Canadian Partnership for Quality Radiotherapy
Technical Quality Control Guidelines
for Medical Linear Accelerators and Multileaf Collimators
          A guidance document on behalf of:
      Canadian Association of Radiation Oncology
      Canadian Organization of Medical Physicists
Canadian Association of Medical Radiation Technologists
          Canadian Partnership Against Cancer
                      July 20, 2016
                    MLA.2016.07.03
                     www.cpqr.ca
Disclaimer
All information contained in this document is intended to be used at the discretion of each individual
centre to help guide quality and safety program improvement. There are no legal standards supporting
this document; specific federal or provincial regulations and license conditions take precedence over the
content of this document. As a living document, the information contained within this document is subject
to change at any time without notice. In no event shall the Canadian Partnership for Quality Radiotherapy
(CPQR) or its partner associations, the Canadian Association of Radiation Oncology (CARO), the Canadian
Organization of Medical Physicists (COMP), and the Canadian Association of Medical Radiation
Technologists (CAMRT), be liable for any damages, losses, expenses, or costs whatsoever arising in
connection with the use of this document.
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
Expert Reviewers
Charles Kirkby
Jack Ady Cancer Centre, Lethbridge, Alberta
Esmaeel Ghasroddashti
Jack Ady Cancer Centre, Lethbridge, Alberta
Crystal Angers
Ottawa Hospital, Ottawa, Ontario
Grace Zeng
Peel Regional Cancer Centre, Trillium Health Partners/Credit Valley Hospital, Mississauga, Ontario
Erin Barnett
Stronach Regional Cancer Centre, Newmarket, Ontario
External Validation Centres
BC Cancer Agency – Fraser Valley Centre, Surrey, British Columbia
BC Cancer Agency – Vancouver Island Centre, Victoria, British Columbia
BC Cancer Agency – Vancouver Centre, Vancouver, British Columbia
Centre de santé et de services sociaux de Trois-Rivières, Trois-Rivières, Quebec
Odette Cancer Centre, Toronto, Ontario
Introduction
The Canadian Partnership for Quality Radiotherapy (CPQR) is an alliance amongst the three key national
professional organizations involved in the delivery of radiation treatment in Canada: the Canadian
Association of Radiation Oncology (CARO), the Canadian Organization of Medical Physicists (COMP), and
the Canadian Association of Medical Radiation Technologists (CAMRT). Financial and strategic backing is
provided by the federal government through the Canadian Partnership Against Cancer (CPAC), a national
resource for advancing cancer prevention and treatment. The mandate of the CPQR is to support the
universal availability of high quality and safe radiotherapy for all Canadians through system performance
improvement and the development of consensus-based guidelines and indicators to aid in radiation
treatment program development and evaluation.
This document contains detailed performance objectives and safety criteria for Medical Linear
Accelerators and Multileaf Collimators. Please refer to the overarching document Technical Quality
Control Guidelines for Canadian Radiation Treatment Centres(1) for a programmatic overview of technical
                                                                                           Page 3 of 15
                                                                                        MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
quality control, and a description of how the performance objectives and criteria listed in this document
should be interpreted.
System Description
Medical linear accelerators (linacs) are cyclic accelerators which accelerate electrons to kinetic energies
from 4 MeV to 25 MeV using non-conservative microwave radio frequency (RF) fields in the frequency
range from 103 MHz (L band) to ~104 MHz (X band), with the vast majority running at 2856 MHz
(S band).(2–5) In a linear accelerator the electrons are accelerated following straight trajectories in special
evacuated structures called accelerating waveguides. Electrons follow a linear path through the same,
relatively low potential difference several times; hence, linacs also fall into the class of cyclic accelerators
just like the other cyclic machines that provide curved paths for the accelerated particles (e.g., betatron).
The high power RF fields used for electron acceleration in the accelerating waveguides, are produced
through the process of decelerating electrons in retarding potentials in special evacuated devices called
magnetrons or klystrons.
