Geoexplorer Downhole: User Manual
Geoexplorer Downhole: User Manual
1 – USER MANUAL
GEOEXPLORER DOWNHOLE
Version 2.2.1
USER MANUAL
(09/ 2022)
     This software is produced for professional and research purposes. The use of
     the software for the purpose of recording geophysical data and / or assessment
     of risks and mechanical characteristics of soils is reserved for qualified and
     specialized personnel such as doctors of engineering, architecture, physics
     and geology, having also followed specialization courses in geophysics and
     structural calculation. All certifications regarding this material are the
     responsibility of the end user.
     The software is provided in its most updated version, without any express or
     implied warranties. The developer of GEOEXPLORER DOWNHOLE is not
     responsible for any loss, injury or damage of any kind in the event that the
     software is used in situations of vital importance and without the necessary
     technical expertise essential for the validation of the data obtained and does
     not accept responsibility in the case of inappropriate use of the product.
Table of contents
1 Introduction                                                                            6
2 Downhole                                                                                6
  2.1 Menu bar                                                                            7
  2.2 Tabs                                                                               11
     2.2.1 Dataset                                                                       12
     2.2.2 Channel selection                                                             16
     2.2.3 Correction                                                                    18
     2.2.4 Picking                                                                       21
     2.2.5 Data control                                                                  24
     2.2.6 Intercept                                                                     26
     2.2.7 Pseudo-interval                                                               30
     2.2.8 Equivalent velocity                                                           32
3 Introduction to data acquisition                                                       33
4 Data Acquisition                                                                       35
  4.2 Preparation of the detection station                                               36
  4.3 Seismic sources                                                                    38
  4.4 Sensor                                                                             39
  4.5 Campaign data acquisition and setup                                                40
  4.6 Execution procedure of detection – Down-Hole acquisition technique                 41
5 Processing                                                                             42
  5.1 Intercept                                                                          45
  5.2 Pseudo-interval                                                                    46
  5.3 Equivalent velocity                                                                47
  5.4 Formulas                                                                           48
  5.5 Bibliography                                                                       49
1 Introduction
     The Downhole module was developed with the aim of being able to verify the quality of the
     data through the management of the spy geophone, to be able to process acquisitions with
     regular and irregular geometries, without limitations, using different processing methods,
     returning the seismic-stratigraphic profile with the speeds and the geophysical parameters that
     can be calculated.
2 Downhole
    Illustration 2-2
     The main page of the software is the Dataset tab where is possible to load in different ways
     the dataset recorded in the field or directly the seismograms ready for the picking.
Illustration 2.1-2
Illustration 2.1-3
Illustration 2.1-4
     This software generate a temporary copy of the uploaded files on which it works. When you
     proceed with saving the work, the program generates a folder with the selected name in the
     desired position. The folder contains the original files, the support files and the processing
     information, thus creating an image relating to the status of the processing process at the time
     of saving.
                     Please note: in addition to making a backup copy of the original files, it is
                     advisable to proceed with periodic backups in different folders, in order to
                     preserve the various processing steps. The software also provides the
                     overwriting function in the same save folder, but we DO NOT RECOMMEND
                     this procedure
                     Please note: Manually modifying the contents of the save folders can lead to
                     anomalous behavior of the software and/or to the loss of processed
                     information. WE STRONGLY NOT RECOMMEND modifying save folders.
Illustration 2.1-6
Illustration 2.1-5
     The Report menu, allows to set all the work specific information for printing the pdf, as well
     as the photos the elaborations
Illustration 2.1-7
The ? menu gives the software information like version and contacts.
Illustration 2.1-8
2.2 Tabs
     The software is structured in tabs which will guide the user through the whole the elaboration
     process:
           1. 2.2.1 Dataset
              In this tab is possible to load a dataset and set the geometry
           2. 2.2.2 Channel selection
              In this tab is possible to select the horizontal channel for each horizontal shot
           3. 2.2.3 Correction
              In tab is possible to evaluate the spy geophone correction and apply it
           4. 2.2.4 Picking
              In this tab is possible to pick the first break
           5. 2.2.5 Data control
              In this tab is possible to revise the picking and see a resume of it
           6. 2.2.6 Intercept
              In this tab is possible to use the intercept method
           7. 2.2.7 Pseudo-interval
              In this tab is possible to use the Pseudo-interval method
           8. 2.2.8 Equivalent velocity
              In this tab is possible to use the Equivalent method
2.2.1 Dataset
     As explained before, on this tab is possible to load a dataset. Selecting one of the files, it is
     possible to see a preview in the box on the right.
