IOL Power Calculation
IOL Power Calculation
Advanced Intraocular
Lens Power Calculations 4
John P. Fang, Warren Hill, Li Wang,
Victor Chang, Douglas D. Koch
from 0.14 to 0.33 mm [24, 29, 30]. At normal Although the immersion technique has been
axial lengths, compression by 0.1 mm results in shown to be more reproducible than the applana-
a postoperative refractive error toward myopia of tion technique, both require mindfulness of the
roughly 0.25 D. Additionally, this method of ul- properties of ultrasound. Axial length is calcu-
trasound biometry is highly operator-dependent. lated from the measured time and the assumed
Because of the extent of the error produced by average speed that sound waves travel through
direct corneal contact, applanation biometry has the eye. Because the speed of ultrasound varies
given way to noncontact methods, which have in different media, the operator must account
been shown to be more reproducible. for prior surgical procedures involving the eye
such as IOL placement, aphakia, or the presence
of silicone oil in the vitreous cavity (Table 4.1).
Length correction can be performed simply us-
4.2.1.2 Immersion Technique ing the following formula:
The currently preferred A-scan method is the
immersion technique, which, if properly per- True length = [corrected velocity/measured ve-
formed, eliminates compression of the globe. locity] × measured length
Although the principles of immersion biometry
are the same as with applanation biometry, the However, using a single velocity for axial
technique is slightly different. The patient lies su- length measurements in eyes with prior sur-
pine with a clear plastic scleral shell placed over gery is much less accurate than correcting each
the cornea and between the eyelids. The shell segment of the eye individually and adding to-
is filled with coupling fluid through which the gether the respective corrected length measure-
probe emits sound waves. Unlike the applanation ments. For example, in an eye with silicone oil,
echogram, the immersion technique produces an the anterior chamber depth would be measured
additional spike corresponding to the probe tip at a velocity of 1,532 m/s, the crystalline lens
(Fig. 4.2). This spike is produced from the tip of thickness at 1,641 m/s, and the vitreous cavity
the probe within the coupling fluid. at either 980 m/s or 1,040 m/s depending on the
Table 4.1 Average velocities under various conditions By adding the CALF to or subtracting it from
for average eye length [16]. PMMA: polymethyl meth- the measured axial length, the true axial length
acrylate is obtained.
Another source of axial length error is that
Condition Velocity (m/s) the ultrasound beam has a larger diameter than
Phakic eye 1,555 the fovea. If most of the beam reflects off a raised
parafoveal area and not the fovea itself, this will
Aphakic eye 1,532 result in an erroneously short axial length read-
4 PMMA pseudophakic 1,556 ing. The parafoveal area may be 0.10–0.16 mm
Silicone pseudophakic 1,476 thicker than the fovea.
In addition to compression and beam width,
Acrylic pseudophakic 1,549
an off-axis reading may also result in a falsely
Phakic silicone oil 1,139 shortened axial length. As mentioned before, the
Aphakic silicone oil 1,052 probe should be positioned so that the magni-
tude of the peaks is greatest. If the last two spikes
Phakic gas 534
are not present (sclera and orbital fat), the beam
may be directed to the optic nerve instead of the
fovea.
density of the silicone oil (1,000 centistokes vs. In the setting of high to extreme axial myopia,
5,000 cSt). The three corrected lengths are then the presence of a posterior staphyloma should be
added together to obtain the true axial length. considered, especially if there is difficulty obtain-
Sect. 4.8 describes in greater detail IOL calcula- ing a distinct retinal spike during A-scan ultraso-
tions in eyes with silicone oil. nography. The incidence of posterior staphyloma
For pseudophakia, using a single instrument increases with increasing axial length, and it is
setting may also lead to significant errors be- likely that nearly all eyes with pathologic myopia
cause IOL implants vary in sound velocity and have some form of posterior staphyloma. Staphy-
thickness (Table 4.2). By using an IOL material- lomata can have a major impact on axial length
specific conversion factor (CF), a corrected axial measurements, as the most posterior portion of
length factor (CALF) can be determined using: the globe (the anatomic axial length) may not
correspond with the center of the macula (the
CF = 1 – (VE/VIOL) refractive axial length). When the fovea is situ-
CALF = CF × T ated on the sloping wall of the staphyloma, it may
where VE = sound velocity being used (such as only be possible to display a high-quality retinal
1,532 m/s), spike when the sound beam is directed eccentric
VIOL = sound velocity of the IOL material being to the fovea, toward the rounded bottom of the
measured, staphyloma. This will result in an erroneously
T = IOL central thickness. long axial length reading. Paradoxically, if the
sound beam is correctly aligned with the refrac- Through noncontact means, the IOL Master
tive axis, measuring to the fovea will often result (Carl Zeiss Meditec, Jena, Germany) emits an
in a poor-quality retinal spike and inconsistent infrared laser beam that is reflected back to the
axial length measurements. instrument from the retinal pigment epithelium.
