Section A
Section A
LABORATORY
QUALITY ASSURANCE/QUALITY CONTROL
6.0 REFERENCES.............................................................................................................................22
6.1 QA/QC General .............................................................................................................. 22
6.2 Method Detection Limits ................................................................................................ 22
6.3 Blank Correction............................................................................................................. 22
A-1
7.0 REVISION HISTORY ...................................................................................................................23
LIST OF TABLES
A-2
Quality Assurance
Revision Date: October 2015
1.0 INTRODUCTION
This chapter specifies essential QA/QC activities to enable laboratories to achieve reproducible results on
an ongoing basis. The first part of Section E ‘Microbiological Examination’ discusses QA/QC as they
pertain to microbiological methods.
Guidelines alone cannot guarantee high quality results. Common sense steps to avoid contamination,
ongoing programs of staff training, and proactive method improvement procedures are also essential.
All laboratories should regularly participate in cooperative inter-laboratory studies, and standard reference
materials should be analyzed frequently, preferably on a blind or semi-blind basis. Large laboratories
should have a full time QA Officer.
The procedures for determining method detection limits and for handling blank corrections are
controversial. Different procedures and calculation algorithms are being used by various laboratories.
A Quality Manual shall be set up that documents all resources, policies and procedures making up the
Quality System. The Quality Manual is to include detailed descriptions of the topics outlined in this
section and clearly define the responsibilities of management, supervisory staff, and laboratory staff with
respect to the quality systems. It shall be reviewed and updated regularly.
The laboratory record system shall be designed to ensure sample, analytical data, and analyst
traceability, including dates and analysts’ initials or signatures. Dated and signed material shall include
forms, instrumental records and printouts, as well as notebooks.
The sample numbering system of the laboratory must be designed to eliminate the possibility of a sample
mix-up.
The record storage system should be designed for easy retrieval. A policy on the length of storage and
disposal of records should be established.
A-3
Potential deficiencies in sample history requirements shall be monitored and noncompliance must be
identified and any affected analytical data flagged.
a. Sample collection
Documentation accompanying all samples should include a test requisition form, or should
comply with chain of custody requirements.
b. Sample containers
Inspect sample containers for damage; record and report temperature on receipt where
appropriate. Where critical, temperature limits should be established and maximum/minimum
thermometers should accompany the samples to document compliance with the limits.
b. Storage conditions
Temperatures of refrigerated and frozen storage facilities shall be monitored and recorded daily.
c. Sample Pre-treatment
Complete records shall be maintained to ensure that potential problems, including cross-
contamination, are traceable.
The hold times and preservation listed below in Table 1 will take precedent over the hold times
and preservation criteria listed in the individual methods. The intent of this table is to ensure that
hold times, sample containers and preservation practices reflect the most current and approved
treatment.
A-4
Table 1. Summary of sample preservation and hold time requirements
BC MOE SAMPLE PRESERVATION & HOLDING TIME REQUIREMENTS(1,2) Version: 06-Nov-2015
Holding
Sample Storage (4)
Parameter Name (3) Preservation Time References
Container Temp
(days)
Water
Physical & Aggregate Properties
Acidity Plastic, Glass ≤6ºC none 14 days APHA
Alkalinity Plastic, Glass ≤6ºC none 14 days APHA
Colour Plastic, Glass ≤6ºC none 3 days APHA / BC MOE
Conductivity Plastic, Glass ≤6ºC none 28 days APHA
pH Plastic, Glass ≤6ºC none 15 minutes APHA
Solids (Total, TSS, TDS, Fixed, Volatile, etc.) Plastic, Glass ≤6ºC none 7 days APHA
Turbidity Plastic, Glass ≤6ºC none 3 days EPA 40CFR 2012 / BC MOE
UV Transmittance Plastic, Glass ≤6ºC none 3 days APHA / BC MOE
Inorganic Non-metallics
Bromide Plastic, Glass no requirement none 28 days EPA 300.1
Chloride Plastic, Glass no requirement none 28 days APHA / EPA 300.1
Chlorate, Bromate Plastic, Glass ≤6ºC 50 mg/L EDA 28 days EPA 317.0
Chlorine, Total Residual (Free Chlorine) Plastic, Glass none none 15 minutes APHA
Chlorite Plastic, Amber Glass ≤6ºC 50 mg/L EDA 14 days EPA 317.0
field NaOH, store in dark 14 days APHA
Cyanide (SAD, WAD) Plastic, Glass ≤6ºC
none 24 hours APHA
Dissolved Oxygen (Winkler Method) Glass BOD bottle ≤6ºC Winkler kit, store in dark 8 hours APHA
Fluoride Plastic no requirement none 28 days APHA / EPA 300.1
H2SO4 28 days APHA
Nitrogen, Nitrate + Nitrite Plastic, Glass ≤6ºC
none 3 days BC MOE
H2SO4 28 days APHA
Nitrogen, Ammonia Plastic, Glass ≤6ºC
none 3 days BC MOE
Nitrogen, Nitrate Plastic, Glass ≤6ºC, do not freeze none 3 days APHA / BC MOE
Nitrogen, Nitrite Plastic, Glass ≤6ºC, do not freeze none 3 days APHA / BC MOE
H2SO4 28 days APHA
Nitrogen, Total Kjeldahl Plastic, Glass ≤6ºC
none 3 days BC MOE
H2SO4 28 days APHA
Nitrogen, Total, Persulfate Method Plastic, Glass ≤6ºC
none 3 days BC MOE
HCl 28 days APHA
Nitrogen, Total, Combustion Method Plastic, Glass ≤6ºC
none 3 days BC MOE
Phosphorus, Dissolved (Orthophosphate) Plastic, Glass ≤6ºC filter (field or lab) 3 days EPA 40CFR 2012 / BC MOE
Phosphorus, Total Reactive (Orthophosphate) Plastic, Glass ≤6ºC none 3 days APHA / BC MOE
filter, H2SO4 28 days APHA
Phosphorus, Total Dissolved Plastic, Glass ≤6ºC
none 3 days BC MOE
H2SO4 28 days APHA
Phosphorus, Total Plastic, Glass ≤6ºC
none 3 days BC MOE
Silica, Reactive Plastic ≤6ºC, do not freeze none 28 days APHA
