Spectros
Spectros
                                        Light Spectrum
Its properties depend on sample composition and concentration. It helps to identify, assess purity,
and quantify the components of the sample by analyzing the pattern of absorption and
transmission of light. It may apply in several sample types, such as monolithic solids, liquids,
glass, powders, and thin films.
Absorbance (A): It, also known as optical density (OD), is the amount of light absorbed by the
object and can be expressed as follows:
Absorbance (A)= -log(T)
Transmittance (T): It is measured by dividing the intensity spectrum of light transmitted through
a sample (I) by the intensity spectrum of light transmitted through the blank (I0).
T= I/Io
UV-Vis Spectroscopy Principle
When a specific wavelength of light hits a molecule, that molecule gets excited. Once the
electron excites, it excites from the ground (lower) energy state to the higher energy state. When
an electron jumps off, it absorbs light energy because electrons in the orbital at a lower energy
state utilize energy to move to a higher energy level.
Energy is neither created nor destroyed but can transform energy from one form to another. On
passing EMR (UV- Vis range 200- 800 nm), only light possessing the precise amount of energy
that can cause transitions from one level to another will absorb because matter’s energy levels
are quantized.
If the energy is utilized, the intensity of light received is lost. At this time, the energy absorbed
by the electrons will equal the energy difference between the two energy levels.
During this stage, electron transition occurs. So, after the interaction of electromagnet radiation,
the spectra received are called absorption spectra. Hence, it is called electron spectroscopy.
Similarly, when electrons in the orbital at a higher energy level move to the ground energy level,
the spectra received are called emissions.
Beer-Lambert Law equation is the principle behind absorbance spectroscopy.
The concentration of the sample can be determined directly from the absorption of spectra
produced by these samples at specific wavelengths using the Beer-Lambert law.
What is Beer-Lambert Law?
When a beam of light allows it to pass through a transparent medium, the rate at which an
intensity decreases with medium thickness is directly proportional to the light beam’s intensity.
According to the Beer-Lambert Law, the absorbance is directly proportional to the concentration
of the substance in the solution. Therefore, a sample’s concentration can also be determined
using UV-visible spectroscopy.
The Beer-Lambert Law can be expressed in the form of the following equation:
A = –log T = –log (I ⁄ Io ) = log(Io ⁄ I)= ecl
A = ecl
Where A = absorbance
l = optical path length of the cell or cuvette or sample holder(cm)
c = concentration of the solution (mol dm-3)
e = molar absorptivity of the compound or molecule in solution, which is constant for a
particular substance at a particular wavelength (dm3 mol-1 cm-1)
Following the Beer-Lambert Law, the plot of absorbance versus concentration should be linear if
the absorbance of a series of sample solutions with known concentrations is measured and
plotted against equivalent concentrations. This graph is known as a calibration graph.
 Instrumentation of UV-Vis Spectroscopy
 The main components of UV- Vis spectrophotometer are:
1. Light Source
2. Wavelength selector
3. Sample container
4. Detectors
  1. Light Source
  It is essential for emitting light in a wide range of wavelengths to work in a UV-Vis
  spectrometer. Commonly, a high-intensity light source used for both UV and Visible ranges is a
  xenon lamp. In contrast to tungsten and halogen lamps, it is less stable and more costly. So, the
  two lamps for this instrument are a deuterium lamp for UV light and a halogen or tungsten lamp
  for visible light as a source of light. The two lamps provide good intensity. While measuring the
  intensity of the light, the spectrometer ought to switch. A smoother transition is possible when
  the switchover occurs between 300 and 350 nm because the light emission for both visible and
  UV light sources is the same amount of light at that wavelength.
 2. Wavelength selector
 In order to allow sample examination using the wavelengths that the light source emits,
 wavelength selection helps to ascertain which wavelength is appropriate for the type of analyte
 and sample. The commonly used wavelength selector in the UV-Vis spectrometer is the
 monochromator. It separates light into a narrow band of wavelength.
 From the entrance slit, radiation of different wavelengths will enter the monochromator. At a
 particular angle, the beam will collide and strike the dispersing element. A monochromator
 contains a prism that separates all different wavelengths of light in a single beam. It bends the
monochromatic light and produces non-linear dispersion. Only single radiation or color of a
specific wavelength will allow it to leave the monochromator and pass through its ultimate chain
or exit slit.
