery operation for laboratory and field use
• Mains plug adapter with wide range input included

SPECIFICATIONS
Vibration frequency 159.2 Hz
Acceleration (rms) 10 m/s²
Velocities (rms) 10 mm/s
at radian frequencies 100s-1 and 1000s-1
Displacements (rms) 10 µm
at radian frequencies 100s-1 and 1000s-1
Maximum weight of test object 600 g
Maximum amplitude error 3 % (10 - 40 °C / 50 100 °F)
± 5 % (-10 - 55 °C / 14 - 130 °F)
Maximum frequency error ± 0,05 %
Transverse vibration < 10 %
Harmonic distortion <1%
Level indication Percent display and piezo beeper
Sensor mounting M5 tapped hole (7mm deep), magnet
Operating temperature range -10 - 55°C (14 - 130°F)
Accumulator type Built-in NiMH battery pack; 7.2 V / 1.6 Ah
Operating time per battery charge Approx. 5 hours with 100 g weight
Charge socket Circular power connector to DIN 45323 (5.5 mm / 2.2 mm)
Charging time Approx. 4 hours
Charge condition indication Bar graph display
Charging voltage 11 - 18 VDC
Charging current <1A
Automatic switch off 10 Minutes
Dimensions (L x W x H) 100 x 100 x 120mm³
Weight 2.2 kg
Accessories supplied Plastic carrying case, mains plug adapter (100 - 240 VAC; 50/60
Hz),thread adapters (M3, M5, M8, 1/4"-28 UNF 10-32)

MONITRAN LTD
Monitor House | 33 Hazlemere Road | Penn | Bucks | HP10 8AD | UK
Tel: +44 (0)1494 816569 Email: info@monitran.com Registered in England No. 2054040
Fax: +44 (0)1494 812256 Web: www.monitran.com Vat No. GB442 4883 40