Sulfate Plastic, Glass ≤6ºC none 28 days APHA / SW846 Ch3 2007
Sulfide Plastic, Glass ≤6ºC ZnAc / NaOH to pH >9 7 days APHA
Metals
1 mL 50% NaOH per 125 mL 30 days EPA 1669
Hexavalent Chromium Plastic, Glass ≤6ºC
none 24 hours APHA
Metals, Total Plastic, Glass no requirement HNO3 (7) 180 days APHA / EPA 200.2
field filter 0.45 um + HNO3,
Metals, Dissolved Plastic, Glass no requirement 180 days APHA
qualify if lab-filtered (7)
(8)
Mercury, Total Glass, PTFE no requirement HCl or BrCl 28 days APHA / EPA 1631E
field filter 0.45 um + HCl or BrCl,
Mercury, Dissolved Glass, PTFE no requirement 28 days APHA / EPA 1631E
qualify if lab-filtered (8)
Aggregate Organics
HNO3, store in dark,
Adsorbable Organic Halides (AOX) Amber Glass ≤6ºC sodium sulfite if chlorinated, 14 days APHA 5320
collect with no headspace
Biochemical Oxygen Demand (BOD) Plastic, Glass ≤6ºC, do not freeze none 3 days APHA / BC MOE
Carbonaceous Biochemical Oxygen Demand (CBOD) Plastic, Glass ≤6ºC, do not freeze none 3 days APHA / BC MOE
filter, H2SO4 or HCl 28 days APHA
Carbon, Dissolved Organic Plastic, Glass ≤6ºC
none 3 days BC MOE
Carbon, Dissolved Inorganic Plastic, Glass ≤6ºC field filter 14 days APHA (alkalinity)
Carbon, Total Organic Plastic, Glass ≤6ºC H2SO4 or HCl 28 days APHA
Carbon, Total Inorganic Plastic, Glass ≤6ºC none 14 days APHA (alkalinity)
H2SO4 (field or lab) 28 days APHA
Chemical Oxygen Demand (COD) Plastic, Glass ≤6ºC
none 3 days BC MOE
Filter Filters: freeze field filter, store in dark Filters: 28 days
Chlorophyll a and Phaeophytin APHA
Dark Plastic, Amber Glass ≤6ºC unfiltered, store in dark 2 days
Surfactants (Methylene Blue Active Substances) Plastic, Glass ≤6ºC none 3 days APHA / BC MOE
Total Phenols (4AAP) Plastic, Glass ≤6ºC H2SO4 28 days APHA
Extractable Hydrocarbons
NaHSO4, HCl, or H2SO4 14 / 40 days EPA 3511
Extractable Hydrocarbons (LEPH, HEPH, EPH) Amber Glass ≤6ºC
none 7 / 40 days SW846 Ch4 2007
Oil and Grease / Mineral Oil and Grease Amber Glass ≤6ºC HCl or H2SO4 28 days EPA 40CFR 2012
Waste Oil Content Amber Glass ≤6ºC none 28 days BC MOE
A-5
BC MOE SAMPLE PRESERVATION & HOLDING TIME REQUIREMENTS(1,2) Version: 04-Jan-2016
Holding
Sample Storage (4)
Parameter Name (3) Preservation Time References
Container Temp
(days)
Individual Organic Compounds
Potassium Dihydrogen Citrate (solid),
EPA 531.2,
~pH 3.8, 9.2-9.5 g/L, 28 days
APHA 6610B
+ 100 mg/L Na2S2O3 if chlorinated
Carbamate Pesticides Amber Glass ≤6ºC ChlorAC buffer, ~pH 3,
1.8mL / 60 mL sample, 28 days EPA 531.1
+ 100 mg/L Na2S2O3 if chlorinated
none,
7 days EPA 8321A
100 mg/L Na2S2O3 if chlorinated
0.5g Ascorbic Acid / L +
14 / 40 days Alberta Env AE130
Chlorinated and Non-chlorinated Phenolics Amber Glass ≤6ºC NaHSO4 or H2SO4
none 7 / 40 days SW846 Ch4 2007
Dioxins / Furans Amber Glass ≤6ºC none unlimited SW846 Ch4 2007
Glyphosate / AMPA Amber Glass or Polypropylene ≤6ºC 100 mg/L Na2S2O3 if chlorinated 14 days APHA 6651B
NaHSO4, HCl, or H2SO4 14 / 40 days EPA 3511
Glycols Glass ≤6ºC
none 7 / 40 days SW846 Ch4 2007
Halogenated Hydrocarbons (Semi-Volatile) Amber Glass ≤6ºC 100 mg/L Na2S2O3 if chlorinated 7 / 40 days SW846 Ch4 2007
HCl (optional), store in dark, APHA 6640A
Herbicides, Acid Extractable Amber Glass ≤6ºC 14 / 21 days
50 mg/L Na2SO3 if chlorinated APHA 6640A
Dark Plastic
Paraquat / Diquat ≤6ºC 100 mg/L Na2S2O3 if chlorinated 7 / 21 days EPA 549.2
(protect from light)
Pesticides (NP, OP, OC) Amber Glass ≤6ºC none 7 / 40 days SW846 Ch4 2007
Polychlorinated Biphenyls (PCBs) Amber Glass ≤6ºC none unlimited SW846 Ch4 2007
NaHSO4, HCl, or H2SO4 14 / 40 days EPA 3511
Polycyclic Aromatic Hydrocarbons (PAHs) Amber Glass ≤6ºC
none 7 / 40 days SW846 Ch4 2007
(0.5g Ascorbic Acid + 0.4g NaOH) / L 14 / 40 days Alberta Env AE129
Resin Acids, Fatty Acids Amber Glass ≤6ºC
none 7 / 40 days SW846 Ch4 2007
43mL Glass VOC Vials 3 mg Na2S2O3 (see BC Lab Manual method
Volatile Organic Compounds (Trihalomethanes) ≤6ºC 14 days BC MOE
(2-3) for more details)
200 mg NaHSO4, or 3 mg Na2S2O3 if
43mL Glass VOC Vials
Volatile Organic Compounds (VOC, BTEX,VH) ≤6ºC chlorinated (see BC Lab Manual method for 14 days BC MOE
(2-3)
other options and details)
Microbiological Parameters
Coliforms, Total, Fecal, and Ecoli Sterile Glass or Plastic <8ºC, do not freeze Na2S2O3 30 hours (5) BC CDC / APHA 9060B
Cryptosporidium, Giardia Sterile Glass or Plastic <8ºC, do not freeze Na2S2O3 96 hours EPA 1623 / APHA 9060B
Enterococcus Sterile Glass or Plastic <8ºC, do not freeze Na2S2O3 30 hours (5) APHA 9060B
Heterotrophic Plate Count Sterile Glass or Plastic <8ºC, do not freeze Na2S2O3 24 hours APHA 9215
Toxicity
Daphnia, Chronic 21day / Chronic EC25 Plastic, Glass (non-toxic) 4±2ºC collect with no headspace 5 days EC EPS 1/RM/14 & 11
Daphnia, LC50 / LT50 Plastic, Glass (non-toxic) 4±2ºC collect with no headspace 5 days EC EPS 1/RM/14 & 11
Microtox Plastic, Glass (non-toxic) 4±2ºC collect with no headspace 3 days EC EPS 1/RM/24
Trout, LC50 Plastic, Glass (non-toxic) 4±2ºC collect with no headspace 5 days EC EPS 1/RM/13 & 9
Trout, LT50 Plastic, Glass (non-toxic) 4±2ºC collect with no headspace 5 days EC EPS 1/RM/13 & 9
A-6
BC MOE SAMPLE PRESERVATION & HOLDING TIME REQUIREMENTS(1,2) Version: 04-Jan-2016
Holding
Sample Storage (4)
Parameter Name (3) Preservation Time References
Container Temp
(days)
Biota
Inorganics
Metals, Total Plastic, Glass freeze (≤ -18C) none 2 years (9) Puget Sound Protocols
Mercury, Total Plastic, Glass freeze (≤ -18C) none 1 year (9) EPA 1631 Appendix
Organics
Semi-Volatile Organic Compounds Glass, PTFE freeze (≤ -18C) none 365 / 40 days Puget Sound Protocols
Volatile Organic Compounds Glass, PTFE freeze (≤ -18C) none 14 days Puget Sound Protocols
Air (Vapours)
VOCs by Canister Sampling SS canister ambient none 30 days EPA TO15
VOCs by Thermal Desorption thermal desorption tube ≤6ºC none 30 days EPA TO17
≤6ºC
VOCs and other Volatile Substances by Charcoal and
see BC Lab Manual Method (or as per applicable none 30 days see BC Lab Manual Method
Miscellaneous Collection Media
reference method)
1
A Director or an Environmental Management Act permit may specify alternate requirements.