Illustration 2.2-1
Illustration 2.2-2:
Illustration 2.2-3:
                     Please note: the Set geometry allows to set the geometry and shot information
                     massively. To do this, select the desired files for editing. (Drag and drop with
                     the mouse or Shift+click). It is recommended to load the data as shot type
                     dataset and edit them before load the second shot type dataset.
     To modify single information just double click on the desired cell. It is possible to change
     Depth, Shot, and enable/disable the file using the checkbox.
                     Please note: double clicking on Shot will change automatically following this
                     pattern P, R, L… Clicking on the Enable cell is possible to switch the
                     habitation from YES to No and vice-versa
     It is possible to browse the table using the arrow keys, and modify the values with the space
     bar.
     On this tab is possible to, in order from top to bottom, to select the P channel, enable the Spy
     geophone option, select the channel of the spy geophone.
     The selected P channel will be used to process P waves
     Enabling the Spy geophone option, the channel selected as spy geophone or the geophone on
     the surface for the quality check of the data will be removed from the display.
     The channels selected as P and as spy geophone will be deleted from the channel selection in
     the Channel selection tab.
                    Please note: it is not allowed to select the same channel both for P and the spy
                    geophone. In that case, an error message will be shown ad it will not be
                    possible to continue with processing.
     Once completed the dataset configuration, click on Order by depth. The dataset will be
     rearranged according to depths and the Check button will be enabled. By clicking on it,
     checks will be carried out on the dataset and, if successful, the next tab will be displayed.
     If there are more shot enabled with the same properties the software will automatically stack
     the seismograms. If there are no enabled seismograms, or if data have not been acquired at a
Illustration 2.2-6
      Illustration 2.2-7
     The buttons over the seismograms:
           •    Automatic select all
           •    Automatic select current
     allow respectively to automatically select the left and right channels for all depths or for the
     current depth.
                       Please note: the software perform the automatic selection analyzing the
                       amplitude of the channel. This selection does not mean is the best solution; we
                       recommend to check manually every depth.
     The button Write P and S create the P and S files that, if the Spy geophone option has been
     selected, will be loaded in the Correction tab, otherwise they will be used to create the final
     seismograms that will be opened on the Picking tab.
     The seismogram windows have the same viewing functions of the GEOEXPLORER
     DoReMi, so we refer to the GEOEXPLORER DoReMi manual for further explanation.
Illustration 2.2-8
     Illustration
     2.2-9
2.2.3 Correction
     In this view it is possible to correct the seismograms of the P, left S and right S waves, shown
     in the right graph, by means of the respective seismograms of the spies, shown in the graph on
     the left.
     Verify button allows to check which channels needs to be corrected according to the selected
     configuration. Those channels will be highlighted in yellow.
     Process button, applies the correction to the seismograms according to the chosen method,
     generates the final corrected files and loads them in the Picking tab.
Illustration 2.2-10
     It is possible to browse the seismogram using the keyboard. By Left and Right arrows the
     channels will be selected. (the currently selected channel is highlighted by a different line
     color). Up and Down arrows move the picking op or down 1 sample at a time. Pag Up and
     Pag Down keys move the picking op or down 1 millisecond at a time. Home and End keys
     move the picking at the beginning or at the end of the window.
     There are 3 correction methods (green box):
           1. None: No correction needed. Process button will be enabled. It allows to proceed to
              Picking without any correction. In that case it is not needed to pick the seismograms of
              the spies.
           2. Selected Channel: The correction is made on the basis of the picking carried out on
              the reference channel (by default the first). A different reference channel can be
              selected for each of the seismograms.