Holladay has described an immersion A/B- The patient is asked to fixate on an internal light
scan approach to axial length measurement in the source to ensure axiality with the fovea. When
setting of a posterior staphyloma [4, 33]. Using a the reflected light is received by the instrument,
horizontal axial B-scan, an immersion echogram the axial length is calculated using a modified
through the posterior fundus is obtained with the Michelson interferometer. There are several ad-
cornea and lens echoes centered while simulta- vantages of optical coherence biometry:
neously displaying void of the optic nerve. The A- 1. Unlike A-scan biometry, the optical coher-
scan vector is then adjusted to pass through the ence biometry can measure pseudophakic,
middle of the cornea as well as the middle of the aphakic, and phakic IOL eyes. It can also mea-
anterior and posterior lens echoes to assure that sure through silicone oil without the need for
the vector will intersect the retina in the region of use of the velocity cenversion equation.
the fovea. Alternatively, as described by Hoffer, if 2. Because optical coherence biometry uses
it is possible to visually identify the center of the a partially coherent light source of a much
macula with a direct ophthalmoscope, the cross shorter wavelength than ultrasound, axial
hair reticule can be used to measure the distance length can be more accurately obtained. Op-
from the center of the macula to the margin of tical coherence biometry has been shown to
the optic nerve head. The A-scan is then posi- reproducibly measure axial length with an ac-
tioned so that measured distance is through the curacy of 0.01 mm.
center of the cornea, the center of the lens, and 3. It permits accurate measurements when pos-
just temporal to the void of the optic nerve on terior staphylomata are present. Since the
simultaneous B-scan. patient fixates along the direction of the mea-
suring beam, the instrument is more likely to
display an accurate axial length to the center
of the macula.
Summary for the Clinician 4. The IOL Master also provides measurements
■ Because the applanation technique re- of corneal power and anterior chamber depth,
quires direct contact with the cornea, enabling the device to perform IOL calcula-
compression will typically cause the axial tions using newer generation formulas, such
length to be falsely shortened. as Haigis and Holladay 2.
■ The speed of ultrasound varies in differ-
ent media. To account for this, the op- The primary limitation of optical biometry is its
erator must alter ultrasound speed set- inability to measure through dense cataracts and
tings for eyes that are pseudophakic or other media opacities that obscure the macula;
aphakic or that contain silicone oil in the due to such opacities or fixation difficulties, ap-
vitreous cavity. proximately 10% of eyes cannot be accurately
■ In the setting of high to extreme axial measured using the IOL Master [21].
myopia, the presence of a posterior When both optical and noncontact ultra-
staphyloma should be considered. sound biometry are available, the authors rely on
the former unless an adequate measurement can-
not be obtained. Both the IOL Master and im-
mersion ultrasound biometry have been shown
to produce a postoperative refractive error close
4.2.2 Optical Coherence Biometry to targeted values. However, the IOL Master is
Introduced in 2000, optical coherence biom- faster and more operator and patient-friendly.
etry has proved to be an exceptionally accurate Though mostly operator-independent, some
and reliable method of measuring axial length. degree of interpretation is still necessary for op-
36 Advanced Intraocular Lens Power Calculations
Fig. 4.4 Corneal surface irregularity shown on the Humphrey topographic map of an eye with epithelial base-
ment disease
38 Advanced Intraocular Lens Power Calculations
thickness, refractive error, and axial length to fur- on the power of the capsular bag IOL (Table 4.3).
ther refine the ELP calculation. The Holladay 2 The important concept is that for stronger intra-
formula is based on previous observations from a ocular lenses, the reduction in power must be
35.000 patient data set and has been shown to be greater. For very low IOL powers, no reduction
advantageous in both long and short eyes. in IOL power is required. Table 4.3 will provide
good results for most, modern posterior cham-
ber IOLs.
Summary for the Clinician
■ The shape of the power prediction curve 4.6 Determining IOL Power
is mostly fixed for each second and third
generation formula.