2
Refer to applicable BC Environmental Laboratory Manual methods for additional detail. Where differences exist between Lab Manual methods and this table, this table takes precedence.
3
Storage temperature applies to storage at the laboratory. For all tests where refrigeration at ≤ 6°C is required at the laboratory, samples should be packed with ice or cold packs to maintain a temperature of ≤10°C
during transport to the laboratory. The storage of <8°C for microbiological samples applies during storage at the laboratory and during transport to the laboratory. To prevent breakage, water samples stored in glass
should not be frozen. Except where indicated by "do not freeze", test results need not be qualified for frozen samples. Labs may apply a "Cooling Initiated" qualifier on reports to indicate where samples were received
above specified storage temperature, but where sampling occurred < 8 hours before arrival at the lab, and where samples were packed appropriately in coolers with ice or cold packs to initiate the cooling process.
4
Hold Times: Single values refer to hold time from sampling to analysis. Where 2 values are given, the first is hold time from sampling to extraction, and the second is hold time from extraction
to analysis. 3 values are given for TCLP semi-volatiles (1st is from collection to TCLP extraction; 2nd is from TCLP extraction to preparative extraction; 3rd is from preparative extraction to analysis).
5
Samples received from remote locations more than 48 hours after collection must not be tested.
6
Methanol extracts are stable for 40 days from sampling. Hermetic samples must be methanol-extracted within 48 hours of sampling or may be frozen at ≤ -7°C within 48 hours of sampling to
extend hold time to 7 days from sampling. Frozen hermetic samples must be extruded into methanol while still predominantly or partially frozen.
7
If not field-preserved, water samples for metals analysis must be acidified at the lab in their original containers by addition of HNO3 (within 14 days of sampling), then equilibrated at least 16 hours prior to sub-sampling
or analysis (otherwise, qualify as "received unpreserved"). This approach is also applicable to dissolved metals if field filtered. Not applicable to mercury.
8
Use only glass or PTFE containers to collect water samples for total or dissolved mercury. For total mercury, field-preserve with HCl or BrCl. For dissolved mercury, field filter and then preserve
with HCl or BrCl. Adding BrCl to original sample container at the laboratory within 28 days of sampling is an acceptable alternative for total mercury and for dissolved mercury (if field-filtered).
if samples are oxidized for 24 hours prior to sub-sampling or analysis. Dissolved mercury should not be lab-filtered. Qualify lab-filtered results for dissolved mercury as "lab-filtered".
9
Freezing is optional for freeze dried tissue samples and for vegetation that is dried prior to digestion and reported on a dry weight basis; in these cases, samples may be stored at ambient temperature, with
a hold time of 28 days for mercury and 6 months for other metals (based on BC MOE soil guidelines).
A-7
BC MOE SAMPLE PRESERVATION & HOLDING TIME REQUIREMENTS
List of References:
Alberta Env AE129 Resin and Fatty Acids in Pulp Mill Effluents and Receiving Waters, Method AE129.0, Alberta Environment, August 1990.
Alberta Env AE130 Chlorinated Phenolic Compounds in Bleached Kraft Mill Effluents and Receiving Waters, Method AE130.0, Alberta Environment,
April 1991.
APHA Standard Methods for the Examination of Water and Wastewater, American Public Health Association (APHA), the American Water Works
Association (AWWA), and the Water Environment Federation (WEF). Primary reference is Section 1060, Collection and Preservation of
Samples, Table 1060:I, Summary of Special Sampling and Handling Requirements, 2011.
APHA 5320 Dissolved Organic Halogen, Method 5320, APHA Standard Methods, 1997.
APHA 6610B Carbamate Pesticides, Method 6610B, High Performance Liquid Chromatography Method, APHA Standard Methods, 2004.
APHA 6640A Acidic Herbicide Compounds, Method 6640A, Introduction, APHA Standard Methods, 2001 (HCl preservation, 14 day hold time).
APHA 6640A Acidic Herbicide Compounds, Method 6640A, Introduction, APHA Standard Methods, 2001 (reference does not use HCl preservation, but
recommends extraction as soon as possible, up to 14 days, cautions about potential analyte degradation).
APHA 6651B Glyphosate Herbicide, Method 6651B, Liquid Chromatographic Post-Column Fluorescence Method, APHA Standard Methods, 2000.
APHA 9060B Microbiological Examination Section, Samples, Method 9060B, Preservation and Storage, APHA Standard Methods, 2006.
APHA 9215 Heterotrophic Plate Count, Method 9215, APHA Standard Methods, 2004.
ASTM D6418-09 Standard Practice for Using the Disposable En Core Sampler for Sampling and Storing Soil for Volatile Organic Analysis.