Various types of linacs are available for clinical use. Some provide x rays only in the low megavoltage range
(4 MV or 6 MV) while others provide both x rays and electrons at various megavoltage energies. A typical
modern high-energy linac will provide two or three photon energies (usually a combination of a low
[4 to 10 MV] and a high [12 to 25 MV] photon beam) and several electron energies (ranging from
4 to 22 MeV).
Included in the scope of this document are multileaf collimators (MLCs); computer-controlled devices
capable of providing photon beam shielding for linear accelerators using high density leaves (typically
tungsten alloy) which are projected into the radiation field.(6–8) In addition to static beam shaping, beam
intensity modulation can also be achieved by adjusting the position of the MLC in the radiation field
between treatment fields (step and shoot, or static intensity-modulated radiation therapy [IMRT]), by
moving the leaves across the field with varying velocities during the beam-on time (dynamic IMRT), or by
varying the dose rate, gantry speed, and MLC leaf positions during arc delivery (volumetric modulated arc
therapy [VMAT]). By doing this, a desired fluence pattern can be approximated within certain physical
limits.
Current MLC systems vary with respect to design, location, and use. They may be installed as a tertiary
device below the secondary collimators, or they may comprise a total or partial replacement of the
secondary collimators. The leaves must provide an acceptable degree of beam attenuation, provide a
large enough field coverage, and must be well integrated with the rest of the collimator shaping system.
In order to minimize penumbra, various design considerations have been devised by manufacturers to
provide focused field shaping.
Computer control is a key component of the MLC, particularly during the delivery of dynamic treatments.
There must be feedback on the leaf position and beam interlock capabilities when leaf misplacement is
                                                                                                 Page 4 of 15
                                                                                              MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
detected. In addition, there must be interlock capabilities to detect leaf carriage positions that could lead
to unintentional irradiation outside the shielded area. Other safety interlocks must recognize the
unintentional use of the MLC in electron mode and incorporate the use of the MLC in port-film mode.
Related Technical Quality Control Guidelines
In order to comprehensively assess medical linear accelerator performance, additional guideline tests, as
outlined in related CPQR Technical Quality Control (TQC) guidelines must also be completed and
documented, as applicable. Related TQC guidelines, available at cpqr.ca, include:
    •    Safety Systems
    •    Major Dosimetry Equipment
    •    Accelerator-Integrated Cone-Beam Systems for Verification Imaging
    •    Patient-Specific Dosimetric Measurements for Modulated Therapies
Test Tables
Table 1: Daily Quality Control Tests
  Designator                             Test                                       Performance
                                                                            Tolerance             Action
 Daily
        DL1      Motion interlock                                                    Functional
        DL2      Couch brakes                                                        Functional
        DL3      Beam interrupt/counters                                             Functional
        DL4      Lasers/crosshairs                                           1 mm                 2 mm
        DL5      Optical distance indicator                                  1 mm                 2 mm
        DL6      Optical back pointer                                        2 mm                 3 mm
        DL7      Field definition: Jaws/MLC leaves                           1 mm                 2 mm
        DL8      Output constancy – photons                                    2%                  3%
        DL9      Output constancy – electrons                                  2%                  3%
                 Dynamic (Varian), Virtual (Siemens) or Universal
     DL10                                                                      2%                  3%
                 (Elekta) wedge factors
                                                                                              Page 5 of 15
                                                                                           MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
Notes on Daily Tests
 DL1           This test establishes that motion-enabling features on the linac (e.g., those that allow
               the gantry to rotate only under desired conditions) are operational. These include
               functionality tests of couch and hand-pendent controls and the proper engagement of
               collision interlocks when touch guards are engaged.
 DL2           A functional test is performed to establish that brakes on the treatment couch engage
               when desired and prevent the couch from floating freely or moving when a small force
               is applied.
 DL3           This test demonstrates (when applicable): the key interlock prevents the linac from
               irradiating; the non-emergency beam interruption system stops the beam; and the
               beam terminates after a predefined number of monitor units as verified by a backup
               monitor unit counter and/or timer if applicable.