Following Corneal
Refractive Surgery
■ Popular third generation two-variable
formulas may also assume that the dis- The true corneal power following corneal refrac-
tance from the corneal vertex to the thin tive surgery is difficult to obtain by any form of
lens equivalent of the IOL is, in part, re- direct measurement. This is because keratometry
lated to the axial length and/or central and topography measure the anterior corneal
corneal power. radius and convert it to total corneal power by
■ The fourth generation IOL power for- assuming a normal relationship between the
mulas address these issues. anterior and posterior corneal curvatures. How-
ever, unlike incisional corneal refractive surgery
for myopia, which flattens both the anterior and
the posterior corneal radius, ablative corneal re-
fractive surgery for myopia primarily alters an-
4.5.3 Capsular Bag to Ciliary Sulcus terior corneal curvature. Additionally, standard
IOL Power Conversion keratometry measures a paracentral region and
Intraocular lens power formulas typically calcu- assumes that this accurately reflects central cor-
late the power of the intraocular lens to be posi- neal power. For these reasons, keratometry and
tioned within the capsular bag. Occasionally, this simulated keratometry by topography typically
is not possible, as with an unanticipated intraop- under-estimate central corneal power following
erative tear in the posterior lens capsule. In order ablative corneal surgery for myopia and overes-
to achieve a similar postoperative refractive re- timate it for corneas that have undergone hyper-
sult with an IOL placed at the plane of the cili- opic ablation.
ary sulcus, a reduction in IOL power is typically There is a second and less commonly recog-
required. nized source of unanticipated postoperative re-
The power adjustment necessary between the fractive error. As a general rule, IOL power cal-
capsular bag and the ciliary sulcus will depend culations following all forms of corneal refractive
surgery should not be run using an uncorrected
two-variable, third-generation formula because
Table 4.3 Intraocular lens (IOL) power correction for they assume that the effective lens position is, in
unanticipated sulcus implantation [13] part, related to central corneal power. By using
axial length and keratometric corneal power to
Capsular bag Ciliary sulcus power estimate the postoperative location of the IOL,
IOL power adjustment or the ELP, the artifact of very flat Ks follow-
ing myopic corneal refractive surgery will cause
+35.00 D to +27.50 D –1.50 D
these formulas to assume a falsely shallow post-
+27.00 D to +17.50 D –1.00 D operative ELP and recommend less IOL power
+17.00 D to +9.50 D –0.50 D than required. To avoid this potential pitfall, the
double K feature of the Holladay 2 formula al-
+9.00 D to -5.00 D No change
lows direct entry of two corneal power values by
40 Advanced Intraocular Lens Power Calculations
checking the box “Previous RK, PRK…”; if the IOLpre + (ΔD / 0.7) = IOLpost
corneal power value before refractive surgery is where IOLpre = the power of the IOL as if no
unknown, the formula will use 43.86 D as the de- LASIK had been performed,
fault preoperative corneal value. Another option ΔD = the refractive change after LASIK at the
is to apply Aramberri’s “double K method” cor- spectacle plane,
rection to the Holladay 1, Hoffer Q or SRK/T for- IOLpost = the estimated power of the IOL to be
mulas [1] or refer to the IOL power adjustment implanted following LASIK.
nomograms published by Koch and Wang [19].
4 Several methods have been proposed to im-
prove the accuracy of IOL power calculation in
eyes following corneal refractive surgery; these
4.6.1.3 Masket IOL
can be divided into those that require preopera-
Power Adjustment Method
tive data and those that do not. Masket [22] has developed another method that
adjusts the IOL power based on the amount of
refractive laser correction. Instead of calculat-
ing IOL power with pre-LASIK data as above,
4.6.1 Methods Requiring this method modifies the predicted IOL power
Historical Data obtained using the patient’s post-laser correction
readings by using the following formula:
4.6.1.1 Clinical History Method
The clinical history method [18] for corneal IOLpost + (ΔD × 0.326) + 0.101 = IOLadj
power estimation requires accurate historical
data and was first described by Holladay as: where IOLpost = the calculated IOL power fol-
lowing ablative corneal refractive surgery,
Kp + SEp - SEa = Ka ΔD = the refractive change after corneal refrac-
tive surgery at the spectacle plane,
where Kp = the average keratometry power be- IOLadj = the adjusted power of the IOL to be im-
fore corneal refractive surgery, planted.