BC CDC British Columbia Centre for Disease Control.
BC MOE British Columbia Ministry of Environment (British Columbia Environmental Laboratory Manual).
Carter Carter, Martin R. and Gregorich, E. G., Soil Sampling and Methods of Analysis, Canadian Society of Soil Science, 2008.
CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method, ISBN 1-896997-01-5, Publication No.
1310, Canadian Council of Ministers of the Environment Inc., 2001.
EC EPS 1/RM/24 Biological Test Method: Toxicity Test Using Luminescent Bacteria, Environment Canada, Report EPS 1/RM/24, November 1992.
EC EPS 1/RM/11 Biological Test Method: Acute Lethality Test Using Daphnia spp., Environment Canada, Report EPS 1/RM/11, July 1990 (with May 1996
amendments).
EC EPS 1/RM/14 Biological Test Method: Biological Test Method: Reference Method for Determining Acute Lethality of Effluents to Daphnia magna,
Environment Canada, Report EPS 1/RM/14 second edition, December 2000.
EC EPS 1/RM/13 Biological Test Method: Reference Method for Determining Acute Lethality of Effluents to Rainbow Trout, Environment Canada, Report EPS
1/RM/13, second edition, December 2000 (with May 2007 amendments).
EC EPS 1/RM/9 Biological Test Method: Acute Lethality Test Using Rainbow Trout, Environment Canada, Report EPS 1/RM/ 9, July 1990 (with May 1996 and
May 2007 amendments).
EPA United States Environmental Protection Agency.
EPA 1311 Toxicity Characteristic Leaching Procedure, SW 846 Method 1311, Revision 0, US EPA Office of Solid Waste, July 1992.
EPA 1631E Method 1631, Revision E: Mercury in Water by Oxidation, Purge and Trap, and Cold Vapor Atomic Fluorescence Spectrometry, US EPA
Office of Water, August 2002.
EPA 1669 Sampling Ambient Water for Trace Metals at EPA Water Quality Criteria Levels, Method 1669, US EPA Office of Water, July 1996.
EPA 200.2 Sample Preparation Procedure for Spectrochemical Determination of Total Recoverable Elements, Method 200.2, Revision 2.8, US EPA Office
of Research and Development, 1994.
EPA 300.1 Determination of Inorganic Anions in Drinking Water by Ion Chromatography, Revision 1.0, US EPA Office of Research and Development,
1997.
EPA 3060A Alkaline Digestion for Hexavalent Chromium, SW846 Method 3060A, Revision 1, US EPA Office of Solid Waste, December 1996.
EPA 317.0 Determination of Inorganic Oxyhalide Disinfection By-Products in Drinking Water Using Ion Chromatography with the Addition of a Postcolumn
Reagent for Trace Bromate Analysis, Revision 2.0, US EPA Office of Research and Development, July 2001.
EPA 3511 Organic Compounds in Water by MicroExtraction, SW 846 Method 3511, Revision 0, US EPA Office of Solid Waste, Nov 2002.
EPA 40CFR 2012 Guidelines Establishing Test Procedures for the Analysis of Pollutants Under the Clean Water Act; Analysis and Sampling Procedures, Final
Rule, 40 CFR Parts 136, 260, et al, Table II - Required Containers, Preservation Techniques, and Holding Times, US EPA , May 18, 2012.
EPA 5035A Closed-System Purge-and-Trap and Extraction for Volatile Organics in Soil and Waste Samples, SW 846 Method 5035A, Draft Revision 1, US
EPA Office of Solid Waste, July 2002.
EPA 531.1 Measurement of N-MethylCarbamoyloximes and N-MethylCarbamates in Water by Direct Aqueous Injection HPLC with PostColumn
Derivatization, Revision 3.1, US EPA Office of Ground Water and Drinking Water, 1995.
EPA 531.2 Measurement of N-MethylCarbamoyloximes and N-MethylCarbamates in Water by Direct Aqueous Injection HPLC with PostColumn
Derivatization, Revision 1.0, US EPA Office of Ground Water and Drinking Water, September 2001.
EPA 549.2 Determination of Diquat and Paraquat in Drinking Water by Liquid-Solid Extraction and High Performance Liquid Chromatography with
Ultraviolet Detection, Method 549.2, Revision 1.0, US EPA Office of Research and Development, June 1997.
EPA 8321A Solvent Extractable Nonvolatile Compounds by High Performance Liquid Chromatography / Thermospray / Mass Spectrometry (HPLC/TS/MS)
or Ultravioloet (UV) Detection, Revision 1, US EPA Office of Solid Waste, Dec 1996.
EPA TO15 Determination Of Volatile Organic Compounds (VOCs) In Air Collected In Specially-Prepared Canisters And Analyzed By Gas
Chromatography/Mass Spectrometry (GC/MS), Compendium Method TO-15, Compendium of Methods for the Determination of Toxic Organic
Compounds in Ambient Air, Second Edition, US EPA Office of Research and Development, January 1999.
EPA TO17 Determination of Volatile Organic Compounds in Ambient Air Using Active Sampling Onto Sorbent Tubes, Compendium Method TO-17,
Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air, Second Edition, US EPA Office of Research and
Development, January 1999.
Puget Sound Protocols Recommended Protocols for Measuring Selected Environmental Variables in Puget Sound, Puget Sound Water Quality Action Team, prepared
for U.S. Environmental Protection Agency (Region 10) and U.S. Army Corps of Engineers, July 1996.
SW846 Ch3 2007 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods, SW846, Final Update IV, Chapter 3, Inorganic Analytes, Revision 4, US
EPA Office of Solid Waste, February 2007.
SW846 Ch4 2007 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods, SW846, Final Update IV, Chapter 4, Organic Analytes, Revision 4, US
EPA Office of Solid Waste, February 2007.
A-8
2.5 Test Methods, Existing
- analysis procedure in sufficient detail such that an experienced analyst, unfamiliar with the
specific test procedure, should be able to perform the analysis
- procedures for the preparation of reagent solutions and calibration standard solutions
- operating instructions for instrumentation, which are supplemental to the manufacturer
operating manuals
- requirement for quality control sample preparation and analysis
- current quality control criteria (i.e., acceptable limits)
Where data are kept on computer files, changes to methodology should be reflected by changes in
related computer codes. A system should be established to review methods and documentation on an
annual basis; the system should include periodic re-validation of methods when warranted, such as when
personnel or equipment are changed.
Methods should be periodically reviewed to ensure that the most up-to-date methods are being used.