 DL4           This test establishes the alignments of crosshairs with appropriate lasers are within the
               specified limits.
 DL5           At gantry angle 0°, the test demonstrates that the optical distance indicator identifies
               the isocentre plane within the specified limits.
 DL6           This test verifies the performance accuracy of the optical back pointer for applicable
               units.
 DL7           Gantry angle 0°, 100 cm source-axis distance (SAD). This test demonstrates the field
               edges are accurately defined by jaws and/or MLC leaves. It is sufficient to confirm a
               predefined field shape using the projected light field at isocentre. Tolerance and action
               levels apply to each edge of a rectangular field at isocentre as defined by the jaws/MLC
               leaves. Note that systems with a tertiary collimation MLC system will require both jaw
               and MLC leaf positions to be verified.
 DL8           Output constancy must be verified for all photon energies in use on the particular
               treatment day. Measurement is to be conducted using standard local geometry using a
               dosimetry system calibrated against the local secondary standard system.
 DL9           Output constancy must be verified for all electron energies in use on the particular
               treatment day. Measurement is to be conducted using standard local geometry using a
               dosimetry system calibrated against the local secondary standard system.
 DL10          Wedge factors for a representative set of dynamic or virtual soft wedges in use on a
               particular treatment day must be verified. Machine design characteristics must be
                                                                                           Page 6 of 15
                                                                                        MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
               considered when determining the representative set. Alternatively, a test cycle
               designed to test the full range of wedges over multiple days may be considered. Daily
               wedge factors for universal wedges are required to ensure functionality and position
               reproducibility.
Table 2: Monthly Quality Control Tests
  Designator                             Test                                   Performance
                                                                      Tolerance               Action
 Monthly
     ML1        Wedge, tray, cone, interlocks                                    Functional
     ML2        Accessories integrity and centring                               Functional
     ML3        Gantry angle readouts                                    0.5°                  1.0°
     ML4        Collimator angle readouts                                0.5°                  1.0°
                Cross-hairs centring/collimator rotation isocentre
     ML5                                                                1 mm                  2 mm
                (mechanical)
     ML6        Couch position readouts                                 1 mm                  2 mm
     ML7        Couch rotation isocentre (mechanical)                   1 mm                  2 mm
     ML8        Couch isocentric angle                                   0.5°                  1.0°
     ML9        Optical distance indicator                              1 mm                  2 mm
    ML10        Relative dosimetry                                       2%                    3%
    ML11        Central axis depth dose reproducibility               1%/2 mm            2%/3 mm
    ML12        Beam profile constancy                                   2%                    3%
    ML13        Light/radiation coincidence                             1 mm                  2 mm
    ML14        Jaw position accuracy                                   1 mm                  2 mm
    ML15        Backup jaw position accuracy (Elekta)                   1 mm                  2 mm
    ML16        MLC leaf position accuracy                              1 mm                  2 mm
    ML17        Dynamic leaf position accuracy (picket fence)          0.5 mm                 1 mm
                                                                        95%                 95%
    ML18        Dynamic MLC fluence delivery
                                                                     ≤ 3%/3 mm           ≤ 5%/3 mm
                Variation of dose rate, gantry speed, MLC leaf        See note:           See note:
    ML19
                speed and position during arc delivery                  ML19                ML19
    ML20        Records                                                          Complete
                                                                                          Page 7 of 15
                                                                                       MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
Notes on Monthly Tests
 ML1           Verify the functionality of latching interlocks (includes verification that electron beams
               cannot be turned on unless the MLC leaves are retracted).
 ML2           Verify the physical integrity and centring of accessories, including wedges, trays, and
               cones, as appropriate.
 ML3           The accuracy of the digital and mechanical (if used clinically) gantry angle readouts must
               be verified for at least 0°, 90°, 180°, and 270°. The coordinate system convention should
               also be verified.