3. Sample Container
In a single-beam spectrophotometer, all the radiation coming from the light source passes
through the sample as one beam. Single-beam spectrophotometers can determine color by
comparing the light sources’ intensities before and after a sample is inserted. The wavelength
range measure is 190–750 nm; however, it may go up to 1100 nm.
In a double-beam spectrophotometer, all the radiation coming from the light source splits into
two beams: one passes through the sample, and the other only passes through the reference.
Similarly, Double-beam spectrophotometers offer a wavelength range of 190 to 1100 nm.
Moreover, the double-beam spectrophotometer measures absorbance versus wavelength or
sample and reference beam ratio.
The reference detector is used to adjust lamp brightness fluctuations for each measurement. After
collecting the sample, the sample detector is measured in the sample position and deducted from
the sample spectrum. It contains both a reference chamber and a sample chamber. The sample is
kept in a flat, transparent container called a cuvette or sample chamber. The solvent in which the
sample dissolves is kept in the reference chamber, also known as the blank. The sample cell’s
choice depends on the path length, shape, size, and transmission characteristics at the desired
wavelength and the relative expense.
For each wavelength, the light intensity passes through the beam separator to the reference
chamber (Io) and sample chamber (I). The intensity of light symbolizes as Io measures the
number of photons per second. When the light passes through the blank solution, it does not
absorb light, referred to as (l). If sample I is less than Io, the sample has absorbed some light.
The absorbance (A) of the sample is related to I and Io according to the following equation:
Absorbance (A)= -log(T)= -log (I/Io)
This equation shows the relationships between absorbance and transmittance.
Also, the fraction I divided by Io is called transmittance (T), which expresses how much light has
passed through a sample.
   T= I/Io
T= I/Io = e–kbc
Where: – I o is the incident intensity
– I is the transmitted intensity
– e is the base of natural logarithms
– k is a constant
– b is the path length (usually in cms).
The lighter the refracted, the more transmittance occurs. The lower the absorbance, the higher
the transmittance.
In UV and visible regions, fused silica or quartz cuvettes are commonly used.
 4. Detector
 Detectors rely on photoelectric coatings or semiconductors. It converts the incoming light from
 the sample into an electric signal or current. The higher the current, the greater the intensity. It
 has the properties of low noise and high sensitivity, so it gives a linear response. Each detector
 has a variety of wavelength ranges and different sensitivity. Finally, The data recorder usually
 plots the absorbance against wavelength (nm) in the UV and visible section of the
 electromagnetic spectrum.
UV-Visible spectroscopy is widely used in the field of analytical chemistry, especially during the
quantitative analysis of a specific analyte. For example, the quantitative analysis of transition
metal ions can be achieved with the help of UV-Visible spectroscopy. Furthermore, the
quantitative analysis of conjugated organic compounds can also be done with the help of UV-
Visible spectroscopy. It can also be noted that this type of spectroscopy can also be carried out
on solid and gaseous analytes in some conditions.
The main disadvantage of utilising a UV-VIS spectrometer is the time it requires to prepare to
use one. Setup is crucial when using UV-VIS spectrometers. Any outside light, electrical noise,
or other outside contaminants that could interfere with the spectrometer’s reading must be
removed from the location.
Because glass and most plastics absorb ultraviolet light, reusable quartz cuvettes were formerly
required for measurements in the ultraviolet range.
What is UV-Vis spectroscopy?
       Ultraviolet (UV) and visible radiation are a small part of the electromagnetic spectrum,
       which includes other forms of radiation such as radio, infrared (IR), cosmic, and X rays.
       Spectroscopy allows the study of how matter interacts with or emits electromagnetic
       radiation. There are different types of spectroscopy, depending on the wavelength range
       that is being measured. UV-Vis spectroscopy uses the ultraviolet and visible regions of
       the electromagnetic spectrum. Infrared spectroscopy uses the lower-energy infrared part
       of the spectrum. In UV-Vis spectroscopy, wavelength is usually expressed in nanometers
       (1 nm = 10-9 m). The UV range normally extends from 100 to 400 nm, with the visible
       range from approximately 400 to 800 nm.