           3. Fixed Value: The correction is made on the basis of a user defined value. In the
              checkbox on the side a different value can be selected for each of the seismogram. If
              the value has not been set for a seismogram, it will be automatically set as the average
              of the picking values of all the channels of the same.
Illustration 2.2-12
                   Please note: the program treats each acquisition as top-down, then assuming
                   that the first channel is the most superficial acquisition.
     Once the correction tab is set, and the picking has been completed, using the Verify button the
     dataset will be analyzed in search of the channels that need to be corrected. These channels
     will be highlighted in yellow. At this point it is possible to repeat the operation until the
     desired result is obtained or press the Process button, which it will apply corrections to the
     dataset and will enable the Picking tab.
     The program checks the difference in ms of each picking of the spy geophone with respect to
     the selected method and translates the corresponding channels in the P and Sh seismograms
     by a value equal to the ms of difference between the picking value of the indicator and the
     value used as a reference (dependent from the selected methodology) if it exceeds the selected
     tolerance.
2.2.4 Picking
     In this tab is possible to perform the first break picking on the various seismograms. These
     seismograms can be opened directly (see chapter 2.1 Menu bar pag 7) or generated by the
     process explained on the previous chapters.
     P pickings are represented in red, S pickings are represented in blue. Furthermore, if enabled
     the option Show Pickings → Both P And S, on the seismogram of the P the pricking of the S is
     shown and vice versa, so as to be able to visually compare the two pickings.
     The table on the right shows the speed values on which the picking has been carried out for
     each channel. You can manually change these values by double clicking on the desired cell.
Illustration 2.2-13
                   Please note: is not mandatory to load the invert & sum and the overlap for the
                   S shot seismogram.
     It is possible, inside the Select seismogram, to select what seismogram to visualize for the
     picking. The seismogram window has the same function of the GEOEXPLORER DoReMi
     seismogram window. To be able to perform the picking just follow the same instruction given
     for the Correction tab.
                       Please note: the picking performed on the
                       seismograms S (Invert Sum and Overlap) is
                       simultaneous. This means is possible to
                       perform the picking only on one of them.
       Illustration 2.2-15
                                                            Illustration 2.2-16
     Set geometry allows to quickly manage the geometry of the acquisition by being able to
     change the position of the shots separately and the depth progression, as well as the height of
     the most superficial measurement.
     The Mirror function allows to flip the acquisition. This function is useful in the case the
     dataset is recorded as down-top.
     Illustration 2.2-17
     On the table is possible to edit the single picking
     value; by double click on the desired cell and is
     possible to edit the picking value and with the enter
     button on the keyboard is possible to apply this
     modification for the new geometrical correction
     computation.
     With the right click you can open the menu where is
     possible to:
           •    Disable cell: disable the selected cell
           •    Enable cell: enable the selected cell
           •    Enable all cells: enable all the cells
           •    Export table: Export the selected table to .csv
Illustration 2.2-18
      Illustration 2.2-19
     The dataset is automatically updated based on the enabled cells
                   Please note: the disabled cells are not used for the
                   computation on the other tabs.
Illustration 2.2-21
2.2.6 Intercept
     In this tab is possible to do the intercept interpretation. The software will plot the enabled
     geometrical corrected pick on the Data Control tab, allowing to draw the segment for P
     arrival and S arrival interdependently.
      Illustration 2.2-22
     With the options Set intercept it is possible to pick the relative P arrival segment or S arrival
     segment.
     Once selected the wave, the software will move the mouse on the axis origin and from there
     will be possible to see the preview of the segment. To set the segment it is need only the left
     click on the desired position. From this second point the software will plot the preview of the
     second segment.
     On every moment with the right click of the mouse it is possible to stop the picking process
     and selecting again the option is possible to start from the last point picked.
     To end the picking is mandatory to pick a point below the last represented depth.
     Placing the mouse over one knee of the intercept, it will be highlighted. Shift + Click on one
     knee of the intercept to reposition the knee in another point of the graph.
     Right clicking along the intercept line will show a context menu
     with the following options (Illustration 2.2-24)
           •    Add Point: Allows the user to position a new point,
                breaking the segment that was clicked.