 ML4           The accuracy of the digital and mechanical (if used clinically) collimator angle readouts
               must be verified for at least 0°, 90°, 180°, and 270°. The coordinate system convention
               should also be verified.
 ML5           This test establishes the correct centring of the crosshairs as well as the mechanical axis
               of rotation of the collimator. Tolerance and action levels refer to the maximum diameter
               of the mechanical isocentre and the maximum displacement of the crosshairs projection
               from the centre of the mechanical isocentre circle.
 ML6           Mechanical and digital couch position readouts must be verified over an appropriate
               clinical range in the directions of the three cardinal axes. Also verify coordinate system
               convention.
 ML7           Isocentric rotation of the couch about the collimator rotation axis must be verified.
               Similar to ML5, the tolerance and action levels refer to the maximum displacement of
               crosshairs projection from the initial position in the isocentre plane.
 ML8           Mechanical and digital couch isocentric rotation angle readouts must be verified over
               the applicable clinical range. Also verify coordinate system convention.
 ML9           A mechanical device, calibrated against the true radiation isocentre, is used to provide
               the base reading for the check of the optical distance indicator. The standards stated in
               the Table apply at the isocentre. The optical distance indicator should be checked over
               a clinically relevant range of source-to-skin distances (SSDs) and gantry angles. The
               tolerance and action levels may be twice as large (i.e., 2 mm and 4 mm) at the clinical
               limits of the optical distance indicator’s range.
 ML10          Using a dosimetry system calibrated against the local secondary standard, the output of
               all clinical beams is checked against yearly reference dosimetry.
                                                                                            Page 8 of 15
                                                                                         MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
 ML11          Measurements are made to confirm that the depth dose has not changed since
               commissioning the unit. Tolerance and action levels are specified in percentages for
               photon beams and in millimetres for electron beams. A single ratio of doses taken at
               clinically relevant depths is sufficient for these measurements. Alternatively, a
               tissue-phantom ratio (TPR) measurement or a check of profile constancy at a shallow
               depth could be used, and the tolerance and action levels adjusted appropriately.
 ML12          This test replaces testing of flatness and symmetry and is intended to be consistent with
               the testing suggested in American Association of Physicists in Medicine (AAPM) protocol
               TG-142.(9) The goal is to ensure that profiles are delivered in a manner consistent with
               that modelled in the associated treatment planning system. Tolerance and action levels
               refer to differences from commissioning (or baseline) profiles as defined in the AAPM
               protocol TG-142.(9) Separate tests are required for all clinically applicable beams.
 ML13          Geometric alignment of the radiation and optical field edges must be established over
               a range of field sizes. Tolerance and action levels apply to each edge of a rectangular
               field.
 ML14          Accuracy of the radiation position of the jaw must be established over a range of jaw
               positions. The number of positions tested shall be determined from the jaw calibration
               method. In conjunction with this test it is important to establish acceptable dose profiles
               for abutting fields at the 0 position. Here the 2 mm action level for each jaw is generally
               not sufficient since in principle, abutting fields could have a difference of up to 4 mm
               between field edges, which can lead to unacceptable peaks or valleys in dose
               distributions. A tolerance of 5% and an action level of 10% in dose profile deviations for
               abutting fields are suggested.
 ML15          Accuracy of the radiation position of the backup jaw must be established over a range
               of positions. The number of positions tested shall be determined from the jaw
               calibration method.
 ML16          Accuracy of the radiation position of the MLC leaf edges must be established over a
               range of MLC positions. The number of MLC positions tested shall be determined from
               the MLC calibration method. For some MLC designs this test may be accomplished by
               evaluating the radiation position of each leaf relative to a reference leaf.
 ML17          For dynamic MLC IMRT, leaf gap accuracy for all leaf pairs is verified via inspection of a
               two-dimensional dose map of a picket fence pattern delivered at gantry angle of 0°.