SEp = the spherical equivalent before corneal re-
fractive surgery,
SEa = the stable spherical equivalent after corneal
refractive surgery,
4.6.1.4 Topographic Corneal
Ka = the estimate of the central corneal power
Power Adjustment Method
after corneal refractive surgery. There are several approaches to modifying post-
LASIK corneal power measurements:
Bc + Pc + SEc – SEs = Ka
4.6.3 Hyperopic Corneal
Refractive Surgery
where Bc = base curve of contact lens in diop- For eyes that have undergone hyperopic LASIK,
ters, it is easier to estimate central corneal power than
Pc = refractive power of contact lens in diopters, for myopic LASIK. This is presumably because
SEc = spherical equivalent with contact lens in the ablation takes place outside the central cor-
place, nea. The average of the 1-mm, and 2-mm an-
SEs = spherical equivalent without contact lens, nular power rings of the Numerical View of the
Ka = estimated corneal power following refrac- Zeiss Humphrey Atlas topographer can serve as
tive surgery. an estimate of central corneal power following
hyperopic LASIK. As an alternative, the adjusted
Unfortunately, the literature now suggests that EffRP of the EyeSys Corneal Analysis System
the hard contact lens method may be less accu- proposed by Drs. Wang, Jackson, and Koch also
rate than originally thought following all forms works well (see Sect. 4.6.1.4) [31].
of ablative corneal refractive surgery [2, 10, 17, Remember that some form of a “double K
32]. Better results may require the use of contact method” is still required for IOL power calcu-
lens designs with posterior curvatures that better lations following hyperopic LASIK in order to
fit the surgically modified corneal surface. avoid an inaccurate estimation of ELP.
42 Advanced Intraocular Lens Power Calculations
matism may be present and may not respond as unacceptably high ametropia, options include
expected to corneal relaxing incisions. IOL exchange, a piggyback IOL, or corneal re-
The higher order optical aberrations and fractive surgery.
multifocality that often accompany the various 2. Defer cataract surgery until the graft has sta-
forms of corneal refractive surgery also remain bilized, preferably after suture removal. Al-
unchanged following cataract surgery. For exam- though more accurate, there would be a delay
ple, third- and fourth-order higher order aber- in visual rehabilitation and the second proce-
rations produced by radial keratotomy can be as dure may cause surgical trauma to the donor
4 much as 35 times normal values. Elevated higher cornea.
order aberrations are also seen following PRK 3. Perform cataract extraction alone without
and LASIK, particularly decentered ablations or IOL implantation in conjunction with the cor-
older treatments with small central optical zones. neal graft. With this approach, there is mini-
Although the positive spherical aberration in- mal risk of trauma to the graft with the second
duced by myopic procedures may be partially procedure. However, it essentially eliminates
ameliorated by implanting an IOL with negative the chance of implanting the IOL in the cap-
asphericity, moderate to high amounts of posi- sular bag.
tive spherical aberration usually remain. The vi-
sual consequence of these aberrations is loss of
best-corrected acuity and contrast sensitivity
Summary for the Clinician
and, understandably, some patients mistakenly ■ Because it is impossible to accurately
expect that cataract surgery will alleviate these predict postoperative central power of
symptoms. Thus, it is important to discuss this the donor graft, there is presently no re-
prior to surgery so that their expectations will be liable method for calculating IOL power
realistic. for eyes undergoing combined corneal
The active use of so many different methods transplantation and cataract removal
of IOL calculation following corneal refractive with IOL implantation.
surgery is eloquent testimony to how far we still
have to go in this area. To minimize the risk of
unexpected postoperative hyperopia, we gen-
erally recommend a refractive target of around
–0.75 D, depending on the refractive status of the
4.8 Silicone Oil
fellow eye. For eyes containing silicone oil, A-scan axial
See Table 4.4 for an example of an intraocular length measurements are best carried out with the
lens calculation following corneal refractive sur- patient seated as upright as possible, especially if
gery. the vitreous cavity is partially filled with silicone
oil. In the upright position, it is more likely that
the silicone oil will remain in contact with the
retina. In the recumbent position, the less dense
4.7 Corneal Transplantation silicone oil will shift away from the retina, toward
There is presently no reliable method for calcu- the anterior segment. This can lead to confusion
lating IOL power for eyes undergoing combined as to the correct interpretation of the position of
corneal transplantation and cataract removal the retinal spike.