           •    Continue Intercept: It allows picking to be resumed from
                the last point entered                                   Illustration 2.2-24
     By enabling the Round depth option, at the time of the picking the depth value is rounded to
     the nearest unit.
     The software will fill the top left table with the depth and velocity information of every
     picked segment for the P and S waves. Picking P and S the software will populate the bottom
     left table with the layer interval computed with the combination of P and s stratigraphy.
     With the options Delete intercepts it will be possible to delete the desired profile.
                       Please note: is not possible to pick a value above the previous in time and
                       depth. For example if you pick 8,1ms is no possible to pick 7ms and if you pick
                       a 7m is no possible to pick a 6m.
     Illustration 2.2-25
     Using the text boxes Offset and Vs eq. It is possible to set the software for the calculation of
     the equivalent Vs. In the offset field it is possible to indicate from what depth to start the
     calculation, while in the Vs eq field the depth at which to calculate the Vs eq is defined in
     meters. The result of the calculation is shown on the right in m/s.
     With the option Enable layer the software will show the stratigraphy and relative modules i
     (see formulas in the chapter 5.4):
           •    Poisson
           •    Density (RHO)
           •    Gamma
           •    Young
           •    Shear
     The option Exclude invalid value allows to do not plot out of range Poisson values.
     (Illustration 2.2-26) The default values are 0,16 the minimum limit, and 0,5 the maximum.
     These values can be changed in the Tools → Settings menu. (see Chapter 2.1 )
     Illustration 2.2-26
     The Tolerance parameter allows the user to define a tolerance value for the layers, if a
     tolerance value greater than 0 is assigned, all the layers with thickness less than the tolerance
     value are ignored as they are not significant. The maximum tolerance value that can be set is
     1m.
2.2.7 Pseudo-interval
     This tab elaborate the dataset as it has been recorded with 2 sensor in the well (true interval).
     The software will compute the time difference between two consecutive measurement to
     evaluate the interval velocity.
                    Please note: if the acquisition has been done with a 1m spacing, and for some
                    reason is missing a depth, or it has been disabled in the Data control tab, the
                    software will compute the interval velocity analysis with the closer
                    measurement, in this case the step will be 2m.
    Illustration 2.2-27
     As for the intercepts it is possible to calculate the Vs eq on the basis of the layers and the
     velocities calculated using the Pseudo-Interval method. The calculation takes place using the
     same methods and with the same Offset and Vs eq values, which are synchronized between
     the two views.
     With the Exclude invalid values option the software will eliminate from the graphical
     representation the values that, according to the verification of the calculations, return outliers.
     Illustration 2.2-28
     On the table is possible to see the result of the velocity analysis (Vp and Vs) and relative
     selected module. From the drop-down menu is possible to select the module.
     With the right click of the mouse on the table is possible to enable/disable cells, row and
     column and copy the data as text or image.
Illustration 2.2-29
     The in hole microseismic investigation is the main technique for the direct determination of
     the elasto-mechanical properties of the subsoil; the prospecting is carried out in a specifically
     equipped hole and allows to determine the propagation speed of the compression or
     longitudinal P waves, and the shear or transverse S waves. This is done by measuring the
     travel time of a wave train from the position of a specific source to the sensor (or sensors)
     prepared and made integral with the ground or in the hole.
     Three alternative methods are available for in hole seismic measurements:
           a) Surface energization and in hole reception, and vice versa.
Illustration 3-3
Illustration 3-4
     The most used microseismic techniques include the following detection techniques:
           a) Down-Hole (DH): the energization station is located on the surface and the sensor
              body are located at a specific depth in the hole.
           b) Up-Hole (UH): the energization station is located in the hole and the sensors are
              placed on the surface, generally at a different distance from the equipped vertical.
           c) Cross-Hole (CH): the energizations are produced in the hole, on a specific vertical
              while the sensors are arranged on a different equipped vertical, at a specific distance
              from the first.
     The Down-Hole and Up-Hole investigation techniques are the best known and most
     widespread, while the Cross-Hole technique is less used, as it presents higher costs for setting
     up the test station.