 ML18          Specific to IMRT, this test demonstrates that the interplay of leaf velocity, gap width,
               gap position, and beam holds combine to deliver a planar dose map consistent with the
                                                                                             Page 9 of 15
                                                                                          MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
               prediction of the treatment planning system. A test plan should consider extreme
               conditions (e.g., the highest levels of modulation used clinically for each leaf pair). An
               acceptable alternative to this test is the regular (more than once per month)
               measurement of patient-specific, dynamic MLC IMRT fields. Tolerance and action levels
               are defined via the gamma metric comparing dose map differences (plan versus
               measurement). Dose maps are defined with region of interest threshold of 10% of the
               maximum dose. Dose differences are global (i.e., with respect to maximum dose).(10)
               Detector resolution must be sufficient to identify performance of individual leaves. As
               with all tests, tolerance and action levels may be tightened at the user’s discretion.
 ML19          The synchronicity of all dynamic parameters during arc delivery. Parameters may be
               evaluated independently, using a subset of the tests described by Ling et al. (11) or
               Bedford and Warrington,(12) or by the repeat delivery of a standard VMAT plan of
               suitable complexity, similar to test ML18. Tolerance and action levels are in reference
               to the consistency of dose delivered at different dose rate, gantry or MLC speeds.
               Tolerance levels should be based on the performance of the linear accelerator, whereas
               action levels should be set to achieve an overall precision consistent with other monthly
               tests (approximately 3%/2 mm from baseline).
 ML20          Documentation relating to the daily quality control checks, preventive maintenance,
               service calls, and subsequent checks must be complete, legible, and the operator
               identified.
Table 3: Annual Quality Control Tests
 Designator                               Test                                   Performance
                                                                         Tolerance           Action
 Annual
     AL1        Profile reproducibility                                     2%                 3%
     AL2        Depth dose reproducibility                                  1%                 2%
     AL3        Reference dosimetry                                         1%                 2%
     AL4        Relative output factor reproducibility                      1%                 2%
     AL5        Wedge transmission factor reproducibility                   1%                 2%
     AL6        Accessory transmission factor reproducibility               1%                 2%
     AL7        Wedge profile reproducibility                               1%                 2%
                Profile and output reproducibility versus gantry
     AL8                                                                    1%                 2%
                angle
                                                                                           Page 10 of 15
                                                                                         MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
       AL9     Monitor chamber linearity                                 1%/1 MU            2%/2 MU
       AL10    End monitor effect                                         0.5 MU              1 MU
       AL11    Collimator rotation isocentre (radiation)                   1 mm               2 mm
               Gantry rotation isocentre (mechanical and
       AL12                                                                1 mm               2 mm
               radiation)
       AL13    Couch rotation isocentre (radiation)                        1 mm               2 mm
               Coincidence of radiation and mechanical
       AL14                                                                1 mm               2 mm
               isocentres
       AL15    Coincidence of axes of rotation                             1 mm               2 mm
       AL16    Couch deflection                                            3 mm               5 mm
       AL17    Leaf transmission (all energies)                             1%                 2%
       AL18    Leakage between leaves (all energies)                        2%                 3%
       AL19    Transmission through abutting leaves                         2%                 3%
       A20     MLC leaf alignment with jaws                                 0.5°                1°
       A21     Dosimetric leaf gap                                        0.2 mm              0.3 mm
       AL22    Independent quality control review                                  Complete
Notes on Annual Tests
 AL1           This test establishes that an appropriate subset of the crossplane and inplane profiles at
               gantry angle 0° are consistent with water-tank measurements made at the time of
               commissioning. Tolerance and action levels refer to differences from commissioning or
               baseline. Measurements should be made for all clinically operable beams.
 AL2           Depth dose scans necessary for calibration protocols (alternatively TPR measurements)
               are also made and used to verify consistency with commissioning/baseline water-tank
               measurements. Tolerance and action levels refer to differences from commissioning or
               baseline. Measurements should be made for all clinically operable beams.