with IOL implantation. This is because it is im- The refractive index of silicone oil is also
possible to accurately predict the central power higher than that of the vitreous, requiring an ad-
of the donor graft. There are several options: justment to IOL power. To prevent the silicone
1. Use a mean corneal power, based on evalua- oil from altering the refractive power of the pos-
tion of prior grafts, as a “best guess” of post- terior surface of the IOL, it is preferable to im-
operative corneal power and proceed with plant polymethyl methacrylate (PMMA) convex-
IOL implantation. In eyes with an acceptable plano lenses, with the plano side oriented toward
postoperative refractive error, additional lens the vitreous cavity and preferably over an intact
surgery will not be required. For eyes with posterior capsule. The additional power that
References 45
11. Hamed AM, Wang L, Misra M, et al. A compara- 24. Olsen T, Nielsen PJ. Immersion versus contact
tive analysis of five methods of determining cor- technique in the measurement of axial length by ul-
neal refractive power in eyes that have undergone trasound. Acta Ophthalmol 1989;67(1):101–102.
myopic laser in situ keratomileusis. Ophthalmol- 25. Patel AS. IOL power selection for eyes with sili-
ogy 2002;109:651–658. cone oil used as vitreous replacement. Abstract
12. Hill WE. The IOLMaster. Tech Ophthalmol #163. Symposium on Cataract and Refractive Sur-
2003;1:62. gery, April 1–5, San Diego, California, 1995;41.
13. Hill WE, Byrne SF. Complex axial length mea- 26. Retzlaff J. A new intraocular lens calcula-
4 surements and unusual IOL Power calculations. tion formula. J Am Intraocul Implant Soc
In: Focal Points – Clinical Modules for Ophthal- 1980;6(2):148–152.
mologists. The American Academy of Ophthal- 27. Sanders D, Retzlaff J, Kraff M, et al. Comparison
mology, San Francisco, 2004;Module 9. of the accuracy of the Binkhorst, Colenbrander,
14. Hoffer KJ. Intraocular lens calculation: the and SRK implant power prediction formulas. J
problem of the short eye. Ophthalmic Surg Am Intraocul Implant Soc 1980;7(4):337–340.
1981;12(4):269–272. 28. Sanders DR, Kraff MC. Improvement of intraocu-
15. Hoffer KJ. The Hoffer Q formula: a comparison lar lens power calculation using empirical data. J
of theoretic and regression formulas. J Cataract Am Intraocul Implant Soc 1980;6(3):263–267.
Refract Surg 1993;19(6):700–712. 29. Schelenz J, Kammann J. Comparison of contact
16. Hoffer KJ. Ultrasound velocities for axial eye and immersion techniques for axial length mea-
length measurement. J Cataract Refract Surg surement and implant power calculation. J Cata-
1994;20(5):554–562. ract Refract Surg 1989;15(4):425–428.
17. Hoffer KJ. Intraocular lens power calculation for 30. Shammas HJ. A comparison of immersion and
eyes after refractive keratotomy. J Refract Surg contact techniques for axial length measurement.
1995;11:490–493. J Am Intraocul Implant Soc 1984;10(4):444–447.
18. Holladay JT. Consultations in refractive surgery 31. Wang L, Jackson DW, Koch DD. Methods of es-
(letter). Refract Corneal Surg 1989;5:203. timating corneal refractive power after hyperopic
19. Koch DD, Wang L. Calculating IOL power in eyes laser in situ keratomileusis. J Cataract Refract
that have had refractive surgery. J Cataract Re- Surg 2002;28:954–961.
fract Surg 2003;29(11):2039–2042. 32. Wang L, Booth MA, Koch DD. Comparison of in-
20. Koch DD, Liu JF, Hyde LL, et al. Refractive com- traocular lens power calculation methods in eyes
plications of cataract surgery after radial keratot- that have undergone laser in-situ keratomileusis.
omy. Am J Ophthalmol 1989;108(6):676–682. Ophthalmology 2004;111(10):1825–1831.
21. Lege BA, Haigis W. Laser interference biometry 33. Zaldiver R, Shultz MC, Davidorf JM, et al. In-
versus ultrasound biometry in certain clinical traocular lens power calculations in patients
conditions. Graefes Arch Clin Exp Ophthalmol with extreme myopia. J Cataract Refract Surg
2004;242(1):8–12. 2000;26:668–674.
22. Masket S. Simple regression formula for intraocu- 34. Zeh WG, Koch DD. Comparison of contact lens
lar lens power adjustment in eyes requiring cata- overrefraction and standard keratometry for mea-
ract surgery after excimer laser photoablation. J suring corneal curvature in eyes with lenticular
Cataract Refract Surg 2006; 32(3):430–434. opacity. J Cataract Refract Surg 1999;25:898–903.
23. Olsen T. Sources of error in intraocular lens
power calculation. J Cataract Refract Surg
1992;18:125–129.