Test methods that are new to the laboratory, methods that have been developed in-house, and changes
and modifications to existing methods must be validated. Validation should include determination of
method detection limits, linear range, precision, recoveries, interference checks, and performance of
equivalency testing where required.
2.7 Equipment
Log books or other records shall be kept which record daily operating, calibration, and setup parameters.
Instrument operating procedures that supplement instructions given in the manufacturer’s operating
manuals should be documented.
A-9
Initially, proficiency in the performance of an analysis new to the analyst should be demonstrated by the
successful analysis of known samples, standards or certified material prior to the analyst being assigned
routine analysis.
Continuing proficiency may be monitored through the analysis of routine quality control samples, certified
materials, and performance evaluation samples, and also by participation in inter-laboratory studies and
blind audits.
All chemicals shall be reagent grade or better, and must meet specifications identified in the test
methods. Procedures for the preparation of reagent solutions and calibration standard solutions shall be
included as part of the applicable analysis method.
Establish and maintain logs that document the preparation of reagents and standard solutions specifying:
A file of certificates of standard solutions that have been obtained from commercial suppliers shall be
kept.
Prior to routine use, and periodically throughout its shelf life, performance of a reagent or standard shall
be verified. Compare performance of standard solutions by analysis of old and new solutions
consecutively. Set criteria for the response of new versus old standard solutions.
All reagent solutions and standard solutions must be properly labelled. Labels shall identify material,
concentration, date prepared and expiry date. Expiry dates will vary depending upon the solution and
concentration. A general guideline for concentrated stock standard solutions is an expiry date of one
year.
Reagent water shall comply with ASTM D 1193-77, Standard Specification for Reagent Water, Type I,
Type II, or Type III, or Standard Methods 18th Edition (1992), Section 1080 Reagent-Grade Water, Type I
or Type II.
Reagent water must be free from chemical and microbiological substances that interfere with analytical
methods. The presence of contaminants may be monitored for each test by the analysis of blank reagent
water with each batch of samples. Microbiological evaluation may be performed by using a Total Plate
Count.
A - 10
2.12 Volumetric Measurement
Delivery volumes of automatic pipettes and diluters shall be checked on a routine basis and records of
results maintained.
Up-to-date documentation shall be kept on all glassware cleaning procedures and requirements.
The effectiveness of cleaning may be monitored by the analysis of blanks using randomly selected
glassware.
Tests in progress and other laboratory operations should not interfere with or lead to contamination of
other tests in progress. Scheduling of tests, and scheduling the use of fume hoods and other apparatus
and facilities may be required.
The identity of samples and the order in which samples are analyzed on each instrument must be
recorded, so that problems such as cross-contamination may be identified.
The accuracy and stability of calibrations is established by setting requirements for the following, as
appropriate:
- equivalent standard/sample reagent backgrounds
- sufficient number of standards
- low standards (less than ten times the detection limit)
- reagent blanks used to zero response
- appropriate curve fit
- internal standards
- normalization standards
- control and verification standards to verify accuracy and stability
- associated control limits and specified corrective action
Appropriate quality control techniques applied to each batch of samples shall be recorded and
documented. QC data must be made available to the client and it is encouraged that batch QC results be
included in reports.
The type and number of quality control samples to be analyzed should be stated in the analysis method,
or in a section in the overall Quality Manual.
A - 11
Due to the diversity of tests and QC sample requirements, it is not possible to define overall
requirements. However, based on an arbitrary sample batch size of ten samples, two examples (not
recommendations) are:
a. Analysis of pH of water:
Control charts for QC samples must be established. Control Limits may be statistically derived based on
data generated from QC samples within each laboratory, but must not exceed Data Quality Objective
values as defined in the individual test methods within this manual. Commonly used statistically derived
limits are ±2s (for warning limits) and ±3s (for control limits) around the mean. Statistical limits are
recommended for use in trend monitoring if not used directly as Control Limits.
Corrective action must be taken when control limits are exceeded, and records of out-of-control events
and actions taken must be maintained. Test results associated with Quality Control data that does not
meet minimum BC MOE Data Quality Objectives must be appropriately qualified in laboratory reports.
Multi-Element Scan (MES) Qualifiers: As the number of analytes in a test method increases, so does the
chance of a DQO exceedance by random chance as opposed to a legitimate method problem. Thus, in
multi-element test methods (also known as Multi-Element Scans), for Laboratory Control Samples, Matrix
Spikes, or Reference Materials, it is considered acceptable to exceed quoted DQOs by up to 10%
(absolute), for up to 10% of the total number of analytes (rounded down) included within the method. For
example, in a PAH scan of seventeen analytes with LCS acceptance limits of 60-130%, 10% of the
analytes reported by the method (i.e. one analyte) may have a recovery exceeding 60-130% by up to
10% absolute (i.e. recovery of 50-140%). Recurring non-random issues with specific parameters must be
addressed, and will be highlighted by ongoing re-validation assessments. Where applicable
exceedances occur, a suitable qualifier (e.g. “MES”) may be applied to test results.
Table 2 presents BC MOE recommended Data Quality Objectives for laboratory duplicates. These DQOs
are applicable to duplicates where at least one of the duplicate results exceeds five times the
Laboratory’s detection limit. These DQOs were derived from Measurement Uncertainty (MU) estimates
obtained from four BC Laboratories represented on the BC Environmental Laboratory Technical Advisory
Committee (BCELTAC).
Laboratory duplicate values that meet these DQOs may be considered acceptable, falling within the
expected range of variability for the indicated parameter. Duplicate values that exceed these DQOs
should be investigated by the laboratory to ensure validity. Where no issues are identified, duplicate
results that exceed DQOs should be quantified to indicate potential heterogeneity issues with the sample.
Table 2 supersedes laboratory duplicate DQO values from within specific BC Lab Manual Methods.
A - 12
Table 2. Recommended data quality objectives for laboratory duplicates
It is good practice to report measurement uncertainty along with measurement results. When
measurement uncertainty is reported, it should be as the 95% confidence interval.
Result ± 95% CI
Many laboratories prefer to report results following the significant figures convention. When that
convention is employed, the previous estimates should be based on its 95% confidence interval where
practical.
Rounding of results should be deferred until the final calculation for reporting purposes. The ability to
store data, to one more significant figure than would normally be the case, is of benefit to those
undertaking statistical studies of data and its provision is encouraged.
The reporting convention of the analyte must be stated in unambiguous terms, but “+” and “-”symbols for
ions need not be shown unless speciation has been effected. For example:
Copper Cu
Nitrate as N
Ortho-phosphate as P
Silica SiO2
Sulfate SO4
A - 13
Units must also be expressed in unambiguous terms, for example:
2.20 Guidelines for Analytical Parameters Determined by Calculation (Ver.: Jan. 8, 2004)
The BC Ministry of Environment has several numerical criteria that incorporate calculated parameters.