       Because light is a form of energy, absorption of light by matter causes the energy content
       of the molecules (or atoms) in the matter to increase. In some molecules and atoms,
       incident photons of UV and visible light have enough energy to cause transitions between
       the different electronic energy levels. The wavelength of light absorbed has the energy
       required to move an electron from a lower energy level to a higher energy level. The
       figure below shows an example of electronic transitions in formaldehyde and the
       wavelengths of light that cause them.
Electronic transitions in formaldehyde. UV light at 187 nm causes excitation of an
electron in the C-O bond, and light at 285 nm wavelength causes excitation and transfer
of an electron from the oxygen atom to the C-O bond.
However, for molecules, vibrational and rotational energy levels are superimposed on the
electronic energy levels. Because many transitions with different energies can occur, the
bands are broadened. The broadening is even greater in solutions owing to solvent-solute
interactions.
     Electronic transitions and UV-visible spectra in molecules (I is intensity and λ is
     wavelength).
UV-Vis spectrophotometry is a versatile technique and has been used for close to a
century in a wide range of fields. UV-Vis spectrophotometers are in common use in
material testing/research, chemistry/petrochemistry, and biotechnology/pharmaceuticals
laboratories.
What are the main components of a UV-Vis spectrophotometer?
       The key components of a UV-Vis spectrophotometer are:
      A light source that generates a broadband of electromagnetic radiation across the UV-
       visible spectrum
      A dispersion device that separates the broadband radiation into wavelengths
      A sample area, where the light passes through or reflects off a sample
      One or more detectors to measure the intensity of the reflected or transmitted radiation
       Other optical components, such as lenses, mirrors, or fiber optics, relay light through the
       instrument.
       Light sources
       See the next question for full details on light sources.
    Monochromators
    All the light sources produce a broad-spectrum white light. To narrow the light down to a
    selected wavelength band, the light is passed through a monochromator, which consists
    of:
   An entrance slit
   A dispersion device, to spread the light into different wavelengths (like a rainbow) and
    allow the selection of a nominated band of wavelengths
   An exit slit where the light of the nominated wavelengths passes through and onto the
    sample
   For more details on how monochromators work and on holographic gratings, filters, plus
    single and double monochromator spectrophotometers, download the Agilent handbook
    on the Basics of UV-Vis Spectrophotometry.
    Sample compartments
    The sample compartment of a UV-Vis spectrophotometer is typically a black-colored box
    with a closing lid. The matt black inside the compartment helps to absorb stray light that
    may enter the compartment. In the sample compartment, the sample is positioned to
    allow the beam from the monochromator to pass through the sample. Glass, plastic, or
    quartz cuvettes are used for liquid samples. Solid samples are held in position by a holder
    attached to the floor of the sample compartment. The light can also be taken out of the
    sample compartment using fiber optics.
    Detectors
    A detector converts the light from the sample into an electrical signal. Like the light
    source, it should give a linear response over a wide wavelength range, with low noise and
    high sensitivity.
    Each detector has a different sensitivity and wavelength range. For systems with multiple
    detectors, the system will switch to the detector corresponding to the required wavelength
    range for the measurement. UV-Vis spectrophotometer detectors include photomultiplier
    tubes (PMT) and silicon diodes (Si). Indium gallium arsenide (InGaAs) photodiodes and
    lead sulfide (PbS) detectors are found on high-performance UV-Vis-NIR systems to
    improve wavelength coverage or sensitivity.
What light sources do UV-Vis spectrophotometers use?
       The ideal light source would yield a constant intensity over all wavelengths with low
       noise and long-term stability of the output. Unfortunately, such a source does not exist.
       Two different light sources have historically been used in UV-Vis spectrophotometers:
      The deuterium arc lamp was used to provide a good intensity continuum in the UV region
       and useful intensity in the visible region.
      The tungsten-halogen lamp yielded good intensity over the entire visible range and part
       of the UV spectrum.
       More recently, a single xenon flash lamp has been used more widely. The use of a xenon
       flash lamp as a single source has significant advantages over the use of the two
       conventional lamps.
       Tungsten-halogen lamp
       The tungsten-halogen lamp uses a filament. When a current is passed through the
       filament, it becomes heated and emits light. The lamp yields good intensity over part of
       the UV spectrum and over the entire visible and NIR range (350 to 3000 nm). This type
       of lamp has very low noise and low drift and typically has a functional life of 10,000 h.
