Stabi

The report dated December 31, 2024, provides baseline stability measurements from a U-3900 Spectrophotometer, indicating a stability of 0.00021 Abs/h, which is outside the acceptable tolerance of 0.0002 Abs/h. The instrument parameters include a time scan measurement type at a wavelength of 500.00 nm with a scan time of 3600 seconds. The report was generated by the service operator and includes details about the instrument's settings and performance.

Uploaded by

Say Goodbye
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
11 views1 page

Stabi

The report dated December 31, 2024, provides baseline stability measurements from a U-3900 Spectrophotometer, indicating a stability of 0.00021 Abs/h, which is outside the acceptable tolerance of 0.0002 Abs/h. The instrument parameters include a time scan measurement type at a wavelength of 500.00 nm with a scan time of 3600 seconds. The report was generated by the service operator and includes details about the instrument's settings and performance.

Uploaded by

Say Goodbye
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Report Date: 14:26:03, 12/31/2024

Baseline Stability
Abs
0.0050
0.0045
0.0040
0.0035
0.0030
0.0025
0.0020
0.0015
0.0010
0.0005
0.0000
-0.0005
-0.0010
-0.0015
-0.0020
-0.0025
-0.0030
-0.0035
-0.0040
-0.0045
-0.0050
s
0 500 1000 1500 2000 2500 3000 3500
Sample: Baseline Stability
File name: BaseStability_20241231142318.GDT
Run Date: 13:23:14, 12/31/2024
Operator: SERVICE
Comment:

Instrument
Model: U-3900 Spectrophotometer
Serial Number: 2338-001
ROM Version: 2J25300 07

Instrument Parameters
Measurement Type: Time Scan
Data Mode: Abs
Scan Time: 3600 s
Wavelength: 500.00 nm
Sampling Interval: 1.0 s
Delay: 0s
Auto Zero before each run: Off
Slit Width: 2 nm
PMT Voltage: Auto
Lamp Change Mode: Auto
Lamp Change Wavelength: 340.00 nm
High Resolution: Off
D2 Lamp: On
WI Lamp: On
R/S Reverse: Off
Path Correct: Off
Path Length: 10.0 mm

Baseline Stability 0.00021 Abs/h (0.00025 - 0.00004) Tolerance within 0.0002 Abs/h NG