Calculated parameters are those that require mathematical operations to be performed on one or more
measured parameter results in order to determine their value. The most common examples, which are
dealt with in this document, are parameters that are calculated by addition or subtraction from pre-
requisite component results.
This document outlines proposed general guidelines for calculating and reporting these calculated
parameters. These rules may be superseded by specific instructions within individual published BC
Environment laboratory methods.
1. For calculation purposes, use a value of zero for any component result that is reported as less than
a detection limit.
3. For parameters that involve both additions and subtractions, a combination of the rules for both
addition and subtraction should be applied.
4. Detection limits for both calculated and directly analyzed parameters (whether MDL’s, RDL’s,
LOQ’s, or otherwise) should generally be provided to two significant digits. Analytical results should
be provided to the same level of significance. This prevents the introduction of rounding error that
may be significant at low levels.
5. These guidelines define procedures for calculation of Method Detection Limits (MDL’s). Limits of
Reporting (reported detection limits) may be higher, for example if based on Level of Quantitation
(LOQ) or Practical Quantitation Limit (PQL).
A summed parameter is one that is obtained by summing two or more parameters that are analyzed
independently.
A - 14
The Method Detection Limit for a Summed Parameter is calculated as follows:
If Resultsummed = ( C1 + C2 + C3 …. )
2 2 2
MDL for Resultsummed = sqrt [ (MDL1) + (MDL2) + (MDL3) + … ]
Where:
Cx = Concentration of parameter X.
MDLx = Method Detection Limit of parameter X.
Special considerations are necessary to determine valid Detection Limits for parameters that are
determined by subtraction. When a result being subtracted approaches the magnitude of the result it is
subtracted from, the difference is subject to a high degree of uncertainty due to the uncertainty in the
original two results. For subtracted parameters, an estimate of the uncertainty in the calculated result is
the only reasonable measure of detection limit.
Except where methods specifically define alternate procedures, the following guidelines are
recommended to determine valid Detection Limits for Subtracted Parameters. There are three different
situations that require consideration:
Where Resultsub = C1 – C2 ;
Use the MDL of the larger result (C1) as the MDL for the subtracted parameter (Resultsub). (Resultsub)
would be reported as < DL.
Use the MDL of the larger result (C1) as the MDL for the subtracted parameter (Resultsub). In this
situation, the uncertainty of the difference of the two results should be acceptably low, and will not
dominate the question of whether or not the subtracted parameter has been detected or not.
In this case, the subtracted result (Resultsub) may be subject to an unacceptably high degree of
uncertainty due to the uncertainty in the component results (C1 and C2). Where practical, it is
recommended that the MDL for the subtracted parameter be determined by adding the uncertainties of
the two results (in quadrature) as follows:
Where Resultsub = C1 – C2
2 2
MDL for Resultsub = sqrt [ (UC1) + (UC2) ]
Where:
UC1 = The laboratory’s expanded (95% confidence level) Measurement Uncertainty (MU)
estimate for result C1 (i.e. MU for parameter 1 at concentration C).
UC2 = The laboratory’s expanded (95% confidence level) Measurement Uncertainty (MU)
estimate for result C2 (i.e. MU for parameter 2 at concentration C).
Under Case 3, if the MDL calculated above is insufficient to meet a data requirement, then an
alternative measurement technique utilizing a direct measurement of the parameter should be sought.
A - 15
Reference:
These guidelines were developed by the BCELQAAC Technical Subcommittee, and were endorsed by
the committee for inclusion in the BC Lab Manual on September 11, 2003.
Where facilities permit, internal audits should be conducted. Performance audits should incorporate blind
or other samples in order to check functioning of laboratory QC procedures. This action may be helpful if
external audits are imposed. Records of audit results and action taken should be kept.
3.1 Introduction
Method Detection Limits (MDLs) are set at the 95% confidence level above zero (or the blank - see
section 4.0). For an infinite number of replications this is:
where SD near zero is standard deviation estimate made within a factor of 10 of the (expected) MDL.
where t1, 0.05 = the one-tailed ‘t’ statistic at p = 0.05 (see Table 1).
The standard deviation, SD, of a low level sample may be used to produce this estimate.
The standard deviation used to estimate σø may be the within-batch standard deviations provided by
duplicate or replicate samples carried through the sample processing steps, or the standard deviations of
between-batch replicates.
A - 16
Calculation of standard deviations should be based on:
The ‘t’ statistic is used to compensate for the tendency of small sample sizes to underestimate variability.
pooled SD = sqrt [ { ∑ (v1 x s12) + (v2 x s22) +..... (vi x si2)} / (v1+v2+ ..... vi) ]
1. Replicate analysis of a bulk sample of the desired matrix known to be homogenous and
stable. Sub-sampling of the bulk sample is performed and each sub-sample is carried
through the entire preparative and analytical step. Concentration is not to exceed 10 times
the estimated MDL, but is preferred to be 1-3 times the estimated MDL. Estimate SD using
formula case 2 for single batch analysis, and formula case 3 for multiple batch analysis.
2. If a bulk sample is not available, prepare a composite sample in the concentration range
required. Continue as in 1.
3. If the above two are not possible, use duplicate analyses from several batches, adhering to
the described concentration range. Use formula in case 1 for estimation of SD.
4. If the above three are not available, use a spiked blank or clean matrix, to produce a large
sample. Take steps to assure the spike is homogenously distributed (e.g. mixing overnight).
Spike to a concentration not greater than 10 times the estimated MDL. Sub-sample this bulk
spiked sample and carry each sample through all preparative and analytical steps. Use
formula case 2 to estimate SD.
5. If facilities do not allow preparation of a bulk spike sample, spike individual blank or clean
matrix samples to a concentration as defined above. Take steps to assure the spike is
homogenously distributed (e.g. mixing overnight). Continue as described above.
A - 17
6. If no blank or clean matrix is available proceed as in 4 or 5 using pure de-ionized water as the
matrix.
7. For methods which employ continuous monitoring of a detector signal, 3 times the noise in
the area of the analytical peak can be used if none of the above can be used. This produces
an estimate of instrument SD and does not take into account the contribution of preparative
and matrix effects.
Section 3.2 demonstrates how method standards may be determined for three cases: from pairs of
duplicates, from replicates in same batch and for pooled standard deviations from replicates in
successive batches.