     The Down-Hole and Up-Hole techniques allow to determine the elastic behavior of soils
     along the direction normal to the ground (P and SH waves), while the Cross-Hole technique
     characterizes the behavior of the subsoil with regard to the propagation of the seismic input on
     the conjugate direction.
     This quickstart explains the correct acquisition method for the Down-Hole technique and the
     processing method.
4 Data Acquisition
     The investigation technique is based on the determination of the time it takes the wave train to
     travel the known distance from the energization point to each of the receiver positions.
     To determine the time taken by the wave train it is necessary to have:
           a) accurate precision in determining the energization instant (trigger) and a control of
              possible “delays” at the trigger (spy geophone), for the identification of the
              energization instant.
           b) high resolution of the signal, through a high signal / noise ratio, which allows the
              recognition of the instant of first arrival of the wave train at the sensor.
           c) adequate sampling frequency, which allows the instant of first arrival of the wave
              train to be determined with the greatest possible accuracy.
           d) length of the recording signal that allows you to sample the wave train at least for a
              duration that appreciates an attenuation of the amplitude of the recorded signal of at
              least 1/3 of that on first arrival.
4.1 Techniques of preparing the test vertical: characteristics and controls
     The vertical test consists of a geognostic survey conducted with continuous or destruction
     coring, with a diameter between 100mm and 140mm, completed with PVC or ABS DN80
     piping, suitably cemented to be made integral with the surrounding ground.
     The pipeline can also be stabilized by filling with fine sand, in particular geological
     conditions; exceptionally and only in particular lithostratigraphic conditions it is possible to
     proceed in non-equipped holes.
     The survey can also be carried out in piezometric verticals or holes equipped for inclinometric
     measurements, also completed with aluminum, fiberglass or plastic pipes, where it is possible
     to make the recording / energization apparatus adhere and make integral (with pistons or
     homologous systems) in the hole with the pipe.
     It is considered particularly useful to know the stratigraphic succession involved up to the
     expected depth of survey.
     Before carrying out the survey, it is necessary to wait for the withdrawal of the cement
     mortar: a waiting period of about 20 days must be foreseen, where mortars with specific
     aging accelerating additives have not been used.
     Checking the practicability of the hole and the integrity of the completion pipeline is the
     preliminary phase to the prospecting procedure.
     The maximum depth in which detection by Down-Hole or Up-Hole techniques is reliable does
     not exceed 50-60m.
     If you do not have certain data on how to complete the equipped forum, it is mandatory to
     carry out an investigation with a "witness" (it is a good idea to proceed with this procedure
     even if you have this data), which does not involve the possible loss instrumental in case it is
     not recoverable the control tool.
  Illustration 4.2-1
      Red box P and SH sources, violet arrow head hole, yellow circle spy geophone.
     The site must be set up in such a way as to be able to carry out and supervise each
     measurement operation, to verify the height of the sensor in the hole, to verify the
     energization procedures and for a quality check in the field.
     The construction site must provide for protection to the instrumentation and operators, in the
     expected climatic conditions, and the foreclosure of the detection area to the uncontrolled
     transit of people and vehicles, for the entire duration of the survey.
     The use of a pulley arranged in axis with the hole allows you to avoid breakages and damage
     to the data cable and the safety cable of the system due to the edge of the hole at the wellhead.
Illustration 4.2-2
Illustration 4.3-1
     The energization station must be equipped with a strike plate for energizing in P waves, and a
     sleeper, suitably ballasted, for energizing in SH waves.
     Each strike station must be equipped with a "trigger" for recording the moment of
     energization; any operational errors that lead to incorrect assessments on the moment of
     energization must be corrected by evaluating the signal with the spy geophone, appropriately
     prepared for this purpose.
     The energization station for P and transverse waves must be placed at a distance usually
     between 2m and 5m from the measurement vertical.
     The sleeper for energizing for shear waves must be suitably ballasted to ensure adherence to
     the ground during energization, for efficient transmission of the signal to the ground.
     The sleeper is placed on the ground previously leveled with an orthogonal direction to the
     source-hole joining; the stress produced is carried out at both ends with the same axis
     orientation.
     The energization at the two ends of the sleeper is necessary to allow the polarization in
     opposition of the shear waves and facilitate their recognition during data processing.