 AL3           A full absolute dosimetry output calibration based on an internationally accepted
               protocol (e.g., AAPM TG-51)(13) must be performed annually on each energy used
               clinically for both photons and electrons. Independence of output with respect to dose
               rate (pulse repetition frequency) must also be established across clinically applicable
               dose rates.
                                                                                           Page 11 of 15
                                                                                         MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
 AL4           An appropriate subset of relative output factors are confirmed to be consistent with
               commissioning measurements.
 AL5           The wedge transmission factors (if applicable) are confirmed to be consistent with
               commissioning measurements.
 AL6           Transmission factors are confirmed to be consistent with commissioning
               measurements. Discretion may be used. Devices where the physical
               composition/dimension can be confirmed not to have changed since a previous
               measurement need not be measured again.
 AL7           This test applies to moving jaw (dynamic and virtual) and universal (Elekta) wedges. This
               test confirms that wedged fields produce profiles that are consistent with baseline data
               through the central 80% of the field for all clinically used wedge angles.
 AL8           This test establishes the independence of output with gantry angle. It requires that
               output be measured under identical conditions (e.g., dosimeter under the same amount
               of buildup material in each position) and that the difference from the gantry at 0°
               position be within the specified limits. In addition to central axis output, beam profiles
               shall be measured at three cardinal gantry angles: 0°, 90°, and 270°. Measurements
               should be made for all clinically operable beams.
 AL9, 10       From a series of radiation measurements with different monitor units the linearity and
               the end monitor effect are determined. The larger of the percentage or absolute value
               is taken as what is applicable. Measurements should be made for all clinically operable
               beams.
 AL11          Commonly measured using a star shot technique this test determines the diameter of
               the circle that encompasses the radiation isocentre of the collimator as it is rotated
               through an appropriate sample of angles within its full range of motion. The diameter
               must be within specifications.
 AL12          This test determines the diameter of both the mechanical and the radiation isocentre
               defined by gantry rotation through its full clinical range of motion. Each diameter must
               be within specifications.
 AL13          This test determines the diameter of the radiation isocentre defined by couch rotation
               through its full clinical range of motion. The diameter must be within specifications.
 AL14          The coincidence of radiation and mechanical isocentres is established for the collimator,
               gantry and couch; and must meet the specified limits.
                                                                                           Page 12 of 15
                                                                                         MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
 AL15          The three axes of rotation (the collimator/MLC, the couch, and the gantry) must meet
               within a sphere of the specified diameter.
 AL16          Couch deflection is measured as a difference in surface position (load versus no load) of
               the couch extended longitudinally at least 30 cm through isocentre. Under “load” is
               considered as a typical patient mass (approximately 70 kg) distributed over the couch
               or placed at the centre. Tolerance and action levels are defined relative to the deflection
               measured at the time of commissioning.
 AL17          The average and maximum MLC leaf transmission is verified in this test for all photon
               energies and compared with the values established at the time of commissioning or the
               values adopted in the treatment planning system. Tolerance and action levels refer to
               changes from the commissioning measurements.
 AL18          The average and maximum leakage between adjacent, closed MLC leaves is verified in
               this test for all photon energies and compared with the values established at the time
               of commissioning or the values adopted in the treatment planning system. Tolerance
               and action levels refer to changes from the commissioning measurements.
 AL19          The average and maximum leakage between abutting closed MLC leaves is verified in
               this test for all photon energies and compared with the values established at the time
               of commissioning or the values adopted in the treatment planning system. Tolerance
               and action levels refer to changes from the commissioning measurements.
 AL20          Use a leaf pattern where one leaf from each leaf bank protrudes well into the field.
               Confirm the leaf edge parallelism with the collimator or solid jaw edge.
 AL21          A dynamic leaf gap test (sometimes referred to as a dosimetric leaf gap test) is
               performed to confirm consistency with baseline measurements. The minimum standard
               is to establish this using a single detector (e.g., an ion chamber) method, although
               methods that calculate separate factors for each leaf pair may be employed. The value
               should be consistent within tolerance for all four cardinal gantry angles.