These appear as Tables 4, 5 and 6 respectively. For each example the data have been organized in a
format convenient for spreadsheet calculation.
MDLs must be determined for each matrix type processed by the laboratory, and must reflect all steps of
the applicable method.
MDLs should be recorded to one non-zero figure. The final MDL values and the data used to estimate
SD should be recorded to two significant figures, to reduce rounding errors.
The limit of quantitation (LOQ) is normally defined as 10 times SD (at MDL). The LOQ is the
concentration at which precise quantitative results are possible (precision approaches ± 30%). At the
MDL, on the other hand, the presence of a detected t analyte is defined with certainty but concentration
estimates are imprecise (precision approaches ±100%).
The MDL definition described here uses a factor of 2 x the one sided ‘t’ factor at p = 0.05, and is
sometimes referred to as a Reliable Detection Limit (RDL). For an infinite sample size, this would be
MDL = 3.29 x SD. This is the sum of two distributions: one centered on 0 (zero) has a 5% risk of
declaring a null quantity as a positive amount, while the other distribution, centered above 0, involves a
5% chance of declaring a measurement as 0, when in fact, it is real (see Table 3).
The MDL definition given at the bottom of Tables 4, 5 and 6 is the USEPA convention. This definition
uses the one-sided ‘t’ factor for p = 0.01. For an infinite sample size, this would set the MDL = 3 x SD.
This is derived from a single distribution, centered on 0 (zero) and having a 1% risk of declaring a null
quantity as a positive amount.
The two MDLs, while derived from different theory, are in fact very close. The USEPA derived MDLs
include lower and upper limits; the CAEAL value always falls within those limits.
A - 18
4.0 BLANKS
While there are few situations where blank correction is appropriate, monitoring of blanks should be
carried out for most samples and blanks should be controlled within defined limits for data to be
acceptable.
Batch blank values shall be recorded on the analytical result sheet with each batch. In addition, a
tabulation of day-to-day blanks shall be maintained and kept current. It will include, at a minimum,
analytical date, preparative and analytical operators’ initials, individual blank values, a reference to the
long term blank and the date range over which it was determined, and any comments explaining unusual
occurrences. If possible a control chart of batch blanks will be maintained.
The long term blank should be estimated, on at least 2 separate occasions, from replicate blank matrices
which have been processed through all the preparative and analytical steps applied to a regular sample.
It is suggested that 10 replicate blanks be processed on each occasion. The long term blank value
should be confirmed at least once each year.
The average and standard deviations of blank readings should be calculated for each occasion. Pooled
blank average and standard deviations should be calculated using the following formula (Taylor, 1987):
sqrt [ ∑ { (v1 x s21) + (v2 x s22) + ... + (vn x s2n) } / (v1 + v2 +... vn) ] (1)
Test for outliers using the Grubbs test (Taylor, 1987). If outliers are present, remove them and
recalculate the average and standard deviation as above.
4.3 Application
Blank correction should be applied to every sample for which a reliable estimate of long term blank levels
has been made (i.e., the blank is in statistical control).
Blank correction should be applied to all samples where the batch method blank(s) exceeds the method
detection limit (MDL), and is less than the long term blank control limits.
Samples with concentrations greater than 20x the batch blank, need not be blank corrected (US EPA,
1990).
If all batch blanks are greater than the MDL and less than the control blank limit, all batch
samples for that parameter should be corrected by subtracting the average batch blank.
A - 19
The control blank limits will be set as the long term average method blank plus 1.64 times the
standard deviation of the long term average blank.
If the long term average blank is <MDL, the control blank limits will be set as MDL plus 1.64 times
the standard deviation of the long term average blank.
This control limit should include approximately 95% of the blank values that will occur.
If the estimate of the long term blank is based on <100 determinations, the appropriate ‘t 0.05’
value from the table of ‘t’ values is substituted for 1.64.
If any batch blank exceeds the long term blank control limit, the batch must be reprocessed.
If this is not possible due, for example, to insufficient sample, or instability of the sample, the
parameter results can be blank corrected, but must be flagged in the report as having "High
blank, subtraction made, accuracy of results may be compromised".
If multiple parameters are analyzed for the same sample preparation batch, and if more than 5%
of the parameters exceed the blank control limits, the batch must be re-processed; e.g., for ICP
total metals scans (33 elements), if more than 2 parameters have blank levels greater than blank
control limits, the batch must be re-processed. Otherwise the parameters greater than blank
control limits can be reported with a flag stating "High blank for parameter XX, subtraction made,
accuracy of results may be compromised".
Batch blank > long term blank control limit reprocess batch
If the batch blank is part of the calibration curve, this effectively corrects for the blank provided the
linearity of the calibration curve is not destroyed.
If the batch blank deviates from the linear best fit line, calculated excluding the blank, by greater
than 50%, the batch blank is too high, and the batch must be re-processed.
If all batch blanks are greater than the MDL and less than 10X the MDL, all batch samples for that
parameter may be corrected by subtracting the average batch blank.
If any batch blank exceeds 10 times the MDL, the batch must be reprocessed.
If this is not possible, due, for example, to insufficient sample, or instability of the sample, the
parameter results can be blank corrected, but must be flagged in the report as having "High
blank, subtraction made, accuracy of results may be compromised".
A - 20
4.4.4 Readable units
Where the batch blank fails the blank control limit test by ≤ one readable unit, the batch blank
shall be considered acceptable.
Batch blank values are to be recorded on the analytical result sheet with each batch. In addition, a
tabulation of day to day blanks shall be maintained and kept current. It will include, at a minimum,
analysis date, preparative and analytical operators initials, individual blank values, a reference to the long
term blank and the date range over which the long term blank was determined as well as any comments
explaining unusual occurrences. If possible a control chart for batch blanks is to be established and
maintained.
Analytical reports to clients must indicate whether or not data has been blank corrected. In addition, the
reports must indicate batch blank levels for the data, regardless of whether blank correction has been
applied.
Blank correction is rarely appropriate when these rules are followed, since at high analyte concentrations
the correction is insignificant and at low levels (near MDL), the blank, whether subtracted or not, merely
increases the uncertainty of the value.
While all laboratories are encouraged to implement a comprehensive QA program, it is recognized that
small laboratories with limited resources may find it economically necessary to implement a less
comprehensive program.
This section is relevant only to small permittee laboratories that participate in the CALA laboratory
proficiency-testing program and who have achieved a PE score greater than 70 for each applicable
variable in the last four studies and who generally analyze only small batches of samples.