     The sleeper must not have metal fittings, bolts or anchors that may produce unwanted "noise"
     at the time of energization.
     The position of the energization points must be maintained and must not be changed
     throughout the test run.
4.4 Sensor
     The forum registration system consists of:
           a) Vertical and horizontal velocimeter (s) or accelerometer (s) with 3 or 5 channels.
                 they are made integral to a specific height along the measurement vertical or
                 arranged in a chain, at a predetermined mutual distance, in the previously chosen
                 recording depth.
     Illustration 4.4-1:
     The receivers must be made integral with the walls of the hole, except for measurements in
     holes filled with water, where suspension receivers can be used.
     The characteristics of the sensors must be adequate for the sensitivity and resolving capacity
     of the signal to be interpreted; ordinary instrumentation provides for the use of velometric
     sensors with a frequency of its own between 8 and 14 Hz with a flat response for frequencies
     above 10 Hz.
     If only one recording group is used, there is 1 vertical sensor and 2 or more horizontal
     sensors, arranged on a horizontal plane at variable angles of 90 ° (2 sensors), 60 ° (3 sensors)
     or 45 ° ( 4 sensors).
     An additional receiver must be set up in the measurement area to check for any "delays" on
     the moment of energization recorded by the dedicated trigger.
     This receiver generally consists of a vertical velocity sensor, kept at a predetermined distance
     by the energization system, which allows you to introduce any corrections on the wave arrival
     times to the different receivers, during subsequent measurements; for this sensor there is no
     need for particular technical specifications.
5 Processing
Illustration 5-1
     It is necessary to pick up the first arrivals on the seismograms of the spy geophone, thanks to
     which the software will check the triggering of the various channels. Once the correction
     method has been chosen and applied, the picking files will be generated.
     Example of trigger error and correction applied through the use of the spy geophone, which
     allows you to evaluate how much to correct the burst x with respect to a reference value
     chosen by the operator:
                                                                      Illustration 5-3
           Illustration 5-2
Illustration 5-4
     The program automatically calculates the geometric correction for each single burst by
     applying the law:
            z
      Tcorr= ×T dove r= √ z + x
                           2   2
            r
Illustration 5-5
Illustration 5-6
5.1 Intercept
     The graphical method
     consists in selecting the
     segment that includes the
     points that line up
     indicating a constant
     speed of propagation. By
     analyzing the inverse of
     the angular coefficient of
     this segment, it is possible
     to evaluate the speeds of
     the layer.
Illustration 5.1-1
Illustration 5.1-2
5.2 Pseudo-interval
     The interval technique is based on the
     fact that the acquisition is performed
     with the use of two sensors, placed at a
     constant distance in the hole. the two
     sensors move along the hole with a
     constant displacement distance.
     The pseudo-interval technique allows
     to analyze the data acquired with a
     single sensor, as if they had been
     acquired as described above.
     Obviously this leads to an error which
     is mitigated by the use of the spy
     geophone.
     The software calculates the interval
     speed following this law:
                Z n−Z n−1
      V=
            Tcorr n−Tcorr n−1
Illustration 5.2-1
Illustration 5.2-2
               Z n−Z 1
     V=
           Tcorr n−Tcorr1
Illustration 5.3-1
Illustration 5.3-2
5.4 Formulas
The formulas used in the software are as follows:
Calculation of correct times:
     •        Correct Times
       z
 Tcorr= ×T dove r= √ z + x
                      2   2
       r
      •       Poisson
  ((VP/VS)2−2)
           2
 2((VP/VS) −1)
          Valid if: | |VP/VS|≠1 and Poisson > 0.16 and Poisson < 0.5
     •        Density (T/m^3)
 0.51(VP 0.19)
     •        Bulk (MPa)
              2  4   2
 ( Density( VP −( )VS ))
                 3
            1000
     •        Young (MPa)
          2                     2     2
     (VS )∗Density∗(3VP −4 VS )
 (                              )/1000
             (VP 2−VS2 )
     •        Shear (MPa)
                    2
 (Density(VS ))
     1000
5.5 Bibliography