 AL22          To ensure redundancy and adequate monitoring, a second qualified medical physicist
               must independently verify the implementation, analysis, and interpretation of the
               quality control tests at least annually.
                                                                                           Page 13 of 15
                                                                                         MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
Acknowledgements
CPQR would like to thank the many people who participated in the production of this guideline. These
include: Kevin Diamond, Laurent Tantôt, and Kyle Malkoske (associate editors); the Quality Assurance and
Radiation Safety Advisory Committee; the COMP Board of Directors, Erika Brown and the CPQR Steering
Committee, and all individuals that submitted comments during the community review of this guideline.
References
1. Canadian Partnership for Quality Radiotherapy. Technical quality control guidelines for Canadian
   radiation treatment centres. 2016 May 1. Available from: http://www.cpqr.ca/programs/technical-
   quality-control
2.   Podgorsak EB, Metcalfe P, Van Dyk J. Medical accelerators. In: Van Dyk J, editor. The modern
     technology of radiation oncology: a compendium for medical physicists and radiation oncologists.
     1st ed. Madison (WI): Medical Physics Publishing; 1999. p. 349–435.
3.   International Electrotechnical Commission. IEC 60976 International standard: medical electronic
     equipment – medical electron accelerators – functional performance characteristics. 2nd ed.
     Geneva (Switzerland): The Commission; 2007.
4.   Karzmark CJ, Morton RJ. A primer on theory and operation of linear accelerators in radiation
     therapy. Madison (WI): Medical Physics Publishing; 1998.
5.   Metcalfe P, Kron T, Hoban P. The physics of radiotherapy x-rays from linear accelerators. Madison
     (WI): Medical Physics Publishing; 1997.
6.   Boyer AL. Basic application of a multileaf collimator. In: Mackie TR, Palta JR, editors. Teletherapy:
     present and future. 1st ed. Madison (WI): Advanced Medical Publishing; 1996. p. 403–444.
7.   Boyer AL, Biggs P, Galvin J, Klein E, LoSasso T, Low D, Mah K, Yu C. Basic applications of multileaf
     collimators – report of Task Group no. 50, Radiation Therapy Committee, AAPM report no. 72.
     Madison (WI): Medical Physics Publishing; 2001.
8.   Boyer AL, Xing L, Xia P. Beam shaping and intensity modulation. In: Van Dyk J, editor. The modern
     technology of radiation oncology: a compendium for medical physicists and radiation oncologists.
     1st ed. Madison (WI): Medical Physics Publishing; 1999. p. 437–479.
9.   Klein EJ, Hanley J, Bayouth, Yin FF, Simon W, Dresser S, Serago C, Aguirre F, Ma L, Arjomandy B, Liu
     C, Sandin C, Holmes T. Task Group 142 report: quality assurance of medical accelerators. Med Phys.
     2009;36(9):4197–212.
10. Van Dyk J, Barnett RB, Cygler JE, Shragge PC. Commissioning and quality assurance of treatment
    planning computers. Int J Radiat Oncol Biol Phys. 1993;26(2),261–73.
                                                                                             Page 14 of 15
                                                                                           MLA.2016.07.03
Technical Quality Control Guidelines for Medical Linear Accelerators and Multileaf Collimators
Part of the Technical Quality Control Guidelines for Canadian Radiation Treatment Centres Suite
11. Ling CC, Zhang P, Archambault Y, Bocanek J, Tang G, Losasso T. Commissioning and quality
    assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys. 2008;72(2):575–
    81.
12. Bedford JL, Warrington AP. Commissioning of volumetric modulated arc therapy (VMAT). Int J Radiat
    Oncol Biol Phys. 2009;73(2):537–45.
13. Almond PR, Biggs PJ, Coursey BM, Hanson WF, Saiful Huq M, Nath MR, Rogers DWO. AAPM’s TG-51
    protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999;
    26(9):1847–70.
                                                                                          Page 15 of 15
                                                                                        MLA.2016.07.03