      The xenon flash lamp emits high intensity light from 185 to 2,500 nm, which means no
      secondary light source is required. The xenon flash lamp may be used for many years
      before requiring replacement and it does not require warmup time, making it a popular
      choice. Xenon flash lamps are used in Agilent Cary 60 and Cary 3500 UV-Vis
      spectrophotometers.
      Cuvettes for measuring liquid samples. From left to right: A standard 10 mm pathlength,
      3 mL cuvette, an ultramicro cell for measuring very low volumes, and a long pathlength
      cuvette for dilute solutions.
      Solid samples can be held in place for simple transmission measurements. They can also
      be measured at various angles of incidence. For more complex measurements, like
      diffuse reflectance or transmission, other accessories may be installed into the sample
      compartment.
      The light can also be taken out of the sample compartment using fiber optics. Fiber optics
      are useful when measuring very large, hot, cold, radioactive, or other dangerous samples.
      As shown below, fiber optics can take the light from the spectrophotometer through a
      fiber-optic probe, to measure solutions outside of the sample compartment. Alternatively,
      a fiber-optic device that allows the measurement of light reflectance, fluorescence, or
      transmission through a solid sample can be used.
   1. Optical cell selection—this is based on desired pathlength, sample volume, and the
      optical properties of the material used in the cuvette windows.
   2. Thermostatting your samples—there are some circumstances that require samples to be
      heated or cooled.
   3. Stirring your sample—this ensures that both solution and temperature homogeneity are
      always maintained.
   4. Measurements at low temperatures—these may cause condensation to form on the
      outside of cuvettes, which can interfere with the measurement.
   5. Solvent transparency for liquid samples or dissolving of solid samples—solvents are
      selected based on sample solubility, stability, pH requirements, and the UV-visible cut-
      off wavelength.
   6. Optimum spectral bandwidth—when measuring a sample, consideration should be given
      to the measurement resolution required.
   7. Stray light—the percentage of radiation reaching the detector whose wavelengths are
      outside the selected spectral band. Most systems are provided with instrument
      performance checks that identify stray light issues.
   8. The linear range of a UV-Vis instrument—understanding the limits of your system allows
      you to avoid measuring samples or performing calibrations that are outside the
      capabilities of your instrument.
   Difference between single beam and double beam spectrophotometer?
1. Design:
       Single-Beam Spectrophotometer: In a single-beam spectrophotometer, only one beam
          of light passes through the sample. The sample is placed in the path of this beam, and the
          intensity of the light transmitted through or absorbed by the sample is measured.
       Double-Beam Spectrophotometer: In a double-beam spectrophotometer, the light beam
          is split into two paths - one passes through the sample, while the other serves as a
          reference and typically passes through a blank solvent or reference material. The two
          beams are then recombined, and the difference in intensity between them is measured.
2. Baseline Stability:
       Single-Beam Spectrophotometer: Since there's no separate reference beam, any
          fluctuations in the light source or detector can affect the accuracy of the measurement.
          Baseline stability can be an issue in single-beam instruments.
       Double-Beam Spectrophotometer: Because it utilizes a separate reference beam,
          double-beam spectrophotometers offer better baseline stability. Any fluctuations in the
          light source or detector are accounted for by the reference beam, resulting in more
          accurate measurements.
3. Accuracy and Precision:
        Single-Beam Spectrophotometer: Single-beam instruments may suffer from
         inaccuracies due to fluctuations in the light source or detector. However, they can still
         provide accurate measurements for many applications with careful calibration and control
         of experimental conditions.
      Double-Beam Spectrophotometer: Double-beam spectrophotometers generally offer
         higher accuracy and precision compared to single-beam instruments, particularly for
         quantitative analysis. The ability to compensate for fluctuations in the light source or
         detector enhances the reliability of measurements.
4. Applications:
      Single-Beam Spectrophotometer: Single-beam instruments are often suitable for
         qualitative analysis, routine measurements, and educational purposes where high
         precision is not critical.
      Double-Beam Spectrophotometer: Double-beam spectrophotometers are preferred for
         quantitative analysis, especially in research laboratories and industries where precise and
         accurate measurements are essential.
   In summary, while both single-beam and double-beam spectrophotometers serve the same
   fundamental purpose of measuring light absorption or transmission by samples, double-beam
   instruments offer enhanced accuracy, precision, and baseline stability due to their design with
   separate reference beams. Therefore, they are often preferred for demanding analytical
   applications.