A small batch is defined as less than 11 samples analyzed at one time as a group using a common blank,
standards and duplicate data. In cases where a laboratory receives only a few samples per day, and
sample stability or treatment permits, it is preferable that samples be stored and run as one small batch
on a weekly basis. If stability does not permit, the samples should be analyzed on a daily basis.
A - 21
5.2 Small Laboratory Quality Control Criteria
a. Calibration control limits of ±2s as a warning limit and ±3s as a control limit are applicable for
quality control standards for most variables.
b. Method blanks
c. Duplicates for water and wastewater must be within the precision limits given in Table 1020:I of
Standard Methods, 18th Edition (1992).
d. Internal recovery control must be within the precision limits given in Table 1020:I of Standard
Methods 18th Edition (1992).
6.0 REFERENCES
Standard Methods for the Examination of Water and Wastewater, AWWA, APHA, 18th Edition, (1992),
Section 1000. (Or more recent edition).
Clark, Malcom J.R., Quality Assurance in Environmental Analysis. Encyclopedia of Analytical Chemistry
(Lemisky, R.A. Meyers (ed.)), John Wiley & Sons Ltd., Chichester, UK, 2000. pp. 3197 – 3227.
Dux, James P., Handbook of Quality Assurance for the Analytical Chemistry Laboratory, Van Nostrad
Reinhold, 1986.
Taylor, J. K., Quality Assurance of Chemical Measurements. p.24, 37, 123ff, 1987.
Pocklington, W.D., Harmonized Protocols for the Adoption of Standardized Analytical Methods and for the
Presentation of their performance characteristics. Pure and Applied Chemistry, 62/1, pp. 149-162,
1996.3
Taylor, J. K., Quality Assurance of chemical measurements. p.24, 37, 123ff, 1987.
U.S. Environmental Protection Agency, Test Methods for Evaluating Solid Waste, Physical/Chemical
Methods, SW 846 - 3rd Edition, Rev. 1 Nov. 1990.
A - 22
7.0 REVISION HISTORY
A - 23
Table 3. Values of the one-tailed ‘t’ statistic at p=0.05, applied to the standard deviation, SD,
appear in the following table
7 1.90
8 1.86
9 1.83
10 1.81
15 1.75
20 1.73
25 1.71
30 1.70
40 1.68
60 1.67
infinity 1.64
For case 2, replicates in same batch, degrees of freedom = n - 1, where n = the number of replicates,
For case 3, pooled SD from replicates in successive batches, degrees of freedom = sum of v1 + v2 + ... vi.
Procedure Results for duplicates analyses of soil samples on different days were recorded.
All data are less than 10x the MDL.
2
Sample # result 1 result 2 difference difference
4486 4100 4600 500 250000
262 5200 5300 100 10000
25074 2600 2200 400 160000
25080 3500 3700 200 40000
25090 1600 1500 100 10000
25099 2000 2300 300 90000
26011 2100 2100 0 0
A - 24
Table 5. Example 2: MDL estimation from replicates in same batch
5 g of sea sand was spiked to 10 µg/g with BTEX standard, mixed, sealed in a purge & trap vial, and retained for
2 days in freezer. The samples were then individually processed. m,p-Xylene was spiked at 20 ug/g.
MDL (USEPA) 4 4 5 3
=2.896 x SD
A - 25
Table 6. Example 3: MDL estimation from pooled standard deviations from replicates in
successive batches
USEPA MDL, µg/L 0.07 lower control limit 0.05; upper control limit 0.12
A - 26
GLOSSARY OF STATISTICAL TERMS USED IN QUALITY ASSURANCE
Accuracy - indicates the closeness of a measured value to a true value; combines bias and precision.
Bias - consistent tendency of measured value to deviate, either positively or negatively, from a true value.
Calibration check standard - standard, usually mid-analytical range, used to monitor the validity of
instrument calibration between periodic full recalibration.
Confidence coefficient - the probability, expressed as a percentage, which a measurement result will
reside in the confidence interval (between the confidence limits).
Confidence interval - the set of possible values within which the true value will reside with a stated
probability level.
Confidence limit - upper or lower boundary values delineating the confidence interval.
Detection limits - various limits are defined; in increasing order these are:
Instrument detection limit (IDL) - the analyte concentration that produces a response five times greater
than the signal/noise ratio of the instrument. This is similar to the “criterion of detection” which is
defined as 1.645 times the standard deviation, s, of blank analyses.
Lower limit of detection (LLD) - also called “detection limit” (DL) and “limit of detection” (LOD) - the
analyte concentration in reagent water that produces an instrument response 2(1.645)s above
the mean of blank analyses. This criterion sets the (maximum?) probability of both Type I and
Type II errors at 5%.
Method detection limit (MDL) - the analyte concentration that, when processed through the complete
method, produces an instrument response with a 99% probability that it is non-blank.
Limit of Quantitation (LOQ) - the analyte concentration that produces an instrument response sufficiently
greater than the response by the blank that it can be measured within specified accuracy limits by
competent laboratories in routine operation. Typically it is regarded as the analyte concentration
that produces a response ten times greater than the standard deviation, s, of the reagent water
blank signal.
Duplicate - least case of replicates (two); in general, while any portion of the analytical protocol can be
duplicated, the term duplicate is usually applied to duplicate samples, i.e., two samples taken at
the same time from the same location.
Internal standard - a pure compound different from, but similar enough to, the analyte, that, when added
at a known concentration to the sample extract immediately prior to instrumental analysis, allows
corrections due to instrument inefficiencies or vagaries.
Laboratory control standard - a standard, optimally certified by an outside authority, used to measure
method bias.
Precision - gauge of the degree of agreement among replicate analyses of a sample, usually expressed
as the standard deviation.
Quality assessment - procedure for determining the quality of laboratory data using internal and external
quality control measures.
Quality assurance - a system of laboratory operation that specifies the measures used to produce data of
documented accuracy.
Quality control - a set of procedures applied to an analytical methodology to demonstrate that the analysis
is in control.
Random error - the deviation experienced in any step of an analytical procedure that can be estimated by
standard statistical techniques.
Replicate - repeated operation of part of an analytical procedure. Two or more (instrumental) analyses
for the same analyte in a processed sample are termed replicate extract analyses.
Significant figures – A convention of reporting numeric results where all digits in result are known without
doubt, except for the right-most figure.
Surrogate (or surrogate standard) - a pure compound different from, but similar enough to, the analyte
that, when added at a known concentration to the sample prior to processing, provides a measure
of the overall efficiency of the method (recovery).
Type I error - also called alpha error, is the probability of deciding an analyte is present when it actually is
absent.
Type II error - also called beta error, is the probability of deciding an analyte is absent when it actually is
present.
A - 27